TIC 300 PRO. Users Manual Mode d emploi Bedienungshandbuch Manual d uso Manual de uso. Non Contact AC Voltage Detector. High Energy Tic Tracer

Similar documents
PY-11 Voltage & Continuity Tester

TIME-10 Process Control Stopwatch

TACH-10 Contact / Non-Contact Tachometer

RS-3 PRO RS-1007 PRO. CAT IV Analog Clamp meter Series. Users Manual. For detailed specifications and ordering info go to

LM-100 LM-120. Light Meters. Users Manual. For detailed specifications and ordering info go to

TX900. Microwave Leakage Detector. 99 Washington Street Melrose, MA Phone Toll Free

PRM-4. Phase Sequence and Motor Rotation Tester. Users Manual. For detailed specifications and ordering info go to

INSP-3 Wiring Inspection Tester

KMP Insulation-Continuity Tester. Users Manual

TMA5 Mini Vane Anemometer

IR-500 Infrared Thermometer / Voltage Detector / Flashlight

THWD-3 and TH-3 Relative Humidity and Temperature Meters

TMULD-300 ULD-300. Ultrasonic Leak Detector. Users Manual. 99 Washington Street Melrose, MA Phone Toll Free

KMP7010. Ground Resistance Tester. Users Manual

LH41A. User Manual. Clamp On Ammeter. For detailed specifications and ordering info go to

CM100. Carbon Monoxide Meter. 99 Washington Street Melrose, MA Phone Toll Free

CO2-100 CO2 Meter. Users Manual Mode d emploi Bedienungshandbuch Manual d Uso Manual de uso

Operator s Manual TC-253A. Temperature Converter

AMB V Insulation Resistance Tester. Users Manual. For detailed specifications and ordering info go to

THWD-5. Relative Humidity Temperature Meter. Users Manual. For detailed specifications and ordering info go to

AC50A. Leakage Clamp Meter. User Manual. For detailed specifications and ordering info go to

362 Clamp Meter. Users Manual

i3000s/i2000 Flex AC Current Probe

TMD90A Dual Input Digital Thermometer

1662/1663/1664 FC. Electrical Installation Tester. Safety Information. Warnings

Users Manual. Phase Rotation Indicator

323/324/325 Clamp Meter

362 Clamp Meter. Users Manual

The following table shows the symbols used on the product and/or in this manual. Hazardous Voltage. Risk of electric shock.

373 Clamp Meter. Users Manual

233 True-rms Remote Display Digital Multimeter Safety Information

1587 FC/1587/1577. Insulation Multimeter. Safety Information. Warnings

302+/303 Clamp Meter. Users Manual

365 Detachable Jaw True-rms Clamp Meter

Instruction Manual VT4 VOLTAGE TESTER. Check out what else you can get from Martindale:

Instruction Manual PC104/PC105 3-PHASE SOCKET TESTER. Check out what else you can get from Martindale:

Calibrator Capabilities Function Range Resolution Measure V dc 28 V 1 mv Measure ma dc Source ma dc

1000FLT. Users Manual. Fluorescent Light Tester

INSPECTOR LINE LOAD SIMULATOR INSTRUCTION MANUAL TASCO, INC.

INSTRUCTION MANUAL CM95/ CM100 TRUE RMS FLEXIBLE CLAMP METER. 1. SAFETY INFORMATION: Always read before proceeding. REMEMBER: SAFETY IS NO ACCIDENT

i400s Calibration Information Introduction

BT508/BT510/BT520/BT521 Battery Analyzer Safety Sheet

DT33B/C/D OPERATOR S MANUAL LIMITED WARRANTY AND LIMITATION OF LIABILITY

Users Manual. Airflow Meter

INSTRUCTION MANUAL DT173 DT175 DIGITAL THERMOMETERS ALWAYS READ THESE INSTRUCTIONS BEFORE PROCEEDING

Visit us at SOLAR-500. Solar Analyzer. Users Manual Mode d emploi Bedienungshandbuch Manual d Uso Manual de uso

Calibration Manual. Detachable Jaw True-rms Clamp Meter

712 RTD Calibrator. Instruction Sheet. Introduction

Go to to register your product and find more information.

