Handbook Of Microbiological Safety Cabinet Type OPTIMALE

Similar documents
Handbook Of Microbiological Safety Cabinet Type BIO II 9 (4 ft) and 12 (3 ft)

TECHINCAL DESCRIPTION BIO BAN

Envair Lab SCS Evo Comfort+ Class II Microbiological Safety Cabinets

Cytotoxic Safety Cabinets ENVAIR eco safe Plus

BSC-1100IIA2-X is designed to meet the operating principles:

SafeFAST Classic. Class II Microbiological Safety Cabinets PROTECTION, SAFETY, RELIABILITY. AND MORE.

SafeFAST Classic. Class II Microbiological Safety Cabinets PROTECTION, SAFETY, RELIABILITY. AND MORE.

Laboratory Equipment MICROBIOLOGICAL SAFETY CABINETS ENVAIR ECO SAFE COMFORT PLUS CLASS II MICROBIOLOGICAL SAFETY CABINET.

Laboratory Equipment MICROBIOLOGICAL SAFETY CABINETS ENVAIR ECO SAFE COMFORT PLUS CLASS II MICROBIOLOGICAL SAFETY CABINET

Operating Instructions and Maintenance Weighting Booth Type PS/PSX EQL

TM Series MICROBIOLOGICAL SAFETY CABINETS. Technical Specifications

Aeolus V. Green Line. Product protection. Vertical Laminar Flow Cabinets. High quality. Ergonomic design. Low noise level.

SafeFAST Top. Class II Microbiological Safety Cabinets PROTECTION, SAFETY, RELIABILITY. AND MORE.

SafeFAST Elite. Class II Microbiological Safety Cabinets PROTECTION, SAFETY, RELIABILITY. AND MORE.

CLASS II BIOLOGICAL SAFETY CABINET

Bio II Advance. Class II Biological Safety Cabinet

Techlab - Matériaux de Référence certifiés et Equipements de Laboratoire -

Angelantoni Life Science. Microbiological safety cabinets Class II. SterilSafe Series

A Thermolab Group Company. HyZone. Class II Microbiological Safety Cabinets. for operator, environment and product protection.

NOTE: The color of the actual product may differ from the color pictured in this catalog due to printing limitation.

Bio II Advance. Class II Biological Safety Cabinet

(Class II Type A2 & B2)

Biological Safety Cabinet

FlowFAST H. Horizontal Laminar Flow Cabinets PROTECTION, SAFETY, RELIABILITY. AND MORE.

ASALAIR BIOHAZARD ATLANTIC CYTOCABINET MODEL: ATLANTIC CYTOCABINET CODE ATLANTIC CYTOCABINET CODE

AS , EN12469 (Microbiological Testing Method)

SafeFAST Top. Class II Microbiological Safety Cabinets PROTECTION, SAFETY, RELIABILITY. AND MORE.

ASALAIR BIOHAZARD ATLANTIC CABINET MICROBIOLOGICAL SAFETY CABINET CLASS II TYPE A2 MODEL:

(Class II Type A2 & B2)

High protection against allergens and contaminants Spacious working area High cleanability and visibility Microprocessor controls and safety alarms

ASALAIR BIOHAZARD ATLANTIC CABINET MICROBIOLOGICAL SAFETY CABINET CLASS II TYPE A2

CLASS II BIOSAFETY CABINET

A. Each cabinet must be constructed and installed in accordance with EN12469:2000.

PCR 6. Operators Manual. Vertical Laminar Airflow Cabinet with U.V. Sterilisation. Issue 05 September 2013

PRODUCT CATALOG CLASS II BIOLOGICAL SAFETY CABINET.

Isolator technology. eco safe Iso Compact. Width (m) EN Isolator. Class. Standard 0.9 / 1.2 II EN

TITAN, ATLAS & SIRIUS SMALL, PORTABLE. economical cabinets. with low energy consumption

Envirco BIOLOGICAL SAFETY CABINET

CLAIRE TOTAL SAFETY CABINET TYPE B2 Safe for chemicals due to Total Exhaust

CytoFAST Elite. Cytotoxic Safety Cabinets PROTECTION, SAFETY, RELIABILITY. AND MORE.

