Intercomparison Between Trapil, France and NEL, UK Using Oil Flow Turbine Meters

Similar documents
Traceability overview

Calibration, Fluid Property Effects Reproducibility and Long Term Stability. All in 20 minutes! Dr Gregor Brown Cameron

DFX Series Liquid Ultrasonic Flow Meters

HTM Series Helical Turbine Meters

Legal Metrology Control of a Liquefied Petroleum Gas Transfer Standard and Field Practice

Coriolis Density Error Compensating for Ambient Temperature Effects

Flow Research and Test Center. Creating resource technology for today and tomorrow.

LIQUID MEASUREMENT STATION DESIGN Class No

Preparing the New EuroLoop Liquid Flow Facility for Accreditation. Jos van der Grinten - NMi EuroLoop

Small hydrocarbon flow calibration facilities at NMIJ

Determination of Volume Correction Factors for FAME and FAME / Mineral-diesel blends

CHEMSYSTEMS. Report Abstract. Petrochemical Market Dynamics Feedstocks

Product Loss During Retail Motor Fuel Dispenser Inspection

RUN ACCUM. TOTAL STOP BAT LOW HIGH

Proving Liquid Ultrasonic Flowmeters Summary of Testing Conducted January 2011

Calibration Laboratory Assessment Service CLAS Certificate Number Page 1 of 5

Carl Stuart Ltd. Tallaght Business Park, Whitestown, Dublin 24. Calibration Laboratory Registration number: 270C

Flow Research and Test Center

INSTALLATION AND OPERATION

DFX series. High-end Ultrasonic Flow Meters for Liquid Hydrocarbons Custody Transfer. 32 beams for ultimate performance

Field test for the comparison of LNG static and dynamic mass measurement methods

Sentinel LCT. Panametrics Ultrasonic Flowmeter for Liquid Custody Transfer Measurement. GE Sensing & Inspection Technologies. Benefits.

Article: Sulfur Testing VPS Quality Approach By Dr Sunil Kumar Laboratory Manager Fujairah, UAE

Copyright Statement FPC International, Inc

PETROLEUM MEASUREMENT MANUAL THE PERFORMANCE OF TURBINE METERS IN LOADING GANTRIES - A REVIEW OF PROVING AND TEST DATA

11/12/2017 Erwin H. Doorenspleet

Copyright Statement FPC International, Inc

Demonstration with optical fibres by Smart Fibres Ltd. Task 15

Model 8800C Vortex Installation Effects English Rev. CA

Copyright Statement FPC International, Inc

Wind Tunnel Measurement Of Aerodynamic Characteristics Of A Generic Eurocopter Helicopter

EI HMC Report Spring API Committee on Petroleum Measurement Standards Meeting

DEPLOYMENT STRATEGIES FOR CLEAN AND FUEL EFFICIENT VEHICLES: EFFECTIVENESS OF INFORMATION AND SENSITIZATION IN INFLUENCING PURCHASE BEHAVIOUR

Technical Data Sheet , Rev HB October Rosemount 8800 Vortex Installation Effects

UPDATE OF THE SURVEY OF SULFUR LEVELS IN COMMERCIAL JET FUEL. Final Report. November 2012

PERFORMANCE OF THE NEW PRIMARY TORQUE STANDARD MACHINE OF INMETRO, BRAZIL

Sentinel LCT. Panametrics Ultrasonic Flowmeter for Liquid Custody Transfer Measurement. GE Measurement & Control. Benefits.

FH8500. Metering Solution for Custody Transfer Applications. One Step Ahead in Ultrasonic Flowmetering. Main Applications

This chapter gives details of the design, development, and characterization of the

Twin Screw Compressor Performance and Its Relationship with Rotor Cutter Blade Shape and Manufacturing Cost

SAE J1939 CAN compatible sensors for hydraulic system monitoring of flow, pressure and temperature on pumps, valves and hydrostatic transmissions

Figure 1 Linear Output Hall Effect Transducer (LOHET TM )

Verifying the accuracy of involute gear measuring machines R.C. Frazer and J. Hu Design Unit, Stephenson Building, University ofnewcastle upon Tyne,

Effect of Compressor Inlet Temperature on Cycle Performance for a Supercritical Carbon Dioxide Brayton Cycle

UK Weighing Federation Technical Articles

EXPERIMENT A Parallel Pump Test Rig

CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF TORQUE METROLOGY (Torque Generating Device Calibration)

