Using the PSD for Backflushing on the Agilent 8890 GC System

Similar documents
Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application

High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph

High-Temperature Simulated Distillation System Based on the 6890N GC Application

Application Note. Abstract. Authors. Environmental Analysis

Optimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System. Jim McCurry Roger Firor Agilent Technologies Wilmington, DE

Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application

Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel

Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI)

High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System

Dual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector

Technical Procedure for Gas Chromatography (GC-FID)

Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis

Analysis of Petroleum Fractions by ASTM D2887

Page 1. Alternate Carrier Gas Considerations and Faster GC Analysis

Detection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector

Simple Heart Cutting with Deans. Capillary Flow Technolgy. Simon Jones Applications Engineer

Setting up SilFlow for BackFlush in your GC

Agilent and ASTM. Update on Recent Activities. Page 1

Method Development for Capillary GC Systems. Slide 1

Application. Gas Chromatography June 1995

GC Inlets. An Overview. Simon Jones GC Applications Engineer

Fast GC. Dial for e-seminar Audio. Slide 4 SPEEDY GC

Optimized Method Development of Large Volume Injection for GC/MS/MS of Food Pesticides

Impact of GC Parameters on The Separation Part 2: Choice of Column Internal Diameter

Application Note. Author. Abstract. Energy & Chemicals - Petrochemicals. Edgar Naegele, Agilent Technologies, Inc. Waldbronn, Germany

Agilent Multimode Inlet for Gas Chromatography

Application Note. Authors. Abstract. Energy & Chemicals

Agilent G2855A Deans Switching System

Analysis of Glycerin and Glycerides in Biodiesel (B100) Using ASTM D6584 and EN Application. Author. Abstract. Introduction

ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column

Refinery Support from the R&D Laboratory Perspective using Fast & Micro Gas Chromatography

Proof of Long-Term, Leak-Free Performance for a Novel Self-tightening GC Column Nut

Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers. Versa HT3. Application Note. Abstract.

Refinery Gas. Analysis by Gas Chromatography WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

GC Analysis of Total Fatty Acid Methyl Esters (FAME) and Methyl Linolenate in Biodiesel Using the Revised EN14103:2011 Method

C2, C3, C4 Monomer Analysis

Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011

GC Analysis of PCBs in Fish Oil by Direct Injection Without Any Sample Cleanup. Philip L. Wylie, Ph.D. Sr. Applications Chemist Agilent Technologies

Oxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels

Model 8610C Gas Chromatograph. GC Chassis Types ECD PID FID / DELCD FPD. Mounts up to Six Detectors and Five Injectors

performance productivity reliability

White Paper. Improving Accuracy and Precision in Crude Oil Boiling Point Distribution Analysis. Introduction. Background Information

Troubleshooting Tips & Tricks for your GC Analyzer & CFT Application

GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585

Typical Gas Chromatographic System

GC Method Compliance and Large Valve Oven Application

Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC

2-D GC/MS and the Deans Switch: Theory and Applications. Fred Feyerherm, Agilent Technologies Inc. Houston, Tx

SELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS

GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS

Performing ASTM 6584 free and total glycerin in BioDiesel using an SRI Gas Chromatograph and PeakSimple software

Simulated Distillation Analyzers, Software, Standards, Consumables, Training

Agilent Multimode Inlet

Practical Steps in GC Troubleshooting

Free and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN and ASTM D6584

GC Best Practices & Troubleshooting. Group/Presentation Title Agilent Restricted Month ##, 200X

Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS)

Characterization of Tiki Torch Fuels

Operation and Applications of Differential Flow Modulation

Agilent 7693A Automated Liquid Sampler

Agilent Solutions for the Analysis of Ethanol- Based Fuels Derived From Biomass. James D. McCurry, Ph.D. Senior Scientist

Technical Procedure for Pyrolysis-Gas Chromatography/Mass Spectrometry (Py-GC-MS)

Analysis of Fatty Acid Methyl Esters (FAMES), and Examination of Biodiesel Samples for these Components, by GCxGC-FID

Application Note. Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D using AC OXYTRACER

Agilent 9000 Intuvo GC Applications. Matthew Giardina, Ph.D. James McCurry, Ph.D. Rebecca Veeneman, Ph. D.

