Experiment No.3: Flow through orifice meter. Background and Theory

Similar documents
Experiment (4): Flow measurement

Fundamentals of Natural Gas Measurement

FUNDAMENTALS OF ORIFICE METERING Ken Embry FMC Measurement Solutions

Measurement of thermocouple emf using potentiometer & milivoltmeter. Milivoltmeter consists of a PMMC type meter. The coil gets magnetised when the

Applied Fluid Mechanics

FLUID FLOW. Introduction

The Discussion of this exercise covers the following points:

CHAPTER 1: INSTRUMENTATION EQUIPMENT MODULE 5: Flow Instrumentation

ORIFICE PLATE flow meters

Exercise 4-1. Flowmeters EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION. Rotameters. How do rotameter tubes work?

ECH 4224L Unit Operations Lab I Fluid Flow FLUID FLOW. Introduction. General Description

The calculation of fluid flow

Low-Cost Pipeline Flow Meter

STEALTH INTERNATIONAL INC. DESIGN REPORT #1001 IBC ENERGY DISSIPATING VALVE FLOW TESTING OF 12 VALVE

CFD Analysis and Comparison of Fluid Flow Through A Single Hole And Multi Hole Orifice Plate

Verabar Velocity Averaging Flow Sensors. True Performance in Flow Measurement

Standards and wall thickness

Familiarize yourself with the pressure loss phenomenon. The Discussion of this exercise covers the following point:

VERIS VERABAR VELOCITY AVERAGING FLOW SENSORS TRUE PERFORMANCE IN FLOW MEASUREMENT

HYDRAULICS. H89.8D - Hydraulic Bench

FLUID FLOW Introduction General Description

UNIT - 3 Friction and Belt Drives

flow measurement solutions RIM10 series rotor insertion flowmeters

Impacts of Short Tube Orifice Flow and Geometrical Parameters on Flow Discharge Coefficient Characteristics

EMCO Primary Elements for Flow Measurement

JUMO flowtrans DP R. Meter run R01 and orifice plate R02. Applications. Brief description. Special features. Customer benefits.

Cross Flow Heat Exchanger H352

Fluid Flow Conditioning

YARWAY NARVIK VEN-TEMP DESUPERHEATER MODEL 25

VORTAB FLOW CONDITIONERS. For Flow Meters, Pumps and Other Flow Profile Critical Equipment

COMPUTATIONAL FLOW MODEL OF WESTFALL'S 2900 MIXER TO BE USED BY CNRL FOR BITUMEN VISCOSITY CONTROL Report R0. By Kimbal A.

Model CO Compact Orifice for use with Foxboro Differential Pressure Transmitters

MECHANICAL EQUIPMENT. Engineering. Theory & Practice. Vibration & Rubber Engineering Solutions

Orifice plates and nozzles. Flow measurement using differential pressure

Yarway Narvik Ven-Temp Desuperheater Model 25

YARWAY NARVIK MODEL 88 SPID SMALL PIPE INLINE DESUPERHEATER

CONTROL VALVE SELECTION AND SIZING (ENGINEERING DESIGN GUIDELINE)

Air Cylinders Drive System Full Stroke Time & Stroke End Velocity. How to Read the Graph

The EPSILON is based on the principle of measuring velocity in the pipe line, therefore the flow measurement is volumetric.

ORIFICE PLATES, FLANGES FLOW NOZZLES, VENTURI TUBES AND ACCESSORIES

FLUID CONTROL RESEARCH INSTITUTE

Comparison between Fluid Viscous Dampers and Friction Damper Devices. Fluid Viscous Dampers (FVD) Friction Damper Device (FDD) Working principle:

FINAL REPORT MARCH 2008

WINTER 14 EXAMINATION Subject Code: Model Answer Page No: 1/20

FLOW CONTROL THROUGH VORTEX SHEDDING INTERACTION OF ONE CYLINDER DOWNSTREAM OF ANOTHER. Jonathan Payton 1, and *Sam M Dakka 2

IFOA Integral Flow Orifice Assembly

Air Flow Measurement Technologies

Control Valves. Topic Highlights. By James Reed. Valve Types Standards and Codes Valve Selection Operation Actuators and Accessories. 4.

