OPERATION and MAINTENANCE INSTRUCTION MANUAL. AEU-10C/SS Power-Tip Systems

Similar documents
OPERATION and MAINTENANCE INSTRUCTION MANUAL. AEU-03SS Porta-Tip Systems

OPERATION and MAINTENANCE INSTRUCTION MANUAL

OPERATION and MAINTENANCE INSTRUCTION MANUAL. AEU-500 Mini-Electric Portable Dental System

OPERATION AND MAINTENANCE INSTRUCTION MANUAL. AEU-17B & AEU-17BV2 Implant / Surgery Systems

OPERATION AND MAINTENANCE INSTRUCTION MANUAL. AEU-707A & AEU-707AV2 Implant / Surgery Systems

OPERATION AND MAINTENANCE INSTRUCTION MANUAL. AEU-707A & AEU-707AV2 Implant / Surgery Systems

OPERATION and MAINTENANCE INSTRUCTION MANUAL. AEU-70A Peristaltic Infusion Pump

OPERATION and MAINTENANCE INSTRUCTION MANUAL. AA-98 AirPac II Compressor 110V

OPERATION and MAINTENANCE INSTRUCTION MANUAL. AEU-7/AEU-7P Peristaltic Water Pump

OPERATION and MAINTENANCE INSTRUCTION MANUAL. ADU-12DCE Task Force Deluxe Portable Pneumatic Dental System

ADC-09S

OPERATION AND MAINTENANCE INSTRUCTION MANUAL. AEU-5000 Dental System

ALU-29CF Portable Light NSN:

OPERATION AND MAINTENANCE MANUAL. AseptiChair. Item ADC-01

OPERATION AND MAINTENANCE INSTRUCTION MANUAL ALU-40MOM PORTABLE DENTAL LAMP

OPERATION AND MAINTENANCE INSTRUCTION MANUAL

OPERATION AND MAINTENANCE INSTRUCTION MANUAL

OPERATION AND MAINTENANCE INSTRUCTION MANUAL ALU-40TC PORTABLE DENTAL LAMP

OPERATION AND MAINTENANCE INSTRUCTION MANUAL ALU-40LED PORTABLE DENTAL LAMP

OPERATION AND MAINTENANCE INSTRUCTION MANUAL

Alternating Current Revolution per minute Protective Earth Ground. Protected from dripping water Fuse Attention, Consult Accompanying Document

SURGICAL POWER & ACCESSORIES. Rx Only PM-X00-710/PM-X IR-12.13

FlexChem Model 404 Rotating Oven

Installation Operation Maintenance Troubleshooting Version2.1. Feb/09. Led II

HTR-202 INSTRUCTION MANUAL. Portable Digital Height Rod. Befour, Inc. 102 Progress Drive Saukville, WI 53080

SERVICE MANUAL & PARTS LIST

Submersible Pond Pump with UV 800 Gallons Per Hour

Operator's Manual. Storage System. Ultrasound Probe Cabinet. Manufactured by:

DECADE 1021 SOLO CHAIR

Model 605 Portable Aspirator

Model NTX7 Series Automatic Battery Charger User s Manual Rev. 1.0 October 17, 2006

Installation/Operating Instructions

82V LITHIUM-ION BATTERY CHARGER GC 400

DCell Suction. Model DM Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority

MAGNAGLIDE AIR OPERATOR VERTICAL MOUNT

Introduction. The words WARNING, CAUTION and NOTE carry special meanings and should be carefully reviewed.

NEUTRALIZATION SYSTEM FOR OPA & GLUTARALDEHYDE MEDI-NEWT. User Manual. Endoscope reprocessing the the way way it it should be be

DECADE 1021 VAC BACK CHAIR

1000 DE BOSS. Installation & Operating Instructions COMMERCIAL FIXTURE READ THESE INSTRUCTIONS BEFORE FIRING UP YOUR BALLAST

8,1,8,1,2,7,2,7,6,3,6,3,4,5,4,5

CATEYE URBAN WIRELESS+

Have a Bright Day! Nicholas Harmon President and CEO

OWNER S MANUAL Please read and understand all precautions prior to use.

