Technician's instructions. QUATTROcare PLUS 2124 A from serial number Always be on the safe side.

Similar documents
Instructions for use. QUATTROcare PLUS 2124 A Always be on the safe side.

Instructions for use Spare cartridge for COMPACTtorque 636 P, 636 CP and POWERtorque 646 B, 646 C , Always be on the safe side.

Technician's instructions. GENTLEpower LUX Contra-angle 25 LP - REF

Operating Instruction K-Control TLC 4955, 4956, Always be on the safe side.

Instructions for use. PHYSIO Evo/Evo F/One

Instructions for use K-ERGOgrip Always on the safe side.

Operating Instruction INTRAmatic Prophy Handpiece 19 ES

Operating instructions K- Control 4960, 4965, Always on the safe side.

Operating Instructions K Always on the safe side.

Podiatry Procedures Chair. For Models: Barrier-Free. User s Guide

Instructions for use DURAtec Prophy REF

C3 Operating Instructions

HEINE NT4 MED V

Installation manual portable distributors

Installation manual wall-mounted distributor

Instructions for use GENTLEmini LUX 4500 BR - REF

basic quattro No /

PURE SINE WAVE DC TO AC POWER INVERTER

Operating instructions

EPS 16 ATEX 1072 X, IECEx EPS X. Operating Instructions. Solenoid coil Type AC19 Magnetspule Typ AC19 Bobine magnétique Type AC19

Catalyser Introduction. 3. Hazard Information. 2. Area of Application. 3.1 Symbology

KeContact P20. User manual

Contents. Introduction Inspection Notes on Safety Precautions. Chapter 1 Overview Product Overview Names of Parts 2

Register your instrument! HeatSealer S100. Operating manual

HANGKAI GROUP HOYMILES MICRO-INVERTER MI-250

User s Manual Power Supply IM E. 5th Edition

AQUATEC R / AQUATEC F / AQUATEC XL. Bathlift Operating instructions

DIGITAL BATTERY TORQUE WRENCH (BC-RAD SELECT) USER GUIDE

Operators Manual. Recirculating Chiller /06/08

Instructions for use GENTLEsilence 6500B - REF

Switching DC Power Supply

c-go 24V/6A 24V/8A 24V/12A

5 IN 1 JUMP START OPERATION & MAINTENANCE INSTRUCTIONS MODEL NO: JS5IN1 PART NO: LS0810

Wilo-Control SC-Fire Jockey

WELDING INVERTER. PEGAS 160 E Smart PEGAS 200 E Smart OPERATING MANUAL. ALFA IN a.s. PEGAS E Smart Manual EN 04

System 3 ZB1PS Device Chassis. Operator s Manual

Lumitester PD-20 Control Kit

NANOPAC-300 & 500 Power Supply. Instruction manual NANOPAC-300 & NANOPAC-500

Technical Documentation

Operating Instructions CYR52

User Guide. Lubricus Lubrication System LUB-D1/LUB-D2/LUB-D3/LUB-D4 (24 VDC)

362 Clamp Meter. Users Manual

Battery Charger JCB- SCH20LI.2

A company of ThyssenKrupp Elevator. ThyssenKrupp Aufzugswerke. Operating Manual. Oil buffer

Operating manual. original operating manual. HDA eco Box 12/24V DC Automatic Dispenser. Translation of the. Item No.:

INSTALLATION AND OPERATING INSTRUCTIONS OF THE INTERNATIONAL ISOBOX SERIES ISOLATION TRANSFORMERS.

INSTALLATION AND OPERATING INSTRUCTIONS OF THE INTERNATIONAL ISOBOX SERIES ISOLATION TRANSFORMERS.

Assembly and Maintenance Manual Type ASNU

FREQUENCY INVERTER INSTALLATION & USER GUIDE

B-RAD Select USER MANUAL TABLE OF CONTENTS

Mounting and Operating Instructions EB 8039 EN. Type 3351 Pneumatic On/off Valve. Type 3351 Pneumatic On/off Valve. Type 3351 Pneumatic On/off Valve

Go to to register your product and find more information.

Accessories for Stand-alone inverter SUNNY ISLAND GENMAN

OPERATION & MAINTENANCE INSTRUCTIONS

Assembly instructions

WATERFLUX 3070 Quick Start

Triple Output Power Supply

CoolFreeze CFX35, CFX40, CFX50, CFX65, CFX65DZ

Operator's Manual. Storage System. Ultrasound Probe Cabinet. Manufactured by:

PTB 14 ATEX 2023 X, IECEx PTB X. Operating Instructions. Solenoid coil Type AC10 Magnetspule Typ AC10 Bobine magnétique Type AC10

BT508/BT510/BT520/BT521 Battery Analyzer Safety Sheet

MAKING MODERN LIVING POSSIBLE. UniLynx Indoor Installation Manual. ULX 1800i ULX 3000i ULX 3600i ULX 5400i SOLAR INVERTERS

Exchange of rollers from the XTS-Mover

Model 1672, 1673 Triple Output Power Supply

Instructions for use. Quick couplings RQ-03, RQ-04, RQ-14, RQ-24, RQ-34

OPERATIONS MANUAL CR Thermoreactor

Adam Equipment DUNE DCT SERIES. (P.N. 9384, Revision B2, June 2013)

OPERATION & MAINTENANCE INSTRUCTIONS

OPERATION & MAINTENANCE INSTRUCTIONS

Stereo Power Amplifier

CO 3-WAY PNEUMATIC VALVE INSTRUCTION MANUAL 2080

Assembly and Maintenance Manual Type AS

Valor 1000W Series Instruction Manual. Serie Valor 1000W Manual de Instrucciones. Série Valor 1000W Guide de I utilisateur

Battery Charger JCB-FCH20LI2

Operating Instructions Flexdip CYA112

Compact System NRGS 11-2 NRGS Original Installation Instructions English

JUMP STARTER OPERATION & MAINTENANCE INSTRUCTIONS MODEL NO: JSM350 PART NO: ORIGINAL INSTRUCTIONS LS1117-ISS 1

Accessories for Wind Power Inverter WINDY BOY PROTECTION BOX 400 / 500 / 600

Pro Booster 802Li. Please read and fully understand the instructions in this manual before operation. Keep this manual safe for future reference.


