Qualification of a precision pattern dispenser

Similar documents
ISO INTERNATIONAL STANDARD

ISO 659 INTERNATIONAL STANDARD. Oilseeds Determination of oil content (Reference method)

Agilent 7693A Automated Liquid Sampler

Dispensette. Testing Instructions (SOP) 1. Introduction. May 2009

National comparison on verification of fuel dispensers

ISO 3679 INTERNATIONAL STANDARD. Determination of flash point Rapid equilibrium closed cup method

The Importance of Liquid Handling Details and Their Impact on your Assays

M-11 MARINE DIESEL ENGINE FUEL INJECTOR

REGULATORY CONTROL OF NUCLEAR FUEL AND CONTROL RODS

Metallic materials Rockwell hardness test. Part 2: Verification and calibration of testing machines and indenters

COMMISSION REGULATION (EU)

ISO INTERNATIONAL STANDARD. Road vehicles Brake lining friction materials Friction behaviour assessment for automotive brake systems

ISO INTERNATIONAL STANDARD

Dilutor Module for TriPlus RSH

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD. Road vehicles Brake lining friction materials Product definition and quality assurance

UK Weighing Federation Technical Articles

EUROPEAN COMMISSION ENTERPRISE AND INDUSTRY DIRECTORATE-GENERAL

Technical Procedure for Gas Chromatography (GC-FID)

ISO INTERNATIONAL STANDARD. Measurement of noise emitted by accelerating road vehicles Engineering method Part 2: L category

AKVdiesel. Automated kinematic Viscometer key features. Fully complies with ASTM D445. Two measuring positions. Remove cap and pour in sample

Agilent 7693A Automated Liquid Sampler

Notification of a Proposal to issue a Certification Memorandum. Rotor Drive System Gearbox TBO Development

Guideline No.M-10(201510) M-10 Marine Diesel Engine Fuel Injection Pump

DRVRMDMOTL Motorized Linear Drive. DRVMDMOTLR Motorized Linear and Rotary Drive. Instruction Manual

PTV TECHNICAL PRESCRIPTIONS. ELASTOMERIC SEALS: Part 1: Vulcanized Rubber FOR. Impartial institute for the monitoring of construction products

WNTE: A regulatory tool for the EU? GRPE Meeting of the Off-Cycle Emissions Working Group. Geneva, June 2006

TSHR 7000 series TN-TS-TX analyzer

Standard Operating Procedure for Detecting MEA Triazine in a Crude Oil Matrix using LONESTAR TM

Part 1: Basic requirements

ISO INTERNATIONAL STANDARD

EN. SKF Grease Spraying System For Large Open Gears. Extend the lifetime of large gear drives

TRANSFORMER SERVICE. ABB Ability inspection for transformers TXplore Oil-filled transformer internal inspection service

Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011

Alliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE

TSHR 6000 series TN/TS/TX analyzer

ISO 1217 INTERNATIONAL STANDARD. Displacement compressors Acceptance tests. Compresseurs volumétriques Essais de réception. Fourth edition

ISO INTERNATIONAL STANDARD. Diesel engines NOx reduction agent AUS 32 Part 1: Quality requirements

EUROPEAN UNION AGENCY FOR RAILWAYS Annex 1 006REC1025 V 1.0

ASTM B117 Testing Quality Control

Implementation procedure for certification and continued airworthiness of Beriev Be-200E and Be-200ES-E

Evaluating and Determining Precision and Accuracy for Nanoliter Volumes of Viscous Solutions on an Automated Liquid Handling Instrument

Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS)

Analysis of WLTP typical driving conditions that affect nonexhaust particle emissions

Thermo Scientific Finnpipette Performance Specifications

5.1 Design and Install Dispenser Transfer Systems Performance Based Skill Assessments 2019

TDG-F-113 CEC New Test Development Proposal for a New Engine Fuels Test Procedure

ISO INTERNATIONAL STANDARD. Tyres Electrical resistance Test method for measuring electrical resistance of tyres on a test rig

Certification Memorandum

Chemical decontamination in nuclear systems radiation protection issues during planning and realization

Notification of a Proposal to issue a Certification Memorandum. Approved Model List Changes

European Tyre and Rim Technical Organisation RETREADED TYRES IMPACT OF CASING AND RETREADING PROCESS ON RETREADED TYRES LABELLED PERFORMANCES

Thermo Scientific Matrix Manual Pipette User Manual

breakthrough versatility

AVL Particle Measurement System Aviation

2 nd ICCT Workshop on Marine Black Carbon Emissions 16/17 Sept, Utrecht. Ralf Oldenburg, Peter Lauer, MAN Diesel & Turbo

TESTING THE UNIFORMITY OF SPRAY DISTRIBUTION UNDER DIFFERENT APPLICATION PARAMETERS

VERITY Purification Systems Build the perfect system to fit your research

Testing of particulate emissions from positive ignition vehicles with direct fuel injection system. Technical Report

Certification Memorandum. Approved Model List Changes

The Analysis of Biodiesel for Inorganic Contaminants, including Sulfur, by ICP-OES

ISO INTERNATIONAL STANDARD. Compressed air Part 5: Test methods for oil vapour and organic solvent content

ISO INTERNATIONAL STANDARD. Diesel engines NOx reduction agent AUS 32 Part 1: Quality requirements

Stability, Linearity and Repeatability of Nitrogen Determination by Flash Combustion using Argon as Carrier Gas

ISO 9129 INTERNATIONAL STANDARD. Motorcycles Measurement methods for moments of inertia. Motocycles Méthodes de mesure des moments d'inertie

MiniLab EL Series for High Performance Engine Development and Maintenance

EUROPEAN NEW CAR ASSESSMENT PROGRAMME (Euro NCAP) CAR SPECIFICATION, SPONSORSHIP, TESTING AND RETESTING PROTOCOL

Perfect Liquid Handling PRECISE RESULTS WITH INNOVATIVE TECHNOLOGY

Infinite F50 family Reliable. Validated. Compact.

