HI-POT/CONTINUITY TESTER Model PLEP01

Similar documents
HT-2000 Dielectric Withstand Tester Volts AC Output

HT Dielectric Withstand Tester Volts AC Output 25mA Leakage Option

MegaPulse 1.2x50-8 PV

MegaPulse. IMPULSE TESTER Implantable Test 1. Instruction Manual

Dielectric Withstand Tester

MegaPulse. Antenna Surge PF 65

GFM-200A. Instruction Manual

HT-25kVdc Dielectric Withstand Tester Volts DC Output

HT Dielectric Withstand Tester Volts AC Output Volts DC Output COMPLIANCE WESTUSA. Instruction Manual

GF-30AC. Ground Circuit Tester COMPLIANCE WESTUSA. Instruction Manual

TESTLINK COMPLIANCE WESTUSA. Remote Actuation Cable Set. Instruction Manual

GFM-40ac COMPLIANCE WESTUSA. Ground Circuit Tester. Instruction Manual

Ensuring the Safety Of Medical Electronics

AVOIDING ELECTROCUTION HAZARDS

INSPECTOR LINE LOAD SIMULATOR INSTRUCTION MANUAL TASCO, INC.

G76x Direct Spark Ignition Controls

ITS-50R TRANSFER SWITCH OWNER S MANUAL

An average of one worker is electrocuted on the job every day There are four main types of electrical injuries:

Component Test Fixture

Transfer Switch TS-50. Owner s Manual

PBA Series Prelube Controls

Electrical Safety. Electrical Safety Webinar. Electrical. Printing Industries Alliance Printing Industries Alliance 1

G600 Series Replacement Intermittent Pilot Ignition Controls

OPERATIONS AND APPLICATIONS MANUAL MODEL NUMBER PM6. 6 KV DC HIGH POTENTIAL TESTER WITH MEGOHMMETER Version 2.0

Electrical Hazards in Construction

User s Manual. Automatic Switch-Mode Battery Charger

Transfer Switch GPTS 30

Valcom Failsafe Unit. 1620ESv2 SERIES. Operation and Maintenance Manual

4000 SYSTEM OPERATING MANUAL

ELECTRICAL. 60 Minutes

G72x Series Direct Spark Ignition Controls

PATENT PENDING. Phone: (877) Operation and Service Manual

USE AND CARE GUIDE DUALCHECK CIRCUIT TESTER

S&C Vista Underground Distribution Switchgear

Fast Charge Option (IEC 335) MegaPulse Surge Testers

1/2 HP SUMP PUMP OWNER'S MANUAL

HE-PSU(4.0) Power Supply & Battery Charger Installation, Commissioning & Operating Manual

model ps600 Address all communications and shipments to: FEDERAL SIGNAL CORPORATION

INSTALLATION INSTRUCTIONS

CLEAN POWER TM CPS Series Operator s Manual

1 AMP CURRENT SOURCE

ELECTRICAL GROUNDING SAFETY PROGRAM

PAD-MOUNTED SWITCHGEAR INSPECTION & MAINTENANCE RECOMMENDATIONS TYPE PSI/II. 15kV 25kV. Qualified Persons

BellMega BM-510. Operation Manual Ver: 1.0 Rev: 1

EXP-EMG-2L-LE6. Explosion Proof Bug Eye Emergency LED Fixture Instruction Manual

High Frequency SineWave Guardian TM

BATTERY CHARGERS NO. 3

K25000 & KD25000 Series

8460SRC Replacement Chassis for Scully ST-6 and Biclops Monitors

TESCO THE EASTERN SPECIALTY COMPANY Date: 05/04/15 Canal Street and Jefferson Avenue Bristol, PA 19007

Manufacturing & Service: Bierer & Associates Inc. Patent No. 6,885,180. ST Series Service Tester Phase Identifier Operating Instructions ST500PGN

Battery Charger Retrofit Kit Q-DCCHG1 for DC Cabinet Retrofit Manual

VARNA Products 15 GPM (57 LPM) Prelube Kit for MTU 4000 Series Marine Engines

E Series CE Approved Intermittent Pilot Ignition Control

D4DL. Guard Lock Safety-door Switch. Model Number Structure

Table of Contents Title Page No.

