Instruction Manual for Microlitre Centrifuge Z 160 M

Similar documents
Instruction Manual. Labnet Spectrafuge 16M Microcentrifuge

INSTRUCTION MANUAL FOR THE CLINICAL 50 CENTRIFUGE

SPECTRAFUGE 6C. User Manual C0060 C V

Clinispin CT20 User Manual

Spectrafuge 24DMicrocentrifuge User Manual C2400 C V C V

Instruction Manual. Labnet Spectrafuge 7M Microcentrifuge

Prism Microcentrifuge

Z100A Clinical Centrifuge

Instruction Manual for Table Top Refrigerated Centrifuge Z 513 K

CONTENTS. 3 MAINTENANCE 3.1 Service and Maintenance Maintenance and cleaning Glass breakage Disinfection

Instruction Manual for Table Top Refrigerated Centrifuge Z 383 K

INSTRUCTION MANUAL FOR BENCH TOP CENTRIFUGE Z 200 A

Operating Manual for Table Top Centrifuge Z 326 K

Operating manual for Table Top Centrifuge Z 32 HK

Hermle Labortechnik GmbH Z366_V1.14. Operating manual for universal Table Top Centrifuge Z 366

Instruction Manual for Compact Centrifuge Z 206 A

INSTRUCTION MANUAL. CAT Place Table Top Centrifuge

Centrifuge. Model XC-2000 Operating Manual

Centrifuge. Model XC-1000 Operating Manual

OPERATIONS MANUAL. Mini Centrifuge Model MCF Certified

Instruction Manual for the Z 216 MK Refrigerated Microcentrifuge

HERMLE L ABORTECHNIK. Universal Centrifuges Z 400 Z 400 K. Universal High Speed Centrifuges Z 383 Z 383 K

Centrifuges and Accessories. Complete Catalogue

Hybex Microsample Incubator

Instruction Manual for Table Top Centrifuge Z 366

Instruction Manual for the Revolutionary Science RS-200 RevSpin Microcentrifuge

3. Operating instructions: Minor 200

Instruction Manual for Microliter Centrifuge Z 233 M-2

Operating Manual Cellspin

Instruction Manual for the Revolutionary Science RS-102 RevSpin Microcentrifuge

High Speed Micro-Centrifuge User manual

Centrifuge Operator / Service Manual

Rescue Pac. Please read and fully understand the instructions in this manual before operation. Keep this manual safe for future reference

The EFL 2000/1 & 2 User Guide Test Sieve Shaker. Contents

Instruction Manual. Vortex Mixers

Operation Manual Supra R22

Bio Lion Table Top Centrifuge XC-L5

PET UNIT DOSE CABINET

OPERATION & MAINTENANCE MANUAL FOR SCEN-206 LED & MULTI-FUNCTION TYPE CENTRIFUGE PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION.

REFRIGERATED HIGH SPEED CENTRIFUGE CONFORM. Equipment. High Speed HERMLE

Bio Lion Table Top Centrifuge XC-H165

materials and workmanship for 2 years. Should the centrifuge require warranty or Phone: or Fax:

Pro Booster 802Li. Please read and fully understand the instructions in this manual before operation. Keep this manual safe for future reference.

D3024 & D3024R High Speed (Refrigerated) Micro-Centrifuge DLAB Laboratory Instruments Limited

VWR Clinical 100 & VWR Compact Star CS4

Register your instrument! MiniSpin /MiniSpin plus. Operating manual

CLEMENTS. Orbital 160 Benchtop Centrifuge. CEN V 50Hz. User Manual. Manual No. CEN Issue 4

Instruction Manual for Table Top Centrifuge Z 366

PET UNIT DOSE TABLE INSTALLATION MANUAL FN: Rev A 1/18

RENA AF371Feeder Operating Manual. Feeder. Operating Manual. Manual Part #: M AF371 Operations Rev

Operating Instructions

Chargestar P24 / P32 Battery Charger Startmaster P300 Starter / Charger

Instruction Manual for Table Top Refrigerated Centrifuge Z 36 HK

Universal Centrifuge 320/320R

DM1424 Hematocrit Centrifuge

Model : LMHS-60 Please read this manual carefully before using the instrument Labnics Equipment

Operation Manual Cole-Parmer MPR16 Multipurpose Refrigerated and MP16 Multipurpose Ambient Centrifuges Models , -25, -50, -55

USER MANUAL. D3024 & D3024R High Speed (Refrigerated) Micro-Centrifuge SCILOGEX, LLC.

