INSTRUCTION MANUAL FOR THE CLINICAL 50 CENTRIFUGE

Similar documents
Z100A Clinical Centrifuge

Instruction Manual. Labnet Spectrafuge 16M Microcentrifuge

Instruction Manual for Microlitre Centrifuge Z 160 M

SPECTRAFUGE 6C. User Manual C0060 C V

Clinispin CT20 User Manual

Instruction Manual. Labnet Spectrafuge 7M Microcentrifuge

Spectrafuge 24DMicrocentrifuge User Manual C2400 C V C V

Prism Microcentrifuge

CONTENTS. 3 MAINTENANCE 3.1 Service and Maintenance Maintenance and cleaning Glass breakage Disinfection

INSTRUCTION MANUAL. CAT Place Table Top Centrifuge

materials and workmanship for 2 years. Should the centrifuge require warranty or Phone: or Fax:

INSTRUCTION MANUAL FOR BENCH TOP CENTRIFUGE Z 200 A

Centrifuge. Model XC-2000 Operating Manual

Centrifuge Operator / Service Manual

OPERATIONS MANUAL. Mini Centrifuge Model MCF Certified

Instruction Manual for the Revolutionary Science RS-200 RevSpin Microcentrifuge

Instruction Manual for the Revolutionary Science RS-102 RevSpin Microcentrifuge

Centrifuge. Model XC-1000 Operating Manual

Instruction Manual for Compact Centrifuge Z 206 A

OPERATION & MAINTENANCE MANUAL FOR SCEN-206 LED & MULTI-FUNCTION TYPE CENTRIFUGE PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION.

Vetlab CombiSpin Centrifuge

PET UNIT DOSE TABLE INSTALLATION MANUAL FN: Rev A 1/18

Battery Chargers Sealed or Valve Regulated Lead Acid Batteries Models: PSC A AND PSC A

CLEMENTS. Orbital 160 Benchtop Centrifuge. CEN V 50Hz. User Manual. Manual No. CEN Issue 4

Operating Manual. Multi Purpose Centrifuge Combi - 514R

Bio Lion Table Top Centrifuge XC-L5

Battery Chargers Sealed or Valve Regulated Lead Acid Batteries Model: PSC A

VWR Clinical 100 & VWR Compact Star CS4

Bio Lion Table Top Centrifuge XC-H165

UV Transilluminator Dual Wavelength (Large)

Technical Information Manual

CLEMENTS. Orbital 260 Benchtop Centrifuge. CEN V 50/60Hz. User Manual. Manual No. CEN Issue 5

Light Source User Guide

200 Shadylane Drive Philipsburg, PA Phone: (814) Fax: (814) Service Manual. Model 642E Centrifuge

58kHz AM Deactivator DB104 Instruction Manual V1.0

Model : LMHS-60 Please read this manual carefully before using the instrument Labnics Equipment

innovations in battery charging

Operating Manual Cellspin

PET UNIT DOSE CABINET

Battery Chargers Sealed or Valve Regulated Lead Acid Batteries Model: PSC AP

Model HPX60 Series Automatic Battery Charger User s Manual Rev. 1.0 October 17, 2006

OWNER'S MANUAL WARNING DANGER. Propane cylinders sold separately. The propane cylinder must be disconnected when this firebowl is not use.

Instruction Manual. Vortex Mixers

UV-AIRE AIR PURIFYING SYSTEM

nual k Voltage Stabilizer N) manual New Brunswick Voltage Stabilizer Operating manual

BT403. A Geno Technology, Inc. (USA) brand name. BT-300 Power Supply. Cat. No. BT

Model: UV-16R/24 READ THESE INSTRUCTIONS CAREFULLY AND COMPLETELY BEFORE PROCEEDING WITH THE INSTALLATION.

HALOGEN FLOODLIGHTS Models CHL1260C & 1260T Part Nos: &

Register your instrument! MiniSpin /MiniSpin plus. Operating manual

1580R User s Manual. Multi-Purpose, High Speed Centrifuge. Gyrozen Co., Ltd.

CENTRIFUGE USER MANUAL 642SYG V.1.0

Owners Manual: - Pumps

Monocellular centrifugal electro-pumps

OPERATOR S MANUAL Please read all the information in this booklet before using the unit.

