Addendum to the Waters Micromass Q-Tof Premier Mass Spectrometer Operator s Guide

Similar documents
StepWave ion guide: cleaning instructions addendum

Waters SQ Detector Operator s Guide

DART-Waters XEVO Interface Manual

Passive Check Valve Installation Instructions for H-Class Systems / Revision A Copyright Waters Corporation 2014 All rights reserved

Mariner Biospectrometry Workstation

Waters Xevo TQ-S Mass Spectrometry System Operator s Overview and Maintenance Guide

ACQUITY QDa Detector Overview and Maintenance Guide

DART-AB Sciex Interface Manual

HP 1946A LC/MSD Replacement Parts

Waters Xevo TQ-S micro Overview and Maintenance Guide

Waters Xevo TQ-S IVD Mass Spectrometry System Operator s Overview and Maintenance Guide

Captivate Illuminated Mirror Light

Installation, Operating & Maintenance Instructions. Series 081, DN (I.D. 2½" - 4") Hot surfaces; do not touch!

SI-200-GIST. DART AB Sciex Interface Manual User Manual. IonSense Inc. 999 Broadway Suite 404 Saugus, MA Manual Revision 2 Document # 7.5.

Specification Grade Contemporary Wall Mount/Technical Installation Data

[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS

HV angle valve with single acting pneumatic actuator and closing spring (NC)

Control Gate Valve with stepper motor actuator

Cable Cubby 1202 and 1402 Installation Guide

INSTRUCTION MANUAL WARNING

Installation, Operating & Maintenance Instructions. Hot surfaces; do not touch!

PTEX extruder pumps Pneumatic extruder pump for drums of grease up to NLGI grade 2

RS-110 Rainfall Sensor Installation Guide

Waters Xevo TQ-XS Mass Spectrometry System

INSTALLATION INSTRUCTIONS LED RETROFIT ASSEMBLY (LRA) Rev F

HV-Angle Valve (Series 264) HV-InlineValve (Series 265) with electromagnetic actuator single acting with closing spring (NC)

Filter Holder Manifold User Guide. 3- and 6-Place Stainless Steel Manifolds 3- and 6-Place Sterility Test Manifolds 3-Place PVC Manifold

IMPORTANT SAFEGUARDS When using electrical equipment, basic safety precautions should always be followed including the following:

HV Gate Valve with pneumatic actuator

Fuel Pump ( ) Table of Contents. Summary. View Related Topic. Summary. General Information Test. Preparatory Steps. Remove.

Consumables and Supplies Required for Repair

Installation, Operating & Maintenance Instructions. HV angle valve with pneumatic actuator single acting with closing spring (NC)

SAFETY TRAINING LEAFLET 04 NITROUS OXIDE

11/30/2017 # F: Engine Oil Leak at Crankshaft Rear Main Oil Seal (Install Revised Crankshaft Rear Main Oil Seal Using Revised Rear Main S

Tensioning Winder 30. CS9160 Eco Solvent Printers

L DURAMAX LBZ EGR DELETE GDP TUNING

Atomic Absorption Spectroscopy

Model 1100 Turbine Flow Meter Repair Kits

881 Compact IC pro. 881 Compact IC pro Anion MCS. Manual EN

HV Angle valve (Series 264) HV Inline valve (Series 265) with pneumatic actuator, single acting with closing spring (NC)

Rapid Digestor Fume Removal System

UHV Angle valve (Series 284) with pneumatic actuator, single acting with closing spring (NC)

ASSEMBLED BY HAND IN THE USA

LA1, LNJ, LG8, LL1, LX9, L82, LL2, LB8, LHO, LG5, LB6, LE2, LQ1, LH7, LC1, L44, LZ4, LZE, LZ9, LGD, LZ8, LZG)

Cable Cubby 650 UT Installation Guide

Lamp House with 2.5 kw lamp

FastPrep FP120 Cell Disrupter INSTRUCTION MANUAL

READ AND FOLLOW ALL SAFETY INSTRUCTIONS 1. DANGER RISK OF SHOCK DISCONNECT POWER BEFORE INSTALLATION

Agilent 210/218 - Isocratic Pump

Alliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE

Document ID# Cadillac CTS

CONGRATULATIONS. on your purchase of a Viridian Grip Laser, the most advanced sighting device available.

