Guide for installers. METTLER TOLEDO MultiRange Explosionsproof weighing system with the ID3sTx weighing terminal.

Similar documents
Operating Instructions Installation Information. METTLER TOLEDO MultiRange Weighing Platforms PUA574 / PUA579(x)

Operating Instructions Installation Information. METTLER TOLEDO MultiRange Weighing Platforms PFA574 / PFA575(x) / PFA579(x)

Operating instructions. METTLER TOLEDO MultiRange Floor scales/pit scales

4.1 Cleaning the exterior. 4.2 Cleaning the interior

Installation information. METTLER TOLEDO MultiRange Floor scales / Pit scales For hazardous areas

2. Putting into operation. 3. Equipotential bonding. 4. Technical data DRF/DSF (stainless steel)

Operating Instructions Installation Information. METTLER TOLEDO MultiRange Weighing Platforms PFA579(x)lift / PFA779lift.

Operating instructions. METTLER TOLEDO MultiRange Table and stand scales. KA15s/KA32s KB60/KCC150/KCC300.

Operating Instructions

Installation Instructions (Translation) Marking. Description. Danger, warning and note symbols. Technical data. ANTARES ExtSet.

Explosion protected terminal boxes type, GHG 72.

Operating Instructions (Translation)

Operating Instructions

SITRANS P measuring instruments for pressure

ILED Dorado. User Manual

ILED Aquarius Illuminated Windsock. User Manual

HIGH VOLTAGE TEF 1060 HV ENCLOSURE WITH A-BLOCK SOLUTIONS USER MANUAL

INSTRUCTION MANUAL (ATEX / IECEx)

103 Differential Pressure Switch

Instruction Manual. Ex-Time 40 / 50

1260 Infinity II Degasser User Manual. 5 Maintenance

LIMIT SWITCH Hawkeye - HX Magnetic Solid State Proximity Sensors Installation, Maintenance and Operating Instructions (HX )

IS1 I/O-Modules. BusRail Series General Information. Symbols. Conformity to Standards. Safety instructions. Function/Characteristics

Operating Instructions

Clarius EP PLUS LED FLOODLIGHT Installation Guide

Operating Instructions in accordance with IECEx Scheme rules, IECEx 02 polykom

Smile 11 RA/-B Smile 12 RF/-G Reset button

The cover screws and all other external metal parts are made of stainless steel (AISI 316L). International approvals

102 Differential Pressure Switch

Centrifugal fan with highly efficient EC drive motor

EStrong-series Emergency Stop

INSTRUCTION MANUAL (ATEX / IECEx)

Assembly. Trunk IN + - S U, I U, I. Spur 1

PHOENIX CONTACT - 01/2008

Zone 1 Linear Range - Installation Guide

Installation information. METTLER TOLEDO IND425 / IND435 / IND445 / IND465 terminals

Explosion Proof Equipment General Overview*

Pneumatic or Electropneumatic Positioner for Rotary Actuators Type Fig. 1 Type 3761 Positioner. Mounting and Operating Instructions EB 8386 EN

Operating Instructions

CERTIFICATE OF COMPLIANCE

27.6 Vdc 1 Amp Switch Mode Power Supply for Fire EN54-4:1997 +A1 +A2

11mm Slot Label S1 = 162, S2 = 170, S3 = 250, S4 = 305, S5 = 330, S6 = 410, S7 = 490, S8 = 570, S9 = 650. Maximum Power Dissipation (Watts) T** T**

Ex m Solenoid Operator Type 0519

Load Cell A special beam load cell has been developed by METTLER TOLEDO to meet hygienic criteria. Self draining and a mirror polish make it unique.

USER MANUAL 6 Position Front Connect Rack for TD Series Modules with 16 Channel Modular Front Panel TDP

EVZ Series (Small body) Explosion protected luminaire Operating Instructions

Linear Luminaire for Fluorescent Lamps

EX Series User Manual

Terminal Boxes in Stainless Steel SLS**.T

Operating Instructions for Float Flow Meter / Monitor. Model: SWK

Shut-off dampers. Type AK-Ex. For low-leakage shut-off of volume flows in potentially. explosive atmospheres (ATEX) PD AK-Ex 1.