User s Manual Power Supply. IM E 2nd Edition. Yokogawa Electric Corporation

XCM12 Mini-Centrifuge

KenkoAir_Purifier manual ENG-FR_FDA 10/12_Air Filter manual 8/15/13 10:44 AM Page 1 OPERATING MANUAL

Instruction Manual for the Z 216 MK Refrigerated Microcentrifuge

D3024 & D3024R High Speed (Refrigerated) Micro-Centrifuge DLAB Laboratory Instruments Limited

The Powermatic User Guide Test Sieve Shaker

UView Mini Transilluminator

Paraffin Dispenser Catalog # PD-120, PD-220 Operator s Manual Version 1.5b, Jan 2007

Via Michelangelo Buonarroti, 52/B Cologno M.se (Mi) USER S MANUAL

200 Shadylane Drive Philipsburg, PA Phone: (814) Fax: (814) Service Manual

Laboratory Centrifuge Model GF614B

Operating Instructions. International English. CCS Dust Extractor

OPERATING INSTRUCTIONS. Note: 6V Charging. Requires Manual Shut Off.

Register your instrument! HeatSealer S100. Operating manual

MyGel Mini. Operating Manual Version 1.2. a division of. Electrophoresis System. 6 Troubleshooting (continued) Problem Causes Resolutions

MODEL 755VES Model 755VES Operator s Manual

High Speed Micro-Centrifuge User manual

OPERATION INSTRUCTIONS

Hybex Microsample Incubator

Safety Precautions. Damages resulting from failure to observe these safety precautions are exempt from any legal claims whatever.

30 Amp Solar Charge Controller

MODEL 853VES Model 853VES Operator s Manual

WELDING INVERTER. PEGAS 160 E Smart PEGAS 200 E Smart OPERATING MANUAL. ALFA IN a.s. PEGAS E Smart Manual EN 04

OPERATION and MAINTENANCE INSTRUCTION MANUAL

MMA 160S ARC/MMA WELDER OPERATION INSTRUCTIONS

QPET, QPETXU Battery Chargers

L-K INDUSTRIES BENCHMARK S MANUAL

Universal Centrifuge 320/320R

Sylvania 9 Color Changing One Plug Tree Tree ID # T5, Item # V

USER S MANUAL TUBE SEALER. Model SE250. No. CAT.SE25010Ce

AIR FILTER MODEL NO: AF1000 OPERATION & MAINTENANCE INSTRUCTIONS PART NO:

INSTRUCTION MANUAL ANGLE GRINDER PT W

DM1424 Hematocrit Centrifuge

DC TO AC POWER INVERTER PWRINV150W INSTRUCTION MANUAL

OPERATION MANUAL HARVEST ROYAL

OPERATION MANUAL HARVEST CAVALIER

GLO GT416. Benchtop Centrifuge. User s Manual. Glotech Co., Ltd

FlexChem Model 404 Rotating Oven

HOT WASHER MODEL NO: KING 125 OPERATION & MAINTENANCE INSTRUCTIONS PART NO: LS1009

MODEL 660 AUTOMATIC FASTENING CENTER OPERATOR S MANUAL

Cincinnati, OH USA

NANOPAC-300 & 500 Power Supply. Instruction manual NANOPAC-300 & NANOPAC-500

ARC 1850 LIST OF FIGURES

ACCUVERT (tm) DISPENSING SYSTEM

User s Manual Power Supply. IM E 3rd Edition. Yokogawa Electric Corporation

Model 700 Microarray Oven

Transcription:

INSTRUCTION MANUAL FOR THE CLINICAL 50 CENTRIFUGE 82013-800 January 2006

INDEX PAGE 1. General Information... 2 1.1 Description... 2 1.2 Safety precautions to be observed before operating the centrifuge... 2 1.3 Technical data... 2 1.4 Accessories... 3 1.5 Warranty... 3 2. Installation... 3 2.1 Unpacking the centrifuge... 3 2.2 Required space... 3 2.3 Cushion Installation... 3 2.4 Centrifuge Installation... 4 3. Operation... 4 3.1 Closing the lid... 4 3.2 Lid release... 4 3.3 Lid lock... 4 3.4 Starting the centrifuge... 4 4. Proper use of the Rotor 4.1 Overloading the rotor... 5 4.2 Balancing the rotor... 5 5. Service and Maintenance... 6 5.1 Service of centrifuge... 6 5.2 Cleaning the centrifuge... 6 5.3 Disinfection... 6 5.4 Cleaning the centrifuge after breakage of glass tubes or bottles... 6 5.5 Replacing fuses... 6 6. Troubleshooting guide... 7 7. Chart for determining RCF (g-value)... 8 1

