Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes Installation and Maintenance Rev. 1 3/04

Similar documents
7777D In-Line Mounting for Durafet II ph Electrodes Installation and Maintenance

7777 Immersion/In-Line Mounting for Durafet II ph Electrodes Installation and Maintenance

7777 Immersion/In-Line Mounting for Meredian II ph/orp Electrodes Installation and Maintenance Manual

7774 Series Stainless Steel Insertion/Removal Assembly for Use with Meredian Electrodes

7774 Series Stainless Steel Insertion/Removal Assembly for Use with Durafet II ph Electrodes Rev. 0 5/99

EQUIVALENT BASIC CIRCUIT

HercuLine 2000 Series Gear Kit Replacement Instructions

U00X ULTRASONIC LEVEL SWITCH. Ultrasonic Liquid Level Switches INSTALLATION AND OPERATIONS MANUAL. For Models: U002, U003 & U004

NEW QUICK CABLE-TO-SENSOR RELEASE,

* * Rev. 9 01/18 Product Manual

CONTACTING CONDUCTIVITY GENERAL PURPOSE SENSORS Instruction Manual

Insertion/Submersion Flow Through Sensor

4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual Rev. 1 7/99

11280A HercuLine Series Actuator. Specification. Non-Contact Position Sensing and Variable Speed Motor Control. Overview. Features. Actuator Operation

Accessories Guide ph/orp/conductivity Sensors

Signet DryLoc ph/orp Electrodes

Signet DryLoc ph/orp Electrodes

SIGNET 2760 DryLoc ph/orp Preamplifier

Signet ph/orp Electrodes

General Purpose ph/orp Sensors

Installation and Operating Manual

Rosemount 402 and 402VP

Signet 3719 ph/orp Wet-Tap

10260 Retrofit to 10260A Non-Contact Sensor Instructions Kits through -512

U00X ULTRASONIC LEVEL SWITCH. Ultrasonic Liquid Level Switches INSTALLATION AND OPERATIONS MANUAL. For Models: U002, U003 & U004

Liquid Conductivity Two & Four Electrode Sensors

ph/orp Sensor Model 399 ph/orp

DL1000 Electrode Mountings

Instruction Manual. Model HC1 Series Electrodless Conductivity Sensors

Signet ph/orp Electrodes

Models 396/396P/396R. Retraction/Submersion/Insertion ph/orp Sensor

Signet 2760 DryLoc ph/orp Preamplifier

10260S Series Actuators Auto-Manual Switch Replacement Instruction

Installation and Service Instructions for EX Series Explosion-proof Switches

General Purpose ph/orp Sensors

INSTALLATION INSTRUCTIONS SWITCH HOUSING SUPPLIED ONLY AS AN ASSEMBLY OF ENCAPSULATED SWITCH & LEADS. FLOAT SUPPLIED ONLY WITH ENCAPSULATED MAGNET.

VSP SERIES Installation, Operation and Maintenance Manual

Special Purpose Transmitters

For additional information, please visit our website at SENSOR SPECIFICATIONS

INSTRUCTIONS PARTS LIST This manual contains important warnings and information. READ AND RETAIN FOR REFERENCE

Microcontroller Unit (MCU) LCD Hour Meters/Counters. LM Series. Datasheet

Signet ph/orp Electrodes

ACTIVE ELECTRODELESS SENSORS Instruction Manual

Signet Conductivity/Resistivity Electrodes

MICRO SWITCH Solenoid Safety Interlock Switches GKL/R Series Issue 2. Datasheet

Signet Conductivity/Resistivity Electrodes

VL Series. General Purpose Compact Limit Switches. Sensing and Control

onlinecomponents.com VL Series General Purpose Compact Limit Switches Sensing and Control

OPERATING INSTRUCTIONS PLEASE READ CAREFULLY

SETRA Operating Instructions Series DPT 230 Pressure Transducers

User s Guide CDCE-90-X SERIES. Conductivity Sensors. Shop online at omega.com SM

Installation Instructions

Signet V(D) to V(D) PVDF Conductivity Electrodes

Load Accessories. Test & Measurement Sensors DESCRIPTION TONGUE SHACKLE FEATURES PULL PLATE FEATURES YOKE SHACKLE FEATURES LOAD BUTTON FEATURES

