NVLAP LAB CODE:

Similar documents
NVLAP LAB CODE:

PHONE: (303) FAX: (970) WEBSITE:

PHONE: (303) FAX: (970) WEBSITE:

PHONE: (303) FAX: (303) WEBSITE:

S BERGIN R BEATTIE Lighting Engineer. Checked Approved

PHONE: (303) FAX: (970) WEBSITE:

PHONE: (303) FAX: (970) WEBSITE:

Test Sample Description Catalog Number: Description:

Test Sample Description Catalog Number: Description:

Report of Test LLIA A-R01*

Test Sample Description Catalog Number: Description:

Report of Test LLIA A

Report of Test LLIA A

Report of Test LLIA A

Report of Test LLIA A-R02*

Osram Sylvania 200 Ballardvale Street Wilmington, Mass

Report of Test LLIA A

LM Test Report. For LIGHT EFFICIENT DESIGN. (Brand Name:N/A) 188 S. Northwest HighwayCary, IL LED Luminaires

LLIA A-R02*

Report of Test LLIA A

Test Sample Description Catalog Number: Description:

LM Test Report. For LIGHT EFFICIENT DESIGN. (Brand Name:N/A) 188 S. Northwest HighwayCary, IL LED Luminaires

LM Test Report. For LIGHT EFFICIENT DESIGN. (Brand Name:N/A) 188 S. Northwest HighwayCary, IL LED Luminaires

LIGHT EFFICIENT DESIGN, LLC

Report of Test LLIA A

Report of Test LLIA A

Calibration Due Date Chroma Programmable AC Source 61604

NVLAP LAB CODE LM Test Report. For LIGHT EFFICIENT DESIGN. (Brand Name:N/A) 188 S. Northwest HighwayCary, IL60013.

NVLAP LAB CODE LM Test Report. For LIGHT EFFICIENT DESIGN. (Brand Name:N/A) 188 S. Northwest HighwayCary, IL60013.

NVLAP LAB CODE LM Test Report. For LIGHT EFFICIENT DESIGN. (Brand Name:N/A) 188 S. Northwest HighwayCary, IL60013.

LIGHT EFFICIENT DESIGN

NVLAP LAB CODE LM Test Report. For LIGHT EFFICIENT DESIGN. (Brand Name:N/A) 188 S. Northwest HighwayCary, IL60013.

8165 E Kaiser Blvd. Anaheim, CA p f NVLAP LAB CODE

NVLAP LAB CODE LM Test Report. For. MaxLite Inc. (Brand Name: MaxLite) 12 York AveWest Caldwell NJ 07006

NVLAP LAB CODE LM Test Report. For. LIGHT EFFICIENT DESIGN (Brand Name:N/A) 188 S. Northwest Highway Cary, IL

NVLAP LAB CODE LM Test Report. For. LIGHT EFFICIENT DESIGN (Brand Name:N/A) 188 S. Northwest Highway Cary, IL

NVLAP LAB CODE LM Test Report. For MAXLITE INC. (Brand Name: N/A) 12 York Ave, West Caldwell, NJ 07006

NVLAP LAB CODE LM Test Report. For. LIGHT EFFICIENT DESIGN (Brand Name:N/A) 188 S. Northwest Highway Cary, IL

Photometric Indoor Test Report

NVLAP LAB CODE LM Test Report. For. LIGHT EFFICIENT DESIGN (Brand Name:N/A) 188 S. Northwest Highway Cary, IL

Photometric Test Report. Overdrive Electronics Pvt. Ltd C-121, Hosiery Complex, Phase2 Extn. Mohit K Mittal, ,

NVLAP LAB CODE LM Test Report. For. LIGHT EFFICIENT DESIGN (Brand Name:N/A) 188 S. Northwest Highway Cary, IL

Integrating Sphere Test Report

Integrating Sphere Test Report

NVLAP LAB CODE LM Test Report. For. MaxLite lnc. (Brand Name: MaxLED) 12 York Ave West Caldwell NJ 07006,USA

8165 E Kaiser Blvd. Anaheim, CA p f NVLAP LAB CODE

Lighting Sciences

NVLAP LAB CODE LM Test Report. For REMPHOS TECHNOLOGIES LLC. (Brand Name: ) 90 Holten St, Danvers, MA 01923,USA.

