GC-101. Testing of Biodiesel Using Gas Chromatography. by Bob Armantrout. Presented at: Local Biodiesel: A Biodiesel Coop Conference

Similar documents
MET-Biodiesel Capillary GC Columns

Application Note. Author. Introduction. Energy and Fuels

Performing ASTM 6584 free and total glycerin in BioDiesel using an SRI Gas Chromatograph and PeakSimple software

Free and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN and ASTM D6584

Determination of Free and Total Glycerin in B100 Biodiesel

AppNote 1/2010 KEYWORDS ABSTRACT. Biodiesel, Automation, ASTM D

Analysis of Glycerin and Glycerides in Biodiesel (B100) Using ASTM D6584 and EN Application. Author. Abstract. Introduction

Determination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105

Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011

Going Green: The Analysis of BioFuels

Technical Procedure for Gas Chromatography (GC-FID)

Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS)

GC Analysis of Total Fatty Acid Methyl Esters (FAME) and Methyl Linolenate in Biodiesel Using the Revised EN14103:2011 Method

ASTM D Standard Specification for Biodiesel Fuel (B 100) Blend Stock for Distillate Fuels

Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593

Model 8610C Gas Chromatograph. GC Chassis Types ECD PID FID / DELCD FPD. Mounts up to Six Detectors and Five Injectors

Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel

High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph

Application Note. Authors. Abstract

Analysis of Petroleum Fractions by ASTM D2887

A Single Method for the Direct Determination of Total Glycerols in All Biodiesels Using Liquid Chromatography and Charged Aerosol Detection

Application Note. Authors. Abstract. Energy & Chemicals

Monitoring Biodiesel Fuel Quality

Simulated Distillation Analyzers, Software, Standards, Consumables, Training

Refinery Gas. Analysis by Gas Chromatography WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Biodiesel Analysis Toolkit

Refinery Support from the R&D Laboratory Perspective using Fast & Micro Gas Chromatography

Application Note. Author. Abstract. Energy & Chemicals - Petrochemicals. Edgar Naegele, Agilent Technologies, Inc. Waldbronn, Germany

Sample preparation Methods Analysis. Biodiesel Analysis guide.

Application. Gas Chromatography June 1995

Environmental Testing Solutions from SRI

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application

EFFECT OF BIODIESEL IMPURITIES ON FILTERABILITY AND PHASE SEPARATION FROM BIODIESEL AND BIODIESEL BLENDS. National Biodiesel Conference 2008

Beverage Grade Carbon Dioxide

Characterization of Tiki Torch Fuels

ASTM D for Denatured Fuel Ethanol Automating Calculations and Reports with Empower 2 Software

Determination of fuel system icing inhibitor content of aviation turbine kerosine by HPLC

This document is a preview generated by EVS

SELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS

Reporter. Determination of Free and Total Glycerin in B100 Biodiesel. Liquid Chromatography Sample Handling Gas Chromatography. Standards Accessories

Detection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector

IFPAC 2003 Dr. Berthold Andres

Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC

Agilent and ASTM. Update on Recent Activities. Page 1

Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application

ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column

4001 Transesterification of castor oil to ricinoleic acid methyl ester

Technical Procedure for Pyrolysis-Gas Chromatography/Mass Spectrometry (Py-GC-MS)

Agilent 7693A Automated Liquid Sampler

Nickel Catalyst Tube Accessory 19205A

Using the PSD for Backflushing on the Agilent 8890 GC System

CERTIFICATE OF ACCREDITATION

Alliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE

Detection of Volatile Organic Compounds in Gasoline and Diesel Using the znose Edward J. Staples, Electronic Sensor Technology

The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector

BIODIESEL PRODUCTION IN A BATCH REACTOR 1. THEORY

Dual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector

The Pennsylvania State University. Earth and Mineral Sciences. College of Energy and Mineral Engineering A COMPREHENSIVE EVALUATION AND DEVELOPMENT OF

Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers. Versa HT3. Application Note. Abstract.