Calibration of Pipe Provers in the field

Technical Services Report

Global Flow Measurement Calibration Consistency

European Fire Pump Listing

THERMOELECTRIC SAMPLE CONDITIONER SYSTEM (TESC)

STEALTH INTERNATIONAL INC. DESIGN REPORT #1001 IBC ENERGY DISSIPATING VALVE FLOW TESTING OF 12 VALVE

Technical Datasheet. HM R Series. Turbine Flow Meters. with Female Inch Threads

GT-Suite Users Conference

KRAL Volumeter - Turbine Fuel and Water Injection Measurement

Technical Data Sheet , Rev HA October Rosemount 8800 Vortex Installation Effects

BAYLAN AUTOMATED TEST BENCH

Overview about research project Energy handling capability

Copyright Statement FPC International, Inc

WP2 - CFD Private Network Meter Commissioning, Proving and Calibration Tests

FUNDAMENTALS OF ORIFICE METERING Ken Embry FMC Measurement Solutions

HERMAN HOFSTEDE. Sales Director, KROHNE Oil & Gas

OIML Weights & Weight Sets

Application Note Original Instructions Development of Gas Fuel Control Systems for Dry Low NOx (DLN) Aero-Derivative Gas Turbines

MILLTRONICS UNIVERSAL SCALE NIVERSAL SCALE Rev. 1.2

PART XV Metering Systems SECTION 2 GUIDE TO GAS METERING SYSTEMS

Copyright Statement FPC International, Inc

Performance of Batteries in Grid Connected Energy Storage Systems. June 2018

FLUID FLOW. Introduction

CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF FORCE METROLOGY:- CALIBRATION OF UNIAXIAL TESTING MACHINES

Improvements to ramp metering system in England: VISSIM modelling of improvements

Technical Report. 10 th Round Robin Test for Multi-Capillary Pressure Drop Calibration Standards (2014)

Impacts of Short Tube Orifice Flow and Geometrical Parameters on Flow Discharge Coefficient Characteristics

VLM20 In-line Vortex Flowmeter

Cross Flow Heat Exchanger H352

F-2600 SERIES INLINE VORTEX FLOW METER

MULTICHANNEL ULTRASONIC FLOW METER AFLOWT UF

Cross Flow Heat Exchanger H352

Flowmeters. The Satam flowmeter range is used for the custody transfer measurement of liquid hydrocarbons, liquid gas and liquid chemicals.

A PORTABLE SYSTEM FOR THE CALIBRATIONOF TRANSDUCERS AND TORQUE WRENCHES: THE CALIBRATION BELL

GMD 13 New gravimetric measuring system with evaluation on location

Optidew Vision Optical Dew-Point Meter

November. Next release: 19 January Sep/AugOct/Sep Nov/Oct Dec 14 December Janu 19 January 2018 Sep/AugOct/Sep Nov/Oct 2017

mass flowmeter for compressed air

Transmission Error in Screw Compressor Rotors

ViscoSense 3D. Viscosity & Density Measurement Enabling Mass Flow Measurement. Product Bulletin TO BE REALLY 1 SURE

F-1500 SERIES INSERTION TURBINE FLOW METER FEATURES Belcher Road South, Largo, FL USA Tel +1 (727) Fax +1 (727)

Small Oil Free Piston Type Compressor For CO2

DaimlerChrysler Alternative Particulate Measurement page 1/8

Dr. Chris Dudfield. Chief Technology Officer Intelligent Energy. Sponsors

RENOVATION OF NMIJ'S 5 MN HYDRAULIC AMPLIFICATION TYPE FORCE STANDARD MACHINE

Copyright Statement FPC International, Inc

Copyright Statement FPC International, Inc

LIFE CYCLE COSTING FOR BATTERIES IN STANDBY APPLICATIONS

Considerations on Flow Regeneration Circuits and Hydraulic Motors Speed Variation at Constant Flow

Is Low Friction Efficient?

