Smith Meter TM Turbine Meters Models 4" Through 24" Sentry TM Series. Service. Contents

Similar documents
Smith Meter TM Turbine Meters Models 3" Through 16" MV Series Service

Form No. P Meter, PD Model SD, ASD, SF, or ASF-4, 4R, 4V, or 4RV-NF Parts List

PRIME 4 Service. Smith Meter Rotary Vane Meters. Contents

Smith PD Rotary Vane Meters F4 PRIME Cover Field Retrofit Procedure Service Bulletin

Smith Meter AccuLoad III AccuLoad III-S and N4 Hardware Worksheet

OrificeMaster Specifications

Issue/Rev. 0.3 (2/04) Bulletin P

Non Ferrous Aircraft Fueling Meters

Models F4-V1, G6-V1 and G6-V3

Smith Meter TM PD Rotary Vane Meters F4-S1/A1-S3/A3 PRIME Upgrade Upgrade and Service

Tools, Mechanical. All or Multiple Products. Parts List. Feeler Gauges. Spiders and Bushings. Feeler Gauge Extension Rod. Cam Setting Bushing

Electronic Preset Delivery System Smith Meter AccuLoad III. Vapor Recovery. Vapor Recovery System. Overview

Additive Injectors. AccuPlus. Maintenance. The Most Trusted Name In Measurement

Model G6-S3. Repeatability ±0.02%

Densitometers L-Dens 427F Ex Density Sensor Specifications

Orifice Fittings and Meter Tubes

Smith Meter PD Meter Meter, PD Series, K12- S1, S3, S5, S6, and S7

Smith Meter PD Meters Meter, PD Series, M16-S3, S5, S6 (1993-Present) Parts List

Model DE-3 Air Release Heads Parts List

Transmitter, Model UPT

Flow Conditioners. We put you first. And keep you ahead. FEATURES OPTIONS. Bulletin SS02007 Issue/Rev. 0.9 (10/14)

MV Series. Smith Meter Turbine Meters. Features. Operating Specifications SPECIFICATIONS. Bulletin SS02016 Issue/Rev. 2.0 (2/18)

MV Series. We put you first. And keep you ahead. SMITH METER TURBINE METERS FEATURES OPERATING SPECIFICATIONS. Bulletin SS02016 Issue/Rev. 1.

Meter, PD Series, K12-S1, S3, S5, S6 and S7

Maintenance Information

Double Case/2-Bladed. Smith Meter PD Rotary Vane Meters. Service. Contents

COYOTE ENTERPRISES, INC. 9/10 BLAST WHEEL MAINTENANCE & ASSEMBLY MANUAL

Transmitter, Model UPT

Maintenance Information

National AirParts, Inc. N400 Belt Drive Alternator

Smith Meter Rotary Vane PD Meters Single and Double Case PD Meters Installation / Operation / Maintenance Manual

IMPORTANT!!!! Read this manual before attempting any installation, wiring or operation.

PD Rotary Vane Meters Double Case/3 Bladed

INSTRUCTION MANUAL IM-422 For HTC/COUPLING ASSEMBLY HC-8088

Smith Meter AccuLoad IV-QT Hardware Worksheet

DeZURIK " BAW AWWA BUTTERFLY VALVES WITH EPOXY-RETAINED SEAT

TR-QS Wafer-Style Turbine Flow Meters Installation, Operating & Maintenance Manual

Liquid Turbine Flowmeter

Industrial Turbo Meters, Sizes 2" through 6"

Smith Meter PD Meters 3" Steel Model E3

3.2 DRIVE TORQUE HUB. Roll, Leak and Brake Testing SECTION 3 - CHASSIS & TURNTABLE. 3-2 JLG Lift

TECHNICAL SERVICE MANUAL GENERAL PURPOSE BRACKET MOUNTED PUMPS SERIES 115 MODELS G, GX2, H and HX4

Section 10 Chapter 17

NOTE: Visit our website at for video repair procedures, under the Tools section.