OPERATING INSTRUCTION MANUAL MS SERIES CONDUCTIVITY CELLS

Installation, Operation and Maintenance of Airflex Quick Release Valve

Defender 3000 Series Base Instruction Manual

HB551 ph Electrodes and Accessories

OPERATING INSTRUCTION MANUAL

V5011N Two-Way Threaded Globe Valve

TUpH Retraction/Submersion/Insertion ph/orp Sensors

1 AMP CURRENT SOURCE

Rosemount 3300HT/3300HTVP/3400HT/ 3400HTVP/3500P/3500VP

BULLETIN NO.ELEC IM121/10A Replaces IM121/09A

OPERATING INSTRUCTION MANUAL

Industrial Pressure Transmitters

NEO-DYN MODEL 100P ENCLOSURE 7 ADJUSTABLE EXPLOSION-PROOF PRESSURE SWITCH

UniMag DT Series Electromagnetic Flow Tubes

3.5mm Coaxial Calibration Kit

Model Q45H/ Residual Chlorine Monitor...

OPERATING INSTRUCTION MANUAL


Analytical Instruments Two-Electrode Conductivity Sensors

Encapsulated Differential ph and ORP Sensors (available in LCP, foreground, or Ryton body materials)

Operating manual. ba55321de LRD Conductivity measuring cell to be installed in pipelines. ba55321de01 12/00

855 Filtrate Pump, Close Coupled and Overhead Mounted Pumps

ABB MEASUREMENT & ANALYTICS DATA SHEET. TBX5 ph, Redox (ORP) sensors with diagnostics

INSTALLATION, CALIBRATION & TROUBLESHOOTING MANUAL

SWITCH HOUSING SUPPLIED ONLY AS AN ASSEMBLY OF ENCAPSULATED SWITCH & LEADS. FLOAT SUPPLIED ONLY WITH ENCAPSULATED MAGNET.

F PRO. 12 Volt Electric Sprayer. Part #AS76 Copyright 2014 by Hydro-Force Rev. 1/14

Retractable ph/orp Sensors

Battery Powered Hydraulic Pump; Kit PN [ ]

10 S e r i e s PRESSURE SWITCH. features

Pressure Transmitter. Informational & Installation Manual

MODEL L-8 HARWIL CORPORATION 541 KINETIC DRIVE, OXNARD, CA TEL: (805) FAX: (805)

OWNER S MANUAL. Model 9880 Intrinsically Safe Air Velocity Meter

Installation Instructions

AP300 Series Combination ph/redox (ORP) Sensors. Measurement made easy. A reliable economic solution for a wide range of applications

Conductivity Sensors. Model 140, 141, 142. Product Data Sheet series/rev.b February 2011

BOILER BLOWDOWN CONDUCTIVITY CONTROLLER (CONTINUOUS METHOD)

Endurance TM Conductivity Sensors

Measurement and Instrumentation

PRG 350 Series Stainless Steel Regulators

Retractable ph/orp Sensors

Component Test Fixture

P70, P72, and P170 Series Controls for High Pressure Applications

SIGNET 2750 ph/orp Sensor ENGLISH

MA-5200 Series MA-5300 Series

KOBOLD DF-MA Flow Transmitter. User Instructions

Lubricator Gun: 10,000 psi (700 bar) Maximum Delivery Pressure when disconnected from Dispenser

Transcription:

Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes 70-82-25-116 Rev. 1 3/04 Industrial Measurement and Control

Copyright, Notices, and Trademarks Printed in U.S.A. Copyright 2004 by Honeywell Revision 1 March 2004 WARRANTY/REMEDY Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship. Contact your local sales office for warranty information. If warranted goods are returned to Honeywell during the period of coverage, Honeywell will repair or replace without charge those items it finds defective. The foregoing is Buyer's sole remedy and is in lieu of all other warranties, expressed or implied, including those of merchantability and fitness for a particular purpose. Specifications may change without notice. The information we supply is believed to be accurate and reliable as of this printing. However, we assume no responsibility for its use. While we provide application assistance personally, through our literature and the Honeywell web site, it is up to the customer to determine the suitability of the product in the application. Industrial Measurement and Control Honeywell 1100 Virginia Drive Fort Washington, PA 19034 Durafet III is a trademark of Honeywell Hastelloy is a trademark of Haynes International, Inc Teflon is a trademark of E. I. DuPont De Nemours & Company, Inc. Other brand or product names are trademarks of their respective owners. ii Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes 3/04