Model 2310 Series. Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority

Model 700 Microarray Oven

RemRest. Service Manual. CAUTION: USA Federal law restricts this device to sale by or on the order of a physician. P/N Rev A

18VDC ESB6 Series Cordless Screwdrivers Operation Manual

PL900 ILLUMINATOR. Fiber-Lite. Operation Manual. Setup. Fiber Optic Connection A PRODUCT OF DOLAN-JENNER INDUSTRIES. Voltage Selection:

INSTRUCTIONS FOR USE. Rx Only

User Manual Stealth s Power Tilt

OPERATING MANUAL POWER DRIVE SYSTEM. Cat. No RevH

Quickdraw. Model 2400 Series. Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority

SUBMERSIBLE MINI-PUMP

THIS PRODUCT IS FOR PROFESSIONAL LABORATORY USE ONLY USER'S MANUAL. WELLS ENGINE UNIT 230 VOLT Product No. U905, U906, U907, U908

Container Fountain Kit with LED Light

Installation/Operating Instructions

Sentinel Enteral Feeding Pump. Model: S-1000-SI. Operating Manual

18VDC ESB6-X Series Cordless Screwdrivers Operation Manual

Installation Operation Maintenance Troubleshooting Version 1, Oct/09. Palm Beach

Illumifalls Patio Pond

LIMITED WARRANTY AIR COMPRESSORS & PORTABLE DENTAL UNITS I NSTRUMENTS TERMS

1416 Cabinet Unit. Installation Operation Maintenance Troubleshooting Version Dec/17

Stryker Visum 300 Ceiling-Mounted Exam Light Assembly, Operations, and Maintenance Manual

Automated I.V. Stand Cat. No OPERATOR S MANUAL. Operation& Maintenance Instructions. Raising the Standard of Care!

Model G14SD Vapor Management System

series USER MANUAL

Infrared Thermocouple Probe

User s Manual Power Supply. IM E 3rd Edition. Yokogawa Electric Corporation

XCEL 250 XCEL 400/700. Xcel Slit Lamp USER'S GUIDE Rev. F Page1

Basic Steel Rollator with 6" Wheels Knockdown Rolling Walker

BASIC STEEL ROLLATOR WITH 6" WHEELS KNOCKDOWN ROLLING WALKER

Submersible Waterfall Pump 1200 Gallons Per Hour

Have a Bright Day! Nicholas Harmon President, Verilux, Inc.

Instruction Manual for 30,000RPM Motor Assembly Model: PH3 30X #

Natural Fiber Planter Fountain

EV-2051-M Electric Motor. Operation and Maintenance Manual

Model HPX60 Series Automatic Battery Charger User s Manual Rev. 1.0 October 17, 2006

ST-70 CONTROL OPERATING MANUAL REVISION DATE: PART#:

Instruction Manual. 8,000RPM Right-Angle Handpiece Model: RZ-15X # for

AEU-14CF Emergency Field Dental Unit

Fluorescent Ring Light with Dimmer INSTRUCTIONS

Premium Pond Pump 330 Gallons Per Hour

Submersible Pond Pump with UV 800 Gallons Per Hour

User s Manual Power Supply. IM E 2nd Edition. Yokogawa Electric Corporation

58kHz AM Deactivator DB104 Instruction Manual V1.0

Hybex Microsample Incubator

EBG Series Surface Preparation Machines

Handheld LED Light Source (HLS) Operator s Manual

The Da-Lite Difference.

User s Manual Power Supply IM E. 5th Edition

Podiatry Procedures Chair. For Models: Barrier-Free. User s Guide

(450 m-925 m) ET/MDPE Tape Water Level Meter (with optional electric rewind)

Sylvania 9 Color Changing One Plug Tree Tree ID # T5, Item # V

Stealth Power i-tilt

RENA AF371Feeder Operating Manual. Feeder. Operating Manual. Manual Part #: M AF371 Operations Rev

INSTRUCTION MANUAL. Tel / Fax SOM-S056S-KE Rev.5

Model. Torque accuracy ±3% No load speed/ rpm Machine screw mm / in. Available Screw. Machine screw mm / in

INSTRUCTION MANUAL AUTO-PUMP 220/240 VAC MODEL #091-9B-220. INPUT :220/240 volt, 50/60 Hz, 1.9/2.2 AMPS OUTPUT: 100 PSI MAXIMUM 1 YEAR WARRANTY