Integrating Sphere Test Report

NVLAP LAB CODE LM Test Report For. Model Different: All construction and rating are the same, except CCT

LED Black Housing High Output Remote Capable

Report of Test LLIA A

NVLAP LAB CODE LM Test Report. For GREEN INOVA LIGHTING TECHNOLOGY (SHENZHEN) LTD. (Brand Name: GI LED LIGHTING)

LM Test Report. LIGHT EFFICIENT DESIGN (Brand Name:N/A) LED Luminaire

NVLAP LAB CODE LM Test Report. For GREEN INOVA LIGHTING TECHNOLOGY (SHENZHEN) LTD. (Brand Name: GI LED LIGHTING)

NVLAP LAB CODE LM Test Report. For. MLS CO., LTD (Brand Name: MLS) NO.1 MULINSEN RD XIAOLAN TOWNSHIP ZHONGSHAN, GUANGDONG CHINA

NVLAP LAB CODE LM Test Report. For GREEN INOVA LIGHTING TECHNOLOGY (SHENZHEN) LTD. (Brand Name: GI LED LIGHTING)

NVLAP LAB CODE LM Test Report. For GREEN INOVA LIGHTING TECHNOLOGY (SHENZHEN) LTD. (Brand Name: GI LED LIGHTING)

Photometric Test Report

NVLAP LAB CODE LM Test Report. For DONGGUAN THAILIGHT SEMICONDUCTOR LIGHTING CO.,LTD. (Brand Name: THAILIGHT)

INDEPENDENT TEST LABORATORY REPORT No

Catalog Number UPC Number Features. General Lumen Output: 2300 Color Temperature: 5000K Cool White CRI: 70+ Housing Color: Black

Photometric Test Report. MaxLite Inc 10 York Avenue, West Caldwell, NJ Ramesh Raghavan, x146,

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Report of Test LLIA A

Catalog Number UPC Number Features. General Lumen Output: 800 Color Temperature: 5000K Cool White CRI: 70+ Housing Color: Black

INDEPENDENT TEST LABORATORY REPORT No

INDEPENDENT TEST LABORATORY REPORT No A

NVLAP LAB CODE LM Test Report. For DONGGUAN THAILIGHT SEMICONDCTOR LIGHTING CO.,LTD. (Brand Name: THAILIGHT)

NVLAP LAB CODE LM Test Report. For DONGGUAN THAILIGHT SEMICONDCTOR LIGHTING CO.,LTD. (Brand Name: THAILIGHT)

NVLAP LAB CODE LM Test Report. For SNC OPTO ELECTRONIC CO.,LTD. (Brand Name: SNC LED)

PHOTOMETRIC TESTING & EVALUATION. Sample Tested Man-D-Tec UNILED-12-4K. Prepared for: Man-D-Tec 8175 E. Paradise Lane Scottsdale, AZ 85260

NVLAP LAB CODE LM Test Report. For SNC OPTO ELECTRONIC CO.,LTD. (Brand Name: SNC LED)

Performance Summary Voltage Current Power Frequency Power Factor Current THD A W Total Luminous Flux

Wet Locations Use. RoHS. Model: VAC / 28 W PF LM 93 LM/W. Alluminium + Tempered Glass 24-45VDC 0.

NVLAP LAB CODE LM Test Report. For HOCAN GROUP CO.,LTD. (Brand Name: SEPICN LED LIGHTING)

8165 E Kaiser Blvd. Anaheim, CA p f NVLAP LAB CODE MaxLite 12 York Ave. West Caldwell, NJ.

LM-79 Test Report. Electrical and Photometric tests as required by IESNA LM-79 test standard

NVLAP LAB CODE LM Test Report. For HOCAN GROUP CO.,LTD. (Brand Name: SEPICN LED LIGHTING)

NVLAP LAB CODE LM Test Report. For DONGGUAN THAILIGHT SEMICONDCTOR LIGHTING CO.,LTD. (Brand Name: THAILIGHT)

LEN Series 2x2 / 2x4 LED

Number Eight Lighting Company 803/K2-HI-15-XX/DIM /FS-P-1-WH

NVLAP LAB CODE LM Test Report. For IGT LIGHTING INC Lincoln St. Suite B, Riverside, CA 92503

Performance Summary Voltage Current Power Frequency Power Factor Current THD A W Total Luminous Flux