CEMS FOR COAL FIRED POWER STATIONS DEVICE CALIBRATION AND MAINTENANCE

ABB MEASUREMENT & ANALYTICS. Predictive Emission Monitoring Systems The new approach for monitoring emissions from industry

Fibertec 8000 Fully automated Crude and Detergent Fibre analysis

MIXPAC TM Technical Services Your partner for successful solutions. Sulzer Chemtech

WELMEC European cooperation in legal metrology

This document is a preview generated by EVS

ISO INTERNATIONAL STANDARD. Tools for moulding Tool specification sheet for diecasting dies

PART XV Metering Systems SECTION 2 GUIDE TO GAS METERING SYSTEMS

PrioVino Premier. Translation of Original Operating Instructions. Status: August First edition January 2018 / PrioVino GmbH

ISO INTERNATIONAL STANDARD

Bruker BioSpin MATCH. Multiple Adjustable Tube Clamp Holder User Manual - Version 005. NMR Solutions. think forward

POSITION PAPER Version 3.0

Certificate: _01 / 29 April 2014

RECYCLABILITY EVALUATION PROTOCOL FOR PE FILMS

COMMISSION OF THE EUROPEAN COMMUNITIES REPORT FROM THE COMMISSION. Quality of petrol and diesel fuel used for road transport in the European Union

S Pyatt. UV glue dispensing studies on the Dima Dotmaster

Performance of Batteries in Grid Connected Energy Storage Systems. June 2018

BACS APPROVED BUREAU SCHEME SUPPORT GUIDELINES

ISO INTERNATIONAL STANDARD

Renewable Fuels Association One Massachusetts Ave. NW Suite 820 Washington, DC Telephone: (202) Fax: (202)

ANNEX MOTOR VEHICLES AND MOTOR VEHICLES' PARTS. Article 1. General Provisions

ISO INTERNATIONAL STANDARD

MB3600-CH30 Laboratory FT-NIR analyzer for biodiesel applications Suitable for production optimization and product quality assessment

Hydrogen Station Equipment Performance Device (HyStEP Device) Specification

REAL WORLD DRIVING. Fuel Efficiency & Emissions Testing. Prepared for the Australian Automobile Association

Design Modification and Optimization of Trolley in an Off-Bearer Mechanism Present In Concrete Block Making Machines

Adam Equipment LHS SERIES. (P.N , Revision A March 2012)

GMD 13 New gravimetric measuring system with evaluation on location

Determination of the Rate of Ignition of Nitrocellulose by Resistance Wire for the Igniter of KV 150 M2

TIER 3 MOTOR VEHICLE FUEL STANDARDS FOR DENATURED FUEL ETHANOL

Note: it is a criminal offence to give false information in this application.

Transcription:

Qualification of a precision pattern dispenser Raf Van Ammel, Katarzyna Sobiech-Matura, Uwe Wätjen 2016 EUR 28268 EN

This publication is a Technical report by the Joint Research Centre, the European Commission s in-house science service. It aims to provide evidence-based scientific support to the European policy-making process. The scientific output expressed does not imply a policy position of the European Commission. Neither the European Commission nor any person acting on behalf of the Commission is responsible for the use which might be made of this publication. Contact information Name: Raf Van Ammel Address: European Commission, Joint Research Centre, Directorate for Nuclear Safety and Security E-mail: raf.van-ammel@ec.europa.eu Tel.: +32 14 571267 JRC Science Hub https://ec.europa.eu/jrc JRC104321 EUR 28268 EN ISBN 978-92-79-64110-7 ISSN 1831-9424 doi:10.2789/1167 European Atomic Energy Community, 2016 Reproduction is authorised provided the source is acknowledged. All images European Atomic Energy Community 2016 How to cite: Van Ammel Raf, Sobiech-Matura Katarzyna, Wätjen Uwe; Qualification of a precision pattern dispenser; EUR 28268 EN; doi:10.2789/1167

Table of contents Abstract... 2 1. Introduction... 3 1.1 General description of the instrument... 4 1.2 Details regarding the construction of the robot... 5 2. Qualification procedure... 9 2.1 Results of the qualification procedure... 9 2.2 Results for air filters preparation... 17 3. Conclusion... 18 References... 19 List of abbreviations and definitions... 20 List of figures... 21 List of tables... 22 Annex 1... 23

