Pinion Products Corporation Inc. ESD and Ionization Instruments for Industry INSTRUCTION MANUAL COMBINATION RESISTIVITY METER, MODEL SRM/RTG

Similar documents
INSTRUCTION MANUAL SURFACE RESISTIVITY METER SRM-110

746 Wrist Strap Tester Installation, Operation and Maintenance

OPERATING INSTRUCTIONS

Physics Experiment 9 Ohm s Law

INSTRUCTION MANUAL CAUTION

General Electrical Information

AUTORANGING RESISTANCE INDICATOR

LAB 7. SERIES AND PARALLEL RESISTORS

Ohm Metrics Continuous Workstation Monitors Model SRM330

AUTO CHARGE 3 STEP AUTOMATIC BATTERY CHARGER

Phase Sequence Tester

OPERATING INSTRUCTIONS

Model Battery Powered Milliohm Meter

VARIABLE RESISTANCE REFERENCE PAR-809C. User Manual

ALPHA 4314 DIGITAL IGNITER TESTERS

Harris IRT Enterprises Digital Resistance Tester Model XP

RF80-K Aircraft Battery Charger / Analyzer

DLRO 10 and DLRO 10X. Digital Microhmmeter. DLRO 10 and DLRO 10X Digital Microhmmeter. NEW interchangeable test lead terminations

MODEL 422 Submersible Pump Controller

DESIGN VERIFICATION TEST REPORT HLE/TSM HLE G-DV/TSM G-DV

CGD8800X CGD8800X220 Combustible Gas Detector

MODEL 901 OPERATING INSTRUCTIONS

AD592/592-10K Non-Room Sensors in BAPI-Box, BAPI-Box 2 or BAPI-Box 4 BA/592-x-(BB, BB2, BB4) Temperature Sensor Instructions & Operation

3000 SERIES DRY-BLOCK HEAT SOURCE

Panel Mount Digital Tachometer Model

Chapter 3. ECE Tools and Concepts

FLUSH EYES IMMEDIATELY WITH WATER. GET MEDICAL HELP FAST. SULFURIC ACID CAN CAUSE BLINDNESS OR SEVERE BURNS.

ACL 780/781 Economy Combo Tester

Standard Operating Procedure Four Point Probe System

EXPERIMENT 4 OHM S LAW, RESISTORS IN SERIES AND PARALLEL

INSTRUCTION MANUAL CAT Bio-Pump Plus

POCKET-TACH 100 MONARCH INSTRUMENT Rev 1.4. Instruction Manual. Printed in the U.S.A. Monarch Instrument 2001 all rights reserved

Digital Multimeter: This handheld device is used by this course to measure voltage and resistance we will not use this to measure current or capacitan

Hand Held Pull Tester Instruction Manual Order No

AD592/592-10K Non-Room Units with a Junction Box, Weatherproof (WP) or Weather Tight (EU) Enclosure

Buck BioAire Model B520 Instruction Manual

Cable Length Meter INSTRUCTION MANUAL

RF80-M. Features. Lead Acid Batteries Multistep Programmable. Automatic Operation Stores 100 Programs

Indicative Battery Testers. Intelligent Battery Tester Chrome Tester OPERATIONS MANUAL. IBT Chrome Tester BT IBTCHR

D-2 Incorporated. JF-1A-RC RampCheck Product Brochure. Part Number: JF-1A-RC Revision Date: 09/11/15 Drawing Number: R6

Utilization of Electric Power Laboratory 3 rd Year G2: Testing & Characteristic of MCCB Used in Commercial and Industrial Applications

model ps600 Address all communications and shipments to: FEDERAL SIGNAL CORPORATION

Larson Systems Inc. Digital Hand Tester User Manual

The D.E. Stearns Company Shreveport Louisiana U.S.A. D. E. STEARNS MODEL 14/20 REGULATED ,000 VOLT HIGH VOLTAGE HOLIDAY DETECTOR

User's Guide. Vane Anemometer. Model Measures Air Velocity & Temperature. Degrees C/F switchable

CruzPro. Volts Amps Amp-Hour Monitor VAH110. Other CruzPro Products. website:

Pull-Off Adhesion Tester

AMETEK Model PPC Series Electronic Pressure Calibrator

CHAPTER 19 DC Circuits Units

QUICK START GUIDE FOR DEMONSTRATION CIRCUIT 1061A LINEAR LI-ION BATTERY CHARGER WITH DUAL SYNCHRONOUS BUCK REGULATOR

BellMega BM-510. Operation Manual Ver: 1.0 Rev: 1

Informational Distribution List:

PIECAL 322 Automated Thermocouple Calibrator Operating Instructions. Product Description

User Guide. Model Insulation Tester / Megohmmeter. Introduction

PHYSICS 111 LABORATORY Experiment #3 Current, Voltage and Resistance in Series and Parallel Circuits

PROCESS INSTRUMENTATION

SECTION 3 BASIC AUTOMATIC CONTROLS UNIT 15 Troubleshooting Basic Controls

USER MANUAL RHF SERIES

Laboratory 2 Electronics Engineering 1270

BUSSMANN. Pre-production and sample electric vehicle power fuses 500 Vdc, A SERIES. Technical Data Effective December 2018.

