Buck BioSlide Model B1020 Instruction Manual

Similar documents
Buck BioAire Model B520 Instruction Manual

Models DP10 & DP20 Series Low Voltage Disconnects User s Manual Rev. 1.1 October 31, 2007

Indoor Air Quality Meter

IAQ-CALC INDOOR AIR QUALITY METER MODEL 7515

Bio-Pump Plus. Operating Instructions. Product # ZBP-200. Description. Specifications

Model NTX7 Series Automatic Battery Charger User s Manual Rev. 1.0 October 17, 2006

Attitude Instructions Ford 6.0 Powerstroke

IAQ-CALC TM Indoor Air Quality Meter

IAQ-CALC TM Indoor Air Quality Meter

Ventilation Testing. Micromanometer. Alnor Model AXD610 AIRFLOW TM Model PVM610. Operation and Service Manual

RCL-7A. User s Guide. Battery Charging Station. Technology in Depth

PLEASE ENSURE THE JUMP STARTER IS FULLY CHARGED PRIOR TO FIRST USE.

Rotating Vane Anemometer

AIR VELOCITY METER ALNOR MODEL AVM410 AIRFLOW INSTRUMENTS MODEL TA410

CRD610 Automatic Fitting Inserter

MICROMANOMETER ALNOR MODEL AXD610 AIRFLOW INSTRUMENTS MODEL PVM610

UniTorq UTQ Quarter-Turn Actuator Installation and Operation Manual

BC-9000 OPERATIONS MANUAL BATTERY CHARGER COFKO ELECTRONICS LLC COPYRIGHT 2014 P/N

OWNER S MANUAL. Model 9880 Intrinsically Safe Air Velocity Meter

Owner s Guide. ca5354

Rotating Vane Anemometer

Installation Power Management Unit Battery Cables and Battery Harness

Owner s Guide CA 4555

ECLIPSE Laundry Dispenser Controller

Dear Customer, User Memo: Please visit us on facebook or twitter! Thank you for purchasing this product.

SB SWITCH CONTROL BOX

SETRA Operating Instructions Series DPT 230 Pressure Transducers

CTFRP Series Power Supplies

MAINFRAME HOT RUNNER TEMPERATURE CONTROL SYSTEMS. Instruction Manual

Flo-Way. Measure Granular or Powdered Material. Stand Alone or Connect to Other Devices for Blending. Used by Successful Producers World-Wide

18VDC ESB6 Series Cordless Screwdrivers Operation Manual

SPC Series. Digital Scale. Operation Manual

Heliporter. Owner s Manual. 150 Series. by Paravion Technology, Inc.

EAGL 1-Touch Laser Level

Type Flexair Pneumatic Pressure Control Valves

VSS-1 VSS-5 VSS-12 VALIDATED SAMPLING SYSTEM INSTRUCTION MANUAL

V-DETECT UTILITY GRADE VOLTAGE DETECTOR

Quick Start Guide. AZ Mount Pro Level and Go Altazimuth Mount Models: #8900, #8920

Model HPX60 Series Automatic Battery Charger User s Manual Rev. 1.0 October 17, 2006

TABLE OF CONTENTS 10.0 LIMITED WARRANTY

Owner s Guide CARS & CA4B5

Broadband PowerShield. CP27U Models. User Manual

Electronic lube meter

7.3L POWERSTROKE BANJO BOLT KIT Fits L Powerstroke Diesel. Installation Guide

Thermaltake Warranty / Support Information

Medi-Flow. Model M101 & M101D Medicator Instructions GETTING TECHNICAL ASSISTANCE

