ACCUSTATION SUBMITTAL MATERIALS

Similar documents
ACCUVALVE MODEL AVT6000 SUBMITTAL MATERIALS MODEL CODE. AccuValve Model AVT6000 AVT6 - - DWG. NO: REVISION: REV. DATE:

ACCUVALVE MODEL AVC6000 SUBMITTAL MATERIALS MODEL CODE. AccuValve Model AVC6000 AVC6 - - DWG. NO: REVISION: REV. DATE:

Accutrol, LLC Product Sheet H-AVC5000. US Patents 6,991,177 & 7,543,759. Specially Made for Vivarium Cage Racks

OPTIONS BLANK = No Options B = BACnet MS/TP D = Internal Transmitter Display (Ref Note 4)

Accutrol, LLC Product Sheet. VTS Series. US Patents 6,991,177 & 7,543,759. Digital Airflow Measurement Station

Accutrol, LLC Product Sheet AVT6000. US Patents 6,991,177 & 7,543,759. Laboratories, Life Sciences, Healthcare

Accutrol, LLC Product Sheet AVC6000. US Patents 6,991,177 & 7,543,759. Laboratories, Life Sciences, Healthcare

Accutrol, LLC Product Sheet AVC5000. US Patents 6,991,177 & 7,543,759. Laboratories, Life Sciences, Healthcare

Acid Resistant AVC6000

Acid Resistant AVT6000

Installation & Operation Manual - Model AV3000

MANUAL INSTALLATION. Venturi FX. VFX Series. v100 Issue Date: 11/22/ Price Industries Limited. All rights reserved.

P2050S-030, 1/2, Electronic Pressure Independent Valve Stainless Steel Ball and Stem, Female NPT Ends

MODELS TSX AND TSX-S SINGLE DUCT ROUND AIR TERMINALS

F6250L, 10, 2-Way Butterfly Valve Resilient Seat, 304 Stainless Steel Disc

F6300L, 12, 2-Way Butterfly Valve Resilient Seat, 304 Stainless Steel Disc

MODELS SGX AND SSX SINGLE DUCT ROUND AIR TERMINALS

INSTALLATION & OPERATION MANUAL. Fan Powered Terminals VAV TERMINALS. Redefine your comfort zone.

F6250L, 10, 2-Way Butterfly Valve Resilient Seat, 304 Stainless Steel Disc

F6200L, 8, 2-Way Butterfly Valve Resilient Seat, 304 Stainless Steel Disc

Mount directly to 1.05 jackshafts. (new ZG-120 bracket shown) EFB24-S N4, EFX24-S N4(H) EFB24-S, EFX24-S (p. 21)

AL-Maraj Central A/C Systems Accessories Mfg Co.

F6150HD, 6, 2-Way Butterfly Valve Resilient Seat, 304 Stainless Steel Disc

AMD-42-TD. Thermal Dispersion Air Measuring Station with VCD-42 Control Damper. Application and Design. Ratings. Features and Control Options

B. Base occupied space sound level estimates on ARI 885. C. Terminal heating coils shall conform to ARI 410.

Installation Instruction ALG Series LED Area Light Version: 3.1

STR Circular VAV Terminal Units

University of Delaware

F6150HD, 6, 2-Way Butterfly Valve Resilient Seat, 304 Stainless Steel Disc

Powers TM Controls EA 338 Electronic Actuator

CFP-LED Contractor Select Flat Panel. Installation Instructions

ACCESSORIES & OPTIONS FOR VENTURI AND VENTURI FX VALVES

Installation Instructions CONSTANT WATT (CW) AND SELF-REGULATING (SR) PIPE TRACING CABLES

Advanced Outdoor Airflow Control, DCV Control and Alarm Solution for OA Applications Document Name: QG_EAC1100_A_R3A

Safety & Installation Instructions

Minimum 27 in-lb torque

Variable Air Volume (VAV) Pressure Independent Control

LED IMPORTANT SAFETY INSTRUCTIONS

COMPACT CAGE RACK VALVES

The IAQ50 Air Damper/Monitor Installation and Maintenance Manual

Versatile and Powerful

Installation Instructions

Minimum 27 in-lb Torque

Installation Instructions Mineral Insulated (M.I.)

