Operating manual. ph / mv controller type M8832N

Similar documents
Table of Contents 1. OVERVIEW OPERATION FUNCTION SPECIFICATIONS INSTRUMENT LAYOUT INSTALLATION...

OD DIGITAL DISSOLVED OXYGEN CONTROLLER POLAROGRAPHIC CELL

EG3000. Generator Electronic Governor Controller Operation Manual

ProMinent. ProMinent D4a Metering Pumps with built-in proportional controllers

ESC2301. Universal Electronic Governor Controller Operation Manual

Electronic Ballast EVG 2000-T

5001TCP SPEED CONTROLLER

LTX RF LEVEL SENSOR. Instruction Manual

5001TCP SPEED CONTROLLER

OD OD DIGITAL DISSOLVED OXYGEN CONTROLLERS

EG1069X. Generator Electronic Governor Controller Operation Manual

ST Wiring diagram. Product description. Temperature controller. Order number

HIGH POWER SOLENOID DRIVER 1

Control and automatic dosing

ENGINE GOVERNING SYSTEMS

Harris IRT Enterprises Digital Resistance Tester Model XP

ENGINE GOVERNING SYSTEMS LSM672 LOAD SHARING MODULE. GOVERNORS AMERICA CORP. 720 Silver Street Agawam, MA , USA MEMBER

FEED WATER REGULATOR EVM-1F FOR CONTINUOS REGULATION

The Temperature Control Company SERIES 2000 / 3020 ANALOG TEMPERATURE CONTROLLERS. Instruction Manual

INSTALLATION & OPERATION MANUAL

ST48-WHUV.102. Wiring diagram. Product description. PID controller. Order number

1/16 DIN Temperature Controller E5CS

SB 2000 PUSH TO SEARCH NEXT STAG E. Aerotech, Inc. FORM: QM 1320

MX431 Generator Automatic Voltage Regulator Operation Manual

PLATINUM SERIES SWITCHBOARD METERS

EG1065X. Compatible with Barber Colman* Dyn1-1065X series * Use for reference purpose only and not a genuine Barber Colman product.

ecotrans Lf 01/02 Microprocessor Transmitter / Switching Device for Conductivity

SP5 INSTALLATION AND SETUP MANUAL

R & D SPECIALTIES ROTROL I USER'S MANUAL

Model 930 Power Control System Instruction Manual

SUM340 SUMMATION BOARD

Instruction Manual. Anderson Instrument Co. Inc. 156 Auriesville Road Fultonville, NY Fax

OD DISSOLVED OXYGEN METER

Combined Ventilation Controller RVWS-T-224HA

Drive electronics pq11

HDS 5105 Amplified pressure sensor/switch

PRO REMOTE Series CAPACITANCE PROBE

Installation Instructions Spector compact Conductivity Transmitter System LRGT 16-1

A3Z OPERATING MANUAL

TABLE OF CONTENTS. 8.3 Online Sensor Specifications. Series TS (Models TS1, TSP, TSH, TSL,TSF, TSB1, TSB2) According to Electromatic factory procedure

Model 322 Automated Thermocouple Calibrator Operating Instructions. Product Description. Practical Instrument Electronics

A6Z OPERATING MANUAL

MPT SERIES OPERATING MANUAL. Caution: To Operate this controller safely, use the Safe Operating Area Design Tool Online at

INSTALLATION INSTRUCTIONS

MODEL ELC-12/40-CVM-D BATTERY CHARGER

Model 422 Automated Thermocouple Calibrator Operating Instructions. Product Description

SWA-2441/2451 PID TEMPERATURE CONTROLLER INSTRUCTION MANUAL

Flowmeter Ultrasonic Flow Metering / Dosing Device. Operating Manual

OPERATING AND MAINTENANCE MANUAL. Primary Current Injection Test Set. 750ADM-H mk2

Operating Instructions. Angle Seat Control Valve. Type 7020

ACO Building Drainage. Microcontroller Dual pump control. Operating Instructions. Version: 15/10/2015

PPT 802, Proximity Probe Transducer System

JUMO ecotrans Lf 01/02

유량계전문제조업체. Electro Magnetic FLOWMETER

PRODUCT SERIES: DPG-2100

SIGNET 2760 DryLoc ph/orp Preamplifier

SP4 DOCUMENTATION. 1. SP4 Reference manual SP4 console.

