L-K INDUSTRIES BENCHMARK S MANUAL

Similar documents
L-K INDUSTRIES. Criterion Water Bath

Centrifuge Operator / Service Manual

Bio Lion Table Top Centrifuge XC-L5

Bio Lion Table Top Centrifuge XC-H165

Z100A Clinical Centrifuge

8 Light Controller. Instruction Manual. With Light Timer 240 Volts. Product # INNOVATING SINCE 1995

230VAC Power Inverter 400W Owner s Manual

INSTRUCTION MANUAL FOR THE CLINICAL 50 CENTRIFUGE

Operation and Instruction Manual

Nature Power Inverters. True Sinewave Inverter Modified Sinewave Inverter. Owner s Manual

Power Inverter 400 MW Owner s Manual

3000 SERIES DRY-BLOCK HEAT SOURCE

200 Shadylane Drive Philipsburg, PA Phone: (814) Fax: (814) Service Manual. Model 642E Centrifuge

Owner s Guide. ca6154

Hydraulic Immediate Need Power Pack

ATD Gallon Pressurized Oil Drain Owner s Manual

Heated Bakery Merchandiser Series. Series: WBC Operation Manual

Installation and Service Manual M² Sync Room Slideout System without Room Lock Connectors on Control Box

will result in death or serious injury. could result in death or serious injury. Indicates a hazardous situation which, if not avoided,

XCM12 Mini-Centrifuge

MODEL 660 AUTOMATIC FASTENING CENTER OPERATOR S MANUAL

OPERATIONS MANUAL. Mini Centrifuge Model MCF Certified

12 Volt Heavy-Duty Air Inflator

Owner s Guide CA 5055

Owner s Guide CARS & CA4B5

Helios. 24 Light 240 Volt Controller. PRODUCT # Instruction Manual.

NANOPAC-300 & 500 Power Supply. Instruction manual NANOPAC-300 & NANOPAC-500

Toshin-Cho Itabashi-ku Tokyo, Japan Tel Fax /06

Centrifuge. Model XC-1000 Operating Manual

8460SRC Replacement Chassis for Scully ST-6 and Biclops Monitors

CA 5054 Owner s Guide

1000-lb Hydraulic Truck Crane

Dip Tester Model DS-03

Instruction Manual. Labnet Spectrafuge 16M Microcentrifuge

Owner s Guide. ca5154

Owner s Guide. ca5354

Transfer Switch GPTS 30

Operator s Safety and Service Manual

Vetlab CombiSpin Centrifuge

Model 20 Constant Speed Blender : 115 Volt : 230 Volt. Instruction Manual. Updated 11/14/2017 Ver. 4.0

Load & Lock Cremators

Model : LHT-002A Please read this manual carefully before using the instrument Labnics Equipment

USER GUIDE TURBOCORD TM PORTABLE CHARGER 240V. AeroVironment EV Solutions

materials and workmanship for 2 years. Should the centrifuge require warranty or Phone: or Fax:

SIP Direct Drive Oil-Lube Air Compressors - Operating & Maintenance Instructions

200 Shadylane Drive Philipsburg, PA Phone: (814) Fax: (814) Service Manual

CRD600 Automatic Fitting Inserter

Owner s Guide PR1BZ PR1BZLR

READ AND SAVE THESE INSTRUCTIONS. High Velocity Restaurant-Duty Utility Set Belt Driven for Roof Mounting

10 Year Limited Warranty

CLEAN ROOM DEVICES, LLC "WHERE TUBING AND FITTINGS COME TOGETHER"

Cooling Module. Instruction Manual

Instruction Manual for the Revolutionary Science RS-200 RevSpin Microcentrifuge

IAQ-CALC INDOOR AIR QUALITY METER MODEL 7515

POWER PINNER RAPID FIRE 7005 RF OPERATOR S MANUAL

POWER PINNER CD HAND WELDER 7300 OPERATOR S MANUAL

Safety, Installation and Service Manual Models 1952, 1953 & 1972

Power. On Your Terms.

Use and Care Guide. Part # SC SV-OS-BZ SC SV-OS-W Model #

SAVE THESE INSTRUCTIONS

100 Gallon 4- Wheel Sprayer

Refinery Supply Company, Inc.