MODEL 905V OPERATING INSTRUCTIONS

TABLE OF CONTENTS. 8.3 Online Sensor Specifications. Series TS (Models TS1, TSP, TSH, TSL,TSF, TSB1, TSB2) According to Electromatic factory procedure

BASEWEST Operating & Maintenance Manual. Model TS-420 Test Set. 28 September 2010

HIGH ACCURACY MICROOHMMETER AND RTD MONITOR MODEL R1L-D1

Product Guide: Series III Pump Control Board Set (RoHS)

DC/DC CONVERTER Designing Battery Charger

UNIT 3: GENErAL ELECTriCAL SySTEM DiAGNOSiS

CONTENTS DIVERSE. Preface SQUEEZE ANALYSER SQA2 RESISTANCE WELDING FORCE MONITOR. Resistance Welding. Squeeze Analyser OPERATING INSTRUCTIONS

PHY132 Practicals Week 5 Student Guide

Digital/Analog Megohmmeter Model 1026 USER MANUAL

MODEL AP/W HIGH VOLTAGE HOLIDAY DETECTOR (NEW PRODUCT RELEASE 2015)

SAMTEC POWER CHARACTERIZATION

METEOROLOGICAL INSTRUMENTS

Powerterm L120C Single Output PSU/Battery Chargers Model C2199A-1 (12V/8A) or Model C2199A-2 (24V/6A)

Virtual Measurements & Control. VC210 Operators Manual

Buck BioSlide Model B1020 Instruction Manual

Condensed Version of Operator s Manual For The VI/VP700 Series Flowmeter and VI800 Series Flowmeter

OPERATING INSTRUCTION D. E. STEARNS MODEL 14/20 REGULATED ,000 VOLT HOLIDAY DETECTOR

12/24 Volt Trailer, Caravan Electrical Tow Plug And Light Tester

AUTO CHARGE AMPS DELUXE AUTOMATIC BATTERY CHARGER. Section 1: Single Battery Systems

MultiCal. Operation Manual for MultiCal mmhga/psia Pressure Module

RF80-M. Features. Lead Acid Batteries Multistep Programmable. Automatic Operation Stores 100 Programs

SC10F Circuits Lab Name:

Model No. SB1B-14 Linear Standby Regulator. B & C Specialty Products P.O. Box B Newton, KS (316)

PosiTest. Pull-Off Adhesion Tester. INSTRUCTION MANUAL v PosiTest AT-M (manual) PosiTest AT-A (automatic)

QUICK START GUIDE FOR DEMONSTRATION CIRCUIT 1020 HIGH EFFICIENCY USB POWER MANAGER + TRIPLE STEP-DOWN DC/DC LTC3555

ESD Static Dissipative Dispensing Components

Activator 282 Aircraft Battery Maintenance Charger Operator Manual

CHAPTER 2. Current and Voltage

Instrument Identification. Customer Instrument ID: S Order #: Calibration Standards TIMER CURRENT

Solar Powered Chameleon Smart Globe

August 2006 Distributor/Dealer Service Instructions for: Safety Recall F33 Directional Signal Bulb Function

Warnings and Cautions. Precision Flow Insertion Mass Flow Meter. Instruction Manual BULLETIN F-41-PFS FR# R

User's Manual. High Voltage Megohmmeter. Analog Insulation Tester plus AC Voltage and Continuity Tests. Model

CBAM Load Share (LSL) Series

Application Note. Monitoring Bearing Temperature with ProPAC

Electricity and Magnetism Module 2 Student Guide

Force Sensing Resistor Technical Guidelines Rev 0 ( ) by M. Wagner

COATING THICKNESS GAGE NOVOTEST TP-1

VSO Thermally Compensated Proportional Valve

Transcription:

Pinion Products Corporation Inc. ESD and Ionization Instruments for Industry 32086 Share 28 Road Los Fresnos, TX 78566 (956) 233-8774 / FAX (956) 233-8924 Email: sales@pinionproducts.net www.pinionproducts.net INSTRUCTION MANUAL COMBINATION RESISTIVITY METER, MODEL SRM/RTG IMPORTANT!! READ THIS MANUAL BEFORE OPERATING THIS UNIT. Page 1 of 5 SRM/RTG Manual 14 December 2007

THEORY OF OPERATION The Pinion/Voyager model SRM/RTG Combination Surface Resistivity meter is a simple to use, highly repeatable measuring device for determining the surface resistivity of materials, as well as resistance to ground. The SRM/RTG incorporates a revolutionary three-point probe measuring technique, enabling accurate, consistent, and repeatable surface measurements. The principle of operation is as follows: A voltage, V (30 Volts), is applied to each of the three outer probe contacts (Figure 1). The center probe contact (spring loaded) provides a ground return of the current supplied by the three outer probe contacts. The surface resistivity of the material being measured is expressed as ten to some power, written as Ohms/Square = 10^x; where x is rounded to the nearest integer by the SRM/RTG with the following exceptions: The 10^5 range covers10^4 to 10^5.5. This allows the first display segment to indicate surface resistivities less than 10^4 Ohms/Square. By studying the relationship between the outer probe contacts and the center probe contact, the following is evident: The diameter of the probe contact is 1/8 inch, and the distance from the outer contact to the center probe is 1-½ inches. The approximate number of squares for a 1/8-inch strip of material between the center probe and one of the outer probes is 12 (Figure 2). Three 12 square areas configured in parallel yield a 4 square resultant area (Figure 3). The configuration in Figure e however still does not accurately include the entire area measured by the SRM/RTG probe arrangement. A more representative depiction of current flow includes the entire circular area (Figure 4). There are more paths in parallel further reducing the number of squares. The resulting number of squares was experimentally determined to be 0.78. Experimental data was taken to determine the effect of fringe current that exists outside the circular area encompassed by the probe arrangement. These fringe affects were measured to be less than 1% (Figure 5). The conductive elastomer probe contacts provide excellent contact to the surface under test. For this reason, great pressure is not needed to perform a measurement. Each probe contact of the SRM/RTG applies a pressure of 40 lbs/in^2, due to the 2 pound weight of the SRM/RTG, which is sufficient. The three-point probe design enables the unit to be balanced on irregular and curved surfaces. SPECIFICATIONS: Display Segment Resistance Range (Ohms/Square) 10^4 <10^4 10^5 10^4.4 10^5.5 10^6 10^5.5 10^6.5 10^7 10^6.5 10^7.5 10^8 10^7, 5 10^8.5 10^9 10^8.5 10^9.5 10^10 10^9.5 10^10.5 10^11 10^10.5 10^11.5 10^12 10^11.5 10^12.5 10^12.5 >10^12.5 Accuracy: Indicated exponent +/-.3 Power supply: Rechargeable NiCad or NiMH batteries (2 x 2.4volt packs). Test Voltage: 30 Volts +/- 2 Volts current limited to 2 ma. Contact Area: 0.0123 sq in/probe Contact Pressure: 40-lbs/sq in/probe. Contact material: Conductive rubber. Dimensions: 3 H x 4 D x 5 W Weight: 2 pounds Calibration Procedure: The SRM/RTG should be calibrated annually. If any step fails as per the following instructions, the unit should be returned to Pinion Products Corporation Inc. for repair and calibration. Pinion Products Corporation Inc. can provide either a standard NIST Calibration with certification traceable to NIST and after calibration data, or a MIL calibration with certification traceable to NIST and both before and after calibration data. 1.) The batteries should be charged before beginning the calibration procedure. Page 2 of 5 SRM/RTG Manual 14 December 2007