About This Document Abstract This manual provides immersion and in-line installation instructions for Durafet III ph Electrodes as part of a 7777DVP electrode assembly. Contacts Contacts World Wide Web The following lists Honeywell s World Wide Web sites that will be of interest to our customers. Honeywell Organization Corporate Industrial Measurement and Control International WWW Address (URL) http://www.honeywell.com http://www.honeywell.com/imc http://www.honeywell.com/business/global.asp Telephone Contact us by telephone at the numbers listed below. Organization Phone Number United States and Canada Honeywell 1-800-423-9883 Tech. Support 1-888-423-9883 Q&A Faxback (TACFACS) 1-800-525-7439 Service Asia Pacific Honeywell Asia Pacific (852) 2829-8298 Hong Kong Europe Honeywell PACE, Brussels, Belgium [32-2] 728-2111 Latin America Honeywell, Sunrise, Florida U.S.A. (854) 845-2600 3/04 Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes iii

Symbol Definitions The following table lists those symbols used in this document to denote certain conditions. Symbol Definition This CAUTION symbol on the equipment refers the user to the Product Manual for additional information. This symbol appears next to required information in the manual. WARNING PERSONAL INJURY: Risk of electrical shock. This symbol warns the user of a potential shock hazard where HAZARDOUS LIVE voltages greater than 30 Vrms, 42.4 Vpeak, or 60 Vdc may be accessible. Failure to comply with these instructions could result in death or serious injury. ATTENTION, Electrostatic Discharge (ESD) hazards. Observe precautions for handling electrostatic sensitive devices Protective Earth (PE) terminal. Provided for connection of the protective earth (green or green/yellow) supply system conductor. Functional earth terminal. Used for non-safety purposes such as noise immunity improvement. NOTE: This connection shall be bonded to protective earth at the source of supply in accordance with national local electrical code requirements. Earth Ground. Functional earth connection. NOTE: This connection shall be bonded to Protective earth at the source of supply in accordance with national and local electrical code requirements. Chassis Ground. Identifies a connection to the chassis or frame of the equipment shall be bonded to Protective Earth at the source of supply in accordance with national and local electrical code requirements. iv Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes 3/04

Contents 1. INTRODUCTION... 1 1.1 Overview... 1 1.2 Compatibility... 1 1.3 Application Restrictions... 2 2. SPECIFICATIONS... 3 2.1 Catalog Suffix Designations... 4 3. DIMENSIONS... 5 4. GENERAL ASSEMBLY INSTRUCTIONS... 7 4.1 Selecting Materials of Construction... 7 4.2 Sealing Pipe Joints... 7 4.3 Durafet III Electrode Preparation... 7 4.4 Cap Adapter Option for Durafet III Electrodes... 7 5. PRESSURE TEST... 8 6. ELECTRODE CABLE CONNECTIONS... 9 6.1 Electrode Cables... 9 6.2 Cable Connection... 9 7. CONNECTING A CAP ADAPTER TO AN INSTRUMENT... 10 7.1 Cap Adapter... 10 7.2 Connection to a 9782 ph Analyzer... 10 7.3 Connection to an APT2000 ph Transmitter... 11 7.4 Connection to an APT4000 ph Analyzer... 12 8. IMMERSION MOUNTING... 13 8.1 General Information... 13 8.2 Direct Electrode-to-Instrument Applications (Cap Adapter)... 15 9. IN-LINE MOUNTING... 17 9.1 General Information... 17 3/04 Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes v

10. MAINTENANCE, STANDARDIZATION, AND REPLACEMENT PARTS... 19 10.1 Maintenance... 19 10.2 Standardization... 19 10.3 Accessories, Replacement Parts, Cables... 20 vi Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes 3/04

Tables Table 5-1 Instructions for Pressure Test 8 Figures Figure 1-1 Temperature vs. Alkali Resistance 2 Figure 3-1 Durafet III (Vario Pin Connector) ph Electrode for In-Line Mountings (Smooth Tip) 5 Figure 3-2 Durafet III (Vario Pin Connector) ph Electrode for In-Line Mountings (Smooth Tip), with cable 5 Figure 3-2 Durafet III (Vario Pin Connector) ph Electrode for Immersion Mountings (Guarded Tip) 6 Figure 3-2 Durafet III (Vario Pin Connector) ph Electrode for Immersion Mountings (Guarded Tip), with cable 6 Table 5-1 Instructions for Pressure Test 8 Figure 7-1 Cap Adapter Cable 10 Figure 7-2 Connecting to an APT2000 ph Transmitter 11 Figure 7-3 Connecting to an APT4000 ph Analyzer 12 Figure 8-1 Suggested Support Arrangements 14 Figure 8-2 Direct Electrode-to-Instrument Connection 16 Figure 9-1 Proper Mounting Angle for Electrode 18 Figure 10-1 Durafet III parts 21 3/04 Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes vii

