Avanti J Series Centrifuges Preinstallation

Similar documents
Preinstallation Requirements for Avanti J Series Centrifuges

Instructions For Use. Avanti J-26 XPI. High-Performance Centrifuge. PN J326XPI-IM-4AD December 2013

J320XP-IM-5. Avanti J-20 XP. Expanded-Performance Centrifuge. Instruction Manual

J326XP-IM-4. Avanti J-26 XP. High-Performance Centrifuge. Instruction Manual

Power Supplies PSS Series

MASTERYS IP+ 10 to 80 kva

Industrial Control Transformers

OPERATIONS MANUAL. Mini Centrifuge Model MCF Certified

INSTALLATION AND OPERATING INSTRUCTIONS OF THE INTERNATIONAL ISOBOX SERIES ISOLATION TRANSFORMERS.

SineWave Guardian TM 380V 600V INSTALLATION GUIDE. Quick Reference. ❶ How to Install Pages 6 17 ❷ Startup/Troubleshooting Pages WARNING

Models beginning with MO, MK, 7B, 6S, 6F or 5B. Product Description. Technical Specifications. Installation Instructions.

Green Power 2.0 MASTERYS GP

1329I Integrated Drive/Motor CE Filter Option Kit

WELLS RURAL ELECTRIC COMPANY Adopted: March 1976 Revised: March 21, 2017 Reviewed: August 21, 1998 RULE NO. 2

Cabinet and Rack Installation

SECTION ENCLOSED SWITCHES AND CIRCUIT BREAKERS

Cabinet and Rack Installation

High Frequency SineWave Guardian TM

Limit Switch D4D- N. Small, Economical Switch Featuring a Positive Opening Mechanism and Conforming to the CE Mark

INSTALLATION AND OPERATING INSTRUCTIONS OF THE INTERNATIONAL ISOBOX SERIES ISOLATION TRANSFORMERS.

INOVA HIGHTECH Ltd. MEP 002/003 Auto Starter Manual

Application Description

Control Power V AC, 50/60 Hz 150-C30NBD 150-C30FHD

Back-UPS ES APC Back-UPS ES, 6 outlet 350VA, 120V, without auto shutdown software

OPTEVA 378 ENHANCED CASH RECYCLER TERMINAL FRONT LOAD LOBBY UNIT PLAN VIEW FRONT VIEW SIDE VIEW

Product Data Sheet ACi 4420 HHR

Cabinet and Rack Installation

UNDERGROUND SERVICES - SECONDARY

GJL 100 A Molded Case Circuit Breaker GJL 75 A Motor Circuit Protector

DELPHYS GP. Green Power 2.0 range 160 to 1000 kva/kw

The Enclosed Altistart 22 Soft Start/Soft Stop Motor Controller

Cabinet and Rack Installation

Mipco Connections for Refrigerated Containers. No One Does It Like Mipco. The Mipco. Advantage. Reliability. Safety. Easy Installation & Service

3 Uninterruptible Power Systems

2NC (slow-action) Direct. Model. opening

APEX DRIVE USER GUIDE ADDENDUM: LVD INSTALLATION INSTRUCTIONS

ICT TRANSFORMERS. Industrial Control Transformers provide a low and safe control voltage for the operation of many electromagnetic devices

MODEL 540 USER MANUAL INCORPORATING MODEL 530 TEMPERATURE CONTROLLER

Rugged steel cabinet construction ensures quiet operation and long term reliability.

ORION PLUS three-phase kVA

M T E C o r p o r a t i o n MATRIX FILTER. SERIES B Volts, 50HZ USER MANUAL PART NO. INSTR REL MTE Corporation

1.2. Reduced Voltage Motor Starters. Contents Description Type S611, Solid-State Soft Starters... Type S801+, Soft Starters. Solid-State Starters

POWER Wave XR Clean Sophisticated Reliable

RCD Air Cooled Condensing Unit (50/60 Hz)

50 Hz Mini/Micro Computer Regulator (MCR) Series Installation and Operation Manual

dv Sentry TM 208V 600V INSTALLATION GUIDE Quick Reference ❶ How to Install Pages 6 14 ❷ Startup/Troubleshooting Pages WARNING

Matrix AP 400V 690V INSTALLATION GUIDE. Quick Reference. ❶ How to Install Pages 6 20 ❷ Startup/Troubleshooting Pages WARNING

