Safety Instructions Proline Promass 300

Similar documents
CERTIFICATE OF COMPLIANCE

Rosemount Product Data Sheet , Rev GA Catalog FIELDBUS PROTOCOL. Hazardous Locations Certifications

Certificate of Compliance

Ex Instruction manual. Rotamass Coriolis mass flow meter. IM 01U10X03-00EN-R, 1st edition, Rev FM-1

Resistance Thermometers Model TR217, with Spring-Loaded Probe Tip

Magnetic-inductive flow meter for hazardous areas

Assembly. Trunk IN + - S U, I U, I. Spur 1

The filling level sensors are designed in accordance to the following standards

Certificate of Compliance

OPERATING INSTRUCTIONS

Terminal Boxes Series 8125

Certificate of Compliance

Certificate of Compliance

Model 8800C Vortex Flowmeter

Certificate of Compliance

E II 3 (1) G Ex na [ia Ga] IIC/IIB T4 Gc

Certificate of Compliance

Operating Instructions

IECEx Certificate of Conformity

Proportional pressure reducing valves type DHRZO and DHRZA standard and ex-proof version, direct operated, ISO 4401 size 06

ANKERSMID Temperature controller ATC 510/520/525 for wall-mounting. Application

Endress+Hauser Wetzer GmbH+Co. KG Obere Wank Nesselwang Germany. Field Temperature Transmitter itemp, Types TMT82, TMT84 and TMT85

XMP ci. Process Pressure Transmitter with HART -communication. Ceramic Sensor. accuracy according to IEC 60770: 0.1 % FSO.

INSTRUCTION MANUAL (ATEX / IECEx)

Resistance thermometer For additional thermowell, flameproof enclosure (Ex d) Model TR10-L

IECEx Certificate of Conformity

Safety instructions VEGAVIB VB6*.GI**C/R/T/Z/N*** IECEx BVS Ex ta, ta/tb, tb IIIC T Da, Da/Db, Db IP66

Thermocouple For additional thermowell, flameproof enclosure (Ex d) Model TC10-L

Data Sheet DS/NGB-EN Rev. B. Multi barriers. Measurement made easy. Fieldbus barrier for PROFIBUS PA for FOUNDATION Fieldbus-H1

Solenoid Valves Type 3963

PLUGS & RECEPTACLES. INNOVATIVE EXPLOSION PROTECTION by R. STAHL

Type Electronic Limit Switch

INSTRUCTION MANUAL (ATEX / IECEx)

Resistance thermometer Without thermowell Model TR10-H

Data Sheet SCHISCHEK EXCOS-P. Pressure sensor 20 Pa Pa. Supplied by. .com. Call us on +44 (0)

VALIADIS S.A. HELLENIC MOTORS

INTERPLANT STANDARD - STEEL INDUSTRY. Specification for Smart Pressure & Differential Pressure Transmitter

TERMINAL BOXES FOR HARSH & HAZARDOUS ENVIRONMENTS

SynergEX Panelboards for Hazardous Locations

Pneumatic control unit for decentralized automation of ELEMENT process valves

Flameproof Level Sensor (Flameproof F/S FP)

Technical Data. Dimensions

JBEP Series Fiberglass Reinforced Polyester Terminal Junction Boxes for Instrumentation Applications Increased Safety. Factory Drilled and Equipped

PF 3100 ION PILOT CARD PRO.FIRE

Terminal Boxes in Stainless Steel SLS**.T

Sensor boxes SRBE Key features and product range overview

RedCos-P Pressure sensor 20 Pa Pa

II 2GD. xxxx. Westlock 2200 Series Quantum - ATEX certified Installation & Operating Instructions

Direct-acting 3/2 way plunger valve

PHOENIX CONTACT - 01/2008

CombiTemp Temperature Measuring System

Ex-proof proportional relief valves Direct or pilot operated, without transducer - ATEX, IECEx, EAC or CULUS

Attachment to Certificate IECEx PTB X

Installation Instructions Inspection Ports

Differential pressure gauge with output signal With integrated working pressure indication (DELTA-trans) Model DPGT40