INSTRUCTION MANUAL CM69 TRMS AC EARTH LEAKAGE CLAMP METER. 1. SAFETY INFORMATION: Always read before proceeding. REMEMBER: SAFETY IS NO ACCIDENT

HIGH VOLTAGE PROXIMITY DETECTOR

41210S CIRCUIT BREAKER FINDER WITH GFCI Instrucciones de uso Detector de Interruptores con GFCI 41210S. Operating Instructions

Users Manual. MegOhmMeter


KSZ 10B Current Probe Calibrator. Instruction Manual

716 Pressure Calibrator

Visit us at

(R86049) WARNING: To reduce the risk of injury, the user must read and understand the operator s manual before using this product.

KSZ 100B Current Probe Calibrator. Instruction Manual

HIGH VOLTAGE PROVING UNIT

80 Series V Digital Multimeter Safety Information

MARTINDALE INSTRUCTIONS ELITE FUSE FINDER KIT ELECTRIC. Trusted by professionals. 4.4 Storage Conditions

51 & 52 Series II. Users Manual. Thermometer. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

PLS SLD. Laser Detector PACIFIC LASER SYSTEMS. For use with your PLS pulsed laser alignment tool. The Professional Standard

60 Watt Industrial LED Low Bay Light

INSTRUCTION MANUAL MARTINDALE CM52 CLAMP METER. ELECTRIC Trusted by professionals. Other products from Martindale:

Instruction Manual. Model 307A Analog Insulation & Continuity Meter

Manufacturing & Service: Bierer & Associates Inc. Patent No. 6,885,180. ST Series Service Tester Phase Identifier Operating Instructions ST500PGN

USE AND CARE GUIDE DUALCHECK CIRCUIT TESTER

User Guide. Model Insulation Tester / Megohmmeter. Introduction

Megohmmeter Test Probe

Model 1672, 1673 Triple Output Power Supply

PK-8270-X/PK-8370-X. Lines and Fittings Kit. Instructions. Introduction. How to Contact Fluke Calibration

Model 1693, 1694 Switching DC Power Supply

Vol-Con XL (61-086) Voltage/Continuity Tester and Vol-Test (61-085) Voltage Tester Instruction Manual WARNING

SLA Battery Capacity Analyzer

Temperature Probe, Battery (installed), Instruction Sheet, Quick Reference Card and Warranty Card.

PK-VLV-ABORT-100M, PK-VLV-ISO-40M, PK-VLV-ISO-100M

Instruction Manual ICM A1 AC/DC Clamp Meter ICM A1

Tachometer. English. Users Manual

BK BK J93 MA BK MA BK MN93-BK MN193-BK MR193-BK SL261-BK SR193-BK

Instruction Manual VT25 / VT28 VOLTAGE TESTER ALWAYS READ THESE INSTRUCTIONS BEFORE PROCEEDING

OPERATION MANUAL MILLIAMP PROCESS CLAMPMETER MODEL - KM 071 LIST OF PRODUCTS

BK BK A BK MA BK MA BK J93 MN93-BK MN193-BK MR193-BK SL261-BK SR193-BK

LEM~flex RR3020 AC Current Probe Operating Instructions Mode d emploi Bedienungshandbuch Manuale Operativo Manual de uso

Switching DC Power Supply

80PK-2A Type K Immersion Probe

INSTRUCTION MANUAL PET Fax: (503)

MicroPod Sub User manual / Bedienungsanleitung / Manual del Usario

Digital/Analog Megohmmeter Model 1026 USER MANUAL

P5510. Users Manual. Pneumatic Comparison Test Pump

Rental Industry Safety Tester Safe Check 5s

Infrared Thermocouple Probe

DIGITAL EARTH RESISTANCE Model KM 2030

Podiatry Procedures Chair. For Models: Barrier-Free. User s Guide

Hybex Microsample Incubator

4-LED Solar-Powered Pendant Shed Light

Grundfos Variable Speed GRUNDFOS INSTRUCTIONS. Installation and Operation

T2 Tester. Calibration Information. Introduction. Safety Information

CP GermanCp 41_CP C. Operating Instructions. CP 30 CP 330 CP 1100 AC/DC current probe /7.10