R4A5. Product Specifications EFFICIENT 14 SEER/ EER AIR CONDITIONER ENVIRONMENTALLY SOUND R 410A REFRIGERANT

3 Uninterruptible Power Systems

Centrifuge Operator / Service Manual

Product Data Sheet ACi 4420 MLR

P66 Series Electronic Fan Speed Controls

Modifiable TITAN Horizontal Motors Accessories and Modifications

Contactors. CL1115/02, CL1215/02, CL1315/02 1, 2 and 3 pole AC and DC NO contactors for voltages up to 1,500 V Catalogue C25.en

MA-5200 Series MA-5300 Series

Product Guide AC Voltage Control Devices. Guide Guide. Popular Variable Transformers and AC Power Supplies.

Features, Benefits and Functions

Installing the Switch Chassis

SOLSTART Instruction Manual

Electric. PN April Series Electric Oven

EVoInnovate EVSE Model Numbers: EVO & EVO User Manual & Installation Guide. Revision 1.0

A. This Section includes ac, enclosed controllers rated 600 V and less, of the following types:

Parker AC10 Frequency Inverter (to 22kW) Easy Start Guide

RVS-AX Instruction Manual

AF series contactors (9 2650)

Switching DC Power Supply

M T E C o r p o r a t i o n MATRIX FILTER. SERIES B Volts, 50HZ USER MANUAL PART NO. INSTR REL MTE Corporation

AF series contactors (9 2650)

Bulletin 1497A Machine Tool Transformers

DNX 3 MCB 4500 A / 6 ka Phase + Neutral, neutral on left side

SOS SERIES SOS1 SOS2. Spares On Site Battery Cabinet Installation Guide rEV3

RX 3 M.C.B A up to 63 A

The Power of Reliability INSTRUCTION MANUAL

VFD Best Practices: Getting the Most from Your VFD Investment

SECTION MOTOR CONTROL

Manual Reset Limit Switch

LED Driver Linear / area fixed output

D4D-N. Safety Limit Switch. Small, Economical Switch Featuring a Positive Opening Mechanism and CE Marking. H Contacts opened by positive opening

EFFICIENCY 17 SEER TWO STAGE AIR CONDITIONER WITH OBSERVER

materials and workmanship for 2 years. Should the centrifuge require warranty or Phone: or Fax:

AFAIR12B. Features. Efficiency: 12 SEER. Warranty: 10-year limited compressor 5-year limited parts. Capacity: 1-1/2 to 5-ton.

R4A4. Product Specifications EFFICIENT 14 SEER AIR CONDITIONER ENVIRONMENTALLY SOUND R 410A REFRIGERANT

1250W 12VDC PowerVerter Ambulance/EMS Inverter/Charger with 2 Hospital Grade Outlets

CSA6. Product Specifications. SoftSound SXC 2600

Recommendations Multipin connector Wired mode (permanent and pulse signal) Positioning mode

Reduced Voltage Motor Starters

PD-Series Ground Fault Circuit Protection

INVERTER FOR DC CRANE CABIN AIR CONDITIONING UNIT 3KW VERSION

WCA4**4. Product Specifications

Safety Limit Switch

Technical Information

ABB central inverters ULTRA-750/1100/ kW to 1560kW

INSTALLATION MANUAL AP60B INSTALLATION MANUAL

24 KVA International Universal IsoTransformer

Inverter User Manual

Uninterruptible Power Systems

Junction Boxes & Enclosures JBES and ECES Series: 316L Stainless Steel Enclosures Increased Safety

ATS22D88Q soft starter-ats22-control 220V-power 230V (22kW)/ V(45kW)

Informational Distribution List:

BRAKE MOTORS. 1. Brake motors TECHNICAL CATALOGUE. M Series / Standard / IEC

Transcription:

J325-TB-001AP February 2009 Avanti J Series Centrifuges Preinstallation INTRODUCTION Purchase of an Avanti J series centrifuge includes installation and Basic Instrument Training for two operators. This guide is provided to assist you in preparing your laboratory and the designated operators so that installation and training can be completed efficiently and successfully. SPECIFICATIONS WEIGHT CONTROL FEATURES Ambient temperature Avanti J-HC.........................15 to 40 C Avanti J-E..........................15 to 35 C All other Avanti J series................16 to 38 C PHYSICAL DATA Electrical supply........................... Class I Maximum heat dissipation into room under steady-state conditions........ 2.0 kw (6900 Btu/h) Humidity restrictions........... <95%, noncondensing Installation (overvoltage) category................ II Pollution degree................................2 * 310 kg (680 lb) Avanti J-30I 290 kg (640 lb) Avanti J-26XP series 300 kg (660 lb) Avanti J-HC 267.4 kg (589 lb) Avanti J-E POWER REQUIREMENTS Nominal Instrument Voltage Rating Line Frequency/ Current Plug Included DIMENSIONS Height (door closed) Avanti J-E.....................91.4 cm (36.0 in.) All other Avanti J series.......... 116 cm (45.5 in.) Height (door open) Avanti J-30I..................141.0 cm (55.5 in.) Avanti J-26XP series............149.0 cm (58.5 in.) Avanti J-HC..................152.0 cm (59.9 in.) Avanti J-E....................152.4 cm (60.0 in.) Width Avanti J-E.....................63.5 cm (25.0 in.) All other Avanti J series..........71.0 cm (28.0 in.) Depth (including air diverter extending from back panel) Avanti J-E....................95.9 cm (37.75 in.) All other Avanti J series....... 102.0 cm (40.25 in.) Single-phase, 200/208/240 V INSTALLATION REQUIREMENTS Installation Category 50/60 Hz/24 A yes Single-phase, 230 V 50 Hz/24 A no Three-phase, 220/ 50 Hz/12 A no 380 V + Neutral This centrifuge is classified as installation category II. * Normally only nonconductive pollution occurs; occasionally, however, a temporary conductivity caused by condensation must be expected. Unbalanced three-phase. Split for single-phase operation internally. 2004 2009 Beckman Coulter, Inc.

Ventilation The centrifuge must have adequate air ventilation to ensure compliance to local requirements for vapors produced during centrifuge operation. Biosafety Considerations If sample containment is required, Beckman Coulter can provide a Class II or III biosafety cabinet specially designed to house the centrifuge. Contact your local Beckman Coulter office for more information. Electrical Requirements Avanti J series centrifuges come with one of three types of power cords. For single-phase 200/208/240 V connections, a permanently attached three-wire UL/CSA approved cord with a NEMA type 6-30P plug is provided. For single-phase 230 V connections, a permanently attached three-wire CENELEC harmonized cord without plug is provided. For three-phase connections, a permanently attached five-wire CENELEC harmonized cord without plug is provided. Terminate the open end of the harmonized cords with a certified single or threephase connector suitable for the power supplied in the country of intended use. Wire Insulation Color Green/Yellow Light Blue Required Wire Connections Terminal Earth-ground Neutral Brown Live or Line Black Live or Line (three-phase (qty 2) connections only) Symbol Harmonized N L North American L L L Power to the centrifuge should originate directly from a main power line transformer at a power source known to be clear of erratic loads, spikes, and electromagnetic interference. Make sure there are properly rated thermal circuit breakers at the wall service panel to protect the centrifuge circuit. If fuses must be used instead of the specified circuit breakers, the fuses may require a rating of greater than 30 amperes (for single-phase centrifuges) or greater than 16 amperes (for three-phase centrifuges). Prior to instrument purchase, the line voltage should be checked with a voltmeter and the appropriate instrument ordered. After delivery, verify that the voltage indicated on the centrifuge s name rating plate (on the back of the control head) agrees with the measured line voltage. If the measured voltage is not within the indicated range, contact your Beckman Coulter Field Service representative. Your representative can rewire the centrifuge to adapt it to the available line voltage. To ensure safety, the instrument should be wired to a remote emergency switch (preferably outside the room where the centrifuge is housed, or adjacent to the exit from that room), in order to disconnect the instrument from the main power source in case of a malfunction. Figure 1 shows the power connection for single-phase Avanti J series centrifuges; Figure 2 shows the power connection for three-phase Avanti J series centrifuges. The single-phase system includes earth ground and two power leads with 30-ampere circuit breakers. The threephase system includes earth ground and three power leads with minimum 16-ampere circuit breakers and common neutral connection. 30-ampere Circuit Breaker Wall Outlet: Hubbell 9330, Bryant 96-30-FR, or Equivalent (NEMA 6-30 R) Earth Ground 30-ampere Circuit Breaker Measured Line Voltage Avanti J series centrifuges require a single grounded outlet, either 50/60 Hz single-phase, 180 V to 264 V, 30 A, or 50 Hz three-phase, 313 V to 457 V plus neutral, 16 A. Install only one Avanti J centrifuge per circuit. Figure 1. Avanti J Series Centrifuge Single-Phase Power Connection 2 Avanti J Series Preinstallation Instructions