CRECOMZR054B00 CRECOMZR055B00. VERTICAL AND HORIZONTAL ECONOMI$ER2 ACCESSORY MEDIUM ROOFTOP UNITS 15 to 27 1/2 TONS WARNING TABLE OF CONTENTS

PHASERS. Operating and Instruction Manual. a n d A C C E S S O R I E S

Notice Regarding this Upgrade. WARNING! Danger Potential

Owner's/Installation Manual

Monicon Instruments Co., Ltd. CHR-1285/2485 CHR-1285/2485 BATTERY CHARGER

E.S.P. Embedded Sensing Probes for Motor Brushes

INSTRUCTIONS FOR THE RELIANCE Fast/Tran TM ARL0909 & ARL0909R

AUTOMATIC DUST COLLECTION FOR SMALL SHOPS INSTRUCTIONS GG500B MOTOR CONTACTOR WITH THERMAL PROTECTION 115VAC COIL

& HIGH CURRENT DC POWER SUPPLIES INSTRUCTION MANUAL

Electrical Safety. Recognizing & Controlling Hazards

XLM 62V Energy Storage Module

PORTABLE POWER SUPPLY MODEL NUMBER VMS-1 VERSION 1.0

STARTING SYSTEMS 8B - 1 STARTING SYSTEMS CONTENTS

OBAE, OBAEXU, ON BOARD Battery Chargers

Electrical Safety. Introduction

Operator s Manual. Automatic Electric Jack Leveling. The leveling system shall only be operated under the following conditions:

Battery Powered Hydraulic Pump; Kit PN [ ]

OutdoorXUPS-600AHV-8364

Test Cabinet (Inspection Box) Instructions. for Testing Accessories on Manually and Electrically Operated EntelliGuard G Low Voltage Circuit Breakers

Operating Instructions - Electric Pow'r-Riser Models

Battery Chargers Sealed or Valve Regulated Lead Acid Batteries Models: PSC A AND PSC A

Freedom egen System End-of- Line Functional Checklist

3 Phase Smart Controller

LESTRONIC II BATTERY CHARGER TAYLOR-DUNN MODEL TYPE 24LC25-8ET

INSTALLATION INSTRUCTIONS FOR FACTORY MUTUAL APPROVED INSTRUMENTS

AGV5-1, AGV5-2 FORM AGV5 OM 11-02

Installation, Operation and Maintenance Manual

User s Manual Rev 1.3 GME

A3Z OPERATING MANUAL

Instruction Manual For Baldor Buffers

8 Light Controller. Instruction Manual. With Light Timer 240 Volts. Product # INNOVATING SINCE 1995

Phenix Technologies Inc. 75 Speicher Drive Accident, Maryland 21520

Instruction Manual for the 110 VAC Draw-off Controller System

HYDRAULIC SPREADER. Max. Pressure: 10,000 PSI (700 BAR) Unit Weight: HS2000 = 4.8 LBS. (2.17 Kg) HS3000 = 22 LBS. (9.98 Kg)

Rental Industry Safety Tester Safe Check 5s

CIVACON GROUND VERIFICATION RACK MONITOR SYSTEM and ASSOCIATED EQUIPMENT

EVS RP6020. Instruction Manual

OBE, OBEXU, ON BOARD Battery Chargers

Unrestricted. S.Mukhopadhyay

Philadelphia Solar Installation and Operation Manual

USER MANUAL. Maxwell Technologies BOOSTCAP 56V UPS Energy Storage Modules. Models: BMOD0130 P056 B02 BMOD0130 P056 B03. Document Number

Mini Multimeter with Non-Contact Voltage Detector (NCV)

GUIDELINES ON ELECTRICAL SAFETY AT WORKPLACES

Inductive. SeekTech Clamp WARNING! Operator s Manual

Transcription:

HI-POT/CONTINUITY TESTER Model PLEP01 Instruction Manual MITCHELL INSTRUMENT CO.

Dear Customer: Congratulations! Mitchell Instrument Co. is please to present you with your Hipot/Continuity Tester. This instrument allows quick and simple field testing of equipment to ensure proper grounding and insulation. Before using the Hipot/Continuity Tester, we suggest that you take a few moments to review this manual. Thank you for your trust and confidence. Rev. 1.0 May 7, 2007 i

Table of Contents An Introduction to Dielectric Withstand and Continuity Testing with the PLEP01...1 Safety Precautions...1 Test Personnel...1 Testing Area...1 Safety Techniques...2 Using the PLEP01 Dielectric Withstand Tester...2 Ground Check on Grounded Products...2 Ground Check Failures...2 High Vo ltage Dielectric Withstand Test...3 High Voltage Dielectric Withstand Test Failures...3 Introduction and Specifications...4 Introduction...4 Specifications...4 Operation...6 Setting up your Tester...6 AC Line Voltage Requirements...6 Front Panel Features...6 Initial Checkout Procedure...10 Operating Techniques...10 Operation Test...10 Testing Completed Products...10 Testing Extension Cords...11 Technical Assistance, Calibration and Repair Service...12 ii

