User manual ma Loop Calibrator Jofra macal

Similar documents
User manual Temperature Calibrator JOFRA CTC-140/320/650 A/B JOFRA CTC-1200 A

User Manual Compact Temperature Calibrator Jofra CTC-155/350/660 A/C

MTC series. Marine Temperature Calibrator. temperature. » Wide temperature range

HPC600. Handheld Pressure Calibrator. Built-in electric pump. pressure

MTC Series. Marine Temperature Calibrator

Reference manual Temperature Calibrator JOFRA MTC-140/320/650 A

Reference Manual Handheld Pressure Calibrator JOFRA HPC400

CTC-series. Compact Temperature Calibrator. temperature

User manual Temperature Calibrator JOFRA CTC-140/320/650 A/B JOFRA CTC-1200 A Copyright 2005 AMETEK DENMARK A/S ( UK)

PTC8001 Temperature Calibrator

Type T Pump. Operation Manual for T-1and T-1-CPF

DT33B/C/D OPERATOR S MANUAL LIMITED WARRANTY AND LIMITATION OF LIABILITY

SafetyPAT Plus A B P/N:

P014 Hydraulic Jack Pump Operation Manual for P014 and System E

WENS30C Loop Calibrator

Model Type T Hydraulic Deadweight Tester

362 Clamp Meter. Users Manual

1.Safe testing IMPORTANT:

WPS600. Hydraulic Pressure Transducer. User s Guide. 600 bar 60 MPa 8700 psi

MY /2-DIGIT DIGITAL MULTIMETER Users Manual

Instruction Manual & Specification

1662/1663/1664 FC. Electrical Installation Tester. Safety Information. Warnings

USER MANUAL. Insulation Tester + DMM. Model MG320

Table of Contents. Title Page UT501/UT502 OPERATING MANUAL. Introduction. Unpacking the Meter. Safety Information. International Electrical Symbols

P014 Hydraulic Jack Pump Operation Manual for P014 and System E

WPS bar, 60 MPa, 8700 psi Hydraulic Pressure Transducer. User s Guide

User- and Reference manual Temperature Calibrator JOFRA ETC-125 A / 400 A / 400 R

Calibrator Capabilities Function Range Resolution Measure V dc 28 V 1 mv Measure ma dc Source ma dc

51 & 52 Series II. Users Manual. Thermometer. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

362 Clamp Meter. Users Manual

User- and Reference manual Temperature Calibrator JOFRA ETC-125 A / 400 A / 400 R Copyright 2005 AMETEK DENMARK A/S ( UK)

707Ex. Users Manual. ma Calibrator

Thermocouple Thermometer Models

KMP Insulation-Continuity Tester. Users Manual

Model ATC-125 Advanced Temperature Calibrator

MARTINDALE INSTRUCTIONS ELITE FUSE FINDER KIT ELECTRIC. Trusted by professionals. 4.4 Storage Conditions

Digital Multimeter User's Manual

INSTRUCTION MANUAL DT173 DT175 DIGITAL THERMOMETERS ALWAYS READ THESE INSTRUCTIONS BEFORE PROCEEDING

EGT MONITOR AND PROBE KIT BG FUEL SYSTEMS McDonald Road Dahlonega, GA (706) A Edition 03

Barnant Company (Type J) (Type K) (Type T)

ASC300. Advanced Signal Calibrator. Process signal calibrator with superior accuracy. Specification Sheet

l The Battery Tester is designed for measuring the l AC four-terminal method to measure the internal

Thermistor Thermometer MODEL NO

Part No. 01/3955 July LS Series Contacting Extensometer. User Manual. LS Series Contacting Extensometer User Manual 1

USER S MANUAL G SERIES USER S MANUAL. Ref.49-MG000EN08

R&S RT-ZC02 R&S RT-ZC03 AC/DC Current Probe User Manual

EZYPAT / EZYPAT+ / SMARTPAT PORTABLE APPLIANCE TESTER INSTRUCTION MANUAL

KMP7010. Ground Resistance Tester. Users Manual

Chek-A-Cell TM TestEquipmentDepot.com. Instruction Manual Rev A 8/12

Item ref: UK & UK MTB01 & MTB02 DIGITAL MULTITESTERS. User Manual

Table of Contents Title Page No.