Quantification and Characterization of Sulfur in Low-Sulfur Reformulated Gasolines by GC-ICP-MS Application

MET-Biodiesel Capillary GC Columns

Application Note. Author. Introduction. Energy and Fuels

Determination of fuel system icing inhibitor content of aviation turbine kerosine by HPLC

AppNote 6/2006. Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) KEYWORDS

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

Optimized Supercritical Fluid Chromatographic Instrumentation for the Analysis of Petroleum Fractions

Analysis and. Separation of Oxygenates in Hydrocarbon Matrices. Simon Jones Application Engineer Folsom, CA

Techniques for Improving the Reproducibility of GC Analysis

PRODUCT SPECIFICATION. Product Specification Crystal 9000 Gas Chromatograph. Chromatec Crystal Laboratory Gas Chromatography System

Meeting the Requirements of EN12916:2006 (IP391/07) Using Agilent 1200 Series HPLC Systems

Productivity Improvements for Dual Column Applications: USP <467> and Blood Alcohol Analysis

Large Volume Injection of Polycyclic Aromatic Hydrocarbons

Restek Corporation 110 Benner Circle, Bellefonte, Pa Barry L. Burger, Neil Johansen, Valerie Gamble, Donald Rhoades

Sulfur Detection at ppb Levels in Light Hydrocarbon Streams

Forensic Blood Alcohol Determination with the Intuvo 9000 GC

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard

[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS

Jaap de Zeeuw Restek Middelburg, The Netherlands. Copyrights: Restek Corporation

Phenomenex Exclusive! Inlet Base Seals. Agilent 5000, 6000 and. Guaranteed fit for Series GCs. Easy Seals

Impurity Testing of Fixed-Dose Combination Drugs Using the Agilent 1290 Infinity II HDR-DAD Impurity Analyzer Solution

ASTM D 6730 Detailed Hydrocarbon Analysis

Determination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105

TECHNICAL REPORT. Introduction. Agilent 1100 HPLC system. Figure 1: HALO columns exhibit UHPLC-like performance at conventional HPLC pressure

Choose for Your Instrument. GC Supplies - Agilent Instrument Quick Pick Guide

Installing the 6850 Direct Valve Column Connector Kit

The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector

Agilent Portfolio. Luca Godina. Technical Marketing Specialist GC-GCMS

Same Column and Gas Type Try Different/Faster Velocities. Same Column, Switch He to H2 Carrier Then Try Faster Velocities

Choose for Your Instrument

Agilent 7693A Automated Liquid Sampler

Agilent 6890 Gas Chromatograph

ANALYSIS OF GASOLINE RANGE HYDROCARBONS ON BP1-PONA

Transcription:

Application Note Petrochemicals Using the PSD for Backflushing on the Agilent 889 GC System Author Brian Fitz Agilent Technologies, Inc. Wilmington, DE, USA. Abstract An Agilent 889 series GC equipped with an Agilent capillary flow technology Deans switch coupled to flame ionization detection and flame photometric detection was used to analyze a heavy hydrocarbon distillate: residual fuel oil. Residual fuel oils typically contain hydrocarbons in the range from C 1 to C 7 with a significant amount of sulfur-containing compounds. To prevent carryover without requiring excessive column bake-out periods, the use of backflushing is necessary. An electronic pneumatic control (EPC) module called the pneumatic switching device (PSD) was used to accomplish both Deans switching and backflushing in a single chromatographic method.

Introduction The use of backflushing in gas chromatography (GC) is essential for obtaining reproducible results in a timely manner when analyzing complex samples containing high boiling point compounds. The benefits of backflushing have been documented extensively 1-. Backflushing is growing in popularity due to ease-of-use improvements such as Agilent capillary flow technology (CFT) devices 4. The recently released Agilent Intuvo 9 GC system delivers easy-to-use backflush capability as a standard option 5,6. With the release of the Intuvo 9 GC system came a newly designed EPC module (available on the 889 GC) called the PSD. The PSD has two pneumatic control channels. The primary channel is a forward pressure-controlled channel. This would typically be used to supply pressure for a backflush or CFT device, similar to the AUX EPC or PCM. The second channel of the PSD (called the purge flow) is an engineered bleed restrictor for the first channel. The purge flow is a user-controlled setpoint with a range of to ml/min with a default setpoint of ml/min. The purge flow has two main functions. First, it allows for better pneumatic control when the PSD is providing low volumetric flow. To supply the engineered bleed, a minimum amount of supply pressure from the primary channel is required. The requisite primary channel pressure ensures that the EPC proportioning valve functions in a stable regime. For example, in a midcolumn backflush configuration, the midpoint pressure source may only provide a few tenths of ml/min of total flow to the second column. Without this purge flow, the valves would not be able to control the flow accurately due to having to control a low supply pressure. To address this issue in previous pneumatic configurations, a bleed restrictor had to be constructed manually by cutting into the pressure line and installing a tee and restrictor. The purge flow engineered into the PSD provides a built-in bleed restrictor. The second function of the purge flow is that it can be held at constant flow with varying input pressures, which helps conserve carrier gas. For example, a typical backflush system uses a fixed restrictor, such as 1 m of 5 µm fused silica tubing. At high pressures (that is, during backflush), the fixed restrictor can have hundreds of ml/min of wasted flow. The PSD will stay at the user defined setpoint (default ml/min) even at high pressures. FPD Plus Secondary, DB-17ht m.5 mm,.15 µm Experimental Figure 1 shows a schematic of the 889 GC system used. The Deans switch was configured to cut between.1 and.4 minutes. This cuts 4,6-dimethydibenzothiophene from column 1 to column for detection with FPD Plus. The multimode inlet (MMI) was used. All analyses used helium as the carrier gas in constant flow mode. See Table 1 for additional instrumental parameters used. Table gives the backflush settings. A simulated distillation separation was performed to analyze the carbon chain distribution in the residual fuel oil. An Agilent J&W DB-HT Sim Dis column (5 m 5 µm,.15 µm) was used. This experiment did not use the Deans switch apparatus. Table lists the parameters used for the simulated distillation method. Table 4 lists relevant consumables used in the experiment. Samples The heavy distillate analyzed was NBS 16c % sulfur in residual fuel oil (RFO). This was diluted 1:4 in toluene, then injected 1 µl splitless. The polyethylene standard (Polywax 5) was diluted to.1 % in toluene, and injected 1 µl splitless. MMI Deans switch CFT PSD Primary, HP-1 m.5 mm,.5 µm Restrictor, deactivated F.S..77 m 1 µm Figure 1. Schematic of the 889 GC system configured with Deans switch with the PSD.