8165 E Kaiser Blvd. Anaheim, CA p f NVLAP LAB CODE Maxlite Jersey Blvd Rancho Cucamonga, CA 91730

T8 4ft 5000K EazyLux EZ8450D7R07

8165 E Kaiser Blvd. Anaheim, CA p f NVLAP LAB CODE

Wet Locations Use VAC / 79 W PF LM 92 LM/W. Alluminium + Tempered Glass 25-42VDC 1.95A 88% CXA PCS

8165 E Kaiser Blvd. Anaheim, CA p f NVLAP LAB CODE

Calibration Due Date Chroma Programmable AC Source 61604

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

LM Test Report. For. Revolution Lighting Technologies, Inc. (Brand Name: ) 2280 Ward Ave. Simi Valley, CA

Report of Test LLIA

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Rated voltage: 100~277 VAC Frequency: 50/60Hz Wattage: 14 W

Report of Test LLIA

Report of Test LLIA

8165 E Kaiser Blvd. Anaheim, CA p f NVLAP LAB CODE

Transcription:

REPORT NUMBER: RAB02540 PAGE: 1 OF 7 CATALOG NUMBER: VAN1LED12NW (SURFACE MOUNT - DROP LENS) LUMINAIRE: WHITE ALUMINUM SHEET METAL HOUSING, ONE WHITE CIRCUIT BOARD WITH 48 LEDS, FROSTED POLYCARBONATE DROP LENS, LENS FROSTED SIDE IN. LAMPS: FORTY-EIGHT WHITE LIGHT EMITTING DIODES (LEDS), VERTICAL BASE-UP POSITION. NOTE: DATA SHOWN IS ABSOLUTE FOR THE SAMPLE PROVIDED. *(SEE PAGE 2 FOR MORE INFORMATION)* CANDELA DISTRIBUTION FLUX 0.0 22.5 45.0 67.5 90.0 0 405 405 405 405 405 5 404 404 404 404 404 38 15 390 390 391 391 392 110 25 362 363 367 370 370 169 35 323 327 334 338 338 208 45 279 286 293 296 295 224 55 229 237 241 247 247 215 65 177 183 185 193 196 185 75 117 120 122 130 131 131 85 53 59 62 62 60 67 90 31 39 43 43 40 95 25 32 35 35 33 37 105 23 26 29 30 29 29 115 20 20 23 26 27 23 125 12 11 16 19 21 14 135 8 6 8 12 14 7 145 4 4 4 6 8 3 155 3 3 3 4 4 1 165 3 3 3 3 3 1 175 1 1 1 1 2 0 180 0 0 0 0 0 ZONAL LUMEN SUMMARY ZONE LUMENS %FIXT 0-30 318 21.7 0-40 526 35.9 0-60 965 65.9 0-90 1348 92.1 90-120 89 6.1 90-130 103 7.0 90-150 114 7.8 90-180 116 7.9 0-180 1464 100.0 TOTAL INPUT WATTS = 11.8 EFFICACY = 124.1 Lm/W CIE TYPE - DIRECT PLANE : 0-DEG 90-DEG SPACING CRITERIA : 1.3 1.3 140 280 420 0 15 30 LEGEND: 0-deg: 45-deg: 90-deg: Checked Approved X.CAO D.WANG-MUNSON 90 75 60 45

REPORT NUMBER: RAB02540 PAGE: 2 OF 7 ADDITIONAL INFORMATION TOTAL INPUT WATTS = 11.795 W AT 120.0 VAC. LED DRIVER: INTEGRAL TEST PROCEDURE: IESNA LM-79-08 ACCREDITED LABORATORY CODE 201058-0 TEST DISTANCE : 28.25 FEET NOTE: THIS REPORT WITH THE USE OF THE NVLAP LOGO SHALL NOT BE USED BY THE CLIENT TO CLAIM PRODUCT CERTIFICATION, APPROVAL, OR ENDORSEMENT BY NVLAP, NIST, OR ANY AGENCY OF THE FEDERAL GOVERNMENT.