Silencers. Transmission and Insertion Loss

Series HCD Airfoil Control Damper

FEASIBILITY STYDY OF CHAIN DRIVE IN WATER HYDRAULIC ROTARY JOINT

Lecture 27: Principles of Burner Design

The Performance Optimization of Rolling Piston Compressors Based on CFD Simulation

Rosemount 285 Annubar Primary Element Series

Heat Transfer Enhancement for Double Pipe Heat Exchanger Using Twisted Wire Brush Inserts

APPLYING VSO OPTION TO SOLUTION AIR HANDLING UNITS

Accutubes. Accuracy True double averaging design +1% of rate accuracy +0.1% repeatability Tested and certified by independent flow labs

STEEMCO-MAS Flowmeter for Steam

Analysis of Exhaust System using AcuSolve

Piston spool valves and poppet valves - A technical comparison of available solenoid valves

A Review of Flowmeters for Water Applications

INSTALLATION AND OPERATION

FUNDAMENTALS OF INSERTION TURBINE METERS Les Bottoms Thermo Electron Corporation, Flow Systems

PRINCIPLES OF OPERATION

Safety cap. Measuring probe. Seal. Operating principle. Advantages of balanced circuits. Balanced circuits have the following principal benefits:

Variable Area Flowmeters Ori-Flowrator TM Meters

High CATALOG. Efficiency. Pilot Operated Safety Valves Series 810 Pop Action Series 820 Modulate Action. The-Safety-Valve.com

Learn how rollers and web handling affect your 100% inspection system s performance

AN EXPLANATION OF CIRCUITS CARTER YH HORIZONTAL CLIMATIC CONTROL CARBURETER

Burner Settings. How to Set and Adjust Burners Why It s Necessary

1 Final Control Elements

The Multivariable Air Flow Meter

Fundamental Training. Flow Con t

White paper: Originally published in ISA InTech Magazine Page 1

Flow nozzle for in-pipe installation, model FLC-FN-PIP Flow nozzle for flange assembly, model FLC-FN-FLN Venturi nozzle, model FLC-FN-VN

Cross Flow Heat Exchanger H352

Piping Systems. J. David Bankston, Jr., and Fred Eugene Baker*

Square Edged Concentric

The Multivariable Air Flow Meter

ORIFICE PLATE METER DIAGNOSTICS

ORIMAS Mass Flowmeter

FlameGuard. Fire Retardant FRP Duct/Pipe. FlameGuard Fiberglass Reinforced Pipe & Fittings

SCHOOL OF COMPUTING, ENGINEERING AND MATHEMATICS SEMESTER 2 EXAMINATIONS 2013/2014 ME110. Aircraft and Automotive Systems

CENTRIFUGAL PUMP: Parallel and Series Operation 11/11/02

Gauges, Sight Glasses and Vacuum Breakers

Air Cylinders Drive System Full Stroke Time & Stroke End Velocity. How to Read the Graph

EXPERIMENTAL INVESTIGATION OF THE FLOWFIELD OF DUCT FLOW WITH AN INCLINED JET INJECTION DIFFERENCE BETWEEN FLOWFIELDS WITH AND WITHOUT A GUIDE VANE

Laboratory structure 1 Base Service Units. CAI. Computer Aided Instruction Software System. INS/SOF. Instructor Software +

Field IT Differential Pressure Flow Elements

Use these modules to gain valuable knowledge about STIHL policies, procedures and products that will be a benefit to you on the job immediately.

5. Combustion of liquid fuels. 5.1 Atomization of fuel

Fluid Power System Model-Based Design. Energy Efficiency. Fluid Power System Model-Based Design Energy Efficiency. K. Craig 1

Gas exchange process for IC-engines: poppet valves, valve timing and variable valve actuation

INSTRUMENTATION CONTROL VALVE SIZING AND SELECTION (ENGINEERING DESIGN GUIDELINE)

Heat Exchangers (Chapter 5)

Fundamentals of Turbine Meters. Paul Honchar Sensus Metering Systems

STREAMLINE LAMINAR FLOW UNIFORM (AXISYMMETRIC) R = 3160 x Q x G t

SmartCone Differential Pressure Flow Meter User Manual

TRANSMITTER ATTACHES TO THIS SURFACE

For installation into ductwork

Transcription:

Experiment No.3: Flow through orifice meter Background and Theory Flow meters are used in the industry to measure the volumetric flow rate of fluids. Differential pressure type flow meters (Head flow meters) measure flow rate by introducing a constriction in the flow. The pressure difference caused by the constriction is correlated to the flow rate using Bernoulli's theorem. If a constriction is placed in a pipe carrying a stream of fuid, there will be an increase in velocity, and hence an increase in kinetic energy,at the point of constriction. From an energy balance as given by Bernoulli s theorem, there must be a corresponding reduction in pressure. Rate of discharge from the constriction can be calculated by knowing this pressure reduction, the area available for flow at the constriction,the density of the fluid and the coefficient of discharge C d. Coefficient of discharge is the ratio of actual flow to the theoretical flow and makes allowances for stream contraction and frictional effects. Venturi meter, orifice meter, and Pitot tube are widely used head flow meters in the industry. The Pitot-static is often used for measuring the local velocity in pipes or ducts. For measuring flow in enclosed ducts or channels, the Venturi meter and orifice meters are more convenient and more frequently used. The Venturi is widely used particularly for large volume liquid and gas flows since it exhibits little pressure loss. However, for smaller pipes orifice meter is a suitable choice. In order to use any of these devices for measurement it is necessary to empirically calibrate them. That is, pass a known volume through the meter and note the reading in order to provide a standard for measuring other quantities. Orificemeter An orifice meter is a differential pressure flow meter which reduces the flow area using an orifice plate. An orifice is a flat plate with a centrally drilled hole machined to a sharp edge. The orifice plate is inserted between two flanges perpendicularly to the flow, so that the flow passes through the hole with the sharp edge of the orifice pointing to the upstream. The relationship between flow rate and pressure drop can be determined using Bernoulli s equation as:

(1) where, Q is the volumetric flow rate, A o is the orifice cross sectional area, p 1 and p 2 are the pressure measured at the upstream and downstream and C d is the discharge coefficient for the orifice. β is the ratio of orifice diameter to the pipe diameter= where is the diameter of the orifice and is the pipe diameter. The fluid contracts and then expands as it moves through the orifice and this results in a pressure drop across the orifice, which can be measured. The magnitude of the pressure drop can be related to the volumetric flow rate. An orifice in a pipeline is shown in figure 1 with a manometer for measuring the drop in pressure (differential) as the fluid passes through the orifice. The minimum cross sectional area of the jet is known as the vena contracta.

How does it work? As the fluid flows thtough the orifice plate the velocity increases, at the expense of pressure head. The pressure drops suddenly as the orifice is passed. It continues to drop until the vena contracta is reached and then gradually increases until at approximately 5 to 8 diameters downstream a maximum pressure point is reached that will be lower than the pressure upstream of the orifice. The decrease in pressure as the fluid passes thru the orifice is a result of the increased velocity of the fluid passing through the reduced area of the orifice. When the velocity decreases as the fluid leaves the orifice the pressure increases and tends to return to its original level. All of the pressure loss is not recovered because of friction and turbulence losses in the stream. The pressure drop across the orifice increases when the rate of flow increases. When there is no flow there is no differential. The differential pressure is proportional to the square of the velocity, it therefore follows that if all other factors remain constant, then the differential pressure is proportional to the square of the rate of flow. Following types of pressure taps can be located for differential pressure measurement:

Corner: pressure taps one each on the upstream and downstream flanges. Radius taps:one pipe diameter upstream side and one and a half pipe diameter on the downstream side. Pipe taps: Upstream side- 2.5pipe diameters and downstream side- 8 pipe diameters. Flange taps: 1 upstream side and 1 downstream side. The analysis of the flow through a restriction (Figure 2) begins with assuming straight, parallel stream lines at cross sections 1 and 2, and the absence of energy losses along the streamline from point 1 to point 2.

The objective is to measure the mass flow rate(,. By continuity (2) Bernoulli s equation may now be applied to a streamline down the centre of the pipe from a point 1 well upstream of the restriction to point 2 in the vena contracta of the jet immediately downstream of the restriction where the streamlines are parallel and the pressure across the duct may therefore be taken to be uniform: = (3) assuming that the duct is horizontal. Combining (3) with (2) gives For a real flow through a restriction, the assumptions above do not hold completely. Further, we cannot easily measure the cross-sectional area of the jet at the vena contracta at cross-section 2 where the streamlines are parallel. These errors in the idealised analysis are accounted for by introducing a single, cover all correction factor, the discharge coefficient, C d, such that

Coefficient of discharge for a given orifice type is a function of the Reynolds number (N Reo ) based on orifice diameter and velocity,and diameter ratio β. At Reynolds number greater than about 30000,the coefficients are substantially constant and independent of β.for square edged or sharp edged concentric circular orifices, the value will fall between 0.595 and 0.62 for vena contracta or radius taps for β upto 0.8 and for flange taps for β upto 0.5.

Coefficient of discharge for square edged circular orifices with corner taps[tuve and Sprenle Instruments(1933)] In summary, the principal advantages of the orifice plate are it is simple and robust standards are well established and comprehensive plates are cheap may be used on gases, liquids and wet mixtures (eg steam) Its principal drawbacks are low dynamic range: maximum to minimum mass flow rates only 4:1 at best performance changes with plate damage or build up of dirt.

affected by upstream swirl large head loss. Because of the large friction losses from the eddies generated by the reexpanding jet below the vena contracta,the pressure recovery in an orificemeter is poor.for a value of β equal to 0.5, the lost head is about 73% of the orifice differential. Comparison of permanent head loss caused by different head meters

Orifice plates and flanges