How will electric vehicles transform the copper industry? 14 March 2018

Theoretical and Experimental Investigation of Compression Loads in Twin Screw Compressor

BARTON 7400 Precision Gas Turbine Flow Meters

Transcription:

Intercomparison Between Trapil, France and NEL, UK Using Oil Flow Turbine Meters A Report for NMSPU, DTI 151 Buckingham Palace Road London SW1W 9SS Project No: OI/3 (OSDC40300) Report No: 305/99 Date: 8 September 1999

This report is issued as part of the contract under which the work has been carried out for the client. NOTES 1 This report may be published in full by the client unless it includes information supplied in confidence by NEL or any third party. Such information, if included within the report, shall be separately designated as confidential by NEL. 2a 2b The prior written consent of NEL shall be obtained by the client before publication by him or any extract from, or abridgement of, this report. The prior written consent of NEL shall be obtained by the client before publication: where such publication is made in connection with any public enquiry, legal proceedings or arbitration where such publication is made in, or in connection with, any company prospectus or similar document where the client has notice that NEL is seeking or intends to seek patent or like protection for any intellectual property produced in the course of rendering the services.

Flow Centre National Engineering Laboratory East Kilbride Glasgow G75 0QU Tel: 01355 220222 Fax: 01355 272999 Intercomparison Between Trapil, France and NEL, UK Using Oil Flow Turbine Meters A Report for NMSPU, DTI 151 Buckingham Palace Road London, SW1W 9SS SUMMARY As part of the DTI 1996 to 1999 flow programme for NMSPU, an intercomparison between NEL and TRAPIL in France was considered a priority as a significant number of flowmeters used in the UK sector of the North Sea are verified and calibrated at TRAPIL. The transfer package consisted of two turbine flowmeters calibrated at NEL and TRAPIL Gas oil and Kerosene. NEL also calibrated in a third higher viscosity oil and at different temperatures to estimate the performance of the package. Agreement in kerosene was extremely good, within 0.1 per cent. Using gas oil some discrepancies showed a wider level of agreement of some 0.25 per cent on one meter while retaining agreement to within 0.1 per cent on the other. It is concluded that basic agreement is good but that more work to widen the intercomparison in Europe should be carried out and that a more stable package be investigated. Prepared by: Mr R Bunch Approved by: Mr R Paton Date: 8 September 1999 for Dr F C Kinghorn Director Report No: 305/99 Page 1 of 13

CONTENTS Page SUMMARY... 1 1 INTRODUCTION... 3 2 THE LABORATORIES 2.1 NEL... 3 2.2 Trapil... 3 3 TEST METERS AND CONDITIONS... 4 4 TEST FACILITIES, MEASUREMENTS AND METHOD 4.1 NEL... 5 4.2 Trapil... 6 5 UNCERTAINTY 5.1 NEL... 6 5.2 Trapil... 7 6 TEST RESULTS... 7 6.1 Kerosine Calibrations... 7 6.2 Gas Oil Calibrations... 8 7 ADDITIONAL CALIBRATION OF METERS BY NEL... 10 8 CONCLUSIONS... 11 9 ACKNOWLEDGEMENT... 12 Report No: 305/99 Page 2 of 13

1 INTRODUCTION The project was initiated under the DTI National Measurement System Policy Unit 1996-1999 flow programme. The objective was to provide an inter-comparison between NEL, UK and Trapil, France. NEL provides the UK National Standard for flow measurement while Trapil is one of the prime calibration facilities for oil flow meters in France. Both laboratories are accredited by their respective National authorities and provide two of the main traceability routes for North Sea oil flow meter calibrations. This report gives an overview of the laboratories and test methods and summarises the results. The full list of tables of results and associated figures pertaining to the inter-comparison is given the report for reference but only the salient inter-comparison graphs are included in this summary. All the tables and figures referenced are available in Microsoft Excel format in NEL report number FC 305/99 -- Inter-comparison Work for 1996-99 Flow Programme. This is available from NEL as a CD-ROM. This CD includes this report and also includes the reports of all other inter-comparisons carried out within the 1996-1999 flow programme. Using Kerosene (substitute) and Gas Oil (substitute) were used as the test fluids The objective of the project was to ensure that industry could have confidence in the calibrations of oil flowmeters carried out in UK or in France. NEL and Trapil provide a significant number of the calibrations of fiscal meters used in Europe and in the North Sea. 2 THE LABORATORIES 2.1 NEL The National Engineering Laboratory (NEL) is an industrial research organisation concerned with many areas of mechanical engineering research. Within NEL, Flow Centre is the holder of the UK National Standards for Flow Measurement. Facilities exist for calibration and research involving water, oil, gas and multiphase flow. All the measurement components are fully traceable to Primary National Standards and most are accredited by United Kingdom Accreditation Service (UKAS). Meters are calibrated using a standing-start-and-finish technique against gravimetric standards. A number of weigh tanks are available from 6 tonne down to 150 kg. The 6 tonne tank was used for this exercise. Temperature is controlled to within ± 1 C. Using this capability of temperature control and the availability of a second oil (7 cst at 20 C), an estimate of the meter s viscosity and temperature coefficients can be measured. 2.2 TRAPIL Societe des Transports Petroliers par Pipeline -- (Trapil) has a flow calibration facility based at a junction of one of the pipeline systems in France and has access to oil storage facilities. It is located to the north of Paris. Trapil is accredited by the Bureau National De Metrologie (B.N.M.) in the field of fluid flow measurement - Dynamic volume and mass measurement. Meters are calibrated against one of two pipe prover loops. Trapil are able to offer the calibration of flowmeters across the range 10 2000 m 3 /h (2.8 556 l/s) over a range of viscosities from 0.5 to 900 cst. All calibrations are carried out at ambient temperatures. Trapil are also able to offer calibrations at pressures up to 20 bar. Report No: 305/99 Page 3 of 13