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE PTC-10 PELTIER TEMPERATURE CONTROL SYSTEM

MATRIX LLC LEVEL CONTROL PANEL

VEGAMET 307 / 308. TIB Technical Information Operating Instructions VEGAMET 308. Signal conditioning instrument. for continuous level measurement

SWA-24X2C PID Temperature Controller with Timer (SWA-2442C/2452C)

PRODUCT INFORMATION BULLETIN

Turbidity and Suspended Solids Transmitter

12 Series Linear Actuators. Operation & Maintenance Manual, Analog Positioner Installation

VP23 3-way proportional pressure control valve Directly-controlled seat valve with µp-driven pressure control G 1/4... G 3/4

SCR Power Controllers

PIECAL 322 Automated Thermocouple Calibrator Operating Instructions. Product Description

ESD5550 / 5570 Series Speed Control Units

DC Variable Speed Drive Panel

MAIN FUNCTION FEATURES. . D = v t /2

System 350 S350P Proportional Plus Integral Temperature Stage Module

The T5220 Series Electro-Pneumatic Transducer converts a DC current or voltage input signal to a directly proportional pneumatic output.

Temperature Controller. TC5+2V4SA Plus USER'S MANUAL

VT-SENSOR AIR VELOCITY AND TEMP

CP-1TT Remote System

MD10. Engine Controller. Installation and User Manual for the MD10 Engine Controller. Full Version

Temperature Controller OVATION 214 User's Guide

Electronic Flow Monitor for liquids

Inlet Controller SB3500 USER'S MANUAL

ESD5100 Series Speed Control Unit PART NUMBER ESD5120 ESD (127)

Superstatic 440. Static Heat- and Cooling Meter. Application

& HIGH CURRENT DC POWER SUPPLIES INSTRUCTION MANUAL

Sensors and Systems for Combustion Engineering

VP23, 3-way proportional pressure control valves Seat valve with μp-driven pressure control

TROUBLESHOOTING AND MAINTAINING ELECTRONIC KILN CONTROL SYSTEMS

Product Manual MNX10010 / REV B MODEL PS03

l The Battery Tester is designed for measuring the l AC four-terminal method to measure the internal

Allan Juul Larsen, MDT LEO9 Prime Serv Piraeus MAN Diesel & Turbo Allan Juul Larsen Examining HCU Events

INSTALLATION, OPERATING, AND MAINTENANCE INSTRUCTIONS DIAGRAMS & PARTS LIST PMC-1 ELECTRO-PNEUMATIC CONTROLLER

INDUSTRIAL CHARGER AUTOMATIC BATTERY CHARGER SERIES 150 KUSSMAUL ELECTRONICS CO., INC. MODEL # XX YEAR WARRANTY INSTRUCTION MANUAL

V-SENSOR LOW AIR VELOCITY SENSOR

User Instructions for WRC1021D. Doc

ESD5500E Series Speed Control Unit

Fisher 2506 and 2516 Receiver Controllers

Manual Installation & Operation

HYGRODYNAMICS DIGITAL DEW POINT MONITOR PC BOARD MODEL 6392N & 6392N2 TABLE OF CONTENTS

SPECIFICATIONS ATTENTION OPERATION DESCRIPTION

Differential pressure sensor

System Electropneumatic Converters (Proportional Valves) Electronic Process Controllers Signal Converters

TECHNICAL DATASHEET#TD1516AX 3 AMP SOLENOID DRIVER

Transcription:

Mostec AG Mess- und Regeltechnik Lausenerstr. 13A CH-4410 Liestal / Switzerland Tel. 061/921 40 90 Fax 061/921 40 83 Operating manual ph / mv controller type M8832N

Description of the «leak proof housing» Warranty: At the front of the splash water proof casing, the inserted plug-in unit is sealed with silicon rubber material. A «SERTO» screw fitting to supply compressed instrument air is mounted on the rear panel. The signal and relay output terminals are also sealed with silicon rubber. As a result, the controllers application range is extended and prevents dirt or gases from entering the casing. Mostec warrants this product to be free of manufacturing defects for a 2-year period after the original date of purchase. Within this period, defective products will be repaired free of charge provided that the defect occured during normal operation. This warranty does not cover damage to the product resulting from ordinary usage such as front panel scratches, against control elements and corrosion, etc... The customer is responsable for paying shipping and packaging charges for products returned under warranty to Mostec. Mostec warrants this product beyond the 2-year warranty period for an additional 2 years in case of long term damages due to improper manufacturing. Such damages as poorly soldered joints or other assembly problems are covered by the warranty. Transportation damages are not covered by the warranty and should be referred to the respective postal delivery service. Mostec AG Mess- und Regeltechnik Lausenerstrasse 13A CH-4410 Liestal Switzerland Tel: 061/921 40 90 Fax: 061/921 40 83 12 1