Abstract The European Commission Joint Research Centre in Geel (JRC-Geel) is on a regular basis preparing radioactive sources. These sources can be used as calibration standards or reference sources in different applications, e.g. proficiency testing of laboratories monitoring radioactivity in the environment organized by JRC-Geel. In order to automate the sample preparation process a precision pattern dispenser was designed to reproducibly dispense radioactive solutions. The set-up should be able to dispense pre-set amounts of stock solution in an automated and reproducible way in predefined positions or patterns. Correct dispensing should be ensured by defining the positions of ampoules or special source holders precisely. The set-up must also be able to dispense pre-set amount of liquid according to designed patterns on air filters having a maximum size of 60x60 cm. During all manipulations no evaporation of stock solution can be afforded in order not to compromise its concentration. The liquids to be handled are mainly acidic solutions so all parts must be resistant to the acidic environment. The number of instrument parts that come in contact with the radioactive solution must be kept to a minimum to limit the amount of radioactive waste and should be easily replaceable. The precision pattern dispenser is composed of a commercially available sample handling unit that makes use of syringes to handle the liquid. This unit is coupled to an XYZ-table where e.g. air filters can be fixed on. A custom-made interface (protocol generator software) is used to design dispensing patterns on filters of different sizes and shapes. The combination of the filter and the pattern, called protocol, is sent to the software steering the sampling unit and the XYZ-table. The mass or volume to be dispensed and the position where it has to be dispensed is easily introducible and made visible in the protocol generator. Qualification tests were conducted to evaluate the performance of the instrument and to assess the compliance with the requirements. The two most important parameters to be tested were trueness and precision. The maximum values according to the specifications were 2 % for trueness and 1 % for precision. Tests were performed by dispensing water into glass ampoules. All other parameters described above were examined and found to be compliant with the specifications. The results of the tests demonstrate that the precision pattern dispenser can dispense liquid within the required specifications. The instrument can be used to spike air filters with an almost homogenous distribution of activity within a predefined area. The precision pattern dispenser makes it possible to prepare sources for different applications in a fast and accurate way. 2

1. Introduction The European Commission Joint Research Centre (JRC) is on a regular basis preparing radioactive sources that are used as calibration standards or reference sources for different applications. It is also organising interlaboratory comparisons for laboratories monitoring radioactivity in the environment. For these purposes samples have to be prepared in a reproducible way. In order to make this process more efficient and accurate an instrument to dispense radioactive solution, called precision pattern dispenser, was installed at the JRC-Geel. In a first stage the basic requirements were specified. The most important requirement was that the instrument must be able to accurately dispense predefined amounts of radioactive solutions in an automated way at certain predefined positions. These known amounts can be dispensed on different sizes and shapes of blank filters or in ampoules or costum made source holders. In this current study the results of the performance tests of the instrument are presented. 3

1.1 General description of the instrument A proposal was made by the company SampleQ [1] proposing a precision pattern dispenser. The precision pattern dispenser consists of two major parts, a commercially available liquid handling unit and a XYZ table. The liquid handling unit is steered by software and can dispense known amounts (introduced as volumes or masses) of liquids in objects that are suspended on the arm of the handling unit. As one of the applications will also be the spiking of air filters, an XYZ table is coupled to the dispensing unit. On this table, air filters can be fixed. The dispensing unit can also dispense pre-set volumes of solution on predefined positions on the fixed air filters. The software in which the combination of positions and volumes, called patterns, can be designed, converts the patterns into a protocol that is steered to the dispensing robot and the XYZ table at the same time. The XYZ table is able to support filters of a maximum size of 60 by 60 cm. The software allows creating patterns within this area. 4

1.2 Details regarding the construction of the robot The precision pattern dispenser is a modular instrument built by SampleQ, Belgium [1]. The core of the instrument is a liquid handling unit that is performing all dispensings. It is a commercially available automatic dispensing system PAL HTC-xt (CTC Analytics AG, Switzerland [2]). Figure 1: Ampoules filling station of the automatic dispensing robot. The dispensing system can dispense pre-set amounts of liquid in an automated and reproducible way. The liquids to be handled are mainly dilute acidic solutions of hydrochloric or nitric acid. The robot has a movable tower in which syringes of different volumes can be mounted. The movements of the tower containing the syringes, are executed in a smooth way to avoid any spillage of liquid to be dispensed during the movements. 5

Figure 2: Syringe positioned inside the tower. The volume of the mounted syringes can vary from 25 µl to 5 ml, in order to be able to dispense any random volume between 20 µl and 5 ml. The amount of liquid to be dispensed at once can be chosen within this volume range. The volume to be dispensed can be easily set in the Chronos software (LEAP Technologies, USA), which is controlling the movements of the tower and the syringe. The tower of the dispensing robot can also move over different modules that are suspended on the main horizontal arm. These items can include: solvent reservoirs (100 ml), waste and wash vials (20 ml), source trays in which different type of ampoules and other source holders can be mounted. In the current set-up 2 holders for the 3 solvent reservoirs having a volume of 100mL each can be suspended. In this way a maximum volume of 600 ml of solvent can be contained. All solvent containers used are closed with septa to prevent evaporation before, during and after dispensing. These septa can be easily replaced if deemed necessary. Figure 3: Solvent reservoirs and wash and waste vials with closed septa. 6

The source trays are custom-made. They fit in the tray support which is suspended on the horizontal arm. In order to prevent them from changing their positions during the dispensing process the fitting of the support trays is made very tight. A chosen type of source holder can be placed in predefined positions in the source tray. The exact positions of all the items suspended on the arm, as well as the positions of the source holders in the source tray can be programmed in the PAL HTC-xt controller panel. In this controller a lot of other parameters influencing the dispensing process can be set, e.g. the speed of the arm, the ejection speed of the liquid, the aspiration speed of the liquid etc. Some of these settings are changed according to the syringe that is used.. Figure 4: Custom-made tray for ampoules. The Chronos software controls all the movements of the tower between the different items suspended on the arm as well as the movements of the plunger of the syringe it contains. The Chronos software can read in the settings from the controller. Some of the default settings set in the controller can still be adjusted in order to dispense the liquid in a smooth and accurate way. During dispensing only the solvent and waste reservoirs and the syringes come in contact with the solution to be dispensed. All these items are resistant to acidic solutions. In this way only these parts need to be replaced if the solution to be dispensed is changed, keeping the amount of radioactive waste generated to a minimum. Another advantage of this method of dispensing is the minimisation of dead volume. Only the needle of the syringe contains a dead volume resulting in only a small amount of liquid waste generated. The XYZ table serves as a movable platform to support and fix filters on which solution can be dispensed. In order to support the filters well, metallic strings are stretched over the whole area of the table every 5 cm. The filter can be fixed and stretched using metallic bars of different length, which in turn are clamped on the magnetic bars positioned on the outer side of the filter. The filter can be centred on the table using the marks on the bottom of the table. Under the metallic strings an adsorbing paper can be introduced to collect eventual spilling. 7

