LAMINAR FLOW HORIZONTAL FLOW BENCH (LF) OPERATION & MAINTENANCE MANUAL

Similar documents
Unimodule M2 Unidirectional Flow Clean Bench/Workstation OPERATION MANUAL

Envirco BIOLOGICAL SAFETY CABINET

LAMINAR DOWNFLOW MODULE OPERATION & MAINTENANCE MANUAL. ENVIRCO Technical Support:

VLF-797 CLEAN BENCH INSTALLATION AND SERVICE MANUAL Volt Model Volt Model

Hospi-Gard IsoClean and IsoClean UV HEPA Filtration Systems INSTALLATION, OPERATION, AND MAINTENANCE MANUAL

Series 301. Horizontal Laminar Flow Clean Bench. C l e a n B e n c h e s. s o l u t i o n s. Series 301 Horizontal Laminar Flow Clean Bench

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

CLEAN AIR WORKSTATION REVERSE FLOW. Installation and Operations Manual

BioPROtect III. and BioPROtect III Jr. Class II Style Walk-In Clean Air and Containment Enclosures.

PURCHASE SPECIFICATIONS FOR NUAIRE PHARMAGARD NU-NR797 NEGATIVE PRESSURE RECIRCULATING STERILE ISOLATOR

FAN POWERED HEPA FILTER UNIT

TXH TOTAL EXHAUST HOOD. Installation and Operations Manual

PHARMACY Products. Compounding Aseptic Isolator (CAI) Positive pressure sterility assurance barrier isolator for complete product protection.

A viewscreen top hinge allows full opening for loading of instrumentation or equipment.

SterilSHIELD. Designed for Aseptic Compounding of Non-Hazardous Pharmaceuticals. SterilSHIELD Barrier Isolator (positive pressure)

PHARMACY Products. Compounding Aseptic Containment Isolator (CACI) A level of safety that is built in, not bolted on. Underwriter Laboratories Listed

Operations & Maintenance Manual DFH-1000 DUCTLESS FUME HOOD

Omniaire 1000V Instruction Manual-8/05

Operator's Manual. Storage System. Ultrasound Probe Cabinet. Manufactured by:

VCM LAMINAR FLOW HOOD VERTICAL

LAMINAR AIRFLOW EQUIPMENT

READ AND SAVE THESE INSTRUCTIONS. ComfortBreeze UV360-1 SYSTEM 24V Ultra-Violet Air Cleaner. Trion


HOWORTH EXFLOW 28 PREVENTS AIRBORNE INFECTION TECHNICAL SPECIFICATION. Exflow 28 Technical Specification. Page 1 of 7

ADVANCED SAFETY ENGINEERING FOR EFFECTIVE HANDLING OF HAZARDOUS PARTICULATE DRUGS AND POWDERS IN A CLASS II, TYPE A2 BIOLOGICAL SAFETY CABINET

NU-430/435 CLASS II, TYPE B2 BIOLOGICAL SAFETY CABINETS

As ept i c I s olator s

READ AND SAVE THESE INSTRUCTIONS. Air Boss MP600M Vertical Air Flow Mist Precipitator Industrial Applications. TRION

OmniAire 2200C Operations and Maintenance Manual

A. Each cabinet must be constructed and installed in accordance with EN12469:2000.

Quest DRY 150. Quest PowerHEPA 500. quest. Read and Save These Instructions. Quest PowerHEPA 500:

Best Products. Best Performance. Best Protection.

HORD HOSPITAL OPERATING ROOM AIR CURTAIN

LABGARD FEATURES

The Da-Lite Difference.