MARTINDALE MARTINDALE IN2001/IN2003 INSULATION & CONTINUITY TESTERS INSTRUCTION MANUAL IN2001/IN2003 INSULATION & CONTINUITY TESTERS

Portable pressure calibrator, Wally Box III Model CPH7600

MODEL 901 OPERATING INSTRUCTIONS

D Dräger Alcotest 3000 Breath Alcohol Measuring Device

Valor 1000W Series Instruction Manual. Serie Valor 1000W Manual de Instrucciones. Série Valor 1000W Guide de I utilisateur

Type T Hydraulic Deadweight Tester

Ulyser Component Maintenance Manual

LTX RF LEVEL SENSOR. Instruction Manual

ASTROAI USER MANUAL DIGITAL MULTIMETER. NOTE: Fully read and understand this manual before using this instrument.


Reference manual Compact Temperatur Calibrator Jofra CTC-140/320/650 A/B Jofra CTC-1200 A

THERMOCOUPLE CALIBRATOR Mode KM-CAL-710

RS-3 PRO RS-1007 PRO. CAT IV Analog Clamp meter Series. Users Manual. For detailed specifications and ordering info go to

Atlas ESR. User Guide. Capacitance and Equivalent Series Resistance Meter. Model ESR60 (Enhanced)

Model UT602/603: OPERATING MANUAL

Instruction Manual. IT-1 Infrared Thermometer

INSTRUCTION MANUAL CM95/ CM100 TRUE RMS FLEXIBLE CLAMP METER. 1. SAFETY INFORMATION: Always read before proceeding. REMEMBER: SAFETY IS NO ACCIDENT

Autoranging Multimeter

DIGITAL INSULATION RESISTANCE TESTER (5KV, 1T )

J O F R A. t e m p e r a t u r e. c a l i b r a t i o n. i n s t r u m e n t s

Instruction Manual. Coating Thickness Gauge. Optionally with Memory

INSP-3 Wiring Inspection Tester

365 Detachable Jaw True-rms Clamp Meter

Visit us at

IN1001/IN1003 ANALOGUE INSULATION & CONTINUITY TESTERS INSTRUCTION MANUAL IN1001/IN1003 ANALOGUE INSULATION & CONTINUITY TESTERS INSTRUCTION MANUAL

1 This instrument must only be used by a competent and trained person and operated in strict accordance with the instructions.

Model ATC-140/250 Advanced Temperature Calibrator

True RMS Autoranging Multimeter

Reference Manual Temperature Calibrator JOFRA ITC-155/320/650 A Copyright 2002 AMETEK DENMARK A/S

Vol-Con XL (61-086) Voltage/Continuity Tester and Vol-Test (61-085) Voltage Tester Instruction Manual WARNING

MX2040, DYNAMIC SIGNAL CHECKER Instruction Manual

Temperature Calibrator

USER MANUAL. 10KV Digital High Voltage Insulation Tester. Model MG500. Additional User Manual Translations available at

P-Series Screw Pump. Operation Manual for P016, P017, and P-018-CPF Pressure Pumps

The following table shows the symbols used on the product and/or in this manual. Hazardous Voltage. Risk of electric shock.

Dycon D1532SM. EN50131/PD6662 Grade 3, 12V 2A Power Supply. Technical Description Installation and Operating Manual DYCON POWER SOLUTIONS LTD

Manual Supplement. For IEC V, CAT I & 600, CAT II Meters only. Serial NO and greater.

BM kv Digital Insulation Tester USER MANUAL

4-20mA DC PROCESS CLAMP METER

Chek-A-Cell TM. Instruction Manual Rev A 8/12

WPS500X 500 psi /34.5 bar Automotive Pressure Transducer. User s Guide

INSTRUCTION MANUAL PORTABLE APPLIANCE TESTER

Reference manual Industrial Temperatur Calibrator Jofra ITC-155/320/650 A

OPERATION MANUAL MILLIAMP PROCESS CLAMPMETER MODEL - KM 071 LIST OF PRODUCTS

COMPONENT-TESTER INSTRUCTION MANUAL

DIGITAL RCD(ELCB) TESTER

(Type J) (Type K) (Type T)