Table 1. Instrumental parameters. Table. Backflush parameters. Gas chromatograph Automatic liquid sampler Inlet type 889 Series GC Agilent 769A automatic liquid sampler (1 µl injection) MMI Oven (post run) 6 C (5 minutes) Inlet temperature 45 C Inlet purge flow 1 ml/min MMI program 1 C (. minutes), 9 C/min to 45 C Oven program 5 C (1 minute), 1 C/min to 5 C (1.5 minutes) Column 1 Agilent J&W DB-1ms UI, m 5 µm,.5 µm, ml/min (helium) Column Agilent J&W DB-17ht, m 5 µm,.15 µm, ml/min (helium) PSD 7 psi (4.5 ml/min column /) Inlet psi (4.5 ml/min column 1) Restrictor.77 m 1 µm deactivated fused silica, ml/min (helium) (controlled through column ) Aux. pressure source PSD purge flow FPD+ Deans switch window Pneumatic switching device (PSD) ml/min (default) Sulfur filter (94 nm) Transfer line: 5 C Emission block: 15 C Air: 6 ml/min Hydrogen: 6 ml/min Nitrogen: 6 ml/min.1 to.4 minutes Table. Simulated distillation parameters. Column Agilent J&W DB-HT Sim Dis, 5 m 5 µm,.15 µm Carrier flow 5 ml/min helium (constant flow) Inlet (MMI) 1 C (. minutes), 9 C/min to 45 C Oven program 4 C (no hold), 1 C/min to 4 C (5 minutes) 45 C Air: 45 ml/min Hydrogen: 4 ml/min Nitrogen: ml/min Table 4. Consumables used. Syringe Blue Line, 5 µl, tapered (p/n G451-86) Liner Ultra Inert, split, glass wool (p/n 519-95) Ferrules Flexible metal ferrules, UltiMetal Plus,.4 mm id (p/n G188-751) Column 1 J&W DB-1ms UI (p/n 1-1UI) Column J&W DB-17ht (p/n 1-181) Software Agilent OpenLab. Results and discussion Figure shows an overlay of the NBS 16c RFO and the Polywax 5 calibration standard obtained with the SIMDIST parameters. The carbon chain distribution of the RFO appears to tail off at the end of the Polywax distribution near C 7, which has a boiling point of 647 C 7. If this sample were analyzed with a typical chromatographic setup using standard capillary columns, there would be significant carryover, as much of the heavy hydrocarbon backbone would not elute. The full range of boiling points in a sample is often not known prior to beginning analysis, but in this case, it shows the need for backflushing. response (pa) 1. 1.5 1..5 C 1 C C 4 NBS 16c residual fuel oil (top) Polywax 5 calibration standard (bottom) 5 1 15 5 5 4 Figure. Chromatogram of NBS 16c overlaid with Polywax 5 calibration standard. The NBS RFO sample contains compounds ranging between C 1 and C 7. C 7