Transcription:

TIC 300 PRO High Energy Tic Tracer Non Contact AC Voltage Detector Users Manual Mode d emploi Bedienungshandbuch Manual d uso Manual de uso

TIC 300 PRO High Energy Tic Tracer Non Contact AC Voltage Detector English Users Manual March 2010, Rev002 2010 Amprobe Test Tools. All rights reserved. Printed in China

Limited Warranty and Limitation of Liability Your Amprobe product will be free from defects in material and workmanship for 1 year from the date of purchase. This warranty does not cover fuses, disposable batteries or damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty on Amprobe s behalf. To obtain service during the warranty period, return the product with proof of purchase to an authorized Amprobe Test Tools Service Center or to an Amprobe dealer or distributor. See Repair Section for details. THIS WARRANTY IS YOUR ONLY REMEDY. ALL OTHER WARRANTIES - WHETHER EXPRESS, IMPLIED OR STAUTORY - INCLUDING IMPLIED WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE OR MERCHANTABILITY, ARE HEREBY DISCLAIMED. MANUFACTURER SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you. Repair All test tools returned for warranty or non-warranty repair or for calibration should be accompanied by the following: your name, company s name, address, telephone number, and proof of purchase. Additionally, please include a brief description of the problem or the service requested and include the test leads with the meter. Non-warranty repair or replacement charges should be remitted in the form of a check, a money order, credit card with expiration date, or a purchase order made payable to Amprobe Test Tools. In-Warranty Repairs and Replacement All Countries Please read the warranty statement and check your battery before requesting repair. During the warranty period any defective test tool can be returned to your Amprobe Test Tools distributor for an exchange for the same or like product. Please check the Where to Buy section on www.amprobe.com for a list of distributors near you. Additionally, in the United States and Canada In-Warranty repair and replacement units can also be sent to a Amprobe Test Tools Service Center (see address below).

Non-Warranty Repairs and Replacement US and Canada Non-warranty repairs in the United States and Canada should be sent to a Amprobe Test Tools Service Center. Call Amprobe Test Tools or inquire at your point of purchase for current repair and replacement rates. In USA In Canada Amprobe Test Tools Amprobe Test Tools Everett, WA 98203 Mississauga, ON L4Z 1X9 Tel: 877-AMPROBE (267-7623) Tel: 905-890-7600 Non-Warranty Repairs and Replacement Europe European non-warranty units can be replaced by your Amprobe Test Tools distributor for a nominalv charge. Please check the Where to Buy section on www.amprobe.com for a list of distributors near you. European Correspondence Address* Amprobe Test Tools Europe In den Engematten 14 79286 Glottertal, Germany Tel.: +49 (0) 7684 8009-0 *(Correspondence only no repair or replacement available from this address. European customers please contact your distributor.)

TIC 300 PRO 1) Voltage sensor 2) ON/ Button 3) Circuit TEST button 4) HIGH range switch 5) HIGH range LED 6) LOW range switch 7) LOW range LED 8) Battery compartment 9) HOT STICK attachment point 10) Detection Blue LEDs indicator

CONTENTS SYMBOLS... 2 UNPACKING AND INSPECTION... 3 INTRODUCTION... 4 Features... 4 OPERATION... 5 ON / Button... 5 TEST Button... 5 Voltage Detection... 6 LOW Range... 6 HIGH Range... 6 Locate breaks in insulated wire & heating elements... 7 Check Outlets and Switches... 7 MAINTENANCE AND REPAIR... 8 Battery Replacement... 8 SPECIFICATION... 9 LOW Voltage... 10 HIGH Voltage... 10 1