CYCLOPS Y Purge Indicator

Description. Fieldbus Optimized Through Junction Boxes

ILED Dorado. User Manual

Intrinsically safe solenoid valves on/off controls - ATEX or IECEX certification

APPLICATION MANUAL. Universal Explosionproof Two-Compartments Instrument Housing Type: XD-ID100, XD-ID100win, XD-ID100H, XD-ID100Hwin

T 3965 EN Type 3965 Solenoid Valve Island To control pneumatic actuators

System Electropneumatic Converters (Proportional Valves) Electronic Process Controllers Signal Converters

Power (W) Tamb 40 o C LX1GAM4LB57 5,700 10, T3 T100 o C T4 T85 o C 13

DMK 331P. Industrial Pressure Transmitter. Pressure Ports With Flush Welded Stainless Steel Diaphragm

Low Flow Transmitter EJX115A and EJA115E

LIMIT SWITCH Hawkeye - HX Magnetic Solid State Proximity Sensors Installation, Maintenance and Operating Instructions (HX )

Vortex Flowmeters VFM 3100 F-T VFM 3100 W-T

Temperature Measurement Resistance thermometers

RedBin-P Pressure switches 5 Pa Pa

IECEx Certificate of Conformity

Reg No: 1999/027771/07 ACCREDITED AND APPROVED TEST LABORATORY IN TERMS OF ARP 0108: REGULATORY REQUIREMENTS FOR EXPLOSION PROTECTED APPARATUS

SITRANS P measuring instruments for pressure

Measuring insert for resistance thermometer Model TR10-A

Fieldbus Segment Protector F2-SP-IC* Features. Assembly. Function. Connection. Zone 2 Div. 2

Magnetic float switch For vertical installation Model FLS (models with Ex approval: 60, AL-ADF)

POSITIONERS and Controls for Complete Valve Automation

Surface thermocouple Model TC50

Load disconnect switch Series 8146/5-V11

Thermocouple For additional thermowell Model TC10-B

SynergEX Panelboards. Installation & Maintenance Information IF Panelboards for Hazardous Locations SAVE THESE INSTRUCTIONS FOR FUTURE REFERENCE

Temperature sensor TE2

HTSB system System overview

SITRANS P measuring instruments for pressure

EG Series Installation Manual

SCHISCHEK REDBIN-D TRANSMITTER

Options guide. Electromagnetic flowmeters. M9xx

IBExU Institut für Sicherheitstechnik GmbH

Mid-West Instrument. Piston Type Model 220. Hazardous Locations. Indicating / Non-Indicating Differential Pressure Switch or Transmitter. Min.

Compact Sheet-steel Emergency Light Fitting Series 6118/1

Thermocouple Without thermowell Model TC10-H

SITRANS T measuring instruments for temperature

The Earth-Rite MGV is designed to enable operators establish safe grounding of their vehicle in accordance with these recommended practices.

BETA SWITCHES FOR HAZARDOUS AREA

DMP 333. Industrial Pressure Transmitter For High Pressure. Stainless Steel Sensor

Mid-West Instrument. Piston Type Model 220. Hazardous Locations. Indicating / Non-Indicating Differential Pressure Switch or Transmitter

Proline Promag 10H, 50H, 53H, 55H

Smart Valve Positioner 700 Series with HART Communication Protocol

Ex m Solenoid Operator Type 0519

Proline Prosonic Flow 91W

SPOOL VALVES solenoid air operated 1/4 to 1/2 tapped body and NAMUR for integrated pilot

Transcription:

XA01472D/06/EN/01.16 71327890 Products Solutions Services Safety Instructions Proline Promass 300 Class I, Zone 1, AEx/Ex de ia IIC/IIB Class I, Zone 1, AEx/Ex d ia IIC/IIB Zone 21, AEx/Ex tb IIIC Txx C C US Document: XA01472D Safety instructions for electrical apparatus for explosionhazardous areas classified according to the National Electrical Code (NEC) and Canadian Electrical Code (CEC) 3

