Leica HI1210 Water Bath

Similar documents
Leica EG1150 C. Cold Plate

Instructions for Use

Leica EG1150 H. Paraffin Embedding Station. Instructions for Use

HistoCore Arcadia H. Paraffin Embedding Station. Instructions for Use

VT1000 S Vibrating-blade microtome

Register your instrument! HeatSealer S100. Operating manual

Paraffin Embedding Station

User s Manual Power Supply. IM E 2nd Edition. Yokogawa Electric Corporation

HEINE NT4 MED V

Catalyser Introduction. 3. Hazard Information. 2. Area of Application. 3.1 Symbology

DC Master 24/ A

User s Manual Power Supply IM E. 5th Edition

RT Series Step Down Transformer for RT Series UPS 6-10kVA UL Input Vac Output Vac User Guide

OPERATIONS MANUAL CR Thermoreactor

Operating manual. High-voltage testing adapter VAS 6558A/30. Operating manual. Customer documentation Revision: 00 Version: 07/2018

Tissue Master. User Manual

18VDC ESB6 Series Cordless Screwdrivers Operation Manual

Tension Meter. Edition FT 03.E. FT Series. Instruction Manual. Valid as of: Please keep the manual for future reference!

Video & Power Over Coax Kit

EN Operating manual. Motorised zone valve. Three-way, 22 mm & 28 mm 3PV2, 3PV8 & VRMH3

HV angle valve with single acting pneumatic actuator and closing spring (NC)

Megohmmeter Test Probe

miconverter 18-Module Rack-Mount Power Chassis User Manual 38 Tesla, Irvine, CA USA Phone: (949) ; Fax: (949) Page 12

Installation, Operating & Maintenance Instructions. HV angle valve with pneumatic actuator single acting with closing spring (NC)

Contents. Introduction Inspection Notes on Safety Precautions. Chapter 1 Overview Product Overview Names of Parts 2

Operating manual. original operating manual. HDA eco Box 12/24V DC Automatic Dispenser. Translation of the. Item No.:

OPERATORS MANUAL FOR USE WITH MAGNETIC/HEATED STIRRER MG200

NANOPAC-300 & 500 Power Supply. Instruction manual NANOPAC-300 & NANOPAC-500

SCOPELITE TIMING LIGHT OPERATING MANUAL

Katalysator Nr /

Mod: KLD6-12/35XLAS-N

UHV Angle valve (Series 284) with pneumatic actuator, single acting with closing spring (NC)

Installation, Operating & Maintenance Instructions. UHV gate valve with pneumatic actuator. Series 108 DN mm (I. D. 2½ 8 )

Operating instructions

Installation, Operating & Maintenance Instructions. HV gate valve with pneumatic actuator. Series 110 DN mm (I. D.

Rental Industry Safety Tester Safe Check 5s

BroadBand PowerShield. 20 AHr Battery. User Manual

Assembly and Maintenance Manual Type ASNU

Installation, Operating & Maintenance Instructions. All-metal gate valve with compact or extended pneumatic actuator

User s Manual Power Supply. IM E 3rd Edition. Yokogawa Electric Corporation

MV Series Motors Operation & Parts Manual

362 Clamp Meter. Users Manual

BK BK A BK MA BK MA BK J93 MN93-BK MN193-BK MR193-BK SL261-BK SR193-BK

Pressure relief valve

Operator's Manual. Storage System. Ultrasound Probe Cabinet. Manufactured by:

XPC-EBP64 External Battery Pack User & Installation Manual Xtreme Power Conversion Corporation. All rights reserved.

Armon Edero. User manual

MV Series Motors Operation & Parts Manual. Models M3V, M5V, M5V-US

Podiatry Procedures Chair. For Models: Barrier-Free. User s Guide

How to get started. with. Kongsberg XP Auto systems. Serial number: Please Read This First

i400s Calibration Information Introduction

Rig Master Power by Mobile Thermo Systems Inc.

Bosch Smart Home. Door/Window Contact Instruction Manual

SPARKSCAN1 HIGH VOLTAGE CLAMP OPERATING MANUAL

Phase Sequence Tester

Invacare. Battery Charger User Manual. Model: POC1 115, POC1 115 AZ, POC1 115 C, POC1 115 EU

EX Series User Manual

18VDC ESB6-X Series Cordless Screwdrivers Operation Manual

Installation, Operating & Maintenance Instructions. Series 081, DN (I.D. 2½" - 4") Hot surfaces; do not touch!