Index Example: Actual value 10 Adapting the controller to a titration curve 9 Alarm circuit 10 Asymmetry (isothermal) 10 Block diagram of instrument 6 Connection example 8 Connections 4 Dead band 10 Dimensions 4 Example of batch operation 11 Example of two pulse diagrams 7 External setpoint 5 / 8 Function of operating controls 5 Front view 3 115/230 VAC line selection 8 Installation dimensions 4 Instrument design 6 Isolation buffer amplifier 6 Leakproof housing 12 mv / ph (gain) 10 P-Band 10 Pt-100 sensor connection 4 Rear view 4 Recorder signal output 4 Setpoint 10 Specifications 3 SU-switch 4 /10 Temperature 10 Terminology 10 T on / T off 10 Triaxial connection 4 Typical applications 7 Given: - Actual-value = 4 ph - Setpoint = 7 ph - P-band = 2 ph - T on = 5 sec. T off = 20 sec. For the controller M8832N, the dead-band is 10% of the P-band, therefore when the actual-value in this example reaches 7±0.2 ph, the controller will not inject reagents, and the instrument no longer controlls the ph. Final conclusion: for more precise controlling, a smaller P-band should be selected. The P-band range is from 7.00 to 5.00 ph [(setpoint) - (P-band)]. Therefore as long as the actual-value is less than 5.00 ph, the base valve will be open constantly and by the addition of base, the actual-value moves slowly to 7 ph. In the moment where the actual-value exceeds 5 ph, the base valve begins to turn off and on (pulse-widthmodulation). In the middle of the P-band, 6.00 ph, the valve s turn-on/off times are 5, respectively 20 seconds. Ton and Toff refer always to the middle of the P-band (7 - (P-band/2). Before and after, these openning and closing times obey a logarithmic function, inside the P-band the amount of base is always less as the actual-value approaches 7 ph, then the openning and closing times will become respectively shorter and longer. At 6.8 ph, the process enters the dead-zone and the actual-value remains constant and no additional base will be added. At this point, the setpoint lies in the steep part of the titration curve, and from a control theory point of view, it is advisable that T on is much greater than T off for the following reasons: - The regulation should not «overshoot», therefore inside the dead band the actualvalue should «stop» and not «overshoot». - The controller should - shortly after injecting - have some time «to look» how the actual-value behaves after this «shoot». The longer the turn-off time is, the less critical is the regulation. In case of non critical regulation, Toff can be reduced just until little overshooting occurs. Thus in batch type operations, expensive process time can be saved. In case of problems concerning controll settings which have not been described in this manual, please contact Mostec AG directly. 2 11

Terminology for the Mostec ph-controller M8832N Front view Actual-value: Asymmetry: Value which is measered from the ph-electrode, compensated for the isothermal intersection point an~ the electrode s temperature offset. = isothermal intersection point. This is the operating point of the electrode at which zero millivolts is measured. This point is usually defined as 7.00 ph. Conventionally, there is an offset of a few mv and therefore it should be adjusted with «Asymmetry». 7.02 mv / ph: = electrode gain. At zero degree Celsius, the electrode should measure 54.196 mvph. In practice, there is a deviation of a few % and therefore it should be adjusted with. Temperature: The ph-electrode has a high positive temperature coefficient. The error is approximately +3.6% per 10 degree Celsius temperature rise. This error can be corrected with the external temperature sensor or with the temperature potentiometer. Setpoint: P-Band: Dead-band: This is the value that the controller should achieve. The range in which the controller demonstrates proportional behaviour (see example on page 11). A zone in which the controller is unable to administer alkaline or acid. The servo units (pumps, valves, etc.) are not in operation. T on : Servo-unit s turn-on-time (see example on page 11). T off : Servo-unit s turn-off-time (see example on page 11). Alarm circuit: SU: The alarm timer supervises the output signals to the servo-units. Should a servo-unit be turned on for a time period which is longer than that pre-selected, an alarm is activated. This can occur when the reagent tank is empty, or a stalled stirrer, or a defective electrode etc. This switch is used to initialize and program the controller. During normal operation, this switch must not be in the SU position or the error message EE2 is displayed and the controller can t work. Specifications Controller principle: Quasi-steady three-point P-controller with preselectable transfer function Range: 0...±1000 mv and 0...14 ph ph zero asymmetry: ±1 ph ph-slope: 50 mv/ph to 62 mv/ph at 0 C ph-temp. compensation: 0-100 C or automatic with Pt-100 platinium sensor, 2-conductor design Temperature Display: 1. Manual 0...100,0 C 2. Pt-100-10... +100,0 C ±1 C Signal input: Floating, with isolation amplifier, common mode voltage max. 500 VDC Input impedance: 10 12 ohm typical Input bias current: 1 pa at 25 C typical Temperature drift in relation to input: 50 mv/ C max. Operating temperature: 10 to 45 C Setpoint values: 0... ±1000 mv or 0...14 ph, adjustable with potentiometer and display Proportional band: 0...1000 mv or 0... 7 ph ln the P-band = Zero position, the controller works as a two-point switching controller Dead band:»10% of «P»-band P-band deviation: ± 2% of end values Valve times: Alarm delays: Relay outputs: Current output: External setpoint: Power supply: Weight: Accessories: Optional equipment: 0.5 to 20 or 0.05 to 2.0 seconds in relation to the mid of P-band 1-100 minutes or 1-100 seconds, can be switched off «UP» valve: change-over contact 1 Amp./230 V, resistive load «DOWN» valve: switch-over contact 1 Amp./230 V, resistive load ALARM: change-over contact 1 Amp./230 V, resistive load 0... 20/4... 20 ma for 0...14/2...12 ph and special ranges. Ri 1 M, max. Ioad 500 ohm at 20 ma 0... 20/4... 20 ma for 0...14/2...12 ph load=51 ohm 115/230V, 50-60Hz» 10 VA approx. 1.4 kg or 3 Lbs Installation kit, 2 keys and manual Leakproof housing Front panel available in different languages Other signal outputs Computer interface RS232C 10 3