Proline Promass 300 XA01472D Proline Promass 300 Table of contents Associated documentation... 4 Manufacturer's certificates... 5 Extended order code... 5 Safety instructions: General... 8 Safety instructions: Installation... 9 Safety instructions: Zone 21, Class II, Class III... 10 Temperature tables... 10 Connection data: Signal circuits... 20 Endress+Hauser 3

XA01472D Proline Promass 300 Associated documentation All documentation is available: On the CD-ROM supplied (not included in the delivery for all device versions). Available for all device versions via: Internet: www.endress.com/deviceviewer Smart phone/tablet: Endress+Hauser Operations App In the Download Area of the Endress+Hauser web site: www.endress.com Download This document is an integral part of the following Operating Instructions: Measuring device Documentation HART FOUNDATION Fieldbus PROFIBUS PA Modbus RS485 Promass A 300 BA01482D BA01515D BA01504D BA01493D Promass E 300 BA01484D BA01517D BA01506D BA01495D Promass F 300 BA01485D BA01518D BA01507D BA01496D Promass H 300 BA01486D BA01519D BA01508D BA01497D Promass I 300 BA01487D BA01520D BA01509D BA01498D Promass O 300 BA01488D BA01521D BA01510D BA01499D Promass P 300 BA01489D BA01522D BA01511D BA01500D Promass Q 300 BA01490D BA01523D BA01512D BA01501D Promass S 300 BA01491D BA01524D BA01513D BA01502D Promass X 300 BA01492D BA01525D BA01514D BA01503D Additional documentation: Document type Contents Documentation code Special documentation Safety instructions Control drawing Remote display and operating module DKX001 Remote display and operating module DKX001: Class I, Zone 1, Zone 21 SD01763D XA01499D As wanted on the nameplate. Please note the documentation associated with the device. 4 Endress+Hauser

Proline Promass 300 XA01472D Manufacturer's certificates Certificate number CSA 16.70087388 Notified body CSA: Canadian Standards Association Extended order code The extended order code is indicated on the nameplate, which is affixed to the device in such a way that it is clearly visible. Additional information about the nameplate is provided in the associated Operating Instructions. Structure of the extended order code * * * * * * _ * * * * * * * * * * + A*B*C*D*E*F*G* Device type Basic specifications Optional specifications * = Spaceholder: At this position, an option (number or letter) selected from the specification is displayed instead of the placeholders. Device type The device and the device design is defined in the "Device type" section (Product root). Basic specifications The features that are absolutely essential for the device (mandatory features) are specified in the basic specifications. The number of positions depends on the number of features available. The selected option of a feature can consist of several positions. Optional specifications The optional specifications describe additional features for the device (optional features). The number of positions depends on the number of features available. The features have a 2-digit structure to aid identification (e.g. JA). The first digit (ID) stands for the feature group and consists of a number or a letter (e.g. J = test, certificate). The second digit constitutes the value that stands for the feature within the group (e.g. A = 3.1 material (wetted parts), inspection certificate). More detailed information about the device is provided in the following tables. These tables describe the individual positions and IDs in the extended order code which are relevant to hazardous locations. Endress+Hauser 5