TSS Single-Duct VAV Terminals

NFB24, NFB24-S, NFX24, NFX24-S On/Off, Spring Return, 24 V

VAV TERMINAL UNIT KYODO-ALLIED TECHNOLOGY PTE LTD

AMD-23-TD. Thermal Dispersion Air Measuring Station with VCD-23 Control Damper. Application. Ratings. Features and Options

TUNDRA 2017 Dual Black Tip Performance Exhaust Part Number: Code: EX3000 Conflicts Kit Contents Item # Qty. Reqd.

PH SERIES SINGLE DUCT ATU PERFORMANCE DATA

AMCA Seal Assures rating of Sound and Air performance. 2 year fan housing warranty, 1 year motor warranty

U00X ULTRASONIC LEVEL SWITCH. Ultrasonic Liquid Level Switches INSTALLATION AND OPERATIONS MANUAL. For Models: U002, U003 & U004

SINGLE DUCT AIR TERMINAL UNITS

FAN POWERED TERMINAL UNITS

Pressure Transmitter. Informational & Installation Manual

Technical specification

OE326-23I-OR VAV/Zone Controller

Laboratory Exhaust Air Terminal

ZG-JSL, ZG-JSLA Jackshaft Retrofit Linkage For AF, NF, LF, NMX and AMX Series Actuators

Industry best 5 year fan housing warranty, 1 year motor warranty

Installation and Operation Instructions DLP Series

Venturi Valve. 2 For more information visit v100. Actuator. Potentiometer. Lever Arm. Airflow Area AIRFLOW.

Evolution. ECM Motor. T Control CONTROLS INC. EVO /ECM-DTC16. Differential Temperature Control. Specifications. Applications

SBH / SBV Sales Guide BLOWER-COILS HORIZONTAL AND VERTICAL

Lube Level Maintainer

MODELS FF & FFC CEILING AND INLINE/CABINET VENTILATORS

MANUAL INSTALLATION. Venturi Valve. VV Series. v203 - Issue Date: 03/25/ Price Industries Limited. All rights reserved.

Installation Instructions

Installation Instructions

Laboratory Exhaust Air Terminal

U00X ULTRASONIC LEVEL SWITCH. Ultrasonic Liquid Level Switches INSTALLATION AND OPERATIONS MANUAL. For Models: U002, U003 & U004

www. ElectricalPartManuals. com Instruction Bulletin ALTIVAR FLEX58 TRX Adjustable Speed Chassis Drive Controllers Installation Guide

Mini Multimeter with Non-Contact Voltage Detector (NCV)

MODELS STXD & STXDe DIRECT DRIVE CENTRIFUGAL ROOF EXHAUSTERS

MODELS FF & FFC CEILING AND INLINE/CABINET VENTILATORS

UV-AIRE AIR PURIFYING SYSTEM

Matrix AP 400V 690V INSTALLATION GUIDE. Quick Reference. ❶ How to Install Pages 6 20 ❷ Startup/Troubleshooting Pages WARNING

GE Measurement & Control

CURRENT TRANSFORMER MODEL N. SS BAND SECURES BRACKET (view of screw clamp band omitted for clarity) 0.25 x 0.50 SLOT (2 PLACES)

Installation Instructions

Chevrolet Camaro (Gen6) Installation Guide PN 11920, 11921, 11924, 11925, 11926, 11927, 11930, 11931

FSAF24(-S) US, FSAF120(-S) US

Assembly instruction. Pendant Light Toccata GENERAL REMARKS ON SAFETY: Technical data: BOM

Installation Instructions Electric Heaters 5 20 kw

F-4600 INLINE ULTRASONIC FLOW METER Installation and Operation Guide

Installation Instructions

Minimum 90 in-lb Torque. Applications. NFB, NFX Series At A Glance. NFB and NFX Series Spring Return Direct Coupled Actuator

