The Analysis of Biodiesel for Inorganic Contaminants, including Sulfur, by ICP-OES

Similar documents
Analysis of biodiesel oil (as per ASTM D6751 & EN 14214) using the Agilent 5100 SVDV ICP-OES

Analysis of Petroleum Samples Using the Teledyne Leeman Labs

Prodigy ICP Application Note: # 1039

Application for Energy and Fuels

Application for Energy and Fuels

Advanced Petrochemical Applications With ICP-OES. Tina Harville, Ph.D. Application Engineer

The Analysis of Biodiesel for Trace Metals and the Development of Certified Biodiesel Standards

Free and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN and ASTM D6584

Analysis of Wear Metals Using the Teledyne Leeman Labs Prodigy Plus ICP-OES

Elemental Analysis of Biofuels with SPECTRO XRF and ICP Spectrometers

GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585

VHG Labs, Inc. Standards and Supplies for Spectrochemical Oil Analysis

The Importance of Biodiesel Meeting the ASTM D6751 Standard. Barbara E Goodrich Manager, Engine Fluids John Deere Power Systems 6 February 2008

A Comparison of Boat Introduction and Direct Injection using the Thermo Scientific ipro 5000 Series Combustion Analyzer

Analysis of Sulfur and Trace Element Content in FAME

Paragon Scientific Ltd Proficiency Testing Scheme Schedule

Project Reference No.: 40S_B_MTECH_007

Chemical and Petrochemical Applications of Microwave Plasma Atomic Emission Spectroscopy (MP-AES)

SPECTRO XRF Report. SPECTRO iq II

Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers. Versa HT3. Application Note. Abstract.

Detection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector

Biodiesel Production from Waste Cooking Oil A Renewable Blend for Diesel Engines

Northville Product Services and Colonial Pipeline B5 Soy Test Data and Report February Soy blended low sulfur diesel

Alberta Innovates - Technology Futures ~ Fuels & Lubricants

ASTM D Standard Specification for Biodiesel Fuel (B 100) Blend Stock for Distillate Fuels

Stability, Linearity and Repeatability of Nitrogen Determination by Flash Combustion using Argon as Carrier Gas

Application Note. Author. Introduction. Energy and Fuels

Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593

Quantification and Characterization of Sulfur in Low-Sulfur Reformulated Gasolines by GC-ICP-MS Application

Monitor Chlorine in Crude at Sub-ppm Levels

lubricants analysis. complete solutions for your lab. Lubricants Analysis

EDXRF APPLICATION NOTE SULFUR IN BUNKER FUEL # 1230

Determination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105

The Role of Chemical Metrology in Transportation Industry

Abstract Process Economics Program Report 251 BIODIESEL PRODUCTION (November 2004)

Biodiesel Business Environment

Standard Test Method for Sulfur in the Analysis Sample of Coal and Coke Using High-Temperature Tube Furnace Combustion

Louis Dreyfus Claypool Holdings, LLC. Biodiesel Production Plant Claypool, Indiana

High Sensitivity UHPLC-DAD Analysis of Azo Dyes using the Agilent 1290 Infinity LC System and the 60 mm Max-Light High Sensitivity Flow Cell

Quantitative Analysis of Chemical Compositions from Various Sources of Crude Glycerine

The Purification Feasibilityof GlycerinProduced During

Analysis of Fatty Acid Methyl Esters (FAMES), and Examination of Biodiesel Samples for these Components, by GCxGC-FID

New Fuels for Oilheat

breakthrough versatility

BQ-9000 Quality Management System. Producer Requirements

Simultaneous Determination of Fatty Acid Methyl Esters Contents in the Biodiesel by HPLC-DAD Method

Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel

Phase Distribution of Ethanol, and Water in Ethyl Esters at K and K

Voting Draft Standard

SCOPE OF ACCREDITATION

Standard Specification for Biodiesel Fuel Blend Stock (B100) for Middle Distillate Fuels 1

Biodiesel Analytics. Important Parameters and their Meaning

Application Note. Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D using AC OXYTRACER

Totally Automated Method for the Determination of Sudan Dyes in Food via On-Line Filtration, SPE and HPLC Analysis

No reason to wait: Start the transition to renewable fuels now!

Biodiesel Analysis Utilizing Mini-Scan - Handheld Analyzer V.C. Gordon PhD, Bonanza Labs

Determination of Free and Total Glycerin in B100 Biodiesel

AVIO 200/500 ICP-OES CONSUMABLES AND SUPPLIES. Consumables and Supplies Reference Guide

Ultra Clean Gas Filters Reference Guide ULTRA CLEAN GAS FILTERS

Alternative Automotive Fuels - Biodiesel

This presentation focuses on Biodiesel, scientifically called FAME (Fatty Acid Methyl Ester); a fuel different in either perspective.