FITTING INSTRUCTIONS FOR CP0404BL/WH NO-CUT AERO CRASH PROTECTORS HUSQVARNA 701 ENDURO/SUPERMOTO 2016-

Installation Instructions Studio Makeup Station

850 Professional IC. AnCat MCS Manual EN

Agilent 1200 Series Manual Injector

Liquid Chromatography Mass Spectrometry LCMS Consumables SSI-LCMS

OmniAire Omnitec Design, Manufacturing Air Purification Equipment Since 1988

Dynojet Research, Inc. All Rights Reserved. Air Fuel Ratio Module Installation and User Guide.

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER AUTOMATIC INSERTION PROBE

Pellicon 2 and 3 Mini Holder. User Guide

SSP-Series Portable Balances Operation Manual

Shock Absorber Rebuild Manual

Rectangular Door L-VAT with pneumatic actuator double acting

USER S GUIDE 2880 SERIES. 3 Safety Instructions. Product View. Assembly 4-5. Operation. Maintenance and Care. Troubleshooting.

Installing a PTV Inlet

TECHNICAL INSTRUCTIONS

Installing an H-200 Integral Right-Angle Module

Agilent G5632 Valve Kit Instructions

260 Solvent Vapor Exit Accessory

G1160A Agilent 1100 Series Multiple Purpose Switching Valve

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard

Preventive Maintenance...

Nota Conductor s Chair

Protector Glove Boxes.

1260 Infinity II Degasser User Manual. 5 Maintenance

COSASCO HYDRAULIC ACCESS FITTING User Manual

Verax 75 and Verax VPA Installation Manual

P5514, P5514-V & P5514-EP Hydraulic Comparison Test Pump

Register your instrument! HeatSealer S100. Operating manual

Sterifil Aseptic System and Sterifil 47 mm Filter Holder User Guide

KD1E Range Potential Indicators

P5510. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

OPERATIONS MANUAL CR Thermoreactor

6890 Installing ¼ Glass Columns

PrioVino Premier. Translation of Original Operating Instructions. Status: August First edition January 2018 / PrioVino GmbH

FITTING INSTRUCTIONS FOR CP0393BL AERO CRASH PROTECTORS SUZUKI GSX-S

Installation, Operating & Maintenance Instructions. HV gate valve with pneumatic actuator. Series 110 DN mm (I. D.

Agilent 210/218 - Binary Pump

Installing a PTV Inlet

UB-2 2 Universal Button

PISTON ACCUMULATOR. Installation and operation manual Revision

NUMBER: S.M. REF.: Listed in Table ENGINE: EPA04 MBE 4000 DATE: April 2009

Waters GCT Premier Mass Spectrometer

P5515, P5515-V, & P5515-EP Hydraulic Comparison Test Pump

SUPER PUMP OUT SYSTEM

Models P, P, P, , DANGER

Fuel Meter. Instructions - Parts List. Contents B. Model , measures in gallons Model , measures in liters. For metering fuels.

OTTO HYDRAULICS GmbH & Co KG SYSTEME UND KOMPONENTEN FÜR ÖLHYDRAULISCHE ANTRIEBE UND STEUERUNGEN SIEMENSSTRASSE 12/ FRIEDRICHSHAFEN

CLIF MOCK TM. Model 5030 Orifice Fitting. User Manual. Manual No , Rev. B

Transcription:

Addendum to the Waters Micromass Q-Tof Premier Mass Spectrometer Operator s Guide 715001174/Revision A Copyright Waters Corporation 2006. All rights reserved.