DESIGN-AIRE SL-RSR-EC. Installation and Operations Manual

Further information: Berriman Associates Model 850P.2 Owners Manual

TUTTLE & BAILEYR. Installation & Service Manual PRODUCT DIMENSIONS. FIGURE 1 - Product Isometric and Dimensional Views

After installing the HEPA filters the discharge air should be monitored to ensure proper filtering efficiency. HEPA Filters

Environmental Series Air Curtains

VERSAFLOW SERIES UNIDIRECTIONAL LAMINAR AIRFLOW COMPOUNDING ASEPTIC ISOLATOR (CAI)

AS , EN12469 (Microbiological Testing Method)

MODEL 950C-NP-UV Installation and Owners Manual

HORD HOSPITAL OPERATING ROOM AIR CURTAIN

Advanced PCO System. Installation Instructions & Consumer Information. Section I. Important Installer Information WARNINGS

Photographer: BRUCE T. MARTIN, LFD Laminar Flow Diffuser

Controlled Environment Equipment Solutions for the Lab

Installation Instructions

Model NTX7 Series Automatic Battery Charger User s Manual Rev. 1.0 October 17, 2006

INSTALLATION, OPERATION, AND MAINTENANCE INSTRUCTIONS. Part Number: P DM

Protector XStream Laboratory Hood is shown with SpillStopper Work Surface , Protector Acid Storage Cabinet and Protector

MICRO SEAL-SERIES HEPA Holding Frames and Housings

Requires Ductwork and Blower

High Performance Ductless Balance Enclosures

A Thermolab Group Company. HyZone. Class II Microbiological Safety Cabinets. for operator, environment and product protection.

Product Service

IV STANDS. Assembly, Usage and Maintenance INSTRUCTION MANUAL

MODEL MC1500 Installation and Operation Manual Important:

is with a variety of standard dimensions. Also our units can be custom engineered to any size.

DESIGN-AIRE SL-RSR-EC. Installation and Operations Manual

High Velocity Series Air Curtains

Quest H5. Quest Dry. Read and Save These Instructions. Installation, Operation and Maintenance Instructions. Quest H5:

INSTALLATION INSTRUCTIONS

PCR Laminar Flow Cabinets

Model G14SD Vapor Management System

INSTALLATION, MAINTENANCE & OPERATION GUIDE. MODEL Sanuvair S300 WHOLE HOUSE HEPA/UV AIR PURIFIER. ITEM # Sanuvair S V EN

OmniAire Portable Air Cleaner OA1200PAC

Purifier Logic Class II Biosafety Cabinets FEATURES & BENEFITS

Operating Manual PFB-2500

Nilfisk-CFM SS Vapor Vacuum Instructions for Use and Spare Parts Manual

FE-WS6 Downflow Ductless Workstation. Operator s Manual

Cytotoxic Safety Cabinets ENVAIR eco safe Plus

AeraMax Pro AM IV. Please read these instructions before use. Do not discard: keep for future reference.

3M Ionized Air Gun 980/980E. User s Guide

Thermo Scientific 1300 Series A2. Class II, Type A2 Biological Safety Cabinets. Good for you good for the environment

INSTALLATION INSTRUCTIONS For Model 855 Wolo Express TM

Installer s Manual. PCO/UV N-Duct Model

Clean Air Products Straight Thru Standard Profile Single Door Air Shower Purchase Specification

EXT HEPA/CARBON FUME HOOD

EXT DUCTLESS FUME HOOD

Integrated Heating System

Tissue Master. User Manual

Stainless Steel Kick Bucket and Stand

Installation Precautions. Use Precautions. Specifications

INSTALLATION INSTRUCTIONS

OWNER S MANUAL C O M M E R C I A L I N D O O R A I R Q U A L I T Y S O L U T I O N S. Total High Efficiency Particulate & Gaseous Filtration System

The Da-Lite Difference.

SafeFAST Top. Class II Microbiological Safety Cabinets PROTECTION, SAFETY, RELIABILITY. AND MORE.

Envair Lab SCS Evo Comfort+ Class II Microbiological Safety Cabinets

3M Ionized Air Gun 980/980E. User s Guide

PHARMAGARD 797 CREATING A SINGLE SOURCE SOLUTION TO ENSURE THE ACCURACY AND STERILITY OF COMPOUNDED PHARMACY PRODUCTS

HARDWARE KIT INSTALLATION GLOVES

FA1 Series Air Filtration System. Operations Manual

Model: UV-16R/24 READ THESE INSTRUCTIONS CAREFULLY AND COMPLETELY BEFORE PROCEEDING WITH THE INSTALLATION.