1. General Information - This manual is intended for use with the Clinical 50 (82013-800) Centrifuge. 1.1 Description The Clinical 50 is a small bench top centrifuge designed for sedimentation of various samples. The Clinical 50 has a built in angle rotor capable of running up to to 4,000 rpm (1,900 x g.) This product is for research only. 1.2 Safety precautions to be observed before operating the centrifuge! Do not operate if: - The centrifuge has not been installed correctly. - The centrifuge is partly dismantled. - Service has been attempted by non-authorized or non-qualified personnel. - Rotors and accessories are not part of the standard range of accessories or have been obtained without the prior permission of the manufacturer. Exception: tubes and bottles made of glass or plastic normally available in the laboratory, and designed for centrifugation. - The centrifuge is to be used in explosive atmospheres. - You intend to centrifuge combustible and explosive materials. - You intend to centrifuge materials which are chemically reactive. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. 1.3 Technical Data Model No. 82013-800 Dimensions: Width 265 mm (11 inch) Depth 330 mm (13.75 inch) Height 245 mm (10.2 inch) Weight 5.5 kg / 12 lbs Max. speed 4,000 rpm Max. volume 90 ml (6 x 15ml) Max. RCF 1,900 x g Maximum Admiss. density 1.2 kg/dm 3 Rotor radius 10.62 cm Timer 0 to 30 minutes or continuos Electrical ratings 120 V/50-60 Hz, 1 Amp The manufacturer reserves the right to make technical modifications 2

1.4 Accessories Accessories supplied with each centrifuge unit: - One 6 x 15ml rotor - One Instruction manual - One Warranty card - Six Rubber Cushions Aditional accessories available: 82013-820 - Combination adapters for 5ml and 7ml tubes, (6pk.) 1.5 Warranty This centrifuge has been subject to thorough testing and quality control. In the unlikely event of any manufacturing faults occuring, our two year warranty (from date of delivery) covers the centrifuge and the rotors. This warranty becomes void in case of incorrect operation, use of non-appropriate spare parts or accessories and non-authorized modification of the rotor or the centrifuge. 2. Installation 2.1 Unpacking the centrifuge The centrifuge is delivered in a carton with two or more formed foam blocks. Please retain the carton and the blocks until it has been established that your unit is working properly. The accessories supplied with the centrifuge should be kept with the instruction manual near the centrifuge's place of installation. 2.2 Required space The centrifuge should be installed on a rigid, even surface. The clinical 50 should only be operated on a stable laboratory table/bench etc. To guarantee sufficient ventilation, please ensure that the centrifuge has at least 15 cm free space on all sides, including the rear. It is recommended that the centrifuge is not situated in a location subject to excessive heat, e.g. strong sunlight, radiators etc. as heat-buildup can occur within the centrifuge bowl. 2.3 Cushion Installation Before inserting any tubes into the centrifuge the rubber cushions must first be properly installed. To install the cushions turn a 10 or 15ml round bottom tube upside down and place the cushion on top. Then remove the rotor sheild and place it on top of the tube while pushing the cushion into the end of the sheild. Then reinsert the shield. 3