3 TEST METERS AND CONDITIONS A twin turbine meter package was used for the inter-comparison. Halliburton Ltd provided the meters for the purposes of the project. The details of the meters are shown below. Meter_1 Manufacturer: Halliburton Type/model: 4-inch turbine flowmeter Serial No: 4SF6903 Date received: 12 August 1998 Date calibrated: 11 November 1998-20 May 1999 Meter_2 Manufacturer: Halliburton Type/model: 4-inch turbine flowmeter Serial No: 4SBF6922 Date received: 12 August 1998 Date calibrated: 11 November 1998-20 May 1999 Calibration Fluid(s) NEL: Approximate Specification Test Fluid_1 Density at 20ºC Viscosity at 20ºC Viscosity at 5ºC Test Fluid_2 Density at 20ºC Viscosity at 20ºC Viscosity at 5ºC Kerosene Substitute 0.7992 kg/l 2.5 cst 3.4 cst Gas Oil Substitute 0.8261 kg/l 7.2 cst 11.2 cst Calibration Fluid(s) Trapil: Approximate Specification Test Fluid_1 Kerosene Substitute Density at 15ºC 0.8023 kg/l Viscosity at 20ºC 1.64 cst Estimated Viscosity at 10ºC is 2.1 cst Test Fluid_2 Gas Oil Density at 15ºC 0.8531 kg/l Viscosity at 20ºC 5.2 cst Estimated Viscosity at 10ºC is 6.4 cst. Estimates are based on NEL curves for similar products and indicated viscosities for each of fluids at a temperature of 20ºC. Report No: 305/99 Page 4 of 13

Test parameters NEL Calibration Range: 7.8 78 l/s Test temperature: 5ºC and 20ºC Pressure downstream: 2 bar gauge Test parameters Trapil Calibration Range: 7.8 78 l/s Test temperature: 8.2 12ºC Pressure downstream: 2 bar gauge Both meters were 100 mm nominal bore turbine flowmeters. Meter_1 was a standard flange type turbine flowmeter. Meter_2 was a wafer type (EZ-IN) turbine flowmeter that was sandwiched between flanges. Both meters were assembled in a package consisting of 20 diameters of 100 mm straight pipe before the meter and 10 diameters of straight pipe after the meter. A flow straightener of the tube bundle type was installed between the two meters to eliminate the possibility of interference from flow profiles from one meter to another. The package was transported as two parts. Each part of the package consisted of the flowmeter with its associated upstream and downstream pipework. Neither part was split during the intercomparison. The two parts were transported separately and then assembled as shown in Fig. 1 at each laboratory. A schematic diagram of the package is given in Fig. 1. 20 d 10 d 20 d 10 d Flow Turbine Flowmeter Flow Straightener Turbine Flowmeter Fig.1 Schematic diagram of meter package 4 TEST FACILITIES, MEASUREMENTS AND METHOD 4.1 NEL The flowmeter package was installed in the NEL National Standards Oil Flow Measurement Facility, lines A (Kerosene) and B (Gas Oil) as shown in Fig. 2. These test lines are identical in all aspects except for the test fluids used. Both are fed from separate storage tanks via suitable pumps to a 6 tonne weighbridge. The oil from the weighbridge is returned to the original storage tank. Report No: 305/99 Page 5 of 13

