Prima OPH User Manual

Similar documents
PRIMA DNT User Manual

XPERT SL 45 Slit Lamp. Instruction Manual

PRECISION MICROSCOPES

Flexible Arm Stand F 360. User Manual

Preface. Scope of Application. General Requirements for Safety THE SAFETY MARKS USED IN THIS INSTRUMENT

AM-3000 Surgical Microscope User s Manual

SCAN OPTICS SO-161-R MICROSCOPE USER MANUAL

MANUAL EXC-120 MICROSCOPE SERIES. 73 Mall Drive, Commack, NY (P) (F)

12 Volt dc supply Connecting the power supply to a 12 Volt dc supply (battery)... 12

SCAN OPTICS SO-5800 MICROSCOPE ASSEMBLY MANUAL

USER MANUAL. SLIT LAMP MICROSCOPE Model: GR-7. Version: 1.0. Gilras LLC

USER MANUAL SLIT LAMP MICROSCOPE

INTRODUCTION...5 PARTS LIST...6 PACKAGING...9

INTRODUCTION...5 PARTS LIST...6 PACKAGING...9 ASSEMBLY INSTRUCTIONS... 10

HALOGEN FLOODLIGHTS Models CHL1260C & 1260T Part Nos: &

Stereo Microscopes Instruction Manual

ULTRA QUARTZ FOLLOWSPOT USER MANUAL

MOVING HEAD. User manual CAUTION!

TECHNICAL MANUAL. FOR BLAST 2 (Bright Light Aiming System Tactical 2) For Rifles and Pistols Equipped with MIL-SPEC-1913 and Picatinny Rails

Mania DC3 user manual

SZ2-ILA INSTRUCTIONS TRANSMITTED ILLUMINATION ATTACHMENT SZ-ADD SZ-POL-2 SZH-CLJ. Optional Modules

User s Manual Dental Microscope LED

XCEL 250 XCEL 400/700. Xcel Slit Lamp USER'S GUIDE Rev. F Page1

Mania EF1 user manual

C.fm Page 1 Thursday, February 21, :28 AM Raptor user manual

ACROBAT SWING STAND MODEL

ALU-29CF Portable Light NSN:

01 / Pralinenfix PFM 2000

Ichip Mystére LED FOLLOWSPOT USER MANUAL

AMCON AUTOMATIC CHART PROJECTOR USER S MANUAL EQ-6002

Snapshot Ok on Dimmer Outdoor OK Sound Activated DMX512 Master/Slave 115V/230V Switch Replaceable Fuse User Serviceable Duty Cycle

OPERATION MANUAL MODEL MODEL MODEL MODEL MODEL NON-CONDUCTIVE ARTICULATING FIBERSCOPE INSTRUMENT TECHNOLOGY, INC.

Installation Operation Maintenance Troubleshooting Version2.1. Feb/09. Led II

Light Source User Guide

Snapshot LX5 USER MANUAL. OK on Dimmer Outdoor OK Sound Activated DMX512 Master/Slave 115V/230V Switch Replaceable Fuse User Serviceable Duty Cycle

Snapshot LX10 USER MANUAL. OK on Dimmer Outdoor OK Sound Activated DMX512 Master/Slave 115V/230V Switch Replaceable Fuse User Serviceable Duty Cycle

Ego 01 Ego 02 Ego 03 Logo Projector user manual

Serial Number: Exer 3/6 Treadmill

Fiber Optic Illuminator. Model 21AC Model 21DC

LED Pinspot 2. Snapshot. User Manual

D.fm Page 1 Wednesday, March 5, :21 PM Raptor user manual

LED Twister II. User manual UK Version 1.0

SO-111 Ophthalmic Microscope User Manual Page 1 of 30 CONTENTS

INSTALLATION INSTRUCTIONS STUDIO S

Twin-Beam Spotlight 3,000,000 Power Series Corded Cordless Rechargeable

INSTALLATION MANUAL MULTI DECK HOT MERCHANDISERS MD PREMIUM. Versions. Solid back level Pass through level

Keeler Spectra Plus. Instructions. Next

ECO Spot 2500 User Manual

STAND MODEL FOR MACH M2

Motion System Components Diagram. Note: #2 Mirror Cover and X-Axis Motor Cover have been removed for visibility. Maintenance.