OPERATORS MANUAL FOR USE WITH MAGNETIC/HEATED STIRRER MG200

1580R User s Manual. Multi-Purpose, High Speed Centrifuge. Gyrozen Co., Ltd.

3 Ton Trolley Jack. Please read and fully understand the instructions in this manual before operation. Keep this manual safe for future reference.

3.25 Ton Heavy Duty Floor Jack

nual k Voltage Stabilizer N) manual New Brunswick Voltage Stabilizer Operating manual

12v / 24v Diesel Transfer Pump Kit

The Powermatic User Guide Test Sieve Shaker

Sigma 2-7. Benchtop centrifuge.

OLYMPIAN MODEL 740 Operation and Service Manual

STARPOINT. Safety instructions. Starpoint VRS (-F) Fax: Phone:

BlueSpin Cryo. (Cat. No. BSD-CRYO)

innovations in battery charging

CLEMENTS. Orbital 260 Benchtop Centrifuge. CEN V 50/60Hz. User Manual. Manual No. CEN Issue 5

MV Series Motors Operation & Parts Manual

BT403. A Geno Technology, Inc. (USA) brand name. BT-300 Power Supply. Cat. No. BT

Car Battery Charger Instructions for Use

HALOGEN FLOODLIGHTS Models CHL1260C & 1260T Part Nos: &

GB/US... LP11 lifting platform. Vers. 6.00

FD Conveyor for FD 260 Tabber

MIKRO R HETTICH DATA SHEET MICROLITRE CENTRIFUGES FASTER RESULTS IN MOLECULAR BIOLOGY FEATURES HIGHLIGHTS

200 Shadylane Drive Philipsburg, PA Phone: (814) Fax: (814) Service Manual. Model 642E Centrifuge

Operating Instructions

NANOPAC-300 & 500 Power Supply. Instruction manual NANOPAC-300 & NANOPAC-500

Model 2310 Series. Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority

6L Oil-less Air Compressor 53103

USER MANUAL. D2012 High Speed Mini Micro-Centrifuge SCILOGEX, LLC.

Quickdraw. Model 2400 Series. Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority

FlexChem Model 404 Rotating Oven

Hand pump PUMP1000-4L-CONTROL. User manual

1 2 3a 3b THREADING CARD 9 10 ABC Office

GLO GT416. Benchtop Centrifuge. User s Manual. Glotech Co., Ltd

Light Source User Guide

UV-cabinet DNA/RNA UVC/T-M-AR

ACROBAT SWING STAND MODEL

80 Litre Suction Oil Drainer With Inspection Chamber

Model 700 Microarray Oven

USER S MANUAL TUBE SEALER. Model SE250. No. CAT.SE25010Ce

D2012 & D2012 plus High Speed Mini Centrifuge

ELECTRIC HOIST MODEL NO: CH2500B, CH4000B OPERATION & MAINTENANCE INSTRUCTIONS PART NO: , LS1010

DM0636 Multi-purpose Centrifuge

Measurements are expressed in millimeters.

Rugged steel cabinet construction ensures quiet operation and long term reliability.

Transcription:

Instruction Manual for Microlitre Centrifuge Z 160 M Brushless Motor

Table of Contents 1. General Information Description Safety Precautions Technical Data Accessories Supplied with Unit Warranty 2. Installation Unpacking the Centrifuge Required Space Installing the Centrifuge 3. Installing the Rotor Rotors and Accessories Mounting and Securing the Angle Rotors Removing the Rotor Overloading the Rotor 4. Operation Closing the Lid Lid Release Lid Lock Speed Selection Selection of Operating Time and Momentary Operation 5. Service and Maintenance Service Cleaning Disinfection Replacing Fuses 6. Troubleshooting 7. Where to call 8. Determination of g-values 9. Conformity Declaration 1

1. General Information This manual provides important operating and safety information for the HERMLE Z 160 M Microcentrifuge. It should be kept near the centrifuge for quick and easy reference. 1.1 Description The HERMLE Z 160 M is a small benchtop centrifuge that is designed for separation of various research samples. The motor is brushless and requires no routine maintenance. The HERMLE Z 160 M is supplied with an 18 x 1.5 ml rotor for microsamples. Adapters are available for tubes smaller than 1.5 ml. The HERMLE Z 160 M reaches speeds of up to 14,000 rpm/16,000 x g. 1.2 Safety precautions Note: All users of the centrifuge must read the Safety Precautions section before attempting to operate the unit! If this equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. Do not operate the HERMLE Z 160 M if any of the following conditions exist: - The centrifuge has not been installed properly. - The centrifuge is partially dismantled. - Service has been attempted by unauthorized or unqualified personnel. - The rotor has not been installed securely on the motor shaft. - Rotors and accessories do not belong to the standard range of accessories and/or permission to use such rotors/accessories has not been obtained from the manufacturer. Exception: Microcentrifuge tubes made of plastic, normally available in the laboratory. - The centrifuge is located in an explosive atmosphere. - Materials to be centrifuged are combustible and/or explosive. - Materials to be centrifuged are chemically reactive. - The rotor load is not properly balanced. 2