Figure 5: The XYZ-table. The maximum deviation between the set position and the executed position of the dispensing is 2 mm in every direction. This is easily obtained by the set-up. All manipulations are performed in such a way that the certified values of the standard solution are not compromised in any way and that no spillage or other forms of (cross-) contamination occur during the whole dispensing process. The whole set up, i.e. the dispensing robot and the XYZ table, is placed in a protective Plexiglas enclosure. This protects the set-up from dust, but keeps it visible. The enclosure can be opened from the front so that all parts are accessible when items like syringes, liquid containers, source holders or filters need to be changed or when service is needed. Figure 6: View on the opened plexiglass enclosure. All relevant data of instrument settings dispensing positions and volumes etc. used during the dispensing are stored electronically and can be easily retrieved. 8

2. Qualification procedure The properties described in the previous paragraph were the subject of the qualification. Two additional parameters were examined being the precision under repeatability conditions, hereafter called repeatability, and trueness in this case the deviation of the actually dispensed amount of liquid from the pre-set amount programmed according to ISO 5725-1 [4]. For both criteria upper limits were set in the tender specifications. The repeatability for any random dispensed volume has to be better than 1 %, while trueness, the bias between the set volume and test results, has to be better than 2 %. Both criteria were examined using syringes of 50, 100, 250, 500 µl and 1 and 5 ml. For each syringe, tests were carried out by dispensing 20, 40, 60 and 80 % of its total volume. For each volume the pre-set volume of water was dispensed 10 times into tarred 5 ml glass ampoules. After dispensing in these 10 ampoules, they were weighed again. The difference in mass gave the amount of liquid dispensed. This mass was corrected for buoyancy taking into account the laboratory temperature, humidity and air pressure. By dividing the calculated mass by the density of the water at the lab temperature (correction of 0.2%) [5], the dispensed volume was obtained. The trueness was calculated as the difference between the average volume obtained from the results of the 10 dispensings and the pre-set volume. The repeatability was calculated as a standard deviation of the 10 dispensed volumes. 2.1 Results of the qualification procedure Dispensing of liquid into glass ampoules was tested. A new set of 10 ampoules of 5 ml was used each time. The ampoules were filled with water using six different syringes (50 µl to 5 ml). The amount of liquid was ranging from 20 µl to 4 ml. A summary of the results is displayed in Table 1 and details of the dispensing are given in Figures 7 to 29. The raw data are added in Annex 1. All 23 tests resulted in a repeatability better than 1%, 22 of them even better than 0.5%. As the instrument will be used mainly under repeatability conditions, the results enhance the trust in the dispensing capabilities. For the trueness assessment, 22 of 23 tests gave results better than 2%. Only the dispensing of 20 µl using a 100 µl syringe showed a value above 2%. This is probably due to the fact that this is close to the edge of the dynamic range of the syringe in combination with the small amount that has to be dispensed. If a dispense volume of 20µL with a trueness better than 2% is necessary, a 50 µl syringe can be used. In general, the volume dispensed is lower than the set volume. To reduce the bias, one can calibrate the syringe with water and in the software a correction factor can be introduced. The correction factor can be applied to the volumes to be dispensed and should correct for the systematic bias. 9

Table 1: Summary of the qualification results Syringe volume (µl) Dispensed volume (µl) Trueness (%) repeatability (%) 50 20 1.11 0.45 30 0.97 0.30 40 0.61 0.18 100 20 3.33 0.88 40 1.57 0.35 60 1.32 0.21 80 1.02 0.16 250 50 1.35 0.38 100 0.64 0.08 150 0.43 0.08 200 0.35 0.05 500 100 0.92 0.12 200 0.25 0.12 300 0.22 0.19 400 0.17 0.18 1000 200 0.55 0.05 400 0.10 0.02 600 0.03 0.02 800 0.16 0.06 5000 1000 0.13 0.19 2000 0.20 0.38 3000 0.55 0.12 4000 0.25 0.06 10

0.02005 0.02 0.01995 0.0199 0.01985 0.0198 0.01975 0.0197 0.01965 0.0196 0.01955 0.0195 Figure 7: 20 µl dispensed using 50 µl syringe 0.0301 0.03 0.0299 0.0298 0.0297 0.0296 0.0295 0.0294 Figure 8: 30 µl dispensed using 50 µl syringe 0.0401 0.04 0.0399 0.0398 0.0397 0.0396 0.0395 0.0394 Figure 9: 40 µl dispensed using 50 µl syringe 11

0.0201 0.0199 0.0197 0.0195 0.0193 0.0191 0.0189 0.0187 Figure 10: 20 µl dispensed using 100 µl syringe 0.0402 0.04 0.0398 0.0396 0.0394 0.0392 0.039 0.0388 Figure 11: 40 µl dispensed using 100 µl syringe 0.0602 0.06 0.0598 0.0596 0.0594 0.0592 0.059 0.0588 Figure 12: 60 µl dispensed using 100 µl syringe 0.0802 0.08 0.0798 0.0796 0.0794 0.0792 0.079 0.0788 Figure 13: 80 µl dispensed using 100 µl syringe 12