SafeFAST Elite. Class II Microbiological Safety Cabinets PROTECTION, SAFETY, RELIABILITY. AND MORE.

NU-430/435 CLASS II, TYPE B2 BIOLOGICAL SAFETY CABINETS

PCR Laminar Flow Cabinets

CytoFAST Elite. Cytotoxic Safety Cabinets PROTECTION, SAFETY, RELIABILITY. AND MORE.

VCM LAMINAR FLOW HOOD VERTICAL

ORION. Safety Class 2 100% exhaust. ALMOST Rocket Science

VERTICAL LAMINAR FLOW HOOD FLV

SafeFAST Premium. Class II Microbiological Safety Cabinets THE LOWEST POWER CONSUMPTION THE LOWEST CO 2 EMISSIONS THE LOWEST NOISE LEVEL

Thermo Scientific 1300 Series A2. Class II, Type A2 Biological Safety Cabinets. Good for you good for the environment

Logic+ Technical Manual and Specifications. Type B2 Biosafety Cabinet Models

topsafe Safety you can afford Microbiological Safety Cabinet Class II Type A2 CERTIFIED EUROPEAN SAFETY AT AFFORDABLE PRICES

Best Products. Best Performance. Best Protection.

Class II Biological Safety Cabinets The World's Most Energy Efficient, Quiet, and Compact Biosafety Cabinet

EuroFlow Series. Operator, product and environmental protection. Class II Biological Safety Cabinet. Superior quality. Proven reliability

GloveFAST Cyto Pharma GloveFAST Cyto Aseptic Isolators PROTECTION, SAFETY, RELIABILITY AND MORE.

Laminar Flow Clean Benches, Horizontal and Vertical

Series 301. Horizontal Laminar Flow Clean Bench. C l e a n B e n c h e s. s o l u t i o n s. Series 301 Horizontal Laminar Flow Clean Bench

HORIZONTAL LAMINAR FLOW HOOD FLO

ASALAIR BIOHAZARD CABINET ATLANTIC 1200 CLASS II TYPE B2 COD

VERTICAL LAMINAR FLOW CABINETS

GloveFAST Cyto/Aseptic. Isolators PROTECTION, SAFETY, RELIABILITY. AND MORE.

Purifier Logic Class II Biosafety Cabinets FEATURES & BENEFITS

FILTRATION FUME CABINET MODELS T800 T1800 OPERATING AND MAINTENANCE MANUAL

Series. Ductless Fume Hoods The Safe, Energy Efficient Solution for Modern Chemistry

TOP QUALITY SOLUTIONS FOR YOUR LABORATORY

Class I Microbiological Safety Cabinets The Leading Energy Efficient, Quiet, and Ergonomic Biosafety Cabinet

BIOSAFETY CABINETS. SURFACE WIDTH: 120cm - 48

VERTICAL LAMINAR FLOW HOOD MODEL 700 PARTICULARLY SUITABLE FOR PCR PROCESS CODE

Appendix C: Biosafety Cabinet Specifications C-1. Product Service

SPECIAL REVERSE HORIZONTAL FLOW ENCLOSURE OPERATING AND MAINTENANCE MANUAL. Issue No: 1

MODEL 950C-NP-UV Installation and Owners Manual

Summary and Features. Model. Main Function Feather:

ADVANCED SAFETY ENGINEERING FOR EFFECTIVE HANDLING OF HAZARDOUS PARTICULATE DRUGS AND POWDERS IN A CLASS II, TYPE A2 BIOLOGICAL SAFETY CABINET

Clean Air Techniek B.V. User manual English Series MSC III

Smart design has its rewards. NEW! Thermo Scientific MSC-Advantage Class II Biological Safety Cabinet. Part of Thermo Fisher Scientific

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

FlowAir CATALOG.

ASALAIR FUME CUPBOARD /R WITH HEPA FILTER

SPH-400 i Series General Manual

MODEL NO.:Hovo 710 PUSH

Class II, Biological Safety Cabinets Probably the Most Advanced Energy-Efficient, Safe, and Ergonomic Biosafety Cabinet in the World

Fume Hoods. Selection Guide. Fume Hoods. Accessories & Options. Clean Benches. Ductless Fume Hoods. Description Filter Work Surface Page.