Owner s Guide CA 4555

BT403. A Geno Technology, Inc. (USA) brand name. BT-300 Power Supply. Cat. No. BT

ROTARY BRUSH CUTTERS THE LEADER OF THE PACK OWNER/OPERATOR SAFETY & INSTRUCTION MANUAL

CRD610 Automatic Fitting Inserter

Pump Sentry. Models 812 PS & 1612 PS INSTALLATION INSTRUCTIONS

Model Combustible Gas Sample Draw Detector Head Operator s Manual

Operator's Manual. Storage System. Ultrasound Probe Cabinet. Manufactured by:

AQUARIUMVAC.COM TOBOQ AQUARIUM WATER PUMP & VACUUM. User Manual V 1.0

FD Conveyor for FD 260 Tabber


OWNER'S MANUAL. Royal Sovereign International Inc. RET Desk Series

ITS-50R TRANSFER SWITCH OWNER S MANUAL

Transfer Switch TS-50. Owner s Manual

PROwatt 150 DC to AC Power Inverter. Owner s Guide

Manual Chain Hoist. Owner s Manual

ProFlo FatBoy

100 Gallon Skid Sprayer

PrO-ASTM Centrifuges. Centurion. C1015 Micro Prime Centrifuge. A Micro that offers: Oil testing. Petroleum testing, ASTM methods

GF1900 SULFURYL FLUORIDE MONITOR

110 Volt/12 Volt Portable Inflator

Model Combustible Gas Sample Draw Detector Head Operator s Manual

GFM-40ac COMPLIANCE WESTUSA. Ground Circuit Tester. Instruction Manual

READ AND SAVE THESE INSTRUCTIONS. Centrifugal Downblast Exhaust Fan Belt Driven for Roof & Wall Mounting

Owner s Manual Drill and Light Weight Portable Utility Pumps

R & D SPECIALTIES ROTROL I USER'S MANUAL

OPERATION MANUAL MODEL MODEL MODEL MODEL MODEL NON-CONDUCTIVE ARTICULATING FIBERSCOPE INSTRUMENT TECHNOLOGY, INC.

60 Watt Industrial LED Low Bay Light

Please read this manual carefully before operating! Unpacking, assembly, and operating videos are available at

USER MANUAL TMT-PH200 IR PREHEATER.

HD18F2M1.

Model H30 Operation Manual

Utility Distribution Systems Installation, Operation, and Maintenance Manual

Blue Air. Commercial Refrigeration Inc. Installation & Operation Manual Glass Door Countertop Refrigerator

WARNINGS, CAUTIONS AND NOTES

6900-( )-( ) HIGH SPEED SANDWICH PRESS

Owner s Manual & Safety Instructions

Operating Instructions for the. In Line Conveyor System West Cabot Boulevard Langhorne, PA w w w.empire-airblast.

Model NTX7 Series Automatic Battery Charger User s Manual Rev. 1.0 October 17, 2006

Transcription:

L-K INDUSTRIES Demand a higher standard. BENCHMARK S MANUAL Heated Laboratory Centrifuge To set the standard for MANUFACTURING QUALITY and customer service in the PETROLEUM TESTING industry. Petroleum Testing Equipment Suppliers Worldwide Since 1930

Limited Product Warranty L-K Industries warrants its manufactured products against defects in materials and workmanship for a period limited to one year from the date of shipment. If purchased from a Distribution Partner, the warranty lasts one year from the in-service date. During the one-year warranty period, L-K Industries shall repair or replace defective equipment free of charge. L-K Industries shall only be liable for repairs or replacements if L-K is contacted immediately following discovery of defect(s). Defective products (under warranty) shall only be returned after contacting and receiving permission from L-K Industries. The warranty does not extend to L-K Industries products that have been misused, neglected, independently modified (without L-K Industries' approval), improperly installed or accidentally damaged. L-K Industries shall not be liable for damage or loss resulting from use of L-K Industries products separately or in combination with other equipment. Contact Information Main Line: 24-Hour Tech Helpline: Fax: Email: Billing & Shipping Address: Website: 713-926-2623 832-588-6369 713-926-7736 sales@lk-ind.com 6952 Lawndale St, Houston, TX 77023 www.lk-ind.com Page. 2

Table of Contents GETTING STARTED Introduction and Safety Precautions Installation Electrical Connection 5 6 7 OPERATION Loading and Powering On Heated Operation Motor Operation Timer Operation 8 9 9 9 MAINTENANCE Replacement Parts & Repair Services Housekeeping Replacement of Fuses 10 10 10 TROUBLESHOOTING Common Issues Motor Lid and Door Assembly 11 11 11 APPENDICES Appendix A: RCF/RPM Conversions** Appendix B: Wiring & Assembly Diagrams Appendix C: Programming Overviews 12 12 12 Page. 3

Getting Started Introduction and Safety Precautions The Benchmark S Heated Centrifuge is intended for general use in test laboratories and refineries, where Class I, Division 2 certification is not required. It operates on 120/240VAC, and comes completely assembled and ready for installation. A power cord can be purchased separately if the unit will not be hard-wired. It is recommended to keep the custom packaging materials for future shipping or storage. For best results, the centrifuge should be housed in a controlled environment. Exposure to solvents (i.e. toluene) may degrade the sealing properties of materials used in the front panel switches. This centrifuge is designed to minimize operator injury and includes a lockable sliding door for safe operation. Carefully following safety guidelines and knowing the limits of this unit will help prevent accidents. REMOVE all foreign objects from the centrifuge bowl. Keep the sliding door CLOSED while samples are spinning. NEVER try to slow the trunnion arm down manually. Never place an object inside the bowl while the trunnion arm is spinning. Allow the trunnion arm to come to a complete stop before removing samples. Turn all switches OFF when not in use. NEVER leave the centrifuge unattended while it is turned on or operating. DISCONNECT the power supply before removing or replacing electrical or mechanical parts. DO NOT leave oily or combustible materials in or around the centrifuge. DO NOT allow unauthorized persons to access the centrifuge. BE AWARE of surroundings. DO NOT operate the unit when fatigued or under the influence of medication, alcohol or illegal substances. DO NOT operate this machine with unbalanced tubes and shields. If a tube has broken inside of a shield, the cushion MUST be replaced. Use CAUTION during heated operation, as the top of the unit is hot to the touch. Page. 4

Getting Started Installation Secure the centrifuge to its operating surface using the hardware supplied, as shown in figure 1-1. Tighten each nut until there is sufficient compression of the rubber shock mount. Allow for optimum working clearance around the unit. Ensure the unit is level by adjusting the shock mounts. Figure 1-1 Schematic of Mounting Bolts 1/4-20 BOLT (4PLACES) RUBBER SHOCK MOUNT (4PLACES) 1/4-20 LOCKING NUT (4 PLACES) Ensure that all collars, cushions and shields are installed properly and in the correct orientation, as shown in figure 1-2. Figure 1-2 A/B/C/D Cushion Orientation CUSHION NOT USED A MODELS B MODELS C MODELS D MODELS Page. 5

Getting Started Electrical Connection If the unit will be hard-wired, remove the access panel on the back of the unit to expose wires. Feed the supply voltage wire through the conduit hub and attach according to local electric code. Replace the back cover. Hard-wiring shall be done by a licensed commercial electrician per local electric code. Wiring details are shown in figure 1-3. Figure 1-3 115VAC/220VAC Model Wiring Diagrams 115VAC MODELS 220VAC MODELS, (European Standards Only) CONDUCT HUB 115VAC LINEIN (BLACK) BACK OF CENTRIFUGE GROUND (GREEN) NEUTRAL (WHITE) CONDUCT HUB 220VAC LINEIN (RED) BACK OF CENTRIFUGE GROUND (GREEN) NEUTRAL (WHITE) If the unit is not hard-wired, a power cord with the appropriate connections shall ONLY be purchased through the L-K Industries warehouse. In geographical regions where power load is prone to fluctuation, power conditioning is recommended to filter out reductions and spikes in power. Failure to do so may void the warranty on the unit. Contact L-K Industries for power/line conditioner recommendations (hard-wire and plug-in options). Page. 6

Operation This unit and all L-K Industries sample tubes are designed and fabricated in compliance with API & ASTM standards. For optimal results, L-K Industries sample tubes (along with matching shields, cushions and collars) should be used. Verified and certified tubes are also available. Before centrifugation, all the tubes, collars, cushions and shields should be visually inspected for damage. Check for oil residue on cushions and ensure that the cushions rest flat inside the bottom of each shield. Damaged parts must not be centrifuged, as they may break and cause further damage to the centrifuge. Figure 2-1 Front Panel Switches 4 PLACE TRUNNION HEAD FOR MODEL A B C D TUBES MOTOR WILL NOT RUN IF DOOR IS OPEN POWER ON PUSH BUTTON POWER ON INDICATOR LAMP EMERGENCY POWER OFF PALM BUTTON TEMPRATURE READOUT ELECTRONIC TECHOMETER WITH LCD READOUT. LCD SHOWS ACTUAL MOTOR SPEED. NOT THEOROTICAL. MOTOR RUN TIMER MOTOR STOP PUSH BUTTON VARIABLE SPEED MOTOR CONTROL TEMPRATURE SETTING HEAT ON/OFF SWITCH MOTOR RUN PUSH BUTTON Loading and Powering On To supply power to the unit, turn the MAIN DISCONNECT switch to the right. The green indicator light should signify that power is on. Visually verify that the EMERGENCY MOTOR OFF SWITCH is not engaged. Open the sliding door and place the filled sample tubes into the shields with collars resting at the top of the shields. Ensure that the tip of each tube rests firmly in the cushion. Inspect inside the bowl for foreign objects and close the door. To start the motor, press the PUSH TO RUN button. To turn off the unit in an emergency, push the EMERGENCY MOTOR OFF SWITCH. Page. 7

Operation Heated Operation For faster performance, preheat the unit by turning the HEAT POWER ON switch to the right and setting the temperature controller to the desired operating temperature. In some conditions, it may be necessary to set the operating temperature approximately 15-20 degrees above the desired sample temperature (maximum 160⁰F). Dry block or water bath heaters may be purchased separately to preheat samples. The temperature probe is located underneath the lid, above the centrifuge bowl. The temperature display indicates the temperature inside the bowl, NOT the samples. To determine sample temperature, a thermometer must be placed inside each sample tube once a test is complete. Thermometers may be purchased separately. Motor Operation Press the PUSH TO RUN button. Verify the motor RPMs in the display, which shows actual RPMs (not theoretical). Turn the SPEED ADJUST control knob to the desired speed (see Appendix A for various RCF values). To prolong centrifuge life, use the lowest feasible speed setting. The maximum allowable speed is 2000 rpm. DO NOT attempt to run the motor with the door open, as the unit is designed to shut off when the door is opened. When the RPM INDICATOR reads 0, wait a few seconds to open the door. DO NOT force the door open. A locking mechanism automatically engages the when the motor is turned on and stays engaged until the motor stops spinning. Verify that the trunnion arm has come to a complete stop. Remove samples and close the door. Keep the door closed at all times, as this minimizes heat loss. Contact the L-K Industries Tech Helpline for instructions for manually bypassing the locking mechanism. The manual bypass should ONLY be done to verify centrifuge speed. Timer Operation The Countdown Timer Display indicates the amount of remaining time on a test. If the countdown time must be changed, the controller must be reprogrammed. The controller will prompt for a code to change the program. Enter code 055. See Appendix C for the Timer Quick Programming Guide. Otherwise, contact the L-K Industries Tech Helpline for timer programming instructions NOTE: On initial startup or after any programming changes, reset the unit before beginning operation. Page. 8

Maintenance Replacement Parts & Repair Services Replacement parts for all centrifuges can be procured through the L-K Industries warehouse (see http://lk-ind.com/centrifuge/benchmarks.html for unit model number, serial number and replacement part numbers) or from a Distribution Partner. L-K Industries also repairs and rebuilds all centrifuge models. For estimates and repairs, first obtain a RMA (return merchandise authorization) number from L-K Industries. Then ship the unit to L-K Industries with a description of any problems and include billing information (unless under warranty), return shipping address and a main contact. Housekeeping Clean the unit after each test. Oil residue can build up and cause difficulty operating the unit. Periodically check the sample tube cushions inside the trunnion shields. Before each test, examine moving parts for wear and stress. Replace parts as necessary. If a tube has broken inside a shield, the cushion MUST be replaced. Glass particles can become embedded in the cushions and cause future breakage of tubes. Two extra cushions are provided. NOTE: B style tubes DO NOT require cushions or collars. Replacement of Fuses There are five fuses located behind the front panel of the centrifuge. These protect from over-current due to voltage spikes or other incidents. If replacement is necessary, disconnect power to the centrifuge. Open the front panel by removing screws and swinging it downward. Replace fuses with 250V slow-blow or fast-acting ceramic fuses that are CSA/UL certified, as indicated in figure 3-1. One extra slow-blow fuse and one fast-acting fuse are provided. Contact the L-K Industries Tech Helpline if assistance is needed in replacing fuses. Figure 3-1 Fuse Types FUSE HOLDER FAST ACTING 5A SLOW-BLOW 5A SLOW-BLOW 20A SLOW-BLOW 20A SLOW-BLOW 20A Page. 9

Troubleshooting Common Issues Issues that may occur during centrifuge operation include excessive vibration, broken tubes, broken or malfunctioning LCD screens, cushions stuck in shields, etc. For issues that cannot be identified and resolved by the user, more technical troubleshooting details can be found on the L-K Industries website. Alternatively, contact the L-K Industries Tech Helpline for assistance. Motor If the motor does not run when the PUSH TO RUN button is pushed, make sure the power to the motor is on (green indicator light is on) and the sliding door is completely closed. In addition, check the fuses in the front panel. Contact the L-K Industries Tech Helpline if assistance is needed in checking fuses. If using the Countdown Timer during operation, make sure there is time remaining. Once the timer runs out, the motor will automatically shut down. If the timer is not being used (i.e. for tests longer than the typical preset time), turn the timer to the OFF position. If the motor does not run, ensure that the door is properly closed. The proximity switch uses a sensor to determine whether the door is closed and closes a circuit allowing power to feed to the motor. When the motor controller is functioning properly, the green indicator light will be lit. If the green light does not turn on when power is supplied properly, there are a number of possibilities for the error. Contact the L-K Industries Tech Helpline for troubleshooting fault codes from the motor controller. Lid and Door Assembly The following are possible sources of malfunction of the lid and door locking mechanism: Sliding Door Add grease for smoother opening and closing. Proximity Switch & Sensor These work together to detect door position and allow power to the motor; failure of either prevents the motor from running. Magnet This is mounted inside the sliding door and aligns with the sensor when the door is closed; can become detached. Solenoid & Locking Pin These work together to lock the door during motor operation; locking pin may be misaligned with the door. Contact the L-K Industries Tech Helpline to troubleshoot problems with the lid and door assembly. Page. 10

Appendices Appendix A: RCF/RPM Conversions** TUBE M1ODEL/ DIST. TIP-TO-TIP RPM* @500 RCF RPM* @600 RCF RPM* @700 RCF RPM* @800 RCF A /16.00 1480 1620 1750 1870 B /17.75 1410 1540 1660 1780 C /19.25 1350 1480 1600 1710 D /18.50 1380 1510 1630 1740 *RPM =265 (RCF/D), where D = tip-to-tip tube diameter in inches **ASTM minimum recommended RCF: 600 Appendix B: Wiring & Assembly Diagrams The following schematics may be found at www.lk-ind.com/downloads 115VAC and 220VAC Wiring Outer Front Panel Schematic Inner Front Panel Schematic Motor Assembly Bowl Heater Assembly Bowl Assembly Door and Locking Mechanism Assembly Front Panel and Lid Assembly Appendix C: Programming Overviews The following programming diagrams may be found at www.lk-ind.com/downloads RPM Quick Programming Timer Quick Programming Page. 11

L-K INDUSTRIES is proud to work with Amazon: www.amazon.com Cole-Parmer: www.coleparmer.com DC Scientific: www.dcscientific.com) Grainger: www.grainger.com) Midland Scientific: www.midlandsci.com) Parkes Scientific: www.parkesscientific.com) Catalog 2016 L-K Industries All Rights Reserved