These instructions are to be left with the user

ECO Spot 2500 User Manual

USER MANUAL LUX-LD30W LED MOVING HEAD

2. Description of Standard NUV Lightsource (350nm-450nm) 3 3. Description of Deep UV (220nm-280nm) and Mid UV (280nm-310nm) Lightsource Systems 3

ACROBAT SWING STAND MODEL

EDGEMERE DUAL CONTROL WIDESPREAD LAVATORY FAUCET INSTALLATION INSTRUCTIONS

LED Pinspot 2. Snapshot. User Manual

A B 0 0 C D E 6 7 G F F H 8 9 K M O O L N I J 1

Introduction. The words WARNING, CAUTION and NOTE carry special meanings and should be carefully reviewed.

6.0 SPECIFICATIONS CONTENTS. Calibration. According to factory procedureeeeeeeeeeeeeee Accuracy*

Light Source User Guide

INSTALLATION INSTRUCTIONS PATIENCE

Operation and care of the Visual Optics Fiber Scopes

INSTRUCTIONS EVIS EXERA XENON LIGHT SOURCE OLYMPUS CLV-160

USER S MANUAL SILT LAMP SL 880

THIS GUIDE IS INTENDED FOR DEALERS AND SOLAR COMFORT TECHNICIANS ONLY AND IS NOT MEANT OR INTENDED TO BE REPRODUCED OR DISTRIBUTED TO THE CONSUMER

Twin-Beam Spotlight 3,000,000 Power Series Corded Cordless Rechargeable

Installation Instructions

evo 4 00 Slit Lamp User s Guide

STARSYS MOBILE POWER CART INSTRUCTIONS FOR USE

Instruction Manual. Labnet Spectrafuge 16M Microcentrifuge

Nimbo Lightweight Posterior Posture Walker

User Manual. Snapshot. Use on Dimmer. Outdoor Use. Sound Activated DMX 512. Master/Slave. 115/230V Power Switch. Replaceable Fuse.

INSTRUCTION MANUAL WARNING

Troubleshooting Guide: 355 Lights (12V)

Maintenance Adjustments

INSTALLATION INSTRUCTIONS FOR THE 45RM1, 45RM2, 45RM4, 45FS1, 45UVFS1 FIBER OPTIC FLAME SCANNERS

shadow focus floodlight SFS-300 INSTRUCTIONS

HTR-202 INSTRUCTION MANUAL. Portable Digital Height Rod. Befour, Inc. 102 Progress Drive Saukville, WI 53080

User Manual SPOTTING SCOPE X 60

OBY-5. User manual MOVING HEAD CAUTION!

Heavy Duty Series (HDS) Installation and user guide

Installation Instructions

AvaLight-DHS Deuterium Halogen Light Source Operating manual

NOTE: Visit our website at for video repair procedures, under the Tools section.

SERVICE MANUAL US. Permobil M300/M400. Power Wheelchair

INSTALLATION INSTRUCTIONS FOR THE PHOENIX FIBER OPTIC FLAME SCANNERS

VECTRIX VX-2 SERVICE MANUAL. Version 1.0/May 2011 VECTRIX, LLC

Installation Instructions Series

Microscope light manual

Connector Systems Inc. SS-20 MACHINE MANUAL

MODEL SCA Installation and Operation Manual Important:

HEAVY DUTY BATTERY BOOSTERS / CHARGERS. MODEL Nos. BC185N & BC205N OPERATING INSTRUCTIONS

OLYMPIAN MODEL 740 Operation and Service Manual

Spotlight 1,000,000 Power Series Cordless Rechargeable

SERIES OPERATION AND MAINTENANCE MANUAL

FL 150 XT-2 B, F, and C

Register your instrument! HeatSealer S100. Operating manual

INSTALLATION INSTRUCTIONS FIBER OPTICS for INSIGHT II FLAME SCANNERS

Section 10: Engraving Machine Cleaning

Transcription:

R Prima OPH User Manual Ophthalmic Microscopy To ensure proper use of this instrument as well as to avoid injury while operating instrument, understanding this manual completely before use is highly recommended.