1.5 t TANK TEST METER 18 m 10 m 6 t TANK FLOW SUPPLY TANK FIG.2 SCHEMATIC DIAGRAM OF TEST CIRCUIT The calibrations were carried out using a standing start and finish gravimetric test method The K-factor (K) was calculated in the following way: K Total pulses from meter = Volume The flowrate was calculated using the time taken to collect the quantity of fluid passed through the meter. 4.2 Trapil The flowmeter package was installed within the Trapil test facility. The test facility comprises of a unidirectional prover loop connected to two, 10 m test benches where the meter package is installed. The loop is a closed circuit consisting of a variable speed circulating pump, test section and prover. The test line is kept full and at constant static pressure using a compensation reservoir (balancing tank) and pressurization pump. The loop is filled with fluid from one of six available storage tanks. 5 UNCERTAINTY 5.1 NEL Using the test method outlined, the accredited standard uncertainty in the measurement of the quantity of fluid passed through the meter is estimated to be 0.03 per cent with a coverage factor k=2. Report No: 305/99 Page 6 of 13

5.2 Trapil Using the test method outlined, the uncertainty in the measurement of the quantity of fluid passed through the meter (Vm) is dependent upon the fluid and also the flowrate. Details are given below. Kerosene flow > 40m 3 /h : 3.7E-4.Vm Kerosene flow < 40m 3 /h : 3.9E-4.Vm Gas Oil flow > 40m 3 /h : 3.8E-4.Vm Gas Oil flow < 40m 3 /h : 4.1E-4.Vm This can be rounded to 0.05%. The coverage has not been provided. 6 TEST RESULTS All the results of the calibrations and the associated graphs are listed at the end of this report. All the tables of results and figures are available in NEL Report FC 305/99 issued as a CD-ROM. Only the figures pertinent to the conclusions are included here. The figure numbers used reference to those provided in the full data set on the CD-ROM. 6.1 Kerosine Calibrations Fig. 3 below shows the calibration of meter_1 in Kerosene at both NEL and Trapil. Two calibrations are shown for NEL (5 C and 20 C) and a single calibration at ambient temperature of between 8.2 C and 12 C for Trapil. This corresponds to oil viscosities of 3.4 cst and 2.5 cst for NEL and 2.1 cst for Trapil. 7.92 7.90 7.88 K-factor P/l 7.86 7.84 0.20% Trapil NEL at 5 C NEL at 20 C 7.82 7.80 0 10 20 30 40 50 60 70 80 90 Flowrate l/s Fig.3 Calibration of Meter_1 in Kerosene Report No: 305/99 Page 7 of 13

The calibration curves shown by all three calibrations are similar with a linearity of some 0.1 per cent from 80 l/s down to 20 l/s. in this region the Trapil results are some 0.12 per cent higher than the NEL results which agree to within 0.025 per cent. Below 20 l/s the k factor drops rapidly and while the three curves follow each other closely some separation between NEL and Trapil is noted but this may be due to the increased linearity and the sensitivity to viscosity at this flow rate. Fig. 4 below shows the calibration of meter_2 for the same test parameters and was calibrated at the same time in series with meter_1. 7.80 7.78 K-factor P/l 7.76 7.74 0.20% Trapil NEL at 5 C NEL at 20 C 7.72 7.70 0 10 20 30 40 50 60 70 80 90 Flowrate l/s Fig.4 Calibration of Meter_2 in Kerosene The NEL results at 20 C and the Trapil results follow the same calibration curve and agree within 0.1 per cent across the range. Both calibrations show repeatability of around 0.02% The NEL 5C calibration shows a more linear calibration curve without the drop in k-factor at both extremes of the flow range shown by the other two curves. 6.2 Gas Oil Calibrations Fig. 5 below shows the calibration of meter_1 in Gas Oil at both NEL and Trapil. Two calibrations are shown for NEL (5 C and 20 C) and a single calibration at ambient temperature of between 8.2 C and 12 C for Trapil. This corresponds to oil viscosities of 11.2 cst and 7.2 cst for NEL and 6.4 cst for Trapil. This set of results shows the Trapil result following a different curve from that demonstrated for the Kerosene calibration. The curve drops slightly at 20 l/s then rises to again at 10 l/s rather than dropping further as shown by the kerosene tests. The NEL 5C result generally follows the same curve shape but the curve is more pronounced with differences from Trapil of up to 0.1 per cent. Report No: 305/99 Page 8 of 13