REPORT NUMBER: RAB02540 PAGE: 3 OF 7 PLANE : 0-DEG 90-DEG LUMINOUS LENGTH : 6.710 4.620 HEIGHT OF SIDE : 2.000 2.000 LUMINANCE DATA IN CANDELA/SQ METER ANGLE AVERAGE AVERAGE AVERAGE IN DEG 0-DEG 45-DEG 90-DEG 45 15193. 15167. 14552. 55 13997. 13799. 13301. 65 12770. 12266. 12021. 75 10696. 9966. 9672. 85 6897. 6867. 5785.

REPORT NUMBER: RAB02540 PAGE: 4 OF 7 CANDELA DISTRIBUTION 0.0 22.5 45.0 67.5 90.0 0.0 405 405 405 405 405 5.0 404 404 404 404 404 10.0 399 399 400 400 399 15.0 390 390 391 391 392 20.0 377 378 380 382 382 25.0 362 363 367 370 370 30.0 344 346 352 355 355 35.0 323 327 334 338 338 40.0 301 305 313 317 315 45.0 279 286 293 296 295 50.0 254 262 268 272 273 55.0 229 237 241 247 247 60.0 204 211 214 221 221 65.0 177 183 185 193 196 70.0 148 152 154 164 166 75.0 117 120 122 130 131 80.0 84 88 90 94 93 85.0 53 59 62 62 60 90.0 31 39 43 43 40 95.0 25 32 35 35 33 100.0 23 28 32 32 30 105.0 23 26 29 30 29 110.0 22 24 26 28 28 115.0 20 20 23 26 27 120.0 15 16 20 23 25 125.0 12 11 16 19 21 130.0 10 8 12 15 18 135.0 8 6 8 12 14 140.0 6 5 5 9 11 145.0 4 4 4 6 8 150.0 3 3 3 4 5 155.0 3 3 3 4 4 160.0 3 3 3 3 3 165.0 3 3 3 3 3 170.0 2 2 2 2 2 175.0 1 1 1 1 2 180.0 0 0 0 0 0

REPORT NUMBER: RAB02540 PAGE: 5 OF 7 ZONAL LUMEN SUMMARY 0-5 10. 5-10 29. 10-15 47. 15-20 63. 20-25 78. 25-30 91. 30-35 101. 35-40 107. 40-45 111. 45-50 113. 50-55 110. 55-60 105. 60-65 98. 65-70 87. 70-75 73. 75-80 57. 80-85 41. 85-90 27. 90-95 20. 95-100 17. 100-105 15. 105-110 14. 110-115 12. 115-120 10. 120-125 8. 125-130 6. 130-135 4. 135-140 3. 140-145 2. 145-150 1. 150-155 1. 155-160 1. 160-165 0. 165-170 0. 170-175 0. 175-180 0.

REPORT NUMBER: RAB02540 PAGE: 6 OF 7 5-DEGREE 10-DEGREE ZONAL LUMEN SUMMARY ZONAL LUMEN SUMMARY 0-5 10 0-10 38 5-10 29 0-20 149 10-15 47 0-30 318 15-20 63 0-40 526 20-25 78 0-50 749 25-30 91 0-60 965 30-35 101 0-70 1149 35-40 107 0-80 1280 40-45 111 0-90 1348 45-50 113 0-100 1384 50-55 110 0-110 1414 55-60 105 0-120 1436 60-65 98 0-130 1451 65-70 87 0-140 1458 70-75 73 0-150 1461 75-80 57 0-160 1463 80-85 41 0-170 1463 85-90 27 0-180 1464 90-95 20 95-100 17 100-105 15 105-110 14 110-115 12 115-120 10 120-125 8 125-130 6 130-135 4 135-140 3 140-145 2 145-150 1 150-155 1 155-160 1 160-165 0 165-170 0 170-175 0 175-180 0