Section 1 An Introduction to Dielectric Withstand and Continuity Testing with the PLEP01 The continuity test/dielectric withstand test is a field test which is used to verify proper operation of electrical equipment used in the work place and on construction sites. The test ensures that the primary circuit power and ground conductors are properly wired and connected for safe operation. It also applies a high-voltage potential between power and ground conductors to make sure that no unintentional leakage or arcing paths exist between power and ground. The test consists of a ground continuity check and a high voltage check. It is non-destructive to the equipment under test, and can be accomplished in a short time. A receptacle is provided for extension cords to allow them to be tested for hipot and ground continuity. Safety Precautions The dielectric withstand test generates voltages of up to 2500 volts ac at potentially lethal current levels. Currents of as little as 5 ma at 120 volts can cause death, and the PLEP01 is capable of generating a lethal output. The PLEP01 has been designed to minimize exposure to high voltages. However, the potential for serious injury or death exists and personnel should be aware when they conduct this test. Test Personnel Personnel require special training to conduct the dielectric withstand test. They should be aware that high voltage is hazardous. Do not wear any metal jewelry while using the PLEP01. The operator should keep the area free of unused leads and equipment. Others should be kept away while testing is being performed. Testing Area The area used for conducting the dielectric withstand test should be as remote as possible from normal work activities. Only personnel actually conducting the test should be allowed in the area, and it should be taped or roped off to preclude casual entry by other employees. In addition, the area should be marked "WARNING - HIGH VOLTAGE TESTING" or the equivalent to warn others of the nature of the testing taking place. The bench being used should be non-conductive, and any exposed metal parts should be tied together and grounded. If a conductive surface must be used, it should be grounded. Because of sparking during a dielectric test failure, it is not safe to conduct dielectric withstand tests in combustible atmospheres. It is imperative that a good ground be provided to the PLEP01. Before connecting the PLEP01, ensure that a low-resistance ground is provided by the building wiring. If the PLEP01 is used on a highresistance grounding circuit, dangerous high voltages may be present to the operator. In addition, the 1

power to the Testing Area should be provided with an easily reached shutoff switch which can be actuated by personnel outside the Area if needed. Safety Techniques The high voltage circuit of the PLEP01 is only energized while the TEST button is being pushed. High voltage can be shut off at any time by releasing the TEST button. The PLEP01 is equipped with a Ground Continuity Check which makes sure that grounded systems have continuity between the ground pin of the power plug and exposed dead-metal parts. This is an important safety feature as well, and high voltage should not be applied to any equipment using a three-wire plug if the ground continuity light is not lit. If equipment using a two-wire plug is being tested, it is imperative that the operator make absolutely sure the ground probe is properly connected to the equipment being tested. If the lead is not properly connected, a dielectric withstand test failure may energize exposed dead metal of the equipment being tested. Additionally, without proper connection of the ground probe, the PLEP01 may not recognize the failure. The test will continue for its normal length of time, and the PLEP01 may show a "PASS". The PLEP01 should be set up to the right and between the operator and the equipment under test. The operator must touch the ground probe to exposed metal of the equipment under test before the TEST button is pushed. Using the PLEP01 Dielectric Withstand Tester The dielectric withstand test involves high voltage and caution should be exercised when using the Tester. The Tester's return lead is connected to ground potential and when properly connected to the equipment being tested, it will guard against the operator contacting high voltage. Always make sure the return lead is firmly connected to exposed dead metal. In the sections below, the two tests are discussed. Ground Check on Grounded Products The Ground check ensures that the grounding pin of the power supply cord is properly connected to the exposed dead metal of the chassis. If this connection is absent, the exposed dead metal of your product could be at line potential indefinitely after an internal wiring fault, causing risk of shock to anyone touching it. If the green Ground Continuity LED lights, the connection between the grounding pin and the exposed dead metal of the equipment being tested is good. Testing can continue. Ground Check Failures If the green Ground Continuity LED does not light when a grounded product is tested, the test should not be continued. The connection between the grounding pin of the attachment plug cap and the exposed dead metal of the chassis has a high resistance or no connection. This indicates a problem with the connection of the PLEP01 to the equipment being tested, or that the ground connection in the equipment being tested is defective. The connection to the tester should be checked, taking care to make a good 2

