INSTRUCTION MANUAL. VARIABLE AREA FLOWMETERS 10A2235 Rev.5 RATOSIGHT FLOW RATE INDICATORS

Similar documents
ABB Automation. Variable Area Dial Flow Indicator Series 10A2227. Series 10A2227 Variable Area Dial Flow Indicator

INSTRUCTION MANUAL. with Parts List FLOW PRODUCTS 10A4500/10A4600 DESIGN LEVEL B V/A MASTER FLOWMETERS PN ABB Instrumentation

Variable Area Flowmeters Ori-Flowrator TM Meters

INSTRUCTION MANUAL. includes Parts List FLOW PRODUCTS 10A6100 DESIGN LEVEL B PURGEMASTER FLOWMETERS PN25104

Variable Area Flowmeters VA Master TM Indicating Flowrator Meters 10A4500

INSTRUCTION MANUAL. INTEGRAL ORIFICE FLOW ELEMENT with Upstream & Downstream Piping and Threaded Connections. 1330L Series, Model A IB-04H104_8

ABB Instrumentation. 50T Series Transmitters Model 54G/A gauge/absolute pressure transmitter for sanitary use. Specification sheet

Model Connections Max. Flow Range (H 2 O) 10A A3220 Horizontal ¼ NPT 100 l/h. 10A3225/55 Vertical ¼ NPT 100 l/h. Vertical 1 NPT Vertical 1 G

ANDERSON GREENWOOD SERIES 500 PILOT OPERATED SAFETY RELIEF VALVES INSTALLATION AND MAINTENANCE INSTRUCTIONS

Shut-off valves and accessories for pressure measuring instruments

Vortex Flowmeter FV4000 (TRIO-WIRL V) Swirl Flowmeter FS4000 (TRIO-WIRL S) For flow rate and volume measurement of liquids, gases and steam

I & M Mark 708ME. Ideal Installation. Start-Up Procedure. Installation & Maintenance Instructions for Mark 708 & Motor Actuator

User Guide IM/TORBAR-EN Rev. D. Averaging pitot tubes

COMPANY. Pneumatic Flow Meters Installation, Operating & Maintenance Manual AW-Lake Company. All rights reserved. Doc ID:PNEUMAN082516

ABB measurement & analytics data sheet. Model M26 Manifold line for 266 pressure transmitters

TUBE FLOW METERS. design features INTERCHANGEABLE. 150 mm Meter with CV TM Valve

DIGIFLO ELECTRONIC FLOWMETER

BYPASS ROTAMETERS FOR MEASURING RATE OF FLOW IN LARGE PIPELINES

2.- HANDLING OF VALVES BEFORE ASSEMBLY 3.- FITTING THE VALVE TO THE REST OF THE ASSEMBLY 5.- PERIODICAL INSPECTION OF THE VALVE AND MAINTENANCE

OPERATION MANUAL. ALUMINUM Models. 316 S.S. Models NTG25 NOMAD TRANS-FLO AIR-OPERATED DOUBLE DIAPHRAGM PUMPS. A JDA Global Company. 1/14 rev.

I & M Mark 708. Ideal Installation. Start-Up Procedure. Installation & Maintenance Instructions for Mark 708 & 14M Actuator and Motor Valve

Field IT Actuators & Positioners. Intelligent Electric Modulating Actuator. Contrac RHD Intelligent Electric Modulating Actuator

Navigator ADS551 Low level dissolved oxygen Replacement / Upgrade procedures wet section spares

Model WEP Electro-pneumatic positioner. WEP positioner A valve control without compromise

KEYSTONE Figure 990/991 Butterfly valves Installation, operation and maintenance instructions

KOBOLD KSK FLOWMETER. User Instructions

OPERATION MANUAL NT50 NOMAD TRANS-FLO AIR-OPERATED DOUBLE DIAPHRAGM PUMPS. A JDA Global Company. 7/11 rev. 1

Rotameters Variable Area Flowmeters

WP Series Hydraulic & Pneumatic Flow Monitors Illustrated Price List

KOBOLD VKA FLOWMETER/SWITCH. User Instructions. Manual-VKA_03-04

KOBOLD DF-MA Flow Transmitter. User Instructions

INSTALLATION, OPERATION AND MAINTENANCE MANUAL (IOM)

FAIRCHILD T5420 DIGITAL-PNEUMATIC TRANSDUCER Installation, Operation and Maintenance Instructions