viii Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes 3/04

Introduction 1. Introduction 1.1 Overview The 7777DVP Series of electrode mountings enables a user to interface a Durafet III ph electrode to a process by either submersing the electrode or mounting it into a ¾ threaded connection. Honeywell's Durafet III Series of industrial electrodes have been designed to provide accurate and stable ph measurements for a wide variety of industrial applications. The Durafet III ph electrode is a non-glass ph electrode. The ph measurement is based in ISFET (Ion Selective Field Effect Transistor) technology, which results in a solid state ph sensing electrode. The Durafet ph electrode is much more rugged than the traditional glass ph electrode. The sturdy ph electrode reduces replacement and inventory costs. The solid state sensing element is packaged in a durable Ryton body that results in a ph electrode that has extended life in a wide variety of process conditions. The ISFET technology also produces an electrode that is up to 10 times faster than glass electrodes. This fast response improves product quality and provides better process control to optimize chemical usage. 1.2 Compatibility Durafet III ph electrodes for the 7777DVP Series are available in a number of options to meet customer application and installation needs. The electrode body is 1 in diameter with ¾ threads at both ends of the electrode. For submersion applications conduit is threaded on to the top of the electrode so that it can be lowered into the process. For in-line applications the electrode is threaded into one leg of a pipe tee. A temperature sensor is mounted internal to the electrode to measure process temperature and provide a temperature signal for automatic (Nernstian) temperature compensation. Two temperature sensors are available: 8550 ohm thermistor and 1000 ohm RTD. The Durafet III ph electrode is compatible with various ph instrumentation: Honeywell 7082 and 9782 Series ph Analyzers with Cap Adapter. APT2000/4000 Series with Cap Adapter. Non-Honeywell instruments with Cap Adapter. Consult your Honeywell sales representative for a list of instruments with this compatibility. 3/04 Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes 1

Introduction 1.3 Application Restrictions Avoid using the Durafet III series ph electrode with these chemicals and applications: Hydroflouric acid High purity water (<10 µs/cm) Hot caustic (see Figure 1-1 below) Durafet ph Sensor Temperature/Alkali Resistance Diminished electrode life may be observed in this area as it is for glass. Temperature, o C 100 80 60 40 20 Long electrode life will be experienced in this area. 0 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 0.04% 0.4% 4% ph sodium hydroxide concentration Figure 1-1 Temperature vs. Alkali Resistance 2 Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes 3/04

Specifications 2. Specifications ¾ NPT Specifications Operating Range 0-14 ph Pressure Rating 345 kpa (50 psig) @ 100 C, 690 kpa (100 psig) @ 50 C Temperature Rating Temperature Sensor Wetted Materials Electrical Connection Acceptable Cables and Lengths (ordered separately) Weight -10 to 110 C 8550 ohm Thermistor and 1000 ohm RTD Ryton body, silicon ISFET die, Ceramic reference junction, Viton media seal, EPM reference frit seals Durafet III: Vario Pin, 11 conductor connector, IP68 rated DirectLine Remote Cable: 20 feet and 50 feet Cap Adapter (for 9782P and APT Series): 20 feet and 50 feet Durafet III: 0. 23 kg (0.5 lb) 3/04 Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes 3

Specifications 2.1 Catalog Suffix Designations Instructions Select the desired Key Number. The arrow to the right marks the selection available. Make one selection each per table using the column below the proper arrow. A dot ( ) denotes unrestricted availability. Key Number I II III - - - KEY NUMBER Description 7777DVP Durafet III Electrode w/vario Pin Connector In-line or Immersion Mtg. Selection 07777DVP Availability TABLE I - Mounting Type, Temperature Sensor, Part No. (Durafet II ph Electrode) Mounting Temperature Sensor Electrode P/N Immersion 8550 Ohm Thermistor 51453503-005 01 (w/tip Guard) 1000 Ohm RTD 51453503-006 02 In-Line 8550 Ohm Thermistor 51453503-001 03 (w/o Tip Guard) 1000 Ohm RTD 51453503-002 04 TABLE II - Instrumentation to Sensor Cables Cable length Cable Type Part Number None 20 ft (6,10 m) Cap Adapter 51453388-001 50 ft (15,24 m) Cap Adapter 51453388-002 07777DVP Selection 00 04 05 TABLE III - Options Tagging None Stainless Steel Customer I.D. Tag - 3 lines, 22 characters/line Certifications None Calibration & Conformance 00 SS 00 CC 4 Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes 3/04