Model 3400 Intrinsically Safe Load Cell

Operating Instructions

SynergEX Panelboards. Installation & Maintenance Information IF Panelboards for Hazardous Locations SAVE THESE INSTRUCTIONS FOR FUTURE REFERENCE

Segurança. Segurança. PSB Heating System Cold applied connection technology. Compulsório. Compulsório. Operation Manual and Installation Instructions

Accessories for Wind Power Inverter WINDY BOY PROTECTION BOX 400 / 500 / 600

Instruction Manual December milltronics RBSS

PROCESS AUTOMATION. Instruction Manual JUNCTION BOXES/ SEGMENT PROTECTORS

Solenoid Valves Type 3963

Linear Luminaire with LED

Magnetic-inductive flow meter for hazardous areas

Single Point Overfill Prevention Control Monitor Model: ST-15C Series Technical Manual

IQL Helios Area-/Streetlight. User Manual

JBEP Series Fiberglass Reinforced Polyester Terminal Junction Boxes for Instrumentation Applications Increased Safety. Factory Drilled and Equipped

Magnetic Level Gauges

Control and Terminal Box Type A21-**-*****

PSU EN54-4 Power Supplies

IBExU Institut für Sicherheitstechnik GmbH

Dimensions shown apply to both Stainless Steel and Mild Steel S1, SF*1 = 162, S2, SF*2, S2L, SF*2L = 170, S3, SF*3 = 250,

Installation Guide. Auxiliary Pressure Control Manifold on 6890 GC Accessory G1570A

Rod probe 11375Z. Technical Information. Conductive level limit detection Partially insulated rod probe for use in conductive liquids

Resistance Thermometers Model TR217, with Spring-Loaded Probe Tip

Leading the way in hazardous areas Features EATON

MAGNETIC FLOAT SWITCH

E II 3 (1) G Ex na [ia Ga] IIC/IIB T4 Gc

UV- TECHNOLOGY. Technical Documentation ELC N6 / ELC N8 / ELC N10

Magnetic Level Gauges

Installation, Operating & Maintenance Instructions. Norbro J-Switch FM 00707

Compact Pressure Transmitter DAN-HH

Electropneumatic Converters i/p Converters Type 6111 Mounting and Operating Instructions EB 6111 EN

Linear Luminaire for Fluorescent Lamps Series EXLUX 6401

Contents. Mounting instructions. 3 Hangers placement...4 Mains connection instructions..5 Safety zones.. 6 Maintenance...7 Warranty card.

INLINE Flow sensor for hazardous area II 1 G/D - II 2 D - II 3 GD - I M1

SIWAREX WL200. Weighing systems. Load cells SIWAREX WL200. Introduction 1. Safety instructions. Description. Application planning 4

Type AK-Ex FOR LOW-LEAKAGE SHUT-OFF OF VOLUME FLOWS IN POTENTIALLY EXPLOSIVE ATMOSPHERES (ATEX)

Installation manual wall-mounted distributor

Type 3761 Pneumatic or Electropneumatic Positioner for Rotary Actuators. Fig. 1 Type 3761 Positioner. Mounting and Operating Instructions EB 8386 EN

Hazardous areas, as found around a service station, are classified into three zones as follow: Table 1: Zone Classification for Vapours

USER MANUAL 6 Position Powered Rack for TD Series Modules TDR 01 AC

OPW Fluid Transfer Group Europe BV

*-JBSC-* Fieldbus junction box with integrated short-circuit current limitation

Operating Instructions

PowerLogic High Density Metering System 4-Meter Enclosure Installation Guide

Operating Instructions. Angle Seat Control Valve. Type 7020

Junction Boxes & Enclosures JBES and ECES Series: 316L Stainless Steel Enclosures Increased Safety

SWB505 MultiMount EN1090

User Manual. Solar Charge Controller 3KW

Additional operating instructions according to ATEX 94/9/EC Volumetric flow controller VM-PRO-Ex

BODAS Pressure sensor PR3 series 10

Intrinsically safe presure transmitters for hazardous environments Type MBS 4201, MBS 4251, MBS 4701 and MBS Technical brochure

Transcription:

Guide for installers METTLER TOLEDO MultiRange Explosionsproof weighing system with the ID3sTx weighing terminal Hazardous zone ID3sTx weighing terminal LWL Safe zone Dual Channel Fiber Optic Converter CL RS D..Tx, PTA459x, PUA579x Ext. battery RWMx PBA430x Ex1 system solution www.mt.com/support

ID3sTx Contents Page 1 Safety instructions... 4 2 System overview... 6 2.1 The explosionproof weighing system... 6 3 Installation... 9 3.1 Installing the system modules... 9 3.2 Connecting the units... 9 3.3 Equipotential bonding... 12 3.4 Fitting connectors on the weighing platform cable... 12 3.5 Installing the Fiber Optic PCB... 13 4 Control drawing... 14 Guide for installers 22000435B 05/05 3

Safety instructions ID3sTx 1 Safety instructions There is an increased risk of injury and damage when the explosionproof ID3sTx weighing terminal is used in a potentially explosive atmosphere. Special care must be taken when working in such hazardous areas. The code of practice is oriented to the "Safe Distribution" concept drawn up by METTLER TOLEDO. Competence Ex approval Operation Battery operation The ID3sTx weighing terminal may only be installed, maintained and repaired by authorized METTLER TOLEDO service personnel. The ID3sTx weighing terminal with built-in power supply unit may only be connected or disconnected to/from the mains by a qualified electrician. No modifications may be made to the terminal and no repair work may be performed on the modules. Any weighing platform or system modules that are used must comply with the specifications contained in the installation instructions. Non-compliant equipment jeopardizes the intrinsic safety of the system, cancels the Ex approval and renders any warranty or product liability claims null and void. The safety of a weighing system including the ID3sTx weighing terminal is only guaranteed when the weighing system is operated, installed and maintained in accordance with the respective instructions. Also comply with the following: the instructions for the system modules, the relevant national regulations and standards, the applicable statutory requirements for electrical equipment installed in hazardous atmospheres in the respective country, all instructions related to safety issued by the operator. The explosion-protected weighing system must be checked to ensure compliance with the requirements for safety before being put into service for the first time, following any service work and every 3 years, at least. Never use protective hoods from other weighing terminals. Always charge the batteries in a safe zone. Install and use METTLER TOLEDO battery chargers in the safe zone. Use the chargers specified by METTLER TOLEDO only. Never try to open or repair batteries. They are intrinsically safe and are irreparable. Recycle defective batteries or dispose of them in the proper manner. 4 Guide for installers 22000435B 05/05

ID3sTx Safety instructions Installation Only install or perform maintenance work on the weighing terminal in the hazardous zone if the following conditions are fulfilled: the operator has issued a permit ("spark permit" or "fire permit"), the area has been rendered safe and the operator's safety coordinator has confirmed that there is no danger, the necessary tools are in position and the operator is wearing any protective clothing that may be required (to prevent the build-up of static electricity). The approval documents (certificates of conformity, manufacturer's declarations) must be available. Only use the cables specified for intrinsically safe circuits by the applicable national regulations and standards to install a weighing system with the ID3sTx weighing terminal. Store cable in such a way as to prevent damage or deterioration. Cable may only be led into the system module enclosure via the earthing cable gland. Ensure that the seals are in their proper positions. If the weighing terminal is used in conjunction with an automatic or manual filling plant, all of the system modules must be equipped with a permanently wired emergency stop circuit, independent of the system circuit, in order to prevent personal injury or damage to other items of equipment. Guide for installers 22000435B 05/05 5

System overview ID3sTx 2 System overview 2.1 The explosionproof weighing system A weighing system designed for operation in hazardous zones 1 and 21 comprises the following components: Hazardous zone Safe zone ID3sTx weighing terminal LWL, max. 300m Dual Channel Fiber Optic Converter D..Tx PTA459x PUA579x Max. 100m Max. 1.5 m Ext. battery CL RS PC Printer 1 weighing machine connection RWMx Ex1 system solution Max. 100m PBA430x Max. 3m ID3sTx weighing terminal The weighing terminal designed for use in a potentially explosive atmosphere is characterized by the following features: Stainless steel enclosure with large high-contrast liquid-crystal display. Numerical keypad to enter default tare values, for example. Power supplied by an internal or external battery or an internal power supply unit. Up to 2 fiber-optic data interfaces for data communication with peripheral equipment, such as printer, PC or setpoint controller. Type of protection IP degree of protection II 2 G EEx ib IIC T4 II 2 D IP65 T 50 C IP65 6 Guide for installers 22000435B 05/05