Copyright notice 2006 WATERS CORPORATION. PRINTED IN THE UNITED STATES OF AMERICA AND IRELAND. ALL RIGHTS RESERVED. THIS DOCUMENT OR PARTS THEREOF MAY NOT BE REPRODUCED IN ANY FORM WITHOUT THE WRITTEN PERMISSION OF THE PUBLISHER. The information in this document is subject to change without notice and should not be construed as a commitment by Waters Corporation. Waters Corporation assumes no responsibility for any errors that may appear in this document. This document is believed to be complete and accurate at the time of publication. In no event shall Waters Corporation be liable for incidental or consequential damages in connection with, or arising from, its use. Waters Corporation 34 Maple Street Milford, MA 01757 USA Trademarks Micromass and Waters are registered trademarks, and MassLynx and Q-Tof Premier are trademarks of Waters Corporation. Other trademarks or registered trademarks are the sole property of their respective owners. Customer comments Please contact us if you have questions, suggestions for improvements, or find errors in this document. Your comments will help us improve the quality, accuracy, and organization of our documentation. You can reach us at tech_comm@waters.com.

Addendum to the Waters Micromass Q-Tof Premier Mass Spectrometer Operator s Guide This addendum details changes that have been made to the Waters Micromass Q-Tof Premier Mass Spectrometer Operator s Guide (71500089402 Revision B). Contents: Topic Page Safety Information 2 Chapter 1 Instrument Description 3 Chapter 7 APCI, APPI and ESCi 5 Chapter 8 Preventative Maintenance 7 Appendix D Materials of Construction and Compliant Solvents 25 1

Safety Information Add the following sub-sections to this section: Solvent Leakage Hazard The system has been designed to be robust and leak-tight. Waters recommends that you perform a hazard analysis, assuming a maximum leak into the laboratory atmosphere of 10% LC eluent. The items identified in Appendix D Materials of Construction and Compliant Solvents may be exposed to solvent; you must evaluate the safety issues involved if the solvents used in your application differ from the solvents normally used with these items. Risque de fuite de solvant Warning: To confirm the integrity of the source exhaust system, the items identified in Chapter 8, in sections Replacing the LockSpray modular source enclosure, probe adjustment flange, and source option O-rings and Replacing the APPI lamp drive assembly O-rings, must be renewed at intervals not exceeding one year. Avertissement: Afin de vérifier l'intégrité du système d'évacuation source, les éléments identifiés dans le chapitre 8, dans les sections Replacing the LockSpray modular source enclosure, probe adjustment flange, and source option O-rings et Replacing the APPI lamp drive assembly O-rings, doivent être renouvelés à des intervalles inférieurs à un an. Le système a été conçu pour être robuste et étanche. Waters recommande que vous exécutiez une analyse de risque, en considérant une fuite maximum de 10% de l'éluant chromatographique dans l'atmosphère du laboratoire. Les articles identifiés dans l Appendice D Materials of Construction and Compliant Solvents peuvent être mis au contact de solvants; si les solvants utilisés dans votre application diffèrent des solvants normalement employés avec ces articles, vous devez évaluer les problèmes de sécurité liés à leur utilisation. 2 Addendum to the Waters Micromass Q-Tof Premier Mass Spectrometer Operator s Guide

Chapter 1 Instrument Description Rear panel Gas connections Delete the existing Gas connections on rear panel figure and replace with the following: Gas connections on rear panel: Collision gas Nitrogen Source waste Syringe waste Vacuum connections to scroll pump Rear panel 3