2.4 Centrifuge Installation 1. Make certain that the timer is set for the "off" position. Before operating the centrifuge, check that the power source corresponds to that on the manufacturer's rating label, then connect the power cord to a properly grounded outlet ATTENTION: Ensure that the timer is in the "OFF" position, before connecting the power cord. 3. Operation! ATTENTION: Never operate the centrifuge if the rotor shows any signs of corrosion or mechanical damage. Never operate with strongly corrosive materials. 3.1 Closing the lid After the rotor has been properly loaded, close the centrifuge lid. 3.2 Lid release The "lid" button must be pushed to open the lid after a centrifuge run has been completed. 3.3 Lid lock The centrifuge can only be started with the lid closed. Do not attempt to open the lid until the rotor comes to a complete stop. 3.4 Starting the centrifuge To start the centrifuge select the desired run time by turning the time knob in a clockwise direction. If performing a continuous run turn the time knob counter clockwise into the "hold" position. Once a run time is selected a given speed must be selected in order to start the centrifuge. The speed (rpm) can be selected between 1,000 and 4,000 rpm by turning the knob to the 1, 2, 3 and 4 speed marks. To select a speed, turn the speed knob to the 2 posistion to activate, then proceed by turning the knob to the desired speed. At the end of the preselected time, the centrifuge stops automatically. To stop the centrifuge prior to the selected time, turn the timer knob to the "OFF" position. NOTE: The clinical centrifuge is calibrated at 120 volts line voltage. Operating the centrifuge at a line voltage other than 120V can significantlly impact the speed settings below 2,000 RPM. 4

4. Proper loading of the rotor 4.1 The rotor has been designed to handle samples with a maximum density of 1.2g/ml. Tubes should not be run with any substance having a density above 1.2g/ml. 4.2 When loading the rotor you should proceed according to sketch A. Fill the tubes equally by eye and insert them into the tube shields of the rotor. The difference in weight between the tubes should not exceed 2-3 grams. A partially loaded rotor may be centrifuged. For example, you may operate a 6-place rotor with 2, 3 or 4 loaded tubes, but it is important that the occupied tube shields are symetrically arranged. (see sketch A). Sketch A: Examples of properly loaded rotors. empty tube shields loaded tube shields 5

5. Service and Maintenance 5.1 Service of centrifuge Ther is no routine maintenance or service required on the centrifuge other than occasional cleaning. 5.2 General cleaning of the centrifuge Always keep the centrifuge housing, rotor, and rotor accessories clean by wiping down with a damp cloth. Do not use harsh cleaners or detergants on the lid or housing surfaces as it may affect the plastic. After cleaning, ensure that all parts are dried thoroughly by hand or in a warm-air-cabinet (max. temp. +50 C). 5.3 Disinfection The centrifuge should be disinfected periodically and before shipping the unit for service. The centrifuge rotor area and rotor shields should be wiped with alcohol or other disinfecting agents. Do not allow the alcohol or disinfectant to sit on the units lid or housing, as it may affect or discolor the plastic. 5.4 Cleaning the centrifuge after breakage of glass tubes or bottles With high g-values, there is a possibility that tube breakage can occur. Should this happen the centrifuge rotor must be thoroughly cleaned and all broken particles removed. If this is not done, damage could result to the centrifuge.be sure to remove the rotor shields and clean thoroughly. If glass or other material has gotten into the centrifuge enclosure, the enclosure must be removed for cleaning. 5.5 Replacing fuses Check the fuse when it is recommended in the "Troubleshooting Guide". The fuse holder is located underneath the unit. Unplug the unit. Unscrew the fuse holder and remove the fuse. Replace a blown fuse with a fuse of equivalant rating and type. 6

6. Troubleshooting guide Centrifuge will not start Possible reason: Solution: No power supply. Check main power supply and centrifuge fuse. Possible reason: Blown fuse. Solution: Check fuse and replace if necessary (see 5.5) Lid cannot be released: Possible reason: Solution: Lid lock is jammed. Push front of lid down while pushing release button. Centrifuge cannot be started, although power is on: Possible reason: Solution: Possible reason: Solution: Lid not closed correctly. Close lid correctly, push down on front of lid. No speed or time has been selected. Check and set the time or speed. Possible reason: Blown fuse. Solution: Check fuse and replace if necessary (see 5.5). Should you have any questions about the centrifuge and its accessories, or require service, please contact VWR International at 800-932-5000. Please have the unit s serial number (located on the back panel of the instrument) available when calling. Do not send in a unit for service without first calling to obtain a repair authorization number and a decontamination form. Should the centrifuge need service be sure to decontaminate it and enclose the decontamination from. The unit should be properly packed to avoid damage. Any damage resulting from improper packing shall be the responcibility of the user. 7

7. Chart for determining g-values See section 1.3 for the correct radius of your rotor. Note the use of tube adapters will change this radius.. To determine the relative centrifugal force (RCF - value). This value is the multiplier to the gravitational pull of the earth (g) 8