USER MANUAL. EC+ Extra Care Bariatric Trauma Stretcher Model ConvaQuip Ind., Inc. P.O. Box 3417 Abilene, TX

Xenon Fiberoptic Lightsource

Transcription:

OPERATION and MAINTENANCE INSTRUCTION MANUAL AEU-10C/SS Power-Tip Systems

TABLE OF CONTENTS: Introduction.........................1 Package Contents....................1 Setting Up the Unit...................2 Operation...........................3 Handpiece..........................4 Troubleshooting......................5 Warranty...................Back Cover SPECIFICATIONS: Console Size: 4.5 x 3.5 x 2 (11 cm x 9 cm x 5 cm) Console Weight: 1.25 lbs (0.56 kg) Motor and Handpiece Size: 6 x 0.8 diameter (max) (15 cm x 2 cm dia.) Motor, Cord and Handpiece Weight: 0.35 lb (0.16 kg) Power Source: 110V AC or Optional 220V AC Max Power Consumption: 15.3V, 6.0W P.O. Box 1548 Woodinville, WA 98072 8333 216 th Street S.E. Woodinville, WA 98072 International (425) 487-3157 Toll Free(800) 426-5913 www.aseptico.com info@aseptico.com To prevent injury to people and damage to property, please heed relevant warnings and remarks. They are marked as follows: WARNING: CAUTION: NOTE: 0459 Aseptico Inc 98072 USA EC REP Advena Ltd HR4 9DQ U.K. Serious injury or death may result if ignored. Damage to property or the environment may result if ignored. Important additional information and hints. WARNING: This device has been tested and found to comply with the emissions requirements of EN 55014:1997. These requirements provide reasonable protection against harmful electro-magnetic interference in a typical medical installation. However, high levels of radio-frequency (RF) emissions from electrical devices, such as cellular phones, may disrupt the performance of this device. To mitigate disruptive electromagnetic interference, position this device away from RF transmitters and other sources of electromagnetic energy.

Your new Aseptico AEU-10 Power-Tip System is one of the finest motor/handpiece combinations available to the professional. Congratulations This system is engineered to provide many years of reliable service. Please read the instructions provided in this manual to receive the best and longest service from your Aseptico equipment. Separate manuals may be provided to cover the operation and maintenance of handpieces or other accessories for your unit. PACKAGE CONTENTS: Electric Control Console Remote Power Cord AM-103A Micro Motor AHP-101 Straight Handpiece Autoclavable Motor Holder AE-7P Foot Pedal (Optional for AEU-10C) 1

SETTING UP THE UNIT: 1. Plug the AEU-10 console into a standard 120 Volt AC grounded power outlet. (220 Volt AC unit is also available) 2. Plug the micro motor into the receptacle on the front of the console labeled 'motor'. Make sure the power switch on the console is turned off. 3. Attach the straight handpiece to the motor by sliding straight onto the motor shaft. Insert a bur or bit into the handpiece collet. Only operate the unit with a bit or blank installed and with the collet in the closed position. 4. Attach AE-7P on/off foot pedal to the receptacle on the back of the unit marked Foot Switch. The AE-7P foot pedal is standard with the AEU-10SS and optional with the AEU-10C. The variable speed AE-14A foot control is also available. Figure 1 ON Indicator On / Off Switch Bit Motor Handpiece Motor Console Motor Holder (Close Collet) AHP-101 Handpiece Foot Pedal Bit Foot Switch Receptacle AC Power Inlet Fuse Holder 2