ID3sTx System overview Intrinsically safe power supply There are 3 intrinsically safe power supplies to the weighing terminal and weighing platform for the ID3sTx weighing terminal: Internal battery Type of protection II 2 G EEx ib IIC T4 When the ID3sTx is used in a zone 21 area, the internal battery may only be operated within the housing of the ID3sTx. External battery The terminal is equipped with a 1.5 m cable to enable connection to an external battery. Type of protection II 2 G EEx ib IIC T4 II 2 D IP65 T 120 C Internal AC power supply unit Type of protection II 2 G EEx m e [ib] IIC T4 Preassembled mains cable 5 m In a hazardous zone, the terminal must be connected up to the mains in accordance with the applicable national installation regulations. When the ID3sTx is used in a zone 21 area, the internal AC power supply unit may only be operated within the housing of the ID3sTx. Weighing platforms x METTLER TOLEDO weighing platforms are available for various maximum loads and readability requirements, equipped with explosionproof extension measuring cells. Weighing platforms manufactured by other companies may only be connected to the terminal if they fulfil the specifications in the terminal connection diagram at the end of this guide for installers. DN Tx, PTA459x, PUA579x Type of protection IP degree of protection Preassembled cable DB Tx, DCS...Tx Type of protection IP degree of protection Preassembled cable II 2 G EEx ia IIC T4 II 2 D IP68 T 80 C IP68 5 m long II 2 G EEx ia IIC T4 II 2 D IP67 T 80 C IP67 5 m long Guide for installers 22000435B 05/05 7

System overview ID3sTx PBA430x Type of protection II 2 G EEx ia IIC T4 II 2 D IP65 T 150 C IP degree of protection IP68, IP69K Preassembled cable 1.5 m capacity 30 kg 2.5 m capacity 60 kg RWM1x (0.5 t / 1 t) Type of protection II 2 G EEx ib IIC T6 II 2 D IP67 T 70 C IP degree of protection IP67 Preassembled cable 5 m long System solution Ex1 is required for RWM1x. Up to 4 RWMx can be connected to this system solution. Preassembled cable: 5 m long Dual Channel Fiber Optic Converter The Dual Channel Fiber Optic Converter has 2 data interfaces; it may only by used in the safe zone. An RS232 or CL terminal is available at each data interface. The weighing terminal must be equipped with one or two fiber optic data interfaces for data transfer. 8 Guide for installers 22000435B 05/05

ID3sTx Installation 3 Installation EXPLOSION HAZARD The explosion-protected weighing system must be installed in accordance with the terminal connection diagram at the end of this guide for installers. 3.1 Installing the system modules 1. Install the weighing platform refer to the operating instructions for the weighing platform. 2. Install the weighing terminal. 3. Install the external battery, where applicable. The battery can be installed permanently refer to the operating instructions for the dimensions. 4. Install the Dual Channel Fiber Optic Converter in the safe zone, where applicable. The converter can be installed permanently refer to the operating instructions for the dimensions. 3.2 Connecting the units Connect the units up in the following order in accordance with the terminal connection diagram: 1. Connect the weighing platform to the weighing terminal. 2. Connect the Dual Channel Fiber Optic Converter, where applicable. 3. Connect the power supply. Once all units have been connected up 1. Connect up equipotential bonding as described in Section 3.3. 2. Close the weighing terminal in such a way that the cover latches audibly into place at all four corners. 3.2.1 Connecting the weighing platform up to the weighing terminal CAUTION Risk of measuring errors Use only shielded cable for the weighing platform. Fit connectors onto customer-specific cable as described in Section 3.4. Guide for installers 22000435B 05/05 9