BT403. A Geno Technology, Inc. (USA) brand name. BT-300 Power Supply. Cat. No. BT

Type Operating Instructions. Bedienungsanleitung Manuel d utilisation. 2/2-Way Solenoid Valve 2/2-Wege-Magnetventil Électrovanne à 2/2 voies

The Brose e-bike system. Instructions for use

LEAD ACID BATTERY CHARGER

110 Volt/12 Volt Portable Inflator

Multi-Mover Charger for model L25

Battery Charger JCB-FCH12Li

Operating Instructions

Leica GKL221 Charger User Manual

Drive Unit e-drive1. Installation instructions 04/2014. English translation of the original German installation instructions

Katalysator Nr /

original operating manual Operating manual Translation of the Item-No.: ,

MICRO-INVERTER OMNIKSOL M-300

Instructions for use INTRAmatic Contra-angle 23 ES

Operating manual. High-voltage testing adapter VAS 6558A/30. Operating manual. Customer documentation Revision: 00 Version: 07/2018

These operating instructions apply for: NCX 380 NCZ 300 NCX 480 NCZ 370 NCX 580 L NCZ 480 NCX 660 K NCZ 560 NCZ 660 NCZ 800

Operating Instructions

The pneumatic applicator complies with the following safety guidelines: CE

PrioVino Premier. Translation of Original Operating Instructions. Status: August First edition January 2018 / PrioVino GmbH

Triple Output High Resolution Power Supply. User Manual

Transcription:

Technician's instructions QUATTROcare PLUS 2124 A - 1.008.3805 from serial number 4000 Always be on the safe side.

Distributed by: KaVo Dental GmbH Bismarckring 39 D-88400 Biberach Phone +49 (0) 7351 56-0 Fax +49 (0) 7351 56-1488 Manufacturer: Kaltenbach & Voigt GmbH Bismarckring 39 D-88400 Biberach www.kavo.com

Table of contents Table of contents 1 User instructions... 5 1.1 User guide... 5 1.1.1 Abbreviations... 5 1.1.2 Symbols... 5 1.1.3 Target group... 5 1.2 Service... 6 1.3 Transportation and storage... 6 1.3.1 Packaging regulations (applicable in Federal Republic of Germany only)... 6 1.3.2 Damage in transit... 6 1.3.3 Storage... 7 2 Safety... 9 2.1 Description of safety instructions... 9 2.1.1 Warning symbol... 9 2.1.2 Structure... 9 2.1.3 Description of hazard levels... 9 2.2 Purpose Proper use... 9 2.2.1 General information... 9 2.2.2 Product-specific... 11 2.3 Safety instructions... 11 2.3.1 General information... 11 2.3.2 Product-specific... 12 3 Product description... 13 3.1 QUATTROcare PLUS 2124 A... 13 3.2 Rating plate... 14 3.3 Technical data... 14 3.4 Scope of delivery... 15 4 First use... 17 4.1 Select location... 17 4.2 Connect the QUATTROcare PLUS 2124 A... 18 4.3 Replacing the QUATTROcare Plus Spray can... 19 4.3.1 Insert the spray canister... 20 4.3.2 Remove the spray canister... 21 4.3.3 Dispose of the spray canister... 22 5 Operation... 23 5.1 Control panel... 23 5.2 Open the front flap... 23 5.3 Mounting instruments... 24 5.4 Start the treatment... 25 5.5 Remove the instruments... 26 5.6 Reposition the treatment couplings... 26 5.7 Chuck Servicing... 27 5.8 Set the blow-out time... 29 6 Check and adjust... 30 6.1 Check and adjust the compressed air input... 30 3 / 60

Table of contents 6.2 Open the service menu... 31 6.3 Run the test for the care product recognition switch... 31 6.4 Adjust the recognition switch... 32 6.5 Check the care product outlet in the individual adapters... 33 6.6 Check the blown air at the adapters... 33 6.7 Check the solenoid valve... 34 6.8 Check for absence of leakage... 34 6.8.1 Box coupling... 34 6.8.2 Oil system... 35 6.8.3 Air system... 35 6.9 Read the number of run cycles... 36 6.10 Check sensor... 37 6.11 Check barcode reader... 38 6.12 Check the chuck maintenance... 38 6.13 Set the blow-out time... 40 6.14 Setting the oil servicing time... 41 7 Bridging the barcode reading function... 42 8 Safety checks pursuant to DIN VDE 701-702 / DIN EN 62638... 43 8.1 Introduction... 43 8.2 General information... 43 8.3 Visual inspection... 43 8.4 General design test... 44 8.5 Pressure/Leakage test... 45 8.5.1 Test equipment... 45 8.5.2 Preparation... 45 8.5.3 Air system... 46 8.5.4 Setting the pressure controller... 47 8.5.5 Oils system - Bottle coupling... 47 8.5.6 Oil system - Solenoid valve to valve block... 48 8.6 Electrical test... 48 8.6.1 Electrical test devices... 48 8.6.2 Protective earth conductor test... 48 8.6.3 Measurement of the insulation resistance... 49 8.6.4 Measurement of the protective earth conductor current... 50 8.6.5 Proof of effectiveness of other protective facilities... 51 8.7 Functional test... 51 8.7.1 Check the output of servicing product in the individual adaptors... 52 8.7.2 Check the blown air in the individual adaptors... 52 8.7.3 Check the solenoid valve... 53 8.8 Analysis and labelling... 53 8.9 Documentation... 55 9 Troubleshooting... 56 10 Tools and consumables... 57 4 / 60

1 User instructions 1 User instructions 1.1 User guide 1.1 User guide CAUTION Incorrect operation. Damage to or malfunction of the unit. Read these Instructions for Use before first start-up. 1.1.1 Abbreviations Abbreviation IfU CI AI TI SC IEC RI RK AS EP EMC PI Explanation Instructions for use Care instructions Assembly instructions Technician's instructions Safety checks International Electrotechnical Commission Repair instructions Retrofitting kit Assembly set Enclosed parts Electromagnetic compatibility Processing instructions 1.1.2 Symbols See the Safety/Warning Symbols section Important information for users and technicians CE mark according to Medical Devices Directive EC 93/42 Action required Wear safety goggles 1.1.3 Target group This document is for service technicians who have been trained by KaVo for the product. 5 / 60

1 User instructions 1.2 Service 1.2 Service KaVo Technical Service If you have any technical questions or complaints, please contact the KaVo Technical Service: +49 (0) 7351 56-1500 service.instrumente@kavo.com KaVo Repair Service For repairs, please contact your local dealer or the KaVo Repair Service directly: +49 (0) 7351 56-1900 service.reparatur@kavo.com KaVo Dental GmbH Customer Service Center Bahnhofstraße 20 D-88445 Warthausen +49 (0) 7351 56-1500 www.kavo.com 1.3 Transportation and storage 1.3.1 Packaging regulations (applicable in Federal Republic of Germany only) Dispose of and recycle the sales packaging appropriately in accordance with current packaging regulations, employing waste management or recycling companies. Comply with the comprehensive return system. KaVo has had its sales packaging licensed for this purpose. Please comply with the regional public waste-disposal system. 1.3.2 Damage in transit KaVo shall not be held liable for damage arising from transportation. The shipment must be checked on arrival. International If the packaging is visibly damaged on delivery, please proceed as follows: Leave the product and packaging in the condition in which you received it. Do not use the product. Record the loss or damage on the delivery receipt. The recipient and the representative of the shipping company must sign this delivery receipt. Without this evidence, the recipient will not be able to assert a claim for damages against the shipping company. If the product is damaged but there was no discernable damage to the packaging upon delivery, proceed as follows: Leave the product and packaging in the condition in which you received it. 6 / 60