CONTENTS 1 INTRODUCTION 1 2 SAFETY INSTRUCTIONS 2 3 SPECIAL INSTRUCTIONS 3 4 UNPACKING 4 5 DESCRIPTION OF PARTS 5 6 SYSTEM DIAGRAM 6 7 INSTRUCTIONS 7-10 8 ELECTRICAL CONNECTIONS 10 9 CONTROLS 11 10 INSTRUCTIONS FOR USING THE MICROSCOPE 12 10.1 SETTING UP MICROSCOPE 12 10.2 SETTING UP MAGNIFICATION 12 11 HOW TO FOCUS THE OBJECT 13 12 CHANGING THE OBJECTIVES AND EYEPIECES 13 13 ADJUSTMENT OF TENSION WHILE USING ACCESSORIES 13 14 DISINFECTION AND STERILIZATION 13 15 CLEANING AND SERVICING 14 16 TROUBLE SHOOTING TABLE 15-16 17 TECHNICAL SPECIFICATION 17-18

1 INTRODUCTION The LABOMED Prima OPH is an Ophthalmic microscope, which is adaptable for different surgical needs without compromise to performance. The microscope provides extremely high optical image quality, good depth of focus and wide field of view for precise surgery. Illumination control, inbuilt tilt, adjustment of the observation head help to reduce the surgeon s work fatigue and allow comfortable use over long period. Salient features of this Microscope are: 1. The observation head can easily be positioned with the help of suspension arm. 2. An advanced 5-step magnification changer allows an optimal magnification for a particular surgery from five different magnifications. 3. Auto home position for X-Y movement can be done by touch of a button. 4. Cold light illumination with a high intensity 150W Halogen lamp is provided using a fiber optic guide for proper illumination. The illumination is further adjustable up to its most suitable brightness using intensity control knob suitably located at the suspension arm, and is easily approachable to the surgeon. 5. The simple lamp changing mechanism allows to change lamp quickly when the lamp in use blows during operation. 6. The power supply to the lamp automatically goes OFF when the suspension arm is raised to its top most position. 7. When the microscope is not in use, the suspension arm can be folded over the main body to store it compactly. 8. Rigid H-form base with castor wheels provides greater stability as well as mobility to the instrument. 1

2 SAFETY INSTRUCTIONS Prima OPH 1. This microscope is manufactured according to the safety norms as per CE regulation and FDA. 2. This microscope is intended for use only as prescribed in this manual. 3. Servicing and repairs are allowed through authorized persons only. 4. Replace burnt out fuses by live fuses of the same type only (rated voltage, rated current, switch off characteristics.) 5. Use mains plug and mains socket both with protective earth conductors only. 6. Do not use force while fixing cable connections. If the male and female parts do not readily connect, make sure that they are appropriate for one another. If any of the connectors are damaged. Please contact the representative. 7. Make sure that inlets and outlets of ventilation system for cooling the housing are open (not covered). 8. The microscope is built for use in dry rooms only. Take care that no fluids enter the microscope components. Do not place any fluid-filled container on top of instrument. 9. Microscope is protected against overheating through a thermal cut-off. 10. The manufacturer will not accept any liability for damage caused by unauthorized persons tampering with the instrument; this will also forfeit any rights to claim warranty. 11. It is recommended to use the instrument only with the accessories supplied. In case you wish to use other accessory, make sure that LABOMED has certified that its use will not impair the safety of instrument. Very important : For identification, service people must know the serial number of the microscope. Intended Use: The Prima OPH is a universal instrument for operative and diagnostic medical use. It is useful for optimal illumination and magnification of the attendance area with the possibility of: 1. Manual variations of magnification 2. Motorized focusing (Z-axis) through foot control 3. Motorized centering of attendance area in XY directions through a Joystick provided on the foot control (applicable to the suitable model) 2

3 SPECIAL INSTRUCTIONS 3.1 Before every use and after re-equipping the instrument 1. Check all Electrical connections. 2. Attach sterilized covers, panels or caps which have been re-moved or opened. 3. Pay special attention to labels on the instrument, such as caution label, warning triangles with exclamation marks or notes. 4. Do not cover nay ventilation openings. 3.2 When instrument is in use 1. Avoid looking directly into the light source, microscope objectives lens or light guide. 2. When the illumination is on, the light guide must be connected at both ends. Otherwise there is a risk of fire or burn injuries. 3. Any kind of radiations has a detrimental effect on biological tissue. This also applies to illuminating the surgical field, therefore adjust the brightness and duration of illumination on the surgical field to the absolute minimum required. 4. Adjust tension of suspension arm as per convenience. 3.3 After every use of the instrument 1. Switch off the Mains sullpy to the instrument. 2. When the microscope is not in use, the suspension arm can be folded over the main body for compact storage. 3

