How to Operate the Nu Plasma MC-ICP-MS. A quick guide

Similar documents
Exercise 3-1. Basic Hydraulic Circuit EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION. Complete hydraulic circuit

HP 1946A LC/MSD Replacement Parts

MINIMIZE WORKFLOW INTERRUPTIONS

Accurate to 0.1 NIST Traceable Advanced PID Algorighm USB Port Free datalogging/control software. Controllers I N S C O N TA G N

Consumables and Supplies Required for Repair

Tube Bender. Machine Type - Tube Bender

Tips for Installation, Calibration and Operation of SmartSwitch Tank Management Systems. Contents: Page

INSTALLATION GUIDE Table of Contents

BMS-LiFePower. 123SmartBMS. Instruction manual

AR2000 Rheometer: Instructions

EPOND SA CP 389 CH Vevey Switzerland.

Updated 12/13/2013 Ver. 2.6

Exercise 3-1. Basic Hydraulic Circuit EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION. Complete hydraulic circuit

BD FACSJazz Users Guide Satchmo Stanford Shared FACS Facility

OWNER'S MANUAL RED LINE NC kv electrostatic powder spray system

Introduction. Figure 1: Labeled picture of the Instron 3367 load frame.

Accurate to 0.1 NIST Traceable Advanced PID Algorithm USB Port Free datalogging/control software

X-Type w/ non-premium sound amplifier installation instructions

User Manual. BMS123 Smart

Radiant High Voltage. Displacement Measurement Fixture. Construction. Introduction. Figure 1

INDIAN INSTITUTE OF TECHNOLOGY KANPUR Kanpur , Uttar Pradesh, India Centre for Lasers and Photonics

Section 3 Technical Information

TWO-WAY LED AUTOMATIC TRANSMISSION REMOTE STARTER. User Guide WARNING

PowerLevel s e r i e s

Clean Water Made Easy

UNITED STATES Ford Customer Relationship Center (FORD) (TDD for the hearing impaired: ) owner.ford.

Atomic Absorption Spectroscopy

TM Vacuum Sputter System SOP

CFM 1050 Reactive Closed Field Magnetron Sputtering System

EFIE Wideband O2 (Electronic Fuel Injector Enhancer) Installation & Operating Instructions.

Figure 1. AirChek XR5000 Air Sample Pump

S-9100, S-9200, S-9300, S-9400, S-9700, S-9800

Operators Manual. Microprocessor Controlled Water Baths /98

Sterling Coilmaster JR TS Issue 3 July 2017

LABORATORY ZERO AIR GENERATOR MODEL N-GC1500 USER S MANUAL

Operating Instructions for Elevator Buffers type LP

Vehicle Security System

January 10, 2017 # Rev. B. Model 464 Mk3 Electronic Pump Control Unit User Guide

Card/Stickers/Pocket (Sleeve) 2016 F-150 Pro Trailer Backup Assist QUICK START GUIDE

Tecomotive - tinycwa User Manual

7500 Greensand Filter Installation & Start Up Guide

PRESSURE SENSOR INSTRUCTIONS

Letter Figures Words SYSTEMS ENGINEERING. Written examination. Monday 19 November 2012

Fleck 2510 Manual Sediment Filter Installation & Start-Up Guide

Induction Power Supplies

MicroCoat. System Operating Manual MC2000 Series. MC785, MC785-WF Spray Valves. US: UK: Mexico:

Operating Instructions EVG620 Aligner

Wiring Main Power connector constant 12V power

Q1. Figure 1 shows a straight wire passing through a piece of card.

Wind Tunnel User Guide V0814

User Guide 1 WAY FM MANUAL TRANSMISSION REMOTE STARTER. Table of Contents. Introduction

A Brief User s Manual of. The Scanning Tunneling Microscope

Analysis of Petroleum Samples Using the Teledyne Leeman Labs

June 25, 2013 # Model 464 Electronic Pump Control Unit User Guide

IMPORTANT! DO NOT THROW AWAY THE SHIPPING CARTON AND PACKING MATERIAL

Tecomotive - tinycwa - User Manual Intercooler Version

Quick GUIDE Web Edition

QUICK INSTALLATION GUIDE

TC62D Installation Instructions

EPAS Desktop Pro Software User Manual

INOVA HIGHTECH Ltd. MEP 002/003 Auto Starter Manual

3.10 Air Conditioning (A/C) Contents

K&L MC200 ELITE WHEEL BALANCER Product Manual - MC Wheel Balancing

BTP12 Two Stage Electric Turbo Pump with Manometer

TWO-WAY LCD AUTOMATIC TRANSMISSION REMOTE STARTER. User Guide

#9040 FUEL TANK SWEEPER

Model 20 Constant Speed Blender : 115 Volt : 230 Volt. Instruction Manual. Updated 11/14/2017 Ver. 4.0

STEALTH CONTROLS. For. By restomod air

TROUBLESHOOTING. 1. What is the model and serial number? If you don t know, how old is it? Give description.