DM01. Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05

INSTRUCTION MANUAL MARTINDALE CM52 CLAMP METER. ELECTRIC Trusted by professionals. Other products from Martindale:

Transcription:

User manual ma Loop Calibrator Jofra macal

User Manual macal ma loop Calibrator Copyright 2005 AMETEK DENMARK A/S

LIST OF CONTENTS 1. INTRODUCTION 1.1 PACKING LIST... 2 2. SAFETY INSTRUCTIONS... 2 3. OPERATION... 4 3.1 DISPLAY AND KEYBOARD... 4 3.2 SWITCHING THE MACAL ON/OFF... 6 3.3 HOW THE CURRENT IS GENERATED... 6 3.4 HOW THE CURRENT IS MEASURED... 7 3.5 HOW THE VOLTAGE IS MEASURED... 7 3.6 HOW NEW PARAMETERS ARE SELECTED (SETUP)... 8 3.6.1 SELECTION OF MODE... 8 3.6.2 SELECTION OF SLOPE TIME... 8 3.6.3 SELECTION OF STEP SIZE... 9 3.6.4 SELECTION OF STEP TIME... 9 3.6.5 SELECTION OR DE-SELECTION OF AUTOMATIC OFF FUNCTION 9 4. ERROR SITUATIONS... 10 4.1 TEST FOR OPEN LOOP... 10 4.2 TEST FOR INCORRECT POLARITY... 10 4.3 TEST FOR OVERRANGE... 10 4.4 INDICATION OF BATTERY STATUS... 11 5. HOW macal IS ADJUSTED (service mode)... 12 6. TECHNICAL SPECIFICATIONS (23 C±3 C)... 14 103887 02 1

1. INTRODUCTION The macal is a loop calibrator for measuring and generating ma. The macal is specially designed for use in industrial plants with 4-20mA signals. The macal can be operated with one hand, and operation is simple and intuitive. 1.1 PACKING LIST When the instrument is received, the box should contain the following: macal Bag 2 test leads with crocodile clip 4 batteries, type LR6, AA 1 USB key containing documentation 2. SAFETY INSTRUCTIONS Read this manual carefully before using the instrument! In order to avoid any personal injuries and/or damage to the instrument all safety instructions and warnings must be observed. Disposal WEEE Directive These calibrators contain Electrical and Electronic circuits and must be recycled or disposed of properly (in accordance with the WEEE Directive 2012/19/EU). WARNING: The calibrator must not be used for any purposes other than those described in this manual. Follow all equipment safety procedures. The calibrator has been designed for interior use only and should not be used in hazardous areas, where vapour or gas leaks, etc. may constitute a danger of explosion. Do not apply more than the rated voltage. See specifications for supported ranges. Do not use the calibrator if it is damaged. Before you use the calibrator, inspect the case. Look for cracks or missing plastic. Pay 2 103887 02

particular attention to the insulation surrounding the connectors. Select the proper function and range for your measurement. Make sure the battery cover is closed and latched before you operate the calibrator. Remove test leads from the calibrator before you open the battery door. Inspect the test leads for damaged insulation or exposed metal. Check test leads continuity. Replace damaged test leads before you use the calibrator. When using the probes, keep your fingers away from the probe contacts. Keep your fingers behind the finger guards on the probes. Do not use the calibrator if it operates abnormally. Protection may be impaired. When in doubt, have the calibrator serviced. Disconnect test leads before changing to another measure or source function. To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator appears. To avoid personal injury or damage to the calibrator, use only the specified replacement parts and do not allow water into the case. Caution Disconnect the power and discharge all high-voltage capacitors before testing resistance or continuity. Use the proper jacks, function and range for your measurement or sourcing application. To avoid damaging the plastic lens and case, do not use solvents or abrasive cleansers. 103887 02 3