HSIV PLOW OPERATOR S MANUAL

VANGUARD A EZ INSTALL OPERATOR S MANUAL

i n s t r u c t i o n m a n u a l

BRAVO Inverter/Battery Charger. Table of Contents

IV. PROOF OF PURCHASE: A warranty claim must be accompanied by proof of the date of purchase.

VELOCICALC Air Velocity Meter

User's Guide. EzFlex Thermo-Anemometer. Model EZ30 UNITS MAX MIN AVG VEL

Owner s Guide. ca6154

Owner s Guide. ca5154

DUSTTRAK AEROSOL MONITOR SOLAR POWER KIT MODEL

INSTALL GUIDE Dodge/RAM 5.7L HEMI

DP-CALC TM Micromanometer

AXD 530 MicroManometer

NEO-DYN MODELS 100T AND 100TC ADJUSTABLE TEMPERATURE SWITCHES; ENCLOSURE 3X FOR WEATHERPROOF APPLICATIONS INSTALLATION AND OPERATION MANUAL

INSTRUCTION MANUAL SURFACE RESISTIVITY METER SRM-110

Outdoor Nema Enclosure For Dotworkz CoolDome Power Supply Product Code: NM-CDPS. User Manual

CAUTION 7.0 WARRANTY CONTENTS. 1.0 Overview... 2

CRD600 Automatic Fitting Inserter

AUTO-BLiP. User Manual Lotus INTELLIGENT DOWNSHIFTS. Version 1.0

CLEAN ROOM DEVICES, LLC "WHERE TUBING AND FITTINGS COME TOGETHER"

ST-70 CONTROL OPERATING MANUAL REVISION DATE: PART#:

LC BATTERY CHARGER OPERATION & MAINTENANCE GUIDE

User Guide IGD Series

TABLE OF CONTENTS NOTES. 1.0 Operating principle... 02

COLT CELL BATTERY CHARGER Installation and Operating Instructions

Electrometer Lead Acid Battery Replacement

Owner s Guide APS596Z

Tension your belts at the speed of light!

CLEAN ROOM DEVICES, LLC "WHERE TUBING AND FITTINGS COME TOGETHER"

PVI 60KW, PVI 82KW, PVI 95KW

Rugged MPPT Solar Energy Harvester

Owner s Guide CA 5055

AUTO-BLiP. User Manual Ford Mustang INTELLIGENT DOWNSHIFTS. Version 1.2

CA 5054 Owner s Guide

Installation Guide Smart-UPS On-Line External Battery Pack XBP48RM1U-LI

NEO-DYN MODEL 100P ENCLOSURE 7 ADJUSTABLE EXPLOSION-PROOF PRESSURE SWITCH

CLEAN ROOM DEVICES, LLC "WHERE TUBING AND FITTINGS COME TOGETHER"

Ratio Feeder DB & DD Series Injectors

INSTALL GUIDE Ford F L

BAK1500 INSTALLATION/OWNER'S MANUAL Compact Amplified Subwoofer

! Warning, refer to accompanying documents.

GARNET. SEE Tester II TM INSTRUMENTS LTD. Model 900M Manual PRINTED IN CANADA

ASM Maxi Poles. Instruction-Parts B Series. For the application of architectural paints and coatings.

NOTES TABLE OF CONTENTS

Low Flow Sample Pump 222 Series

AC Irrigation and Propagation Controllers I Four Station, 5006-I and 5006-IP Six Station

Owner s Guide ca4054 ca4554

Battery Charger Retrofit Kit Q-DCCHG1 for DC Cabinet Retrofit Manual

Series 2100 Operating Instructions

JUMP&GO. Portable Jump Start & Power Supply OWNER S MANUAL

Field Sensitive Magnetic Proximity Cycle Switch

This manual is valid for the valve ordering numbers: E01/ E01/ E01/ E01/08

DC Distribution Panel User Guide # S

Superchips Model 2704 MAX MicroTuner GM Trucks with 6.6L Duramax Diesel Engines Vehicle Programming Instructions

Transcription:

A.P. BUCK, INC. Buck BioSlide Model B1020 Instruction Manual 10-20LPM Constant Flow Bioaerosol Sampling Pump For Microscope Slides Page Quick Guide 3 Product Description and Features 4 A.P. BUCK, INC. 7101 Presidents Drive Suite 110 Orlando, FL 32809 USA Tel 1-800-330-2825 407-851-8602 Fax 407-851-8910 apbuck@apbuck.com www.apbuck.com A.P. BUCK, INC. MAN-BIOSLIDE REV-A 12/18/2003 Manual Part No. APB-107036 Additional Features and Specifications 5 Operating Instructions BioSlide Model B1020 6 Timing Routines and Battery Charging 7 Parts & Accessories 8 Service Information and Technical Support 9 Warranty 10

NOTES:

WARRANTY The seller warrants to the Purchaser that any equipment manufactured by it and bearing its name plate to be free from defects in material or workmanship, under proper and normal use and service, as follows: if, at any time within 1 (one) year from the date of sale, the Purchaser notifies the Seller that in his opinion, the equipment is defective, and returns the equipment to the Seller's originating factory prepaid, and the Seller's inspection finds the equipment to be defective in material or workmanship, the Seller will promptly correct it by either, at its option, repairing any defective part or material or replacing it free of charge and return shipped lowest cost transportation prepaid (if Purchaser requests premium transportation, Purchaser will be billed for transportation costs). If inspection by the Seller does not disclose any defect in material or workmanship, the Seller's regular charges will apply. This warranty shall be effective only if installation and maintenance is in accordance with our instructions and written notice of a defect is given to the Seller within such period. This warranty is exclusive and is in lieu of any other warranties, written, oral or implied; specifically without limitation, there is no warranty of merchantability or fitness for any purpose. The liability of the Seller shall be limited to the repair or the replacement of materials or parts as above set forth. BioSlide Quick Guide BUCK BioSlide Bioaerosol Air Sampling Pump A.P. BUCK, INC. Power ON (for Continuous Mode press key for 4 seconds) Power OFF Press key for 1 sec. ON OFF 1 2 Press Pre-Set Sample time After power ON 5 10 A.P. Buck, Inc. CAL POWER CALIBRATE CONTINUOUS COMPLETE Calibration Mode press key for 2 seconds use arrow keys to adjust flow, press Cal key to set flow LOW BATTERY Status LED s confirm all pump functions LIMITATION OF LIABILITY The seller shall not be liable for any claim for consequential loss or damage arising or alleged to have risen from any delay in delivery, malfunction or failure of the equipment. The Seller's liability for any other loss or damage arising out of or connected with the manufacture, sale or use of the equipment sold, including damage due to negligence, shall not in any event exceed the price of the equipment supplied by us. A.P. Buck, Inc. reserves the right to make changes at any time, without notice, in prices, colors, materials, specifications and models, and to discontinue models. Seller makes no warranty for merchantability of purpose. Copyright 2003 A.P. Buck, Inc. This operating manual and the data enclosed herein are not to be reproduced or used, in whole or in part, by anyone without written permission of A.P. Buck, Inc. Buck BioSlide is shown with the APB-706220 Calibration Rotameter. Adjust flowrate on key pad under CAL mode up or down until rotameter ball is centered on 15LPM mark. All sample flow calibrations are performed with a Buck Gel-Impaction slide installed. Page 10 Page 3