4 UNPACKING Prima OPH The appliance is delivered in sub-assembled groups along with one set of Installation Kit and one instruction / service manual. Please check following at the time of unpacking 1. Mobile supporting base with castor wheels, or as ordered 2. Column, as ordered 3. Illumination Box 4. Swivel arm and Suspension arm assembly 5. Cover for swivel arm 6. X-Y motorized coupling; if ordered 7. Z-axis motorized focusing attachment with magnichanger assembly 8. Foot controls, (with Joystick in case of XY model) 9. Observation Head 10. Paired Eye Pieces; as ordered 11. Power Cord 12. Set of sterilizable caps 13. Installation Kit a) Allen Wrench 3.00mm b) Allen Wrench 4.00mm c) Allen Wrench 5.00mm d) Allen Wrench 6.00mm e) Allen Wrench 8.00mm f) Spanner-12/ 13mm 12. Instruction cum Operating Manual / Service Manual 4

5 DESCRIPTION OF PARTS 14 13 12 11 9 15 18 17 23 10 7 8 5 1 2 19 3 14 13 12 11 9 21 20 15 16 18 17 23 10 6 7 8 1 5 2 19 3 20 21 1. Observation Head 2. Eyepieces 3. CMO - Common Main Objective 4. Fibre Optic Cable 5. X-Y Coupling 13. Swivel Arm 14. Swivel Arm Locking Knob 15. Illumination Box 16. Lamp Changing Knob 17. On/Off switch for Foot Control 6. Z axis coupling 18. Connectors 7. Intensity Control Knob 19. Pillar 8. Coupling Locking Knob 20. Moveable Base 9. Up/Down Movement Control Knob 21. Castor Wheel with Lock 10. Cable Cover 22. Foot Control with Joystick 11. Suspension Arm 23. Tension Adjustment 12. Suspension Arm Locking Knob 5

6 SYSTEM DIAGRAM A616500-811 Ad- 613400-820 613800-815 612086-600 1271094 6122012 6122016 6122020 A612086-400 A6122015-800 6136000-900 6135000-800 6134105-102 6134105 6134100-800 6133176-00 6133201-000 6133251-000 6135000-809 6135000-909 6137300 6137500 6137200 6165000-817 6137100 6

7 INSTRUCTIONS Prima OPH 7.1 Pull out the base from the wooden pallet by removing two nuts with the help of spanner provided. 7.2 Remove allen screws from bottom side of the pillar. 7.3 Put the pillar on the base and align it with the pillar guide pin (A). After aligning the screw holes with the guiding holes on base, fix the screws removed in Fig 2 using allen wrench 8mm. Fig.2 and 3. Fig. 2 7.4 Remove safety cap from topside of the pillar. 7.5 Place swivel arm on the pillar by sliding in the groove and rotate the suspension arm by 90 as shown in Fig.4. Fig. 3 Fig. 4 7

Prima OPH 7.6 Firmly tight the safety lock screw provided on the suspension arm by using allen wrench 3mm. See fig 5. 7.7 Remove four allen screws from the illumination box before mounting it on pillar. 7.8 Fix the illumination box on the pillar only after having the connectors passed through the hole provided on the H- block. Refer to fig 6. Fig. 5 7.9 Secure the illumination box with four allen screws provided using wrench 5mm. See fig 7. 7.10 Connect two appropriate male female connectors provided for electrical connection. See fig 8. Do not pull the cables too much to join the connector s. Additional force can damage wire connections. Fig. 6 A B Fig. 7 A Fig. 8 8

Prima OPH 7.11 Install the microscope coupling to the suspension arm prior to installing the head and eyepieces. Before installing, loosen the coupling locking knob (A) and safety position screw (B).Slide in the carrier shaft to the suspension arm and screw it with locking nut. Secure the locking knob (A) and safety position screw (C). B A C 7.12 Remove the caution label from the suspension arm. Unscrew safety nut (A) from the carrier and guide the shaft smoothly into the bush of suspension arm as shown in Fig. 9 Fig. 9 B A 7.13 Remove the cable cover by unscrewing two allen screws provided at the lower side. Run the fibre optic cable and the cable meant for intensity control knob and XYZ control through the tray. 7.14 Set the cable cover on the suspension arm by aligning the guide screw with the groove provided on the cover by using allen wrench 3mm.(Fig. 12) Fig. 10 7.15 Remove the swivel arm locking knobs A & B. Remove the self tapping screws fixed on the top and bottom side of cover for swivel arm. Fix these panels on swivel arm and fix back the screws removed earlier. Make sure that wire for Fibre optic is passed through while fixing the panels. See fig 11 and 12.. Fig. 11 A B 9 Fig. 12

