SERVICE MANUAL VIO 50 C ELECTROSURGERY

Similar documents
Service Manual. VIO 300 S V 1.0.x. VIO 200 S V 1.2.x. V 1.1.x V 1.2.x 02.08

The Erbe workstation configurable for any speciality

APC 3. power your VIO 3 ARGON PLASMA COAGULATION

ZE motor-protective relay Overload monitoring of EEx e motors

TIMER INTERFACE USER MANUAL

Type Operating Instructions. Bedienungsanleitung Manuel d utilisation. 2/2-Way Solenoid Valve 2/2-Wege-Magnetventil Électrovanne à 2/2 voies

KeContact P20. User manual

nual k Voltage Stabilizer N) manual New Brunswick Voltage Stabilizer Operating manual

Type Operating Instructions. Bedienungsanleitung Manuel d utilisation. 2/2-Way Solenoid Valve 2/2-Wege-Magnetventil Électrovanne à 2/2 voies

User s Manual. Automatic Switch-Mode Battery Charger

Installation instructions. Adapter AD 01 AD 02 AD 03

Operating instructions Flow monitors SI / / 2010

c-go 24V/6A 24V/8A 24V/12A

Operating instructions Flow monitors SI6600 SI6700 SI /02 01/2011

USE AND CARE GUIDE DUALCHECK CIRCUIT TESTER

HST -LS Interlocking device (Translation of Original Manual)

INSTRUCTION MANUAL PORTABLE APPLIANCE TESTER

HP21 SERVICE SUPPLEMENT UNIT INFORMATION. TSC6 Two-Speed Control

Battery Charger XT EXTENDED. For lead-acid batteries. User Manual and Guide to professional charging of starter and deep cycle batteries.

Return electrodes ELECTROSURGERY

Type 0283, Operating Instructions. Bedienungsanleitung Manuel d utilisation. 2/2-way solenoid valve 2/2-Wege-Magnetventil Électrovanne 2/2 voies

Operating manual. High-voltage testing adapter VAS 6558A/30. Operating manual. Customer documentation Revision: 00 Version: 07/2018

User Manual Rittal PMC UPS 6kVA

! WARNING To avoid risk of electrical shock, personal injury or death; disconnect power to oven before servicing, unless testing requires power.

A company of ThyssenKrupp Elevator. ThyssenKrupp Aufzugswerke. Operating Manual. Oil buffer

Pressure chlorine changeover unit C 7520

Operating instruction and technical description

Instruction Manual. Q Series 2 Motor Control Modules - OEM

NT-500 Technical Support Guide. For use by qualified company approved Technicians

Technical Documentation

IVTM Installation Manual

Operating instructions Flow monitors SI5000 SI / / 2007

Operating instructions Flow monitors SI5002 SI5003 SI / / 2012

Tester for Motor-Driven Relays Type Siemens SICO 1225 M A N U A L

Rental Industry Safety Tester Safe Check 5s

VAPORIX-PCM. Technical Documentation. Corrective control module for connection to VAPORIX-Control. Version: 3 Edition: Art.

Operating Instructions K Always on the safe side.

Instruments and accessories. for waterjet surgery WATERJET SURGERY

Instruction Manual. Electrobloc EBL 40 B

Installing Power Components

INSTRUCTION MANUAL FOR THE THERMOCOUPLE ATTACHMENT UNIT (TAU)

DESCRIPTION & OPERATION

EQUIPMENT Operation Manual

Operating instructions Flow monitor SI / / 2014

USER MANUAL. Power GUARD UPS. Uninterruptible Power System

KEYLESS ENTRY SYSTEM & TIRE PRESSURE MONITOR ACCESSORIES & EQUIPMENT General Motors Corp. - Remote Keyless Entry System

HWP Protection Board

Warmbier Germany, Wrist Strap & Footwear Combo Tester

Information I Manuals I service. AIR premium x2. Operating Instructions _a I 03/2012

User s Guide. EM3 AEM Monitor

Accessories for Stand-alone inverter SUNNY ISLAND GENMAN

BM-780 II. Service Manual V 2.1

HST-BL-2830MS & HST-BL-2830MS-USA

Battery Control Center - Diesel

User Manual Digital Energy Uninterruptible Power Supply ML Series UPS VA GE Digital Energy Power Quality

Instruction Manual. Caravan Charging System CSV 404. Table of contents

BMW i. Freude am Fahren. BMW i Wallbox Plus. Owner's manual

Match 19" GE Digital Energy. Uninterruptible Power Supply VA. Technology for the Digital World. Match 19" UPS.

Operating Manual (Edition 04/2004) sinamics. Braking Module / Braking Resistor SINAMICS G130

1 AMP CURRENT SOURCE

Contents. Introduction Inspection Notes on Safety Precautions. Chapter 1 Overview Product Overview Names of Parts 2

CAUTION All safety information must be followed as provided in Service Manual

Uninterruptible Power System

METROLOGIC INSTRUMENTS, INC. MX001 Industrial Control Interface Installation and User s Guide

Sense7-series Non-contact coded safety switch

ST48-WHUV.102. Wiring diagram. Product description. PID controller. Order number

Date received into service; / /.

Interactive Whiteboard. User Guide. Before operating the unit, please read this manual thoroughly, and retain it for future reference

DC-PMM/24V Power Management Modul für Positionier-Motor POSMO A Power Management Module for Positioning Motor POSMO A

M200. Battery Charger. For lead-acid batteries. User Manual and Guide to professional charging of starter and deep cycle batteries.