Rear view Adapting the controller to a titration curve If fast acting electric valves are used for a given process, it is preferable to keep the valve switching times or the time ratio between «T on» and «T off» as low as possible. If hydraulic or pneumatic valves are required, the periods have to be increased because of the relatively large opening and closing times of these valves. The added or delivered reagent quantities must have a certain ratio to the volume of the process vessel to make sure that the reaction does not last too long and to avoid too much reagent being added, because it would cause the control system to oscillate. Power line 115/230 VAC with protective ground Relay change over output for: - Acid ( ) - Base ( ) - Alarm Terminal no. 11,14 and 17= normally open contact Terminal no. 12,15 and 18= normally closed contact Analogue output for: Recorders, remote displays, process control computers etc. Terminals for remote PT-IOO temperature probe in the process vessel External analogue setpoint input 0... 20 ma or 4... 20 ma Remote setpoint selec. ON/OFF: Jumper ¾ = remote setpoint active, without jumper = internal setpoint +24VDC against terminal no. 3 max. Ioad 10 ma 4 132 1 Reference terminal (= Signal ground of the input isolation amplifier) 2 Ground terminal, to ground the outer most shield of the triax coaxial cable 3 ph-probe 4 Set-Up switch. Used to program the controller. Must not be at SU during operation. If the error message EE2 is displayed, the switch is in the wrong position. Programming the controller will not be documented in the users manual, this will be factory set only. In case of relatively flat titration curves in which the ph value is roughly proportional to the reagent added, «T on» should be set to equal «T off». However, if the titration curve is steep at the endpoint, «T on» must always be shorter than «T off». In extreme cases, «T on» = minimum and «T off» = maximum must be selected. Optimum time conditions are to be determined empirically. During the first run, «T on» is kept very small as compared to «T off». If the neutralization lasts relatively long, «T on» must be, increased until there is no «overshoot» of the control process (opening of the other valve when the setpoint value is exceeded). By changing the P-band together with «T on» and «T off», the controller can be adapted to any desired titration without causing the system to oscillate. Front panel cut-out: (Units are mm) * Distance from one instrument to the next 4 9