BMS16 v cell Lithium Battery Management System

User manual MTS-32. QUI-1013 Version 1.5 January

DEFIE (Digital Electronic Fuel Injector Enhancer) Installation & Operating Instructions.

Pressure chlorine changeover unit C 7520

CF-10 Option Operating Instructions

INSTALLATION AND OPERATION INSTRUCTION

RESISTIVE LOAD XP300C

CURTIS TOLEDO. AF Series Compressors VS models with VFD WARNING

OWNER'S MANUAL RED LINE ProCoater PC201IF 100 kv electrostatic powder spray system

QUICK START GUIDE FOR DEMONSTRATION CIRCUIT 551A-B LITHIUM-ION BATTERY CHARGER WITH CHARGE TERMINATION

DENTONVACUUM DESKV DESK V - SPUTTER/ETCH UNIT AND ACCESSORIES

AS-1774 / 1774U. User Guide 5-BUTTON AUTOMATIC / MANUAL TRANSMISSION REMOTE STARTER

XeVision XeRay XV-LX50 HID Searchlight

CHEMINSTRUMENTS HOT ROLL LAMINATOR MODEL HL-100, HL-101 OPERATING INSTRUCTIONS

Advanced Petrochemical Applications With ICP-OES. Tina Harville, Ph.D. Application Engineer

Redox Potentials and the Lead Acid Cell Minneapolis Community and Tech. College v I. Introduction. Part I

5900e Carbon Filter Installation & Start-Up Guide

PowerSTAR PS-2024-D. Maximum Power Point Tracking Solar Regulator. w w w. r o c s o l i d. c o m. a u. Contents

ELECTRONIC POSITIVE AIR SHUTOFF

Vessels and Distillation Equipment

MicroCoat System Operating Manual MC4000 Series MC785M, MC785M-WF Spray Valves

Scientific vapor pumps

Troubleshooting. CryoSmart 1 - CryoSmart 20

U-5000AT Ultrasonic Nebulizer Operator s Manual

CONTROLLER DIAGNOSTIC GUIDE

BOSS 302S. Technical Information and Tuning Guidelines ZF Sachs Dampers for Ford Racing BOSS 302S

Performance DSG Upgrade Installation Instructions using the Byteshooter tool version 1.4

Dielectric Withstand Tester

Installation Instructions for VTU

LG Air conditioning CAC and Multi Split unit Fault code sheet Universal and Multi Split Units

Centrifuge Operator / Service Manual

Transcription:

How to Operate the Nu Plasma MC-ICP-MS A quick guide By C. Dalpé, W.F. McDonough, R. Rudnick ICP-MS Laboratory (version 1.2) UMD, May 2002 Note: This is a complement reading of the Nu Plasma manual

How to Start 1- Switch on the Compact computer and the monitor (Login Name = Nu- Plasma; Password = plasma$16a). NOTE: The computer might already be ON. 2- Single click with the left mouse button on the Plasma.exe program located in the toolbar at the bottom screen. Go to Control menu and select Preferred. 3- Check the torch, gas connections (Cool, Aux, Neb) and cones for any unusual arrangements or conditions. (You might want to clean the cones before beginning your session. Check that the safety control switch for the hood is in the position released (unlocked). Move the torch back to the rear closure, under Plasma Control window, and inspect visually the sample cone. In this case you must, however, received a cleaning demonstration before hand from Claude or Bill. *** Please *** wear gloves when working underneath the Plasma Hood. This will avoid having grease from hands on the glassware and cones. Figure 1 (not available yet) 4- Set up your sample introduction configuration to the torch (i.e., Scott double pass, Cyclonic Spray chamber, Aridus, LASER, etc.). Connect the red wire from the electronic relay located above the torchbox and the stainless steel Plasma Shield. The Plasma Shield should be located symmetrically to the RF Coil (~1 mm apart) and inserted between the Si-glass torch and the Sibonnet. Figure 2 (not available yet) 5- Close the hood and make sure it s well locked with the pneumatic piston using the HOOD LOCK chrome toggle switch on the instrument bench (the small PVC grey box). Figure 3 (not available yet) 6- Check room temperature (should be below 20-22 C 67-71 F). With the plasma and extraction rotary pump off. If not call CCMS HVAC Control at x52222 and specify a temperature problem for CHEM-0225 room.