BAK1500 INSTALLATION/OWNER'S MANUAL Compact Amplified Subwoofer

VP-4124/VP-4124-E 24/48 VOLT DC SWITCHING POWER SUPPLY

Installation, Operating & Maintenance Instructions. Hot surfaces; do not touch!

UZK12.261, UZO GENERAL DESCRIPTION 2. SHORT-FORM DATA 3. ORDER NUMBERS 4. MARKINGS BATTERY MODULE BATTERY MODULE 12V, 26AH 1/10.

Switching DC Power Supply

650mA Phase Dimmable LED Driver

OPERATION AND MAINTENANCE

PY-11 Voltage & Continuity Tester

Type Operating Instructions. Bedienungsanleitung Manuel d utilisation. 2/2-Way Solenoid Valve 2/2-Wege-Magnetventil Électrovanne à 2/2 voies

EPS 16 ATEX 1121 X, IECEx EPS X Solenoid coil Type 06xx Magnetspule Typ 06xx Bobine magnétique Type 06xx. Operating Instructions

Assembly and Maintenance Manual Type AS

HV Angle valve (Series 264) HV Inline valve (Series 265) with pneumatic actuator, single acting with closing spring (NC)

RENA AF371Feeder Operating Manual. Feeder. Operating Manual. Manual Part #: M AF371 Operations Rev

BT403. A Geno Technology, Inc. (USA) brand name. BT-300 Power Supply. Cat. No. BT

FUNNEL CAKE FRYER Instruction Manual Models: 8078 / FC-4, 8082 / FC-6, and 8090 / FC-4

HV-Angle Valve (Series 264) HV-InlineValve (Series 265) with electromagnetic actuator single acting with closing spring (NC)

TBX10A INSTALLATION/OWNER'S MANUAL 10" Sealed Enclosure with Built-in Amplifier

Smart-UPS RT External Battery Pack Stack/Rack-Mount 6U

DALI 650mA LED Driver

Operating Instructions for PAC800 Battery Charger

WELDING INVERTER. PEGAS 160 E Smart PEGAS 200 E Smart OPERATING MANUAL. ALFA IN a.s. PEGAS E Smart Manual EN 04

3. Operating instructions: Minor 200

900 PEAK AMP PORTABLE JUMP STARTER

PL900 ILLUMINATOR. Fiber-Lite. Operation Manual. Setup. Fiber Optic Connection A PRODUCT OF DOLAN-JENNER INDUSTRIES. Voltage Selection:

1000mA - 650mA Phase Dimmable LED Driver

Exchange of rollers from the XTS-Mover

RS-3 PRO RS-1007 PRO. CAT IV Analog Clamp meter Series. Users Manual. For detailed specifications and ordering info go to

PRM-4. Phase Sequence and Motor Rotation Tester. Users Manual. For detailed specifications and ordering info go to

OPERATING INSTRUCTIONS

SAFETY INFORMATION & INSTALLATION GUIDE RMX Programmable Power Supplies. About the RMX Safety Information & Installation Guide

EVS RP6020. Instruction Manual

DUSTTRAK AEROSOL MONITOR SOLAR POWER KIT MODEL

System 3 ZB1PS Device Chassis. Operator s Manual

Fiber Optic Illuminator. Model 21AC Model 21DC

Type Operating Instructions. Bedienungsanleitung Manuel d utilisation. 2/2-Way Solenoid Valve 2/2-Wege-Magnetventil Électrovanne à 2/2 voies

EPS 16 ATEX 1072 X, IECEx EPS X. Operating Instructions. Solenoid coil Type AC19 Magnetspule Typ AC19 Bobine magnétique Type AC19

02OWNER S MANUAL 06 TROUBLESHOOTING 08LIMITED WARRANTY

Lab Rotators and Low Cost Orbital Shakers

Broadband PowerShield. External Battery Pack 20 Ah 12 V Battery. User Manual

Medimachine II. Dissociator and Homogenizator USER MANUAL. (code )

RP Instruction Manual

Transcription:

Instructions for Use Leica HI1210 Water Bath Leica HI1210 V 3.3, English 07/2016 Order No.: 14 0415 80101 RevG Always keep this manual with the instrument. Read carefully before working with the instrument.