1 User instructions 1.3 Transportation and storage Do not use a damaged product. Report the damage to the shipping company without delay, definitely no later than 7 days after delivery. If the recipient fails to comply with any of the above-mentioned obligations, the damage will be considered to have arisen after delivery (in accordance with CMR law, Chapter 5, Art. 30). In Germany If the packaging is visibly damaged on delivery, please proceed as follows: Leave the product and packaging in the condition in which you received it. Do not use the product. Record the loss or damage on the delivery receipt. The recipient and the representative of the shipping company must sign this delivery receipt. Report the damage to the shipping company. Report the damage to KaVo. Consult with KaVo first, before returning a damaged product. Send the signed delivery receipt to KaVo. If the product is damaged but there was no discernable damage to the packaging upon delivery, proceed as follows: Leave the product and packaging in the condition in which you received it. Do not use a damaged product. Report the damage to the shipping company without delay, definitely no later than 7 days after delivery. Report the damage to KaVo. If the recipient fails to comply with any of the above-mentioned obligations, the damage will be considered to have arisen after delivery (in accordance with the General German Freight Forwarders Terms and Conditions, Art. 28). 1.3.3 Storage Please keep the packaging in case you need to return the product for servicing or repair. The symbols printed on the outside are for transportation and storage, and have the following meaning: Temperature range -20 to 50 (-4 to 122 o F) 7 / 60

1 User instructions 1.3 Transportation and storage Humidity 5% to 95% non-condensing Air pressure 700 to 1,060 hpa Transport upright Protect from impact Protect from moisture 15 kg max Permissible stacking load: 15 kg (33 pounds) Disposal 8 / 60

2 Safety 2 Safety 2.1 Description of safety instructions 2.1 Description of safety instructions 2.1.1 Warning symbol Warning symbol 2.1.2 Structure DANGER The introduction describes the type and source of the hazard. This section describes potential consequences of non-compliance. The optional step includes necessary measures for hazard prevention. 2.1.3 Description of hazard levels Safety instructions distinguishing between three hazard levels are used in this document to prevent personal and property damage. CAUTION CAUTION indicates a hazardous situation that can cause damage to property or mild to moderate injuries. WARNING WARNING indicates a hazardous situation that can lead to serious or fatal injury. DANGER DANGER indicates a maximal hazard due to a situation that can directly cause death or fatal injury. 2.2 Purpose Proper use 2.2.1 General information CAUTION Improper maintenance or repair of the unit. Damage to and malfunction of the unit. Repair and servicing work on the electronic part of the unit may be done only by skilled staff or KaVo-trained technicians. Use original KaVo spare parts only. 9 / 60

2 Safety 2.2 Purpose Proper use CAUTION Inappropriate installation location of the unit. Damage to and malfunction of the unit. The unit is designed for indoor use only. Do not use the unit in areas where there is an explosion hazard. The user must ensure that the unit works properly and is in satisfactory condition before each use. According to these provisions, the user is responsible for: exclusive use of equipment that is operating correctly. compliance with the specified intended use. protecting himself and others against hazards. Compliance with the warnings on the device and in the Instructions for Use is mandatory. The following persons are authorised to repair and service the KaVo product: Technicians of KaVo branch offices after appropriate product training. Specifically KaVo-trained technicians of KaVo franchised dealers. The applicable national legal regulations must be observed during the use of the device, in particular the following: Applicable regulations governing the connection and start-up of medical devices. Current occupational safety regulations. Current accident prevention regulations. KaVo shall not be responsible for damage caused by: external influences, poor media quality or faulty installation. The use of incorrect information. repair work carried out incorrectly. Disposal of electronic and electrical devices According to EC directive 2002/96 concerning used electrical and electronic devices, this product is subject to the cited directive and must be disposed of accordingly within Europe. For more information, please visit www.kavo.com or contact your specialised dental supplier. For final disposal, please contact: International For country-specific information on disposal, please contact your dental supplier. In Germany To return an electrical device, you need to proceed as follows: On the homepage of enretec GmbH www.enretec.de, you can download a form for a disposal order under the menu item eom, or you can complete it as an online order. 10 / 60

2 Safety 2.3 Safety instructions Please complete the disposal order entering the necessary data and submit it as an online order or by Fax +49 (0) 3304 3919-590. The following contact options are also available for questions and for initiating a disposal order: Phone +49 (0) 3304 3919-500 enretec GmbH, Geschäftsbereich eomrecycling, Kanalstraße 17, 16727 Velten, Germany pickup@eomrecycling.com A unit that is not permanently installed in the office will be picked up. A permanently installed unit will be picked up at the curb at your address on the agreed date. The owner or user of the device will have to bear the costs of disassembly, transportation and packaging. 2.2.2 Product-specific The intended use of the KaVo QUATTROcare is for internal cleaning, i.e, purging of old lubricant, for the maintenance of rotating dental and surgical instruments. KaVo QUATTROcare spray can is a component of the KaVo QUATTROcare unit and is specifically designed to fit this unit only. The lubricant spray can is unable to be used manually. Compliance with the preceding and subsequent re-processing steps according to the recommendations of the Medical Devices Law and RKI guidelines is required although the QUATTROcare PLUS 2124 A services the straight and contra-angle handpieces and turbines to be re-processed and cleans the inner parts from mechanical abrasion. 2.3 Safety instructions 2.3.1 General information WARNING Electrical power. Electrical shock. Damage to or malfunction of the unit. Connect the device to a properly grounded and fuse-protected power source only. Check if the information on the rating plate of the unit (voltage, frequency) is consistent with the available electrical power supply. The rating plate is attached on the back panel of the device. Supply the device only with the voltage specified in the instructions for use, on the rating plate, and on the device. Only use extension cords that are permissible for the indicated rated voltage and are fitted with a protective conductor. Disconnect the unit from the power supply when you leave your practice. 11 / 60

2 Safety 2.3 Safety instructions 2.3.2 Product-specific CAUTION Compressed air. Malfunction of the device due to unsupervised supply of compressed air. When you leave your practice, disconnect the instrument from the compressed air connection. CAUTION Contaminated compressed air. Damage to and malfunction of the instrument. In general, ensure dry, clean oil-free compressed air in accordance with EN ISO 7494-2. CAUTION Overheating. If the instrument with the inserted spray canister overheats above 50 (122 o F), there is a danger of explosion. Protect the instrument from overheating above 50 (122 o F). 12 / 60

3 Product description 3 Product description 3.1 QUATTROcare PLUS 2124 A 3.1 QUATTROcare PLUS 2124 A 1 3x INTRAmatic service couplings 2 1x MULTIflex service couplings 3 Control panel and display 4 Door of spray can chamber 5 Chuck service adaptor 6 QUATTROcare Plus Spray canister 7 Collecting pan 8 Front door Back of QUATTROcare PLUS 2124 A 1 Rating plate 2 Compressed air connection 3 Input of mains voltage 4 Fuse 5 Main switch 13 / 60