2.) Remove the 4 screws, two each side, from the unit, and set the cover to the side. Leave the wires attached so that there is access to the printed circuit board. Turn the unit ON. 3.) All voltage measurements are made relative to the unit ground. The test point for ground is adjacent to the 15 and +15 volt test points. 4.) Use a 4-½ digit digital voltmeter such as a Keithley Model 175 Multimeter to measure the voltage on the 30VTP. Record this voltage as V30. 5.) Measure and record the voltage at the +15V test point as V15. 6.) Calculate and record I(ref) using the formula: I(ref)= V30 *.04054 7.) Calculate and record R (tp1) using the formula R (tp1) = (V15-.236)/ (I (ref)). 8.) Calculate and record I(O) using the formula I(O) = I(ref) x 1.016), (exponent 3) 9.) Calculate and record I(6) using the formula I(6) = I(ref) *.992), (exponent 9) 10.) Turn the unit off and remove the shorting jumper connected to TP1 and its adjacent pin. Measure the resistance between the +15V test point and TP1 lead closest to R7. Adjust P1 until the measured resistance is equal to the calculated value RTP1 +/- 1%. 11.) Replace the jumper on TP1 and the adjacent pin. Turn the unit on. Connect a current source (e.g. Keithley Model 220) to the side of R6 connected to the large round pad in the center of the board. Apply a current equal to the calculated I (O). Measure the voltage on the test pin marked OUT and adjust P2 until the measured voltage is 0.000 VDC +/-.050 VDC. 12.) Apply a current equal to calculated value I(6) to the same point as in step 11. Measure the voltage on the test pin marked OUT, and adjust P3 until the measured voltage is 6.000 VDC +/-.050 VDC. 13.) Apply a current equal to.003^-9 amperes. Measure the voltage on the test pin marked OUT, and adjust P6 for a reading of 8.5 VDC +/-.05VDC. The 10^12.5 led should be illuminated. 14.) Apply a current equal to.03^-9 amperes. Measure the voltage on the test pin marked OUT, and verify a reading of approximately 7.5 VDC. The 10^12 led should be illuminated. 15.) Measure the voltage on the pin marked +8.000 VDC and adjust P4 for a reading of 8.000 VDC +/-.050 VDC. 16.) Measure the voltage on the pin marked 2.000 VDC and adjust P5 for a reading of 2.000 VDC +/-.050 VDC. 17.) The calibration procedure is now complete. Place the cover back on the base and replace the screws. Operating Instructions: DO NOT operate the SRM/RTG with the battery charger connected. The charger will introduce electrical noise which will render readings inaccurate. Turn off the SRM/RTG when it is not being used, and charge ONLY with the supplied charger when the Low Battery indicator is illuminated, or at the end of the day s use. Charge for a minimum of 12 hours. Do NOT leave the charger connected continuously. SRM mode 1.) The SRM/RTG is extremely easy to operate. Simply turn the unit on in the SRM mode, and place it on the surface to be measured. The illuminated display segment indicates the resistivity on the surface being measured. To insure accurate readings, place small or thin samples on non-conductive surfaces (> 10^14 ohms/sq). Ideally, a clean glass surface. 2.) Measurements in the 10^11 to 10^12.5 may require a few seconds for the illuminated display to stabilize. RTG Mode 1.) Connect the supplied cable from the black banana jack on the rear of the unit to your ESD ground connection, turn the unit on in the RTG mode, and place it on the surface to be measured and read the display. 2.) Measurements in the 10^11 to 10^12.5 may require a few seconds for the illuminated display to stabilize. 3.) The indicated reading will be resistance to ground readings. 4.) If grounding cable is not connected to a proper ground, several LEDs in the display may illuminate. This is normal. Page 3 of 5 SRM/RTG Manual 14 December 2007

Calibration Check: 1.) This is a Quick calibration check to verify the SRM/RTG is performing within tolerance. 2.) Connect a voltmeter between each of the three outer probe contacts relative to ground (one at a time). The voltage should equal 30V +/- 2 V at each probe. 3.) Connect a programmable current source to the center probe contact and ground of the SRM/RTG. 4.) Apply the currents in the following table and check that the appropriate display segment is illuminated. Current (Amperes) Display Segment Tolerance 1.45 * 10^-3 <10^4 1% 4.95 * 10^-4 10^5 1% 3.77 * 10^-5 10^6 1% 3.84 * 10^-6 10^7 1% 3.85 * 10^-7 10^8 1% 3.85 * 10^-8 10^9 1% 3.85 * 10^-9 10^10 1% 3.85 * 10^-10 10^11 1% 3.85 * 10^-11 10^12 1% 6.00 * 10^-12 10^12.5 35% 5.) The calibration check is now complete. If any step failed, the SRM/RTG requires calibration. Please notify Pinion Products Corporation Inc. that the SRM/RTG failed its calibration check, and information will be supplied on the procedure for having the SRM/RTG calibrated. Available accessories: SFS-100 Conductive Rubber probes (set of 4) BFS-100 1 Diameter Conductive Rubber Probes (set of 4) 6550-0011 Replacement Battery Pack (each SRM contains two battery packs) 6120-0035 Replacement Battery Charger for SRM 4000-0001 Carry Case Please visit our website for details at www.pinionproducts.net Page 4 of 5 SRM/RTG Manual 14 December 2007

Page 5 of 5 SRM/RTG Manual 14 December 2007