0.0502 0.05 0.0498 0.0496 0.0494 0.0492 0.049 0.0488 Figure 14: 50 µl dispensed using 250 µl syringe 0.1002 0.1 0.0998 0.0996 0.0994 0.0992 0.099 Figure 15: 100 µl dispensed using 250 µl syringe 0.1502 0.15 0.1498 0.1496 0.1494 0.1492 0.149 Figure 16: 150 µl dispensed using 250 µl syringe 0.2002 0.2 0.1998 0.1996 0.1994 0.1992 0.199 Figure 17: 200 µl dispensed using 250 µl syringe 13

0.1004 0.1 0.0996 0.0992 0.0988 0.0984 Figure 18: 100 µl dispensed using 500 µl syringe 0.2002 0.2 0.1998 0.1996 0.1994 0.1992 0.199 0.1988 0 1 2 3 4 5 6 7 8 9 Figure 19: 200 µl dispensed using 500 µl syringe 0.3005 0.3 0.2995 0.299 0.2985 0.298 0.2975 Figure 20: 300 µl dispensed using 500 µl syringe 0.4005 0.4 0.3995 0.399 0.3985 0.398 0.3975 Figure 21: 400 µl dispensed using 500 µl syringe 14

0.2004 0.2 0.1996 0.1992 0.1988 0.1984 Figure 22: 200 µl dispensed using 1000 µl syringe 0.4002 0.4 0.3998 0.3996 0.3994 0.3992 Figure 23: 400 µl dispensed using 1000 µl syringe 0.6002 0.6 0.5998 0.5996 0.5994 Figure 24. 600 µl dispensed using 1000 µl syringe 0.8005 0.8 0.7995 0.799 0.7985 0.798 0.7975 0.797 Figure 25: 800 µl dispensed using 1000 µl syringe 15

1.002 1.001 1 0.999 0.998 0.997 0.996 0.995 Figure 26: 1000 µl dispensed using 5000 µl syringe 2.005 2 1.995 1.99 1.985 1.98 1.975 0 1 2 3 4 5 6 Figure 27: 2000 µl dispensed using 5000 µl syringe 3.005 3 2.995 2.99 2.985 2.98 2.975 Figure 28: 3000 µl dispensed using 5000 µl s yringe 4.002 4 3.998 3.996 3.994 3.992 3.99 3.988 3.986 3.984 3.982 Figure 29: 4000 µl dispensed using 5000 µl syringe 16

2.2 Results for air filters preparation In order to dispense liquid on a filter, dispensing patterns are designed in the custommade software application called Protocol Generator (Flamac, Belgium [3]). To enter the application one has to log in as an operator and has to select a folder in which all relevant data of the project will be stored. In the software, a filter model can be created with a certain width and length (inserted as X-size and Y-size) for a rectangular filter and diameter for a circular filter. Also the thickness of the filter (inserted as filter support height) can be introduced. On these filters one has to define an active area smaller than the filter size in which a dispensing pattern can be created. Two kinds of patterns, rectangular and circular, can be created independent of the shape of the filter. The pattern is a set of predefined positions where a chosen amount of liquid has to be dispensed by the device. The software allows saving different sizes of filters together with their active area, the patterns to be dispensed and combination of filter and patterns. All these files created by the software are saved in a selected folder and can be read in at a later stage. Once a combination of a filter and a dispensing pattern (called protocol) is generated in the software, it can be sent to the Chronos software (controlling the dispensing robot) and the motor controlling the XYZ table in order to execute the dispensing according to the inserted protocol. Figure 30: View of the protocol interface 17

3. Conclusion For the purchase of a precision pattern dispenser the minimum requirements were laid down. The minimum requirements were met with the solution that was proposed by sampleq. After installation the precision pattern dispenser tests were carried out to test the repeatability and trueness of the dispensing. The tests show that the instrument performed the dispensing within the target values set for both parameters. The precision pattern dispenser is qualified for use. 18

References [1] SampleQ, Avenue J.E. Lenoir 2,B-1348 Louvain-la-Neuve, Belgium, http://www.sampleq.com/ [2] CTC Analytics AG,Industriestrasse 20, CH-4222 Zwingen, Switzerland, http://www.ctc.ch/ [3] Flamac a division of SIM, Technologiepark-Zwijnaarde 903, 9052 Gent, http://www.flamac.be/ [4] ISO 5725-1:1994(en): Accuracy (trueness and precision) of measurement methods and results Part 1: General principles and definitions [5] Handbook of Chemistry and Physics, 53rd Edition, p. F4 19

List of abbreviations and definitions JRC-Geel: European Commission Joint Research Centre in Geel 20

List of figures Figure 1: Ampoules filling station of the automatic dispensing robot.... 5 Figure 2: Syringe positioned inside the tower.... 6 Figure 3: Solvent reservoirs and wash and waste vials with closed septa.... 6 Figure 4: Custom-made tray for ampoules.... 7 Figure 5: The XYZ-table.... 8 Figure 6: View on the opened plexiglass enclosure.... 8 Figure 7: 20 µl dispensed using 50 µl syringe... 11 Figure 8: 30 µl dispensed using 50 µl syringe... 11 Figure 9: 40 µl dispensed using 50 µl syringe... 11 Figure 10: 20 µl dispensed using 100 µl syringe... 12 Figure 11: 40 µl dispensed using 100 µl syringe... 12 Figure 12: 60 µl dispensed using 100 µl syringe... 12 Figure 13: 80 µl dispensed using 100 µl syringe... 12 Figure 14: 50 µl dispensed using 250 µl syringe... 13 Figure 15: 100 µl dispensed using 250 µl syringe... 13 Figure 16: 150 µl dispensed using 250 µl syringe... 13 Figure 17: 200 µl dispensed using 250 µl syringe... 13 Figure 18: 100 µl dispensed using 500 µl syringe... 14 Figure 19: 200 µl dispensed using 500 µl syringe... 14 Figure 20: 300 µl dispensed using 500 µl syringe... 14 Figure 21: 400 µl dispensed using 500 µl syringe... 14 Figure 22: 200 µl dispensed using 1000 µl syringe... 15 Figure 23: 400 µl dispensed using 1000 µl syringe... 15 Figure 24. 600 µl dispensed using 1000 µl syringe... 15 Figure 25: 800 µl dispensed using 1000 µl syringe... 15 Figure 26: 1000 µl dispensed using 5000 µl syringe... 16 Figure 27: 2000 µl dispensed using 5000 µl syringe... 16 Figure 28: 3000 µl dispensed using 5000 µl s yringe... 16 Figure 29: 4000 µl dispensed using 5000 µl syringe... 16 Figure 30: View of the protocol interface... 17 21