REPORT NUMBER: RAB02540 PAGE: 7 OF 7 COEFFICIENTS OF UTILIZATION - ZONAL CAVITY METHOD EFFECTIVE FLOOR CAVITY REFLECTANCE 0.20 RC 80 70 50 30 10 0 RW 70 50 30 10 70 50 30 10 50 30 10 50 30 10 50 30 10 0 0 117117117117 114114114114 107107107 100100100 95 95 95 92 1 105100 95 90 102 97 92 88 91 87 84 86 83 80 81 78 76 73 2 95 86 78 72 91 83 76 71 78 73 68 74 69 65 70 66 62 60 3 86 75 66 59 83 73 64 58 68 62 56 64 59 54 61 56 52 49 4 78 66 57 49 75 64 55 49 60 53 47 57 51 46 54 49 44 42 5 72 59 49 42 69 57 48 42 54 46 40 51 44 39 48 43 38 36 6 66 52 43 37 64 51 42 36 48 41 35 46 39 34 44 38 33 31 7 61 47 38 32 59 46 38 32 44 36 31 42 35 30 40 34 29 27 8 57 43 34 28 55 42 34 28 40 33 27 38 32 27 36 31 26 24 9 53 40 31 25 51 39 31 25 37 30 25 35 29 24 34 28 24 22 10 50 36 28 23 48 36 28 23 34 27 22 33 26 22 31 25 21 19 ALL CANDELA, LUMENS, LUMINANCE, AND VCP VALUES IN THIS REPORT ARE BASED ON ABSOLUTE PHOTOMETRY. THE COEFFICIENT OF UTILIZATION VALUES ARE BASED ON THE TOTAL ABSOLUTE LUMEN OUTPUT OF THIS LUMINAIRE SAMPLE.

REPORT NUMBER: RAB02550 Page 1 of 4 DATE: 10/4/2016 CATALOG NUMBER: VAN1LED12NW (SURFACE MOUNT - DROP LENS) ADDRESS: 170 LUDLOW AVE, NORTHVALE, NJ 07647 LUMINAIRE: WHITE ALUMINUM SHEET METAL HOUSING, ONE WHITE CIRCUIT BOARD WITH 48 LEDS, FROSTED POLYCARBONATE DROP LENS, LENS FROSTED SIDE IN. LAMP: DRIVER: NOTE: FORTY-EIGHT WHITE LIGHT EMITTING DIODES (LEDS), VERTICAL BASE-UP POSITION. INTEGRAL DATA SHOWN IS ABSOLUTE FOR THE SAMPLE PROVIDED AT RATED INPUT VOLTAGE (120.0 VAC, 60Hz) TO THE TEST SAMPLE. Calibration Due: INSTRUMENTS: GWINSTEK PROGRAMMABLE AC POWER SOURCE APS-7100 N/A CHROMA PROGRAMMABLE DIGITAL POWER METER MODEL 66202 2/26/17 OCEAN OPTICS QE65PRO Spectroradiometer 9/16/17 RAB 2.0 meter Diameter Integrating Sphere, 4PI Geometry 9/16/17 OBJECT OF TEST: PROCEDURE: *NOTE: RESULTS: Measure the Total Radiant Flux*, Spectral Power Distribution (SPD), Correlated Color Temperature (CCT), Color Rendering Indices (CRIa,1-14), Chromaticity Coordinates (x,y; u v ), ANSI C78.377 Duv, and electrical data including ANSI C82.77-2002 Power Factor (PF) and Total Harmonic Distortion (THD) to the test sample. Report Off-State Power. The test sample was provided by the customer and had an unknown number of burn hours. The test sample was mounted inside the integrating sphere and allowed to stabilize. After stabilization occurred, measurements were taken. In order to measure mean performance, multiple data sets were recorded and averaged. Readings were taken with the test sample operating at 120.0 VAC input in a 25 +/-1 degree Celsius free air ambient and in accordance with IESNA LM-79-08. All data are traceable to the National Institute of Standards and Technology. Off-State Power was reported with no voltage applied to the sample. Proper calibration of integrating spheres for measuring total flux output of non-directional samples will produce reliable, repeatable results within the calibration tolerances of the equipment used. However, measurement of test samples with significant self absorption and/or directional output, even when these effects are compensated for, are likely to have a greater variation in results compared to the flux output calculated from a goniophotometric exploration since these artifacts do not affect the goniophotometric results. (continued subsequent pages) Checked Approved X.CAO D.WANG-MUNSON Lighting Engineer THIS REPORT IS BASED ON PUBLISHED INDUSTRY PROCEDURES. FIELD PERFORMANCE MAY DIFFER FROM LABORATORY PERFORMANCE.