Dimensions 3. Dimensions mm inches Figure 3-1 Durafet III (Vario Pin Connector) ph Electrode for In-Line Mountings (Smooth Tip) Figure 3-2 Durafet III (Vario Pin Connector) ph Electrode for In-Line Mountings (Smooth Tip), with cable 3/04 Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes 5

Dimensions Figure 3-3 Durafet III (Vario Pin Connector) ph Electrode for Immersion Mountings (Guarded Tip) Figure 3-4 Durafet III (Vario Pin Connector) ph Electrode for Immersion Mountings (Guarded Tip), with cable 6 Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes 3/04

General Assembly Instructions 4. General Assembly Instructions 4.1 Selecting Materials of Construction Electrode The materials used in the construction of the DurafetIII ph electrode are listed in the specifications. Materials of wetted parts must be compatible with the process temperature and corrosion conditions. User-supplied components Pipe couplings, and tees must be supplied by the user. Select materials that are compatible with the process temperature and corrosion conditions. 4.2 Sealing Pipe Joints When making pipe joints apply Teflon tape pipe sealant to male threads. Wrap the threads with the tape overlapping by 50% on each wrap. Start the wrap at the end of the pipe and wrap in the direction of the thread at least two turns. When installing an electrode, apply Teflon tape to the threads, then hand-tighten the electrode in the fitting only until snug. If necessary to stop a leak, a wrench may be applied to the electrode s wrenching flats to further tighten the joint only until the leak has stopped. Always leave at least one thread on the electrode showing outside the fitting. CAUTION When tightening an electrode never exceed 15 ft-lb applied torque. Applying too much torque can result in damage to the electrode. 4.3 Durafet III Electrode Preparation Prepare Durafet III electrode as described in the electrode manual, part number 70-82-25-115. 4.4 Cap Adapter Option for Durafet III Electrodes The Cap Adapter option for Durafet III electrodes eliminates the need to separately mount a preamplifier. The Cap Adapter is a preamplifier electronics module that is integral with the electrode-to-instrument cable. The preamplifier PCA is molded into the cap of the quick-disconnect connector. The cable has tinned leads that are ready to be wired directly to 9782/7082 Analyzer or APT2000/4000 Transmitters and Analyzers. The Cap Adapter mounts directly on the Durafet III electrode. The electrode has a key that mates with the keyway in the connector on the cable. The cap is then threaded onto the electrode. The cap should be handtightened and engage the O-ring at the top of the electrode. The knurled fitting at the top of the cap should be hand-tightened to seal an internal O-ring around the cable. When properly threaded onto the electrode the connection is waterproof. 3/04 Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes 7

Pressure Test 5. Pressure Test When to perform the test Before submerging an assembly do a low-pressure test to check the various seals made during the assembly operation. How to perform the test Table 5-1 lists the steps for pressure testing the assembly. Table 5-1 Instructions for Pressure Test Step 1 Double back the cable into the immersion pipe. Action 2 Connect a source of low-pressure air to the immersion pipe. [Approximately 69 kpa (10 psi) will simulate immersion in 6.10 m (20 ft) of water.] 3 Immerse the assembly in a shallow tank of water and look for any indication of air bubbles streaming from the assembly. 4 To seal leaking connections: Use extra Teflon tape on pipe joints. Use silicone grease on O-ring seals. Make the connecting rings on the preamplifier snug. 8 Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes 3/04