3 Product description 3.2 Rating plate 3.2 Rating plate Type SN REF QUATTROcare PLUS 2124 A Year of manufacture - serial number Material number Fuse-protected with T1000 ma Comply with the Instructions for use CE mark CSA mark Labelling according to 2002/96/EC 3.3 Technical data Device dimensions Width Depth Height Weight 350 mm (1.15 feet) 230 mm (0.75 feet) 380 mm (1.25 feet) 10 kg (22 pounds) 14 / 60

Electrical supply 3 Product description 3.4 Scope of delivery Line voltage Mains frequency Power consumption Overvoltage category 100 to 240 VAC 50 or 60 Hz 41 VA II Mains voltage fluctuation ±10 % Compressed Air Compressed air Compressed air consumption Air requirements 0.4 to 0.6 MPa Approximately 50 NL/min Dry, oil free, dirt free, uncontaminated in accordance with EN ISO 7494-2 Operating environment Temperature 15 to 40 o C (59 to 104 o F) Humidity 25 to 90 % Degree of soiling 2 Permissible to a maximum 2000 m above sea level (6560 feet altitude) Transportation and storage conditions Storage temperature 0 to 25 o C (32 to 77 o F) Transport temperature -18 to 40 o C (-0 to 104 o F) Rel. humidity Air pressure 5 to 95 % non-condensing 700 to 1060 hpa 3.4 Scope of delivery The package includes the articles listed below. If any of these articles is missing, please contact the supplier immediately so that the missing article can be delivered. 15 / 60

3 Product description 3.4 Scope of delivery 1 Chuck service adaptor 2 QUATTROcare Plus Spray canister 3 Compressed air coupling 4 Instructions for use 5 O-ring MULTIflex (4 units) 6 O-ring INTRAmatic (3 units) 7 Compressed air hose 8 Power cable 9 Service coupling MULTIflex 10 Service coupling INTRAmatic (3 units) 11 QUATTROcare PLUS 2124 A 16 / 60

4 First use 4 First use 4.1 Select location 4.1 Select location The QUATTROcare Plus Spray canister contains explosive propellant gases. The following instructions and setup conditions therefore need to be observed when setting up the unit. CAUTION Propellant gases of the QUATTROcare Plus spray canister. Explosive, flammable vapour/air mixtures may be produced during use. Risk of fire from overheating. Do not expose the unit to direct sunlight. Do not place the unit right next to a source of heat. Keep away from ignition sources, and do not use open fire close by. Ensure sufficient ventilation of the premises. CAUTION Danger of overheating. Damage to and malfunction of the device. Do not cover or obstruct the vents and openings on the unit. Maintain a minimum distance of 20 cm (8 inches) between the unit and the surfaces of walls or cabinets. Tests have demonstrated that the unit meets the requirements of EC Directive 2004/108/EC concerning electromagnetic compatibility. Even though the unit does not emit interfering radiation, other equipment (e.g., ultrasonic cleaners) can interfere with the unit. Keep the unit away from possible sources of interference. 17 / 60

4 First use 4.2 Connect the QUATTROcare PLUS 2124 A The unit must be set up on a flat, level surface that is insensitive to oil and water. The unit must not be set up or operated on an angled surface. Installation requirement: It is mandatory to maintain a minimal distance of 20 cm (8 inches) 3 between the unit and wall or cabinet surfaces. Well-ventilated rooms 1 are a suitable installation location. Do not expose to direct sunlight. Do not set up the unit in cabinets. Maintain a minimum distance of 50 cm (20 inches) 2 from electrical plugs, light switches, and sources of ignition. Only operate the device in a clean environment such as that of a laboratory or dental practice. 4.2 Connect the QUATTROcare PLUS 2124 A CAUTION Contaminated compressed air. Damage to and malfunction of the device. The compressed air must be free of oil, dirt, and contamination. If needed: Use a compressor with a dry air system, if required, possible install an upstream air filter (on the compressor) or blow-out the lines before connecting them. 18 / 60

4 First use 4.3 Replacing the QUATTROcare Plus Spray can CAUTION Use of power cables other than original KaVo parts. Damage to or malfunction of the unit. Use original KaVo power cables only (for material numbers, see chapter on "Consumables and tools"). Back of QUATTROcare PLUS 2124 A A T1000 ma fuse is installed 4. Plug the power cable into the electric socket 3. Plug the compressed air hose onto the compressed air connector 1. Compressed air hose must not be kinked. Turn the main switch 5 on. 4.3 Replacing the QUATTROcare Plus Spray can CAUTION Use of servicing products other than original KaVo products. The use of third-party products can shorten the product life, cause explosion, damage or malfunction. Always service KaVo products with original KaVo service products. KaVo advises explicitly that original KaVo service products must be used exclusively. Only QUATTROcare Plus spray canisters bearing a bar code may be used. The QUATTROcare PLUS unit checks if the original spray can is being used. If a spray canister without a barcode or with a false bar code is inserted, the servicing device will indicate this as an error by means of a flashing LED. See also: 2 Troubleshooting, Page 0 QUATTROcare plus Spray package with 6 canisters can be ordered using the following Mat. no. 1.005.4525. 19 / 60

4 First use 4.3 Replacing the QUATTROcare Plus Spray can 4.3.1 Insert the spray canister Wear safety goggles when you insert or replace the spray can. Please note the safety data sheet of the spray can. The safety data sheet of the QUATTROcare Plus Spray can be viewed at www.kavo.com. Open the side door of the unit. Swing the adaptor on the retention clip outward. Place the spray can on the retention clip and position it towards the thread. Screw the spray can into the adaptor in clockwise direction and tighten until you notice a resistance. Do not jam it. ð The sealing ring seals with respect to the canister connector. 20 / 60

4 First use 4.3 Replacing the QUATTROcare Plus Spray can Check the screw connection. After insertion of the spray can check visually, tactile, acoustically and by touch for absence of leakage. The spray can must rest flat against the adaptor. There must be no gap between the spray can and the adaptor. The connection must be tightly sealed. If any oil or gas leaks, you need to use a different spray can. Swing the canister and adaptor back into the instrument. Close the side door of the instrument. 4.3.2 Remove the spray canister Open the side door of the unit. 21 / 60

4 First use 4.3 Replacing the QUATTROcare Plus Spray can Swing the spray can and the adaptor out of the device. Place the spray can on the retention clip and unscrew it counterclockwise from the adaptor. Close the side door of the unit. 4.3.3 Dispose of the spray canister Dispose of the spray can in empty condition. Waste code according to the European waste catalogue: 15 01 04 - metals, packaging, empty. See also: 2 Safety data sheet QUATTROcare Plus Spray. 22 / 60