7- Check compressed air pressure (should read 80 psi from the gauge located inside the lab along the corridor wall behind the ELEMENT2). The yellow handle should be parallel with the copper pipe. Trip 000008 means that the compressed air is off or too low Figure 4 8- Switch on the RF generator located behind the instrument bench (the white switch in front of the RFPP unit). Figure 5 (not available yet) 9- Select the fan extraction knob to maximum speed. Figure x (not available yet) 10- Check the Ar-dewar (tank) main gas valve located inside the gas cabinet (the one labelled GAS USE). Pressure should be maintained between 110-120 psi using the stainless steel regulator (SS regulator) and the Pressure Builder valve. The last one is located behind the GAS USE exit port. What to do if the pressure on the SS regulator is less than 80-85 psi: Set the Pressure Builder knob to 120-140 psi (chrome knob that can be turned clockwise or counterclockwise). Then open the valve located below written Pressure Builder. Stay there and watch the pressure increase with time on the SS regulator. Control the amount of building pressure with the Pressure Builder valve and the position of the chrome knob to 95-115 psi. Figure 6 Need another Ar-dewar: You will have to call Roberts Oxygen Co. at: 301-927-5600 and ask for a new 200 liter Ar-dewar delivery (part# RLN178). They provide next day Ar-dewar delivery. The price for a 200 liter of liquid Ar from Roberts Ox. is approx. $140.65 (can be paid by VISA on Roberta s credit card using the PlasmaLab account). For the high-pressure backup cylinders, the part number is R104D3 at a cost of ~$80 (300 ft 3 of Zero Grade Argon). Their price includes the Ar gas delivery at the gas cabinet (dewar or cylinders). 11- Go to next room 0227 (i.e., ICP Preparation Lab). Check the distilled water level reservoir from the front of the big JULABO chiller (level should be between min and max). Switch on the chiller using the green knob located at the rear top on your left and then go back to the instrument room.

Figures 7a, -b 12- Check the vacuum conditions on Pirani-1 ( 1.5E-3 mbar), Pirani-3 gauge ( 2.5E-4 mbar), Penning gauge ( 3.6E-8 mbar) and the ionic pump ( 1.9E- 9 mbar) (refer to the NU Plasma Vacuum Schematic plastic-coated colour picture for further information). If vacuum state is satisfied then proceed ahead. If not, then check the turbo pump speed and ion pump condition (not in standby condition / please refer to the manual for more details or contact me at home: CDalpe @ 9-301-441-1392 or call the factory @ 011-44- 1978-661-304). There is also a section below concerning the different steps to follow when the instrument has experienced a power shutdown. 13- Double-click with the left mouse button the Cool Gas valve gray square at the bottom right of torch schematic (inside the Plasma Control window) in order to flush trapped air inside the torch. This will help plasma ignition. For the Aridus user: Turn on the Aridus from behind the beige box and install the spray chamber + the neb (T1H from CETAC or PFA from Elemental Scientific Inc.; referred as ES) on the Aridus. Click on the Open Valve 1 inside the Switches menu from the Nu Plasma software and then regulate the pressure to neb via the black regulator on the Aridus table to 0 psi (i.e., no pressure). Preset the Sweep gas flow rate between 0.8-1.0 L/min. If you use the CETAC spray chamber you must select the set heater temperature to 70 C max (VERY IMPORTANT). If you use the ES spray chamber (PFA) you can select the heater temperature up to a maximum of 100-105 C (normally we run at 100 C). To change the temperature heater for the Spray Chamber press the blue set button and then simultaneously press the up or down grey arrow button to reset to the new temperature. Wait until both heaters reach their setting point temperature (i.e., 70 or 100 C for the spray chamber and 160ºC for the desolvator oven ). During this stage set the neb gas flow rate from the Plasma software to 0 L/min. No need to set the mass flow controller when you connect the Aridus. If you are not ready to ignite the plasma close the neb selenoid valve by removing the tick located beside Neb Gas under Plasma inside the Digital Control window. By doing this you will avoid wasting Ar gas. Figures 8a, -b, -c, -d For the Cyclonic Spray Chamber: Make sure that you connect correctly the cooling system of the NESLAB chiller (Tygon tubing from the RTE-101 unit), the drain to the Peripump