SYMBOLS W X T P Refer to the manual Caution Dangerous Voltage Double insulated Complies with EU directives ; Conforms to relevant Australian standards. = Do not dispose of this product as unsorted municipal waste. For Use By Trained Personnel Only Anyone using this instrument should be knowledgeable and trained about the risks involved with measuring medium and high voltage, especially in an industrial setting, and the importance of taking safety precautions and of testing the instrument before and after using it to ensure that it is in good working condition. WWARNING and PRECAUTIONS Use extreme caution when testing live electrical circuits due to risk of injury from electrical shock. Do not exceed the maximum overload limits (see specifications) nor the limits marked on the instrument itself. Never test voltage more than 122 kv ac rms. When using the HIGH range, the TIC 410A HOT STICK or equivalent extension pole MUST be used. Comply with local and national safety requirements. Use proper protective equipment as required by local and national authorities. 2

Observe the proper safety precautions when working with voltages above 33 VAC rms to avoid electrical shock hazard. Before and after hazardous voltage measurements, test the Tic 300 PRO by pressing the TEST button and/or on a known source such as line voltage to determine proper operation. Never ground yourself when taking measurements. Do not touch exposed circuit elements. Do not operate the instrument in an explosive atmosphere. Do not expose this product to rain or moisture. This increases the risk of fire or electric shock. Do not rely on this tester for shielded wire. Do not let the unit make contact with a live line with voltage higher than 20 kv. 3 phase feeder cables with conductors close to each other may self-cancel the electric field and not be detected by your Tic tracer. Verify that the phase conductors are separated by at least 15 (381mm) before testing for AC voltage. Unpacking and Inspection Your shipping carton should include: 1 TIC 300 PRO 1 9V Alkaline Battery 1 Users Manual 1 Carrying Case If any of the items are damaged or missing, return the complete package to the place of purchase for an exchange. 3

INTRODUCTION The TIC 300 PRO is an instrument for safely checking the presence of an AC voltage without making any disconnection in cables, wall sockets, fuses, circuit breakers, junction boxes, etc. No current flow is needed for correct operation. Rugged and reliable for utility, industrial and mine safety applications, The TIC 300 PRO warns against dangerous voltages in low and medium voltage applications up to 122 KV without contacting the energized conductor. Features Utility tool for checking transmission lines, power distribution equipment, down power lines, fuses, and load break connectors Lower voltage setting for checking voltage presence in breaker panels, breakers, power outlets and wiring Verifies presence of Voltage from 30 VAC to 122,000 VAC (122kV) Non-contact AC voltage detection Visual and audible voltage indication Self-test verifies that tester works properly Ergonomic design with a convenient handle Drop-proof to 6-FT Used with Hot Stick TIC 410A for voltage higher than 1500V 4

Operation ON / Button When the TIC 300 PRO is activated, the blue LED s and beeper will pulse at a slow rate about twice a second. The red LED of the range (HIGH) will turn ON. TEST Button The TEST button should be used prior to testing actual power circuits. The TEST function will cause the blue LEDs to pulse at a continuous rate and the beeper tone to stop either on LOW or HIGH mode. OSHA listed clearance distances for working with voltage. National or local requirements may be different. Line voltage (kv) Phase to ground Phase to phase phase to phase (ft-in) (m) (ft-in) (m) 0.03 to 1.0 note1 note note1 note1 1.1 to 15.0 2-1 0.64 2-2.66 15.1 to 36.0 2-4 0.72 2-7 0.77 36.1 to 46.0 2-7 0.77 2-10 0.77 46.1 to 72.5 3-0 0.90 3-6 1.05 72.6 to 121 3-2 0.95 4-3 1.29 Note1. No contact necessary WCaution: If a three phase circuit is being checked and one phase is open, the voltage from another phase maybe detected and will give a faulty indication.always make sure that the phase cables are separated by at least 15 (381mm) 5

Voltage Detection LOW Range 1. Push ON/ push button to turn the unit ON 2. The high voltage detection circuit is activated.a twice per second beep will be heard and the blue LEDs will flash at the same rate. 3. Push LOW range push button. The unit is ready for low voltage detection 4. Place the probe tip of the TIC 300 PRO near, but not touching the low (<1000 V) voltage line. See Fig.1 5. If voltage is present, the blue LEDs will blink at a faster rate and the beeper sound will increase at the same rate. A continuous visual rate without a beeper tone will register when the probe tip is too close to the voltage line. HIGH Range (HOT STICK must be used) (See Fig.5) 1. Push the HIGH range push button 2. Approach the probe tip near, but not touching the high voltage (> 1500V and <122 kv) conductor. 3. If voltage is present, the blue LEDs will blink at a faster rate and the beeper sound will increase at the same rate. A continuous visual rate without a beeper tone will register when the probe tip is too close to the high voltage line. Locate breaks in insulated wire and heating elements. (See Fig.2) 1. Push the ON/ push button to turn the unit on 6

2. Push the LOW range push button 3. Make sure the wire or the heating element is energized 4. Place the tip of the probe near the wire and start tracing along it. When the fast or continuous beep stops, this indicates the point of the break or open Check Outlets and Switches. (See Figs 3 & 4) 1. Push the ON/ push button to turn the unit on 2. Push the LOW range push button 3. Place the probe tip of the TIC 300 PRO against the outlet or switch contact. 4. Increase beeping indicates voltage presence 7

MAINTENANCE AND REPAIR If the TIC 300 PRO will not turn on, check the battery. If there appears to be a malfunction during the operation of the voltage detector, the following steps should be performed in order to isolate the cause of the problem: 1. Check the battery. 2. Review the operating instructions for possible mistakes in operating procedure. Except for the replacement of the battery, repair of the voltage detector should be performed only by a Factory Authorized Service Center or by other qualified instrument service personnel. The front panel and case can be cleaned with a mild solution of detergent and water. Apply sparingly with a soft cloth and allow drying completely before using. Do not use aromatic hydrocarbons or chlorinated solvents for cleaning. Battery Replacement Use phillips blade screwdriver to remove the battery hatch screw. Lift the hatch. Remove the battery and replace it with a 9 volt alkaline battery. Make sure the seal is in place. Close the battery hatch and replace the screw. 8

SPECIFICATION Operating voltage LOW Range: 30 to 1500 VAC; HIGH Range: 1500 VAC to 122 kvac Operating Temperature: 0 C to +52 C(32 F to +125 F); RH < 90% Storage Temperature:0 C to +52 C(32 F to +125 F) battery removed. Environment: < 2000m, outdoor operation Power: 9 volt alkaline battery Power Consumption: : 340uA Low Detection mode: 39 ma High Detection mode:39 ma Duty Cycle: Continuous Response Time: Instantaneous Dimensions: 345(L) x 75(W) x 47(H)mm (13.5 x 3.0 x 1.8 in) Weight: 243g (0.5 lb.) Optional Accessory: TIC 410A Hot Stick P. EMC: EN 61326-1:2006 This product complies with requirements of the following European Community Directives: 2004/108/EC (Electromagnetic Compatibility) and 2006/95/EC (Low Voltage) as amended by 93/68/EEC (CE Marking). However, electrical noise or intense electromagnetic fields in the vicinity of the equipment may disturb the measurement circuit. Measuring instruments will also respond to unwanted signals that may be present within the measurement circuit. Users should exercise care and take appropriate precautions to avoid misleading results when making measurements in the presence of electronic interference. 9

Electrical: Sensitivity: typical as listed on the table below LOW Voltage Voltage (Phase to Ground) Detection Range Average 30V 2.7 75V 9.7 100V 1 1 120V 1 6 220V 2 4 500V 4 2 1000V 6 3 1500V 7 9 HIGH Voltage Voltage (Phase to Ground) Detection Range Average 1.5 kv 1 2 2.0 kv 2 5 4.0 kv 3 10.0 kv 6 3 25.0 kv 9 4 50.0 kv 13 0 75.0 kv 14 8 100.0 kv 19 8 122.0 kv 22 0 10

Fig.1 Voltage detection in a line Fig.2 Finding a break or open 11

Fig.3 Voltage detection in a outlet ON ON ON ON ON ON Fig.4 Finding a breaker 12

HOT STICK Fig.5 Voltage detection in transmission lines 13