Prima OPH 7.19 Assemble fibre optic cable into the hole provided at the backside of magnichanger assembly by gently pushing it in, as shown in fig 13. A click sound will confirm correct position. 7.20 Plug in the wire for foot control into the illumination box. See fig 14. Fig. 13 Fig. 14 8 Electrical Connections 8.1 Adjusting mains Voltage: Connect the power cable to the AC inlet socket provided on the illumination box and check the voltage settings at the voltage selector switch. The line voltage of the electrical system is set in the factory the rated line voltage of the country of destination which must be either 110V or 220V AC. The line voltage available at the site of installation must lie within the admissible voltage range. If this is not the case you must not operate the system. 8.2 Connections of Various Controls: 8.2.1 Connect 7 pin connector from X-Y coupling assembly to terminal of electrical box. Fig. 15 8.2.2 Connect 4 pin connector from Z-axis motorized control assembly to terminal of electrical box. 8.2.3 Connect 12-pin connector from foot control to terminal of electrical box. This completes your electrical connections. 10

Prima OPH 9 Controls 9.1 ON/OFF Switch 9.1 ON/OFF switch It is located on the Right hand side of illumination box 1 indicates ON position of microscope. At ON position green LED glows and cooling fan starts running. Keep the intensity control knob at minimum level before changing / replacing the lamp. To save burning life of lamp, switch OFF the appliance if the microscope is not in use for longer time. 9.2 Lamp changing knob It is located on the right hand side of illumination box. If lamp blows off during the use of microscope, the Prima OPH is provided with a spare lamp. Lamp changing knob connects to the Lamp II for smooth running of the operation by turning it anti-clockwise. 9.3 Intensity control knob It is located on the suspension arm. Brightness of field of view can be adjusted as per user comfort using intensity control knob. 9.4 Brakes Locks the stand from unwanted movement by pressing down the two brakes provided on caster wheels. To unlock press upper portion of brake. See Figure 9. 9.5 Swivel arm locking knob This knob helps you to lock the movement of swivel arm at the desired position after initial focusing of the attendance area by turning it clockwise. 9.6 Suspension Arm locking knob This knob locks the movement of suspension arm at any desired position by turning it clockwise. 9.7 Coupling locking knob This knob is located on the suspension arm and helps to lock the position of observation head at any desired point by turning it clockwise. 9.8 Up / Down Movement Control Knob Locks the UP/Down movement of Suspension arm at a desired position after initial focusing of the attendance area. Fine focusing is done using F.O.V. 3 Fig. 16 Fig. 17 1 2 11

10 INSTRUCTIONS FOR USING THE MICROSCOPE 10.1 Setting up of Microscope: 10.1.1Lock all the brakes on base wheels after setting up of microscope on the attendance area for stability. 10.1.2 Adjust tension on suspension arm using tension adjustment screw as per your convenience by turning the knob clock wise or anti-clock wise. 10.1.3 Lock the Up & Down movement of suspension arm using locking knob after coarse focusing of the attendance area. 10.1.4 Adjust the eye distance as per IPD scale according to your convenience. Fig. 18 10.2 Setting up of magnification (Ref. 18) 1. Adjust to highest magnification with one of the rotating knobs provided at magnification changer. 2. Fine focusing is done through F.O.V. knob 3. Absolute centering of observation area in field of view can be done by manual handle(21c) 4. Make sure that the magnification changer is engaged in the index point at the click stop position. 5. Use filter changing knob (18d) to change filters. 12

11 How to focus the object 12 1. Adjust both the eyepieces to 0 diopter adjustment. 2. Adjust IPD of the observation head using IPD scale. 3. Bring highest magnification factor in the click stop position using one of the knobs provided at the magnichanger unit. By doing so observation area will remain parfocalized in all magnifications. 4. Fine focusing is done by using foot control by pressing on up and down paddles. Changing the objectives/ eyepieces 1. The objectives can be taken out by rotating it in anticlockwise direction. It can be threated in by rotating in clockwise direction. 2. To install the Eyepieces, insert in the Eyetubes of observation head. 3. A range of objectives/eyepieces can be selected by choice. Fig. 19 13 Adjustment of Tension while using Accessories After Supplementary accessories are mounted, the additional load of suspension arm must be compensated by adjusting tension on tension control screw provided on suspension arm by moving it clock wise or anticlockwise. Fig. 20 14 Disinfection and Sterilization For Diagnosis: Fig. 21 13 Moisten smooth cotton with antiseptic fluid (for example Sagrotan P); when required, clean often touched parts, like rotating knobs, handles and so on. After surgery:- Sterilizable polymer covers are provided on every part that require to be touched during operation. Sterilize them after every use of the instrument.