3. OPERATION 3.1 DISPLAY AND KEYBOARD The various keys on the keyboard have the following functions (cf. Fig. 1 on the cover of the user s guide): ITEM KEYBOARD Description: ON/OFF Switching instrument on/off. % ma / SETUP Alternating between readout in ma and %. If the key is held down for more than 2 sec., a change to SETUP takes place. Under SETUP this key is used to select next parameter. AUTO AUTO ramp or AUTO-STEP function is started or stopped. DOWN ARROW Less key. If the key is held down, the key press is repeated. UP ARROW More key. If the key is held down, the key press is repeated. STEP The STEP function is activated or de-activated. FUNC. (FUNCTION) Change-over takes place between SOURCE ma, MEASURE ma and MEASURE Volt function. PWR. (POWER) Change-over takes place between EXT. (the +24VDC of the current loop is generated by an external power supply) and INT. (the +24VDC of the current loop is generated by the macal itself). The various segments of the display indicate the following (cf. Fig. 1 on the cover of the user s guide): ITEM DISPLAY SEGMENT Description: Shows numerical value in ma, %, V or sec., and for readout of error codes etc. EXT (EXTERNAL) External loop supply is selected. INT (INTERNAL) Internal loop supply is selected, the internal +24VDC of the macal being used to supply the current loop with power. MEASURE Measurement of current or voltage is selected (indicated by unit). SOURCE Generation of current is selected. STEP The STEP function is active. When the segment flashes, this indicates that the AUTO- STEP function is activated. %, s, V or ma Indication of unit for readout numerical values. 4 103887 02

ITEM DISPLAY SEGMENT Description: LOW BAT (LOW BATTERY) When this segment is lit, the voltage in the batteries is less than 3.5 V, and they should be replaced. When the segment flashes, the voltage is less than 3.1 V and the instrument cannot be operated. 0-20 0-20mA is selected as the range, which affects the readout in %, as well as STEP values. 4-20 4-20mA is selected as the range, which affects the readout in %, as well as STEP values FLOW VALVE FLOW transmitters are selected, which affects the readout in %, as well as STEP values. Proportional transmitters are selected, which affects STEP values and the AUTO-STEP function. ITEM CONNECTIONS Description: - Connection of black test lead - + Connection of red test lead + Connectio230mA.n to mains adaptor 6V DC, 230mA. Note : The mains adapter does not charge rechargeable batteries. 103887 02 5

3.2 SWITCHING THE macal ON/OFF Press and the unit is switched on/off. When the unit is switched on, the present battery voltage appears in the display for approx. 1 second. Fit the red test lead into the connector marked +. Fit the black lead into the connector marked -. The macal is now ready for use. The unit automatically starts up in the function (SOURCE ma, MEASURE ma or MEASURE Volt) which was active when it was last switched off. 3.3 HOW THE CURRENT IS GENERATED Connect the macal to the instrument to be tested as shown in Fig. 3 or Fig. 4 on the cover of the user s guide and switch the unit on. Press until the display shows SOURCE ma or SOURCE%. Press for selection of INT or EXT (the present selection is shown in the display). If INT. is selected, the +24VDC of the current loop is generated by the macal itself. If EXT is selected, the +24VDC of the current loop is generated by an external power supply. Press for selection of readout in ma or %. The display has the following appearance when 4mA is generated with internal loop supply: Press and the generated current is increased. Press and the generated current is decreased. Press for activation of pre-defined current values (values depend on selection of MODE and STEP SIZE). 6 103887 02

Press for automatic changing of current value. If STEP is selected (the AUTO-STEP function), change-over takes place automatically between the pre-defined current values. For proportional transmitters (VALVE), change-over takes place between 3.8-4 - 4.2-4... and 19-20 - 21-20... ma (the range is selected using the up arrow or down arrow keys). The macal now generates the current shown in the display. 3.4 HOW THE CURRENT IS MEASURED Connect the macal to the transmitter to be tested as shown in Fig. 5 or Fig. 6 on the cover of the user s guide and switch the unit on. Press until the display shows MEASURE ma or MEASURE % Press for selection of INT or EXT (the present selection is shown in the display). If INT is selected, the +24VDC of the current loop is generated by the macal itself. If EXT is selected, the +24VDC of the current loop is generated by an external power supply. Press for selection of readout in ma or %. The display thus has the following appearance in the measurement of 20mA in the current loop. The current is printed out with a resolution of 0.001mA: The macal now measures the current shown in the display. 3.5 HOW THE VOLTAGE IS MEASURED Connect the macal to the instrument to be tested as shown in Fig. 7 on the cover of the user s guide and switch on the unit. Press until the display shows MEASURE V. Readout in % cannot be selected. The display thus has the following appearance in the measurement in V. The voltage is printed out with a resolution of 0.1Volt: 103887 02 7