Figure A shows an overlay of three replicate injections of the NBS 16c RFO separated on the J&W DB-1ms UI column and detected by with the Deans switch configuration. These injections did not use backflush. The separation ended at a final temperature of 5 C, near the upper operating range of both the J&W DB-1ms UI and J&W DB-17ht columns. The final peak that eluted was C 6. Each subsequent injection shows a growing baseline toward the end of the chromatogram, indicating that the sample is not fully eluting from the previous injection, and carryover is occurring. Comparing the chromatograms in Figure A to the chromatograms in Figure, it is clear that a significant portion of the sample remains on the column (the portion from C 6 to C 7 ). The region from.1 to.4 minutes marked in Figure A was cut to the second column for detection with the FPD Plus. Figure B shows the cut region from (A) separated on the secondary column (J&W DB-17ht) and detected with the FPD Plus. The tallest peak is 4,6-dimethyldibenzothiophene (4,6-DMDBT) with two unidentified smaller peaks on either side. The retention time shifts significantly, and the area precision is poor. This is a common side-effect of having a large amount of carryover between runs, as evidenced by the increasing baseline in A. Figure C shows a no-inject blank after the three injections of the RFO. The heart cut still occurs at.1 to.4 minutes, and a small peak of 4,6-DMDBT appears in the FPD Plus channel. There is still significant carryover on the channel as evidenced by the growing baseline at the end of the chromatogram. response (pa) FPD Plus response (15 pa) Detector response 1 4 1 5 1 15 5 1 4. 1.5 1..5 1 A B FPD Plus 4,6-Dimethyldibenzothiophene Zoom Cut to FPD Plus Injection (top) Injection (middle) Injection 1 (bottom) 5 1 15 5 4 C 1 FPD Plus ( 15pA) 4,6-DMDBT carryover Heavy analyte (>C ) carryover (pa) 5 1 15 5 Figure. A) Overlay of three injections of NBS 16c RFO with a narrow heart-cut.1 to.4 minutes and no backflush. B) Overlay of the three cuts from (A) of 4,6-DMDBT detected with FPD Plus. C) No-inject blank run after three injections of (A). C 6 4

Figure 4A shows an overlay of three replicate injections of the NBS 16c RFO with the same experimental parameters as Figure, but with backflush. See Table for backflush parameters. During the backflush, the PSD is held at 7 psi to backflush column 1 with 4.5 ml/min of flow (towards the inlet). The purge flow is held at ml/min. A fixed restrictor of 1 m 5 µm would allow nearly 5 ml/min of flow during the 7 psi backflush. The PSD provides significant savings in gas flow. The end of the separation, from 5 to minutes in Figure 4A, is very reproducible. There is no increase in baseline, as seen in the nonbackflush chromatogram in Figure A. Figure 4B shows the cut region at.1.4 minutes from Figure 4A. The retention time and area precision are remarkably improved. Figure 4C shows a no-inject blank following the three injections of the RFO using backflush. There is no visible carryover of heavy analytes in the channel. This shows that the backflush works well. response (pa) FPD Plus response (15 pa) 1 4 A 1 Cut to FPD Plus 5 1 15 5 1 4. 1.5 1..5 1 B 5 1 15 5 4 C FPD Plus 4,6-Dimethyldibenzothiophene Zoom Detector response 1 FPD Plus ( 15 pa) (pa) 5 1 15 5 Figure 4. A) Overlay of three injections of NBS 16c RFO with a narrow heart-cut at.1.4 minutes with backflush. B) Overlay of the three cuts from (A) of 4,6-DMDBT detected with FPD Plus. C) No inject blank run following three injections of NBS 16c RFO from (A). 5

Conclusion The 889 GC system coupled with a Deans switch to and FPD Plus with backflushing is shown to provide reproducible analyses of a heavy hydrocarbon distillate: an RFO sample with a carbon chain distribution from C 1 to C 7. The PSD provided backflush capability with significantly reduced carrier gas consumption due to the fixed purge flow. The use of backflushing can extend column lifetime due to not requiring extended high temperature bake-outs. This also helps increase sample throughput due to shorter run to run times. References 1. Tranchida, P. Q.; et al. Heart cutting multidimensional gas chromatography: A review of recent evolution, applications, and future prospects. A. Chem. Acta 1, 716, 66 75.. Seeley, J. V. Recent advances in flow-controlled multidimensional gas chromatography, J. Chromatogr. A 1, 155, 4 7.. Meng, C-K. Improving productivity and extending column life with backflush, Agilent Technologies Application Brief, publication number 5989-618EN, 6. 4. Agilent CFT Backflush Brochure, publication number 5989-948EN, 1. 5. Westland, J. Examining maximum residue levels for multiresidue pesticides in jasmine rice. Agilent Technologies Application Note, publication number 5991 99EN, 18. 6. Westland, J. Meeting European Union maximum residue level regulations for pesticides in tea and honey. Agilent Technologies Application Note, publication number 5991-98, 18. 7. ASTM Standard D65-15, 15, Standard Test Method for Boiling Range Distribution of Petroleum Distillates in Boiling Range from 174 F to 7 F by Gas Chromatography, ASTM International, West Conshohocken, PA. DOI 1.15/D65-15. www.agilent.com/chem This information is subject to change without notice. Agilent Technologies, Inc. 18 Printed in the USA, December 14, 18 5994-55EN