Agilent 7693A Automated Liquid Sampler

Troubleshooting Tips & Tricks for your GC Analyzer & CFT Application

GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585

GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS

Technical Procedure for the Examination of Arson Evidence

Agilent 1290 Infinity Autosampler

A NOVEL CONTINUOUS-FLOW REACTOR USING A REACTIVE DISTILLATION TECHNIQUE FOR BIODIESEL PRODUCTION

Optimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System. Jim McCurry Roger Firor Agilent Technologies Wilmington, DE

MB3600-CH30 Laboratory FT-NIR analyzer for biodiesel applications Suitable for production optimization and product quality assessment

Louis Dreyfus Claypool Holdings, LLC. Biodiesel Production Plant Claypool, Indiana

[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS

Agilent Multimode Inlet for Gas Chromatography

Phase Distribution of Ethanol, and Water in Ethyl Esters at K and K

Biodiesel Analysis Utilizing Mini-Scan - Handheld Analyzer V.C. Gordon PhD, Bonanza Labs

Application Note. Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D using AC OXYTRACER

Ultrafast GC Performance in the Real World: Multi Lab Studies for Repeatability & Reproducibility

Agilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve

Choose for Your Instrument. GC Supplies - Agilent Instrument Quick Pick Guide

DANI Transformer Oil Gas Analyzer

Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI)

High-Temperature Simulated Distillation System Based on the 6890N GC Application

Oxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels

Determination of phase diagram of reaction system of biodiesel

Meeting the Requirements of EN12916:2006 (IP391/07) Using Agilent 1200 Series HPLC Systems

CHEMICAL ENGINEERING LABORATORY CHEG 4137W/4139W. Biodiesel Reaction Kinetics

Analytical and Testing Instruments for Renewable Fuels. Shimadzu s Analytical Solutions for the Renewable Fuels Industry

Optimized Method for Analysis of Commercial and Prepared Biodiesel using UltraPerformance Convergence Chromatography (UPC 2 )

GC Supplies - Varian/Bruker Instrument Quick Pick Guide

Nickel Catalyst Tube on a 6820 GC, Accessory G4318A

AZ Chrom s.r.o.

Tennessee Department of Agriculture

Agilent Multimode Inlet

Practical Steps in GC Troubleshooting

2007 B100 Quality Survey Results

Choose for Your Instrument. GC Supplies - Bruker (Varian) Instrument Quick Pick Guide

GC Instruments GC Columns GC Accessories

X-CELL Analytical. Biofuels Analysis. St. Francis Xavier University Analytical Services Lab. Equipment Specification Brochure

ii) PC based works station with monitor and printer having below mentioned specifications:

Method Development for Capillary GC Systems. Slide 1

White Paper. Improving Accuracy and Precision in Crude Oil Boiling Point Distribution Analysis. Introduction. Background Information

PRODUCT MANUAL. CARBONATE REMOVAL DEVICE 200 CRD 200 (4 mm) (P/N ) CRD 200 (2 mm) (P/N )

Transcription:

GC-101 Testing of Biodiesel Using Gas Chromatography by Bob Armantrout (Image Research Machines plc) Presented at: Local Biodiesel: A Biodiesel Coop Conference Strategies For Community-based Biodiesel Groups July 14 July 16, 2006

Triglycerides to Methyl Ester (Image Research Machines plc) www.biology.iastate.edu

Why Glycerin is Bad in Biodiesel From: US Department of Energy Biodiesel Handling and Use Guidelines, Second Edition, March 2006 Free and total glycerin numbers measure the amount of unconverted or partially converted fats and by-product glycerin present in the fuel. Incomplete conversion of the fats and oils into biodiesel can lead to high total glycerin. Incomplete removal of glycerin can lead to high free glycerin and total glycerin. If these numbers are too high, storage tank, fuel system, and engine fouling can occur. Fuels that exceed these limits are highly likely to cause filter plugging and other problems.

How Glycerin is Removed from Biodiesel Converting the triglycerides, diglycerides, and monoglycerides to methyl esters to reduce the levels of bound glycerin. Washing the esters to remove glycerin.