Product Description The Buck BioSlide Model B1020 is a bioaerosol sampling pump designed to draw air into a chamber containing a specially prepared 1 X 3 inch microscope slide (known as the BUCK Gel- Impaction Slide). The built-in impaction slit in the top door of the sampler is designed to mimic the collection technique of the spore traps such a the Buck BioAire Cassettes. The Gel-Impaction Slide contains the collection adhesive. A microprocessor controls the BioSlide Model B1020 s flow and provides preset timing routines for sample collection. The pump is capable of flow rates from 10 to 20 Liters Per Minute (LPM). The flows are adjustable from the keypad in (CAL) Calibration Mode. The calibrated flow rate will be maintained at a constant flow throughout the collection period. Four timing routines are selectable from the keypad and continuous flow is achieved by pressing the ON key for 4 seconds. The OFF key manually stops the pump from sampling. The pump uses nickel cadmium (NiCad) rechargeable batteries and will operate up to five hours on a fully charged pack. The standard battery charger supplied with the pump charges the batteries in 16 hours. An optional FastOne (APB-601900) One Hour charger is available. A/C powered operation can be conducted with the supplied A/C Charger. Product Features 10-20LPM Adjustable Constant Flow Range 5 hours continuous operation at 15LPM with Buck Gel-Impaction Slide or a standard microscope slide 1 X 3 coated with stopcock grease Small and lightweight Battery powered, rechargeable with standard charger or optional FastOne one-hour charger Tripod Mount on bottom of unit Convenient timers for 1, 2, 5, or 10 minute samples End-of-Sample audible and visual indicators Easy keypad calibration with supplied Calibration Rotameter and sampling media in-line for greatest accuracy Page 4 Service Information You must obtain an RMA number prior to returning any product. Obtain your RMA number by calling A.P. Buck, Inc. Customer Service at 1-800-330-BUCK or 407-851-8602. To expedite service and repairs, have your Customer ID handy and visit our website, www.apbuck.com, and request an RMA# online. Please ensure that all products returned to A.P. Buck, Inc. contain no hazardous materials. Any obviously contaminated product received will be returned to the customer. All products scheduled for service must be received within 30 days of the RMA number issuance date. Unauthorized products will be returned to the customer. For all work not covered under warranty, A.P. Buck, Inc. will repair any instrument for the cost of parts and labor as quoted. If major components must be replaced, A.P. Buck, Inc. will notify the customer before proceeding with repairs. When returning any instrument for service, please include a Purchase Order marked: Repair Cost Not To exceed $250.00 Without Customer Authorization. Please provide the following information with your instrument: Company Name: Address: Telephone: Fax: Contact Name: Serial Number(s): Date of Purchase: Service Required or Description of Problem: TECHNICAL SUPPORT SERVICES Technical Assistance: Phone (407) 851-8602 Fax: (407) 851-8910 Email: apbuck @apbuck.com Web site: www.apbuck.com Hours: Monday-Friday 8:00AM to 4:30 PM (EST) Page 9

Parts and Accessories BUCK BioSlide Pump Model B1020, (APB-707000) complete with standard 120VAC adapter/charger, instruction manual, and APB-706220 Calibration Rotameter Assembly BUCK BioSlide Pump Model B1020, (APB-707020) complete with standard 230VAC adapter/charger, instruction manual, and APB-706220 Calibration Rotameter Assembly BUCK FastOne Charger, optional, 90-240VAC (APB-601900) Single Station Automatic One Hour Charger BUCK Standard Charger, supplied, 120VAC (APB-603900) Designed to charge the BUCK BioSlide Pump Battery Pack (max 16 hours charge time) BUCK Standard Charger, supplied, 230VAC (APB-603910) Designed to charge the Buck BioSlide Pump Battery Pack (max 16 hours charge time) BUCK BioSlide Rechargeable Battery Pack 4.8VDC (APB-108104) The pack of four cells can easily be changed by opening the case and lifting the pack off the Velcro strip, unplugging the connector and installing the new pack Calibration Rotameter Assembly (APB-706220) fits on top of BioSlide pump flowrate setting and pump at calibration. Tripod for BioSlide support (ALO-9804-08) supports BioSlide for breathing zone and other elevated samples Buck Gel-Impaction Slides (SAM-500910) a microscope slide 1 x 3 inches, box of 25 Audible Alert ON/OFF The arrow keys will turn off the audible beep in areas where sound is not appropriate. With the pump power ON, press the Down Arrow Key for 4 (four) seconds or until two beeps acknowledge the disabling of the beeper. Press the Up Arrow Key for 4 (four) seconds or until three beeps signify the audible beeper is ON. Low Battery Warning When sampling pressing the down arrow key will light up the LED s, located on the right side of the unit, based on the battery charge. If all 5 LED light the batteries are 100%. If 4 light is 80%, 3 light =60%, etc. The low battery LED will begin flashing when there is only enough power to collect a 10 minute sample. Sample Timing Lock-Out Once the Red low battery warning LED stops blinking and becomes Solid Red, none of the timing keys will function and a new sampling cannot be started. At this time the pump must be recharged or plugged into the charger to continue sampling. Tripod Mount A female connector located on the bottom of the unit provides easy connection to most standard tripods. Model: Flow Range Accuracy Specifications BioSlide-B1020 10 to 20LPM ± 5% at Set Point Run Time 5 hours @ 15LPM Size 4 H x 6 W x 5.25D (10.2 cm H x 15.2 cm W x 13.3 cm D) Weight 29 oz (826 g) Battery Voltage Compatibility: Recharge Time Marks 4.8VDC (120 or 230VAC standard charger included) Buck Gel-Impaction Slides 16 hours (1 hour with optional FastOne 90-240VAC Charger) Page 8 * BioSlide and FastOne are trademarks of A.P. Buck, Inc. Page 5