Electrode Cable Connections 6. Electrode Cable Connections 6.1 Electrode Cables The Durafet III ph electrode is supplied with a connector at the electrode top to provide a quick disconnect for easy electrode replacement or with an integral cable. CAUTION ESD sensitive devices inside Durafet III electrodes and Cap Adapter Cables. Cable connectors The electrode cable terminates with a female Vario Pin connector on one end and tinned leads at the instrument end. Avoid contamination The connectors are constructed with high resistance insulating material which can be contaminated if exposed to oil and salts from bare hands. Avoid contaminating the internal area of the connectors. CAUTION Do not allow liquids or other foreign matter to contact the cable connectors. Save the protective cap supplied with electrodes with connectors and put the cap on the connector whenever the cable is not installed on an electrode. How to clean the connectors The connectors can be cleaned by wiping with a cotton swab moistened with isopropyl or grain alcohol. Allow connectors to dry completely before making connections. 6.2 Cable Connection Make sure electrode connector and cable connector are clean and dry. Align key way on VarioPin connector of electrode with tab inside mating connector on cable. Press cable connector onto electrode firmly. Tighten knurled bushing of cable connector by hand to ensure waterproof seal. 3/04 Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes 9

Connecting a Cap Adapter to an Instrument 7. Connecting a Cap Adapter to an Instrument 7.1 Cap Adapter The Cap Adapter is an integral part of the electrode cable. It is essentially a preamplifier that does not require separate mounting. The output from the Cap Adapter can be connected directly to a ph instrument. The Cap Adapter and cable for Durafet III electrodes are available in lengths of 20' and 50'. One end of the Cap Adapter cable is the mating connector to the Vario Pin of the Durafet III electrode. The other end of the cable terminates with tinned leads. The tinned leads connect to the input terminals of the ph instrument as described below. Figure 7-1 Cap Adapter Cable 7.2 Connection to a 9782 ph Analyzer The Cap Adapter is connected to a 9782P-01 analyzer. (See Instruction Manual #70-82-25-73 9782 ph/orp Multifunction Analyzer/Controller, for a more detailed description of the 9782.) Connect the tinned leads of the Cap Adapter cable as follows: Wire Color Orange 9782 Terminal PA Blue V+ Green V White Black Red Yellow White/Black Red/Black SC SC TH Ground Screw Not Used TH 10 Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes 3/04

Connecting a Cap Adapter to an Instrument 7.3 Connection to an APT2000 ph Transmitter The Cap Adapter can be connected to the APT2000 ph Transmitter. (See the APT2000 ph Transmitter Manual, part number 70-82-25-92, for more detailed description of the APT2000.) +3 V/0.5 ma GND -3 V/0.4 ma Do not connect + 4 to 20 ma _ T1 T2 N.C. Aux. el. Reference el. Meas. el. 19 18 17 16 15 14 8 7 6 5 4 2 1 BLUE BLACK GREEN WHITE RED/BK YELLOW WT/BK NOTE: Orange and Red wires are not typically connected. These should be clipped and electrically sealed to avoid possible contact with other conductors. Figure 7-2 Connecting to an APT2000 ph Transmitter 3/04 Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes 11

Connecting a Cap Adapter to an Instrument 7.4 Connection to an APT4000 ph Analyzer The Cap Adapter can be connected to the APT4000 ph Analyzer. (See the APT4000 ph Analyzer Manual, part number 70-82-25-103, for more detailed description of the APT4000.) Figure 7-3 Connecting to an APT4000 ph Analyzer 12 Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes 3/04

Immersion Mounting 8. Immersion Mounting 8.1 General Information Intended Use The system can be used in a variety of configurations to accommodate many techniques for support, immersion, and removal of the electrode in a process solution. A variety of mounting configurations are used according to the process application. By using accessory parts such as pipe, pipe fittings and cable grips, an immersion assembly can be built to suit a specific application. Minimum immersion depth In all cases, the electrode body should be immersed a minimum of 5.08 cm (2 in.) into the process liquid to ensure proper temperature compensation. Suggested support arrangements Figure 8-1 illustrates possible support arrangements for immersion mounting of the Durafet III electrode. 3/04 Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes 13

Immersion Mounting Figure 8-1 Suggested Support Arrangements 14 Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes 3/04

Immersion Mounting 8.2 Direct Electrode-to-Instrument Applications (Cap Adapter) Application Figure 8-2 illustrates this configuration. Materials required The material supplied with this catalog number is listed below. Quantity Item 1 Durafet III Electrode with slotted tip and cable Materials supplied by the user are listed below. Quantity Item 1 Length of 3/4 in. Sch. 40 pipe, threaded on both ends. Pipe length to be determined by user. When planning pipe length, allow enough cable between the pipe and the instrument to permit removal for servicing. 2 3/4 in. NPT metal Sch. 40 pipe couplings 1 3/4 in. NPT cable grip for 1/8 in. diameter cable Assembly Assemble the materials as shown in Figure 8-2. CAUTION Do not push excess electrode cable into the immersion pipe or pipe coupling. Dimensions For mounting dimensions, see Section 2. Cable electrode to instrument Cable electrode to instrument as described in Section 6. Pressure test Perform a submersible pressure test (see Section 5). Mounting Mount the assembly. See Figure 8-1 for mounting suggestions. 3/04 Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes 15