1.3 Technical Data Type HERMLE Z 160 M Dimensions: Width 20 cm Depth 22 cm Height 18 cm Weight: 4.5 Maximum speed 14,000 rpm Maximum volume 18 x 1.5/2.0 ml Maximum RCF 15,994 x g Admiss. density 1.2 kg/dm 3 Electrical rating 230 V, 50/60 Hz, 0.6 A Fuse 1.6 AT 1.4 Accessories supplied with each centrifuge unit Each unit is supplied with - 1 instruction manual, - 1 power cord. Some models are supplied with a rotor screw wrench. 1.5 Warranty This centrifuge has been subject to thorough testing and quality control. In the unlikely event of a manufacturing fault, our two years warranty (from the date of delivery) covers the centrifuge and the rotor. This warranty becomes invalid in case of incorrect operation, use of nonstandard spare parts or accessories and unauthorized modification of the rotor or the centrifuge. 3

2. Installation 2.1 Unpacking the centrifuge Before unpacking the centrifuge, inspect the outside of the carton for any shipping damage. The centrifuge is delivered in an carton with protective cushions. Remove the centrifuge from the carton. Please retain the carton and the cushions until it has been established that your unit is working properly. Inspect the centrifuge for any visible signs of shipping damage. Shipping damage is the responsibility of the transportation carrier. Any claims for damage must be filed within 48 hours. The accessories supplied with the centrifuge should be kept with the instruction manual near the centrifuge s place of installation. 2.2 Required space The centrifuge should be installed on a rigid, even surface. The HERMLE Z 160 M should be operated on a stable laboratory table, cabinet, etc. To guarantee sufficient ventilation, ensure that the centrifuge has at least 15 cm of free space on all sides, including the rear. It is recommended that the centrifuge is not located in a position subject to excessive heat such as strong sunlight, radiators or near the exhaust of a compressor, as heat buildup can occur within the rotor chamber. 2.3 Installation Make certain the timer is set to the off position. Before operating the centrifuge, check that the power supply corresponds to that on the manufacturer s rating label, then connect the power cord to the centrifuge and the power supply. ATTENTION: The timer must be in the OFF position before connecting the power cord. Failing to place the timer in the off position may result in damage to the centrifuge and injury to personnel. 4

3. Installing the rotors 3.1 Rotors and accessories The following accessories are available for the HERMLE Z 160 M: Angle rotor for 18 x 1.5 ml Order no. Tube measurement: Maximum speed: Centrifuging radius: RCF (g-value): Included with unit 1.5 ml (10 x 40 mm) 14,000 rpm 7.3 cm 15,994 x g Adapter for 0.5 ml Order no. 704.005 Tube measurements: Ø 8 x 30 mm Maximum speed: 14,000 rpm Centrifuging radius: 6.6 cm RCF (g-value): 13.584 x g Adapter for 0.4 ml Order no. 704.004 Tube measurements: Ø 6 x 45 mm Maximum speed: 14,000 rpm Centrifuging radius: 7.3 cm RCF (g-value): 15,994 x g Adapter for 2 x 8 x 0,2 ml PCR stripes Order no. 33-5059 Maximum Speed: 14,000 rpm When the rotor begins to accelerate, the lid lock indicator light turns on and the lid button becomes inoperable. Do not attempt to open the lid until the lid lock indicator turns off. 3.2 Mounting and securing the angle rotor Remove the rotor screw from the motor shaft by turning it counter clockwise. Clean the motor shaft and the rotor mounting hole (see figure 1). Place the rotor on the motor shaft ensuring that the crosspin (figure 2) aligns correctly with the rotor slot (see figure 1). Note: Figures 1 and 2 can be found on the following pages. Install the rotor screw on the motor shaft by turning it clockwise. Hold the rotor with one 5