Delete the Exhausts, Source waste and LC waste and Connecting the waste bottle sections and replace with the following: Source waste Warning: LC solvents and analytes may be carried in the nitrogen exhaust, which must be vented via the nitrogen exhaust waste bottle and laboratory exhaust system. The laboratory exhaust system must provide a minimum vacuum of 2 millibar below atmospheric pressure (negative pressure). Warning: To avoid the build-up of hazardous gases, do not place the nitrogen exhaust waste bottle in an enclosed cabinet. This connection carries the nitrogen gas from the source; it is connected to a nitrogen exhaust waste bottle which, in turn, is connected to the laboratory exhaust system. The nitrogen exhaust waste bottle must be located in an area where it is visible, so that you can monitor and empty it, and then perform a leak test on it, at regular intervals. See also: Emptying the nitrogen exhaust waste bottle. Nitrogen exhaust waste bottle: To laboratory exhaust port From instrument source waste connection 4 Chapter 1 Instrument Description

Syringe waste Liquid spilled into the syringe pump drains from this connection. Chapter 7 APCI, APPI and ESCi IonSabre APCI operation Installing IonSABRE LockSpray To install IonSABRE LockSpray: Add the following before step 6: Caution: To avoid damage, do not apply any force to the source enclosure door while the door is open. Add the following before step 10: Caution: When refitting the LockSpray motor assembly, the securing thumbscrews must each be sequentially tightened a small amount until they are all fully tight; this ensures that the LockSpray motor assembly is uniformly seated on the source enclosure. IonSabre APCI operation 5

APPI and dual APPI operation Delete the above title and replace with: APPI and dual APPI/APCI operation To install APPI/dual APPI: Delete the above title and replace with: To install APPI/dual APPI/APCI: Add the following before step 6: Caution: To avoid damage, do not apply any force to the source enclosure door while the door is open. Add the following before step 10: Caution: The securing thumbscrews must each be sequentially tightened a small amount until they are all fully tight; this ensures that the APPI lamp drive assembly is uniformly seated on the source enclosure. ESCi operation Add the following before step 3: Caution: To avoid damage, do not apply any force to the source enclosure door while the door is open. Add the following before step 8: Caution: When refitting the LockSpray motor assembly, the securing thumbscrews must each be sequentially tightened a small amount until they are all fully tight; this ensures that the LockSpray motor assembly is uniformly seated on the source enclosure. 6 Chapter 7 APCI, APPI and ESCi

Chapter 8 Preventative Maintenance Maintenance schedule Add the following to the Maintenance schedule table: Procedure Replace the LockSpray modular source enclosure, probe adjustment flange, and source option O-rings. Replace the APPI lamp drive assembly O-rings. Empty the nitrogen exhaust bottle. Frequency Annually. Annually. As required. Add the following section after the Safety and handling section: Replacing the LockSpray modular source enclosure, probe adjustment flange, and source option O-rings Warning: To confirm the integrity of the source exhaust system, the O-rings listed below must be renewed at intervals not exceeding one year. To confirm the integrity of the source exhaust system, the following O-rings must be renewed at intervals not exceeding one year: Source enclosure door O-ring Source enclosure door glass O-ring Source enclosure housing O-ring Source enclosure side flange O-ring Probe adjustment flange O-ring Note: To complete this procedure, you will be required to perform a pressure test on the source, as described in the Waters Micromass Source Pressure Test Unit Operator s Guide. Maintenance schedule 7

To remove the source enclosure from the instrument: Warning: The source components may be contaminated with biohazardous and/or toxic materials. Always wear nitrile gloves while performing this procedure. Warning: The source may be hot; to avoid burns, take great care while working with this component. 1. Remove the probe from the source. See also: To remove the probe from the source: on page 8-6. Caution: To avoid damage, do not apply any force to the source enclosure door while it is open. 2. Unfasten the source enclosure door s securing clips and open the door. 3. Remove the currently-fitted modular source option or blank side flange from the source enclosure. See also: To change the source configuration: on page 1-12, steps 1 to 4. 4. Remove the source enclosure from the instrument. See also: To remove the ion source and ion block: on page 8-9, steps 1 to 4. 8 Chapter 8 Preventative Maintenance

To remove the source enclosure and probe adjustment flange O-rings: Warning: The source components may be contaminated with biohazardous and/or toxic materials. Always wear nitrile gloves while performing this procedure. Source enclosure door: Source enclosure door glass O-ring Door glass retaining clip Source enclosure door O-ring Caution: To avoid damage, do not use a metal tool to remove any of the O-rings. 1. Carefully remove the source enclosure door O-ring from the source enclosure door. 2. Use a hex (Allen) key to remove the four bolts securing the source enclosure door glass retaining clips to the source enclosure door. 3. Remove the four source enclosure door glass retaining clips from the source enclosure door. 4. Remove the source enclosure door glass from the source enclosure door. 5. Carefully remove the source enclosure door glass O-ring from the source enclosure door. Maintenance schedule 9

Probe adjustment flange, and source enclosure side flange and housing O-rings: Probe adjustment flange O-ring Source enclosure housing O-ring Source enclosure side flange O-ring 6. Carefully remove the probe adjustment flange O-ring from the probe adjustment flange. 7. Carefully remove the source enclosure side flange O-ring. 8. Carefully remove the source enclosure housing O-ring. Warning: The O-rings may be contaminated with biohazardous and/or toxic materials. Ensure that they are correctly disposed of according to local environmental regulations. 9. Dispose of the O-rings in accordance with local environmental regulations. 10 Chapter 8 Preventative Maintenance

To fit the replacement source enclosure and probe adjustment flange O-rings: Warning: The source components may be contaminated with biohazardous and/or toxic materials. Always wear nitrile gloves while performing this procedure. 1. Ensure that all the grooves for the O-rings are free from dirt and hairs. Tip: If contamination is present, use an appropriate solvent, applied to a lint-free cloth, to carefully clean the grooves. 2. Fit the new source enclosure housing O-ring to the source enclosure. 3. Fit the new source enclosure side flange O-ring to the source enclosure. 4. Fit the new probe adjustment flange O-ring to the probe adjustment flange. 5. Fit the new source enclosure door glass O-ring to the source enclosure door. 6. Fit the source enclosure door glass to the source enclosure door. 7. Fit the four source enclosure door glass retaining clips to the source enclosure door. 8. Use a hex (Allen) key to fit and tighten the four bolts securing the source enclosure door glass retaining clips to the source enclosure door. Caution: Ensure that the source enclosure door O-ring tail is correctly located in its groove when fitting the O-ring to the source enclosure door. 9. Fit the new source enclosure door O-ring to the source enclosure door. 10. Ensuring that the wires to the microswitch do not become trapped between the source enclosure and the pumping block, fit the source enclosure to the pumping block. Caution: The source enclosure and source option securing screws must each be sequentially tightened a small amount until they are all fully tight; this ensures that the source enclosure is uniformly seated on the pumping block. 11. Tighten the three captive source enclosure securing screws. 12. Fit the required modular source option or blank side flange to the source enclosure, and tighten the four source option securing screws. Maintenance schedule 11

13. Connect the appropriate gas and electrical supplies to the module. 14. Close the source enclosure door and fasten the securing clips. 15. Install the ESI or APCI probe, as required. 16. Start up the instrument. See also: Starting the Q-Tof Premier on page A-2. Warning: To confirm the integrity of the source exhaust system, a source pressure test must be performed, as described in the Waters Micromass Source Pressure Test Unit Operator s Guide. 17. Perform a source pressure test. Add the following section after the Replacing the LockSpray modular source enclosure, probe adjustment flange, and source option O-rings section: Emptying the nitrogen exhaust waste bottle Warning: The waste liquid in the nitrogen exhaust waste bottle comprises LC solvents and analytes. Always wear nitrile gloves while handling the nitrogen exhaust waste bottle, and ensure that the waste liquid is correctly disposed of according to local environmental regulations. The nitrogen exhaust waste bottle in the nitrogen exhaust line must be emptied before it is completely full. See also: Chapter 1, Source waste. 12 Chapter 8 Preventative Maintenance

Nitrogen exhaust waste bottle: To laboratory exhaust port From instrument exhaust connection To empty the nitrogen exhaust waste bottle: 1. In the MassLynx Tune window, click Press for Standby and confirm that the adjacent instrument status indicator shows red. 2. In the MassLynx Tune window, click to stop the nitrogen flow. 3. Disconnect the instrument exhaust and laboratory exhaust system lines from the nitrogen exhaust waste bottle. 4. Dispose of the waste liquid in accordance with local environmental regulations. 5. Connect the instrument exhaust and laboratory exhaust system lines to the nitrogen exhaust waste bottle. 6. In the MassLynx Tune window, click to start the nitrogen flow. 7. In the Source page, set Desolvation (l/hr) to 1200. 8. Set Cone (l/hr) to 300. Maintenance schedule 13

Warning: To confirm the integrity of the source exhaust system, the following leak test must be performed. Caution: To avoid damage to the instrument, Snoop (or equivalent) leak detector liquid must only be used only for the purpose described in the following step; it must not be used on any other part of the instrument. 9. Use Snoop (or equivalent) leak detector liquid to ensure that there are no leaks at the instrument exhaust and laboratory exhaust system line connections. Cleaning the Source components To remove the sample cone: Add the following before step 5: Caution: To avoid damage, do not apply any force to the source enclosure door when it is open. Fitting the ion block and ion source enclosure To fit the ion block and source enclosure: Delete step 5 and replace with: 5. Ensuring that the wires to the microswitch do not become trapped between the source enclosure and the pumping block, fit the source enclosure to the pumping block. Add the following before step 6: Caution: The source enclosure securing screws must each be sequentially tightened a small amount until they are all fully tight; this ensures that the source enclosure is uniformly seated on the pumping block. 14 Chapter 8 Preventative Maintenance

Fitting the sample cone: Add the following after step 11: 12. Start-up the instrument. Warning: To confirm the integrity of the source exhaust system, if any O-ring detailed in Replacing the LockSpray modular source enclosure, probe adjustment flange, and source option O-rings has been replaced, you must perform a source pressure test as described in the Waters Micromass Source Pressure Test Unit Operator s Guide. 13. If any O-ring detailed in Replacing the LockSpray modular source enclosure, probe adjustment flange, and source option O-rings has been replaced, perform a source pressure test. Replacing the ion block cartridge heater To refit the ion block: Delete step 7 and replace with: 7. Ensuring that the wires to the microswitch do not become trapped between the source enclosure and the pumping block, fit the source enclosure to the pumping block. Add the following before step 8: Caution: The securing screws must each be sequentially tightened a small amount until they are all fully tight; this ensures that the source enclosure is uniformly seated on the pumping block. Replacing the ESI probe capillary Installing the new capillary To install the new capillary: Delete step 22 and replace with: 22. Refit the probe to the instrument. Replacing the ion block cartridge heater 15

Delete step 25 and replace with: 25. Use the probe adjuster knob to adjust the capillary so that the capillary protrudes approximately 0.5 mm from the end of the probe. 26. While infusing a known compound, tune the capillary length to the position that gives the best sensitivity. Maintaining the APPI lamp Changing the lamp bulb Delete this section and replace with the following: Changing the UV lamp bulb To change the UV lamp bulb: Warning: The probe and source may be contaminated with biologically hazardous materials. Wear nitrile gloves at all times while handling the components. Warning: To avoid electric shock, ensure that the instrument is in standby while working with the instrument s front access door open. Warning: The probe and source may be hot. To avoid burns, take great care while working with the instrument s front access door open. 16 Chapter 8 Preventative Maintenance

APPI lamp drive assembly: Source enclosure side plate PEEK insulator Repeller electrode Repeller electrode retaining screw Insulator retaining screw (2-off) 1. On the MassLynx Tune window, click, and confirm that the instrument status indicator is red. 2. Wait for three minutes for the desolvation gas to cool the probe and source. 3. Click to stop the nitrogen flow. 4. Remove the APPI lamp drive assembly from the source enclosure. 5. Remove the repeller electrode retaining screw. 6. Remove the repeller electrode from the PEEK insulator. 7. Remove the cover cap from the back of the lamp drive assembly. Maintaining the APPI lamp 17

Changing the UV lamp bulb: APPI lamp bulb Nylon tube Caution: To avoid breaking the bulb, do not use a screwdriver to push the bulb out of the lamp drive assembly. 8. Insert an appropriate tool (for example, a length of 4-mm nylon tube) through the back of the lamp drive assembly, and push the bulb forward. 9. Withdraw the bulb from the lamp drive assembly. 10. Insert the new bulb into the lamp drive assembly. 11. Refit the cover cap to the lamp drive assembly. 12. Fit the repeller electrode to the PEEK insulator. Cover cap (removed from the lamp drive assembly) 13. Fit and tighten the repeller electrode retaining screw. 14. Fully retract the lamp s mounting shaft in the APPI lamp drive assembly. 18 Chapter 8 Preventative Maintenance

15. Fit the APPI lamp drive assembly to the source enclosure and perform a source pressure test. Cleaning the lamp window The transmission of the high-energy photons responsible for APPI relies on the cleanliness of the magnesium fluoride lamp window. The window should be cleaned to keep the surface clear of contamination and avoid reduced sensitivity. To clean the lamp window: Warning: To confirm the integrity of the source exhaust system, you must perform a pressure test on the source, as described in the Waters Micromass Source Pressure Test Unit Operator s Guide. Warning: The probe and source may be contaminated with biologically hazardous materials. Wear nitrile gloves at all times while handling the components. Warning: To avoid electric shock, ensure that the instrument is in standby while working with the instrument s front access door open. Warning: The probe and source may be hot. To avoid burns, take great care while working with the instrument s front access door open. 1. On the MassLynx Tune window, click, and confirm that the instrument status indicator is red. 2. Wait for three minutes for the desolvation gas to cool the probe and source. 3. Click to stop the nitrogen flow. Maintaining the APPI lamp 19

Caution: To avoid damage, do not apply any force to the source enclosure door while it is open. 4. Open the instrument s front access door. 5. Unfasten the source enclosure door s securing clips, and open the door. 6. Use methanol or isopropyl alcohol, applied to a lint-free cloth, to carefully clean the lamp window. 7. Close the source enclosure door, and fasten the securing clips. Replacing the APPI lamp drive assembly O-rings Warning: To confirm the integrity of the source exhaust system, the APPI lamp drive assembly O-rings listed below must be renewed at intervals not exceeding one year, as described in this section. The following APPI lamp drive assembly O-rings must be renewed at intervals not exceeding one year: Lamp bulb sealing O-ring Mounting shaft O-rings Lamp mounting flange O-ring Note: To complete this procedure, you will be required to perform a pressure test on the source, as described in the Waters Micromass Source Pressure Test Unit Operator s Guide. To remove the APPI lamp drive assembly O-rings: Warning: The probe and source may be contaminated with biologically hazardous materials. Wear nitrile gloves at all times while handling the components. Warning: To avoid electric shock, ensure that the instrument is in standby before commencing this procedure. 1. Remove the lamp bulb. See also: To change the UV lamp bulb. 20 Chapter 8 Preventative Maintenance

2. Use a 2.5-mm hex (Allen) key to remove the two insulator retaining screws. 3. Remove the PEEK insulator from the mounting shaft. Caution: To avoid damage, do not use a metal tool to remove any of the O-rings. 4. Carefully remove the lamp bulb sealing O-ring from inside the end of the mounting shaft. Lamp bulb sealing O-ring: Lamp bulb sealing O-ring 5. Extract the mounting shaft from the APPI source enclosure side plate. 6. Carefully remove the mounting shaft O-rings from inside the lamp mounting flange aperture. Maintaining the APPI lamp 21