KOBOLD SV FLOWMETER/SWITCH. User Instructions

introduction < STANDARDS >

Actuators & Positioners Electro-Pneumatic Converters I/P Signal Converters - TEIP 11 Model without Booster Stage

Rosemount 402 and 402VP

Product Manual. S&S V-100 Ball Valves 2 though 8 Inch Designs. Introduction: Installation:

7430 Series Hastelloy / Monel / Stainless Steel Installation Instructions

INSPECTION & MAINTENANCE BULLETIN ARI Plug Type Angle Valves

Fisher D2T FloPro Control Valve

Installation Instructions

Type T205B Balanced Tank Blanketing Regulator

WPS500X 500 psi /34.5 bar Automotive Pressure Transducer. User s Guide

ABB MEASUREMENT & ANALYTICS DATA SHEET. TEIP11 I/P signal converter for standard signals

Valtek Flow Boosters

Product Manual. CVS V-100 Ball Valves 2 though 8 Inch Designs. Introduction: Installation:

ABB Instrumentation. Thermal Mass Flowmeter Sensyflow. Intelligent solutions for accurate gas mass flow measurement

MCV106A. Hydraulic Displacment Control-PV DESCRIPTION FEATURES ORDERING INFORMATION. BLN Issued: March 1991

Hydraulic Motors Repair Instructions

Introduction. Installation. I & M Mark V-100 Series. Installation & Maintenance Instructions for the Mark V-100 Series Control Valves

VSI INDUSTRIAL BALL VALVES SERIES 7100 SERIES 8100 SERIES 7200 SERIES 8300 SERIES 7400 SERIES 8400 SERIES 8000 SERIES 8500

Technical Data. Flow Meter Schwertberg, Poneggenstr. 5 Austria/Europe Fax +43/7262/ Tel.

PROMAG SR SERIES SEAL-LESS CENTRIFUGAL PUMPS

WESTATE DIESEL SYSTEMS

A-4000 Series Oil Removal and Pressure-Reducing Stations and Air Compressor Accessories

OPERATION MANUAL NT80 NOMAD TRANS-FLO AIR-OPERATED DOUBLE DIAPHRAGM PUMPS. A JDA Global Company. 5/13 rev.2

Maintenance Manual WATER SUMP CONTROL VALVE F532B

Endura AZ10 Oxygen probe Sensor sub-assembly tested spares kit

Welker Sampler. Installation, Operation, & Maintenance Manual. Model GSS-4HP

Variable Area Flowmeters Armored Purgemeter

ANDERSON GREENWOOD. Before installation these instructions must be fully read and understood.

COMPANY. Liquid Flow Meters Installation, Operating & Maintenance Manual AW-Lake Company. All rights reserved. Doc ID:FLOWLIQUIDMETERMAN082516

Rotameters Variable Area Flowmeters

Bottom Loading & Recirculation Adapters

The perfect solution for wellhead flow measurement with CoriolisMaster, ProcessMaster and VortexMaster

WPS500X. 500 psi / 34.5 bar Automotive Pressure Transducer. User s Guide

VA-908x Series Electric Rotary Actuators for Two-Position and Modulating Service

SUPPLEMENTAL MANUAL FOR THE INSTALLATION, OPERATION AND MAINTENANCE OF THE XL234 FILTER VESSEL

OBSOLETE DOCUMENT. PRX Series Pilots for Pilot-Operated Pressure Reducing Regulators. PRX Series. Introduction. Product Description

Flow Fittings. for Electrochemical Sensors. B Operating Manual /

TECHNICAL SERVICE MANUAL

SM Series. High Pressure - All-Metal Flowmeter and Switch

INSTRUCTION MANUAL MODEL 213, 214, 215 and 216 POSITIVE DISPLACEMENT FLOWMETERS

8-Position Cat5E EMT Shielded Slim-Line Modular Plug

Flow Fittings. for Electrochemical Sensors. B Operating Manual /

Installation, Operation, and Maintenance Manual

ZDM Positive Displacement Flow Meter User Instructions

IMPORTANT OPERATING CONDITIONS. Failure to comply with any of these conditions invalidates the warranty. STANDARD CONFIGURATIONS

Model 8329 Table of Contents

Air Trap TATSU2. Copyright 2013 by TLV CO., LTD. All rights reserved ISO 9001/ ISO M-02 (TATSU2) 7 August 2013.