Immersion Mounting Arrangement of components Figure 8-2 illustrates the configuration of components used for direct connection of the electrode to the instrument without an external preamplifier module or external preamp adapter. Figure 8-2 Direct Electrode-to-Instrument Connection 16 Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes 3/04

In-Line Mounting 9. In-Line Mounting 9.1 General Information Choice of pipe tee This mounting allows the electrode body to be inserted directly into a Schedule 40, 3/4 in. NPT metal pipe tee in a pipe line. The electrode will NOT fit into standard PLASTIC pipe tees. Use a metal tee or the special Honeywell schedule 80 CPVC 3/4 in. pipe tee, Honeywell part number 31120167. Avoiding damage to the electrode The Durafet III electrode for in-line mounting has an exposed sensor. Exercise care when inserting or removing the electrode from the pipe tee to prevent damage to the sensor surface. Ensuring accurate temperature sensing When process temperature varies considerably from ambient temperature, insulate the entire Durafet III electrode body to ensure accurate process temperature sensing. Orientation of electrode For reliable measurement, the electrode must be immersed in the process fluid. Therefore, orientation of the electrode vertical to the horizon is not recommended, as this orientation may prevent sufficient depth penetration to reliably immerse the sensor. The electrode mounting angle should be at least 45 degrees from vertical as shown in Figure 9-1. If solids are present in the process fluid, avoid angles exceeding 90 degrees from vertical to minimize accumulation of solids around the sensor. 3/04 Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes 17

In-Line Mounting Figure 9-1 Proper Mounting Angle for Electrode 18 Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes 3/04

Maintenance, Standardization, and Replacement Parts 10. Maintenance, Standardization, and Replacement Parts 10.1 Maintenance Keeping electrode moist When an electrode is removed from the process for any reason, assure that it does not become dry and remain dry for more than a short period of time. The electrode may require more frequent maintenance if used in a batch treatment installation which leaves the electrode dry between batches, or if it is exposed to process fluids that leave a deposit on the surface of the sensor chip (measuring electrode) and reference electrode. The manual supplied with the electrode contains instructions for cleaning the electrode. Thoroughly rinse the electrode with water after any type of cleaning. In addition to periodic cleaning, other electrode maintenance includes electrode performance checks, and treatment for a clogged junction or severely dry electrode. CAUTION Read the manual (part number 70-82-25-115) supplied with the electrode before attempting any maintenance procedures. Improper handling of the electrode can result in damage that will affect accuracy. 10.2 Standardization In addition to electrode maintenance, successful ph measurement relies on periodic standardizing of the measurement instrument and its electrode system. (This is required because all electrodes do not produce exactly the same potential in a solution of known ph. A periodic corrective adjustment eliminates any deviation from the standard value.) Establish regular intervals for standardizing according to conditions and experience. Procedures for standardizing are given in the measuring instrument instructions. 3/04 Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes 19

Maintenance, Standardization, and Replacement Parts 10.3 Accessories, Replacement Parts, Cables Description Part Number Accessories NIST Buffer 4.01 ph 31103001 NIST Buffer 8.86 ph 31103002 NIST Buffer 9.18 ph 31103003 Reference Gel refill kit 51205807-002 Replacement Parts Guarded Electrode Tip (immersion) see Figure 10-1, item 1 51204992-001 Smooth Electrode Tip (In-line) see Figure 10-1, item 2 51204993-001 Frit O-Ring (small) see Figure 10-1, item 3 51198302-001 Frit O-Ring (large) see Figure 10-1, item 4 31074364 ESD Protective cap see Figure 10-1, item 5 51500474-003 Storage Cap see Figure 10-1, item 6 51453164-001 Cables 20 Cap Adapter cable 51453388-001 50 Cap Adapter cable 51453388-002 20 Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes 3/04

Maintenance, Standardization, and Replacement Parts Figure 10-1 Durafet III parts 3/04 Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes 21

22 Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes 3/04

Industrial Measurement and Control Honeywell 1100 Virginia Drive Fort Washington, PA 19034 70-82-25-116 Rev. 1 0304 Printed in USA www.honeywell.com/imc