The macal now measures the voltage shown in the display. 3.6 HOW NEW PARAMETERS ARE SELECTED (SETUP) Press for approx. 2 seconds. The present selection of MODE is now shown in the display, e.g. ma. Press to cancel a selection. SETUP is exited and it is necessary to start all over again. This option can be selected at any time. 3.6.1 Selection of MODE It is possible to choose between 5 different MODES: 4-20mA LIN 0-20mA LIN 4-20mA FLOW 0-20mA FLOW 4-20mA VALVE LIN - % calculated linearly. FLOW - % calculated logarithmically. VALVE - % not calculated. AUTO-STEP function 3.8-4 - 4.2-4... or 19-20 - 21-20... ma. Press or for selection of new MODE. Press to save the present selection. 3.6.2 Selection of SLOPE TIME After MODE has been selected, SLOP TIME (ramp time in the case of AUTO) can be changed. The text s is shown on the display for approx. 1 second. Press or for setting of SLOPE TIME. The time can be adjusted between 30-999 S. Press to save the present selection. 8 103887 02

3.6.3 Selection of STEP SIZE After SLOPE TIME has been selected, STEP SIZE (step size in % of range) can be selected for the transmitter types LIN and FLOW. If MODE 4-20mA VALVE is selected, a fixed STEP SIZE is applied. The text is shown in the display for approx. 1 second. Press or for setting of STEP SIZE. A choice can be made between the step sizes 5%, 10%, 20%, 25%, 50% and 100%. Press to save the present selection. 3.6.4 Selection of STEP TIME After STEP SIZE has been selected, STEP TIME (the time between changes in the AUTO-STEP function) can be changed. Either s or s is shown in the display. Press or for setting of STEP TIME. Press to save the present selection. 3.6.5 Selection or de-selection of automatic OFF function After STEP TIME has been selected, it is possible to select or de-select the AUTOMATIC OFF function (macal automatically switched off approx. 30 minutes after a key has last been activated). The text is shown in the display for approx. 1 second. Alongside, the text or appears. Press and AUTOMATIC OFF is selected. appears in the display. Press and AUTOMATIC OFF is deselected. appears in the display. Press to save the present selection. SETUP is exited and the macal is now ready for normal operation with new parameters. 103887 02 9

4. ERROR SITUATIONS 4.1 TEST FOR OPEN LOOP The text CAUSE REMEDY flashes in the display: The unit is unable to generate the desired current. The current loop is probably open or selection of PWR is incorrect. Check whether instrument is connected correctly. Check that PWR selection is correct. 4.2 TEST FOR INCORRECT POLARITY The text flashes in the display: CAUSE The unit measures a negative current. REMEDY The test leads + and - are probably to be changed over. 4.3 TEST FOR OVERRANGE The text CAUSE REMEDY flashes in the display: The unit measures a current which is greater than 24mA or a voltage which is greater than +45Volts or less than -45Volts. Measure only voltages in the range -45 to +45 Volts. Measure only in current loops, where the current is less than 24mA. Check that the instrument is connected properly. See Fig. 3 to Fig. 7 on the cover of the user s guide. 10 103887 02

4.4 INDICATION OF BATTERY STATUS The text LOW BAT lights up in the display: The text LOW BAT flashes in the display: CAUSE LOW BAT lights up: The batteries are about to be used up. However, the instrument can still be used. LOW BAT flashes: REMEDY The batteries are used up. The instrument cannot be used. The display goes out. However, the keyed-in values are retained. LOW BAT lights up: Change the batteries. See Fig. 2 on the cover of the user s guide. LOW BAT flashes: The batteries must be changed. See Fig. 2 on the cover of the user s guide. The 4 used LR6, AA batteries are removed and new ones inserted. 103887 02 11

5. HOW macal IS ADJUSTED (service mode) Connect the macal to an external precision instrument (with an accuracy higher than 0.02% of F.S.) as shown in Fig. 3 on the cover of the user s guide. Hold down and press 3 images appear in the display: The first display image shows the software version. After approx. 1 second, the serial number of the pcb is shown. After another approx. 1 second, the text is shown (code must be keyed in). Press to exit service mode. This option can be selected at any time. Key in the following code: Press 3 times. Press 5 times. Press Twice. Press Once. The following display image appears: Press or until the display shows the same value as that measured with the external precision instrument. Press to approve the value. 12 103887 02