List of tables Table 1: Summary of the qualification results... 10 Table 2: Results of dispensing with a 50 µl syringe: 20, 30 and 40 µl... 23 Table 3: Results of dispensing with a 100 µl syringe: 20, 40, 60 and 80 µl... 24 Table 4: Results of dispensing with a 250 µl syringe: 50, 100, 150 and 200 µl... 25 Table 5: Results of dispensing with a 500 µl syringe: 100, 200, 300 and 400 µl... 26 Table 6: Results of dispensing with a 1000 µl syringe: 200, 400, 600 and 800 µl... 27 Table 7: Results of dispensing with a 5000 µl syringe: 1000, 2000, 3000 and 4000 µl 28 22

Annex 1 Results of the ampoules weighing Table 2: Results of dispensing with a 50 µl syringe: 20, 30 and 40 µl Set Dispensed Deviation Unc. # Volume volume from the set (ml) sample (ml) (ml) value 1 0.02 0.019802 0.000085 0.99% 2 0.02 0.019672 0.000113 1.64% 3 0.02 0.019890 0.000087 0.55% 4 0.02 0.019832 0.000113 0.84% 5 0.02 0.019804 0.000099 0.98% 6 0.02 0.019782 0.000121 1.09% 7 0.02 0.019870 0.000102 0.65% 8 0.02 0.019637 0.000113 1.81% 9 0.02 0.019660 0.000098 1.70% 10 0.02 0.019840 0.000118 0.80% Mean 0.019779 1.11% Stdev 9.1E-05 0.45% 1 0.03 0.029610 0.000112 1.30% 2 0.03 0.029812 0.000104 0.63% 3 0.03 0.029654 0.000095 1.15% 4 0.03 0.029830 0.000105 0.57% 5 0.03 0.029780 0.000081 0.73% 6 0.03 0.029570 0.000079 1.43% 7 0.03 0.029704 0.000071 0.99% 8 0.03 0.029720 0.000112 0.93% 9 0.03 0.029762 0.000070 0.79% 10 0.03 0.029640 0.000072 1.20% Mean 0.029708 0.97% Stdev 8.9E-05 0.30% 1 0.04 0.039780 0.000116 0.55% 2 0.04 0.039664 0.000098 0.84% 3 0.04 0.039784 0.000101 0.54% 4 0.04 0.039672 0.000115 0.82% 5 0.04 0.039835 0.000111 0.41% 6 0.04 0.039644 0.000130 0.89% 7 0.04 0.039821 0.000118 0.45% 8 0.04 0.039748 0.000116 0.63% 9 0.04 0.039811 0.000131 0.47% 10 0.04 0.039811 0.000149 0.47% Mean 0.039757 0.61% Stdev 7.1E-05 0.18% 23

Table 3: Results of dispensing with a 100 µl syringe: 20, 40, 60 and 80 µl Set Volume (ml) Dispensed volume (ml) Unc. (ml) Deviation from the set value # sample 1 0.02 0.019009 0.000103 4.95% 2 0.02 0.019262 0.000070 3.69% 3 0.02 0.019246 0.000084 3.77% 4 0.02 0.019497 0.000073 2.52% 5 0.02 0.019501 0.000067 2.50% 6 0.02 0.019505 0.000099 2.48% 7 0.02 0.019451 0.000068 2.75% 8 0.02 0.019294 0.000067 3.53% 9 0.02 0.019461 0.000068 2.70% 10 0.02 0.019116 0.000080 4.42% Mean 0.019334 3.33% Stdev 1.8E-04 0.88% 1 0.04 0.039094 0.000077 2.26% 2 0.04 0.039413 0.000074 1.47% 3 0.04 0.039520 0.000068 1.20% 4 0.04 0.039381 0.000069 1.55% 5 0.04 0.039395 0.000069 1.51% 6 0.04 0.039588 0.000080 1.03% 7 0.04 0.039417 0.000076 1.46% 8 0.04 0.039265 0.000068 1.84% 9 0.04 0.039411 0.000071 1.47% 10 0.04 0.039247 0.000070 1.88% Mean 0.039373 1.57% Stdev 1.4E-04 0.35% 1 0.06 0.058928 0.000089 1.59% 2 0.06 0.058996 0.000093 1.48% 3 0.06 0.059210 0.000087 1.12% 4 0.06 0.059232 0.000095 1.08% 5 0.06 0.059094 0.000095 1.31% 6 0.06 0.059120 0.000088 1.27% 7 0.06 0.059178 0.000088 1.17% 8 0.06 0.058908 0.000086 1.63% 9 0.06 0.059004 0.000087 1.47% 10 0.06 0.059258 0.000092 1.04% Mean 0.059210 1.32% Stdev 1.3E-04 0.21% 1 0.08 0.078899 0.000074 1.18% 2 0.08 0.078997 0.000082 1.06% 3 0.08 0.079185 0.000077 0.82% 4 0.08 0.079211 0.000069 0.79% 5 0.08 0.078875 0.000078 1.21% 6 0.08 0.078907 0.000073 1.17% 7 0.08 0.078969 0.000082 1.09% 8 0.08 0.078963 0.000069 1.10% 9 0.08 0.079103 0.000077 0.93% 10 0.08 0.079155 0.000077 0.86% Mean 0.079183 1.02% Stdev 1.3E-04 0.16% 24

