Instruction Manual ICM A1 AC/DC Clamp Meter ICM A1

Similar documents
Instruction Manual. IM-191 Magnetic Field Meter

Instruction Manual IVT-10 / IVT-20 Voltage Tester

INSTRUCTION MANUAL IDM67 DIGITAL MULTIMETER

ICA 39T INSTRUCTION MANUAL ICA39T AC/DC CURRENT TRANSDUCER

INSTRUCTION MANUAL IDM97II DIGITAL MULTIMETER EN FR IT DE

INSTRUCTION MANUAL IDM93N DIGITAL MULTIMETER EN FR IT DE

365 Detachable Jaw True-rms Clamp Meter

INSTRUCTION MANUAL IDM91E DIGITAL MULTIMETER EN FR IT DE JP

362 Clamp Meter. Users Manual

362 Clamp Meter. Users Manual

Instruction Manual. IT-1 Infrared Thermometer

323/324/325 Clamp Meter

373 Clamp Meter. Users Manual

302+/303 Clamp Meter. Users Manual

RS-3 PRO RS-1007 PRO. CAT IV Analog Clamp meter Series. Users Manual. For detailed specifications and ordering info go to

LH41A. User Manual. Clamp On Ammeter. For detailed specifications and ordering info go to

Users Manual. Phase Rotation Indicator

BT508/BT510/BT520/BT521 Battery Analyzer Safety Sheet

i400s Calibration Information Introduction

INSTRUCTION MANUAL CM69 TRMS AC EARTH LEAKAGE CLAMP METER. 1. SAFETY INFORMATION: Always read before proceeding. REMEMBER: SAFETY IS NO ACCIDENT

DT33B/C/D OPERATOR S MANUAL LIMITED WARRANTY AND LIMITATION OF LIABILITY

Calibration Manual. Detachable Jaw True-rms Clamp Meter

INSTRUCTION MANUAL CM95/ CM100 TRUE RMS FLEXIBLE CLAMP METER. 1. SAFETY INFORMATION: Always read before proceeding. REMEMBER: SAFETY IS NO ACCIDENT

i3000s/i2000 Flex AC Current Probe

The following table shows the symbols used on the product and/or in this manual. Hazardous Voltage. Risk of electric shock.

Go to to register your product and find more information.

USER MANUAL. 200A AC Mini Clamp Meter. Model MA130. Please visit for user manual translations

Visit us at

Calibrator Capabilities Function Range Resolution Measure V dc 28 V 1 mv Measure ma dc Source ma dc

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com INSTRUCTION MANUAL PSMR1

MY /2-DIGIT DIGITAL MULTIMETER Users Manual

TX900. Microwave Leakage Detector. 99 Washington Street Melrose, MA Phone Toll Free

1587 FC/1587/1577. Insulation Multimeter. Safety Information. Warnings

KMP Insulation-Continuity Tester. Users Manual

233 True-rms Remote Display Digital Multimeter Safety Information

1662/1663/1664 FC. Electrical Installation Tester. Safety Information. Warnings

CP GermanCp 41_CP C. Operating Instructions. CP 30 CP 330 CP 1100 AC/DC current probe /7.10

R&S RT-ZC02 R&S RT-ZC03 AC/DC Current Probe User Manual

Dawson DDM190D. Digital Multimeter User s Manual

Users Manual. MegOhmMeter

PRM-4. Phase Sequence and Motor Rotation Tester. Users Manual. For detailed specifications and ordering info go to

1 INTRODUCTION SPECIFICATIONS Electrical Data General Data... 3

LM-100 LM-120. Light Meters. Users Manual. For detailed specifications and ordering info go to

AC/DC CURRENT TRANSDUCER INSTRUCTION MANUAL

Table of Contents. Title Page UT501/UT502 OPERATING MANUAL. Introduction. Unpacking the Meter. Safety Information. International Electrical Symbols

ASTROAI USER MANUAL DIGITAL MULTIMETER. NOTE: Fully read and understand this manual before using this instrument.