Model ESMH MOTORIZED FORCE MEASUREMENT TEST STAND. User s Guide

PFS Rotameter. Variable Area Flowmeter. Instruction Manual PFS-RM200, RM230, RM250

1.4 N/1.4 D. Operating Instructions. 580 Mayer Street, Building #7, Bridgeville, PA phone fax

Instruction Manual. Fudge Puppy Display Case

CRITICAL ENVIRONMENT DIFFUSERS

Biological Safety Cabinet

Transcription:

LAMINAR FLOW HORIZONTAL FLOW BENCH (LF) OPERATION & MAINTENANCE MANUAL ENVIRCO Technical Support: 800-884-0002

Table of Contents WARNING... 3 INTRODUCTION... 3 FEATURES... 4 APPLICATIONS... 4 ASSEMBLY... 5 SERVICE... Fluorescent Lamp Replacement... 5 Fluorescent Fixture Ballast Replacement... 5 Prefilter Replacement... 6 HEPA Filter Replacement... 6 Blower Replacement... 6 Rate of Flow Adjustment... 6 WIRING DIAGRAMS... LF Console... 7 LF Table... 7 ENGINEERING DATA... 8 PARTS LIST... Model LF-430... 9 Model LF-530... 9 Model LF-630... 10 Model LF-830... 10 PARTS LOCATION... 11 LIMITED WARRANTY... 11 2 Innovators in Clean Air Technology www.envirco.com Operation & Maintenance

WARNING To reduce the risk of fire, electrical shock, or injury to persons, observe the following: Installation work and electrical wiring must be done by qualified person(s) in accordance with all applicable codes and standards, including fire-rated construction. When cutting or drilling into wall or ceiling, do not damage electrical wiring and other hidden utilities. If this unit is to be installed over a tub or shower, it must be marked as appropriate for the application. Use this unit only in the manner intended by the manufacturer. If you have any questions, please contact: ENVIRCO CORPORATION 101 McNeill Road Sanford, NC 27330 Tel: 919.775.2201 Fax: 800.458.2379 Toll Free: 800.884.0002 Email: customerservice@envirco.com Before servicing or cleaning the unit, switch power off at service panel and lock out service panel to prevent power from being switched on accidentally. INTRODUCTION The LF Horizontal Laminar Flow Clean Bench is a completely self-contained, multipurpose, vibration-free clean bench that supplies a laminar flow pattern of HEPA filtered air. The clean bench provides a Class 100 (M 3.5) environment. The unique design concept of the LF Horizontal Laminar Flow Clean Bench consists of separate plenum and table modules that virtually isolates all vibration transmission from the filter/blower console to the work surface. Air enters the LF Horizontal Laminar Flow Clean Bench through a prefilter which removes most of the large particles. The air is then evenly distributed across the High Efficiency Particulate Air (HEPA) filter by blower. The HEPA filter removes 99.99% of all particles 0.3 micros and larger in size and imparts a unidirectional or laminar flow pattern to the air. The air moves through the work area at a uniform velocity of 90 feet per minute (+/-20%), carrying away all airborne contamination generated by activities within the work area. All samples or equipment within the work area are continually bathed in the ultra-clean air. Innovators in Clean Air Technology www.envirco.com Operation & Maintenance 3

FEATURES Fulfills requirements of current Federal Standard 209. Washable polyurethane foam prefilter, easily removable for maintenance. HEPA final filter, 99.99% efficient on particulates 0.3 microns and larger. Designed to use a permanent nonhardening flexible compound sealant in conjunction with a neoprene gasket. Filters are easily accessible through the front of the cabinet. Quiet operation: 65 +/- 3 dba operator s station with a 48 dba ambient. Virtually isolates all vibration transmission from the filter/blower console to the work surface. Dynamically balanced direct drive blower mounted on vibration dampened, shock absorbent rubber mounts with additional resilient cushion-faced mating surfaces. Continuous duty motor(s) with sealed-for-life bearings and built-in thermal overload protection. Solid state speed control provides for output of blower. Three prong grounded electrical plug for standard 115-volt receptacle. Blower and filter plenum is constructed of all welded construction, painted with a non-particulating white polyurethane paint. All seams are sealed. All units are factory tested for leaks, sound level, vibration, airflow, light level, and mechanical function. The LF is a horizontal flow workstation, designed for product protection only. The work surface is constructed of pressed wood laminated on all sides with a non-particulating, non-glare white Formica. (Stainless Steel is optional.) Adjustable leveling feet are located on the workbench. A duplex receptacle is provided on the face of the right hand work surface leg. The blower/motor and light switch are located on front of light canopy for easy access. APPLICATIONS The high-performance, two-piece LF Clean Bench has successfully demonstrated its versatility in a wide variety of applications and industries. Optical Assembly Microscopy Analysis Critical Sample Preparation Pharmaceutical Preparation Tissue Culture Microelectronic Fabrication and Assembly 4 Innovators in Clean Air Technology www.envirco.com Operation & Maintenance