Connection example N L1 PE Power line 115/230 VAC 50/60 Hz Alarm device Base valve, pump or power relay Acid valve, pump or power relay Operation of the controller Function of the operating controls 1a «mv» measuring range: When this switch is in the «mv» position, it is possible to control redox or other processes which generate mv signals to indicate actual values. With the instrument set to this measuring range, «ASYM», «mv/ph», «Temp. C» controls are out of operation. 1b «mv» measuring range: In this range, the actual value is the ph signal that has been derived from the formula UpH(mV)= 54.196 [1+(0.003661 Tc)] 6 7 «Setpoint» potentiometer: This control is used to set the «mv» or «ph» setpoint. «P-band» potentiometer: Serves to adjust the width of the proportional band (amplification). A low P-band means strong amplification and vice versa. 8 «Auto/Manual» switch: This switch is used to control valves either through the controller or manually. In manual mode the valves can be controlled by two push 10a Alarm circuit: The opening of one of the two valves (in the «Auto» operating mode only) starts a delay timer which can be adjusted from one to 100 minutes or 1 to 100 seconds. If a valve remains open for a period that is longer than the preselected period, an alarm signal is triggered. Depending on the application, this can have the following meanings: 1. The ph-electrode is defective. 2. Reagent tank is empty. 3. Defective circuitry. 4. Defective controller or wiring to controller. 5. Defective stirrer. 6. Process flowrate in waste water treatment plant is too high. 13 Change over switch to display the actual temperature in C. When the switch «Pt-100» is in the UP posihon then the accurate manual temperature setting is displayed. In the DOWN (Pt-100) position it displays the temperature of the remote Pt-100 sensor. 14 LED remote controlled. This lamp is on when the controller is remotely con trolled by the RS232C or the RS485 interface from a host computer. Analogue instrument, recorder or process computer Pt-100 platinum temperatur sensor 1a 1b 14 11 2 7.02 13 3 10b 10a Analogue external setpoint Jumper/switch to select internal/ external setpoint 12a 12b 4 9 5 8 7 6 Power supply board. Selecting the proper line voltage 115/230 VAC. View to the component side of the PC board. 115 VAC 50/60 Hz: Jumper B2 and B3 placed 230 VAC 50/60 Hz: Jumper B1 placed in relation to 0 C The ph setpoint can be set anywhere between the full operating range of 0-14 ph. In this case the temperature dependence, electrode slope and asymmetry can be compensated for. «Manual/Pt-100» switch: This switch is used to change from manual temperature compensation to a platinium sensor located in the process liquid (in most cases, this sensor is housed next to the ph electrode). If the switch is in the «Pt- 100» position and the sensor is not wired to the instrument, the instrument works with a temperature of 20 C. «Temp.» potentiometer: A manual control to compensate the temperature coefficient of the ph electrode. «Asymmetry» potentiometer: Is used to adjust the zero mv point of the ph signal (ph = 7) «Slope» potentiometer: Serves to adjust the slope of the ph electrode by means of a calibration (buffer) solution. buttons. Two light emitting diodes indicate the valve position. If a given lamp lights up, the corresponding valve is open. «Up» means process to more positive redox potentials or higher ph values. «Down» means process to more negative redox potentials or lower ph values. «T on /T off» potentiometer: These two controls are used to adjust the valve opening or closing time in relation to the midpoint of the P-band. If both potentiometers are adjusted to the same value, the open and close times of the corresponding valve are equal and in the center of the P-band. In this case, the transfer function of the system would be linear. These controls make it possible to adjust the controller to the shape of a titration curve so that even with a titration curve having a steep end-point the transfer behavior remains linear and the control system does not begin to oscillate. The maximum nonlinearity ratio is about 1 to 40 (0.5 to 20 seconds). The dead band of this P-controller in which none of the two valves is actuated, represents about 10% of the P-band. 7. Valve is stuck in the open or close position, etc. The alarm is displayed by a LED at the front panel. If the alarm circuit is not used, it can be switched off. Digital LED display: It displays the actual values, the temperature, the setpoint value and error messages. Pressing this button changes the display from actual value to setpoint value. Remote control light. This LED is on when the controllers setpoint comes from an extemal source as an analogue current signal with a jumper installed between the terminal 3 and 4. The intemal set point on the front panel is off. 8 5 2 3 4 5 9 10b 11 12a 12b

Instrument design Typical application of the M8832N controller The system consists of the following components: 1 High impedance FET amplifier 2 Isolation amplifier for the ph-probe signal 3 16-channel A-D converter 4 Front panel with controls and digital display 5 Micro computer 6 D-A converter 7 Relays for valves and the alarm 8 Power supply 9 Option: Interface with several standard configurations to remote control the controller or to read back process values. All the electronic components are located on four printed circuit boards. The rear panel has two terminal strips with 10- and 12-pole screw plugs to connect the controller to the external wiring. A 13-mm ph-plug with teflon insulation is used for the probe signal. The integrated isolation amplifier isolates the ph-probe signal from the control processor ground. The processor ground is floating against protective ground. The printed circuit boards are build in a plastic case which is shielded with a conductive nickel coating. The transparent polycarbonate door, fitted with rubber gasket and lock, is splash water proof. 6 7