On Line UPS. LUC 1000E / LUC 2000E / LUC 3000E User Manual

SERVICE MANUAL AIR PURIFIER SP-300A

Appendix C: Biosafety Cabinet Specifications C-2. Product Service

TXH TOTAL EXHAUST HOOD. Installation and Operations Manual

Laminar Flow Clean Benches, Horizontal and Vertical The Leading Solution for Research Laboratories

Contamination Control Laboratories (CCL) Pty Ltd

Compounding Safety Solutions.

CG INLET CEILING. Inlet ceiling with the laminar discharge flow for operating rooms

Class II biological safety cabinets

Operating Manual PFB-2500

Inlet Controller TC5-ITA USER'S MANUAL. M rev. 02 K rev. 00

External Dimensions (mm) Internal Dimension (mm) Material. Side Panel Pre-filter Main Filter Air Speed Noise Glass Door

Biological Safety Cabinet. Feature & HEPA-filter efficiency

Biological Safety Cabinet Class II A2

MPPT Solar Integrated Inverters (500VA~2500VA) User Manual THE SOLAR TECH CO.,LTD. Solar Integrated Inverter Series 500VA~2500VA 13.

Laminar Flow Clean Benches, Horizontal and Vertical The Leading Solution for Research Laboratories

KENYA MARINE AND FISHERIES RESEARCH INSTITUTE

BATTERY OPERATED DIGITAL MILLI-OHMMETER

Transcription:

Handbook Of Microbiological Safety Cabinet Type ADS LAMINAIRE S.A. : 35-37 Rue Baudin 93315 LE PRE SAINT GERVAIS Cedex France Tel: +33 (0) 1 48 44 74 69 - Fax: +33 (0) 1 48 44 14 64 - E Mail : info@adslaminaire.com - Web : http://www.adslaminaire.com Capital 80 000 - SIRET 69204043900022 APE 2825 Z T.V.A. FR 65692 040 439

Type Dear Sirs, You have just acquired an and we congratulate you. This cabinet guarantees you: - To work in sterile conditions. - The protection of the operator. - The protection of the surrounding environment from all risks of contamination. Your cabinet is recommended for the manipulations of pathogenic germs (levels 2 and 3). The model (12) is built according to the European Norm EN 12469. It is controlled and certified by LNE with the NF mark: FRENCH STANDARD INSTITUTION (AFNOR) ELECTED ORGANISM Tour EUROPE - Cedex 7 Laboratoire National d Essai (LNE) 92 080 PARIS LA DEFENSE Service Certification Tél: 00 33 / 1 42 91 55 55 1, rue Gaston Boissier Minitel: 3616 Marque NF 75 724 PARIS cedex 15 We wish you a good utilisation and we are staying at your complete disposal for further details. Eric FITOUSSI Bernard BIJAOUI Managing Director - 2 -

Table of Contents I CONSTRUCTION... 4 II OPERATING DESCRIPTION SECURITY... 6 III FRONT PANEL... 8 IV USING... 10 V USING THE... 15 VI RECOMMANDATION AND ADVISES... 16 VII TECHNICAL DATA... 17 VIII ELECTRICAL SKETCH... 18 IX MAINTENANCE... 19 X GUARANTEE AGREMENT... 20-3 -

I CONSTRUCTION 1 FRAME The external frame is manufactured with epoxy painted steel. The internal frame is completely in 304 L stainless steel smooth finish and thickness of 15/10 This unit is very strong and insensitive to the deformations and alterations. 2 WORK SURFACE Removable, it is made of one part white TRESPA panel. 3 - AIR FLOW BARRIER This barrier is created through a Perforated Grid in stainless steel 304L. This avoids cross contamination. 4 - RECEPTACLE Placed under the work surface, its structure in Stainless steel 304L permits to contain liquids. It can be cleaned by easily removing. 5 - LIGHTING A fluorescent lighting is fixed outside the work volume on the front panel. The level is 600-800 LUX on the work surface. 6 - SASH WINDOW A front sash window panel (Triplex Glass) acts as a protective shield. The standard working position is 220 mm high above the work surface. Three other positions of the window panel are possible: - Standard Position for normal use - Total Opening position (orange colour button) - Total Closing till the bottom of the work surface. - 4 -