Mounting shaft O-rings: APPI source enclosure side plate Mounting shaft O-rings Lamp mounting flange 7. Use a 2.5-mm hex (Allen) key to unscrew the four captive lamp mounting flange securing screws. Lamp mounting flange: APPI source enclosure side plate Lamp mounting flange Securing screw 8. Remove the lamp mounting flange from the APPI source enclosure side plate. 9. Carefully remove the lamp mounting flange O-ring from the APPI source enclosure side plate. 22 Chapter 8 Preventative Maintenance

Lamp mounting flange O-ring: APPI source enclosure side plate Lamp mounting flange O-ring Warning: The O-rings may be contaminated with biohazardous and/or toxic materials. Ensure that they are correctly disposed of according to local environmental regulations. 10. Dispose of the O-rings in accordance with local environmental regulations. To fit the new APPI lamp drive assembly O-rings: 1. Ensure that all the grooves for the O-rings are free from dirt and hairs. Tip: If contamination is present, use an appropriate solvent, applied to a lint-free cloth, to carefully clean the grooves. 2. Fit the new lamp mounting flange O-ring to the APPI source enclosure side plate. 3. Fit the lamp mounting flange to the APPI source enclosure side plate. 4. Use a 2.5-mm hex (Allen) key to tighten the four captive lamp mounting flange securing screws. Maintaining the APPI lamp 23

Caution: Take care not to damage the APPI lamp drive assembly mounting shaft O-rings when fitting them to the lamp mounting flange. 5. Fit the APPI lamp drive assembly mounting shaft O-rings to the lamp mounting flange. Tip: To fit each mounting shaft O-ring, first seat the O-ring in the small radius at the bottom of the groove in the source enclosure side plate, then use a suitable tool, having a circular cross-section, to roll the remainder of the O-ring in the groove. 6. Insert the mounting shaft in the lamp mounting flange. 7. Fit the lamp bulb sealing O-ring to the mounting shaft. 8. Fit the PEEK insulator to the mounting shaft. 9. Fit and tighten the two insulator retaining screws. 10. Insert the new bulb into the lamp drive assembly. 11. Refit the cover cap to the lamp drive assembly. 12. Fit the repeller electrode to the PEEK insulator. 13. Fit and tighten the repeller electrode retaining screw. 14. Fully retract the lamp s mounting shaft in the APPI lamp drive assembly. 15. Fit the APPI lamp drive assembly onto the right side of the source enclosure. Caution: The securing thumbscrews must each be sequentially tightened a small amount until they are all fully tight; this ensures that the APPI lamp drive assembly is uniformly seated on the source enclosure. 16. Fit and tighten the three securing thumbscrews. 17. Close the source enclosure door and fasten the door securing clips. 18. Connect the APPI drive cable (part number 4170075CC1), between the APPI lamp drive assembly APPI electrical connection and the APPI connector on the instrument front panel. 24 Chapter 8 Preventative Maintenance

19. Connect the HT cable (part number M955286BC1), between the APPI lamp drive assembly s HV electrical connection and the HV connector on the instrument front panel. Warning: To confirm the integrity of the source exhaust system, you must perform a pressure test on the source, as described in the Waters Micromass Source Pressure Test Unit Operator s Guide. 20. Perform a pressure test on the source. Appendix D Materials of Construction and Compliant Solvents Items exposed to solvent Add the following to the table Items exposed to solvent : Item APPI lamp drive assembly: Mounting shaft Repeller electrode Insulator Lamp window Material Stainless steel Stainless steel PEEK Magnesium fluoride Items exposed to solvent 25

26 Appendix D Materials of Construction and Compliant Solvents