IMPORTANT NOTE The information, numerical data, notes and value judgments contained in this manual represent the current state of scientific knowledge and state-of-the-art technology as we understand it following thorough investigation in this field. We are under no obligation to update the present manual periodically and on an ongoing basis according to the latest technical developments, nor to provide our customers with additional copies, updates etc. of this manual. To the extent permitted in accordance with the national legal system as applicable in each individual case, we shall not be held liable for erroneous statements, drawings, technical illustrations etc. contained in this manual. In particular, no liability whatsoever is accepted for any financial loss or consequential damage caused by or related to compliance with statements or other information in this manual. Statements, drawings, illustrations and other information regarding the contents or technical details of the present Instructions for Use are not to be considered warranted characteristics of our products. These are determined only by the contract provisions agreed between ourselves and our customers. Leica reserves the right to change technical specifications as well as manufacturing processes without prior notice. Only in this way is it possible to continuously improve the technology and manufacturing techniques used in our products. This document is protected under copyright laws. All copyrights to this documentation are held by Leica Biosystems Nussloch GmbH. Any reproduction of text and illustrations (or of any parts thereof) by means of print, photocopy, microfiche, web cam or other methods including any electronic systems and media requires express prior permission in writing by Leica Biosystems Nussloch GmbH. For the instrument serial number and year of manufacture, please refer to the nameplate on the back of the instrument. Leica Biosystems Nussloch GmbH Leica Biosystems Nussloch GmbH Heidelberger Str. 17-19 D-69226 Nussloch Germany Phone: +49 (0)62 24 143-0 Fax: +49 (0)6224 143-268 Internet: http://www.leicabiosystems.com Leica HI 1210 Water Bath 3

Table of Contents 1. Important Information...5 1.1 Symbols and their meanings... 5 1.2 Qualification of personnel... 6 1.3 Intended use of instrument... 6 1.4 Instrument type... 6 2. Safety...7 2.1 Safety notes... 7 2.2 Warnings... 8 3. Instrument Components and Specifications...9 3.1 Technical data... 9 4. Instrument Setup...10 4.1 Location conditions... 10 4.2 Standard delivery... 10 4.3 Unpacking and setting up the HI1210... 11 4.4 Setting up the power supply... 11 5. Operation...12 5.1 Control panel fields on the instrument... 12 5.2 Switching the instrument on... 12 5.3 Setting the temperature... 13 5.4 Switching off the instrument... 13 6. Cleaning and Maintenance...14 6.1 Cleaning the instrument... 14 6.2 Replacing fuses... 14 7. Warranty and Service...15 8. Decontamination Certificate (Master)...16 4 Instructions for Use V 3.3 RevG 07/2016

1. Important Information 1.1 Symbols and their meanings Warnings appear in a gray box and are marked by a warning triangle. Notes, i.e. important user information, appear in a gray box and are marked by an information symbol. (5) Numbers and parentheses refer to item numbers in the illustrations. START Function keys to be pressed on the instrument touch screen are in boldprint capital letters. Instrument surfaces which become hot during operation are marked with this symbol. Avoid direct contact with these surfaces - they may cause burns. Environmental protection symbol of the China RoHS directive. The number in the symbol indicates the "Environment-friendly Use Period" of the product. The symbol is used if a substance restricted in China is used in excess of the maximum permitted limit. Symbol for labeling electrical and electronic equipment in accordance with Section 7 of the German Electrical and Electronic Equipment Act (ElektroG). ElektroG is the law regarding the sale, return and environmentally sound disposal of electrical and electronic equipment. The package contents are fragile and must be handled with care. The package must be kept in a dry environment. Indicates the correct upright position of the package. It allows maximum 3 stacks layers. Manufacturer Date of manufacture Indicates the temperature range permitted for transporting the package. Minimum -29 C, Maximum +50 C SN REF Observe the instructions for use! Serial number Order no. The CE labeling shows that the product corresponds to one or more applicable European directives. Caution, consult the instructions for use for cautionary information. Indicates the temperature range permitted for storing the package. Minimum +5 C, Maximum +50 C Indicates the humidity range permitted for storing and transporting the package. Minimum 10 % r.h., Maximum 85 % r.h Indicates the item can be recycled where correct facilities exist. Leica HI 1210 Water Bath 5