ASSEMBLY 1. The LF comes completely assembled in a single open framed crate. The unit is comprised of two components, the filter/blower console and the workstation. 2. Carefully remove top and four sides of the crate. 3. Remove the protective plastic covering from the unit. 4. Cut and carefully remove binding straps. 5. Remove wooden blocks holding unit in place. 6. Remove unit from shipping skid. 7. Remove prismatic light diffuser and light tubes (placed in a separate box) and insert on underside of light canopy after removing and cardboard packing. 8. Move the rear console unit into its permanent location. 9. Remove prefilter to expose the motor and blower. 10. Remove cardboard shipping blocks in blower* and replace the prefilter. 11. Slide the front workstation in against the rear console. Adjust leveling feet if necessary to maintain alignment of the two components. 12. Insert the male connector of cord on workstation into the female receptacle on the console. 13. Plug power cord into 115 VAC grounded receptacle. 14. Wipe interior surfaces of the work area with a mild disinfectant solution. 15. Turn on unit and let run for a few hours before using as a Class 100 clean environment. Note: If you encounter any problems or need technical assistance, please call 800.884.0002. Unit is factory tested. On-site test and certification is recommended to assure proper operation. CAUTION: FAILURE TO REMOVE THE CARDBOARD SHIPPING BLOCK MAY CAUSE DAMAGE TO THE BLOWER AND/OR MOTOR THAT IS NOT COVERED BY THE LIMITED WARRANTY. SERVICE IMPORTANT: DISCONNECT ELECTRICAL CORD BEFORE ATTEMPTING ANY SERVICE. Fluorescent Lamp Replacement 1. Remove the prismatic diffuser set on the underside of the lamp canopy. 2. Replace the lamps 3. Reassemble by reversing the above procedure. Fluorescent Fixture Ballast Replacement Proceed as above, and then disassemble the fixture by removing the two screws, which hold the plate directly above the lamps. The ballast is then exposed and can be replaced by removing the wire nut connectors and the ballast mounting screws. Install new ballast in reverse order. Innovators in Clean Air Technology www.envirco.com Operation & Maintenance 5

Prefilter Replacement This filter may be washed or vacuumed. If care is taken to keep the prefilter in good condition, long life may be achieved for the HEPA filter. Never operate the unit without a prefilter. 1. Lift the prefilter out. 2. Clean the prefilter. 3. Reinstall cleaned prefilter or install new prefilter. HEPA Filter Replacement Note: Extreme care must be taken when replacing a new filter so as not to puncture media. Filter should be handled by its sides at all times. The life of the HEPA filter depends on prefilter maintenance and the ambient conditions. If the prefilter is clean, the speed control is turned all the way up, and the air drops below 72 FPM, then the filter should be changed. 1. Disconnect the plug from console. 2. Remove the work bench assembly from the front of the console assembly. 3. Lift the perforated air diffuser up and out. 4. Remove the hardware and angle clips, holding the filter in place. 5. Remove the filter by pulling straight out. 6. Clean the sealant from the filter mounting frame. 7. Install new filter by reversing above procedure. Blower Replacement 1. Disconnect power cord. 2. Lift the foam prefilter up and out. 3. Disconnect wires from motor. 4. Remove screws holding blower mounts to vibration isolators. 5. Remove blower from console. 6. Reassemble by reversing the above procedures. Rate of Flow Adjustment 1. Follow steps 1-2 under Prefilter Replacement to expose the speed control. 2. Adjust the speed control knob to obtain desired rate of flow. The solid state speed control is located in the utility box next to the blower. 3. Reassemble by reversing the above procedure. 6 Innovators in Clean Air Technology www.envirco.com Operation & Maintenance