TABLE OF CONTENTS GENERAL INFORMATION

Maintenance Information

MODEL 1100 TURBINE FLOW METER

CAB TILT HYDRAULIC SYSTEM

3, 4 & 6 (DN 80, 100 & 150) CLASS 150 & 300 STANDARD BORE SERIES 7150, 730S & 7300 FLANGED BALL VALVES EQUIPPED WITH KEROTEST HANDLE ASSEMBLIES

EZ-IN Series Turbine Flowmeter

LJ20 Distributor - Disassembly Inspection Repair

SUSPENSION 2-1 SUSPENSION TABLE OF CONTENTS

TECHNICAL SERVICE MANUAL

TECHNICAL SERVICE MANUAL GENERAL PURPOSE JACKETED PUMPS SERIES 230 MODELS HX4, KK, LQ, Q, M, N

TECHNICAL SERVICE MANUAL

MANUAL TRANSAXLE Return to Main Table of Contents

INSTRUCTION MANUAL AND PARTS LIST FOR PG/RG3D_-187, 218, 250 and 312 SERIES PUMPS

SuperTrac. Axle. Service & Maintenance. Manual

Table of Contents Visual Inspection and Neutralizing... 3 Disassembly

REMOVAL & INSTALLATION

INSTALLATION, OPERATION, MAINTENANCE MANUAL FOR MANUALLY OPERATED STOP CHECK VALVE

DEMCO Resilient-Seated Butterfly Valve

INTAKE/EXHAUST SYSTEM 7 C TURBOCHARGER

INSPECTION & MAINTENANCE BULLETIN ARI 1301/1302 1" Plug Type Angle Valves

Example Maintenance Schedule 1. The attached document is an example of an acceptable maintenance schedule.

Tooling Assistance Center

DRIVE AXLE Volvo 960 DESCRIPTION & OPERATION AXLE IDENTIFICATION DRIVE AXLES Volvo Differentials & Axle Shafts

CONTENTS. VIKING PUMP, INC. A Unit of IDEX Corporation Cedar Falls, IA USA SECTION TSM 710.1

RSW PRODUCTS, INC Cromwell Houston, Texas Tel: Fax:

Maintenance Instructions

INSTRUCTION MANUAL. for BUTTERFLY VALVES

Fluid-O-Tech ROTOFLOW ROTARY VANE PUMP REBUILD MANUAL

PRODUCT SERVICE MANUAL. BK6DHZ(C)-250, 275, 312 and 400 PUMPS

Maintenance Information

REAR AXLE AND SUSPENSION RA 1

3" Steel Model E3. We put you first. And keep you ahead.

ProLine. 44 Mower. for 120 Traction Unit. Model No & Up. Operator s Manual

Maintenance Information

Vickers. Service Data. Vane Pumps. Quiet Intra-Vane Single Pumps. 25V(T)-35V(T)-45V(T) Series 20 Design 282/283 Series

AVK SERIES 41 SWING CHECK VALVE FIELD MAINTENANCE AND INSTRUCTION MANUAL FOR SWING CHECK VALVES 3" - 12"

HIGH PRESSURE CONTROL VALVE R2L ACTUATOR

REAR AXLE <FWD> GROUP 27A 27A-1 CONTENTS GENERAL INFORMATION... 27A-2 SPECIAL TOOLS... 27A-4 ON-VEHICLE SERVICE... 27A-5 SPECIFICATIONS...

4.2 WATER PUMP (GEAR CASE MOUNTED AND LATER) (GCM)

Model 1100 Turbine Flow Meter

Southwest Windpower Instruction Sheet AIR-X Circuit Replacement Kit

Valtek Spring Cylinder Linear Actuators

AVK SERIES 41 SWING CHECK VALVE FIELD MAINTENANCE AND INSTRUCTION MANUAL FOR SWING CHECK VALVES 3" - 12"

Rheometer Calibration Fixture Instruction Manual Model 280 & 286

Type A10. Falk Freedom Disc Couplings Installation and Maintenance. Type A10 Sizes 85 thru 8770 (Page 1 of 5)

TRANSMISSION AND TRANSFER CASE

PRODUCT SERVICE MANUAL FOR BK12DHZ PUMPS

Bray/ VAAS Slurry Series Knife Gate Valve 760/762/765/766/767/768 Series Operation and Maintenance Manual

Installation, Operation and Maintenance Manual RGS Rev. 0 November Bettis RGS F-Series Spring-Return Actuators

Amerigear SF Spindle

INSTRUCTIONS Installation Operation Repair

WKM Model MB DynaCentric High Performance Butterfly Valve

Maintenance Information

73000 Series Masoneilan* Sweep Angle Control Valves Models XX-73471

PLACING THE MOTOR IN SERVICE

Vane Pumps. Single Stage Single and Double Vane Pumps. Vickers. Overhaul Manual

Transcription:

Smith Meter TM Turbine Meters Models 4" Through 24" Sentry TM Series Service Issue/Rev. 0.3 (8/02) Bulletin MN02004 Contents Section 1 General Information... Page 2 Return Procedures... Page 2 Section 2 Service... Page 3 Section 3 Troubleshooting Guide... Page 8 Section 4 Maintenance and Calibration Record... Page 9 Section 5 Related Publications... Page 10 The Most Trusted Name In Measurement

Section 1 General Information It is suggested that a detailed record be maintained for each meter. Nameplate data, progressive totalizer readings, meter factor, parts used, and other similar information provide background material for scheduling a preventive maintenance program. An increase in meter factor drift against throughput can be used as the basis for making an inspection. The time of the first inspection must be based on the operating conditions imposed by the installation. Flow rate, lubrication properties of the fluid, and the possibility of abrasive contaminants are points to consider. Then, at the time of inspection, the condition of the meter should indicate whether the inspection interval could be lengthened or shortened. All parts, as they are removed, should be thoroughly inspected and, if necessary, cleaned in solvent. Before any disassembly is performed on the meter, be sure the trouble is in the meter. 1. Check that the meter is being operated within the proper flow rate. Refer to the nameplate on the side of the meter. 2. Check pick-up coil and preamplifier for proper operation. An oscilloscope will be necessary to determine wave form of the pick-up coil. Refer to Bulletin SS02001 for pick-up coil specifications and Page 3 for preamplifier information. 3. Check for electronic totalizer malfunction. 4. Check that the system flow temperature has stabilized, that there are no abnormal pressure fluctuations, and that there is proper system back pressure. 5. Check valve operation in the system. 6. Check for air in the system. After all other accessories have been checked and found to be operating normally, the turbine meter should be checked. Before beginning any service procedure, stop product flow, drain the line in the direction of flow (if possible), and relieve the pressure from the system. Remove the Sentry Series Turbine Meter from the line and take it to a clean area for disassembly. Return Procedures No material can be returned from either a distributor or customer without receiving a properly executed Return Material Authorization (RMA). Once the RMA is received, the material may be returned in accordance with the instructions contained in the RMA. Authorization for the return of either new or used material can only be granted by the Customer Service Department of FMC Measurement Solutions. If the material to be returned is new and unused, the customer should supply the invoice number and/or the shipping order number of the original purchase. For any units that have been installed and could contain a residue of product, it is the responsibility of the customer to properly flush and, if necessary, neutralize the inside of the equipment in question. If not properly accomplished, the customer must assume all responsibility for any injuries, property damages, or violations of state or local statutes. All items to be returned must be freight prepaid to the FMC Technologies Measurement Solutions, Inc. Parts Operation, 1602 Wagner Avenue, Erie, Pennsylvania 16514 and shipped in accordance with all rules and regulations of the Department of Transportation and Environmental Protection Agency. 2 MN02004 Issue/Rev. 0.3 (8/02)

Section 2 Service Removal and Replacement of Pick-Up Coil With Meter in Product Line 1. Make certain power has been disconnected to meter. 2. Remove cover from junction box by turning counterclockwise. 3. If a preamplifier is being used, disconnect pick-up coil wires from Terminals 6 and 7 of PA-6 Preamplifier. 4. Remove junction box from meter by turning counterclockwise, Figure 1. Caution: Tighten pick-up coil finger tight. Over tightening will result in damage to coil. Removal and Replacement of Preamplifier With Meter in Product Line 1. Remove pick-up coil as outlined above, Steps 1 through 5. 2. For PA-6 removal, remove #6-32 round head screw, Figures 3 and 4, to permit removal of metal clip, Figure 5. Figure 3 Figure 1 5. Remove pick-up coil from meter by turning counterclockwise, Figure 2. 6. Check resistance of coil (nominal dc coil resistance - 1,020 Ω) to determine if coil is faulty. 7. For reassembly, reverse procedure. Figure 4 Figure 2 Figure 5 Issue/Rev. 0.3 (8/02) MN02004 3