Users Manual. Accessories Motor and Phase Rotation Indicator ships with the following items: 3 test leads 3 test probes 3 alligator clips 9 V battery

PY-11 Voltage & Continuity Tester

Instruction Manual. 99 Washington Street Melrose, MA Fax TestEquipmentDepot.com. Model 308A Digital Insulation & Continuity Meter

Low Amp Probe EETA308A

Phase Sequence Tester

51 & 52 Series II. Users Manual. Thermometer. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

Users Manual. Airflow Meter

1000FLT. Users Manual. Fluorescent Light Tester

80 Series V Digital Multimeter Safety Information

MAX. AC 600V HOLD V M H.V. LIVE CIRCUIT DIGITAL INSULATION TESTER DO NOT OPERATE IF LIT POWER DATA HOLD TEST ON / OFF 250V 500V 1000V OFF CAT.

KMP7010. Ground Resistance Tester. Users Manual


MM300. INSTRUCTION MANUAL Manual-Ranging Digital Multimeter

TIC 300 PRO. Users Manual Mode d emploi Bedienungshandbuch Manual d uso Manual de uso. Non Contact AC Voltage Detector. High Energy Tic Tracer

User's Manual. High Voltage Megohmmeter. Analog Insulation Tester plus AC Voltage and Continuity Tests. Model

MW Manuel d utilisation User s Manual. Multimètre / Contrôleur d isolement Insulation tester / Multimeter M9090 M00

INSTRUCTION MANUAL DM5B Fax: (503)

Safety design conforming to the following provisions

4-WIRE EARTH RESISTANCE TESTER C1 ES P1 S. Max. AC 200V MICROPROCESSOR CONTROLLED TEST. Switch ON, TEST, SCROLL RESULTS 20Ω 200Ω 2000Ω OFF

Cable Length Meter INSTRUCTION MANUAL

716 Pressure Calibrator

Temperature Calibrator

DIGITAL INSULATION RESISTANCE TESTER (5KV, 1T )

SL101 Manual_SL101 User Manual 05/10/ :06 Page 1. Instruction Manual

OPERATION MANUAL MILLIAMP PROCESS CLAMPMETER MODEL - KM 071 LIST OF PRODUCTS

INSPECTOR LINE LOAD SIMULATOR INSTRUCTION MANUAL TASCO, INC.

4-20mA DC PROCESS CLAMP METER

SAM990DW DIGITAL PSYCHROMETER GENERAL TOOLS & INSTRUMENTS USER S MANUAL. Please read this manual carefully and thoroughly before using this product.

OPERATING INSTRUCTIONS

THERMOCOUPLE CALIBRATOR Mode KM-CAL-710

Table of Contents Title Page No.

For technical questions and replacement parts, please call

Instruction Manual VT4 VOLTAGE TESTER. Check out what else you can get from Martindale:

User Guide. Model Insulation Tester / Megohmmeter. Introduction

PD-20 HIGH VOLTAGE PROBE INSTRUCTION MANUAL

717 Series. Pressure Calibrators. Instruction Sheet. Introduction

Item ref: UK & UK MTB01 & MTB02 DIGITAL MULTITESTERS. User Manual

AMB V Insulation Resistance Tester. Users Manual. For detailed specifications and ordering info go to

TIME-10 Process Control Stopwatch

INSTRUCTION MANUAL DT173 DT175 DIGITAL THERMOMETERS ALWAYS READ THESE INSTRUCTIONS BEFORE PROCEEDING

719 Series. Pressure Calibrator with Electric Pump. Product Overview (E)

User's Manual. Earth Ground Resistance Tester. Model Introduction

OPERATING INSTRUCTIONS

Nature Power Inverters. True Sinewave Inverter Modified Sinewave Inverter. Owner s Manual

Tachometer. English. Users Manual

Instruction Manual PC104/PC105 3-PHASE SOCKET TESTER. Check out what else you can get from Martindale:

AC/DC CURRENT TRANSDUCER 600A 10mV/1A, 1mV/1A

SLA Battery Capacity Analyzer

DIGITAL RCD(ELCB) TESTER

MOdel No: MM19.V3 WARNING CAUTION 1. SAFETY INSTRUCTIONS

Dawson DAN100. Analog Multimeter User s Manual

True RMS Autoranging Multimeter

INSP-3 Wiring Inspection Tester

Autoranging Industrial Multimeter

Transcription:

Instruction Manual ICM A1 AC/DC Clamp Meter EN ICM A1

Warning...................................... Safety sheet.............................................. Read First Safety Information - Use the instrument only as specified in this manual or the protection by the instrument might be impaired. - Always use proper terminals, switch position, and range for measurements. - Verify the instrument s operation by measure ing a known voltage. If in doubt, have the instrument serviced. - Do not apply more than the rated voltage, as marked on the instrument. - To avoid false readings that can lead to electric shock and injury, replace battery as soon as low battery indicator. - Do not use the instrument around explosive gas or vapor. - To reduce the risk of fire or electric shock do not expose the instrument to rain or moisture. - Individual protective equipment should be used if HAZARDOUS LIVE parts in the installation where measurement is to be carried out could be ACCESSIBLE. - Do not allow fingers to protrude beyond the Tactile Barrier when fitting or removing the instrument from around a Hazardous Live conductor, as this may cause a shock. 1

Caution Do not expose the instrument to extremes in temperature or high humidity. Symbols as marked on the instrument and instruction manual Risk of electric shock See instruction card DC measurement AC measurement Equipment protected by double or reinforced insulation < Battery Earth Conforms to EU directives E Application around and removal from hazardous live conductors is permitted Do not dispose this electrical or electronic product in domestic household waste. 2

Power on / off Auto Power Off 3

Auto Power Off Disable Hold down and press ACA & DCA CAT.Ⅲ600V with respect to earth for the jaw Barrier for Hand Guard The barrier on the JAW is indicating the limit of safe access of the handheld part, do not hold over the barrier when in normal use. 4

Auto ACA&DCA Detection / ACA&DCA Due to the high sensitivity of the instrument, perform DCA Zero in the same direction as measurement to avoid interference by an external magnetic field. 5

Data Hold The display will flash continuously if the measured signal is larger by 50counts than the display reading. However, it cannot detect across the AC and DC Current. 6

DCA ZERO Inrush Current The instrument calculates the RMS value for 100ms as detecting a more than 5A current. 7

Low Pass Filter The cut -off frequency of the low pass filter is about 160Hz with attenuation characteristic of approx-24db/octave. Backlight on/off 8

Voltsense The number of dashes displaying on the LCD indicates the electric field intensity. If no indication, voltage could still be present. 9

Battery Replacement Maintenance Do not attempt to repair this instrument. It contains no user-serviceable parts. Repair or servicing should only be performed by qualified personnel. Cleaning Periodically wipe the case with a dry cloth and detergent. Do not use abrasives or solvents. 10

Specifications 1-1 General Specifications Display Count : 6000 count Overrange Display : OL or -OL Conversion Rate : 2 times / second Dimensions (W x H x D) : 60 x 147 x 31.5 mm Weight : 140 g Power requirement : LR44 Size Button Battery 1.5V x 2 Battery Life : 20 hours. Maximum Conductor Size : 22 diameter LVD : EN61010-1, EN61010-2-030, EN61010-2-032 EMC : EN61326-1 Installation category : CAT.Ⅲ. 600V. CAT. Application field Ⅰ The circuits is not connected to mains. Ⅱ The circuits is directly connected to Low-voltage installation. Ⅲ The building installation. Ⅳ The source of the Low-voltage installation. 11

1-2 Environmental Conditions Indoor Use. Maximum operating altitude : 2000m (6562 ft) Operating temperature : 0 C ~ 30 C, 80%RH 30 C ~ 40 C, 75%RH 40 C ~ 50 C, 45%RH Storage temperature : -20 to +60 C, 0 to 80% RH (no batteries). Temperature coefficient : 0.2 x (Specified accuracy) / C, < 18 C, > 28 C Pollution Degree : 2 Shock vibration : MIL-PRF-28800F for A class 2 Instrument Drop Protection : 4 Feet Drop to hardwood on concrete Floor 12