The following display image appears: Press or until the display shows the same value as is measured with the external precision instrument. Press to approve the value. The macal now carries out a self-calibration and adapts its circuits to that measured. Switch on the macal. It is now adjusted and ready for use. 103887 02 13

6. TECHNICAL SPECIFICATIONS (23 C±3 C) Current generation : Range : 0-24mA Resolution : 0.01mA Accuracy (k=2) : 0.05% F.S. Max. load : 900 Ohm Current measurement : Range : 0-24mA Resolution : 0.001mA Accuracy (k=2) : 0.05% F.S. Input impedance : 10 Ohm Voltage measurement : Range : OV to +45V Resolution : 0.1V Accuracy : 0.5V Operation temperature : -10 C to +40 C Storage temperature : -20 C to +50 C Electromagnetic : Designed for use in basic environment electromagnetic environment as defined in EN61326-1 : 2013. Battery life : 200 hours with external loop pwr. 20 hours with 12mA output/internal loop pwr. Battery types : 4 types LR6, AA, 1.5V Alkaline Mains adapter : 103950, 6V DC/230mA EURO with 1.3 mm DC female connector 103957, 6V DC/230mA UK, with 1.3mm DC female connector 103964, 6V DC/240mA US/JAPAN with 1.3 mm DC female connector Dimensions l x w x h : 174 x 66 x 26mm Weight (incl. batteries, bag and test leads) : 355 gr. 14 103887 02

Standard complied with Warrenty : (2014/30/EU), EN61326-1 : 2013 Electrical equipment for measurement, control and laboratory use EMC requirements According to current terms of sale and Delivery. The guarantee only applies to factory faults and ceases to apply if the instrument has been subjected to unauthorised interference and/or incorrect operation. 103887 02 15

AMETEK Sensors, Test & Calibration A business unit of AMETEK Measurement & Calibration Technologies Division offering the following industry leading brands for test and calibration instrumentation. JOFRA Calibration Instruments Temperature Calibrators Portable dry-block calibrators, precision thermometers and liquid baths. Temperature sensors for industrial and marine use. Pressure Calibrators Convenient electronic systems ranging from -25 mbar to 1000 bar - fully temperaturecompensated for problem-free and accurate field use. Signal Instruments Process signal measurement and simulation for easy control loop calibration and measurement tasks. M&G Deadweight Testers & Pumps Pneumatic floating-ball or hydraulic piston dead weight testers with accuracies to 0.015% of reading. Pressure generators delivering up to 1,000 bar. Crystal Pressure Digital pressure gauges and calibrators that are accurate, easy-to-use and reliable. Designed for use in the harshest environments; most products carry an IS, IP67 and DNV rating. Lloyd Materials Testing Materials testing machines and software that guarantees expert materials testing solutions. Also covering Texture Analysers to perform rapid, general food testing and detailed texture analysis on a diverse range of foods and cosmetics. Davenport Polymer Test Equipment Allows measurement and characterization of moisture-sensitive PET polymers and polymer density. Chatillon Force Measurement The hand held force gauges and motorized testers have earned their reputation for quality, reliability and accuracy and they represent the de facto standard for force measurement. Newage Hardness Testing Hardness testers, durometers, optical systems and software for data acquisition and analysis. United Kingdom Tel +44 (0)1243 833 302 caluk.sales@ametek.com France Tel +33 (0)1 30 68 89 40 general.lloyd-instruments@ametek.fr Germany Tel +49 (0)2159 9136 510 info.mct-de@ametek.de Denmark Tel +45 4816 8000 jofra@ametek.com USA Florida Tel +1 (800) 527 9999 cal.info@ametek.com California Tel +1 (800) 444 1850 crystal@ametek.com India Tel +91 22 2836 4750 jofra@ametek.com www.ametekcalibration.com Singapore Tel +65 6484 2388 jofra@ametek.com China Shanghai Tel +86 21 5868 5111 Beijing Tel +86 10 8526 2111 jofra.sales@ametek.com.cn Information in this document is subject to change without notice. 2016, by AMETEK, Inc., www.ametek.com. All rights reserved.