5 Operation 5 Operation 5.1 Control panel 5.1 Control panel 1 LED spray can control 2 LED control light main switch (on/off) 3 LED display 4 Start button for chuck service 5 Start button for service procedure 5.2 Open the front flap CAUTION Objects on or in the front door. Placing objects on or in the front door can damage the unit. Do not place any objects on or in the front door. Grasp the recesses on the left and right of the front door. Pull forward and fold down to open it. 23 / 60

5 Operation 5.3 Mounting instruments 5.3 Mounting instruments Remove the handpieces from the unit as soon as the servicing is completed. Heads of the "22" series cannot be serviced because their necks are too thin. These heads must be serviced together with the corresponding shanks. Not every adaptor has to be used. The LED display shows the following: Adaptor on the far left corresponds to LED display 1. Adaptor on the far right corresponds to LED display 4. Mount turbines on the MULTIflex service coupling (service coupling on the far right 4) and lock them into place. Mount straight and contra-angle handpieces on the INTRAmatic service coupling (service coupling on the far left 1, second from the left 2 or third from the left 3) and lock them into place. Working with LUX handpieces, make sure the latch is in the correct position. Mount heads on the service coupling for INTRA heads (for Special accessories, see chapter on "Consumables and tools"). Push back the locking sleeve 1 and snap it into place. 24 / 60

5 Operation 5.4 Start the treatment When installing the service coupling for INTRA heads, make sure to push the actuating sleeve2 (for Special accessories, see chapter on "Consumables and tools") onto the associated flap to ensure proper servicing. Push the front door upwards and close it. 5.4 Start the treatment The LED of the spray canister control1 lighting up indicates that the spray can is empty. When the main switch is on, the LED 2 lights up. Press the start key 4. The automatic servicing process starts. The automatic servicing time per handpiece is 15-40 seconds depending on the set blow-out time. After the servicing process is completed, the LEDs 3 of the couplings in use light up and an acoustic signal is emitted. The LEDs 3 continue to be lit until the front door is opened. The service procedure is now completed. Open the front door. See also: 2 5.2 Open the front flap, Page 23 The LED display 3 is turned off. 25 / 60

5 Operation 5.5 Remove the instruments 5.5 Remove the instruments The exterior of handpieces needs to be wiped off thoroughly with a soft absorbent cloth before removing them. Pull the turbine in the axial direction of the service coupling, and take it off. Press on the latch with a thumb and take the straight or contra-angle handpiece off by simultaneously pulling in the direction of the axis of the service coupling. Push back the locking sleeve and take off the head. 5.6 Reposition the treatment couplings The MULTIflex coupling must always be positioned on servicing station 4 (far right). Couplings with a 4-hole standard connector for third-party products can also be mounted. See chapter on "Consumables and tools". The unit is configured as follows at the time of delivery: 3 x service couplings for INTRAmatic 26 / 60

1 x service coupling for MULTIflex 5 Operation 5.7 Chuck Servicing Unscrew the union nut clockwise in the direction of the arrow, and take the service coupling off. Attach the service coupling as selected and fasten it turning the union nut counterclockwise against the direction of the arrow. 5.7 Chuck Servicing Handpieces must be taken off the service couplings before the chuck service can be started and performed. Please cover the head region of the handpiece with a paper tissue to prevent any contamination of the ambient area during chuck service. Close the front door and press the chuck service button 2 for at least three seconds until the spray canister control LED 1 flashes three times consecutively. ð The device is in chuck service mode. 27 / 60

5 Operation 5.7 Chuck Servicing A holder for accommodating the chuck service adaptor is provided in the side door. The service adaptor is included in the accessories package at the time of delivery. Take the chuck service coupling out and attach it to coupling service point 4 (on the far right). A MULTIflex adaptor must be mounted on the coupling. Press the handpiece, with the chuck to be serviced leading, against the tip of the chuck service coupling. If this is done with insufficient pressure, there is a risk of the chuck service adaptor being pushed off the service coupling during the servicing and of the servicing agent leaking. Press the chuck service button 2. ð The spray is then briefly injected through the chuck, completing the service. Close the chuck service mode. Place the handpieces in the QUATTROcare PLUS 2124 A, close the front door and start theservice procedure. Alternatively, after three minutes with no service procedure running, the device automatically switches back to normal service mode. 28 / 60

5.8 Set the blow-out time 5 Operation 5.8 Set the blow-out time With the QUATTROcare PLUS 2124 A the blow-out time for straight and contra-angle handpieces and turbines can be changed. Hold down the start key 2. Turn the main switch3 on. ð LEDs 1 to 4 of the LED display1 light up one after the other. The blow-out time is as follows for: LED 1: 15 seconds LED 2: 20 seconds LED 3: 30 seconds LED 4: 40 seconds Release the start key 2 for the desired blow-out time. ð An acoustical signal is emitted if a blow-out time was set on the unit. 29 / 60

6 Check and adjust 6.1 Check and adjust the compressed air input 6 Check and adjust 6.1 Check and adjust the compressed air input Remove the two Torx 3x10 screws, and remove the rear wall from the device. 4.50- Separate the compressed air connection between the compressed air filter and the pressure controller at the swivel banjo coupling of the pressure controller. Connect the digital manometer with the swivel banjo coupling. Connect the compressed air supply. Compare the display on the digital manometer with the specification 4.5 ± 0.2 bar (65 ± 3 psi). If necessary, adjust the pressure setting on the pressure controller. Restore the hose connection filter pressure controller. Re-attach the back wall. 30 / 60

6.2 Open the service menu 6 Check and adjust 6.2 Open the service menu Hold down the flap 10 and start button 5, and turn on the main switch6. Press and hold the flap 10 and start button 5 until all LEDs are shining simultaneously. The service menu is activated. Point 7.3 Detection switch care agent test must be carried out in order to proceed to the principal level. 6.3 Run the test for the care product recognition switch Press all flaps 7 to 10 one after the other. ð The function is acknowledged with a tone and a corresponding LED display 3. The Standby LED 2 flashes. 31 / 60

6 Check and adjust 6.4 Adjust the recognition switch 6.4 Adjust the recognition switch The recognition switches are under flaps 7 to 10. Hold and press the flaps 7 and 10 and press start button 5. Place the handpieces on the couplings and lock it in place. Push the front flap 11 up, and leave a gap approx. 2 cm wide. ð The function of the recognition switch is acknowledged with a LED display 3. The numbering of the adapters corresponds to the numbers of the LED displays 3. If a recognition switch is not activated, push the corresponding flap up and turn the screw underneath to the left. Only close the front flap 11 when all the recognition switches are activated. Press the start button 5 to switch to the main level. ð The Standby LED 2 flashes. 32 / 60

6 Check and adjust 6.5 Check the care product outlet in the individual adapters 6.5 Check the care product outlet in the individual adapters Oil is leaking. Keep the flap 9 pressed and simultaneously press the start button 5. Press all flaps 7 to 10 one after the other. The light barrier (oil sensor) is disabled during this process. ð The solenoid valves are enabled and the service product is dispensed for as long as the corresponding flap and the start button 5 are pressed. Press the start button 5 to switch to the main level. ð The "Stand-by" LED 2 flashes. 6.6 Check the blown air at the adapters Keep the flap 10 pressed and simultaneously press the start button 5. 33 / 60