and the micro-concentric nebulizer to the spray chamber. Installed correctly the Black/White Tygon tubing around the perilstatic pump (pump turn clockwise) and the end part inside the Waste bottle (the one that has an hole in its lid). For the Scott-Double pass + Peltier Cooling system: Connect properly the neb and drain to the spray chamber. Then make sure the Peltier option in the Plasma Control panel selected. You can modify the cooling temperature by pressing the up or down arrows (normal setting point around 4 ºC). The 4 mm blue polyurethane tubing must be connected to the base of the Peltier heat sink in order to remove the extra heat during the run before switching ON the large Julabo chiller. The large Julabo chiller is responsible to cool down the heat sink. 14- Insert the Polymer, PTFE or PFA sampler tubing into MilliQ water or 2% HNO 3. 15- The instrument is now ready for ignition. 16- Press the green START button inside the Plasma Control window from the Nu Plasma software and watch the process. Should take approx. 90-120 sec. 17- It might be possible that you will need to adjust the Air Capacitor position (up or down arrows) during the ignition. However, I strongly recommend that you allow the software the first three trials. A good guess for the Air Capacitor and the Vacuum Capacitor should be around 95-100 and 6500-6600, respectively, when a new plasma shield is used. 18- After ignition: For Aridus users: - Increase the Ar-sweep gas with the CETAC mass flow controller to 2.5 L/min and the neb pressure, with the black regulator, to 40 psi. for the PFA-ES nebulizer or 60-80 psi for the T1H-CETAC nebulizer. No need at this point to increase the nitrogen (or helium) additional gas. See Figure (not available yet) - the software will go into a start up routine where the backing valve will close, the first LOS will open, the backing valve will reopen, and then the High Tension (HT) will switch on. 19- Wait 10-20 sec. before switching on the Plasma Shield ON (under the Switches window).

20- You must open manually the second LOS from the Valve Control window if you want to check at this point the instrument signal intensity. To do this double-click on the dark grey symbol representing the second LOS (if you don t have any response at this stage please check if you still have compressed air again; see Pt. 7 above). Quad 1 = set value for your isotopic system (e.g., Quad 1 for Li = -0.3). Quad 2 = set value for your isotopic system (e.g., Quad 2 for Li = 46). Those Quad values can be modified for other isotopic systems under Setup menu in Quad Lens Setup. Also please make sure that those lenses are set to 0 volt: LIN 1 = 0.00 LIN 2 = 0.00 LIN Corr = 0.00 Cubic = 0.00 Q15 = 0.00 Q16 = 0.00 under Deflector and HT Supplies Control QUADS). 21- A trip message such as: 0000001 or 0000101 or 0000100 or an ERROR message such as: 00200000 means the neb flow rate generated from the Aridus is too low. Most of the time this is created by a discharge between HV1 and HV2. Adjust the neb pressure (black regulator) and sweep gas according to the step #17 above and then reset the trip by clicking under Protection menu the Reset Trips bar. If you still have the appearance of the trip, switch off the PlasmaShield (under Switches), switch off the HT (under Digital Control Power), reset the trip again (should disappear at this point if not please call CDalpe). Then switch on the HT wait 10-15 sec switch on the PlasmaShield. The instrument should be fine from then. You should never get involve with this hassle if you use a regular spray chamber such as the Scott double pass or the Cyclonic models. 22- Let the instrument sit in this configuration for the next 20-30 minutes in order for the instrument to achieve its new vacuum conditions and plasma stability. 23- Typical Signal Intensity: Signal intensity can vary from day to day basis. However, it is strongly dependent regarding the sample system introduction

you use, the neb aspiration rate (Micromist, PFA-Elemental Scientific, T1H- Cetac, etc.), tubing I.D., gas flow, quality of cones, tuning configurations, etc. 24- Tuning steps: I strongly recommend to start with: - Sweep Gas flow rate (between 2.00-5.00 L/min for Ar and 8-25 ml/min for N, no miracle values). Make sure if you use nitrogen that the Nitrogen cylinder valve is open and the regulator is set to 50 psi. (cylinder located at the left of the big chiller in 0227). - the source (torch position controlled via the Up/Down In/Out Backward/Forward arrows under Plasma Control window, neb pressure/flow rate (neb regulator or neb mass flow controller), then - HV1 (the plateau intensity should be achieved between 4002-4008 volts) - HV2 - Deflectors Source H1 and Source V1 - HV3 - Deflector Source H2 Then into the transfer chamber: - HV6 - Deflectors Transfer V2 - Deflectror Transfer H1 - HV5 - Deflector Tranfer V1 (Please refer to the plastic-coated Location of Nu Plasma Ion Beam Steering and Focusing Potentials color diagram). 25- For further instruction please refer to the white binder entitled Nu Plasma Operating Manual located in my office.