1. Important Information The Regulatory Compliance Mark (RCM) indicates a device s compliance with applicable ACMA technical standards of New Zealand and Australia - that is, for telecommunications, radio communications, EMC and EME. 1.2 Qualification of personnel 1.4 Instrument type All information provided in these Instructions for Use applies only to the instrument type indicated on the cover page. A nameplate indicating the instrument serial number is attached to the rear side of the instrument. The Leica HI1210 may be operated by trained laboratory personnel only. All laboratory personnel designated to operate this instrument must read these Instructions for Use carefully and must be familiar with all technical features of the instrument before attempting to operate it. 1.3 Intended use of instrument The Leica HI1210 is a paraffin flotation bath for flattening and drying cut tissue samples used in all fields of biomedical research and routine diagnostics. It may only be used by Leica Biosystems Nussloch GmbH for operating approved instrument accessories. Any use of the instrument other than its designated use is considered improper. Failure to adhere to these instructions may result in an accident, personal injury, damage to the instrument or accessory equipment. Proper and intended use includes compliance with all inspection and maintenance instructions, along with the observance of all instructions in the Instructions for Use. Fig. 1 Fig. 1 is provided as an example only and shows a valid nameplate for this instrument with the necessary information about instrument type and power requirement. The precise data for the various versions is specified in Chapter 3.1 "Technical data". 6 Instructions for Use V 3.3 RevG 07/2016

2. Safety 2.1 Safety notes The safety and caution notes in this chapter must be observed at all times. Be sure to read these notes even if you are already familiar with the operation and use of other Leica products. The protective devices located on the instrument and the accessories must not be removed or modified. Only qualified service personnel authorized by Leica may repair the instrument and access its internal components. Residual risks The instrument has been designed and constructed with the latest state-of-the-art technology and according to recognized standards and regulations with regard to safety technology. Operating or handling the instrument incorrectly can place the user or other personnel at risk of injury or can cause damage to the instrument or other property. The instrument may be used only as intended and only if all of its safety features are in proper working condition. Malfunctions which could impede safety must be remedied immediately. Only original spare parts and permitted original accessories may be used. The instrument MUST be connected to a grounded power socket. Use only a provided power cable that is intended for the local power supply. These Instructions for Use include important instructions and information related to the operating safety and maintenance of the instrument. The Instructions for Use are an important part of the product, and must be read carefully prior to startup and use and must always be kept near the instrument. These Instructions for Use must be appropriately supplemented as required by the existing regulations on accident prevention and environmental safety in the operator's country. This instrument has been built and tested in accordance with the safety requirements for electrical equipment for measurement, control, and laboratory use. To maintain this condition and ensure safe operation, the user must observe all notes and warnings contained in these Instructions for Use. The instrument's CE certificate can be found on the Internet at: http://www.leicabiosystems.com Leica HI 1210 Water Bath 7

2. Safety 2.1 Safety notes (continued) In order to properly service the instrument, it is indispensable to read the User Manual before switching on the instrument and to familiarize yourself with all technical details. Before connecting the instrument to the line voltage, ensure that the electrical power requirements of your laboratory match the values on the nameplate of the instrument. When installing the power cable, always be sure to route it so that it cannot contact the heated surfaces of the instrument at any time. The instrument is designed for indoor use only. The instrument must be switched off and unplugged from the power supply during all repair and service work. During operation, the instrument surface can be very hot. Risk of injury! Do not store any combustible and flammable substances near the instrument. The instrument may be operated with non-flammable liquids (preferably distilled water) only! Operate the instrument in filled condition only. Minimal filling level: 1 cm 2.2 Warnings The safety devices installed in this instrument by the manufacturer only constitute the basis for accident prevention. Operating the instrument safely is, above all, the responsibility of the owner, as well as the designated personnel who operate, service or repair the instrument. To ensure trouble-free operation of the instrument, make sure to comply with the following instructions and warnings. Please note that electrostatic charge may result through direct or indirect contact with the Leica HI1210. 8 Instructions for Use V 3.3 RevG 07/2016

3. Instrument Components and Specifications 3.1 Technical data Nominal supply voltages: Nominal frequency: Nominal power: Main fuse: Dimensions: Empty weight: Relative humidity: Two factory-preset voltages (not user-adjustable): 100-120 VAC + 10 % 230-240 VAC + 10 % 50/60 Hz 350 VA 2 melting fuses, 5x20 mm, UL-approved For 100-120 V: 5x20 mm, 2x T 5A L250 VAC For 230-240 V: 5x20 mm, 2x T 2,5A L250 VAC 350x310x100 mm 3.6 kg (w/o packaging) 20-80 %, non-condensing IEC 1010 classification: Protection class: 1 Pollution degree: 2 Overvoltage category: II Operating temperature range: +15 C to +40 C Temperature control range: Room temperature to 75 C Operating elevation: IP protection class (IEC 60529): maximum of 2000 m above sea level IP20 Leica HI 1210 Water Bath 9