1-3 Electrical Specifications Accuracy is given as ±(% of reading + counts of least significant digit) at 23 C ± 5 C,with relative humidity Less than 80% R.H. ACV and ACA specifications are ac coupled, true R.M.S. The crest factor may be up to 3.0 as 4000 counts. For non-sinusoidal waveforms, Additional Accuracy by Crest Factor (C.F.) : Add 3.0% for C.F. 1.0 ~ 2.0. Add 5.0% for C.F. 2.0 ~ 2.5. Add 7.0% for C.F. 2.5 ~ 3.0. Position Error of Clamp: ±1.5% of LCD reading. DC Current Range Resolution Accuracy 60.00A (1) 0.01A ±(1.5% +10D) (2) 300.0A 0.1A ±(1.5% + 5D) All specifications are valid from 5% to 100% of each range. (1) There is less than 0.3A variation as measuring in different directions. (2) Add 10D to accuracy in Auto AC & DC Sense Mode. 13

AC Current Range Resolution 60.00A 0.01A 300.0A 0.1A Accuracy (50 ~ 100Hz) Accuracy (100 ~ 400Hz) ±(1.5% + 5D) ±(2.5% +5D) All specifications are valid from 5% to 100% of each range. Frequency Response : 50 ~ 400Hz (Sine Wave) Low-pass Filter Range Resolution Accuracy (50Hz/60Hz) 60.00A 0.01A ±(3.5%+5D) 300.0A 0.1A ±(3.5%+5D) All specifications are valid from 5% to 100% of each range. Cut-off Frequency (-3dB) : Approx. 160Hz Attenuation Characteristic : Approx. -24dB / Oct 14

Inrush Current Range Resolution 300.0A 0.1A Integration Time : 100ms Trigger Current : 5.0A VoltSense Voltage Range : 80V ~ 600V (At the tip of clamp) 15

Limited Warranty This meter is warranted to the original purchaser against defects in material and workmanship for 3 years from the date of purchase. During this warranty period, RS Components will, at its option, replace or repair the defective unit, subject to verification of the defect or malfunction. This warranty does not cover fuses, disposable batteries, or damage from abuse, neglect, accident, unauthorized repair, alteration, contamination, or abnormal conditions of operation or handling. Any implied warranties arising out of the sale of this product, including but not limited to implied warranties of merchantability and fitness for a particular purpose, are limited to the above. RS Components shall not be liable for loss of use of the instrument or Other incidental or consequential damages, expenses, or economic loss, or for any claim or claims for such damage, expense or economic loss. Some states or countries laws vary, so the above limitations or exclusions may not apply to you. For full terms and conditions, refer to the RS website. 16

Africa RS Components SA P.O. Box 12182, Vorna Valley, 1686 20 Indianapolis Street, Kyalami Business Park, Kyalami, Midrand, South Africa www.rs-components.com Asia RS Components Pte Ltd. 31 Tech Park Crescent Singapore 638040 www.rs-components.com China RS Components Ltd. Suite 23 A-C, East Sea Business Centre Phase 2, No. 618 Yan'an Eastern Road Shanghai, 200001, China www.rs-components.com Europe RS Components Ltd. PO Box 99, Corby, Northants. NN17 9RS, United Kingdom www.rs-components.com Japan RS Components Ltd. West Tower (12th Floor), Yokohama Business Park, 134 Godocho, Hodogaya, Yokohama, Kanagawa 240-0005, Japan www.rs-components.com U.S.A Allied Electronics 7151 Jack Newell Blvd. S. Fort Worth, Texas 76118, U.S.A. www.alliedelec.com South America RS Componentes Limitada Av. Pdte. Eduardo Frei M. 6001-71 Centro Empresas El Cortijo Conchali, Santiago, Chile www.rs-components.com