REPORT NUMBER: RAB02550 Page 2 of 4 DATE: 10/4/2016 CATALOG NUMBER: VAN1LED12NW (SURFACE MOUNT - DROP LENS) RESULTS: SPECTRORADIOMETRIC Observer Chromaticity Ordinate x Chromaticity Ordinate y Observer Chromaticity Ordinate u Chromaticity Ordinate v Correlated Color Temp CCT (K) Color Rendering Index (CRIa) Color Rendering Index 1 (Light greyish red) Color Rendering Index 2 (Dark greyish yellow) Color Rendering Index 3 (Strong yellowish green) Color Rendering Index 4 (Moderate yellowish green) Color Rendering Index 5 (Light bluish green) Color Rendering Index 6 (Light blue) Color Rendering Index 7 (Light violet) Color Rendering Index 8 (Light reddish purple) Color Rendering Index 9 (Strong red) Color Rendering Index 10 (Strong yellow) Color Rendering Index 11 (Strong green) Color Rendering Index 12 (Strong blue) Color Rendering Index 13 (Light yellowish pink (skin)) Color Rendering Index 14 (Moderate olive green (leaf)) ANSI C78.377-2008 Duv Total Radiant Flux (milliwatts) CIE 1931 2 degree 0.3880 0.3848 CIE 1976 2 degree 0.2269 0.5062 3861 83 81 87 93 83 81 83 87 67 12 71 82 63 82 96 0.002 4443 * *NOTE: ELECTRICAL FOR SPECTRORADIOMETRIC TEST Input Voltage (Volts AC ) Input Current (Amps AC ) Input Power (Watts) Input Power Factor (%) Input Current THD (%) Input Voltage THD (%) Off-State Power (Watts) 120.0 0.099 11.8 99.4 11.1 0.2 0.0 Proper calibration of integrating spheres for measuring total flux output of non-directional samples will produce reliable, repeatable results within the calibration tolerances of the equipment used. However, measurement of test samples with significant self absorption and/or directional output, even when these effects are compensated for, are likely to have a greater variation in results compared to the flux output calculated from a goniophotometric exploration since these artifacts do not affect the goniophotometric results. THIS REPORT IS BASED ON PUBLISHED INDUSTRY PROCEDURES. FIELD PERFORMANCE MAY DIFFER FROM LABORATORY PERFORMANCE.

REPORT NUMBER: RAB02550 Page 3 of 4 DATE: 10/4/2016 CATALOG NUMBER: VAN1LED12NW (SURFACE MOUNT - DROP LENS) RESULTS: Wavelength mw per nm Wavelength mw per nm Wavelength mw per nm 380 0.079 515 16.053 650 14.057 385 0.085 520 16.970 655 12.594 390 0.092 525 17.710 660 11.201 395 0.104 530 18.302 665 9.870 400 0.141 535 18.853 670 8.673 405 0.224 540 19.377 675 7.555 410 0.418 545 19.905 680 6.579 415 0.810 550 20.512 685 5.714 420 1.519 555 21.134 690 4.944 425 2.828 560 21.741 695 4.252 430 5.111 565 22.395 700 3.648 435 8.327 570 23.059 705 3.121 440 13.958 575 23.641 710 2.676 445 22.925 580 24.185 715 2.294 450 26.211 585 24.635 720 1.964 455 18.374 590 24.985 725 1.670 460 12.622 595 25.126 730 1.426 465 10.169 600 25.003 735 1.217 470 7.482 605 24.761 740 1.039 475 6.084 610 24.256 745 0.885 480 6.156 615 23.531 750 0.759 485 6.845 620 22.598 755 0.652 490 8.087 625 21.408 760 0.560 495 9.856 630 20.010 765 0.480 500 11.715 635 18.582 770 0.410 505 13.455 640 17.076 775 0.354 510 14.829 645 15.559 780 0.053 27.5 Output vs. Wavelength 22.0 Output (mw/nm) 16.5 11.0 5.5 0.0 380.0 420.0 460.0 500.0 540.0 580.0 620.0 660.0 700.0 740.0 780.0 Wavelength (nm) THIS REPORT IS BASED ON PUBLISHED INDUSTRY PROCEDURES. FIELD PERFORMANCE MAY DIFFER FROM LABORATORY PERFORMANCE.

REPORT NUMBER: RAB02550 Page 4 of 4 DATE: 10/4/2016 CATALOG NUMBER: VAN1LED12NW (SURFACE MOUNT - DROP LENS) CIE Chromaticity Diagram THIS REPORT IS BASED ON PUBLISHED INDUSTRY PROCEDURES. FIELD PERFORMANCE MAY DIFFER FROM LABORATORY PERFORMANCE.