15 CLEANING AND SERVICING Prima ENT 15.1 Cleaning of optical surfaces: Remove coarse dirt particle with a clean dry air from optics outer surfaces (Objectives, Eyepieces). Moisten smooth cotton cloth with lens cleaning agent and wipe on the lens surface gently starting from the middle of the lens to the outer edge. 15.2 Cleaning of mechanical surfaces: All mechanical surfaces of the equipment can be cleaned by wiping with a moist cloth. Don t use any aggressive or abrasive cleaning agents. Any household dish washing fluid can be used for cleaning residue. 15.3 Servicing: Service whenever required, inform after-sale service. 14

16 TROUBLESHOOTING TABLE Prima OPH Problem Possible Cause Remedy No Illumination at all Power cable not plugged Plug in power cable. Power switch not pressed Defective instruments fuse Press power switch Change instrument fuse Defective power cable Line power failure Failure of suspension system electronics Light guide not properly inserted in lamp or microscope. Change power cable Contact in-house Techician Contact service dept. Insert light guide to the maximum intensity. Insufficient Brightness level set too Adjust brightness using the Illumination low brightness control knob. Light guide not properly inserted in lamp or microscope. Defective light guide (illumination not uniform) Insert light guide to the maximum illumination Chance light guide. Halogen lamp in the Knob for brightness control Turn the knob clock wise illumination system not operated remains dark, and fan is running. Lamp module has no Insert lamp module properly. contact. Defective halogen lamp. Switch to backup lamp. Halogen lamp goes constantly off and on Ventillation slots are Ventillation slots must be clear, during operation. covered or contaminated clean them if necessary. 15

Prima OPH Problem Possible Cause Remedy Thermal cut-off in lamp housing is contaminated. Clean thermal cut-off with a dry brush; blow it clean, if necessary Defective fan. Failure of system electronics. Contact service dept. Illuminate surgical field using an Or illuminator. Contact service dept. Up & Down motion Friction adjustment screw Loosen friction adjustment of screw on on suspension system screw on suspension system suspension system tightened too firmly. as require too stiff Microscope Unstable Brakes on wheels not Use Brakes. used. No image visible in Magnichanger is not Index magnichanger properly. field of view. indexed properly. 16

17 SPECIFICATIONS Prima OPH Technical Data (Specifications) Binocular tubes Eyepieces 45 inclined WF 10x/16mm with eye guards; optional WF 12.5x/16mm; WF 16x/16mm Apochromatic Magnichanger 0.4x, 0.6x, 1.0x, 1.6x & 2.5x Objective Light source Built-in filters Vertical movement of arm Microscope carriers f=200, motorized fine focus 15V/150W halogen, twin lamp Green & Blue 600 mm XY coupling or basic Ophthalmic carrier (motorized fine focus) Accessories Type: Base (Dimensions): Assistant Binocular attachment, ivu OP Beam splitter integrated digital camera Module; Double Beam splitter; 0-210 inclinable ergo tube Floor stand 775mm width 762mm length Stand Height: Weight of complete microscope: Elevation Stroke: Stand Height in Horizontal Position: Field of illumination with F.O.V. f=200mm objective: 1676mm 72 Kg. Approx. 600mm 1100mm Ø 70mm 17

Dimensions 18

www.laboamerica.com Our policy is one of continuous development. Labo America, Inc., reserves the right to change design and specifications without prior notice. Labo America Inc. 920 Auburn Court Fremont CA 94538 U.S.A. Telephone: 510 445 1257 Fax: 510 445 1317 sales@laboamerica.com LABOMED and Prima OPH are registered trademarks of Labo America, Inc. With a policy of continuous development, Labo America, Inc. reserves the right to change design and specifications without prior notice. 2010 Labo America, Inc. 6135000-990A 02-2010 ISO 9001 : 2008 File No. A9020