SPANS AND RANGES. Lower Range Limits 150 in. H 2 O ( 38 kpa) 800 in. H 2 O ( 200 kpa)

Maintenance Manual. Hydrant Coupler. F250 Series

Types 1808 and 1808A Pilot-Operated Relief Valves or Backpressure Regulators

I & M 8000 Series. Ideal Installation Schematic. Preferred Installation. Trouble Shooting

Product Manual. S&S V-100 Ball Valves 2 through 12 Inch Designs. Introduction: Installation:

INSTRUCTION MANUAL AND PARTS LIST FOR SERIES 8L-630J AND 630M WARNING

FGM1190 Glass Tube Variable Area Flowmeters

Technical Data TYPE T124 & T134 TEMPERATURE PILOT SPENCE ENGINEERING COMPANY, INC. 150 COLDENHAM ROAD, WALDEN, NY SD 4512A

A-4000 Series Oil Removal and Pressure Reducing Stations & Air Compressor Accessories

WPS600. Hydraulic Pressure Transducer. User s Guide. 600 bar 60 MPa 8700 psi

INSTALLATION/OPERATION/MAINTENANCE INSTRUCTIONS FOR ARCHON MODELS WD2010L, WD2010, WD2010H WASHDOWN STATIONS. ARCHON Industries, Inc.

Y690A Series Pressure Reducing Regulators

MEMFlo MFTV METAL BODIED FLOW METER WITH CLEAR TOP

See pages 2, 7 and 8 for ATEX ratings. AIR VALVE No-lube, no-stall design 30(51) 35(00) U.S. Gallons per minute

Before installation these instructions must be fully read and understood

PFS Rotameter. Variable Area Flowmeter. Instruction Manual PFS-RM200, RM230, RM250

EJECTORS GENERAL OPERATION & MAINTENANCE MANUAL

M9000-3xx Series Weather Shield Enclosures

860 and 960 Series Spring-Loaded Pressure/ Vacuum Relief Valve (ATEX Approved)

Transcription:

INSTRUCTION MANUAL VARIABLE AREA FLOWMETERS 10A2235 Rev.5 1) Glass meter tubes have been designed to operate up to maximum design working pressures listed herein. This not to be construed as a certification that the tubes will not break at any pressure. Inherent material limitations can result in tube breakage due to conditions beyond our control. For example, glass is a brittle material which may break upon impact; glass if subjected to thermal shock may break; glass is notch sensitive in that scratches, nicks or cracks may result in breakage when pressurized; incorrect installation or faulty operating methods can cause tube breakage regardless of operating pressure. RATOSIGHT FLOW RATE INDICATORS 2) Glass meter tubes are not recommended for either hot or strong alkalies, fluorine, hydrofluoric acid, steam or water over 200 o F (93 o C). Glass meter tubes should be periodically inspected for signs of wear. Erosion, stress cracks, nicks or deep scratches provide early warning for tube replacement. With certain fluids, the glass may erode evenly so wear is not visibly noticeable. If wear is suspected, the tube should be replaced in order to eliminate this potential cause of meter tube breakage. 3) It is important that all materials of construction be compatible with the service to which the Meter is applied. It is especially important that "O" ring material be compatible with the process fluid. Glass meter tube breakage can occur in those meters using an "O" ring as an internal seal if the improper material is used. For example: VITON "O" RINGS MUST NEVER BE USED FOR AMMONIA SERVICE; THE CORROSIVE ATTACK OF AMMONIA ON VITON IS EXTREME, CAUSING THE "O" RING TO SWELL TO THE EXTENT OF BREAKING THE GLASS ME- TER TUBE. 4) The Meter should never be subjected to excessive vibration. Avoid the use of quick acting devices in the fluid stream in order to prevent shock waves, associated with such devices, from damaging the meter. 5) The use of a pressure relief valve and/or a rupture disc is recommended in the pipeline containing the Meter and located such to preclude glass meter tube breakage in the event of an overpressurization of the line. 6) When applied to a high pressure gas cylinder, at least two stepdown pressure regulators are to be used between the Meter and the cylinder. 7) Remove pressure from the Meter before attempting to remove the meter tube. 8) Be sure the parts that serve to lock the meter end fittings in place are secure. This should be checked before the Meter is put into service or returned to service after maintenance. Loose end fittings may result in glass meter tube breakage. 9) The glass meter tube should be periodically inspected and replaced if cracked, nicked, scratched or worn.