6 Check and adjust 6.7 Check the solenoid valve Press all flaps 7 to 10 one after the other. ð The solenoid valves are enabled for as long as the corresponding flap and the start button 5 are pressed. Air exits for as long as the start button 5 is pressed. Press the start button 5 to switch to the main level. ð The "Stand-by" LED 2 flashes. 6.7 Check the solenoid valve Keep flap 9 and flap 10 pressed and simultaneously press the start button 5. Check all functions one after the other. ð The solenoid valves are active as long as the respective flap and the start button 5 are pressed. Press the start button 5 to switch to the main level. ð "Stand-by" LED flashes. 6.8 Check for absence of leakage 6.8.1 Box coupling 34 / 60

6 Check and adjust 6.8 Check for absence of leakage Screw the test adaptor 12 (Mat. no. 1.005.6156) into the bottle coupling. Connect the hose coupling to the leakage air from test instrument 2182 KL (Mat. no. 0.411.8540). Pressurisation with 5 ± 0.2 bar (73 ± 3 psi). The compressed air should not decrease for 20 seconds. 6.8.2 Oil system Open/turn on the 2/2-way valve Press and hold flaps 9 and 10, and press the start button 5. ð "Stand-by LED 2 lights up permanently. Press and hold flaps 9 and 10. Activating the start button 5 turns the valve on and off (toggle). Screw the test adaptor 12 (Mat. no. 1.005.6156) into the bottle coupling. Connect the hose coupling to the leakage air from test instrument 2182 KL (Mat. no. 0.411.8540). Pressurisation with 5 ± 0.2 bar (73 ± 3 psi). The compressed air should not decrease for 20 seconds. 6.8.3 Air system 35 / 60

6 Check and adjust 6.9 Read the number of run cycles Apply the compressed air hose on the pressure controller at the swivel banjo coupling. Connect the compressed air hose to the leakage air from test instrument 2182 KL (Mat. no. 0.411.8540). Pressurisation with 5 ± 0.2 bar (73 ± 3 psi). The compressed air should not decrease for 20 seconds. Press the start button 5 to switch to the main level. ð "Stand-by" LED 2 flashes. 6.9 Read the number of run cycles Hold flap 8 and 10 and press the start button 5. ð The LED display 3 shows the number of run cycles. Read out the number of run cycles: LED 1 and LED 2 simultaneously stand for hundred thousands LED 3 and LED 4 simultaneously stand for ten thousands LED 1 stands for thousands LED 2 stands for hundreds LED 3 stands for tens LED 4 stands for ones The LED is active for 1 second and stops for 1 second. The counting always starts with the largest unit. The pause is always 2 seconds long when switching from thousands to hundreds and tens or ones. After the counting is over, the menu automatically switches back to the main level of the service menu. To leave the service menu, turn off the device at the main switch 6! 36 / 60

6.10 Check sensor 6 Check and adjust 6.10 Check sensor Place the handpieces on the service couplings and let them lock in place. The system (oil hoses, valve block, sensor) must be filled with oil spray. Screw the test adaptor 12 (Mat. no. 1.005.6156) into the bottle coupling. Connect the test adaptor 12 to the adaptor hose 13 (Mat. no. 1.008.6967) and to the turbine connection of the test instrument 2182 KL (Mat. no. 0.411.8540). Pressurisation with 3.8 bar +- 0.2 bar (55 ± 3 psi). Start the servicing programme with the start button 5. The spray bottle control LED 1 must flash by the fourth handpiece/ servicing process, at the latest. 37 / 60

6 Check and adjust 6.11 Check barcode reader 6.11 Check barcode reader Insert original KaVo spray canister with barcode. Start the conditioning cycle with at least one allocated position. Reader reads the imprinted code, The conditioning cycle must start now. The message "Barcode defective", displayed by the running light of the LED may not appear. However this can be activated if no spray canister is inserted. The spray canister control indicator could light up. 6.12 Check the chuck maintenance Switch the device on. 38 / 60

6 Check and adjust 6.12 Check the chuck maintenance Grip the recesses in the left and right of the front flap, and pull it forward to open it and fold it down. Handpieces must be taken off the service couplings before the chuck service can be started and performed. Open the side door and remove the clamping chuck service coupling from the side door. 39 / 60

6 Check and adjust 6.13 Set the blow-out time Attach the clamping chuck service coupling to the right-hand service coupling (MULTIflex service coupling). Cover the tip with a cloth and activate the start button for the chuck servicing process in the control panel. ð A brief jet of spray should be ejected from the clamping chuck service coupling. ð All four LEDs in the control panel light up green simultaneously. Close the chuck service mode. Option 1: Place the dental handpieces in the QUATTROcare PLUS 2124 A, close the front door and start theservice procedure. Option 2: After three minutes with no service procedure running, the device automatically switches back to normal service mode. 6.13 Set the blow-out time With the QUATTROcare PLUS 2124 A the blow-out time for straight and contra-angle handpieces and turbines can be changed. Hold down the start key 5, and turn on the main switch6. For the duration of 5 seconds, the blow-out time that is set/saved in the instrument is indicated by the LED display 3. Release the start button 5 within the 5 seconds. The "old" setting is retained. A beep sounds for 3 seconds. Press the start button 5 for longer than 5 seconds. The new adjustment of the oil blow-out time begins. The LEDs 1, 2, 3 and 4 light up sequentially for 1.5 seconds. 40 / 60

6 Check and adjust 6.14 Setting the oil servicing time Release the start button 5 for the corresponding blow-out time of the respective LED display 3. The selected blow-out time will be accepted and saved. A beep sounds for 3 seconds. The blow-out time is at: - LED 1: 15 seconds - LED 2: 20 seconds - LED 3: 30 seconds - LED 4: 40 seconds 6.14 Setting the oil servicing time With the QUATTROcare, the oil service time can be changed for the handpieces and contra-angle handpieces and turbines. Hold down the clamping chuck service start button 4, and turn on the main switch6. For the duration of 5 seconds, the oil service time that is set/saved in the instrument is indicated by the LED display 3. Release the chuck service start button 4 within the 5 seconds. The "old" setting is retained. A beep sounds for 3 seconds. Press the chuck service start button 4 for longer than 5 seconds. The new adjustment of the oil service time begins. The LEDs 1, 2 and 3 light up sequentially for 1.5 seconds. Release the clamping chuck service start button 4 as soon as LED 3 lights up that corresponds to the desired oil servicing time. The selected oil service time will be accepted and saved. A beep sounds for 3 seconds. The oil service is at: - LED 1: 100 msec - LED 2: 150 msec - LED 3: 200 msec - LED 4: not allocated 41 / 60