Installation ID3sTx Weighing platforms with preassembled cable 1. Open the weighing terminal. 2. With partially certified scales: Remove the installation cable from the weighing platform terminal. 3. Draw the preassembled weighing platform cable in, making sure that the cable is routed correctly and that the seals are fitted properly. 4. Secure the weighing platform cable in the housing with the pre-mounted cable holders. 5. Connect the cable to the weighing terminal in accordance with the terminal connection diagram. METTLER TOLEDO weighing platforms without preassembled cable Remove the load plate from the weighing platform, connect the cable to the junction box in accordance with the terminal connection diagram and replace the load plate. Comply with the following instructions when using weighing platforms manufactured by other companies 1. Wind the weighing platform cable round the ferrite core three times. Bring the ferrite core as close to the terminal enclosure as possible. 2. Secure the shield cable to the stud bolt. The shield cable must not be routed through the ferrite core! 3.2.2 Connecting the Dual Channel Fiber Optic Converter The Dual Channel Fiber Optic Converter must be installed in the safe zone. Only use the power supply unit specified by METTLER TOLEDO. A weighing system with the Dual Channel Fiber Optic Converter may not be operated in a zone 21 area! EXPLOSION HAZARD Potentially explosive substances in the hazardous zone may be ignited by intensive incident light radiation. The incident light radiation in a hazardous zone must not exceed 0.4526 mw/mm 2. Disconnect the weighing terminal from the power supply before connecting the Dual Channel Fiber Optic Converter to the weighing terminal. 10 Guide for installers 22000435B 05/05

ID3sTx Installation Connecting the fiber-optic cable Preassembled METTLER TOLEDO fiber-optic cables are available in various lengths (max. 300 m). Two cables are required for a bi-directional link. CAUTION Kinks in fiber-optic cables render them useless! 1. Remove the installation cable from the interface terminal. 2. Draw two fiber-optic cables in, making sure that they are routed correctly without any kinks, and that the seals are fitted properly. 3. Connect the two fiber-optic cables to terminals J2 (Transmit) and J3 (Receive) on the Fiber Optic PCB. 4. Screw-fit the threaded bush to the rear panel of the weighing terminal. 5. Cross the two fiber-optic cables over and connect them to terminals J3 (Transmit) and J4 (Receive). Use terminals J5 (Transmit) and J6 (Receive) accordingly for Channel 2. Fiber Optic Print Dual Channel Fiber Optic Converter Comm1 J2 (Transmit) J3 (Receive) J3 (Transmit) J4 (Receive) Channel 1 Fiber Optic Print J2 (Transmit) J5 (Transmit) Comm2 Channel 2 J3 (Receive) J6 (Receive) 3.2.3 Connecting the power supply Fitting the internal battery 1. Place a charged internal battery in the battery compartment on the right-hand side of the enclosure. 2. Tighten the knurled screws securing the cover of the battery compartment, making sure that the seal is fitted properly. Guide for installers 22000435B 05/05 11

Installation ID3sTx Connecting the external battery The weighing terminal is supplied with a 1.5 m long cable to connect it up to an external battery with a bayonet fitting. 1. Connect the bayonet fitting on the battery cable to the charged external battery. 2. Make the equipotential bonding connection. Connecting the internal power supply unit EXPLOSION HAZARD The unit must be connected to the mains by a qualified electrician in accordance with the terminal connection diagram and the applicable national regulations. 3.3 Equipotential bonding Note Equipotential bonding must be installed by a qualified electrician authorized by the operating company. The function performed by METTLER TOLEDO Service is of a purely supervisory and advisory nature. Connect up equipotential bonding (PA) for all items of equipment (ID3sTx, weighing platform, external battery) in accordance with the terminal connection diagram and the applicable national regulations and standards. Ensure that: the enclosures of all items of equipment are connected to the same potential via the bonding (PA) terminals, no compensating current flows via the shielding on the intrinsically safe cable. 3.4 Fitting connectors on the weighing platform cable Customer-specific weighing platform cables must be preassembled in the following manner: Max. length Dimension A (ID3sTx) Dimension B (weighing platform) 100 m 210 mm 50 mm End ferrules with plastic collars, crimped connection Earthing cable gland Dimension A ID card retaining ring Cable (3 x 2 x 0.75 mm 2 ) in accordance with the applicable national regulations for intrinsically safe circuits Cable shield Slide sleeve between wires and shield Dimension B 12 Guide for installers 22000435B 05/05