! WARNING To avoid risk of electrical shock, personal injury or death; disconnect power to oven before servicing, unless testing requires it.

TURBOTRONIK NT 151/361

DC POWER SUPPLY ALIMENTATION C.C.

Installation and operating instructions. Solar charge controller MPPT 10 A / 20 A Z Z

OPERATING INSTRUCTIONS PLEASE READ CAREFULLY

IQS300 Series Quasi-Sine Inverter. Installation & Operation Manual

SCHNITZ MOTORSPORTS USER MANUAL AND INSTALLATION GUIDE PRO-MOD BATTERY VOLTS DIAGNOSTICS NOS PULSE FREQUENCY NOS DELAY TIME IN SECONDS

Operating instructions Flow monitor SI / / 2013

BMW i. Freude am Fahren. BMW i Wallbox Plus. Instructions for use

SCHNITZ MOTORSPORTS PNC-202, 2-STAGE PROGRESSIVE NITROUS CONTROLLER USER MANUAL AND INSTALLATION GUIDE NOS PULSE FREQUENCY

User Manual. Solar Charge Controller 3KW

SINAMICS G130. Motor reactors. Operating Instructions 05/2010 SINAMICS

! WARNING To avoid risk of electrical shock, personal injury, or death, disconnect power to range before servicing, unless testing requires power.

Declaration of Conformity

User Manual. Digital Energy Uninterruptible Power Supply ML Series UPS VA. GE Digital Energy Power Quality. GE imagination at work

TABLE OF CONTENTS. 8.3 Online Sensor Specifications. Series TS (Models TS1, TSP, TSH, TSL,TSF, TSB1, TSB2) According to Electromatic factory procedure

TIC 300 PRO. Users Manual Mode d emploi Bedienungshandbuch Manual d uso Manual de uso. Non Contact AC Voltage Detector. High Energy Tic Tracer

Products for specific requirements

Siemens Drives & PLCs

OnGuard Display Operating Instructions

MODEL 520 REMOTE START ENGINE MANAGEMENT SYSTEM

##CFG BWOP Manuals## Astramax Mobile Models Assembly Guide

EN Operating manual. Motorised zone valve. Three-way, 22 mm & 28 mm 3PV2, 3PV8 & VRMH3

Wilo-Control SC-Fire Jockey

Superchips Model 2704 MAX MicroTuner GM Trucks with 6.6L Duramax Diesel Engines Vehicle Programming Instructions

Operating instructions Rotary actuators M135 M140 M150 M180. June 2012 / / EN

ABB i-bus EIB EIB Power Supply Units SV/S SV/S SU/S

3200NT Timer Service Manual

RADIO FREQUENCY SYSTEMS

! WARNING To avoid risk of electrical shock, personal injury, or death, disconnect power to range before servicing, unless testing requires power.

Transcription:

SERVICE MANUAL VIO 100 C V 1.0.x VIO 50 C V 1.0.x ELECTROSURGERY 80116-221

SERVICE MANUAL VIO 100 C VIO 50 C

Registered trademarks of Erbe Elektromedizin GmbH: AUTO CUT, AXUS, BICISION, BiClamp, classiccoag, classic- CUT, CLEVERCAP, DeCo, drycut, endocut, endocoag, ERBE, erbe power your performance., ERBECRYO, ERBEFLO, ERBEFLO AeroRinse, ERBEJET, ERBELift, ERBOKRYO, FIAPC, forcedcoag, Hybrid knife, HybridKnife, ICC 200, ICC 80, KYRON, NESSY, NESSY Ω, PLURA, preciseapc, Preflow, pulsedapc, ReMode, REMODE, softcoag, spraycoag, swiftcoag, The Color Blue, thermoseal, twincoag, VIO. Service Manual Art. No. 80116-221 All rights to this manual, in particular rights of duplication, dissemination and translation, are reserved. No part of this manual may be reproduced in any form (by photocopying, microfilming or other methods) or processed, duplicated or disseminated by the use of electronic systems without the written consent of Erbe Elektromedizin GmbH. The information contained in this manual may be amended or supplemented without prior notice and represents no obligation on the part of Erbe Elektromedizin GmbH. Printed by Erbe Elektromedizin Printed in Germany Copyright Erbe Elektromedizin GmbH, Tübingen 2017

Table of Contents Table of Contents 1 Safety information....................................... 7 Classification of the safety information................................. 7 Knowledge of the User Manual....................................... 7 Protection from the risk of electric shock............................... 7 Electrostatically sensitive components................................. 8 Liability and warranty............................................... 8 2 Controls................................................ 9 Controls at the front................................................ 9 Controls at the rear................................................ 10 3 Technical Data.......................................... 11 4 Service mode........................................... 13 General information............................................... 13 Setup overview................................................... 13 Service overview.................................................. 16 Call up Service mode............................................... 19 Exit Service mode................................................. 19 5 Remedying malfunctions................................. 21 Safety information................................................ 21 Error messages................................................... 21 6 Maintenance and servicing................................ 27 Who is allowed to perform servicing and maintenance work?............. 27 What is a technical safety check?.................................... 27 How often does a technical safety check have to be performed?........... 27 Technical safety check step by step................................. 28 Safety information.............................................. 28 Testing and measuring equipment................................. 29 User manual and visual inspections................................ 30 Tests to be conducted in accordance with the national specifications and regulations................................................ 30 DC resistance.................................................. 31 Performance tests.............................................. 33 Footswitch activation............................................ 34 Fingerswitch activation.......................................... 35 AUTO START mode (for VIO 100 C only)............................. 36 HF power output CUT............................................ 37 HF power output COAGULATE..................................... 39 Monitor circuits................................................. 42 5 / 44

Table of Contents 6 / 44

1 Safety information CHAPTER 1 Safety information Classification of the safety information WARNING! The WARNING! safety indication refers to a risk of personal injury. CAUTION! The CAUTION! safety indication refers to a risk of damage to property. ATTENTION! The ATTENTION! safety indication refers to a risk which can cause equipment to become unserviceable. IMPORTANT! The IMPORTANT! designation indicates application information and other particularly important information. Knowledge of the User Manual The User Manual for this unit constitutes an integral part of this Service Manual. For performing servicing activities it is assumed that the reader has knowledge of the User Manual, especially procedures for installation, putting into operation, and handling. Protection from the risk of electric shock WARNING! The supply voltage must match the voltage specified on the rating plate. Connect the unit / the equipment cart to a properly installed grounded outlet. Only use the Erbe power cord or an equivalent power cord for this purpose. The power cord must bear the national test symbol. For safety reasons, multiple outlets and extension cords should not be used. If their use is unavoidable, they also must be provided with proper grounding. WARNING! Unplug the power cord from the outlet before exchanging parts of the unit or cleaning it. WARNING! Do not plug a wet power cord into the unit or into an outlet. 7 / 44

1 Safety information WARNING! Do not touch any unprotected wires or conductive surfaces while the unit is opened and under voltage. WARNING! Blown line fuses may only be replaced by a competent technician. Only replacement fuses of the rating specified on the unit's name plate may be used. Before resuming operation the unit must be subjected to a performance test by a competent technician. Electrostatically sensitive components CAUTION! This unit contains electrostatically sensitive components. Work at an anti-static workplace while repairing the unit. Wear a grounding armband while working with electrostatically sensitive components. Hold the circuit boards by their non-conducting corners. Use an anti-static container for transporting electrostatically sensitive components and the circuit boards. Liability and warranty ATTENTION! Adjustments, tests, modifications, maintenance and repair work may only be performed by Erbe or persons trained by Erbe. If the work is not performed by trained persons, Erbe accepts no liability and warranty rights become void. It is recommended that the technical safety check also be performed by Erbe or persons trained by Erbe. ATTENTION! Only use original Erbe spare parts. The manufacturer accepts no liability and the warranty rights becomes void if original spare parts are not used. 8 / 44

2 Controls CHAPTER 2 Controls IMPORTANT! This chapter contains an overview of the controls of the unit(s). The relevant User Manual for the unit(s), knowledge of which is assumed for servicing work, provides detailed information about how to use the unit(s). Controls at the front VIO 100 C 2 5 8 3 6 9 4 7 1 Fig. 2-1 1 Power Switch 2 Selection button for programs 3 yellow selection button (CUT) 4 blue selection button (COAG) 5 Selection button for activation type 6 yellow Plus/Minus buttons (CUT) 7 blue Plus/Minus buttons (COAG) 8 Focus button for bipolar receptacle 9 Focus button for monopolar receptacle Footswitch indicator light AUTO START indicator light Return electrode indicator light 9 / 44

2 Controls Controls at the rear 1 2 3 Fig. 2-2 1 Potential equalization terminal 2 Footswitch socket 3 Power supply module with fuses 10 / 44

3 Technical Data CHAPTER 3 Technical Data Power connection Rated supply voltage 100 V to 240 V (± 10 %) Rated supply frequency Line current Power input in standby mode Power input with max. HF output Terminal for grounding (potential equalization) Power fuses 50 / 60 Hz Max. 2.0 A < 15 watts 190 watts / 200 VA Yes T 4 A H / 250 V Operating mode Intermittent operation Duty cycle 25% (10 s ON / 30 s OFF) Dimensions and weight Width x height x depth Weight 280 x 135 x 300 mm 4.0 kg Ambient conditions for transport and storage of unit Temperature -40 C to + 70 C Relative humidity 10% 95% Ambient conditions for operation of unit Temperature +10 C to + 40 C Relative humidity 15% 80%, noncondensing 11 / 44

3 Technical Data Acclimatizing If the unit has been stored or transported at temperatures below +10 C or above +40 C, the unit will require approx. 3 hours to acclimatize at room temperature. Standards Classification according to EC Directive 93/42/EEC Protection class as per EN 60 601-1 Type as per EN 60 601-1 II b I CF 12 / 44

4 Service mode CHAPTER 4 Service mode General information This unit has a service mode with two levels: Level 1 = Setup (adjustment of settings that affect the operation of the unit, e.g. display brightness) Level 2 = Service (adjustment of settings that affect the operation of the unit + service settings + test programs) Setup overview No. Designation Description Use 1 Volume for CUT activation Volume setting of the CUT activation tones in 10 levels. Activate setup: Press the yellow selection button. The current setting appears in the blue display. Change settings: Press the blue Plus/Minus buttons. The setting change appears in the blue display. Save setting change: Exit setup: Press and hold the top left selection button until a check mark in the adjacent display indicates the end of the process. Press the yellow selection button again. The settings display goes out in the blue display. 13 / 44

4 Service mode No. Designation Description Use 2 Volume for COAG activation Volume setting of the COAG activation tones in 10 levels. Activate setup: Press the yellow selection button. The current setting appears in the blue display. Change settings: Press the blue Plus/Minus buttons. The setting change appears in the blue display. Save setting change: Press and hold the top left selection button until a check mark in the adjacent display indicates the end of the process. Exit setup: Press the yellow selection button again. The settings display goes out in the blue display. 3 Volume button Volume setting of the buttons in 10 levels. Activate setup: Press the yellow selection button. The current setting appears in the blue display. Change settings: Press the blue Plus/Minus buttons. The setting change appears in the blue display. Save setting change: Exit setup: Press and hold the top left selection button until a check mark in the adjacent display indicates the end of the process. Press the yellow selection button again. The settings display goes out in the blue display. 4 Warning tone volume Volume setting of the warning tones in 8 levels. Activate setup: Press the yellow selection button. The current setting appears in the blue display. Change settings: Press the blue Plus/Minus buttons. The setting change appears in the blue display. Save setting change: Press and hold the top left selection button until a check mark in the adjacent display indicates the end of the process. Exit setup: Press the yellow selection button again. The settings display goes out in the blue display. 14 / 44

4 Service mode No. Designation Description Use 5 Display brightness Selection of the display brightness in 10 levels. Activate setup: Press the yellow selection button. The current setting appears in the blue display. Change settings: Press the blue Plus/Minus buttons. The setting change appears in the blue display. Save setting change: Press and hold the top left selection button until a check mark in the adjacent display indicates the end of the process. Exit setup: Press the yellow selection button again. The settings display goes out in the blue display. 15 / 44

4 Service mode Service overview ATTENTION! In this table, service programs are described that are used for basic servicing work. In addition, the service level provides a series of test programs with a number of diagnostic and inspection options in the unit. Test programs are described in the repair instructions and may not be used without precise knowledge of their function. No. Designation Description Use 1 5 Setup programs Repeat of setup programs 1 5 from service level 1. 6 19 Not assigned. 20 Reset Front panel reset to default values (=default values for programs, setup and service settings, clearing the error list). Perform reset: Press the yellow selection button. The S (for Service mode) appears with an asterisk in the top left display. Now press and hold the top left selection button until a check mark appears in the top left display. 21 Error list Saves up to 20 errors. Display error list: Press the yellow selection button. The error number appears in the yellow display, the associated error code in the blue display. The yellow Plus/Minus buttons can be used to switch between errors. Clear error list: Press and hold the top left selection button until a check mark in the adjacent display indicates the end of the process. Exit service program. Press the yellow selection button until only the number of the service program appears in the yellow display. 16 / 44

4 Service mode No. Designation Description Use 22 Activation period limit Setting of the unit activation period between 1 and 99 seconds. Activate service program: Change settings: Press the yellow selection button. The current setting appears in the blue display. Press the blue Plus/Minus buttons. The setting change appears in the blue display. Save setting change: Press and hold the top left selection button until a check mark in the adjacent display indicates the end of the process. Exit service program. Press the yellow selection button until only the number of the service program appears in the yellow display. 23 Test program -> Description, see repair instructions. 24 Test program -> Description, see repair instructions. 25 Software version Display of the software version of the unit. Display software version: Press the yellow selection button. The three-digit software version appears (first and second digit in the yellow display, third digit in the blue display). Exit service program: Press the yellow selection button until only the number of the service program appears in the yellow display. 26 Test program -> Description, see repair instructions. 27 Activation type Setting of the AUTO START activation, ON or OFF. Set activation type: Press the yellow selection button. The current setting appears in the top right display (AUTO ON or AUTO OFF). The top right selection button is used to switch between settings. Save setting change: Press and hold the top left selection button until a check mark in the adjacent display indicates the end of the process. Exit service program. Press the yellow selection button until only the number of the service program appears in the yellow display. 17 / 44

4 Service mode No. Designation Description Use 28 Return electrode resistance Display of return electrode resistance values. Display NE resistance: Press the yellow selection button. The current NE resistance appears in the blue display. Exit service program: Press the yellow selection button until only the number of the service program appears in the yellow display. 29 Test program -> Description, see repair instructions. 30 Test program -> Description, see repair instructions. 31 Test program -> Description, see repair instructions. 18 / 44

4 Service mode Call up Service mode Call up Setup VIO 100 C 1. When switching on the unit, press and hold the top left selection button. The unit switches on to the Setup and shows <Setup 1> in the yellow display. The yellow Plus/Minus buttons can be used to switch between setup settings. For more about setup settings, see table on page 13. Call up Service VIO 100 C 1. When switching on the unit, press and hold the top left selection button. The unit switches to the Setup. VIO 100 C 2. Press and hold the three buttons as illustrated above. The unit switches on to the service level and shows an <S> in the top left display. The yellow Plus/Minus buttons can be used to switch between service programs. For more about service programs, see table on page 16. Exit Service mode Switch the unit off. 19 / 44

4 Service mode 20 / 44

5 Remedying malfunctions CHAPTER 5 Remedying malfunctions Safety information ATTENTION! Adjustments, technical tests, modifications, maintenance and repair work may only be performed by Erbe or persons trained by Erbe. If the work is not performed by trained persons, Erbe accepts no liability and warranty rights become void. Error messages Error code Description Additional information Action 1 Reaching the activation period limit. If the activation period limit has been set too short, service program 22 can be used to extend it. 2 NE monitoring returns an illegal value at or during monopolar activation. Check whether the NE is applied correctly and connected to the unit. Check whether the NE cable is connected correctly and in good condition. If the error still occurs, notify Erbe Service. 3 Activating the unit without prior acknowledgement of the front panel setting. 4 Activation of CUT with no defined power setting. 5 Activation of COAG with no defined power setting. Press any button to acknowledge the settings of the front panel and then reactivate the unit. Select CUT power setting > 0 watts before activating the unit. Select COAG power setting > 0 watts before activating the unit. 21 / 44

5 Remedying malfunctions Error code Description Additional information Action 6 A button on the front panel is pressed when an activation request occurs. Do not activate the unit when making settings on the front panel. If the error still occurs, notify Erbe Service. 7 Restart the unit. If the error occurs again, notify Erbe Service. 8 Several activation signals are pending simultaneously. Actuate only one activation signal (footswitch or fingerswitch). Check the correct function of the fingerswitches and footswitches. If the error still occurs, notify Erbe Service. 9 Activation source (=footswitch) was not assigned. The footswitch must first be assigned to an output socket before the footswitch can be used for activation. Assign the footswitch of the required output socket. 10 18 Restart the unit. If the error occurs again, notify Erbe Service. 19 After assigning the AUTO START function, an activation was detected within a period of 200 ms. Occurs when an electrically conductive contact exists between the electrodes of the connected instruments during AUTO START assignment. Check the connecting cable to the forceps or forceps for shortcircuit. Ensure that there is no electrically conductive contact between the forcep branches during AUTO START assignment; place the forceps on an insulated base and open the forcep branches (electrical contact is interrupted). If the error still occurs, notify Erbe Service. 22 / 44

5 Remedying malfunctions Error code Description Additional information Action 20 The CUT footswitch pedal was pressed when switching on the unit. Do not press the CUT footswitch pedal when switching on the unit. Check the correct function of the footswitch. If the error still occurs, notify Erbe Service. 21 The COAG footswitch pedal was pressed when switching on the unit. Do not press the COAG footswitch pedal when switching on the unit. Check the correct function of the footswitch. If the error still occurs, notify Erbe Service. 22 The CUT fingerswitch was pressed when switching on the unit. Do not press the CUT fingerswitch when switching on the unit. Check the correct function of the fingerswitch. 23 The COAG fingerswitch was pressed when switching on the unit. If the error still occurs, notify Erbe Service. Do not press the COAG fingerswitch when switching on the unit. 24 The selection button (CUT) was pressed when switching on the unit. 25 Only for V 1.0.0: The selection button (COAG) was pressed when switching on the unit. 26 The selection button for activation type was pressed when switching on the unit. Check the correct function of the fingerswitch. If the error still occurs, notify Erbe Service. Check whether the corresponding button was pressed when switching on the unit. If not, notify Erbe Service. Check whether the corresponding button was pressed when switching on the unit. If not, notify Erbe Service. Check whether the corresponding button was pressed when switching on the unit. If not, notify Erbe Service. 23 / 44

5 Remedying malfunctions Error code Description Additional information Action 27 The Plus selection button (CUT) was pressed when switching on the unit. 28 The Minus selection button (CUT) was pressed when switching on the unit. 29 The Plus selection button (COAG) was pressed when switching on the unit. 30 The Minus selection button (COAG) was pressed when switching on the unit. 31 The focus button for bipolar receptacle was pressed when switching on the unit. Check whether the corresponding button was pressed when switching on the unit. If not, notify Erbe Service. Check whether the corresponding button was pressed when switching on the unit. If not, notify Erbe Service. Check whether the corresponding button was pressed when switching on the unit. If not, notify Erbe Service. Check whether the corresponding button was pressed when switching on the unit. If not, notify Erbe Service. Check whether the corresponding button was pressed when switching on the unit. If not, notify Erbe Service. 32 The focus button for monopolar receptacle was pressed when switching on the unit. Check whether the corresponding button was pressed when switching on the unit. If not, notify Erbe Service. 33 More than one button was pressed when switching on the unit. Check whether buttons were pressed when switching on the unit. If not, notify Erbe Service. 34 42 Restart the unit. If the error occurs again, notify Erbe Service. 43 The unit temperature is outside the operational range. If the temperature is too high, allow the unit to cool down. If the temperature is too low, allow the unit to acclimate to room temperature. If the error still occurs, notify Erbe Service. 44 50 Restart the unit. If the error occurs again, notify Erbe Service. 24 / 44

5 Remedying malfunctions Error code Description Additional information Action 60 + 61 Restart the unit. If the error occurs again, notify Erbe Service. 62 The user interface settings are not stored in the EEPROM. Default programs are used (and the default settings for Setup and Service). 63 Error when writing to the external EEPROM. 64 Error when reading the external EEPROM. User programs cannot be saved. Attempt to resave user programs. If not possible, notify Erbe Service. Attempt to save the user program again. If not possible, notify Erbe Service. Attempt to read the user program again. If not possible, notify Erbe Service. 70 Restart the unit. If the error occurs again, notify Erbe Service. 80 + 81 Restart the unit. If the error occurs again, notify Erbe Service. 94 Microcontroller Flash data is invalid. Invalid software may have been installed. Have the mainboard reprogrammed (by Technical Service Germany only). 25 / 44

5 Remedying malfunctions 26 / 44

6 Maintenance and servicing CHAPTER 6 Maintenance and servicing Who is allowed to perform servicing and maintenance work? ATTENTION! Adjustments, tests, modifications, maintenance and repair work may only be performed by Erbe or persons trained by Erbe. If the work is not performed by trained persons, Erbe accepts no liability and warranty rights become void. It is recommended that the technical safety check also be performed by Erbe or persons trained by Erbe. What is a technical safety check? IMPORTANT! The technical safety check is a preventive measure to examine whether the device is safe and ready for operation. In order to perform the various tests the current specifications and regulations of the particular country and the instructions in this service manual must be observed. IMPORTANT! How often does a technical safety check have to be performed? Erbe recommends performing a technical safety check after every repair, but at least once a year. 27 / 44

6 Maintenance and servicing Technical safety check step by step For simplification the device to be tested is referred to below as the "test specimen". WARNING! Safety information For safety reasons (personnel protection) the test specimen should generally be operated by a suitable isolating transformer. An exceptional case is the tests for grounded conductor resistance, ground leakage current, and patient leakage current, in which the test specimen is supplied with current via the safety tester. ATTENTION! In the event of a fault occurring in the test specimen or individual components during the technical safety check the test steps taken so far no longer apply. Remedy the defect and repeat the technical safety check from the beginning. ATTENTION! The specifications made in this chapter for the test programs allow the service technician to perform the safety check. In addition, the test programs provide a number of diagnostic and inspection options in the unit; detailed descriptions are available in the repair instructions. Erbe expressly states that the full functional range of the test programs may not be used without precise knowledge of the test programs. IMPORTANT! It is assumed that the user knows how to operate the test specimen, the test equipment, the measuring equipment, and auxiliary test equipment. The test instructions only apply in conjunction with the relevant test steps. IMPORTANT! Test equipment, measuring equipment, and auxiliary test equipment (cables, test boxes, etc.) are listed separately at the beginning of each test unit. Where Erbe article numbers are specified, only original Erbe test equipment, measuring equipment, and auxiliary test equipment may be used. IMPORTANT! The test report for the technical safety check can be requested from Erbe Technical Service Tübingen. For the address see address sheet on last page. 28 / 44

6 Maintenance and servicing IMPORTANT! Testing and measuring equipment The following list contains the testing and measuring equipment recommended by Erbe for servicing. Where Erbe article numbers are specified, only original Erbe testing and measuring equipment should be used. Erbe Art. No. Description Safety tester (with insulation testing > 500 V DC) HF power meter (RL: 50 Ohm 5200 Ohm, measurement tolerance +7 % / -5 %) 20189-107 Two pedal footswitch VIO C 20190-045 Electrode handle ICC/ACC 20192-127 Patient cable AE 20192-110 Patient cable AE, international Laboratory measuring cable 20196-045 Bipolar cable 20196-053 Bipolar cable, international 2Pin 22 20100-034 Adapter cable bipolar 20194-070 Patient cable NE 20194-075 Patient cable NE, international 20100-033 Adapter cable NE 20100-035 Cable LF leakage current 20100-038 Cable LF leakage current, international 20100-101 VIO Testbox Symmetry/Resistance (NE asymmetry/critical resistance) 20100-102 VIO Testbox Auto Start/Auto Stop (bipolar start/ stop) 29 / 44

6 Maintenance and servicing User manual and visual inspections Test specimen and accessories (where enclosed) undamaged externally. User manual present. All labels on the test specimen (conformity declaration mark, rating plate, and all wording) present and readily legible. Tests to be conducted in accordance with the national specifications and regulations Grounded conductor test Ground leakage current measurement Patient leakage current measurement Ground terminal to chassis. Ground terminal to potential equalization pin. Ground leakage current, normal condition (N.C.). Ground leakage current, single-fault condition (S.F.C.). Patient leakage current, normal condition (N.C.). Patient leakage current, single-fault condition (S.F.C.). IMPORTANT! Close the output relay of the unit for testing the patient leakage current. Procedure: Only for V 1.0.0: 1. Call up service level (see page 19). 2. Use the Plus/Minus button to select and the yellow selection button to activate the test program <Setup 24>. 3. Close the relay of the bipolar receptacle. Use the Plus/Minus buttons next to the CUT and COAG displays and perform the following settings: CUT display: <Setup 1> (= bipolar receptacle) COAG display: Bipolar relay closed / monopolar relay open 4. Close the relay of the monopolar receptacle. Use the Plus/Minus buttons next to the CUT and COAG displays and perform the following settings: CUT display: <Setup 2> (= monopolar receptacle) COAG display: Bipolar and monopolar relay closed 5. Perform the measurement. 6. Switch the test specimen off. 30 / 44

6 Maintenance and servicing From V 1.0.1: 1. Press and hold the blue selection button (COAG) when switching on. The unit jumps via quick access to the appropriate test program. 2. All relays close. Press the blue selection button (COAG) again. Both relays (bipolar + monopolar) are closed or opened together. The switching state of the relays is shown as a symbol in the COAG display: COAG display: Bipolar and monopolar relay closed 3. Perform the measurement. 4. Switch the test specimen off. DC resistance Testing and measuring equipment Erbe Art. No. Description 20192-127 Patient cable AE 20192-110 or Patient cable AE, international Laboratory measuring cable 20190-045 Electrode handle ICC/ACC 20194-070 Patient cable NE or 20194-075 Patient cable NE, international 20100-033 Adapter cable NE Safety tester (with insulation testing > 500 V DC) 31 / 44

6 Maintenance and servicing Test setup WARNING! Across the measuring lines there is the DC voltage of 500 V! In order to avoid injuries, only switch on the test specimen and safety tester when all the electrical connections have been made. + + Fig. 6-1 The test specimen is connected to the power supply via the power cord. The NE socket of the test specimen is connected to the safety tester via the patient cable NE with the adapter cord. The AE socket of the test specimen is connected to the safety tester via the patient cable AE and the electrode handle with the laboratory measuring cable. Test procedure 1. Start safety tester in the "Insulation resistance" function The measured value displayed should be >200 MOhm. 2. Start the test specimen and call up test program <Setup 24>. 3. Close all output relays of the test specimen in the test program <Setup 24>. 4. Determine insulation resistance using the safety tester. The measured value must be >2 MOhm. 5. Document the measured value. 6. Exit the test program <Setup 24>. Switch the test specimen off. 32 / 44

6 Maintenance and servicing Performance tests Testing and measuring equipment Erbe Art. No. Description 20189-107 Two pedal footswitch VIO C 20194-070 Patient cable NE or 20194-075 Patient cable NE, international 20100-033 Adapter cable NE Test setup The test specimen is connected to the power supply via the power cord. The dual-pedal footswitch is connected. The NE patient cable with the shorted adapter cable is connected to the NE socket on the test specimen. Test procedure Power switch 1. Check power switch for smooth operation. The power switch must be easy to operate and must neither stick nor scrape. 2. Press power switch. The power switch must snap into the "ON" position and the test specimen must perform a system start. Start routine / acknowledgement tone Pushbuttons / acknowledgement tone when pressed Display / LEDs 1. The test specimen must perform the system start without error message(s). 2. The test specimen must emit an acoustic signal (acknowledgement tone) during the self-test. 1. Check all the pushbuttons (focus buttons, selection buttons, +/ buttons) on the test specimen to make sure they are working properly. Press each button at least twice. When pressing the buttons there must be an acoustic signal every time (= acknowledgement tone). Display: 1. Check the backlighting of the display to make sure it is operating properly. If backlighting is faulty, no image will be visible. LEDs: 1. When starting the system, all LEDs must briefly light up at the same intensity. 33 / 44

6 Maintenance and servicing Activation tones 1. Press CUT pedal on the dual-pedal footswitch at least twice. When pressing the pedal there must be an acoustic signal every time (=acknowledgement tone). 2. Press COAG pedal on the dual-pedal footswitch at least twice. When pressing the pedal there must be an acoustic signal every time (=acknowledgement tone). Footswitch activation Testing and measuring equipment Erbe Art. No. Description 20189-107 Two pedal footswitch VIO C Test setup The test specimen is connected to the power supply via the power cord. The dual-pedal footswitch is connected. Test procedure Call up test program Dual-pedal footswitch activation CUT / COAG 1. Call up service level (see page 19). 2. Use the Plus/Minus button to select and the yellow selection button to activate the test program <Setup 26>. 1. Press the CUT pedal on the two pedal footswitch. The "X" character must appear in the left segment in the CUT display 1 during activation. 2. Press the COAG pedal on the two pedal footswitch. The "X" character must appear in the right segment in the CUT display during activation. 3. Remain in the Test program. 1. Legend: " " not yet activated, "X" successful activation, "0" = activation finished 34 / 44

6 Maintenance and servicing Fingerswitch activation Testing and measuring equipment Erbe Art. No. Description 20192-127 Patient cable AE or 20192-110 Patient cable AE, international 20190-045 Electrode handle ICC/ACC Test setup The test specimen is connected to the power supply via the power cord. Test procedure Electrode handle Activation CUT / COAG 1. Connect the electrode handle with patient cable AE to the Monopolar socket of the test specimen. 2. Press the CUT button on the electrode handle. The "X" character must appear in the left segment in the COAG display 2 during activation. 3. Press the COAG button on the electrode handle. The "X" character must appear in the right segment in the COAG display during activation. 4. Remove the electrode handle. Exit test program 1. Switch the test specimen off. 2. Legend: " " not yet activated, "X" successful activation, "0" = activation finished 35 / 44

6 Maintenance and servicing AUTO START mode (for VIO 100 C only) Testing and measuring equipment Erbe Art. No. Description 20196-045 Bipolar cable or 20196-053 Bipolar cable, international 2Pin 22 20100-034 Adapter cable bipolar 20100-102 VIO Testbox Auto Start/Auto Stop (bipolar start/stop) Test setup The test specimen is connected to the power supply via the power cord. Test procedure 1. Switch on the test specimen without it being connected to the test box. 2. Activate AUTO START if necessary. To do so: Call up service level (see page 19) Use the Plus/Minus button to select and the yellow selection button to activate <Setup 27>. Use the top right selection button to set <AUTO ON>. Save the setting. Press and hold the top left selection button until a check mark in the adjacent display indicates the end of the process. Exit Service mode. Switch the test specimen off. 3. Switch on the test specimen without it being connected to the test box and set it to: BIPOLAR SOFT, 50 Watt, AUTO START 36 / 44

6 Maintenance and servicing + + BP1 1 T1 T2 2 BP2 X X X X Testbox 20100-102 4. Use the bipolar cable and adapter to connect the test specimen's bipolar socket to the test box (see image). The test specimen must start the activation automatically. If this does not occur, press the T1 button on the test box. The test specimen must start the activation automatically. 5. Press button T2. The test specimen must terminate activation. HF power output CUT Testing and measuring equipment Erbe Art. No. Description 20192-127 Patient cable AE or 20192-110 Patient cable AE, international Laboratory measuring cable 20190-045 Electrode handle ICC/ACC 20194-070 Patient cable NE or 20194-075 Patient cable NE, international 20100-033 Adapter cable NE HF power meter 20189-107 Two pedal footswitch VIO C 37 / 44

6 Maintenance and servicing Test setup + + AE NE HF power meter Fig. 6-2 The test specimen is connected to the power supply via the power cord. The levels of power are determined with the HF power meter. The measuring cables are plugged into the HF power meter direct. AUTO CUT DRY CUT Test procedure for VIO 50 C 1. Set the test specimen to: AUTO CUT, 50 Watt 2. Set HF power meter to: RL = 500 ohms 3. Activate test specimen via CUT button on the electrode handle. 4. Determine and document measured value. The tolerance range is 40 to 60 watts. 1. Set the test specimen to: DRY CUT, 50 Watt 2. Set HF power meter to: RL = 500 ohms 3. Activate test specimen via CUT button on the electrode handle. 4. Determine and document measured value. The tolerance range is 40 to 60 watts. Test procedure for VIO 100 C 1. Set the test specimen to: AUTO CUT, 100 Watt 2. Set HF power meter to: RL = 500 ohms 3. Activate test specimen via CUT button on the electrode handle. 4. Determine and document the measured value. The tolerance range is 80 to 120 watts. 38 / 44

6 Maintenance and servicing 1. Set the test specimen to: DRY CUT, 100 Watt 2. Set HF power meter to: RL = 500 ohms 3. Activate test specimen via CUT button on the electrode handle. 4. Determine and document the measured value. The tolerance range is 80 to 120 watts. HF power output COAGULATE Testing and measuring equipment Erbe Art. No. Description 20192-127 Patient cable AE or 20192-110 Patient cable AE, international 20190-045 Electrode handle ICC/ACC Laboratory measuring cable 20194-070 Patient cable NE or 20194-075 Patient cable NE, international 20100-033 Adapter cable NE 20196-045 Bipolar cable or 20196-053 Bipolar cable, international 2Pin 22 20100-034 Adapter cable bipolar HF power meter 20189-107 Two pedal footswitch VIO C 39 / 44

6 Maintenance and servicing SOFT COAG FORCED COAG Test setup + + AE NE HF power meter Fig. 6-3 The test specimen is connected to the power supply via the power cord. The levels of power are determined with the HF power meter. The measuring cables are plugged into the HF power meter direct. Test procedure for VIO 50 C 1. Set the test specimen to: SOFT COAG, 50 Watt 2. Set HF power meter to: RL = 100 Ohm 3. Activate test specimen via COAG button on the electrode handle. 4. Determine and document measured value. The tolerance range is 40 to 60 watts. 1. Set the test specimen to: FORCED COAG, 50 Watt 2. Set HF power meter to: RL = 500 ohms 3. Activate test specimen via COAG button on the electrode handle. 4. Determine and document measured value. The tolerance range is 40 to 60 watts. Test procedure for VIO 100 C 1. Set the test specimen to: SOFT COAG, 80 Watt 2. Set HF power meter to: RL = 100 Ohm 3. Activate test specimen via COAG button on the electrode handle. 4. Determine and document the measured value. The tolerance range is 64 to 96 watts. 40 / 44

6 Maintenance and servicing 1. Set the test specimen to: FORCED COAG, 50 Watt 2. Set HF power meter to: RL = 500 ohms 3. Activate test specimen via COAG button on the electrode handle. 4. Determine and document measured value. The tolerance range is 40 to 60 watts. BIPOLAR SOFT COAG Test setup + + Bipolar HF power meter Fig. 6-4 The test specimen is connected to the power supply via the power cord. The levels of power are determined with the HF power meter. The measuring cables are plugged into the HF power meter direct. The dual-pedal footswitch is connected. Test procedure for VIO 50 C 1. Set the test specimen to: BIPOLAR (=BIPOLAR SOFT COAG), 50 Watt 2. Set HF power meter to: RL = 100 Ohm 3. Activate test specimen via COAG pedal on the footswitch. 4. Determine and document measured value. The tolerance range is 40 to 60 watts. Test procedure for VIO 100 C 1. Set the test specimen to: BIPOLAR (=BIPOLAR SOFT COAG), 80 Watt 2. Set HF power meter to: RL = 100 Ohm 3. Activate test specimen via COAG pedal on the footswitch. 4. Determine and document the measured value. The tolerance range is 64 to 96 watts. 41 / 44

6 Maintenance and servicing Monitor circuits NE monitoring of critical resistance for dual surfaced return electrodes Testing and measuring equipment Erbe Art. No. Description 20194-070 Patient cable NE or 20194-075 Patient cable NE, international 20100-033 Adapter cable NE 20100-101 VIO Testbox Symmetry/Resistance (NE asymmetry/critical resistance) 20189-107 Two pedal footswitch VIO C Test set-up and test procedure 1st test step Testbox 20100-101 1 2 3 4 5 + + 6 9 7 8 10 11 12 Fig. 6-5 The test specimen is connected to the power supply via the power cord. The test is performed without a load. The NE socket of the test specimen is connected to the VIO Testbox via the patient cable NE with the adapter cable. The dual-pedal footswitch is connected. 1. Set the test specimen to: AUTO CUT, 10 Watt 2. The return electrode indicator light must light up red on the test specimen. 3. Activate AUTO CUT via the footswitch. The test specimen must inhibit activation and emit or display an optical and acoustic warning. 42 / 44

6 Maintenance and servicing 2nd test step Testbox 20100-101 1 2 3 4 5 + + 6 9 7 8 10 11 12 Fig. 6-6 The test specimen is connected to the power supply via the power cord. The test is performed without a load. The NE socket of the test specimen is connected to the VIO Testbox via the patient cable NE with the adapter cable. The dual-pedal footswitch is connected. 1. Keep the setting on the test specimen. 2. The return electrode indicator light must light up green on the test specimen. 3. Activate AUTO CUT via the footswitch. It must be possible to activate the test specimen without error or warning signals. 43 / 44

6 Maintenance and servicing 44 / 44