ASTM Specification for Glycerin D6751-06e1 stipulates: Free Glycerin 0.020% mass or less Total Glycerin 0.240% mass or less The total glycerin method is used to determine the level of glycerin in the fuel and includes the free glycerin and the glycerin portion of any unreacted or partially reacted oil or fat. Low levels of total glycerin ensure that high conversion of the oil or fat into its mono-alkyl esters has taken place. High levels of mono-, di-, and triglycerides can cause injector deposits and may adversely affect cold weather operation and filter plugging. (ASTM 6751-06e)

The Equipment Hewlett Packard - Model 5890A-II Gas Chromatograph Unit includes an autosampler, column, pc, software, and gas regulators AIR will provide delivery, installation, familiarization and a 6-month warranty AIR will provide service outside of warranty for a fee. Total quote price: $13,382.15 (Oct. 2005) Work will begin upon receipt of purchase commitment, leadtime is 1 to 2 weeks.

What the GC does Precisely measures the percent mass of free glycerin, mono, di and triglycerides in a sample of methyl esters.

How the GC works A Gas Chromatograph (GC) is a chemical analysis instrument for separating chemicals in a sample. It uses a thin capillary fiber known as the column, through which different chemicals pass at different rates depending on various chemical and physical properties. The function of the column is to separate different components, causing each one to exit the column at a different time. As the chemicals exit the end of the column, they are detected and identified electronically. (from Wikipedia)

How the GC works An FID is an ion detector which uses an air-hydrogen flame to produce ions. As components elute from the GC's column they pass through the flame and are burned, producing ions. The ions produce an electric current, which is the signal output of the detector. The greater the concentration of the component, the more ions are produced, and the greater the current. (from Wikipedia)

ASTM D6584 prescribes the method for GC setup, sample prep and chromatogram analysis. 4. Summary of Test Method 4.1 The sample is analyzed by gas chromatography, after silyating with N-methyl-N-trimethylsilyltrifluoracetamide (MSTFA). Calibration is achieved by the use of two internal standards and four reference materials. Mono-, di-, and triglycerides are determined by comparing to monoolein, diolein, and triolein standards respectively. Average conversion factors are applied to the mono-, di-, and triglycerides to calculate the bonded glycerin content of the sample. Copyright by ASTM Int'l

Maintenance and Repair Calibration Use Standards to calibrate GC for glycerol, monolein, diolein, butanetriol, tricaprin. Retention time (RT) changes indicate instrument status Decreasing RTs are are typically caused by column aging Increasing RTs typically caused by leaks in septa, connections, low gas pressures. Repair Septa replacement Column replacement Syringe replacement

Other Equipment and Supplies Additional Lab Ware required to use the GC: Analytical Balance - $1500 Pipetter 100 microliter (µl) $250 15ml vials for sample prep 2ml vials with septa top Compressed Gases Helium, Hydrogen, Compressed Air USP grade, tanks run about $50 to $70 each Chemical Standards, Carriers Calibration Standards monolein, diolein, triolein Butanetriol and Tricaprin - $150 each per 5ml Heptane used as a carrier for sample Hexane for syringe cleaning

Facility Considerations Sturdy Desk to set GC on Dedicated power circuit Small refrigerator for storage of standards Fire Cabinet for storage of reagents Proper storage of gas bottles

Equipment and Material Suppliers Analytical Instrument Recycle www.aironline.com Telephone: 303-215-0468 Toll Free: 888-909-0468 (U.S.) Sigma-Aldrich/Supelco www.sigmaaldrich.com Telephone: 800-325-3010 (U.S.) 6390 Joyce Dr. Suite 200W Golden, CO 80403

Photo Essay of GC Testing Process

Sample Preparation (logbook & glassware) Weigh the Sample (analytical balance)

Lab refrigerator Add Internal Standards (micropipetter )- 100 microliters (µl) each of MSTFA, and internal standards 1 & 2 Add heptane 8 ml

Load sample to GC vial Place vial in GC autosampler

Load Sample Data into GC Software Start the Run

Analysis of Chromatograms

Approximate Relative Retention Times

Calculating Glycerin Levels

Chromatograms The Good

Chromatograms The Good

Chromatograms The Good

Chromatogram The Bad

Chromatogram The Ugly

Understanding your Processes Conversion recipe development

Understanding your Processes Tracking glycerin removal through washing

Verifying your testing process Lab cross checking is a good way to assure your lab equipment and procedures are providing valid results

Verifying your testing process

Third Party GC Testing Magellan will test biodiesel for Free & Total Glycerin for under $100 A 10 ml sample is required

Third Party GC Testing Intertek will test biodiesel for Free & Total Glycerin for $275 A 10 ml sample is required