hand and hand-tighten the rotor screw. Use an adjustable or ¼ in. wrench (some units are supplied with a wrench) to tighten the screw one additional quarter turn. When loading the rotor, refer to figure 3 (located on the following page). Loading in the pattern indicated will ensure a balanced load. Tubes to be loaded should be filled equally by eye. The difference in weight between the tubes should not exceed 0.1 gram. A partially loaded rotor may be centrifuged. For example, you may want to operate the 18- place rotor with only 2 or 4 tubes loaded. In this case, it is important that the occupied holes are opposite each other (refer to figure 3). 3.3 Removing the rotor Using an adjustable or ¼ in. wrench (some units are supplied with a wrench) loosen the screw and remove the rotor retaining screw/washer assembly by turning it counter clockwise. Lift the rotor directly upward in a straight vertical motion. Caution: Be sure to secure the rotor screw and tighten with a wrench before further operation. Figure 1: Motor shaft and rotor mounting hole right! Figure 2: Motor shaft and rotor mounting hole wrong! 6

Figure 3: Loading the rotor 3.4 Overloading rotors All approved rotors are listed with their maximum speed and maximum filling weight in table 2: permissible net weight (see APPENDIX P. IV). The maximum load permitted for a rotor, which is determined by the manufacturer, as well as the maximum speed allowed for this rotor (see label on rotor), must not be exceeded. The liquids the rotors are loaded with, should have an max. homogeneous density of 1,2 g/ml or less when the rotor is running at maximum speed. In order to spin liquids with a higher density, the speed has to be reduced according to the following formula: 1.2 Reduced speed n red = 1.2 x max. speed (n max ) of the rotor Example: n red = x 4.000 = 3.360 rpm If In case of any questions, please contact the manufacturer! 4. Operation ATTENTION: Never attempt to operate the centrifuge with rotors or adapters that show signs of corrosion or mechanical damage. Never centrifuge strongly corrosive materials that may damage the rotors or accessories. 4.1 Closing the lid After the rotor has been properly secured and loaded, close the centrifuge lid by holding it down for two seconds in order to engage the lid latch. 7

4.2 Lid release At the end of a run, the centrifuge lid will automatically pop open. WARNING: Do not attempt to open the lid of any centrifuge until the unit has come to a complete stop. In the event of a power failure or malfunction, it may be necessary to open the lid manually: 1. Disconnect the power cord from the wall socket power supply. 2. Remove the plastic plug, located on the left side of the unit, below the quick button. 3. Pull the cord (attached to the plug) to open the lid lock manually. 4.3 Lid lock The centrifuge can be started only with the lid securely closed. When the rotor begins to accelerate, the lid lock indicator light turns on and the lid button becomes inoperable. Do not attempt to open the lid until the lid lock indicator turns off. 4.4 Speed selection The speed (rpm) can be selected to 14,000 rpm with the knob labelled speed. The scale is directly proportional to the speed a setting of 9 corresponds to 9,000 rpm, a setting of 13 corresponds to 13,000 rpm, etc. 4.5 Selection of operating time and momentary operation Operation of the centrifuge begins when the timer knob is turned clockwise to set a run time. For run times less than 5 minutes, turn the knob clockwise past the halfway point and then counter-clockwise to the desired time. For run times longer than 5 minutes, turn the knob clockwise to the desired time. When the selected time expires, the centrifuge will stop automatically. To stop the centrifuge prior to the expiration of set time, turn the timer knob to the zero position. The centrifuge may be operated manually by pressing and holding the quick button. The centrifuge will continue to run as long as the button is depressed. 8

Some models have a timer with a hold position. For continuous operation, turn the timer knob firmly to the left. The centrifuge will continue to run until the knob is turned back to the zero position. Note: The timer knob may be turned in either direction during operation of the centrifuge without damage to the timer mechanism. Figure 4: HERMLE Laufzeit Deckelschlo Minuten Dauerlauf 0 Z 160 M control panel layout quick Drehzah min -1 x 1 1 1 1 1 1 9 8 2 7 3 4 5 6 2 3 2 1 5 10 Deckel 9

5. Service and Maintenance 5.1 Centrifuge service The brushless motor in the HERMLE Z 160 M requires no routine maintenance. Any required service should be performed only by authorized, qualified personnel. Repairs performed by unauthorized personnel may void the warranty. 5.2 Cleaning the centrifuge Always keep the centrifuge housing, rotor chamber, rotors and rotor accessories clean. All parts should be wiped down periodically with a soft cloth. For more thorough cleaning, use a neutral cleaning agent (ph between 6 and 8) applied to a soft cloth. Excessive amounts of liquid should be avoided. Liquid should not come into contact with the motor. After cleaning, ensure that all parts are dried thoroughly by hand or in a warm air cabinet (max. temp. 50 C). 5.3 Disinfection Should a spill of infectious materials occur within the rotor or rotor chamber, the unit should be disinfected. This should be performed by qualified personnel with proper protective equipment. 5.4 Replacing fuses Check the fuse when it is recommended in the Troubleshooting Guide located in this manual. The fuse holder is located in the power inlet on the rear of the unit. Disconnect the power cord from the power inlet. Open the fuse holder drawer by inserting a small screwdriver under the tab and prying it open. Remove the innermost (operative) fuse from its retaining tabs and replace if necessary. A spare fuse is located in the outermost chamber of the fuse drawer. Replace only with a fuse of exactly the same value as the original. (Fuse type may be found in the Technical Data section of this manual.) 10