Measured Line Voltage Phase to Phase 16-ampere Circuit Breaker Neutral 16-ampere Circuit Breaker Additional Requirements for Three-Phase Power Connections For three-phase power service to the Avanti J, use the Y connected three-phase configuration shown in Figure 3. Also note the following information: 16-ampere Circuit Breaker Figure 2. Avanti J Series Centrifuge Three-Phase Y Power Connection Earth Ground The steady state current draw of the centrifuge can be as high as 12 amperes per phase, depending on the voltage. Inrush current to the centrifuge is 100 amperes for up to 2 seconds during the refrigeration start period. Circuit breakers, whether thermal or magnetic actuating, must be a motor start delay type.! WARNING To reduce the risk of electrical shock, this equipment uses a three-wire or five-wire electrical cord and plug to connect the centrifuge to earth-ground. To preserve this safety feature, make sure that the matching wall outlet receptacle is properly wired and earthgrounded. Check that the line voltage agrees with the voltage listed on the name rating plate affixed to the centrifuge. The centrifuge can tolerate a drop of 15 volts during the start period at low line voltage (see Figure 3 for the acceptable voltage ranges). The main power supply cable for the centrifuge must be of sufficient wire gauge to provide this condition. The wire gauge required at a particular site must be determined by a facilities engineer at that site. Space Requirements Dimensions for the Avanti J-26XP series, and Avanti J-30I centrifuges are shown in Figure 4; for the Avanti J-HC centrifuge in Figure 5, and for the Avanti J-E in Figure 6. In addition to space for the instrument itself, allow a 7.7-cm (3-in.) clearance on each side of the instrument and a 16-cm (6.25-in.) clearance behind the instrument for air circulation. Correct Configuration Y Incorrect Configuration Delta Safety Ground 312 456 180 264 N Circuit Common (Neutral) Figure 3. Correct and Incorrect Three-Phase Configurations Avanti J Series Preinstallation Instructions 3

cm in. 86 34 Air Exit Power Cord 1.8 m 6.0 ft Air Diverter 16 Depth: 6.25 Air Intake 86 34 116 45.5 141 or 55.5 149 58.5 (see SPECIFICATIONS) 71 28 Figure 4. Rear View and Dimensions of the Avanti J-26XP Series and Avanti J-30I Centrifuges cm in. 86 34 152 59.9 Air Exit Power Cable 1.8 m 6.0 ft. Air Diverter Depth 6.25 16.0 Air Intake 84.6 33.3 86 34 116 45.5 71 28 Figure 5. Rear View and Dimensions of the Avanti J-HC Centrifuge Figure 6. Rear View and Dimensions of the Avanti J-E Centrifuge 4 Avanti J Series Preinstallation Instructions