7 - VENTILATION This cabinet is equipped with an adjustable and direct centrifugal fan designed to maintain the air flow velocity at low rotation speed (ECM technology: principle hereunder) In order to avoid any vibration, engine and turbine are balanced together and the blower is mounted on rubber vibration isolators. 8 - FILTRATION HEPA Filters are installed with an efficiency of 99.999% for particles greater than 0.3µ (classification H14 DOP tested). Those filters are designed for laminar air flow blowing (ISO 5, Class 100, US Norm) and exhaust. Media: Frame: Gasket: Glass Fiber Aluminium frame and two white epoxy grids (both sides of the filter). Neoprene Gasket. 9 - AUTOMATIC AIR FLOW REGULATION Air Flow Velocity is maintained constant automatically at 0,4 m/s +/-10% whatever the clogging of the filters. Use of ECM Fans allows an automatic flow rate regulation due to the microprocessor integrated inside the fan. - Fan for S 9 and 12: ECM DP 9.7 - Fan for S 16 and 18: ECM DP 9.9 10- Electrical Plug On the back wall and right side, you benefit from 1 European electrical plug without pin. - 5 -

II OPERATING DESCRIPTION SECURITY 1 OPERATING DESCRIPTION The vertical laminar air flow ISO 5 (class 100) is issued from the first absolute filter above the work surface, sweeping all the work volume in order to keep the handling out of any contamination. The 30% of air flow rate exhausted through the second HEPA filter (located on the top of the booth) is compensated by new air sucked through a grid in the forepart of the work surface. This is creating a real air flow protection for the user. Rejected air flow Absolut filters (H14) Vertical laminar air flow (ISO5) 2 SECURITY The use of the Optimale is possible only when the key position is ON : in position arrêt with window closed, the cabinet is OFF. - 6 -

When the Optimale is in function (key in position ON), you benefit from the following safe principle: In order to protect the operator, the regulation system is designed to maintain the air flow constant whatever the clogging of the filter. A spot light and a buzzer alarm inform that you have to remove the HEPA Filters in a delay of 3 to 4 weeks, when it is clogged. Other alarm prevents when sash window position is higher or lower than the recommendation for normal use. Nevertheless you can stop the alarm if you need a longer delay with the sash window completely open (use the switch board behind the front panel for filters access: change position n 2 on the switch. 3 - PACKAGING AND TRANSPORT In case of long time without using you must package the cabinet to avoid climatically variations in respecting the conditions stipulated in the part IV of this present Handbook. If the has to be transported, it has to undergo no shock. - 7 -

III FRONT PANEL A. PRESENTATION Up & Down window Menu selection Menu Access ENTER ON/OFF EXIT Spare Contact Display the downflow Velocity Light Display the inflow velocity Display the filter pressure Standby UV Display the messages as alarm messages Display the filter lifetime - 8 -

B. BUTTONS On/Off : Switch on and switch off the cabinet, which means the fan and all functions. Allows to exit from a menu or a sub menu or to cancel a function. Sleep : Switch on and switch off the Sleep mode. : Up & down the sash windows. Use it to navigate into the menu or to select a choice. Light : Switch on and switch off the main light UV : Switch on and switch off the UV Lamp. Only available during the Sleep Mode. Spare contact : Switch on and switch off a contact in the main board ready to connect any accessory (solenoid, electrical socket, ) - 9 -

IV USING A. STARTING Push ON/OFF Button to turn ON. If no password has been recorded, the cabinet starts directly. If a password has been recorded (Factory settings) you need to enter it. The Factory settings password is 0001. You can change it in the User Menu. The panel displays Wait and the sash window stays closed until the velocity is OK. Then, the window raises itself to the correct height. - 10 -