4. Instrument Setup 4.1 Location conditions When installing the power cable, always be sure to route it so that it cannot contact the heated surfaces of the instrument at any time. The power socket to which the instrument is to be connected to has to be located near the instrument and easily accessible. The power supply must be at a distance no greater than the length of the power cable an extension cable must NOT be used. The substrate must be largely vibration-free and have sufficient load capacity and rigidity for the weight of the instrument. Avoid impacts, bright direct light, and excessive temperature fluctuations. The instrument must be connected to a suitable power socket.use only the provided power cable, which is intended for the local power supply. 4.2 Standard delivery The Leica HI1210 standard delivery includes: 1 Leica HI1210 basic instrument 1 Dust cover... 14 0415 41178 1 Set of fuses: 2 fuses, T2.5A... 14 6943 02500 2 fuses, T5A... 14 6943 05000 1 Instructions for Use, printed (German/English, with Language CD 14 0415 80200)... 14 0415 80001 The country specific power cord needs to be ordered separately. Please find a list of all power cords available for your device on our website www.leicabiosystems.com within the product section. Check the delivery carefully against the packing list, delivery note and your order. Should you find any discrepancies, please contact your Leica sales office without delay. 10 Instructions for Use V 3.3 RevG 07/2016

4. Instrument Setup 4.3 Unpacking and setting up the HI1210 Set the carton on a flat surface, cut through the adhesive tape (Fig. 2) and open the carton (Fig. 3). Fig. 2 Remove the accessories and carefully lift the instrument out of the carton (Fig. 4). Fig. 3 Hold the instrument firmly with one hand, pull off both foam sections (Fig. 5) and remove the dust cover. õõ 1 Fig. 4 2 Fig. 5 4.4 Setting up the power supply Before connecting the power cable, make sure that the power switch (3) on the rear of the instrument (Fig. 6) is switched to"0" = OFF. Use the instrument with the provided power cable only. Insert the connector of the power cable (Fig. 5, 2) into the connection socket (4) and plug the power plug into the power socket. 3 4 Fig. 6 Leica HI 1210 Water Bath 11

5. Operation 5.1 Control panel fields on the instrument 5 = POWER LED 6 = SET Is illuminated when the instrument is switched on with the main power switch. Indicates the selected target temperature. 7 = Set key: lower temperature. 8 = Set key: increase temperature. 9 = RUN/STOP 10 = Display Switch instrument on or off. 5 6 7 8 9 10 Fig. 7 5.2 Switching the instrument on Before switching the instrument on, fill the water bath with a sufficient quantity of distilled water. Switch on the instrument with the power switch on the left rear panel of the instrument (Fig. 6, 3). The green LED display POWER on the front panel is then illuminated. 12 Instructions for Use V 3.3 RevG 07/2016

5. Operation 5.3 Setting the temperature Pressing the RUN/STOP button on the front panel brings the temperature of the instrument to the latest stored target value. The displayed temperature corresponds to the current actual value. The set target temperature is shown in the display (Fig. 7, 10). To adjust to the desired temperature, hold down the corresponding set key (Fig. 7, 7+8). The display changes slowly for the first 8 digital steps, then more quickly after that. The temperature display refers to the target value. After the desired temperature is reached, release the set key. The displayed value is saved automatically. This stored target value will continue to be displayed for about 2 seconds, then the current actual temperature display will appear again. The target temperature can be checked by pressing the SET key (Fig. 7, 6). The target value remains saved both when the instrument is switched off (RUN/STOP key or power switch) and in case of power failure and disconnection from the power supply. 5.4 Switching off the instrument The instrument can be switched off by pressing the RUN/STOP key. You do not have to switch the main power switch on or off each time for everyday routine use. Leica HI 1210 Water Bath 13