XA01472D Proline Promass 300 Device type Item Order code for Selected option Description 1 Instrument family 8 Coriolis flowmeter 2 Sensor A, E, F, H, I, O, P, Q, S, X Sensor type 3 Transmitter 3 Transmitter type: 4-wire, compact version 4 Generation index B Platform generation 5, 6 Nominal diameter 1 to 350 1: 01 2: 02 to 350: 3E, 3F, 3R Nominal diameter of sensor Basic specifications Position 1, 2 Order code for "Approval" Option selected Position 4, 5 Order code for "Output, input 1" Option selected Type of protection Transmitter Sensor C1 BA, BB, GA, LA, NA, RA, SA, MA CA, CB, HA, TA Cl.I, Zone 1, AEx/Ex de ia IIB... Gb Zone 21, AEx/Ex tb IIIC T** C Db Cl.I, Zone 1, AEx/Ex de ia [ia Ga] IIB... Gb Zone 21, AEx/Ex tb [ia Da] IIIC T** C Db Cl.I, Zone 1, AEx/Ex ia IIB... Gb Zone 21, AEx/ Ex ia tb IIIC T** C Db C2 BA, BB, GA, LA, NA, RA, SA, MA CA, CB, HA, TA Cl.I, Zone 1, AEx/Ex de ia IIC... Gb Zone 21, AEx/Ex tb IIIC T** C Db Cl.I, Zone 1, AEx/Ex de ia [ia Ga] IIC... Gb Zone 21, AEx/Ex tb [ia Da] IIIC T** C Db Cl.I, Zone 1, AEx/Ex ia IIC... Gb Zone 21, AEx/ Ex ia tb IIIC T** C Db C3 BA, BB, GA, LA, NA, RA, SA, MA CA, CB, HA, TA Cl.I, Zone 1, AEx/Ex d ia IIB... Gb Zone 21, AEx/Ex tb IIIC T** C Db Cl.I, Zone 1, AEx/Ex d ia [ia Ga] IIB... Gb Zone 21, AEx/Ex tb [ia Da] IIIC T** C Db Cl.I, Zone 1, AEx/Ex ia IIB... Gb Zone 21, AEx/ Ex ia tb IIIC T** C Db C4 BA, BB, GA, LA, NA, RA, SA, MA CA, CB, HA, TA Cl.I, Zone 1, AEx/Ex d ia IIC... Gb Zone 21, AEx/Ex tb IIIC T** C Db Cl.I, Zone 1, AEx/Ex d ia [ia Ga] IIC... Gb Zone 21, AEx/Ex tb [ia Da] IIIC T** C Db Cl.I, Zone 1, AEx/Ex ia IIC... Gb Zone 21, AEx/ Ex ia tb IIIC T** C Db 6 Endress+Hauser

Proline Promass 300 XA01472D Position Order code for Option selected Description 4, 5 Output, input 1 BA 4-20mA HART CA GA HA MA SA TA 4-20mA HART Ex-i PROFIBUS PA PROFIBUS PA Ex-i Modbus RS485 FOUNDATION Fieldbus FOUNDATION Fieldbus Ex-i 6 Output, input 2 A W/o; B C D E F G H I J 4-20mA 4-20mA Ex-i Configurable I/O initial setting off Pulse/freq./switch output Pulse output, phase-shifted Pulse/freq./switch output Ex-i Relay 4-20mA input Status input 7 Output, input 3 A W/o; B C D E F G H I J 4-20mA 4-20mA Ex-i Configurable I/O initial setting off Pulse/freq./switch output Pulse output, phase-shifted Pulse/freq./switch output Ex-i Relay 4-20mA input Status input 8 Display, Operation A W/o; via communication F G 4-line, illuminated; touch control 4-line, illuminated; touch control + WLAN M W/o; prepared for remote display DKX001 1) Endress+Hauser 7

XA01472D Proline Promass 300 Position Order code for Option selected Description 17, 18 Device model A1 1 O Separate, with remote display DKX001, 4-line, illuminated; 10m/ 30ft cable; touch control 1) DKX001 is approved according to CSA 160686-70030937. Optional specifications ID Order code for Option selected Description Jx Test, Certificate JP Ambient temperature of measuring device 50 C Safety instructions: General Staff must meet the following conditions for mounting, electrical installation, commissioning and maintenance of the device: Be suitably qualified for their role and the tasks they perform. Be trained in explosion protection. Be familiar with national regulations (e.g. CEC or NEC). Install the device according to the manufacturer's instructions and national regulations. Do not operate the device outside the specified electrical, thermal and mechanical parameters. Only use the device in media to which the wetted materials have sufficient durability. Refer to the temperature tables for the relationship between the permitted ambient temperature for the sensor and/or transmitter, depending on the range of application, and the temperature classes. Modifications to the device can affect the explosion protection and must be carried out by staff authorized to perform such work by Endress+Hauser. When using in hybrid mixtures (gas and dust occurring simultaneously), additional measures should be taken. Please see approval body. Open the device only if one of the following conditions is met: An explosive atmosphere is not present. A waiting time of 10 minutes is observed after switching off the power supply. Observe all the technical data of the device (see nameplate). LWARNING Substitution of components is not permitted. Substitution of components may impair intrinsic safety. 8 Endress+Hauser