Section 2 Service (continued) 3. Disconnect output wiring and remove preamp from junction box, Figure 6. 4. For reassembly, reverse procedures. 8. Remove downstream stator, Figure 9. In a unidirectional turbine meter, the stator is located by a Woodruff key. Care should be taken that the Woodruff key is not lost. In a bidirectional turbine meter, there is no Woodruff key; the two stators are located by two deflector rings. Figure 6 Removal and Servicing of Meter lnternals 1. Drain product from line. Drain in direction of flow (if possible). 2. Make certain power has been disconnected to meter. 3. Remove junction box cover, disconnect output wiring and conduit where required. 4. Make certain meter is properly supported before removing bolts and nuts. 5. With meter removed, check straightening section to make certain tube bundle is clean and free from damage. 6. If necessary to remove pick-up coil and preamplifier, refer to procedures outlined above. 7. Remove cotter pin and slotted nut from downstream end of meter, Figures 7 and 8. Figure 9 9. Inspect the holes in the downstream stator fins on the thrust bearing end, Figure 10, to make certain they are open and the inside of the stator is clean. Inspect thrust washer for signs of roughness, grooves, or cracks. Figure 10 Figure 7 Figure 8 If thrust washer shows signs of wear but is not cracked, turn washer over. In some of the older turbine meter designs, it will be necessary to apply a small amount of heat to the stator around the thrust washer area to remove the washer. On models after serial number 229T, the thrust washer is secured in place with a small amount of Locktite No. 72-41. This can also be removed by applying a small amount of heat around the washer. The washer can then be turned over for reuse. If the new bearing surface shows evidence of Locktite (yellowish color), it can be removed with acetone. 4 MN02004 Issue/Rev. 0.3 (8/02)

Section 2 Service (continued) 10. Remove rotor and platform bearings, Figure 11, using extreme care that the platform bearing is not permitted to fall from the rotor since, due to the brittleness of the bearing, a fall could result in severe damage. Inspect outside surface of bearing for grooves, chipping, roughness, and cracks, as well as cracks in end surfaces. When removing the rotor from a bidirectional turbine meter, it is necessary to remove the downstream deflector ring. Remove the set screw and slide the ring out. When reinserted, the ring must be relocated exactly in its former position as the ring properly positions the downstream stator in relation to the upstream stator. Figure 12 12. Remove upstream stator with shaft and inspect as described in Step 9. On most unidirectional models, holes are in nut end of stator (Figure 13). Figure 11 Inspect inside surface and end surfaces of rotor bearing for same defects as for platform bearing. If bearing surfaces show no defects, the bearing will be suitable for reuse in most cases. Bearing wear is minimal and bearing replacement is usually required only due to damaged surfaces. 11. Should rotor bearing require replacing, it is suggested that rotor assembly be returned to FMC Measurement Solutions for rework. See Return Procedures, Page 2. Rotor bearing replacement can be done in the field by the following repair procedure. a. With proper tool and fixture, Figure 12, the old rotor bearing may be forced out of the rotor hub with an arbor press. Care must be taken in the selection of tool and fixture to insure a straight line or axial press only. Push the bearing out with a smooth, steady motion. b. The new bearing may be installed with the same equipment used to remove the old bearing. Center the bearing carefully in line with the hub bore before pressing. The bearing material has been selected for its hardness and wear characteristics, but may be chipped or broken if mistreated. Press the bearing into the rotor hub to achieve equal protrusion within 0.005" from both hub faces. Figure 13 In some bidirectional turbine meters, two downstream stators are used. Therefore, the holes will be located as shown in Figure 10. 13. For a unidirectional turbine meter, it is not necessary to remove deflector ring in order to service meter internals. If ring is removed, it must be relocated exactly in its former position when reinserted as ring properly positions upstream stator in relation to downstream stator. If necessary to remove, remove set screw and slide ring out from upstream end of meter, Figure 14. Note: It will be necessary to remove one deflector ring when servicing a bidirectional turbine meter. Issue/Rev. 0.3 (8/02) MN02004 5