ID3sTx Installation 1. Cut the cable to length and strip the ends of the cable according to dimensions A/B. 2. Cut the shield back to 7 mm. 3. Strip the insulation off the ends of the wires. 4. Use a crimping tool to fit ferrules onto the ends of the wires. 5. Slide the three rear parts of the earthing cable gland onto the cable. 6. Slide a sleeve between the wires and the shield, being careful not to damage the insulation on the wires! 7. Slide the front part of the cable gland onto the cable and screw it onto the rear section. 3.5 Installing the Fiber Optic PCB J3 J2 Channel 1 Channel 2 1. Open the weighing terminal and remove the blanking plug. 2. Fit the Fiber Optic PCB into one of the slots provided and use the screws supplied with the PCB to secure. 3. Connect the Fiber Optic PCB to the main PCB. To do this, connect the lead supplied with the PCB to terminals J1 on the Fiber Optic PCB and to J3 (Channel 1) or J4 (Channel 2) on the main PCB. 4. Connect the fiber-optic cable as described in section 3.2.2. 5. Secure the fiber-optic cable in the housing with the pre-mounted cable holder. Guide for installers 22000435B 05/05 13

14 Guide for installers 22000435B 05/05 Terminal connection diagramm Table of color codes Exc+ Sen+ Sig+ Sig - Sen - Exc - DNx PTA459x PUA579x RWMx grey yellow white brown green pink B Junction box D..Tx PBA430x PTA459x PUA579x RWMx A B DBx DCSx blue green white red grey black 2) PA Exc+ Sen+ Sig+ Sig Sen Exc A PBA430x green blue white red brown black 1) max. 100 m Ex installed values U i I i P i L i C i 7.14 V 298 ma 0.53 W 0.3 mh 13.4 µf 3) Internal battery PE 2 1 Mains terminal 5) P1 Exc+ Sen+ Sig+ Sig Sen Exc P2 GND +12V Low Bat GND Power supply 230 V AC 1) Cable in accordance with the applicable national regulations for intrinsically safe circuits - Cable 3 x 2 x 0.75 mm 2, shielded and paired - Cable entry via earthing cable gland - Cable in accordance with Section 3.4 of the Guide for installers 2) Equipotential bonding (PA) connected in accordance with the applicable national regulations. Measures must be taken to ensure that the bonding terminals are used to connect all items of equipment to the same potential. No compensating current may flow via the shielding of the intrinsically safe cable. 3) Power supply 230 V AC mains connection in accordance with the applicable national regulations; refer to the rating plate for information concerning mains voltage and frequency 4) Dual-channel fiber-optic converter mains connection in accordance with the applicable national regulations; refer to the rating plate for information concerning mains voltage and frequency 5) Take the relevant creepage distances and clearances into consideration when installing the 12 V DC lines. P4 P3 ID3sTx weighing terminal B A Ausgabe Fiber Optic I/O PCB J3 J4 / Änderung Ersatz für: / Hazardous zone 03/08 Datum Comm1 Power density 0.4526 mw/mm 2 Fiber Optic I/O PCB Comm2 Power density 0.4526 mw/mm 2 External battery 05/05 Schultz Schultz Name METTLER-TOLEDO Bearb. Gepr. 2) PA Datum 05/97 05/97 2) PA Ersetzt durch: Name Schultz Schultz Mettler-Toledo (Albstadt) Waagen und Systeme D-72458 Albstadt Safe zone Dual Channel Fiber Optic Converter Maßstab J3 (Transmit) J4 (Receive) J5 (Transmit) J6 (Receive) Benennung Leitzahl 4) Mains terminal ID3sTx terminal connection 22000249 A3 CL or RS232 CL or RS232 4 Control drawing Control drawing ID3sTx

22000435B Subject to technical changes Mettler-Toledo (Albstadt) GmbH 05/05 Printed in Germany 22000435B Mettler-Toledo (Albstadt) GmbH D-72458 Albstadt Tel. ++49-7431-14 0, Fax ++49-7431-14 232 www.mt.com/support