The air diverter, shown in Figures 4, 5, and 6 should touch (or nearly touch) the wall behind the centrifuge. The power cord should be placed to one side of the air diverter. To avoid damaging the power cord when installing or moving the centrifuge, be sure to move the cord out of the way before pushing the centrifuge against the wall. SUPPLIES FOR INSTALLATION Avanti J series centrifuges come with a ship kit, which includes the following: Anchoring kit (includes hardware and installation instructions) Instruction manual and accompanying documents Foot pedal pad (not included with Avanti J-E) Air filter We recommend that you order sufficient quantities of your required labware in advance to avoid delay in instrument operation. For your convenience, all Avanti J series centrifuge supplies and accessories are described in the Beckman Coulter High Performance, High Speed, High Capacity Rotors, Tubes & Accessories catalog (BR-8102, available at www.beckmancoulter.com). SECURING THE CENTRIFUGE TO THE FLOOR For the Avanti J series centrifuge to meet CE requirements, it must be secured to the floor using the anchoring hardware kit (part number 367459 for the Avanti J-HC, 369168 for the Avanti J-E, or 364996 for all other models) shipped with the centrifuge. This will prevent the centrifuge from moving in the unlikely event of a rotor mishap. Instructions for installing the anchoring hardware (publication J3HC-TB-002 for the Avanti J-HC, SJ-TB-003 for the Avanti J-E, or J325-TB-003 for all other models) are packaged with the kit. Two steps are required for installation: 1. Drilling holes in the laboratory floor where the centrifuge will be located. The anchoring instructions include a full-sized template which shows where these holes must be drilled in relation to the wall behind the centrifuge. This step must be done by the customer. NOTE The Beckman Coulter representative installing the centrifuge will need clearances of approximately 61 cm (2 ft) on each side of the instrument to attach the shackles through the brackets. After installation, the required clearances are 7.6 cm (3 in.) at the sides and 16 cm (6.25 in.) at the back of the instrument. 2. Attaching the mounting brackets to the floor, and inserting shackles through the floor brackets and the brackets on the centrifuge frame. This step will be done by the Beckman Coulter Field Service representative during installation. NOTE Beckman Coulter representatives are not equipped to drill holes in your floor. The holes must be drilled before installation. BIO-SAFETY LEVEL 3 INSTALLATION For laboratories with epoxy aggregate floors, such as BSL-3 labs, a non-invasive installation kit (PN 393830) is available for Avanti J series centrifuges (except for the Avanti J-HC). The kit which consists of an adhesivebacked mounting plate, is CSA certified for use on epoxy aggregate floors only. ROTOR USE INFORMATION! CAUTION Do not use the Beckman Coulter JA-10, JS-7.5, JA-14, or JS-13 rotors in an Avanti J series centrifuge before reading this information. Avanti J Series Preinstallation Instructions 5

Rotors used in Avanti J series centrifuges must have drive pins in the rotor drive hole (see Figure 7). These drive pins engage with the centrifuge drive hub to ensure that the rotor does not slip during acceleration. Some Beckman Coulter rotors, including the JA-10, JS-7.5, JA-14, and the JS-13, were originally manufactured without drive pins because these rotors did not need pins to run in J2 series centrifuges. Check all J2 series rotors for drive pins before using them in an Avanti J series centrifuge. To check for drive pins, hold the rotor up or turn it on its side and look into the drive hole. If you do not see two metal pins near the top of the hole, do not use the rotor in the Avanti J. Contact Beckman Coulter Field Service for information on returning the rotor to the factory for upgrading.! CAUTION JS-4.2 and JS-4.2A rotors manufactured before January, 2000 must be modified before use in the Avanti J-HC. (Rotors manufactured before January, 1997 cannot be modified.) Contact Beckman Coulter Field Service for more information. OPERATOR TRAINING REQUIREMENTS Immediately following the installation of your Avanti J series centrifuge, a Beckman Coulter Field Service representative will train a maximum of two operators in basic instrument operation. Additional operators can be trained during later sessions (maximum of two per training session). Contact Beckman Coulter Field Service for information. The successful implementation of your new Avanti J series centrifuge depends upon the training that your operators receive. To ensure a focused and optimized training session, the following guidelines have been established: Basic Instrument Training requires a maximum of two (2) hours. It is imperative that the operators are available and uninterrupted throughout this initial training period. No more than two operators can be trained on the instrument at any given time. Learning is accelerated when the operators have hands-on instruction. With only one centrifuge control panel, this cannot be achieved with more than two operators. BECKMAN Rotor Drive Hole Drive Pins (Angled pins shown. Pins can also be vertical or horizontal.) Avanti J Centrifuge Drive Spindle Hub Figure 7. Checking the Rotor for Drive Pins 6 Avanti J Series Preinstallation Instructions

Preinstallation Checklist The following is a list of items necessary for installation of the Avanti J series centrifuge system. Use this checklist to verify the items that you have and other items as you receive them. To ensure a successful installation, please confirm that you have these items prior to scheduling the installation with your local Beckman Coulter service office. AVANTI J SERIES CENTRIFUGE Avanti J Series Centrifuge Power requirements Space requirements Centrifuge instruction manual; preinstallation instructions Anchoring kit and anchoring instructions OPERATOR Up to 2 operators available for 2 hours training at time of installation ROTORS/CONSUMABLES Which rotors will be available at time of installation? What type of tubes and bottles will be on hand at time of installation? SERVICE Has a service contract been ordered? Instrument certification? Avanti J Series Preinstallation Instructions 7

Beckman Coulter, Inc. 250 S. Kraemer Blvd. Brea, California 9282 Sales and Service: 1-800-742-2345 Internet: www.beckmancoulter.com 2009 Beckman Coulter, Inc. All rights reserved