Table 4: Results of dispensing with a 250 µl syringe: 50, 100, 150 and 200 µl # sample Set Volume (ml) Dispensed volume (ml) U (ml) Deviation from the set value 1 0.05 0.049187 0.000076 1.63% 2 0.05 0.049016 0.000068 1.97% 3 0.05 0.049307 0.000071 1.39% 4 0.05 0.049401 0.000101 1.20% 5 0.05 0.049460 0.000066 1.08% 6 0.05 0.049269 0.000069 1.46% 7 0.05 0.049351 0.000068 1.30% 8 0.05 0.049110 0.000079 1.78% 9 0.05 0.049492 0.000066 1.02% 10 0.05 0.049648 0.000068 0.70% Mean 0.049324 1.35% Stdev 1.9E-04 0.38% 1 0.1 0.099435 0.000090 0.57% 2 0.1 0.099288 0.000069 0.71% 3 0.1 0.099318 0.000070 0.68% 4 0.1 0.099304 0.000067 0.70% 5 0.1 0.099455 0.000068 0.55% 6 0.1 0.099511 0.000070 0.49% 7 0.1 0.099320 0.000066 0.68% 8 0.1 0.099318 0.000073 0.68% 9 0.1 0.099354 0.000068 0.65% 10 0.1 0.099286 0.000081 0.71% Mean 0.099359 0.64% Stdev 7.9E-05 0.08% 1 0.15 0.149390 0.000075 0.41% 2 0.15 0.149253 0.000072 0.50% 3 0.15 0.149327 0.000075 0.45% 4 0.15 0.149442 0.000068 0.37% 5 0.15 0.149546 0.000065 0.30% 6 0.15 0.149470 0.000070 0.35% 7 0.15 0.149368 0.000096 0.42% 8 0.15 0.149323 0.000067 0.45% 9 0.15 0.149171 0.000077 0.55% 10 0.15 0.149241 0.000076 0.51% Mean 0.149353 0.43% Stdev 1.1E-04 0.08% 1 0.2 0.199270 0.000089 0.36% 2 0.2 0.199194 0.000073 0.40% 3 0.2 0.199294 0.000068 0.35% 4 0.2 0.199343 0.000068 0.33% 5 0.2 0.199302 0.000069 0.35% 6 0.2 0.199323 0.000071 0.34% 7 0.2 0.199176 0.000067 0.41% 8 0.2 0.199353 0.000101 0.32% 9 0.2 0.199499 0.000080 0.25% 10 0.2 0.199192 0.000065 0.40% Mean 0.199295 0.35% Stdev 9.6E-05 0.05% 25

Table 5: Results of dispensing with a 500 µl syringe: 100, 200, 300 and 400 µl # sample Set Volume (ml) Dispensed volume (ml) U (ml) Deviation from the set value 1 0.1 0.099050 0.000067 0.95% 2 0.1 0.099078 0.000068 0.92% 3 0.1 0.098985 0.000067 1.01% 4 0.1 0.099110 0.000072 0.89% 5 0.1 0.099348 0.000066 0.65% 6 0.1 0.099035 0.000066 0.96% 7 0.1 0.099166 0.000077 0.83% 8 0.1 0.098901 0.000070 1.10% 9 0.1 0.099037 0.000067 0.96% 10 0.1 0.099106 0.000065 0.89% Mean 0.099082 0.92% Stdev 1.2E-04 0.12% 1 0.2 0.199822 0.000068 0.09% 2 0.2 0.199704 0.000069 0.15% 3 0.2 0.199377 0.000073 0.31% 4 0.2 0.199582 0.000075 0.21% 5 0.2 0.198960 0.000071 0.52% 6 0.2 0.199533 0.000070 0.23% 7 0.2 0.199395 0.000069 0.30% 8 0.2 0.199634 0.000073 0.18% 9 0.2 0.199437 0.000067 0.28% Mean 0.199494 0.25% Stdev 2.5E-04 0.12% 1 0.3 0.299787 0.000068 0.07% 2 0.3 0.299620 0.000067 0.13% 3 0.3 0.299624 0.000070 0.13% 4 0.3 0.299353 0.000068 0.22% 5 0.3 0.299692 0.000071 0.10% 6 0.3 0.299102 0.000067 0.30% 7 0.3 0.299458 0.000070 0.18% 8 0.3 0.297881 0.000067 0.71% 9 0.3 0.299106 0.000071 0.30% 10 0.3 0.299636 0.000066 0.12% Mean 0.299326 0.22% Stdev 5.6E-04 0.19% 1 0.4 0.398818 0.000065 0.30% 2 0.4 0.399994 0.000082 0.00% 3 0.4 0.400132 0.000068-0.03% 4 0.4 0.399406 0.000067 0.15% 5 0.4 0.399464 0.000066 0.13% 6 0.4 0.398840 0.000066 0.29% 7 0.4 0.399454 0.000065 0.14% 8 0.4 0.397757 0.000086 0.56% 9 0.4 0.399839 0.000067 0.04% 10 0.4 0.399580 0.000078 0.10% Mean 0.399328 0.17% Stdev 7.0E-04 0.18% 26