Proline Promass 300 XA01472D Safety instructions: Installation In the event of potentially explosive vapor/air mixtures, only operate the device under atmospheric conditions. Temperature: 20 to +60 C Pressure: 80 to 110 kpa (0.8 to 1.1 bar) Air with normal oxygen content, usually 21 % (V/V) If no potentially explosive mixtures are present, or if additional protective measures have been taken, the device may also be operated under non-atmospheric conditions in accordance with the manufacturer's specifications. Continuous service temperature of the connecting cable: 40 to +80 C ( 50 to +80 C for optional specification, ID Jx (Test, Certificate) = JP); but at least according to the service temperature range of the application taking into account additional influences of the process conditions (,min and,max + 20 K). Only use certified cable entries suitable for the application. Observe selection criteria as per CEC or NEC. The following applies when connecting the transmitter with a connection compartment in Ex d: Only use separately certified cable and wire entries (Ex d IIC) which are suitable for operating temperatures up to 85 C and for IP 67 protection. If using conduit entries, the associated sealing mechanisms must be mounted directly on the housing. The following applies when connecting the transmitter with a connection compartment in Ex e: Only use separately certified cable and wire entries and sealing plugs (Ex e II) which are suitable for operating temperatures up to 85 C and for IP 67 protection. The cables must be routed such that they are securely seated, and sufficient strain relief must be ensured. When the measuring device is connected, attention must be paid to explosion protection at the transmitter. In potentially explosive atmospheres: Do not disconnect the electrical connection of the power supply circuit when energized. Do not open the connection compartment cover when energized. Install the transmitter circuit wiring according to Canadian Electrical Code (CEC) respective National Electrical Code (NEC) using threaded conduit or other wiring methods in accordance with articles 500 to 510. When connecting through a conduit entry approved for this purpose, mount the associated sealing unit directly at the housing. Seal unused entry glands with approved sealing plugs that correspond to the type of protection. The plastic transport sealing plug supplied does not meet this requirement and must therefore be replaced during installation. Only use certified sealing plugs. The metal sealing plugs supplied meet this requirement. Endress+Hauser 9

XA01472D Proline Promass 300 Intrinsic safety Observe the guidelines for interconnecting intrinsically safe circuits (e.g., Proof of Intrinsic Safety). When the intrinsically safe Ex ia circuits of the device are connected to certified intrinsically safe Ex ib circuits with explosion group IIC or IIB: The type of protection changes to Ex ib IIC or Ex ib IIB. The device can be connected to the remote display and operating module DKX001 with Ex ia explosion protection; refer to the Special Documentation and Ex documentation. If the remote display and operating module DKX001 is used, the internal display unit must be removed from transmitter! Potential equalization Integrate the device into the local potential equalization. If the ground connection has been established via the pipe as specified, it is also possible to integrate the sensor into the potential equalization system via the pipe. Safety instructions: Zone 21, Class II, Class III To ensure dust-tightness, securely seal the transmitter housing, cable entries and sealing plugs. Only open the transmitter housing briefly, ensuring that no dust or moisture enters the housing. Only use certified cable entries. The metal cable entries supplied meet this requirement. When connecting the transmitter to the remote display and operating module DKX001: Connection values 21 Temperature tables Ambient temperature Minimum ambient temperature: = 40 C Optional specification, ID Jx (Test, Certificate) = JP = 50 C Maximum ambient temperature: = +60 C depending on the medium temperature and temperature class 10 Endress+Hauser