connection. The unit should then be retested. If a failing result is repeated, the connection between the power supply cord ground pin and the chassis of the equipment being tested should be checked and reworked until a passing result is obtained. Do not use equipment with a three-wire plug that has failed the Ground Continuity Test. It is unsafe. High Voltage Dielectric Withstand Test This test checks for insulation system breakdowns between the primary and ground circuits. The dielectric withstand test is conducted by shorting the line and neutral conductors of the power supply cord and applying high voltage between them and the exposed dead metal of the equipment being tested. The test is continued while the TEST button is pressed. High Voltage Dielectric Withstand Test Failures If the red Hipot Fail LED lights, a problem has been found with the insulation system between primary and ground, or the leakage current between line and ground is more than the currently set leakage current limit, set at 5mA at the factory.. The equipment under test should be removed from service until the problem can be determined. Do not use equipment that has failed this test. It may be unsafe. 3

Section 2 Introduction and Specifications Introduction This manual contains complete operating, maintenance and calibration instructions for the Mitchell Instrument Co. Model PLEP01 Dielectric Withstand Tester. Specifications Specifications for the PLEP01 are listed in Table 1. The instrument is a bench-type Dielectric Withstand Tester with AC Output, designed for production line testing. The PLEP01 features automatic one button operation, with numerous safety features designed to protect the operator:? The Return Lead is directly connected to ground potential for operator safety.? In case of trouble, the test can be immediately terminated at any time by releasing the TEST button.? If a hipot or leakage failure is encountered, the high voltage is immediately cut to protect the operator. The Hipot Fail LED on the front panel lights to alert the operator. Convenience and testing features include:? Operator instructions are printed on the side panel for quick reference.? Three wire extension cords, two wire (double insulated), and three-wire (grounded) products can be tested with one tester. Your Tester is warranted for a period of one year upon shipment of the instrument to the original purchaser. 4

ELECTRICAL Output PRETEST: 500 Volts AC. METAL CASED: 1000 Volts AC. DOUBLE INSULATED: 2500 Volts AC. 5 ma preset. Leakage Current Pass/Fail Criteria: Ground Continuity: 3 ohm ± 5% (Metal Cased only). Leakage Current: Internally adjustable 0.5-5 ma. Dielectric Breakdown: Separate high frequency detection circuit for breakdown spike detection. Duty cycle 100 %. ENVIRONMENTAL Operating Temperature Relative Humidity Range GENERAL Input power requirements Weight Failure Indication Operation Instructions 15-40?C. 0-90% non-condensing. Model PLEP01: 114-127 volts, 50/60 Hz. 3 lbs. Visual, provided by red LED on front panel. Hipot Test automatically terminated on failure. Provided on side panel. Table 1. PLEP01 Specifications 5

Section 3 Operation This section describes how to set up and make measurements with your Tester. We recommend that you read the entire section carefully so that you can use all of the features of your Tester. Setting up your Tester Your Tester is shipped with this manual. If reshipment of the instrument is necessary, be sure that adequate protection is provided to prevent damage during shipment. We recommend that the instrument be surrounded by at least one inch of shock-absorbing material on all sides of the container. Remove the Tester from its container and place it on a test bench. AC Line Voltage Requirements The PLEP01 can only be used on voltages noted in the Specifications. Operation at other voltages will cause improper operation and may permanently damage the PLEP01. Front Panel Features Before using your Tester, take a few minutes to become familiar with the use of its controls, indicators and connectors. The front panel features of the PLEP01 are shown in Figure 2 and described in Table 2. 6

7

ITEM NAME FUNCTION 1 Power LED When lit, indicates that the PLEP01 is ready to test. 2 Continuity LED When lit, indicates that the PLEP01 is properly connected to a Metal Cased product, and the product has passed the Continuity Test. 3 High Voltage LED Lit when TEST Button is pressed. Indicates that the hipot test is being conducted with no failures. 4 Hipot Fail LED Indicates failure of the hipot test. High voltage output is terminated. 5 Voltage Select Switch Selects which Hipot Test voltage of 500, 1000 or 2500 volts will be applied to the Hipot Receptacle when the TEST button is pressed. 6 Ground Probe Used during Continuity and Hipot Testing to touch all exposed metal parts of the equipment being tested. 7 Hipot Receptacle Equipment or extension cord under test to be plugged into this receptacle. 8 Extension Cord Receptacle Extension cord under test to be plugged into this receptacle. 9 TEST Button While pressed, allows selected Hipot Voltage to be present on the Hipot Receptacle. Releasing button stops test. 10 Power Cord Plug into correctly rated source of supply. See Specifications. Table 2. Controls, Indicators, Connectors - Model PLEP01 Front Panel 8