The main 20ºC calibration from NEL demonstrates a very different curve and while agreeing closely at 80l/s with the other tests drops rapidly to give a 0.3 per cent difference at 10L/s. Repeatability for all three calibrations is excellent. 7.94 7.92 7.90 K-factor P/l 7.88 7.86 7.84 0.20% Trapil NEL at 5 C NEL at 20 C 7.82 0 10 20 30 40 50 60 70 80 90 Flowrate l/s Fig.5 Calibration of Meter_1 in Gas Oil Fig. 6 below shows the calibration of meter_2 for the same test parameters and was calibrated at the same time in series with meter_1. 8.05 8.00 0.20% Trapil NEL at 5 C NEL at 20 C 7.95 K-factor P/l 7.90 7.85 7.80 7.75 0 10 20 30 40 50 60 70 80 90 Flowrate l/s Fig.6 Calibration of Meter_2 in Gas Oil Report No: 305/99 Page 9 of 13

In this case the NEL 20 C and the Trapil calibration agree very closely, with the NEL 5 C about 0.1 percent higher. All the results show a curve which is linear from 80 l/s down to 15 l/s. Below 15 l/s the k-factor rises steeply. The above results have not had any corrections made for either temperature effects or viscosity. 7 ADDITIONAL CALIBRATION OF METERS BY NEL NEL carried out some additional calibrations upon the meter package to better understand the performance in relation to both temperature and viscosity. This involved calibration of the meters in a third oil (Velocite) at varying temperatures (12 C, 20 C and 45 C) as well as an additional calibration in both original test fluids at a higher temperature (45 C). Details of the additional calibrations and their associated graphs can be found on the CD-ROM. Fig. 29 below shows the calibration of meter_1 over all three-test fluids at all three temperatures. Details of the fluid viscosities are given on the graph to allow an assimilation of the data provided. 8.30 K-factor P/l 8.20 8.10 8.00 Kero 20 C, 2.5cst Kero 5 C, 3.4 cst Gas Oil 20 C, 7.2cst Gas Oil 5 C, 11.2 cst Kero 45 C, 1.6 cst Gas Oil 45 C, 3.5 cst Velocite 20 C, 19 cst Velocite 12 C, 33 cst Velocite 45 C, 9 cst 7.90 7.80 7.70 0 10 20 30 40 50 60 70 80 90 Flowrate l/s Fig.29 NEL calibration of Meter_1 The calibration curves show the effects of both viscosity and temperature upon meter_1 with the greatest effects being shown for the calibration using the thicker viscosity oil. This is especially true at flowrates below 25 l/s where the calibration curves rises rapidly as the viscosity increases. This trend changes as the oil viscosities thin and from the graph it can be seen that the curves actually drop off as the flowrates decrease. At flowrates above 25 l/s the general linearity of the meter is within the 0.15% as would be expected for each calibration. The overall agreement between each separate calibration is 0.6. Fig. 30 below shows the calibration of meter_2 for the same test parameters. Report No: 305/99 Page 10 of 13

8.20 K-factor P/l 8.10 8.00 7.90 Kero 20 C, 2.5 cst Kero 5 C, 3.4 cst Gas Oil 20 C, 7.2 cst Gas Oil 5 C, 11.2 cst Kero 45 C, 1.6cSt Gas Oil 45 C, 3.5 cst Velocite 20 C, 19 cst Velocite 12 C, 33 cst Velocite 45 C, 9 cst 7.80 7.70 7.60 0 10 20 30 40 50 60 70 80 90 Flowrate l/s Fig. 30 NEL Calibration of Meter_2 The calibration curves show the effects of both viscosity and temperature upon meter_2 with the greatest effects being shown for the calibration using the thicker viscosity oil. This is especially true at flowrates below 20 l/s where the calibration curves rises rapidly as the viscosity increases. This trend changes as the oil viscosities thin and from the graph it can be seen that the curves actually drop off as the flowrates decrease. At flowrates above 20 l/s the general linearity of the meter is within the 0.15% as would be expected for each calibration. The overall agreement between each separate calibration is 0.6%. 8 CONCLUSIONS 1) The agreement between Trapil and NEL for meter_1 in Kerosene was found to be within 0.1%, with the Gas Oil results within 0.25%. 2) For meter_1 the agreement using gas oil was within 0.25 per cent but this is mostly due to the 20C 7.6 cst result which does not fit well with the other data. 3) The agreement between Trapil and NEL for meter_2 in both Kerosene and Gas Oil was found to be better than 0.1%. 4) Repeatability for both meters was better than ±0.02% in both labs. 5) Variations in viscosity and temperature on the meters leads to variations in K-factor of up to 0.6% over the range 1.6 to 33 cst, and 5 to 45 C. Some general trends could be discerned from the data, but it was not considered practical to derive and apply corrections to the results of the intercomparison. 6) With the level of agreement showing agreement around 0.1% future intercomparisons should concentrate on widening the covering of commercial laboratories in France (Fos-Sur-Mer and Faure-Harman are examples) and other similar laboratories in Europe. Report No: 305/99 Page 11 of 13