Installing the Switch Chassis

LED IMPORTANT SAFETY INSTRUCTIONS

Nature Power Inverters. True Sinewave Inverter Modified Sinewave Inverter. Owner s Manual

Installation Instructions

Dual Duct Variable Air Volume Terminal (Model TDS)

F-2600 SERIES INLINE VORTEX FLOW METER

INSTALLATION INSTRUCTIONS K-SERIES UNDER CABINET HOOD

Installation Instructions

Laboratory Room Single Duct Supply Air Terminal

OPTIMAX RADIANT FIELD RETROFIT INSTRUCTIONS

F-2700 SERIES INSERTION VORTEX FLOW METER

MA-5200 Series MA-5300 Series

MAx-305 & MAx-318 Series MAx-405 Through MAx-419 Series Two-Position Actuators

Transcription:

ACCUSTATION SUBMITTAL MODEL CODE WARNING: NOT FOR USE WITH PERCHLORIC ACID MATERIALS HOUSING MATERIAL 2 = 304SS, 20 GAUGE 3 = 316SS, 20 GAUGE 4 = ALUMINUM, 16 GAUGE SIZE 06 = 06" DIAMETER 08 = 08" DIAMETER 10 = 10" DIAMETER 12 = 12" DIAMETER 14 = 14" DIAMETER 18 = 12"x18" RECTANGULAR 24 = 12"x24" RECTANGULAR 36 = 12"x36" RECTANGULAR 48 = 12"x48" RECTANGULAR VTS - OPTIONS BLANK = NO OPTIONS B = BACnet MS/TP F = FLANGES (REF NOTE 1) I = INSULATION Model Material Designator (2) 304SS (3) 316SS (4) Aluminum Housing 304 Stainless Steel 316 Stainless Steel Al. Alloy 02-H32 Compression Section 304 Stainless Steel 316 Stainless Steel Al. Alloy 02-H32 Static Regain Section 304 Stainless Steel 316 Stainless Steel Al. Alloy 02-H32 End Plate 304 Stainless Steel 316 Stainless Steel Galvanized Steel Vortex Sensors Materials Exposed to the Airstream UL94-V0 UL94-V0 UL94-VO Sensor Tubing Polyurethane, Ether-based Polyurethane, Ether-based Polyurethane, Ether-based Compression Seals Viton Rubber Viton Rubber EPDM Rubber Machine Screws 304 Stainless Steel 316 Stainless Steel 304 Stainless Steel Rivets 304SS 316SS 304SS MODEL CODE NOTES: 1) Standard AccuStation flanges do NOT include predrilled holes. Reference Accutrol Flange Detail Submittal Drawing for details. OPERATING RANGE SIZE AND WEIGHT Model Code Min. Flow Measured Full Scale Range CFM L/S CMH CFM L/S CMH VTSX06 30 14 1 31 149 3 VTSX08 80 38 136 800 378 139 VTSX10 120 7 204 1300 613 2209 VTSX12 180 8 306 1790 84 3041 VTSX14 20 118 42 270 1298 4672 VTSX18 260 123 442 3200 110 437 VTSX24 30 16 9 4000 1888 6796 VTSX36 20 24 883 6400 3020 10874 VTSX48 700 330 1189 8000 377 1392 Dimensions (Reference Sheet 2) Weight Model Code "D" or "W" "L" "H" Stainless Steel Aluminum in. mm in. mm in. mm Lbs. kg Lbs. kg VTSX06.88 149 22 9 10 24 11.0 7 3.2 VTSX08 7.88 200 24 610 13 330 12.4 8 3.6 VTSX10 9.88 20 24 610 1 381 16 7.3 10 4. VTSX12 11.88 300 27 686 17 432 22 10.0 12.4 VTSX14 13.88 30 30 762 19 483 2 11.3 1 6.8 VTSX18 17.88 44 30 762 19 483 37 16.8 20 9.1 VTSX24 23.88 607 30 762 19 483 43 19. 23 10.4 VTSX36 3.88 911 30 762 19 483 88 39.9 0 22.7 VTSX48 47.88 1216 30 762 19 483 98 44. 8 26.3 2-21-19 SHEET: 1 OF:

ROUND AccuStation VTSX06 VTSX08 VTSX10 VTSX12 VTSX14 RECTANGULAR AccuStation VTSX18 VTSX24 Access Cover 19" 483mm 11.88" 302mm Access Cover Housing Inlet View W 30" 762mm VTSX36 VTSX48 L D 30" 762mm 11" (280mm) 11.88" 302mm 6" 12mm H W 7" 178mm Inlet Discharge 22" 9mm Mounting Strut 2-21-19 SHEET: 2 OF:

ROUND: INSTALLATION INSTRUCTIONS 1. Read all instructions prior to beginning installation. NOTE: For detailed installation instructions, refer to the AccuStation Installation & Operation Manual. 2. Verify the tag number located on the AccuStation label matches the HVAC schedule. WARNING: Wear eye protection, protective gloves and clothing suitable for working with sheet metal which may have sharp edges. 3. Locate the duct section which the AccuStation is servicing and select a suitable mounting location for it. NOTES: The AccuStation does not require straight inlet duct runs to operate properly, however it's always best to locate the AccuStation away from transitions and bends to minimize impact on system static pressure. Be sure to select a location that will provide a minimum clearance of 14 inches (36 mm) unobstructed access to the transmitter and access cover. The AccuStation is not position sensitive. It can be installed in any plane or rotational axis without having impact on the performance. 4. Provide an opening in the selected duct section sized appropriately for the AccuStation being installed. NOTE: A slip-fit AccuStation will require an opening approximately 2" (0.8 mm) smaller than the AccuStation length, whereas a flanged AccuStation will require an opening the same length as the AccuStation. Reference Sheet 1 for AccuStation dimensions.. Install duct hangers within 12 inches (30 mm) from each end of the AccuStation. Reference Sheet 1 for AccuStation weights. WARNING: Use duct hangers and hardware designed to support the total load of AccuStation and associated duct sections. Failure to do so may result in serious personal injury or death. 6. Install the AccuStation into the duct in accordance with the Airflow Direction Label located on the AccuStation. Position the AccuStation for easy access to the transmitter side then secure to duct per the appropriate Figure below. NOTE: Screws, nuts, fasteners, duct sealant, hangers, and gaskets are not provided by Accutrol LLC. Standard Slip-fit AccuStation Secured Using Tek Screws Secure AccuStation to duct at both ends using tek screws and duct sealant. Standard Slip-fit AccuStation Secured Using Draw Bands (Draw Bands are Sold Separately) Reference Accutrol Draw Band Clamp Submittal Drawing for Details After sealing joints with appropriate type of tape, secure both ends using draw band clamps. Flanged AccuStation Option F Secured Using Companion Flanges (Companion Flanges are Sold Separately) Reference Accutrol Flange Option Submittal Drawing for Details Install companion flanges to duct ends and secure to duct. Apply duct sealant and/or gasket to flange face. Install AccuStation and secure flanges using tekscrews. 14" (36mm) 14" (36mm) Figure 1 Figure 2 14" (36mm) Figure 3 2-21-19 SHEET: 3 OF:

RECTANGULAR: INSTALLATION INSTRUCTIONS 1. Read all instructions completely before installing the AccuStation. WARNING: Wear eye protection, protective gloves and clothing suitable for working with sheet metal which may have sharp edges. 2. Verify the tag number located on the AccuStation label matches the HVAC schedule. 3. Select optimum mounting location for the AccuStation. NOTE: The AccuStation does not require straight inlet duct runs to operate properly, however it s always best to locate any duct device away from transitions and bends to minimize impact on system static pressure. 4. Allow a minimum clearance of 14 inches (36 mm) unobstructed access to the transmitter, and access cover. NOTE: Rectangular AccuStations are normally installed with the access side facing downwards for easy access. However, the AccuStation is not position sensitive. It can be installed in any plane or rotational axis without having impact on the performance.. To support the weight of the AccuStation, install duct hangers within 12 inches (30 mm) of AccuStation connections. The 12"x36" and 12"x48" AccuStations include an integral mounting strut which shall be used to support the AccuStation in addition to the duct hangers. Reference Sheet 1 for AccuStation weights. WARNING: Use duct hangers and hardware designed to support the total load of the AccuStation and associated duct sections. Failure to do so may result in serious personal injury or death. 6. After the duct section is properly supported to carry the weight of the AccuStation, install into the duct in accordance with the Airflow Direction Label located on the AccuStation. Position AccuStation so the transmitter, actuator and access cover are easily accessible. For 12x36" and 12x48" AccuStations, attach the integral mounting bracket to threaded rod or duct hangers capable of supporting AccuStation weight. 7. Reference the appropriate diagram to the right for installation details. NOTE: Screws, nuts, fasteners, duct sealant, hangers, companion flanges and gaskets are not provided by Accutrol LLC. Figure 1 Standard Slip-fit AccuStation Using Tek Screws Figure 2 Flanged AccuStation Option F Using Companion Flanges Figure 3 12x36" and 12x48" AccuStations RECTANGULAR: INSTALLATION DIAGRAMS Support AccuStation Support Struts WARNING: Dual AccuStations are provided with an integral mounting strut to help support the weight of the AccuStation. In addition to supporting the duct within 12" of the AccuStation, each end of the integral mounting strut shall be secured to the building structure using properly rated hardware and methods in accordance with local building codes. Failure to do so may result in serious personal injury or death. Secure both ends using tek screws and duct sealant 14" (36mm) Mounting Apply Sealant and/or Gasket, then Secure both ends using tek-screws 14" (36mm) Support Apply Sealant Support Support Here 2-21-19 SHEET: 4 OF

TRANSMITTER WIRING INSTRUCTIONS Connect the power wires to the per the diagram. CAUTION: Maintain polarity if the power source is used to power multiple devices otherwise equipment may be damaged. CAUTION: Do not use the transmitter enclosure as a junction box. The only wires entering the transmitter enclosure shall be wires connected to the transmitter otherwise equipment may be damaged. Optional BACnet MS/TP Module Connect the 2-wire EIA 48 BACnet MS/TP bus to the transmitter per the diagram. Network bias and EOL termination are not provided by the transmitter. Remote Monitor AFM (Sold Separately) Plug one end of the factory cable into the back of the remote monitor and plug the other end into the transmitter Display Port. VTM Remote Monitor (sold separately) AIRFLOW MONITOR WARNING: During installation, testing, servicing and troubleshooting of this product, it may be necessary to work with live electrical components. This work shall be performed by a licensed electrician or qualified individual who has been properly trained in handling live electrical equipment. Failure to follow all electrical safety precautions when exposed to live electrical components may result in serious injury or death. TRANSMITTER ELECTRICAL SPECIFICATIONS Power Supply: 24 VAC +/- 20% 0/60 Hz, 4 VA max (8. VA max with remote monitor) 24 VDC +/- 10% 1. W max (3. W max with remote monitor) Output Signal: 0-20mA, 4-20mA, 0-10V, 2-10V, 0-V, or 1-V (Software Config.) 12-bit Resolution Capable of driving 1K-ohm load Configuration: USB Type C Connector Power Source Switch: Selects alternate power source for configuration when main power is not available by drawing V from PC connected to USB config. port BACnet MS/TP (Optional): EIA 48 2-wire BACnet MS/TP, Galvanically Isolated Data Rates 9600, 19200, 38400, 7600, 76800 and 11200 1/8 Unit Load Receiver Input Impedance Network bias and EOL Termination not provided within the Remote Monitor (Sold Separately): LCD, 2 lines x 8 characters with white LED backlight Includes USB Configuration Port and Factory Cable Terminal Plugs: 3-position, vertical pluggable, screw on top, wire size 12-30 AWG EIA 48 BACnet MS/TP Bus 24V Power Common Output Signal BACnet MS/TP (Optional) 2-21-19 SHEET: OF: