Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard

Similar documents
Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard

For more information about Agilent Technologies services please visit our web site using the following URL

Customer Information. Service Engineer s Responsibilities. Agilent Cary 7000 UMS/UMA/Solids Autosampler

Please exercise care when using solvents close to the instrument. It could damage the vapor sensor present in the ELSD.

Troubleshooting Tips & Tricks for your GC Analyzer & CFT Application

Installing a Purged Packed Inlet

Agilent 6890 Gas Chromatograph

6890 CHECKOUT PROCEDURE TCD (THERMAL CONDUCTIVITY) DETECTOR

Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593

Page 1. Alternate Carrier Gas Considerations and Faster GC Analysis

Agilent 7890 Series Gas Chromatograph

Technical Procedure for Gas Chromatography (GC-FID)

1310 Inlets. 1 of 36. Jun Illustrated Parts Breakdown Agilent 6890 Gas Chromatograph Service Manual

Installing a Purged Packed (PP) Inlet

Setting up SilFlow for BackFlush in your GC

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application

Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)

Technical Procedure for Pyrolysis-Gas Chromatography/Mass Spectrometry (Py-GC-MS)

6890 Installing ¼ Glass Columns

Agilent 6850 Series II Network GC System G2630A Performance Specifications

Installing a Cool On-Column Inlet

MSD EI and CI Source Cleaning and Installation. Technical Overview. Introduction

Ultra Clean Gas Filters Reference Guide ULTRA CLEAN GAS FILTERS

SPECIFICATION. Bench top triple quadrupole Mass Spectrometry System coupled to a gas performance Chromatography (GC-MS/MS)

Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS)

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

GC Best Practices & Troubleshooting. Group/Presentation Title Agilent Restricted Month ##, 200X

High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph

Typical Gas Chromatographic System

SELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS

Installation Guide. Volatiles Interface for 6890 Gas Chromatograph Accessory G2319A

Agilent and ASTM. Update on Recent Activities. Page 1

Using the PSD for Backflushing on the Agilent 8890 GC System

GCMS-QP2010 Series. Consumables. Shimadzu Gas Chromatograph/Mass Spectrometer

Installation Guide. Flame Ionization Detector on a 6850 GC Accessory G2621

GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585

Installing the 6850 Direct Valve Column Connector Kit

Hot Injection Adapter Kit - Installation Guide

Agilent G2647A (Single) and G2648A (Dual) Flame Photometric Detector Update Kits

Installing a PTV Inlet

Model 8610C Gas Chromatograph. GC Chassis Types ECD PID FID / DELCD FPD. Mounts up to Six Detectors and Five Injectors

Choose for Your Instrument

Installing a PTV Inlet

GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS

Agilent G3510A/G3511A Multimode Inlet

Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI)

GC Inlets. An Overview. Simon Jones GC Applications Engineer

Agilent 5975T. Mobile Lab Hardware Installation. Agilent Technologies

260 Solvent Vapor Exit Accessory

HP 1946A LC/MSD Replacement Parts

Application Note. Abstract. Authors. Environmental Analysis

Practical Steps in GC Troubleshooting

Performing ASTM 6584 free and total glycerin in BioDiesel using an SRI Gas Chromatograph and PeakSimple software

Oil Free Vacuum Pumps for Mass Spectrometry

GC Method Compliance and Large Valve Oven Application

Agilent Multimode Inlet for Gas Chromatography

Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application

Proof of Long-Term, Leak-Free Performance for a Novel Self-tightening GC Column Nut

Nickel Catalyst Tube Accessory 19205A

Choose for Your Instrument. GC Supplies - Agilent Instrument Quick Pick Guide

Installing an Auxiliary Pressure Control Module

Agilent 7820A Gas Chromatograph

Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis

Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011

Agilent 7693A Automated Liquid Sampler

Installation Guide. Thermal Conductivity Detector on a 6850 GC Accessory G2623A

High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System

C2, C3, C4 Monomer Analysis

Agilent InfinityLab Quick Change Valve G4235A Instructions

Agilent 7693A Automated Liquid Sampler

Installing the Flame Ionization Detector EPC Flow Control Manifold

2-D GC/MS and the Deans Switch: Theory and Applications. Fred Feyerherm, Agilent Technologies Inc. Houston, Tx

Agilent G3348B Flame Photometric Detector

Detection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector

PRODUCT SPECIFICATION. Product Specification Crystal 9000 Gas Chromatograph. Chromatec Crystal Laboratory Gas Chromatography System

Installing an Auxiliary Pressure Control Module

Fast GC. Dial for e-seminar Audio. Slide 4 SPEEDY GC

Genuine Restek Replacement Parts For Agilent 5890 GCs

trace samples >2µL 4.0mm 6.5mm x 78.5mm accommodates more polar solvents and lower MW solvents 5.0mm 6.5mm x 78.5mm 2.0mm 6.5mm x 78.

HPLC SYSTEM S 500 SERIES

Installing the CO 2 Cryogenic Oven Cooling Kit

Agilent 7697A Headspace Sampler

Impact of GC Parameters on The Separation Part 2: Choice of Column Internal Diameter

Method Development for Capillary GC Systems. Slide 1

Agilent s GC Analyzer Solutions: Introduction to the Controlled Substances Analyzer

Agilent Multimode Inlet

Agilent G1554A On-Column Capillary Inlet with EPC

10 Maintenance. Agilent Technologies. This chapter describes the maintenance of the module Infinity Binary Pump User Manual

IFPAC 2003 Dr. Berthold Andres

DANI Transformer Oil Gas Analyzer

Refinery Gas. Analysis by Gas Chromatography WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

S 1125 HPLC Pump System

Forensic Blood Alcohol Determination with the Intuvo 9000 GC

AZ Chrom s.r.o.

Analysis of Petroleum Fractions by ASTM D2887

A g i l e n t A G a s C h r o m a t o g r a p h Service Manual

Beverage Grade Carbon Dioxide

ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column

Agilent 1290 Infinity Autosampler

Transcription:

7890 GC Standard Agilent Preventive Maintenance provides factory recommended service for your analytical systems to assure reliable operation and the accuracy of your results. Delivered by highly-trained and certified service engineers using genuine Agilent parts and supplies, Agilent Preventive Maintenance provides everything you need to reduce unplanned downtime and keep your systems operating at their peak. For more information about Agilent Technologies GC Support please visit our web site using the following URL: http://www.chem.agilent.com/en-us/technical-support/instruments-systems/gas- Chromatography/Pages/default.aspx Customer Information Customers should provide all necessary operating supplies upon request of the engineer. A customer representative should be available to the engineer while performing the preventive maintenance procedures. Any parts, not included in the Parts Lists section of this document, are not part of the recommended Preventive Maintenance service, nor are they included in the price of this service. If a system requires the use of additional or special procedures and/or parts for the instrument service, then these must be ordered separately and charged as a repair, which may incur additional costs. Service Engineer s Responsibilities Only complete/printout pages that relate to the system or module being serviced. Complete empty fields with the relevant information. Complete the relevant checkboxes in the checklist using a X or tick mark in the checkbox. Complete Not Applicable check boxes to indicate services not delivered, as needed. Complete the PM service in the order of the tasks listed. Complete the Service Review section together with the customer. Additional Instruction Notes Check for any active service notes for this unit. If there are any applicable Safety or Modification Recommended Service notes, plan to implement the changes on this unit before doing any qualification service. Do not implement firmware updates, unless you get approval from the customer and are sure that they are compatible with the instrument control software. Issued: 20-Jan-2015, Revision: A.01.09 Copyright 1 2011 Agilent Technologies Page 1 of 7

7890 GC Standard System Information Guidance Check this box if an instrument configuration report is attached instead of completing the table. Instrument system name and ID Instrument system site and location List system component product numbers List the serial numbers of each component 1. 1. 2. 2. 3. 3. 4. 4. 5. 5. 6. 6. 7. 7. 8. 8. 9. 9. 10. 10. Preparation Discuss any specific issues with the customer prior to starting. Review the instrument logbook. Save instrument control settings before starting the procedure. Perform general inspection of system for cleanliness Check for proper installation of safety-related parts, assemblies, sensors etc. Check for required firmware updates and verify with customers if they would like it installed. Before starting the following procedures, record the Detector Signal Output(s) in the results table. If the GC is turned OFF or in a service mode, comparing the detector outputs before and after the service is not possible. Issued: 20-Jan-2015, Revision: A.01.09 Copyright 2 2011 Agilent Technologies Page 2 of 7

7890 GC Standard Clean and inspect GC Unplug power cord from the power source. Open GC covers and vacuum/remove any dust/debris. Pay particular attention to cooling fans. Inspect internal connectors for proper contact and placement. Reconnect Power to the GC. Power the GC on and verify the power on self-test passed. Verify oven motor spins freely and turns on with the oven door closed; off when the door is opened. Verify operation of all other fans - the inlet and EPC cooling fans. Verify oven intake/outlet flap assembly is operating smoothly while heating and cooling the oven Inlet and detector consumable replacement For the inlets installed, perform inlet maintenance as defined in the 7890 manual Maintaining Your GC - for the inlet(s) installed. Replace the split vent trap on units with these inlets: Split/Splitless Capillary (SSL), Multi-Mode Inlet (MMI), Programmed Temperature Vaporizer (PTV), Volatiles Interface (VI). If the GC includes a Flame Ionization Detector (FID), replace the jet. If the ignitor shows any buildup of sample or corrosion, replace the ignitor. Examine the FID collector and castle assemblies for contamination clean as necessary. Zero Sensors and Leak test Zero all pressure sensors per the procedure in the 7890 Advanced User Guide. Perform inlet pressure decay test(s) as defined in the 7890 Advanced User Guide. If the PM is done in preparation for an OQ/PV, then the pressure decay test defined within that protocol can be used for the PM. Record if test passed or failed in the results table. ALS Maintenance Section NOT applicable Check all cabling and configuration settings between GC, tray, and injectors. Vacuum or removed any dust, especially around fans. Check operation of all fans. Check syringe for smooth plunger operation. Check for smooth operation of the needle support rod clean if necessary Check for correct operation of syringe volume stops. Issued: 20-Jan-2015, Revision: A.01.09 Copyright 3 2011 Agilent Technologies Page 3 of 7

7890 GC Standard Restore Instrument Restore the normal operating conditions using the Keyboard or Data System. Check and record the post PM detector signal output values. Results should be similar or lower than the detector outputs recorded prior to PM. Perform a chemical checkout. If this is a routine PM, inject the customer s sample using the ALS if applicable. This will act as a final checkout of both the ALS and the GC. Guidance If the PM service is performed prior to a qualification service, then use the qualification procedure as a guide for final instrument set up and checkout. Issued: 20-Jan-2015, Revision: A.01.09 Copyright 4 2011 Agilent Technologies Page 4 of 7

7890 GC Standard Service Review Attach available reports/printouts of all tests to this documentation. Record the PM service activity in the customer's instrument records/logbook Update/reset instrument maintenance counters as appropriate Affix the PM sticker to the system or instrument logbook based on the customer's request. Complete the Service Review Comments section below if there are additional comments Review the service and any test results with the customer. If the Instrument firmware was updated, record the details of the change in the Service Engineer's Comments box below or if necessary, in the customer's IQ records. 7890 GC Test Results Table Detector Signal Outputs Before PM service After PM service Front detector output Back detector output AUX detector output Pressure decay test Expected result Actual result or N/A Front inlet pressure decay test Back inlet pressure decay test Pass Pass Issued: 20-Jan-2015, Revision: A.01.09 Copyright 5 2011 Agilent Technologies Page 5 of 7

7890 GC Standard 7890 GC Parts List Table The following kits are recommended for capillary and purged packed inlets. If this is a general PM and the customer has a preferred set of consumables, you may use the customer's consumables. Part Description Part Number Model# where used Quantity Consumed SSL Capillary Inlet PM kit, Splitless 5188-6497 G3440A SSL Capillary Inlet PM kit, split 5188-6496 G3440A PP Inlet PM kit 5188-6498 G3440A Split vent trap PM kit, single cartridge (for MMI, PTV & VI) 5188-6495 G3440A MMI Cleaning Kit G3510-60820 G3440A PTV Septumless Head Rebuild Kit 5182-9747 G3440A PTV Septumless Head Teflon Guide 5182-9748 G3440A Ignitor (glow plug) assembly with O-ring 19231-60680 G3440A FID Collector Rebuild/Cleaning Kit G1531-67000 G3440A FID Collector Replacement Kit G1531-67001 G3440A Standard.011-inch FID Jet for capillary FID base G1531-80560 G3440A Standard.018-inch FID Jet for packed column with packed FID base 18710-20119 G3440A Standard.011-inch FID Jet for capillary column with packed FID base 19244-80560 Service Engineer Comments (optional) G3440A If there are any specific points you wish to note as part of performing the service or other items of interest for the customer, please write in this box. Issued: 20-Jan-2015, Revision: A.01.09 Copyright 6 2011 Agilent Technologies Page 6 of 7

7890 GC Standard. Other Important Customer Web Links How to get information on your product: Literature Library - http://www.agilent.com/chem/library Need to know more? - www.agilent.com/chem/education Need supplies? - www.agilent.com/chem/supplies Service Completion Service request number Date service completed Agilent signature Customer signature Document part number: G3430-90004 Issued: 20-Jan-2015, Revision: A.01.09 Copyright 7 2011 Agilent Technologies Page 7 of 7

Agilent 5977-5975-5973 Series [Standard/Interim] Agilent Preventive Maintenance provides factory recommended service for your analytical systems to assure reliable operation and the accuracy of your results. Delivered by highly-trained and certified service engineers using genuine Agilent parts and supplies, Agilent Preventive Maintenance provides everything you need to reduce unplanned downtime and keep your systems operating at their peak. For more information about Agilent Technologies services please visit our web site using the following URL http://www.chem.agilent.com/en-us/products/services/pages/default.aspx Customer Information Customers should provide all necessary operating supplies upon request of the engineer. A customer representative should be available to the engineer while performing the preventive maintenance procedures. Any parts, not included in the Parts Lists section of this document, are not part of the recommended Preventive Maintenance service, nor are they included in the price of this service. If a system requires the use of additional or special procedures and/or parts for the instrument service, then these must be ordered separately and charged as a repair, which may incur additional costs. Service Engineer s Responsibilities Only complete/printout pages that relate to the system or module being serviced. Complete empty fields with the relevant information Complete the relevant checkboxes in the checklist using a X or tick mark in the checkbox. Complete Not Applicable check boxes to indicate services not delivered, as needed Complete the PM service in the order of the tasks listed. Complete the Service Review section together with the customer Additional Instruction Notes Preventive maintenance is a factory recommended procedure designed to reduce the likelihood of electro-mechanical failures. Failure to perform preventive maintenance may reduce the long term reliability of certain instruments and systems. Two preventative maintenances (PMs) per year are recommended, the Major PM service will be performed annually with a Interim PM performed 6 months after the Major PM. Issued: 01-Dec-2012 Revision: A.02 Copyright 2012 G3870-90304 Agilent Technologies Select pages for required products

Agilent 5977-5975-5973 Series [Standard/Interim] System Information Instrument System Name/I.D: Instrument Location: Record the list of system component product numbers below. List the serial numbers of the components present in the system below. 1. 1. 2. 2. 3. 3. 4. 4. 5. 5. 6. 6. 7. 7. 8. 8. 9. 9. 10. 10. Guidance: Check box if instrument configuration report is attached instead of completing the table above. Preparation Discuss any specific issues with the customer prior to starting. Review the instrument logbook. Save instrument control settings before starting the procedure. Perform general inspection of system for cleanliness Check for proper installation of safety-related parts, assemblies, sensors etc Check for required firmware updates and verify with customers if they would like it installed. Interim Preventive Maintenance, 6 months, oil based foreline pumps Customer Responsibilities Customers should ensure that all necessary operating supplies, consumables and usage dependent items such as gases, vials, syringes, calibrant solution and solvents required for the successful preventive maintenance are available. A customer representative should be available while the preventive maintenance procedure is being performed. Important notice for customers The customer should complete the following before the Support Provider arrives on site: Perform an autotune and retain the printed tune report Parts required: 6040-0834 Rough pump fluid, 1L Issued: 01-Dec-2012 Revision: A.02 Copyright 2012 G3870-90304 Agilent Technologies Select pages for required products

Agilent 5977-5975-5973 Series [Standard/Interim] Discuss any problems the customer is having with the instrument Review customer maintenance records and exclude maintenance on recently serviced items. Review the recent autotune report. This will give a starting point for evaluating spectral peaks, baseline noise, peak shape, mass assignments and resolution. GCMS Record instrument model no. Record instrument serial no. Record current vacuum readings (if applicable): Rough Vacuum: Manifold Vacuum: Type of column installed Check manually that you have calibration peaks Vent the instrument. Perform general system inspection: Inspect vacuum hoses, pump exhaust tubing and power cords for excessive wear Look for any obvious external damage or problems. Note any obvious external damage or problems. Clean air intake (cosmetic cover may need to be removed). Verify system line voltage meets instrument specifications: Measured voltage: Wet Mechanical vacuum pumps Check for evidence of oil leakage Check pump gasket for leakage Drain and replace Mechanical pump oil Discuss with customer the need for more frequent oil changes if the oil is dirty. Demonstrate ballast, if requested. System post-check Pump system back down. Leak Check Checkout Tune Tune Evaluation Systems using Hydrogen will not be able to run a sensitivity check the same day. Units must go through a bake out overnight. If the system is a CI unit the CI verifications could be done and EI will follow the checkout of the CI. Autotune Performed [EI Only]. Issued: 01-Dec-2012 Revision: A.02 Copyright 2012 G3870-90304 Agilent Technologies Select pages for required products

Agilent 5977-5975-5973 Series [Standard/Interim] Change to the Tune context and verify that all temperatures, pressures, and gas flows reach tune method set points. Record current vacuum readings: Rough Vacuum: Manifold Vacuum: High Vacuum: Check manually that you have calibration peaks Standard Preventive Maintenance, yearly Customer Responsibilities Customers should ensure that all necessary operating supplies, consumables and usage dependent items such as gases, vials, syringes, calibrant solution and solvents required for the successful preventive maintenance are available. A customer representative should be available while the preventive maintenance procedure is being performed. Important notice for customers The customer should complete the following before the Support Provider arrives on site: Perform a positive ion autotune and negative ion autotune and retain the printed tune report Parts required: MSD Maintenance supplies Description Part number Abrasive paper, 30 µm 5061-5896 Alumina powder 393706201 Cloths, clean (package of 300) 05980-60051 Cloths, cleaning (package of 300 9310-4828 Cotton swabs (package of 100) 5080-5400 Foreline pump oil, Inland 45 6040-0834 Gloves, clean, large 8650-0030 Gloves, clean, small 8650-0029 Grease, Apiezon L, high vacuum 6040-0289 Filters and Seals Description Part number Gas Clean Filter kit GC/MS 1/8 in (complete replacement kit) CP17974 Gas Clean GS/MS Filter CP17973 Chemical Ionization Gas Purifier (CI systems) G1999-80410 Oil Mist Filter, 3/8 BSP Male Threads 5975-5973 G1099-80039 Filter cartridge,oil mist elimin, 2/pk - 5977 2710100200 Exhaust silencer kit for IDP Series - 5977 EXSLRIDP-3 Replacement filter element for IDP Series silencer - 5977 REPLSLRFILTER2 Issued: 01-Dec-2012 Revision: A.02 Copyright 2012 G3870-90304 Agilent Technologies Select pages for required products

Agilent 5977-5975-5973 Series [Standard/Interim] Replacement tip seal kit, IDP3-5977 IDP3TS Discuss any problems the customer is having with the instrument Review customer maintenance records and exclude maintenance on recently serviced items. Review the recent autotune report. This will give a starting point for evaluating spectral peaks, baseline noise, peak shape, mass assignments and resolution. GCMS Record instrument model no. Record instrument serial no. Record current vacuum readings (if applicable): Rough Vacuum: Manifold Vacuum: Type of column installed Check manually that you have calibration peaks Vent the instrument. Perform general system inspection: Inspect vacuum hoses, pump exhaust tubing and power cords for excessive wear Look for any obvious external damage or problems. Note any obvious external damage or problems. Clean air intake (cosmetic cover may need to be removed). Verify system line voltage meets instrument specifications: Measured voltage: Wet Mechanical vacuum pumps Check for evidence of oil leakage Check pump gasket for leakage Drain and replace Mechanical pump oil Discuss with customer the need for more frequent oil changes if the oil is dirty. Demonstrate ballast, if requested. Dry Mechanical vacuum pumps Uninstall existing Dry Mechanical Pump Replace Dry Mechanical pump with a replacement pump (if applicable). Replace the tip seal for the IDP3 Re-install new Dry Mechanical pump. Issued: 01-Dec-2012 Revision: A.02 Copyright 2012 G3870-90304 Agilent Technologies Select pages for required products

Agilent 5977-5975-5973 Series [Standard/Interim] Diffusion pumps Drain and replace diffusion pump oil Replace gas mist filter elements Replace gas filters nitrogen, Helium methane Remove the ion source assembly from the vacuum manifold. Disassemble and clean the ion source assembly. Follow the documented ion source assembly cleaning procedure. System post-check Pump system back down. Leak Check Checkout Tune Tune Evaluation Systems using Hydrogen will not be able to run a sensitivity check the same day. Units must go through a bake out overnight. If the system is a CI unit the CI verifications could be done and EI will follow the checkout of the CI. Autotune Performed [EI Only]. Change to the Tune context and verify that all temperatures, pressures, and gas flows reach tune method set points. Record current vacuum readings: Rough Vacuum: Manifold Vacuum: High Vacuum: Check manually that you have calibration peaks Guidance: If the PM service is performed prior to a qualification service, then use the qualification procedure as a guide for final instrument set up and checkout. Issued: 01-Dec-2012 Revision: A.02 Copyright 2012 G3870-90304 Agilent Technologies Select pages for required products

Agilent 5977-5975-5973 Series [Standard/Interim] Service Review Attach available reports/printouts of all tests to this documentation. Record the PM service activity in the customer s instrument records/logbook Update/reset instrument maintenance counters as appropriate Affix the PM sticker to the system or instrument logbook based on the customer s request. Complete the Service Review Comments section below if there are additional comments Review the service and any test results with the customer. If the Instrument firmware was updated, record the details of the change in the Service Engineer s Comments box below or if necessary, in the customer s IQ records. Agilent 5977 Series Test Results Table: Test Description Expected Test Result Actual Test Result Agilent 5977 Series Parts List Table: ο Section NOT Applicable Part Description Part Number Product/Model # where used Quantity Consumed Issued: 01-Dec-2012 Revision: A.02 Copyright 2012 G3870-90304 Agilent Technologies Select pages for required products

Agilent 5977-5975-5973 Series [Standard/Interim] Service Engineer Comments (optional): If there are any specific points you wish to note as part of delivering the PM service including any follow-up activities, specific observations made or other items of interest for the customer, please write in this box. Other Important Customer Web Links How to get information on your product: Literature Library - http://www.agilent.com/chem/library Need to know more? - www.agilent.com/chem/education Need technical support? - www.agilent.com/chem/techsupp Need supplies? - www.agilent.com/chem/supplies Service Completion Service Request number. Agilent Signature Date service completed Customer Signature.. Total no. of pages for this document:. Issued: 01-Dec-2012 Revision: A.02 Copyright 2012 G3870-90304 Agilent Technologies Select pages for required products