Push ON/OFF Button to turn OFF the cabinet. If no password has been recorded, the cabinet will stop directly. If a password has been recorded (Factory settings) you need to enter it. The Factory settings password is 0001. You can change it in the User Menu. Then, the window closes automatically (an alert message displays) and the light switches off. B. SLEEP MODE Push Sleep Button during 2 seconds to switch the cabinet in Sleep mode. When the moon is highlighted, it s OK. If no password has been recorded (Factory settings), the cabinet switch in Sleep mode directly. If a password has been recorded you need to enter it before. Then, the window closes automatically (an alert message displays) and the light switches off. It is possible to switch ON the light if necessary. The screen display that you need to push Sleep button 2s more to come back to the normal mode. If a password has been recorded you need to enter it. After few seconds, the flower icon appears in standard mode and the airflow reduces itself at the value recorded in the Program menu. - 11 -

When the cabinet comes back to the normal use, the sash window stays closed until the velocity is OK. Then, the window raises itself to the correct height. C. ACCESSORIES Window : Push to open and close the sash window. If the button is released the window stop moving. When opening, the windows stops automatically at the standard height. Maintenance : To open the window at the maximum height, still pressing the during 2 s. Then it starts opening until the button is released. If a password as been recorded in the Window Lock menu, a password will be asked to permit the opening, Light : press the light button, to switch on or switch off the light. The icon display little rays when the light is ON. - 12 -

UV : Press the Light button, to switch on or switch off the UV Lamp. UV can be activated only in the Sleep Mode. If the cabinet is in Normal Mode, this message appears : Sleep mode for UV The icon display little rays when the light is ON. Spare Contact: Press the Contact button, to switch on or switch off the spare contact. D. ALARMS Clogging Alarm : When the filter is clogged, the alarm beeps and this message appears Filter Clogged Incorrect Flow Alarm : When the velocity is out of the usual range, the alarm beeps and this message appears speed too low or speed to high - 13 -

Window s Alarm : When the window is not in the right position, the alarm beeps and this message appears window too low or window to high Fan Alarm : When the fan is out of service, the alarm beeps and this message appears Fan Default Maintenance Alarm : After calibration due date, this message is displayed: Call After Sales Service E. Bugs If the cabinet presents any bug, you can try to reinitialize it to fix the problem. To reinitialize the cabinet unplug then plug it again to the power jack. Then push Up & Down buttons together during 3 second while the flower appears. - 14 -

V USING THE CONDITIONS : The conditions of the environment hereunder have to be respected for a good operation of your MSC (Microbiological Safety Cabinet): Ambient temperature: + 5 C to + 40 C. Humidity: 30% to 95%. Agents of pollution: Do not install the MSC in a place where the quantity of pollutants is abnormal. Radiation: The can not be exposed to important radiations. OPERATING PROCEDURE OF THE CABINET: Placed on its support legs or on an existing work bench (pallet or stable horizontal table). Connect the Booth on current mono 220 V+T -15 Press the touch M.A (1), fan starts to blow. You shall wait 2 hours (time of purge) before the first utilisation. Sash Window : Press on the touch RID (6), then on (5) to go up or (7) to go down.. To stop the sash window in intermediate position, press on RID (6). The engine will stop automatically at the right high fixed in factory. Light : press on ECL (3). Press a second time to switch off UV Light : To switch on the Ultraviolet Lighting, press on U.V (10), only when the sash window is on the bottom position. Standby mode : when the sash window is maintained on bottom position ((7)) the cabinet is automatically on Standby Mode with lower velocity. To return to the standard mode, push RID (6) button. Alarm : If COL (8) and ALM (9) get in function, you have to remove imperatively the absolute filters which are clogged. For all starter or new starter after a prolonged stop, wait 2 to 3 hours of purge before utilisation. - 15 -