õ 6. Cleaning and Maintenance 6.1 Cleaning the instrument Before cleaning the instrument, turn off using the power switch (Fig. 6, 3) and disconnect from the power supply. The instrument has a continuous front panel and no lateral cooling slots, so the surfaces are smooth and easy to clean. All surface materials are resistant to common laboratory detergents. The instrument must be switched off and have cooled down for cleaning. Do not use organic solvents (e.g. alcohol or xylene)! 6.2 Replacing fuses Switch the unit off and unplug it. Use only the supplied replacement fuses. Both fuses must have the same rating (check the imprint!). Fuses must always be switched out in pairs. Insert a small screwdriver into the cutout (Fig. 8, 11) and carefully pry out the fuse insert (12). Remove the fuse holder and replace the defective fuses with two new ones (13). Then reinsert the fuse holder and press it into place. 13 11 12 Fig. 8 14 Instructions for Use V 3.3 RevG 07/2016

7. Warranty and Service Warranty Service information Leica Biosystems Nussloch GmbH guarantees that the contractual product delivered has been subjected to a comprehensive quality control procedure based on the Leica in-house testing standards, and that the product is faultless and complies with all technical specifications and/ or characteristics warranted. The scope of the warranty is based on the content of the concluded agreement. The warranty terms of your Leica sales organization or the organization from which you have purchased the contractual product shall apply exclusively. If you require technical service or replacement parts, please contact your Leica sales office or dealer who sold the product. The following inputs are required: Model name and serial number of the instrument. Location of the instrument and name of the person to contact. Reason for the service call. Date of delivery. Shutdown and disposal of the instrument The instrument or parts of the instrument must be disposed of in compliance with the local laws. Special attention should be paid to the lithium cell of the electronic circuit board! Leica HI 1210 Water Bath 15

8. Decontamination Certificate (Master) Dear Customer, Any product that is to be returned to Leica Biosystems or serviced on site, must be cleaned and decontaminated in the appropriate manner. Since it is not possible to decontaminate for prion diseases, such as CJD, BSE, CWD etc., equipment exposed to specimens containing prion diseases cannot be returned to Leica Biosystems for repair. On-site repair of prion contaminated equipment will only be conducted after the Field Service Engineer has been educated in the risks, instructed in the policies and procedures of the institution, and provided with personal protective equipment. Please fill out this confirmation carefully and enclose a copy with the instrument. Attach the confirmation to the outside of the flight case or hand it directly to the service technician. Packages will not be opened, nor servicing commenced until the Company or service engineer have received a satisfactory certificate. Should returned goods be considered a hazard by the Company, they will be returned immediately to the customer at his/her expense. Note: Microtome knives must be in boxes. Mandatory information: Fields marked with * are mandatory. Depending on whether the instrument is contaminated, please also complete either section A or section B. Nameplate information Model (see nameplate)* SN (see nameplate)* REF (see nameplate)* A B Yes Yes No 1 Tick Box A if applicable. Otherwise please complete all parts of B, providing further information as requested or appropriate. This equipment has not been in contact with unfixed biological samples. This equipment has been exposed internally or externally to hazardous materials as indicated below: Please provide further detail here: Blood, body fluids, pathological samples Other biohazards Chemicals/substances hazardous to health Other hazards Yes No 2 Radioactivity This equipment has been cleaned and decontaminated: If yes, give details of the method: Please provide further detail here: If no**, please indicate why not: Yes No 3 ** Such equipment must not be returned without the written agreement of Leica Biosystems. The equipment has been prepared to ensure safe handling/transportation. Whenever possible, please use the original transportation case/box. Important - to avoid refusal of shipment: 16 Instructions for Use V 3.3 RevG 07/2016

8. Decontamination Certificate (Master) Important - to avoid refusal of shipment: Place one copy in the unit prior to packaging, or hand it over to the service engineer. Customer assumes all responsibility for the immediate return shipment of articles sent to Leica without proper decontamination documentation. If you have any further questions, please call your local Leica organization. Leica Internal Use: If applicable, note corresponding Job and RAN-/RGA-Number: Job Sheet No.: BU Return Authorization Number: SU Return Goods Authorization: Signature/Date* Institute* Name* Position* Department* Address* email Phone* Fax Leica Biosystems Nussloch GmbH Heidelberger Str. 17-19 69226 Nussloch, Germany Phone: ++49 (0) 6224 143 0 Fax: ++49 (0) 6224 143 268 www.leicabiosystems.com Leica HI 1210 Water Bath 17

Notes 18 Instructions for Use V 3.3 RevG 07/2016

www.leicabiosystems.com Leica Biosystems Nussloch GmbH Heidelberger Straße 17-19 D- 69226 Nussloch Phone: +49 - (0) 6224-143 0 Fax: +49 - (0) 6224-143 268 Web: www.leicabiosystems.com