Section 2 Service (continued) Figure 14 Reassembly of lnternals 1. Inspect to make certain all parts are clean and free from defects, especially replacement parts. Make sure the rotor bearing is properly aligned in the rotor (equal amounts of the bearing should be on each side of the rotor). 2. Place upstream stator on shaft and install shaft nut. Note. Before installing the nut on the shaft, the mating surfaces of the nut, the stator, and the threads of the shaft should be cleaned and lubricated with SAE grade 30 oil (or equivalent). The shaft nut should be properly tightened and secured with the cotter pin. See Table 1 for proper torque values. Table 1 - Shaft Nut Torque Values Meter Torque Required Size Thread Size (Lubricated) ±5% 4" #10-32 UNF-2 22 in-lb 6", 8" 5/16"-18 UNC-2 8 ft-lb 10", 12" 1/2"-13 UNC-2 32 ft-lb 16" 3/4"-16 UNF-2 120 ft-lb 18" 7/8"-14 UNF-2 115 ft-lb 20" 1"-12 UNF-2 170 ft-lb 24" 1-1/4"-12 UNF-2 250 ft-lb 3. Apply blueing or similar compound to downstream side of rotor hub that matches O.D. of stator, Figure 15. Rotor is marked IN on upstream side of rotor. 4. Slip platform bearing on shaft with approximately 1/4" of shaft showing. Lubricate O.D. of platform bearing with light machine oil. 5. Assemble rotor to platform bearing and gently push bearing and rotor into housing until positioned by stator. Do not bump bearing ends against stator. In a bidirectional turbine meter, reinsert deflector ring. Carefully relocate the ring in its exact former position and insert the set screws, but do not over tighten them because this could cause distortion of the deflector ring. Figure 15 6. Assemble downstream stator. In a unidirectional turbine meter, the stator is positioned with the Woodruff key, Figure 16. In a bidirectional turbine meter, the stator is positioned with the deflector ring. Assemble shaft nut; refer to Step 2. Figure 16 7. With air inserted through upstream end of meter, spin rotor at a reasonable speed not to exceed 100 rpm for 1 minute. 8. Disassemble downstream stator and inspect hub for blue markings. If markings are detected, this indicates rotor is rubbing stator which will affect meter linearity. This can be caused by the rotor bearing not having equal protrusion within 0.005" from the rotor hub, or there may be burrs or high spots on the rotor. First, check the rotor bearing and align as outlined on Page 5, Step llb. If this does not correct the problem, remove high or blue spots on stator, Figure 15, by light filing. Repeat this inspection until rotor runs smoothly and then assemble internals completely. 9. Position turbine meter on its upstream flange. Spin rotor with a rod, Figure 17, and visually observe if rotor spins free and returns to rest with smooth rotation. If any drag or friction is detected, this indicates upstream face of rotor O.D. is rubbing against O.D. face of deflector ring. This condition will affect meter 6 MN02004 Issue/Rev. 0.3 (8/02)

Section 2 Service (continued) linearity and may be corrected by repositioning rotor bearing in rotor. Repositioning is accomplished in accordance with procedures in Step 11, Page 5. This procedure can be repeated from the opposite side for a bidirectional turbine meter. 10. Place turbine meter in horizontal position, Figure 18. Additional Notes Check to make certain stators are positioned at 45 to each other, Figure 19. Tests have proven that when stators are positioned in this manner, meter accuracy is improved Upstream Stator Downstream Stator 45 (Approx.) Figure 17 Hold either up or downstream stator and move internal laterally. Assembly should move approximately 0.010" to 0.032". If assembly has lateral movement or end play, this indicates platform bearing is securely locked in place. No end play indicates platform bearing is loose and Figure 18 can rotate about shaft which will result in meter inaccuracy and premature service. Disassemble meter, determine fault, and correct. Figure 19 When transporting meter, make certain meter is not subjected to any severe shock as bearings may be damaged. Cover flange openings to protect internals. Note: When a meter is disassembled and reassembled, it should be recalibrated for optimum accuracy. Test After meter has been inserted into the product line, meter must be proved in order to obtain new meter factor. This is usually accomplished by the use of a displacementtype prover. During testing, should repeatability not be obtainable, check to make certain that pick-up coil is tight. If coil is tight and results are still not satisfactory and assuming prover is functioning properly, it may be necessary to recheck rotor and platform bearing for possible damage during installation. Proving the meter should be under the same conditions of flow rate, pressure, temperature, and liquid characteristics as exist during normal operation of the meter. For meter proving procedures, refer to API Manual of Petroleum Measurement Standards, Chapters 4 and 12. Issue/Rev. 0.3 (8/02) MN02004 7