VI RECOMMANDATION AND ADVISES Following the Norms Recommendations You should manipulate only when the sash window is in normal position of use (automatically stopped at the right high when you open the window) and at a distance of 13 cm on the work surface from the front grids (or window). After Installation and the first start of the cabinet, you can use and work easily after each interruption or standby mode as described hereunder: - Standby mode : when the sash window is maintained on bottom position ((7)) the cabinet is automatically on Standby Mode with lower velocity. To return to the standard mode, push RID (6) button. - During the night or week- end the position of the window is closed; thus the cabinet is automatically in Standby Mode (half normal Velocity). The spot lights for V and M/A are still ON. This is the normal and recommended position for standby: you needn t to switch off completely the cabinet, just let it in Standby with red light V in position ON. - In case where you switch off the cabinet by pushing M/A button: the spot lights are off but an air flow follows sweeping when the window is completely closed (it is depending of the position of one Switch on the electronic board behind the Plexiglas Cover after the white front panel). IMPORTANT: You benefit from an automatic air flow re-start after a prolonged stop if you have a general electrical power cut in your building). (Sash Window: Press on the touch RID (6), then on (5) to GO UP or (7) to GO DOWN. To stop the sash window in intermediate position, press on RID (6). The engine will stop automatically at the right high fixed in factory) Light : press on ECL (3). Press a second time to switch off Alarms : If COL (8) and ALM (9) get in function simultaneously, you have to remove imperatively the absolute filters which are clogged. Others Alarm : Alarm clogging optical and acoustic (hepa filter) Alarm fan dysfunction optical and acoustic Alarm on sash window to ensure the right height An alarm permanently tests the exhaust air and the air flow barrier. The alarm of the air flow is in function while the normal air flow velocity is not reached. - 16 -

VII TECHNICAL DATA 9 12 16 18 Weight 190 Kg 230 Kg 300 Kg 350 Kg Recycling Airflow 790 m 3 /h 1050 m 3 /h 1310 m 3 /h 1580 m 3 /h Extracted Airflow 350 m 3 /h 470 m 3 /h 580 m 3 /h 700 m 3 /h Effective Length 921 mm 1226 mm 1531 mm 1836 mm Overall Length 971 mm 1276 mm 1581 mm 1886 mm Fan Type DD 9/7 ECM DP 9/7 ECM DP 9/9 ECM DP 9/9 ECM Blowing HEPA Filter M6/9L M6/12L M6/15 L M6/18 L Exhaust HEPA Filter M3/6/90 M3/7/90 M3/9/90 2 x M3/6/90 Light 900 mm / 30 W 1200 mm / 36 W 1500 mm / 58 W 1500 mm / 58 W UV Light UV 450/G15T8 UV 900 / G30 T8 UV 900 / G30 T8 UV 900 / G30 T8 Sash Window Thickness 8 mm 960 x 700 1265 x 700 1570 x 700 1875 x 700 For any : DIMENSIONS OVERALL EFFECTIVE Depth: 786 616 Height: 1400 650 Electricity: mono 220 + T - 15A - 50 Hz Noise level: < 64 db according NF control Level of lighting : superior to 750 LUX ADS Laminaire is allowed to change any reference in the. - 17 -

VIII ELECTRICAL SKETCH - 18 -

IX MAINTENANCE WARNING: Before accessing to the cabinet components, you have to cut the current of the MSC. Before changing HEPA Filters, you must proceed to a decontamination of Cabinet. As soon as the diode COL is switch ON, you have to change the two HEPA filters. Before their replacement, anticipate the sterilization of the cabinet. Thanks to a very easy access and a simple fixation system, HEPA filters are easily removable. Blowing Filter: Exhaust Filter: Fan Motor: Unscrew from inside the booth, under the filter. Filter is free. Do the reverse in maintaining the new filter on the top for screwing. The access of the filter is direct behind the front panel. Unscrew and take out HEPA filter to replace the new one. Without maintenance. Fluorescent lighting: Direct access behind front panel. IMPORTANT HEPA Filters replacement must be made by a Specialist with validation and control. After long time of no use, proceed to a control and validation before your manipulations For usual Control and Test, you benefit from 2 inlets for DOP testing: refer to the photo hereunder You can use Formaldehyde Fumes to decontaminate the cabinet - 19 -

X GUARANTEE AGREMENT The is guaranteed One Year for spare parts, except for wrong use and disposables. CONDITIONS OF THE GUARANTY: - During the period of guaranty, the customers will benefit in case of failure, from free spare parts replacement. - The guarantee is not available for: - In case of consecutive damages referring to a wrong utilisation or a default of maintenance. (non respect of orders) or an external cause (damages of water, fire, and so on C.f. the insurance of your establishment. ) - In case of external intervention, other than ADS LAMINAIRE company during the period of the guaranty. - 20 -