Operating Instructions Model B1020 Sampling Setup 1. Fully charge the pump to be assured of the maximum sampling time. 2. Place a Buck Gel-Impaction Slide carefully into the slide slot on the bottom section of the sampling chamber. Close the top section. 3. Turn the pump on by pressing the ON key. At this moment the BioSlide pump will actuate all the indicating LED lights for one second and beep twice to acknowledge the turn on. This turn on process provides confirmation to the operator that the indicating lights and beeper are functioning properly and that the unit has adequate battery power to operate. Calibrate and Set Flow Rate 1. Place the Rotameter Calibrator on top of the sampling chamber as shown. 2. The CAL key, when pressed, will turn on the pump motor to approximately 15LPM with the microscope slide installed. 3. Use the Arrow Keys to move the flow rate up or down until the red ball of the Rotameter is centered at the 15LPM mark. The pump requires 10-15 seconds to stabilize after the arrow keys are pressed. 4. Once the flow is selected, press the CAL key to stop the pump and store in memory the flow rate selected. This will remain as the flow rate until the calibration is changed by pressing the CAL key and repeating Steps 1-4. The BioSlide Model B1020 is now ready to collect a sample. For greatest accuracy, calibrate with each use, or at least daily. The Rotameter Calibrator is a Secondary Flow Standard and can be used anytime to verify flow rate. Note: a dedicated slide may be used to set the flow rate or verify the flow rate is correct. Wall Charger APB-603900 Timing Routines Sampling 1. Place the sampling slide into the sampling chamber of the pump. 2. Turn on the pump by pressing the ON Key (the green POWER LED is lighted). 3. Start the sampling by pressing one of the timing keys (1, 2, 5, or 10 minutes). LED will light. The BioSlide Model B1020 will automatically run to the calibrated flow for the selected time. End of Sampling At the conclusion of the sampling, the pump will stop. The COMPLETE LED will light and the timer LED will be flashing to indicate which timing routine was selected. Auto Power Off for Battery Conservation The BioSlide Model B1020 turns off 5 minutes after being turned ON if no keys have been pressed. The pump automatically turns off approximately 30 minutes after any sampling routine if no other key is pressed Battery Charging The standard wall charger supplied will charge the batteries in approximately 16 hours or run pump with dead battery. FastOne Charger APB-601900 (Optional) The FastOne Charger will charge the pump s batteries in approximately one hour. This charger uses a negative delta voltage technique which provides a fully charged pack before going to a trickle mode. LED lights on the charger indicate the status of the charging. Page 6 Page 7