9

Initial Checkout Procedure The following procedure will verify that the Tester is working correctly. We recommend that this procedure be conducted periodically to ensure proper operation of the Tester. CAUTION High voltage (up to 1000 Volts AC) generated by the PLEP01 is exposed during this test. A risk of shock exists. Exercise care when using the PLEP01. 1. Connect the Tester to a proper source of supply using the included 18 AWG power supply cord. The green Power LED should light. 2. Touch the tip of the Ground Probe to the U-shaped grounding pin of the Hipot Receptacle. The green Ground Continuity LED should light. 3. Remove the Ground Probe from the Hipot Receptacle. The green Ground Continuity LED should go out. 4. Carefully hold the insulated portion of the Ground Probe. High voltage will be exposed during this test. 5. Select the 1000V/Metal Cased setting of the Voltage Switch. Hold down the test button. The orange High Voltage LED should light. 6. Carefully insert the Ground Probe into one of the parallel openings in the Hipot Receptacle. The orange High Voltage LED shoud go out, and the Red Hipot Fail LED should light. Passage of these tests indicates that the PLEP01 is functioning properly and that it is safe to use. If the results of the performance test are not in accordance with the above, service is required. Remove the PLEP01 from service and contact the manufacturer for servicing information. Operating Techniques The following paragraphs describe how to operate your PLEP01 Dielectric Withstand Tester. Before proceeding with testing, the PLEP01 Tester operation should be verified. CAUTION: High voltage is generated by the PLEP01. Although the chassis of the equipment under test is grounded by the PLEP01, a risk of shock exists. Exercise care when using the PLEP01. Operation Test We recommend that the operation of the PLEP01 be audited periodically by conducting the performance test described above in the "Initial Checkout Procedure" section. Testing Completed Products This section describes how the PLEP01 is used to conduct a hipot/continuity test. Before continuing, we recommend that you read Section 1, "An Introduction to Dielectric Withstand Testing with the PLEP01". It contains valuable safety, operation and test result evaluation information which can help you conduct the test safely and correctly. The test can be stopped immediately at any time by releasing the TEST button. 1. Connect the PLEP01 to a correctly rated source of supply. The green Power On LED should light, indicating that the PLEP01 is ready to test. 2. Make sure the power switch of the unit being tested is ON. 3. Plug the equipment under test into the Hipot Receptacle of the PLEP01. 4. Using the Voltage switch, select the proper test voltage. Avoid using test voltage higher than recommended below, or permanent damage to the equipment under test may result: a. 500 V:Operational testing of new equipment before energizing. b. 1000 V: For testing a product with a three-wire power supply cord and/or a metal enclosure. c. 2500 V: For testing double insulated products. 5. Touch the Ground Probe to exposed metal on the piece of equipment being tested. 6. Press the TEST button, and hold for test duration. 7. If the orange High Voltage lamp stays lit, the equipment under test passes the tests. 10

8. If the red Hipot Fail LED lights, a hipot or leakage current failure has occurred. Put the equipment under test aside and do not use it until the problem has been fixed. 9. Release the TEST button. The PLEP01 has completed the test. Testing Extension Cords This section describes how the PLEP01 is used to conduct a hipot/continuity test to test three-wire extension cords with grounding pins.. Before continuing, we recommend that you read Section 1, "An Introduction to Dielectric Withstand Testing with the PLEP01". It contains valuable safety, operation and test result evaluation information which can help you conduct the test safely and correctly. The test can be stopped immediately at any time by releasing the TEST button. 1. Connect the PLEP01 to a correctly rated source of supply. The green Power On LED should light, indicating that the PLEP01 is ready to test. 2. Plug the extension cord into the Hipot Receptacle and the Extension Cord Receptacle of the PLEP01. 3. Using the Voltage switch, select the 1000 volt range. Avoid using test voltage higher than recommended. The Ground Probe is not used in this test. 4. Press the TEST button, and hold for test duration. 5. If both the green Ground Continuity LED and the orange High Voltage LED stay lit, the extension cord passes the tests. 6. If the red Hipot Fail LED lights or the green Ground Continuity LED is not lit, a hipot or ground continuity failure has occurred. Do not use the cord until the problem has been found and corrected. 7. Release the TEST button. The PLEP01 has completed the test. 11

Section 4 Technical Assistance, Calibration and Repair Service For Technical Assistance Phone: (800) 748-6224 Technical Assistance is available between the hours of 8:30 AM and 4:30 PM Pacific Time. We are also available to repair and calibrate the tester as needed. 12