Proline Promass 300 XA01472D Medium temperature Minimum medium temperature: Promass A, F, H, I, P, Q, S, X: T m = 50 C Promass E, O: T m = 40 C Promass F, Q with cryogenic temperature version (order code for "Measuring tube material", Option LA): T m = 200 C Maximum medium temperature: T m for depending on the maximum ambient temperature ( ) = The maximum permitted medium temperatures in brackets only apply if the sensor is installed in such a way that the transmitter is not mounted above the sensor and free convection can occur on all sides. Compact version Without thermal insulation Promass A T m 1 4 50 205 50 95 130 150 205 205 60 95 130 150 205 205 Promass E T m 8 50 50 150 50 100 130 130 150 150 55 80 100 130 150 150 60 (80) (100) (130) (150) (150) 80 50 150 50 75 110 150 150 150 55 75 110 150 150 150 60 (75) (110) (150) (150) (150) Endress+Hauser 11

XA01472D Proline Promass 300 Promass F 1) T m 08 15 50 150 50 95 130 150 150 150 60 95 130 150 150 150 50 240 50 95 130 160 240 240 60 95 130 160 (240) (240) 50 150 2) 50 95 100 150 150 150 60 95 100 150 (150) (150) 25 50 50 150 50 95 130 150 150 150 60 95 130 150 150 150 50 240 50 95 130 160 240 240 60 95 130 160 (240) (240) 50 150 2) 50 95 100 150 150 150 60 95 100 150 (150) (150) 50 350 45 85 115 175 275 350 60 85 115 175 275 350 80 50 350 45 85 115 175 275 350 60 85 115 175 275 350 80 250 50 150 50 75 110 150 150 150 60 75 110 150 150 150 50 240 50 75 110 170 240 240 60 75 110 170 (240) (240) 50 150 2) 50 75 110 150 150 150 60 75 110 150 150 150 1) Maximum temperature range see nameplate 2) Cryogenic temperature version: T m = -196 150 C Promass H T m 8 50 150 1) 50 65 100 150 150 150 60 65 100 150 150 150 12 Endress+Hauser

Proline Promass 300 XA01472D T m 8 50 205 2) 50 65 100 160 205 205 60 65 100 160 205 205 15 50 50 150 1) 50 75 115 150 150 150 60 75 115 150 150 150 15 50 50 205 2) 50 75 115 180 205 205 60 75 115 180 205 205 1) Measuring tube material tantalum 2) Measuring tube material zirconium Promass I T m 8, 15, 15FB, 25 25FB, 40, 40FB, 50 50 150 50 95 130 150 150 150 60 95 120 (150) (150) (150) 50 150 50 85 120 150 150 150 60 85 120 (150) (150) (150) 50FB, 80 50 150 50 85 120 150 150 150 FB = Full bore 60 85 120 (150) (150) (150) Promass O T m 80 250 50 205 50 75 110 170 205 205 55 75 110 170 205 205 60 75 110 170 (205) (205) Endress+Hauser 13

XA01472D Proline Promass 300 Promass P 1) T m 8 45 150 45 65 100 150 150 150 60 65 100 150 150 150 45 205 45 65 100 160 205 205 60 65 100 160 205 205 15 50 50 150 50 75 115 150 150 150 60 75 115 150 150 150 50 205 50 75 115 180 205 205 60 75 115 180 205 205 1) Maximum temperature range see nameplate Promass Q 1) T m 25 100 50 205 50 75 110 160 205 205 60 75 110 160 205 205 25 100 50 150 2) 50 75 110 150 150 150 60 75 110 150 150 150 1) Maximum temperature range see nameplate 2) Cryogenic temperature version: T m = -196 150 C Promass S T m 8 45 150 45 65 100 150 150 150 60 65 100 150 150 150 15 50 50 150 50 75 115 150 150 150 60 75 115 150 150 150 14 Endress+Hauser

Proline Promass 300 XA01472D Promass X T m 350 50 180 50 90 120 170 180 180 55 90 120 170 180 180 60 (90) (120) (170) (180) (180) Promass A With thermal insulation according to Endress+Hauser specifications For information on the thermal insulation of the device, see the "Thermal insulation" section of the Operating Instructions T m 1 4 50 205 50 95 130 150 205 205 55 (95) (130) (150) (205) (205) Promass E T m 8 50 50 150 50 100 130 130 150 150 55 (100) (130) (130) (150) (150) 80 45 150 50 75 110 150 150 150 50 75 110 150 150 150 55 (75) (110) (150) (150) (150) Promass F 1) T m 08 15 50 150 50 95 130 150 150 150 60 95 110 (150) (150) (150) 50 240 50 95 130 160 240 240 Endress+Hauser 15