6. Troubleshooting guide Please refer to this guide before calling for service. Centrifuge will not start Possible reason: Solution: Possible reason: Solution: No power supply Check that power is being supplied to the outlet. Check that the power cord is correctly plugged into the wall outlet and the back of the centrifuge. Check that the power cord is intact not damaged. Blown fuse Check fuse and replace if necessary. Lid lock will not release Possible reason: Solution: Possible reason: Solution: Possible reason: Solution: Possible reason: Solution: Defective lid lock Open manually and have unit serviced No power from PC board Call for service Lid lock is jammed Call for service Centrifuge is not receiving power See Centrifuge will not start section Centrifuge cannot be started, although power is on Possible reason: Lid not closed correctly Solution: Close lid correctly Possible reason: Solution: No speed or time has been selected Set speed and time 11

7. Where to call Should you have any questions about the HERMLE Z 160 M or its accessories, please call Hermle Labortechnik s Customer Service Department at +49-7426-96 22 55. Customer Service is staffed from 8:00 am to 12:00 am and from 1:00 pm to 4:00 pm, Monday through Friday. Our 24 hour fax number is +49-7426-96 22 49. Should your HERMLE Z 160 M require service, please call our above Customer Service Department. Please be sure to have your serial number (located on the back panel of the instrument) available when calling. 12

8. Determination of g-values When using the 1.5 ml rotor the radius is 7.3 cm. See section 3.1 for the correct radius when using adapters and smaller tubes. Spinning radius in cm 50 Speed (RPM) per minute 20000 45 40 The relative centrifugal force (g-force) can be estimated by using the chart on this page or by applying the following formula: 15000 Relative centrifugal force in g 35 10000 30 n g = 11.18 x r x ( )2 30000 20000 25 10000 6000 20 18 where: r = radius in cm n = speed in rpm 5000 4000 3000 2000 1000 4000 3000 16 14 12 The radius from the centre of the rotation axis to the bottom of the outermost portion of the tube should be used. RCF is expressed relative to the earth s gravity. 500 400 300 200 2000 10 9 8 To use this chart, find the radius value on the radius scale. Place the edge of a ruler on the value. Place the right side edge of the ruler on the speed scale at the desired speed. The estimated RCF can then be read from the RCF scale where the ruler edge passes through it. This chart can also be used to determine the proper speed for desired RCF value. 100 50 20 1000 500 7 10 6 3 5 200 13

9. Declaration of Conformity EG Konformitätserklärung EC Conformity Declaration Hermle Labortechnik GmbH - Siemensstr. 25 - D-78564 Wehingen Germany Das bezeichnete Produkt entspricht den einschlägigen grundlegenden Anforderungen der aufgeführten EG-Richtlinien und Normen. Bei einer nicht mit uns abgestimmten Änderung des Produktes oder einer nicht bestimmungsgemäßen Anwendung verliert diese Erklärung ihre Gültigkeit. The Product named below fulfills the relevant fundamental requirements of the EC directives and standards listed. In the case of unauthorized modifications to the product or an unintended use this declaration becomes invalid. Produkttyp Product type Laborzentrifugen mit Zubehör Laboratory centrifuge with accessories Typenbezeichnung Typ designation Z 160 M Einschlägige EG-Richtlinien / Normen Relevant EC directives / standards RL 2006/95/EC, 2004/108/EC, RoHS II 2011-65-EC, WEEE 2002/96/EC IEC 61010-1:2001 Ed. 2, EN 61000-3-2, EN 61000-3-3 Wehingen, den 15.10.2015 Alexander Hermle Geschäftsführer, Managing Director 14

HERMLE Labortechnik GmbH Siemensstraße 25 78564 Wehingen Tel: 0 74 26-96 22-17 Fax: 0 74 26-96 22-49 Email: vertrieb@hermle-labortechnik.de Internet: http://www.hermle-labortechnik.de Technische Änderungen vorbehalten. HERMLE Labortechnik GmbH 2015 15