OPERATION & MAINTENANCE MANUAL FOR SCEN-206 LED & MULTI-FUNCTION TYPE CENTRIFUGE PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION.

Similar documents
LCD Centrifuge MDCEN-302

Bio Lion Table Top Centrifuge XC-L5

1580R User s Manual. Multi-Purpose, High Speed Centrifuge. Gyrozen Co., Ltd.

GLO GT416. Benchtop Centrifuge. User s Manual. Glotech Co., Ltd

Operating Manual Cellspin

Ultra 5.0 Manufactured By Hanil Science Industrial Co., Ltd. In Korea

Bio Lion Table Top Centrifuge XC-H165

1236R User s Manual. Multi-Purpose, High Speed Centrifuge

624R User s Manual. Compact Bench-Top Refrigerated Centrifuge

Instruction Manual. Labnet Spectrafuge 16M Microcentrifuge

INSTRUCTION MANUAL FOR THE CLINICAL 50 CENTRIFUGE

CONTENTS. 3 MAINTENANCE 3.1 Service and Maintenance Maintenance and cleaning Glass breakage Disinfection

Model : LMHS-40 Please read this manual carefully before using the instrument Labnics Equipment

OPERATIONS MANUAL. Mini Centrifuge Model MCF Certified

BlueSpin Cryo. (Cat. No. BSD-CRYO)

Operation Manual Supra R22

USER MANUAL. D2012 High Speed Mini Micro-Centrifuge SCILOGEX, LLC.

Multi-Purpose, High Speed Centrifuge 1248R/1580R User s Manual

Instruction Manual for Table Top Centrifuge Z 366

Centrifuge. Model XC-2000 Operating Manual

D3024 & D3024R High Speed (Refrigerated) Micro-Centrifuge DLAB Laboratory Instruments Limited

Distribution Limited. Petrol Power Float Operator s Manual BT900 FOR YOUR SAFETY READ AND UNDERSTAND THE ENTIRE MANUAL BEFORE OPERATING THIS MACHINE

Instruction Manual for Table Top Refrigerated Centrifuge Z 36 HK

D2012 & D2012 plus High Speed Mini Centrifuge

DM1424 Hematocrit Centrifuge

High Speed Micro-Centrifuge User manual

Heavy Duty Hoist WARNING. General Safety Precautions. Thank you for purchasing a

USER MANUAL. D3024 & D3024R High Speed (Refrigerated) Micro-Centrifuge SCILOGEX, LLC.

L550 Table-Top Low-Speed Large Capacity Centrifuge

INSTRUCTION MANUAL FOR BENCH TOP CENTRIFUGE Z 200 A

This chapter describes how to perform periodic inspection and maintenance to ensure the long operating life of the chamber.

OPERATING MANUAL. Mega 17R HIGH SPEED REFRIGERATED CENTRIFUGE HANIL SCIENCE INDUSTRIAL CO., LTD. PRINTED IN KOREA

DM0636 Multi-purpose Centrifuge

Three Lobes Roots Blower. Instruction Manual

PrO-ASTM Centrifuges. Centurion. C1015 Micro Prime Centrifuge. A Micro that offers: Oil testing. Petroleum testing, ASTM methods

MODEL 853VES Model 853VES Operator s Manual

Z100A Clinical Centrifuge

Clinical Centrifuge. Before using centrifuge, please carefully read this user manual for efficient operation and safety.

CLEMENTS. Orbital 160 Benchtop Centrifuge. CEN V 50Hz. User Manual. Manual No. CEN Issue 4

Centrifuge Operator / Service Manual

Neofuge 13/13R. Laboratory Centrifuge. Applications. Heal Force leads you to healthier life

Supra R30. High Speed Centrifuges. High Speed Centrifuge. Supra R30 / Supra R22 / Supra R17 / Supra R12

materials and workmanship for 2 years. Should the centrifuge require warranty or Phone: or Fax:

air-cooled 18,000 rpm 25,356 xg 24,000 rpm 45,078 xg refrigerated Table-Top High-Speed Centrifuges High performance instrument for the daily routine

PrO-Cyto Centrifuges 61

Spectrafuge 24DMicrocentrifuge User Manual C2400 C V C V

MODEL 853VES. Model 853VES Operator s Manual. Laboratory Centrifuge IVD

WARNING. Electric Recovery Winch. General Safety Precautions

USER MANUAL PRF DUO. Clinical Centrifuge. Before using centrifuge, please carefully read this user manual for efficient operation and safety.

CLEMENTS. Orbital 260 Benchtop Centrifuge. CEN V 50/60Hz. User Manual. Manual No. CEN Issue 5

REFRIGERATED CENTRIFUGE. LRF-A2 Series

Max. 22,000 rpm / 48,700 xg, full 3 liter capacity

Rotor. Max. RCF(xG) angle. 45 o 1210xG. LCEN-200/201, Chlinical Centrifuges, 8 Places Or 12 Places Of 15ml Tubes. Max.

Large Capacity, High-Speed Centrifuge

WIND TURBINE ELECTRICAL AUTOCONTROLLOR

Instruction Manual for the Revolutionary Science RS-200 RevSpin Microcentrifuge

INSTRUCTIONS MANUAL CLAM BALL / CLAM CAP

SPECTRAFUGE 6C. User Manual C0060 C V

Centrifuge. Model XC-1000 Operating Manual

Multi-Purpose Rotor & Micro Rotor can be used in this convenient unit.

Extremely quiet due to unique flexible drive system. Super High-Speed Centrifuge. Refrigerated - 30,000 rpm / 70,434 xg

BLOOD BANK REFRIGERATED CENTRIFUGE

ATV WINCH. Thank you for purchasing a

USER MANUAL. GASOLINE SCOOTER (4 Stroke engine) OFF ROAD USE ONLY. Read these instruction carefully before using

EBA 21 Small Centrifuge

XCM12 Mini-Centrifuge

REFRIGERATED CENTRIFUGE. LRF-A1 Series

Clinispin CT20 User Manual

Instruction manual for our serial Screw conveyor

4x750 ml Ventilated Refrigerated. Centrifuges

Service Manual. Extractor Model XR28QP. For The. For: Troubleshooting Adjustments

Vetlab CombiSpin Centrifuge

Model : LMHS-60 Please read this manual carefully before using the instrument Labnics Equipment

SLIDING DOOR OPERATOR INSTRUCTION MANUAL

Immunity: EN capacity 75 µl Capillaries. 1,5 x 75 mm. Rotor and Lid as evaluation disc incl.

Since Discover economical and innovative solutions.

Universal Centrifuge 320/320R

Company Profile Hunan Xiang Yi Laboratory Instrument Development Co., Ltd is high technology enterprises mainly engaged in manufacturing centrifuge

Heal Force Multipurpose Centrifugeses

I. General Safety Precautions

Register your instrument! MiniSpin /MiniSpin plus. Operating manual

swing-out rotor MULTIBAS+

FTS600P FTS800P FTS1200P

Instruction Manual for the Revolutionary Science RS-102 RevSpin Microcentrifuge

MODEL 755VES Model 755VES Operator s Manual

MIKRO R HETTICH DATA SHEET MICROLITRE CENTRIFUGES FASTER RESULTS IN MOLECULAR BIOLOGY FEATURES HIGHLIGHTS

Scandlight 1500W DMX STROBE USER S MANUAL

SAC SERIES CONTENTS TRIPLE-INTERVAL HIGH PRECISION COUNTING SCALE OPERATION MANUAL 1. INSTALLATION 2. SPECIFICATIONS

Self-Recovery Winch WARNING. General Safety Precautions

Instruction Manual for Compact Centrifuge Z 206 A

User s Manual. Automatic Switch-Mode Battery Charger

Prism Microcentrifuge

MICRO CENTRIFUGES. Experts in Laboratory Equipment

目录

Comfy Dry Bath Incubator

It s Your Turn. Eppendorf Centrifuge 5702, 5702 R and 5702 RH

Bench Centrifuge or Bench Centrifuge with Trolley

HIGH SPEED CENTRIFUGE

EBA 21. Small Centrifuge

PrO-Vet Centrifuges. Centurion. Veterinary, Equine & Cytology. Scientific Limited

Transcription:

OPERATION & MAINTENANCE MANUAL FOR SCEN-206 LED & MULTI-FUNCTION TYPE CENTRIFUGE PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION. M.R.C.LTD. Hagavish 3 st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529

INDEX (Ⅰ) SPECIFICATION---------------------------------------------------------------- 1 (Ⅱ) ROTOR AND ACCESSORY---------------------------------------------------- 2 (Ⅲ) INSTALLATION------------------------------------------------------------------ 3 (Ⅳ) OPERATION--------------------------------------------------------------------- 4 (Ⅴ) REPAIR---------------------------------------------------------------------------- 8 (Ⅵ) PARTS LIST----------------------------------------------------------------------- 10 (Ⅶ) CIRCUIT DIAGRAM------------------------------------------------------------ 11 (Ⅰ) SPECIFICATION Speed (rpm) Max. centrifugal Force Max. Capacity Electronic Control 0~6,000 LED display 3,980 x g 300 ml Microprocessor control system 10 memory programs available Imbalance and cut off power system Motor at over heat protective device Safety Device Triple balance system Automatic Lid lock, lift cover and power cut off Glass fuse 6A Timer 1 ~ 99 min. digital time setting Weight (kg) 10 Dimension (cm) 29 x 32 x 24.5 Brake System Power Power Consumption Automatically brake system 110 / 220V, 50 / 60Hz 150W 1

(Ⅱ) ROTOR AND ACCESSORY 1. Centrifugal force calculation: Use its speed and radius to calculate force as following formula. RCF= (rpm) 2 x 0.00001118 x R R: radius (cm) RCF: Relative Centrifugal Force ( x g ) 2. Acceptable speed and load: Please refer to the form (Ⅱ-4) to see the max. speed each model of rotor and tube carrier can be load. The load weight including: rotor, tube carrier, washer, samples, and so on. 3. Sterilizing and cleaning rotor: Rotor is made of stainless steel or steel or aluminum with high density coating and highly resistant to corrosion, but not fit to alkaline and strong acid substance. Note: Don t heat the rotor to a high temperature (100 ) to clean or sterilize it. 4. Rotor Specifications: Rotor Capacity (ml x pcs) Max. RCF Tube size (φx l ) mm (x g) Max. Min. Cat. No. RA-1508 15 x 8 3,940 16.5 x 122 15 x 90 2-5052-11 RA-1512S 15 x 12 3,940 16.5 x 122 15 x 90 2-5052-21 RA-5004 50 x 4 3,980 29.5 x 115 27.5 x 90 2-5052-12 RA-5006 50 x 6 3,980 29.5 x 115 27.5 x 90 2-5052-22 REA-1520 1.5 x 20 3,140 11 x 48 10 x 40 2-5052-13 ** Tubes are not attachable. ** Before using the centrifuge, please must refer to the rotor s max. speed setting value. 2

(Ⅲ) INSTALLATION 1. When you take out centrifuge from carton, please check if there is any scratch, press trace and damage on knob or not. If yes, please contact local dealer immediately. 2. Please install the centrifuge on a horizontal and stable table or bench. If not, it will affect anti-vibration efficiency. 3. Please use power voltage as following range. To avoid electric shock please make sure to connect ground wire to the ground terminal. Power voltage Acceptable voltage range 100V / 110V 90V ~ 110V / 99V ~ 121V 220V / 240V 198V ~ 242V / 216V ~ 264V 4. Keep it away from dusty and moist working place. 5. Avoid the same power source with the large electric consumption machine at same time. 6. As less as better to use brake system, it will last more motor s life. 7. After dismantle or clean rotor, please must screw it tightly. 3

(Ⅳ) OPERATION 1. Tube & Sample Setting (1) The tubes in use must be placed in even number such as 2, 4, 6, 8. Glass tube must be placed in carrier completely which are positioned symmetrically toward rotor center. When you use tubes in odd number, it is necessary to put same volume of water in a separate tube and then place such tubes into carrier for dummy use. (2) Please consider the speed and centrifugal force the tubes can bear or not, or it will cause the tubes easy broken from rotation. (3) Please note tube length, if it is too long than the top of tube will hit the rotor and break. If you have to put longer tubes in angle rotor, please place an interval between tubes. (4) Adapter is available for tube s diameter much smaller than carrier s, it can prevent tube broken from rotation. correct incorrect Adapter 50cc 15cc 15cc 6cc Caution: a. Adapter should be placed into tube carrier properly, if it is slanted, then tube is difficult to take out from carrier. If glass tube is broken or if rotor elapsed over one year, then rotor needs to be renewed. Damaged rotor or rotor containing broken glass will cause glass tube broken easily. b. Carrier should be placed onto rotor properly, if not, it will be imbalanced and destroy centrifuge, meanwhile to check if carriers swing normally by hand first. c. Stainless steel tube carrier can be used for acidity and alkalinity tube samples. (please purchase separately) 4

2. Door Lock & Cover (1) Please close cover tightly, otherwise rotor can not rotate, cover can not be opened during rotation. (2) Plug in the socket and switch it on, then press OPEN button and Manual safe button to open cover. (3) If power cut off and cover can t be opened, then use a small rod in more than 5 cm length to put into the hole at the right side of centrifuge and press manual safe button at the same time, then cover can be opened. 3. Outline: See-through window Cover Emergency open hole Manual safety button Control panel Rubber stepping-stone 5

4. Operation Panel: (01) START: After setting time and speed, press it and rotor rotates immediately.. (02) STOP: For emergency stop. (03) BRAKE: It is with auto-braking function, when presetting time is due. (04) OPEN: It is used at speed under 100rpm. (05) RPM: Press RPM (flash), adjust or then press RPM to finish it. (06) TIME: Press TIME (flash), adjust or then press TIME to finish it. (07) INCREASE: to add speed or time. (08) DECREASE: to reduce speed or time. (09) RECALL: to display settled speed and time. (10) STORE: to remember settled speed and time. After setting speed and time, press STORE twice. If you need presetting function, press STORE first then press PROG to select demand figure then press STORE. (11) PROG: to display speed and time. Press RECALL (flash) then press PROG to select demand figure, then press RECALL again to confirm. (12) DOOR LAMP: when cover is open, it lights up. (13) BRAKE LAMP: when it lights up, it will decrease speed automatically. (14) IMBAL: when it lights up, it is imbalanced on rotor and stop running. You have to press OPEN to open cover and check problem. (15) RUN: it lights up during rotation. (16) DISPLAY: to show speed, time and settled figure. (17) Transient centrifugal function: it can reach specified speed in transient time without speed and time limited. 6

5. Operation Procedure (01) Insert power plug and switch power to ON at the right back side of centrifuge (02) Choose operation parameter and press START to begin operating. Meanwhile, the light of RUN lighted up automatically. (03) When there is a balance error, the light of IMBAL lighted up and stop operating. Then press OPEN to check the rotor. (04) Press BRAKE to set the time of brake. (05) When operating time is up, the buzzer will sound automatically. (06) During rotation, the OPEN button can not work. (07) When the cover is open, the light of DOOR will light up and the START can not work. (08) Speed and time can be changed during operation, but it will not re-countdown. (09) LED display time, speed and signal of breakdown. (10) Emergency open hole can be used when electric power cut. (11) The last setting parameter will be stored and display next time. (12) Parameters will not be removed from memory storage when power cut off. (13) To set speed, please refer to the form (Ⅱ-4). Caution: 1. Please refer to operation panel, there are 10 sets of program available to store the desired rpm and time. When you start to run the centrifuge next time, it will perform the last running program. These programs don t disappear at power off and on. 2. During rotation, if you want to change speed or time, then press RPM or TIME and adjust or then press RPM or TIME to finish it separately. 3. During rotation, please don t exceed the max. speed (refer page 2) of the centrifuge and place substandard rotor or carrier. The centrifuge is without anti-explosion, so please don t use explosive material. 7

(Ⅴ) REPAIR 1. Checking before and after operation * Check to see if there is any material or water inside chamber and keep it clean. * Check to see if the rotor is loose or not; if yes, after screw it then use it again. * Check if carriers can swing smoothly, if not, please clean the rotor. * Check if carrier is hung well, if not, then stop it and contact with local distributor. * Check if cover is properly close and screws fixed or not; if not, then stop it and contact with local distributor. * Check if the ground wire is connected well, lamps and switch are available. 2. Daily checking: Clean stain in chamber, check if there is any scratch, deformation, and corrosion on rotor or carrier. If yes, please contact local dealer for further checking. 3. Yearly checking: Check if rotor fixing nut is loose or not; and the indication value of speed meter can match real value or not, if not, please contact local dealer. 4. Cleaning: When sample is split over rotor or carrier, please use cloth with little detergent to wipe and dry it. When sterilization is made on rotor or carrier, do not heat it to a temperature exceeding 100. 5. Trouble shooting: symptom reason Power Electrical Circuit Cover problem Motor problem Imbalance Main lamp is not light up * * * Fuse blows very often * * * Motor does not drive * * * * * Set rpm can not be attained * * * Big vibration * * Unusual noise * * 8

Repair methods: Cause Details of cause Repairing way Power a. Power cord is not connected. b. Power voltage is not proper. Checking power plug, replace fuse Lid problem Lid can not be closed PC board Parts can not connect well Contact local dealer Contact local dealer Motor problem Over- heat, carbon brush use out Stop using it and wait for cool, replace carbon brush or contact local dealer. Imbalance Carrier or tube is put improperly. Please refer operation description 6. Signal of display: Signal Status Description Repairing Err-1 Hot Motor can not drive Motor stops during operation Sensor can not detect well Motor is over-heat Motor stops running Carbon brush connect badly. Wire matched badly. The outlet of controller is out of order. Carbon brush connect badly. Wire matched badly. The outlet of controller is out of order. Sensor is out of order Wire matched badly. The outlet of controller is out of order. Temp. switch skips away. Temp. switch is out of order. Wire matched badly. 1. check carbon brush 2. check wiring 3. check the outlet of controller 4. replace controller 1. check carbon brush 2. check wiring 3. check the outlet of controller 4. replace controller 1. check sensor or replace new one 2. check wiring 3. check the outlet of controller 4. replace controller After temp. is cooler then operate it again. 1. check temp. switch or replace new one. 2. check wiring. 9

7. Machine Check: a. Check motor brush and PC board: please open cover then dismantle rotor and loosen 5 screws at the baseboard and take apart the body and baseboard. Then you can check both easily. b. Fuse is under the power plug, please take off power cord and use 一 type of screw driver to open it then check if fuse burnt or not. There is one reserved fuse available attached inside. Fuse position Reserved fuse Fuse in use (Ⅵ) PARTS LIST ITEMS QUANTITY MOTOR 1 ROTOR 1 SPEED SENSOR 1 CARBON BRUSH 2 PC BOARD 1 KEY BOARD 1 FILTER CAPACITY 1 BRAKE RESISTANCE 1 POWER SWITCH 1 SOLENOID 1 FUSE 6A 2 LID MICRO SWITCH 1 TRANSFORMER 1 10

(Ⅶ) CIRCUIT DIAGRAM 11