How to Shut-off the Plasma 1- Clean the Polymer/PTFE tubing of the neb with 2 or 5% nitric acid for 90 sec and then with MilliQ water for another 90 sec. CETAC recommends to aspirate air for their Aridus for another 90 sec in order to dry the desolvator cell before shutting down the Aridus. 2- Switch OFF the Plasma Shield from the Switches window. 3- Press the red STOP button from the Plasma Control window. Aridus Users: - Press the Close Valve 1 from the Switches window under Plasma software (this will close the argon line supplying the desolvator cell, i.e. sweep gas, and the nebulizer gas, i.e. the black regulator). Turn counterclockwise the black knob on top of the small regulator until you feel no friction; i.e. minimum pressure. - Switch off the power supply of the ASX-100 (the red knob located behind the autosampler). - Switch off the Aridus using the black switch located behind the unit (located below the Ar In gas tubing). - Close all main gas valves of each high pressure cylinders located in 0227 (Nitrogen and/or Helium; clockwise). 4- Switch off the big JULABO chiller in room 0227 (the green knob located near the top left, on front of the chiller). 5- Switch off the RF generator located behind the instrument bench (the white switch in front of the RFPP unit). Figure x (not available yet) 6- Select the fan extraction knob to minimum speed. Figure x (not avalable yet) 7- Close the green GAS USE valve from the Ar-dewar (turn clockwise). Close the Pressure Builder valve (if it was open). Figure 6

8- Export your data to the computer via the network. 9- Shutdown the Window NT on the Gateway computer. The monitor will turn to sleep mode 20 sec. after you shut down the computer. 10- Remove Polymer or PTFE tubings from any solution to avoid spillage during night. 11- Release the tension on each tubing attached to the perilstatic pump (Gilson Peripump; Aridus Peripump located behind the unit). Figure 10 (not available yet) 12- I recommend to remove the spray chamber + neb from the Aridus heater and then to store both parts under the fume hood in 0227. Be careful the spray chamber might be very HOT. Please cover the hole of the spray chamber oven with Parafilm + masking tape to avoid falling debris from the HVAC system.

How to Reset the Instrument Configurations After a Short Power Outage 1- From the Nu Plasma software go to the SETUP, than under preferences unclick Run OFF Line. 2- Again from the SETUP menu click on Mass Spec and than answer YES (it will take a few seconds). 3- Again from the SETUP menu go to Constants, Instrument, and then Initialise. Answer YES. 4- Under the Protection menu bar clear the tick beside the Instrument Settings and the Vacuum. 5- Go to Setup, Contants, Tuning, Initialise (e.g., Lithium-2 (1mm spacer.tun; for lithium setting conditions). 6- The #$@ Trip is still there don t panic it can be worse The problem can be resolved if you reset manually the Discriminator Voltage on each Ion Counter (IC1, IC2, IC3). To do this you must go under the IC Supplies window and then under Disc you insert manually the new values of 0.110 (minus), -0.190 (minus), -0.160 (minus) Volts for IC0, IC1 and IC2, respectively (instead of 0.000 volt). 7- Go to Protection and click again on Reset Trip. The trip should be erased from then. 8- Ion Pump. You must restart the Ion Pump Power Supply manually following those steps: - Walk behind the instrument near the RF PP Generator. The Ion Pump Power Supply is located just above the RF PP written Digital MPC for Multiple Pump Control. From the Supply 1 (top green Digital Display ) it should be written at this stage Standby *** (if not then the Ion Pumps should work properly) *** Figure 10 (not available yet)

This means that the Ion Power Supply do not provide any voltage to the three Ion Pumps. From the white keyboard press: - (1) High Voltage Operate (dial #9) - (2) Select on with the Display Select button. - (3) Press then the Enter white button. The voltage should increase gradually up to 6-7 kv (approx. 6800 Volts). 9- You should be ready to operate the instrument in normal condition from then. Go to Point # 10 (under How to Start the Nu Plasma MC-ICP-MS ) to verify the vacuum conditions (mostly the Analyser ). You might have to restart the Penning gauge manually under Digital Control and then tick the Penning gauge box. It will switch to OFF state if the vacuum condition inside the Transfer chamber is above 5E-6 mbar. 10- Good luck! *** END *** Dr. Claude Dalpé 301-405-7504 (office); 301-405-5350 (lab); 301-441-1392 (home)