WIRING DIAGRAMS LF Console LF Table Innovators in Clean Air Technology www.envirco.com Operation & Maintenance 7

ENGINEERING DATA Model No. Dimension in Inches (cm) A B C D E F Avg. CFM @ 90 FPM BTU/Hr Power Req d Amps @ 115V # of Motor/ Blowers 1/2 HP Ship Wt. Lbs. (kg) LF-430 51.00 (129.54) 65.00 (165.10) 24.00 (60.96) 44.00 (111.76) 900 3000 7.5 1 260 (117.93) LF-530 63.00 (160.02) 65.00 (165.10) 24.00 (60.96) 44.00 (111.76) 1125 3000 7.5 1 280 (127.01) LF-630 75.00 (190.50) 65.00 (165.10) 24.00 (60.96) 44.00 (111.76) 1350 3000 7.5 1 300 (136.17) LF-830 102.00 (259.08) 65.00 (165.10) 24.00 (60.96) 44.00 (111.76) 1800 6000 15 2 315 (142.88) 8 Innovators in Clean Air Technology www.envirco.com Operation & Maintenance

PARTS LIST Model LF-430 (P/N 10552) Item Description Qty/Unit Part No. A Light Lens Package 1 33500 B Control Panel 1 20230 C Plexiglas Sides 2 33483 D Diffuser Screen 1 33566 E HEPA Filter (30x48x5 7/8) 1 69262-008 F Work Surface 1 33477 G Duplex Electrical Outlet 1 66750 H Motor 1 61657 I Power Cord 1 63042-008 J Solid State Speed Control 1 63982-002 J* Solid State Speed Control 1 S23132-15 K Blower Supports (LH/RH) 2 61812/61813 L Blower 1 62592 M Vibration Isolators 4 62605 N Capacitor 1 61485 O Prefilter (24x33.75) 1 62981-003 P Leveling Feet 4 60440 Q Lamps (T12) 2 61918 Q Lamps (T8) 2 63849-002 R Motor/Blower Assy 1 S23091 Model LF-530 (P/N 10553) Item Description Qty/Unit Part No. A Light Lens Package 1 33501 B Control Panel 1 20230 C Plexiglas Sides 2 33483 D Diffuser Screen 1 33567 E HEPA Filter (30x60x5 7/8) 1 69262-007 F Work Surface 1 33478 G Duplex Electrical Outlet 1 66750 H Motor 1 61657 I Power Cord 1 63042-008 J Solid State Speed Control 1 63982-002 J* Solid State Speed Control 1 S23132-15 K Blower Supports (LH/RH) 2 61812/61813 L Blower 1 62592 M Vibration Isolators 4 62605 N Capacitor 1 61485 O Prefilter (24x33.75) 1 62981-003 P Leveling Feet 4 60440 Q Lamps (T12) 2 61918 Q Lamps (T8) 2 63849-002 R Motor/Blower Assy 1 S23091 * Units built before 06/05 Innovators in Clean Air Technology www.envirco.com Operation & Maintenance 9