Section 3 Troubleshooting Guide For maximum accuracy, the meter must be recalibrated after each repair. Problem Probable Cause Corrective Action Overdelivery Dirty bearings. Clean. (or under registration) Worn bearings. Replace. Corrosion, erosion. Minor: Reprove meter. Extreme: Replace damaged parts. Bent blades. Replace rotor. Missing rotor blade. Replace rotor. Underdelivery Deposits (wax, for example). Clean. or over registration) Clogging of tube bundles. Remove and clean. Bent blades. Replace rotor. Cavitation. Higher back pressure. Entrained air or gas. Eliminate source or add air eliminator. Noise on transmission line. Add or check shielded cable. Eliminate source. Check that shield is grounded at instrument end only. Decrease sensitivity on instrument front end. Add suppression circuit to preamplifier. Non-Repeatability Damaged or missing blade. Replace rotor. Electrical noise interference. See Underdelivery. Damaged thrust washer. Replace. Deposits on the bearings. Clean or use slofted bearings. Plugged stator fin holes. Clean. Problems with the prover. Check prover operation. Leaks in the system. Check system for leaks. Meter Factor Shift Refer to Probable Causes of Refer to action to be taken for overdelivery and Overdelivery, Underdelivery. underdelivery. Output Without Flow Noise. See Underdelivery. Line vibration transmission Eliminate source of vibration. through pick-up coil. With pick-up coil only, desensitize coil by inserting a non-ferrous spacer(s)(0.030" increments) in pick-up coil boss. With preamp, refer to Preamplifier Bulletin SS02006 to desensitize. 8 MN02004 Issue/Rev. 0.3 (8/02)

Section 4 Maintenance and Calibration Record Maintenance and Calibration Record Meter No. Meter No. Serial No. Serial No. Product Product Flow Rate Flow Rate "K" Factor "K" Factor Date Proved By Date Proved By Remarks Remarks Meter No. Meter No. Serial No. Serial No. Product Product Flow Rate Flow Rate "K" Factor "K" Factor Date Proved By Date Proved By Remarks Remarks Issue/Rev. 0.3 (8/02) MN02004 9

Section 5 Related Publications The following literature can be obtained from FMC Technologies Measurement Solutions, Inc. Literature Fulfillment at johno@gohrs.com or online at www.fmctechnologies.com/measurementsolutions. When requesting literature from Literature Fulfillment, please reference the appropriate bulletin number and title. 4" Thru 24" Sentry Series Specifications... Bulletin SS02001 Flow Straightening Assemblies Specifications... Bulletin SS02007 Preamplifier PA-6 Specifications... Bulletin SS02012 Sentry Series Installation/Operation...Bulletin MN02003 Parts List (Form No.) 1 4"... P0586 6" Low Flow... P0587 6"... P0588 8"... P0589 10"... P0590 12"... P0591 16"... P0592 18"... P0593 20"... P0594 1 The latest edition is indicated by a 2-digit postscript (e.g.,.01,.02, etc.). Revisions included in MN02004 Issue/Rev. 0.3 (8/02): Pages 3-4 changed Preamplifier from PA-4 to PA-6. The specifications contained herein are subject to change without notice and any user of said specifications should verify from the manufacturer that the specifications are currently in effect. Otherwise, the manufacturer assumes no responsibility for the use of specifications which may have been changed and are no longer in effect. Headquarters: 1803 Gears Road, Houston, TX 77067 USA, Phone: 281/260-2190, Fax: 281/260-2191 Gas Measurement Products: Erie, PA USA Phone 814/898-5000 Thetford, England Phone (44) 1842-82-2900 Kongsberg, Norway Phone (47) 32/286-700 Buenos Aires, Argentina Phone 54 (11) 4312-4736 Integrated Measurement Systems: Corpus Christi, TX USA Phone 361/289-3400 Kongsberg, Norway Phone (47) 32/286-700 San Juan, Puerto Rico Phone 787/274-3760 United Arab Emirates, Dubai Phone 971 +4/331-3646 Visit our website at www.fmctechnologies.com Liquid Measurement Products: Erie, PA USA Phone 814/898-5000 Los Angeles, CA USA Phone 661/702-8660 Slough, England Phone (44) 1753-57-1515 Ellerbek, Germany Phone (49) 4101-3040 Barcelona, Spain Phone (34) 93/201-0989 Moscow, Russia Phone (7) 495/564-8705 Melbourne, Australia Phone (61) 3/9807-2818 Printed in U.S.A. 6/93 FMC Measurement Solutions. All rights reserved. MN02004 Issue/Rev. 0.3 (8/02) Beijing, China Phone (86) 10/6500-2251 Singapore Phone (65) 6861-3011 Chennai, India Phone (91) 44/450-4400