7 Bridging the barcode reading function 6.14 Setting the oil servicing time 7 Bridging the barcode reading function If the barcode of a spray canister is not readable (foreign canister, contaminated and/or scratched), the spray canister from QUATTROcare PLUS will not be detected. The barcode reading function can be bridged via a key combination. Activate bridging Press and hold down the start button for conditioning process and the start button for the chuck conditioning process simultaneously. Switch on the device using the main switch 6. Press and hold down the start button 5 and start button for the conditioning process 4 until you hear the signal. Switch off the device using the main switch 6 and switch on again. ð The barcode reading function is bridged. ð The LED "Oil" flashes during the conditioning process. ð Now, a spray canister with an illegible or invalid barcode can be used. Limitation Max.1500 single place conditioning cycles can be carried out with the bridging function. These 1500 conditioning cycles are divided into three blocks of 500 conditioning cycles each. Hence the bridging function can be called three times, at the most. Consequently, 1500 conditioning cycles can be carried out with spray canisters that do not comply with the specifications. The QUATTROcare PLUS switches to disturbance when the max. number of 1500 conditioning cycles is reached. The disturbance can only be rectified by a service technician. Deactivating the bridging status The bridging function is automatically deactivated after the insertion of a KaVo QUAT TROcare PLUS spray canister with a valid/ legible barcode. The care cycles activated in the bridging function remain stored. The oil control LED 1 extinguishes. If an original KaVo spray canister is used after a random number of conditioning cycles, but fewer than 500, the timer will stop and will not be reactivated at this point until a spray canister is used that is not the original KaVo canister. After the maximum 1500 cycles in bridging mode have run, this function from QUATTROcare PLUS will switch and the KaVo technical service must be notified. 42 / 60

8 Safety checks pursuant to DIN VDE 701-702 / DIN EN 62638 8.1 Introduction 8 Safety checks pursuant to DIN VDE 701-702 / DIN EN 62638 8.1 Introduction The basic requirements for electrical safety checks to be carried out during re-testing of laboratory devices pursuant to VDE 0701-702 are specified in the following. The re-tests have to be carried out by professionals in electrical engineering or by individuals trained in electrical engineering and supervised by a professional in electrical engineering. A professional in electrical engineering is defined to be anyone whose training, knowledge and experience as well as knowledge of the pertinent regulations allow him or her to assess and recognise possible hazards. The professional in electrical engineering in charge of testing has to make a decision whether nor not further individual tests are required in order to attain the safety target. If any of the following failures or damage is/are detected during the re-test, the testing process must be discontinued and the device must be labelled accordingly and blocked from further use or submitted for repair of the defect. Damage, signs of inappropriate interventions or modifications that could lead to a reduction in safety Damage that may have occurred during the use of the device Occurrence of functional deficiency 8.2 General information The QUATTROcare servicing unit must be subjected to a service check including safety check ("STK")every 5 years. This may only be done by a professional trained by KaVo, or in a shop trained by KaVo. The QUATTROcare Plus Spray canister contains explosive propellant gas. For the safety testing of the QUATTROcare servicing unit, please comply with the set-up conditions. Please comply with the Instructions for Use. Do not expose the QUATTROcare to direct sunlight. Do not place the QUATTROcare right next to a source of heat. Keep away from ignition sources, and do not use open fire in the vicinity of the unit. Ensure that the room is adequately ventilated. 8.3 Visual inspection Nameplate on the rear of the instrument: existing and undamaged the rating data for the following electrical safety check and pressure test Accept and record the product and manufacturer designation as well as the serial number in the test document s on the spray chamber door: Observe symbols of the instructions for use existing and legible 43 / 60

8 Safety checks pursuant to DIN VDE 701-702 / DIN EN 62638 8.4 General design test Symbol wear safety goggles existing and legible Safety instructions existing and legible in 7 languages Control panel: Writing is complete and legible 1 LED spray can control 2 LED control light main switch (on/off) 3 LED display 4 Start button for chuck service 5 Start button for service procedure Reading the servicing cycles from the memory. 8.4 General design test Check the unit for signs of external damage. Check on the stable positioning of the unit. Check the instrument legs for completeness and damage. Check whether all screws under the floor plate for completeness and firm seat. Check whether there are signs of overload, overheating or other inappropriate use or operation. Check the equipment for entirety (power cable, collection trough, service couplings, O-rings). Check for intended use/ selection of electrical cables and fluid lines. Check condition of power plugs, power cables, protection against bending, and damage to the insulation. Check power switch for condition and function. Check the fluid ports and lines for condition and damage. Check the switches, control units and operating elements as described below: Check that the front flap securely closes and is easy to open with two hands. Check that the position switch of the service coupling moves easily. Check the O-rings on the service coupling for completeness and damage. Check the care agent canister as described below: 44 / 60

8 Safety checks pursuant to DIN VDE 701-702 / DIN EN 62638 8.5 Pressure/Leakage test The canister for the care agent can be easily screwed in or off. Check the thread for damage. Check the gasket for completeness and damage. Check the function of the pivoting mechanism. The touchable, conductive parts, which are taken into account for the PE, insulation and leakage current testing, must be determined. 8.5 Pressure/Leakage test Check the tubing and screw connections with sealing points, pressurised with compressed air and pressure from the care agent canister, for water tightness. 8.5.1 Test equipment Name Material number Test device 2182 KL 0.411.8540 Test adaptor 1.005.6156 QUATTROcare Plus Spray can 1.005.3843 Digital manometer 8.5.2 Preparation To gain access for testing of the internal fluid hoses, remove the two Torx screws 3x10 and take off the rear wall. Check all the hoses and their connections for proper seat on the nipples, brittleness and cracks, etc. Replace any hose that exhibits defects immediately. Check the seat of the plugs and contacts on the electrical cables. Check the insulation of the cables at critical points such as edges and corners. Check the seat of the plugs and contacts on the electrical cables. Check the insulation of the cables at critical points such as edges and corners. 45 / 60

8 Safety checks pursuant to DIN VDE 701-702 / DIN EN 62638 8.5 Pressure/Leakage test 8.5.3 Air system Test device 2182KL Carry out the leakage test with test device 2182KL (Mat. no. 0.411.8540). Connection to air leakage test socket on the test device. Test pressure: 5 ± 0.2 bar (73 ± 3 psi) Measuring time for pressure drop: 20 seconds Connect the test device to the compressed air connection socket. Open the pressure controller by turning it clockwise to the limit stop (inside the QUATTROcare PLUS unit). Start the test sequence on the test device 2182KL. ð The pressure of the compressed air must remain constant. The hoses must be attached safely to the nipples after the testing of the air system. A visual inspection must be carried out. 46 / 60