9 ACKNOWLEDGEMENT NEL would like to thank Halliburton for their assistance with the supply of two turbine flowmeters. Appreciation is also shown to the staff at Trapil for their co-operation in this project and their welcome during the visit to the facility. List of Tables and Figures provided in NEL Report No 305/99; entitled Inter-comparison Work for 1996-99 Flow Programme. Table 1 Calibration of Meter package at NEL in Kerosene at 20 C. Table 2 Calibration of Meter package at NEL in Kerosene at 5 C. Table 3 Calibration of Meter package at NEL in Gas Oil at 20 C. Table 4 Calibration of Meter package at NEL in Gas Oil at 5 C. Table 5 Calibration of Meter package at Trapil in Kerosene at ambient temp. Table 6 Calibration of Meter package at Trapil in Gas Oil at ambient temp. Table 7 Calibration of Meter package at NEL in Kerosene at 45 C. Table 8 Calibration of Meter package at NEL in Gas Oil at 45 C. Table 9 Calibration of Meter package at NEL in Velocite at 20 C. Table 10 Calibration of Meter package at NEL in Velocite at 12 C. Table 11 Calibration of Meter package at NEL in Velocite at 45 C. Fig. 1 Schematic diagram of meter package. Fig. 2 Schematic diagram of NEL test facility. Fig. 3 Calibration of Meter_1 in Kerosene at both facilities. Fig. 4 Calibration of Meter_2 in Kerosene at both facilities. Fig. 5 Calibration of Meter_1 in Gas Oil at both facilities. Fig. 6 Calibration of Meter_2 in Gas Oil at both facilities. Fig. 7 NEL Calibration of Meter_1 in Kerosene at 20 C. Fig. 8 NEL Calibration of Meter_2 in Kerosene at 20 C. Fig. 9 NEL Calibration of Meter_1 in Kerosene at 5 C. Fig. 10 NEL Calibration of Meter_2 in Kerosene at 5 C. Fig. 11 NEL Calibration of Meter_1 in Gas Oil at 20 C. Fig. 12 NEL Calibration of Meter_2 in Gas Oil at 20 C. Fig. 13 NEL Calibration of Meter_1 in Gas Oil at 5 C. Fig. 14 NEL Calibration of Meter_2 in Gas Oil at 5 C. Fig. 15 Trapil Calibration of Meter_1 in Kerosene at ambient temp. Fig. 16 Trapil Calibration of Meter_2 in Kerosene at ambient temp. Fig. 17 Trapil Calibration of Meter_1 in Gas Oil at ambient temp. Fig. 18 Trapil Calibration of Meter_2 in Gas Oil at ambient temp. Fig. 19 NEL Calibration of Meter_1 in Kerosene at 45 C. Fig. 20 NEL Calibration of Meter_2 in Kerosene at 45 C. Fig. 21 NEL Calibration of Meter_1 in Gas Oil at 45 C. Fig. 22 NEL Calibration of Meter_2 in Gas Oil at 45 C. Fig. 23 NEL Calibration of Meter_1 in Velocite at 20 C. Fig. 24 NEL Calibration of Meter_2 in Velocite at 20 C. Fig. 25 NEL Calibration of Meter_1 in Velocite at 12 C. Report No: 305/99 Page 12 of 13

Fig. 26 NEL Calibration of Meter_2 in Velocite at 12 C. Fig. 27 NEL Calibration of Meter_1 in Velocite at 45 C. Fig. 28 NEL Calibration of Meter_2 in Velocite at 45 C. Fig. 29 NEL Calibration of Meter_1. Fig. 30 NEL Calibration of Meter_2. Report No: 305/99 Page 13 of 13