Table 6: Results of dispensing with a 1000 µl syringe: 200, 400, 600 and 800 µl # sample Set Volume (ml) Dispensed volume (ml) U (ml) Deviation from the set value 1 0.2 0.199086 0.000074 0.46% 2 0.2 0.198823 0.000067 0.59% 3 0.2 0.198945 0.000080 0.53% 4 0.2 0.198745 0.000069 0.63% 5 0.2 0.198843 0.000069 0.58% 6 0.2 0.198891 0.000071 0.55% 7 0.2 0.198891 0.000066 0.55% 8 0.2 0.199060 0.000070 0.47% 9 0.2 0.198795 0.000065 0.60% 10 0.2 0.198855 0.000067 0.57% Mean 0.198893 0.55% Stdev 1.1E-04 0.05% 1 0.4 0.399388 0.000070 0.15% 2 0.4 0.399550 0.000075 0.11% 3 0.4 0.399578 0.000076 0.11% 4 0.4 0.399542 0.000069 0.11% 5 0.4 0.399618 0.000077 0.10% 6 0.4 0.399592 0.000072 0.10% 7 0.4 0.399584 0.000069 0.10% 8 0.4 0.399681 0.000073 0.08% 9 0.4 0.399679 0.000070 0.08% 10 0.4 0.399663 0.000072 0.08% Mean 0.399587485 0.10% Stdev 8.6E-05 0.02% 1 0.6 0.599603 0.000081 0.07% 2 0.6 0.599727 0.000075 0.05% 3 0.6 0.599796 0.000077 0.03% 4 0.6 0.599715 0.000075 0.05% 5 0.6 0.599776 0.000101 0.04% 6 0.6 0.599942 0.000076 0.01% 7 0.6 0.599780 0.000076 0.04% 8 0.6 0.599729 0.000078 0.05% 9 0.6 0.599990 0.000092 0.00% 10 0.6 0.599860 0.000079 0.02% Mean 0.599792 0.03% Stdev 1.1E-04 0.02% 1 0.8 0.798858 1.997845 0.14% 2 0.8 0.799054 1.998242 0.12% 3 0.8 0.799128 1.998264 0.11% 4 0.8 0.799353 1.998696 0.08% 5 0.8 0.798917 1.998199 0.14% 6 0.8 0.798525 1.997375 0.18% 7 0.8 0.797670 1.995948 0.29% 8 0.8 0.798717 1.997936 0.16% 9 0.8 0.798205 1.996961 0.22% 10 0.8 0.799112 1.998329 0.11% Mean 0.798754 0.16% Stdev 5.0E-04 0.06% 27

Table 7: Results of dispensing with a 5000 µl syringe: 1000, 2000, 3000 and 4000 µl # sample Set Volume (ml) Dispensed volume (ml) U (ml) Deviation from the set value 1 1 0.996359 0.000124 0.36% 2 1 0.997122 0.000071 0.29% 3 1 0.995462 0.000073 0.45% 4 1 0.999205 0.000072 0.08% 5 1 0.999005 0.000070 0.10% 6 1 0.998411 0.000097 0.16% 7 1 0.998722 0.000084 0.13% 8 1 1.000220 0.000078-0.02% 9 1 1.000854 0.000076-0.09% 10 1 1.001173 0.000072-0.12% Mean 0.998653 0.13% Stdev 1.9E-03 0.19% 1 2 1.998192 0.000026 0.09% 2 2 1.999380 0.000026 0.03% 3 2 1.999396 0.000026 0.03% 4 2 1.999494 0.000028 0.03% 5 2 1.999452 0.000028 0.03% Mean 1.996090 0.20% Stdev 7.6E-03 0.38% 1 3 2.980831 0.000024 0.64% 2 3 2.982638 0.000023 0.58% 3 3 2.981697 0.000025 0.61% 4 3 2.978159 0.000025 0.73% 5 3 2.980186 0.000023 0.66% 6 3 2.982949 0.000030 0.57% 7 3 2.985927 0.000023 0.47% 8 3 2.988139 0.000023 0.40% 9 3 2.989800 0.000023 0.34% 10 3 2.984046 0.000023 0.53% Mean 2.983437 0.55% Stdev 3.6E-03 0.12% 1 4 3.984627 0.000023 0.38% 2 4 3.987231 0.000028 0.32% 3 4 3.991572 0.000023 0.21% 4 4 3.992467 0.000023 0.19% 5 4 3.991347 0.000023 0.22% 6 4 3.990057 0.000027 0.25% 7 4 3.990002 0.000024 0.25% 8 4 3.990607 0.000024 0.23% 9 4 3.991022 0.000037 0.22% 10 4 3.990314 0.000030 0.24% Mean 3.989925 0.25% Stdev 2.3E-03 0.06% 28

Europe Direct is a service to help you find answers to your questions about the European Union Free phone number (*): 00 800 6 7 8 9 10 11 (*) Certain mobile telephone operators do not allow access to 00 800 numbers or these calls may be billed. A great deal of additional information on the European Union is available on the Internet. It can be accessed through the Europa server http://europa.eu How to obtain EU publications Our publications are available from EU Bookshop (http://bookshop.europa.eu), where you can place an order with the sales agent of your choice. The Publications Office has a worldwide network of sales agents. You can obtain their contact details by sending a fax to (352) 29 29-42758.

LC-NA-28268-EN-N JRC Mission doi:10.2789/1167 ISBN 978-92-79-64110-7 2