Figure 2 POWER SWITCH 'ON' INDICATOR HAND/FOOT SELECTOR SPEED DIAL MOTOR RECEPTACLE FWD/REV SELECTOR OPERATION: 1. Turn the power switch to the 'ON' position. When switched on, the L.E.D. 'ON' indicator should light. 2. Select the proper running speed by turning the 'SPEED' dial. Turn the dial clockwise to increase speed, and counterclockwise to decrease speed. 3. When running the handpiece, DO NOT force or stall the motor. If the motor stalls due to excessive pressure, release pressure immediately and use a light touch. 4. If equipped with a foot pedal, select Foot with hand/foot selector and press foot pedal to operate. Foot pedal must be unplugged and selector must be moved to Hand to return to hand mode. NOTE: If motor stops due to an open collet or excessive pressure: 1. Turn Power Switch off 2. Let unit cool down for 2-4 min. 3. Turn Power Switch on 4. Resume use WARNING: Do not attempt to disassemble the motor or motor connector. Do not oil or lubricate the motor. Do not attach a handpiece to the motor while the motor is running. Do not bend motor cord sharply. Failure to comply with any of the above instructions may void your warranty CLEANING THE CONSOLE: The exterior of the console may be cleaned by wiping with a soft cloth moistened with mild detergent or disinfecting solution. 3

HANDPIECE: The Aseptico AEU-10 System includes the AHP-101 straight handpiece. This handpiece features a quick-twist chuck that will accept any standard HP (3/32") shank bit (or bur) or any Doriot style attachment. To Install Bit: 1. Open collet by turning grip 1/4 turn as shown in Fig. 4 until you feel a click. The gap between the grip and handpiece retainer will narrow when the collet is open. 2. Insert bit fully into collet. 3. Close collet by turning grip 1/4 turn as shown in Fig. 3 To Remove Bit: 1. Open collet by turning grip 1/4 turn as shown in Fig. 4 until you feel a click. 2. Push bit back into collet and turn to release. 3. Pull bit straight out of collet. Insert Bur Close Collet Bit Collet Grip Gap Retainer Figure 3 NOTE: Before use, check to insure that bit is fully inserted and grip is firmly secured in locked position. Push, Turn, and Pull Bur Figure 4 Open Collet Sterilization: The AHP-101 steam autoclaving sterilization procedure provided below has been independently validated to achieve a sterility assurance level (SAL) of 10-6. This handpiece must be sterilized before initial use and between each patient use per the following procedure: Scrub dirt and other debris from handpiece and wipe clean with a cotton swab or cloth. Autoclave at 132 C (270 F) for ten minutes. Do not rinse or submerge in any solution, or use ultrasonic cleaners. Do not lubricate. The sealed bearings in this handpiece eliminate the need for lubrication. 4

TROUBLESHOOTING: Problem: On Indicator does not light when turned on: On Indicator lights when turned on, but handpiece does not turn: On Indicator dims and goes out, and/or motor slows and stops working. Correction: Check console to power connection. Check that voltage is proper voltage. Check fuse. If blown, replace with 5x20mm 0.5A/250V fast-acting fuse. Depress foot pedal (if used). Check foot pedal connection (if used). Check motor plug connection. Test for possible bad handpiece by removing handpiece from motor and then running motor by itself. Check that Hand/Foot Selector switch is placed fully in either the Hand or Foot positions and isn t set halfway between. Check that a bit is properly seated in the handpiece and the collet is closed. Test for possible motor/cable short by removing motor cable from motor receptacle. Test for possible bad bearings in the motor by removing the handpiece from the motor then running motor by itself. 5

WARRANTY Aseptico warrants its products against defects in material or workmanship for a period of two (2) years, from date of original invoice. Some handpieces are warranted for one year under the same conditions. Other handpieces and expendable components, such as air turbines and light bulbs, are covered by shorter warranty periods, or have no warranty. Aseptico's sole obligation under product warranty is (at its sole option and discretion) to repair or replace any defective component or product in part or whole. Aseptico shall be the sole arbiter of such action. In the event of alleged defect under warranty, the purchaser is to notify Aseptico's Customer Service Department promptly. Customer Service will provide instructions, usually directing that the product be returned for service. Shipment to Aseptico and the cost thereof is always the responsibility of the purchaser. Accidental misuse, inappropriate installation, or failure to perform directed maintenance voids the warranty. Aseptico does not assume, under this warranty, any risks or liabilities arising from the clinical use of its products, whether or not such use involves coincidental utilization of products manufactured by others. P/N: 420270 Rev. I ECO 12576 03/2011 P.O. Box 1548 Woodinville, WA 98072 8333 216th Street SE Woodinville, WA 98072 (425) 487-3157 (800) 426-5913 www.aseptico.com info@aseptico.com PRINTED IN THE U.S.A.