Model LF-630 (P/N 10554) Item Description Qty/Unit Part No. A Light Lens Package 1 33502 B Control Panel 1 20230 C Plexiglas Sides 2 33483 D Diffuser Screen 1 33568 E HEPA Filter (30x72x5 7/8) 1 69262-006 F Work Surface 1 33479 G Duplex Electrical Outlet 1 66750 H Motor 1 61657 I Power Cord 1 63042-008 J Solid State Speed Control 1 63982-002 J* Solid State Speed Control 1 S23132-15 K Blower Supports (LH/RH) 2 61812/61813 L Blower 1 62592 M Vibration Isolators 4 62605 N Capacitor 1 61485 O Prefilter (24x47.75) 1 62981-019 P Leveling Feet 4 60440 Q Lamps (T12) 2 61918 Q Lamps (T8) 2 63849-002 R Motor/Blower Assy 1 S23091 Model LF-830 (P/N 10555) Item Description Qty/Unit Part No. A Light Lens Package 1 33503 B Control Panel 1 20230 C Plexiglas Sides 2 33483 D Diffuser Screen 1 33569 E HEPA Filter (30x48x5 7/8) 1 69262-008 F Work Surface 1 33480 G Duplex Electrical Outlet 1 66750 H Motor 1 61657 I Power Cord 1 63042-009 J Solid State Speed Control 1 24201-002 J* Solid State Speed Control 1 23132-25 K Blower Supports (LH/RH) 2 61812/61813 L Blower 1 62592 M Vibration Isolators 4 62605 N Capacitor 1 61485 O Prefilter (24x33.75) 1 62981-003 P Leveling Feet 4 60440 Q Lamps (T12) 2 61918 Q Lamps (T8) 2 63849-002 R Motor/Blower Assy 1 S23091 * Units built before 06/05 10 Innovators in Clean Air Technology www.envirco.com Operation & Maintenance

PARTS LOCATION LIMITED WARRANTY ENVIRCO CORPORATION warrants all equipment manufactured by it and bearing its name, will be free from defects in materials and workmanship normal use. ENVIRCO s obligation under this warranty is to repair or replace any parts, which are defective for a period of one year after invoice date, provided that ENVIRCO receives written notice of such defect. For a period of 90 days after invoice date, ENVIRCO may effect such repairs or replacement, via ENVIRCO Service Personnel, at the equipment s installation site provided that the equipment is located in the continental United States and that ENVIRCO receives written notice of such defect. In all other cases, Buyer agrees to assume any incidental expenses, including, but not limited to, the cost of transporting the defective equipment to ENVIRCO s repair facilities. It is expressly agreed and understood that repair or replacement of defective parts is Buyer s exclusive remedy under this warranty and that incidental and consequential damages may not be recovered. Excluded from this warranty are certain expendable items such as light tubes, filters, etc., as well as damage due to abuse or accident. Except for this warranty, ENVIRCO makes no warranty, express or implied, including but not limited to, the Warranty of Merchantability, or the Warranty of Fitness for a Particular Purpose. ENVIRCO is not responsible for any warranties made by another manufacturer. Innovators in Clean Air Technology www.envirco.com Operation & Maintenance 11

Clean air solutions built for you. MORE INFORMATION AVAILABLE AT W W W. E N V I R C O. C O M Industrial Hospital & Healthcare Pharmaceutical Laboratory & Research Cleanroom Products Hospital & Healthcare Pharmaceutical & Laboratory & Research»» MAC 10 Original»» IsoClean and IsoClean Medical Device»» 100-Plus Horizontal or»» MAC 10 IQ with Ultraviolet Light»» Unimodule M2 Vertical Laminar Flow Workstation»» MAC 10 IQ LAF»» AirCeil»» MAC 10 LEAC»» Hospi-Gard Room»» MAC 10 LEAC2 Pressure Monitor Vertical Flow Workstation»» MAC 10 LEDC Enviramedic Products»» Laminar Downflow Module»» Ducted Ceiling Module:»» HOR Horizontal Flow DCM & RSR»» AC or DC Control Systems»» MAC-T Ceiling Grid System Enclosure»» METD 100% Exhausted Vertical Flow Clean Bench»» Unidirectional Flow Horizontal Flow Bench (LF)»» TT Table Top Horizontal Flow Clean Bench»» EnviraLab Sterility Module: ESM»» HCF Horizontal Flow Surgery Room»» VOR Vertical Flow Surgery Isolator ENVIRCO MAC 10 is a registered trademark of ENVIRCO, USA. Specifications subject to modifications or changes without notice. US patents 4,560,395 and 5,470,363. Other patents issued and pending in foreign countries. ENVIRCO 101 McNeill Road Sanford, NC 27330, USA tel: 919.775.2201 toll free: 800.884.0002 fax: 800.458.2379 www.envirco.com email: info@envirco.com Copyright 2017 ENVIRCO. All Rights Reserved. Form P/N#: 63959 Rev. 02/17