8 Safety checks pursuant to DIN VDE 701-702 / DIN EN 62638 8.5 Pressure/Leakage test 8.5.4 Setting the pressure controller 4.50- Separate the compressed air connection between the compressed air filter and the pressure controller at the swivel banjo coupling of the pressure controller. Connect the digital manometer with the swivel banjo coupling. Connect the compressed air supply. Compare the display on the digital manometer with the specification 4.5 ± 0.2 bar (65 ± 3 psi). If necessary, adjust the pressure setting on the pressure controller. Restore the tube connection filter to the pressure control. Check for water tightness. See also: 2 Air system, Page 0 Attach the rear wall. 8.5.5 Oils system - Bottle coupling Screw the test instrument with test adaptor 12 into the canister coupling. 47 / 60

8 Safety checks pursuant to DIN VDE 701-702 / DIN EN 62638 8.6 Electrical test Start the test sequence on the test device 2182KL. ð The pressure of the compressed air must remain constant. 8.5.6 Oil system - Solenoid valve to valve block Connect the device to the mains power and open the service menu. Open/turn on the 2/2-way valve. Press and hold flaps 9 and 10, and press the start button 5. ð "Stand-By"- LED 2 shines continuously. Press and hold flaps 9 and 10 and simultaneously press start button 5. Activating the start button 5 turns the valve on and off (toggle). Screw the test adaptor 12 into the bottle coupling. Connect the tube coupling to the leakage air from the test instrument 2182KL. Start the test sequence on the test device 2182KL. ð The pressure of the compressed air must remain constant. 8.6 Electrical test The QUATTROcare PLUS is a protection class I device for a supply voltage (operating voltage) of 240 VAC 50 Hz to 100 VAC 60 Hz. 8.6.1 Electrical test devices Test devices must comply with VDE 0404; 0413-1, 0413-2; 0413-4. 8.6.2 Protective earth conductor test The orderly condition of the electrical connection of the port of the device for the protective earth conductor (earthing contact of the power plug, if applicable) and any part that is connected to the protective earth conductor that might be touched must be demonstrated as follows: Optical inspection of the earthing circuit Measurement of the resistance of the protective earth conductor during which various sections and the input sites of each cable included in the measurement must be moved 48 / 60

8 Safety checks pursuant to DIN VDE 701-702 / DIN EN 62638 8.6 Electrical test Limit Test current Test set-up < 0.3 Ω 0.2 A Measuring circuit C1a or C1b must be used. Device with protective earth conductor and plug connection. Measuring points: Protective earth conductor of power plug against chassis (the chassis is the fastening screw on the power input module). Requirements of test instruments The measuring voltage can be either DC or AC. The open-circuit voltage must be no more than 24 V and no less than 4 V. The measuring current within the measuring range between 0.2 and 1.99 Ω may not fall below 0.2 A. A lower test current has no cleaning effect on the contact site (test instrument in accordance with VDE 0413-1 and -4). 8.6.3 Measurement of the insulation resistance With the measurement of the insulation resistance, primarily the condition of the insulations as well as the creepage distances of the instruments should be evaluated. Moreover, it can be tested whether the safety of the instruments is impaired by contamination and/or ingress of moisture. The insulation resistance must be measured between the active parts and any touchable conductive part, including the protective earth conductor. The measuring circuits listed in the table must be used. If applicable, the measurements must be carried out for various switch positions. 49 / 60

8 Safety checks pursuant to DIN VDE 701-702 / DIN EN 62638 8.6 Electrical test Test voltage Measuring current Limit Test set-up At least 500 V DC At least 1 ma Protection class 1 without heating elements 1.0 MΩ 2a,.. 2d Device with protective earth conductor and plug connection as well as touchable conductive parts that are not connected to the protective earth conductor. Measuring points Mains circuit: Phases on the power plug together against device mass and the screw sites on the rear and front walls of the housing as well as the pivoting mechanism/ bottle coupling. Requirements of test instruments The value of the measuring voltage of the measuring direction must be at least equivalent to the rated voltage of the instruments top be tested against earth, but not less than 500 V DC. The measuring current must be at least 1 ma. (Test devices pursuant to VDE 0413-1 and -2). 8.6.4 Measurement of the protective earth conductor current Sum of all currents flowing through the insulations of a device to the protective earth conductor, depends on the circuiting in the device. The protective earth conductor current must be measured on all devices fitted with a protective earth conductor. If a one-phase device can be connected to the supply circuit independent of its polarity (non-polarised connection plug, connection cable having no plug), the measurement must be made in all positions of the connector or connection cable. All switches, regulators and similar means must be closed during the measurement in order to fully include all active parts in the measurement. If applicable, the measurements must be carried out for various switch positions. 50 / 60

8 Safety checks pursuant to DIN VDE 701-702 / DIN EN 62638 8.7 Functional test Limit Protective earth conductor current < 3.5 ma Test set-up 3a, 3b and 3d, or 3c Measuring procedure for equivalent leakage current Measuring points In the protective earth conductor with the device operational (run servicing cycle). Requirements of test instruments The internal resistance of the protective earth conductor current measuring facility must not exceed 5 Ω. The measuring range must at least range from 0.25 ma to 19 ma. The test facility must evaluate the measured current as an effective value independent of curve shape. (Test devices pursuant to VDE 0404-4). 8.6.5 Proof of effectiveness of other protective facilities Check fuse for correct rating - T1000 ma. 8.7 Functional test Connect the device to the electrical power and the compressed air supply to the compressed air input on the device. Screw the QUATTROcare PLUS spray canister into the canister coupling. Switch the device on. 51 / 60

8 Safety checks pursuant to DIN VDE 701-702 / DIN EN 62638 8.7 Functional test 8.7.1 Check the output of servicing product in the individual adaptors Open the service menu first. 7 8 9 10 2 5 Oil is leaking. Keep the flap 9 pressed and simultaneously press the start button 5. Press all flaps 7 to 10 one after the other. The light barrier (oil sensor) is disabled during this process. ð The solenoid valves are enabled and the service product is dispensed for as long as the corresponding flap and the start button 5 are pressed. Press the start button 5 to switch to the main level. ð The "Stand-by" LED 2 flashes. 8.7.2 Check the blown air in the individual adaptors 7 8 9 10 2 5 52 / 60

8 Safety checks pursuant to DIN VDE 701-702 / DIN EN 62638 8.8 Analysis and labelling Keep the flap 10 pressed and simultaneously press the start button 5. Press all flaps 7 to 10 one after the other. ð The solenoid valves are enabled for as long as the corresponding flap and the start button 5 are pressed. Air exits for as long as the start button 5 is pressed. Press the start button 5 to switch to the main level. ð The "Stand-by" LED 2 flashes. 8.7.3 Check the solenoid valve 7 8 9 10 2 5 Keep flap 9 and flap 10 pressed and simultaneously press the start button 5. Check all functions one after the other. ð The solenoid valves are active as long as the respective flap and the start button 5 are pressed. Press the start button 5 to switch to the main level. ð "Stand-by" LED flashes. The safety labels must be checked after each completed individual test. 8.8 Analysis and labelling The device passes the test if it passes all individual parts of the test. The respective instrument should be labelled accordingly. 53 / 60