XA01472D Proline Promass 300 1) T m 55 (95) (130) (160) (240) (240) 60 95 110 110 110 110 25 50 50 150 50 95 130 150 150 150 60 95 110 (150) (150) (150) 50 240 50 95 130 160 240 240 55 (95) (130) (160) (240) (240) 60 95 110 110 110 110 50 350 45 85 115 175 275 350 60 85 115 175 275 350 80 50 350 45 85 115 175 275 350 60 85 115 175 275 350 80 250 50 150 50 75 110 150 150 150 60 75 110 (150) (150) (150) 50 240 50 75 110 170 240 240 55 (75) (110) (170) (240) (240) 60 75 110 110 110 110 1) Maximum temperature range see nameplate Promass H T m 8 50 150 1) 50 65 100 150 150 150 55 (65) (100) (150) (150) (150) 60 65 100 100 100 100 8 50 205 2) 50 65 100 160 205 205 55 (65) (100) (160) (205) (205) 60 65 100 100 100 100 15 50 50 150 1) 50 75 115 150 150 150 55 (75) (115) (150) (150) (150) 60 75 115 115 115 115 16 Endress+Hauser

Proline Promass 300 XA01472D T m 15 50 50 205 2) 50 75 115 180 205 205 55 (75) (115) (180) (205) (205) 60 75 115 115 115 115 1) Measuring tube material tantalum 2) Measuring tube material zirconium Promass I T m 8, 15, 15FB, 25 25FB, 40, 40FB, 50 50 150 50 95 130 150 150 150 60 95 120 (150) (150) (150) 50 150 50 85 120 150 150 150 60 85 120 (150) (150) (150) 50FB, 80 50 150 50 85 120 150 150 150 FB = Full bore 60 85 120 (150) (150) (150) Promass O T m 80 250 50 205 50 75 110 170 205 205 55 (75) (110) (170) (205) (205) Promass P 1) T m 8 45 150 45 65 100 150 150 150 50 65 100 150 150 150 60 65 100 125 (150) (150) 45 205 45 65 100 160 205 205 Endress+Hauser 17

XA01472D Proline Promass 300 1) T m 50 65 100 160 205 205 60 65 100 115 (205) (205) 15 50 50 150 50 75 115 150 150 150 60 75 115 125 (150) (150) 50 205 50 75 115 180 205 205 60 75 115 (150) (150) (150) 1) Maximum temperature range see nameplate Promass Q 1) T m 25 100 50 205 50 75 110 160 205 205 55 (75) (110) (160) (205) (205) 25 100 50 150 2) 50 75 110 150 150 150 55 (75) (110) (150) (150) (150) 1) Maximum temperature range see nameplate 2) Cryogenic temperature version: T m = -196 150 C Promass S T m 8 45 150 45 65 100 150 150 150 50 65 100 150 150 150 60 65 100 125 (150) (150) 15 50 50 150 50 75 115 150 150 150 60 75 115 125 (150) (150) 18 Endress+Hauser

Proline Promass 300 XA01472D Promass X T m 350 50 180 50 90 120 170 180 180 55 (90) (120) (170) (180) (180) With thermal insulation without Endress+Hauser specifications 1 A B 1 Position of the reference point for temperature measurement A Standard version B Extended temperature version, cryogenic temperature version, hightemperature version 1 Reference point A0031198 T m 59 72 75 76 77 77 Endress+Hauser 19