Description The RATOSIGHT flow rate indictaor operates on the variable area principle to indicate the flow rate of a liquid or a gas. The basic meter components, as shown in Figure 1, are a tapered glass tube, a float, and a metal body. A scale is provided on the glass metering tube. The float is guided by flutes formed on the inside surface of the tube; the height of the float in the tube is proportional to the flow rate, so that the flow rate can be read from the scale at a point opposite the grooved horizontal line of the float. The scale on the meter tube is direct reading in GPM, water or SCFM, air. Special scales may be furnished, when specified. In addition to providing a direct indication of flow rate, an alarm extension may be installed on the Meter to open or close alarm circuits at predetermined flow rates. A magnet, connected to the float by an extension, is moved by the float and actuates a switch at a pre-determined point, within the range of the Meter. For details regarding the alarm extension, refer to ABB 55AR1000 Instruction Manual. FIGURE 1. CROSS SECTION OF RATOSIGHT FLOW RATE INDICATOR PRESSURE RATING, METER ASSEMBLY @ 100 o F (38 o C) * Maximum Design Pressure, in psig & MPa Connection Size Meter Tube Inlet Pressure Rating Maximum Temperature NPTI Size (Inches) (Inches) psig MPa o F o C 1/2 1/4 200 1.4 250 121 1/2 1/2 175 1.2 250 121 1 1 125 0.86 250 121 1 1/2 1 1/2 100 0.70 250 121 2 2 75 0.52 150 60 * For pressure rating at other temperatures, consult ABB. NOTE The warning was changed since this Bulletin was last published as PN24289A Read these instructions before starting installation; save these instructions for future reference. Copyright 2002 ABB Company. [April 2002] 2

Installation The RATOSIGHT flow rate indicator is normally furnished for horizontal connections. However, vertical connections or combination horizontal-vertical connections may be made by changing the locations of the pipe plugs. Thus an over alarm extension can be added to the Meter using a horizontal outlet. If the Meter has a vertical inlet, a nipple and union should be added to facilitate future removal of the float, The various piping arrangements are shown in Figure 2. Refer to Figure 4 for the outline dimensions of the Meter. Install the Meter in the piping system so that the tube is vertical. The Meter should be as free as possible from piping strains. When installing large indicators, use brackets or piping supports to eliminate piping strains. If the float is received separately from the instrument, insert it in the tube after removing the inlet fitting. Figure 3 shows the correct float orientation and the reading edges. If the RATOSIGHT flow rate indictor is equipped with an alarm extension, complete the wiring as discussed in the alarm s Instruction Bulletin. WARNING An operator protective shield is required when this meter is used for compressed gases or liquids above their boiling point. A sheet of polycarbonate can be used as a shield. Alternately, an ABB series 10A4500 meter with a protective shield included or an all metal ABB meter can be substituted. FIGURE 2. PIPING ARRANGEMENTS Operation Start flow thru the Meter slowly to avoid excessive flow surges which might damage the float or tube. Flow rate is read from the scale at a point opposite the grooved horizontal line on the float. When millimeter or other special scales are furnished, a calibration curve may be supplied. By entering the indicated value on this curve, the actual flow rate can be obtained. If an alarm extension has been furnished, refer to the appropriate Instruction Manual to set the alarm point. Maintenance Normal maintenance consists of cleaning the float and tube to maintain good visibility. If the meter inlet is connected vertically, the inlet piping must be separated to remove the float. Remove the float and tube by taking the inlet fitting from the bottom of the meter. Caution Exercise care when removing the inlet fitting as the float and tube are free to fall out of the meter body. Clean the float and tube carefully with a suitable solvent (detergent and water are usually satisfactory), then replace the parts. To disassemble the meter completely, remove the float as outlined above, then the outlet adaptor ring may be withdrawn from the meter body. During meter re-assembly it is recommended that new "O" rings be used. 3

FIGURE 3. FLOAT READING EDGES FIGURE 4. OUTLINE DIMENSIONS 4

PN24289B The Company s policy is one of continuous product improvement and the right is reserved to modify the information contained herein without notice. 2002 ABB Inc. Printed in USA ABB Inc. 125 East County Line Road Warminster, PA 18974 USA Tel. 215-674-6000 FAX: 215-674-7183 ABB Instrumentation Ltd Howard Road, St. Neots Cambs. England, PE19 3EU Tel. +44 (0) 1480-475-321 FAX: +44 (0) 1480-217-948 ABB Instrumentation S.p.A Via Sempione 243 20016 Pero (Milano) Italy Tel: +39 (02) 33928 1 Fax: +39 (02) 33928 240 ABB Automation Products GmbH Industriestr. 28 D-65760 Eschborn Germany Tel: +49 (0) 6196 800 0 Fax: +49 (0) 6196 800 1849