XA01472D Proline Promass 300 Connection data: Signal circuits The following tables contain specifications which are dependent on the transmitter type and its input and output assignment. Compare the following specifications with those on the nameplate of the transmitter. Terminal assignment Transmitter: supply voltage, input/outputs HART Supply voltage Input/output 1 Input/output 2 Input/output 3 1 (+) 2 ( ) 26 (+) 27 ( ) 24 (+) 25 ( ) 22 (+) 23 ( ) Device-specific terminal assignment: adhesive label in terminal cover. FOUNDATION Fieldbus Supply voltage Input/output 1 Input/output 2 Input/output 3 1 (+) 2 ( ) 26 (A) 27 (B) 24 (+) 25 ( ) 22 (+) 23 ( ) Device-specific terminal assignment: adhesive label in terminal cover. PROFIBUS PA Supply voltage Input/output 1 Input/output 2 Input/output 3 1 (+) 2 ( ) 26 (B) 27 (A) 24 (+) 25 ( ) 22 (+) 23 ( ) Device-specific terminal assignment: adhesive label in terminal cover. Modbus RS485 Supply voltage Input/output 1 Input/output 2 Input/output 3 1 (+) 2 ( ) 26 (B) 27 (A) 24 (+) 25 ( ) 22 (+) 23 ( ) Device-specific terminal assignment: adhesive label in terminal cover. 20 Endress+Hauser

Proline Promass 300 XA01472D Remote display and operating module DKX001 1 Vcc Gnd 81 82 A 83 B 84 2 81 82 83 84 3 Vcc Gnd A B A0027518 1 Remote display and operating module DKX001 2 Connecting cable 3 Measuring device Safety-related values Order code for "Output; input 1" Option BA Output type Current output 4 to 20 ma HART U N = 30 V U M = 250 V Safety-related values "Output; input 1" 26 (+) 27 ( ) Option GA PROFIBUS PA U N = 32 V U M = 250 V Option MA Modbus RS485 U N = 30 V U M = 250 V Option SA FOUNDATION Fieldbus U N = 32 V U M = 250 V Order code for "Output; input 2"; "Output; input 3" Option B Option D Output type Current output 4 to 20 ma User configurable input/output Safety-related values Output; input 2 Output; input 3 24 (+) 25 ( ) 22 (+) 23 ( ) U N = 30 V U M = 250 V U N = 30 V U M = 250 V Endress+Hauser 21

XA01472D Proline Promass 300 Order code for "Output; input 2"; "Output; input 3" Option E Output type Pulse/frequency/ switch output Safety-related values Output; input 2 Output; input 3 24 (+) 25 ( ) 22 (+) 23 ( ) U N = 30 V U M = 250 V Option F Double pulse output U N = 30 V U M = 250 V Option H Relay output U N = 30 V I N = 100 ma DC/500 ma AC U M = 250 V Option I Current input 4 to 20 ma U N = 30 V U M = 250 V Option J Status input U N = 30 V U M = 250 V Intrinsically safe values Order code for "Output; input 1" Output type Intrinsically safe values "Output; input 1" 26 (+) 27 ( ) Option CA Current output 4-20 ma HART Ex i U i = 30 V l i = 100 ma P i = 1.25 W L i = 0 C i = 0 Option HA PROFIBUS PA Ex i Ex ia U i = 30 V l i = 570 ma P i = 8.5 W L i = 10 µh C i = 5 nf Option TA FOUNDATION Fieldbus Ex i Ex ia U i = 30 V l i = 570 ma P i = 8.5 W L i = 10 µh C i = 5 nf 22 Endress+Hauser

Proline Promass 300 XA01472D Order code for "Output; input 2"; "Output; input 3" Output type Intrinsically safe values Output; input 2 Output; input 3 24 (+) 25 ( ) 22 (+) 23 ( ) Option C Current output 4-20 ma Ex i U i = 30 V l i = 100 ma P i = 1.25 W L i = 0 C i = 0 Option G Pulse/frequency/ switch output Ex i U i = 30 V l i = 100 ma P i = 1.25 W L i = 0 C i = 0 Remote display DKX001 Basic specification, position 1, 2 Approval Terminal assignment Basic specification, position 8 Display; Operation Option O Option C2, C4 81, 82, 83, 84 A connecting cable with the value L/R 24 µh/ω and C cable 1 000 nf must be used for the version for connecting to the remote display DKX001 or ODKX001. Endress+Hauser 23

www.addresses.endress.com