ElvaX Sulfur in Oil analyzer

Refinery and Natural Gas Analysis

Sulfur Detection at ppb Levels in Light Hydrocarbon Streams

SOUTH AFRICAN NATIONAL STANDARD

Joint Oil Analysis Program Spectrometer Standards VHG Labs Inc. Qualification Report For D19-0, D3-100 and D12-XXX Series Standards

Application Note. Authors. Abstract. Energy & Chemicals

ISBN SANS 342:2006 Edition 4 SOUTH AFRICAN NATIONAL STANDARD Automotive diesel fuel Published by Standards South Africa 1 dr lategan roa

Metals Determination in Biodiesel (B100) by ICP-OES with Microwave Assisted Acid Digestion

Meeting the Requirements of EN12916:2006 (IP391/07) Using Agilent 1200 Series HPLC Systems

DaimlerChrysler Alternative Particulate Measurement page 1/8

Author: Vincenzo Piemonte, Associate Professor, University UCBM Rome (Italy)

MB3600-CH30 Laboratory FT-NIR analyzer for biodiesel applications Suitable for production optimization and product quality assessment

Biodiesel Quality in Germany

Determining the Ethanol Content of Denatured Fuel Ethanol Using Near Infrared. Gulf Coast Conference Patrick Ritz PAC LP

Product Quality Bulletin

TIER 3 MOTOR VEHICLE FUEL STANDARDS FOR DENATURED FUEL ETHANOL

Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011

Joint Oil Analysis Program Spectrometer Standards SCP Science (Conostan) Qualification Report For D19-0, D3-100, and D12-XXX Series Standards

Biodiesel and SmartWay Grow and Go Go. EPA-MMTA Fuel-Saving Seminar June 15, 2007

(TOYOTA DO BRASIL LTDA)

Calibration & Verification Standards

Contents. Purpose. 1.Measurement Condition. 2.Sample. 3.Standard Requirement. 4.Contents of evaluation. 5.Result. 6.Conclusion. 7.Notices.

UPDATE OF THE SURVEY OF SULFUR LEVELS IN COMMERCIAL JET FUEL. Final Report. November 2012

BIODIESEL PRODUCTION BY A CONTINUOUS PROCESS USING A HETEROGENEOUS CATALYST

Experimental Investigation and Modeling of Liquid-Liquid Equilibria in Biodiesel + Glycerol + Methanol

Guidance on Changes to ASTM D1655 and Def Stan to Increase Allowable FAME Limits to 50 mg/kg

Elemental Analysis Advanced

SpectrAA Report. 2:04 AM 11/25/2014 Page 1 of 22

Article: Sulfur Testing VPS Quality Approach By Dr Sunil Kumar Laboratory Manager Fujairah, UAE

Agilent Solutions for the Petrochemical and Oleochemical Industries

Annex to the Accreditation Certificate D-PL according to DIN EN ISO/IEC 17025:2005

TULSION BIODIESEL PRODUCTION: WET VS. DRY WHICH METHOD SHOULD YOU USE?

Combustion Properties of Alternative Liquid Fuels

BQ-9000 Quality Management System. Accredited Producer Requirements

OPTIMIZATION OF BIODIESEL PRODCUTION FROM TRANSESTERIFICATION OF WASTE COOKING OILS USING ALKALINE CATALYSTS

C196-E081C. Ultra High Performance Liquid Chromatograph Nexera X2. Specifications

Green chemistry in the first year lab: Using biodiesel to teach general chemistry principles. Overview:

A Single Method for the Direct Determination of Total Glycerols in All Biodiesels Using Liquid Chromatography and Charged Aerosol Detection

Biodiesel is NOT raw vegetable oil or SVO (Straight Vegetable Oil) or refined oil or filtered used cooking oil.

Transcription:

application Note Biofuels Authors Zoe A. Grosser, Ph.D. Lee J. Davidowski, Ph.D. Pamela Wee PerkinElmer, Inc. Bridgeport Avenue Shelton, CT USA The Analysis of Biodiesel for Inorganic Contaminants, including Sulfur, by ICP-OES Introduction The alternative fuels market has been growing over the past few years and is expected to continue. Biodiesel production increased from 9 million tons/year in 2007 to 11.1 million tons/year in 2008, an increase of almost 20%. Although the fluctuating price of conventional petroleum-based fuel influences the rate of development and acceptance, it is generally acknowledged that petroleumbased fuel will run out. Governments around the world, such as the U.S., Brazil, China, and India have set goals for increased biofuel use and research into processing of raw materials that minimize competition with food and water is continuing.

Table 1. European and U.S. Specifications for Metals in Biodiesel (1,2). Element EN14214 A astm D6751 comments Limit L limit (mg/kg) Methods (mg/kg) Methods S 10.0 20846 (UVFL) 15.0 D5453 (UVFL) D1266 (lamp), D2622 (XRF), 20884 (XRF) D3120 (microcoul), D4294 (XRF) P 10.0 14107 (ICP-OES) 10.0 D4951 (ICP-OES) Na K Ca Mg sum 5.0 14108 (AA) sum 5.0 14109 (AA) sum 5.0 14538 (ICP-OES) sum 5.0 14538 (ICP-OES) UOP 391 UOP 389 Biodiesel is one example of an alternative fuel. It can be used in diesel engines in a 20% mix with conventional fuel without any modification to the engine. The emissions are lower and the fuel replaces some of the lubricity lost in removing sulfur to comply with new diesel fuel regulations. Sulfur limits in conventional diesel fuel have been lowered from 500 ppm to less than 15 ppm in the U.S. and less than 10 ppm in Germany and other EU countries. Measurements of sulfur and other metals in biodiesel are important to ensure adequate performance of the fuel. High Levels of Na and K can form a soap solution and are common catalysts of the biodiesel reaction. If not removed they can cause instability and filter clogging. Ca and Mg can also form soap and the creation of problems similar to those caused by Na and K. Sulfur and phosphorous are both considered carryover elements from vegetable oil P from phospholipids S from glucosinilates Sulfur and phosphorous are both potential catalyst poisons In the U.S. and Europe, standards have been established proving acceptable concentrations of these elements in biodiesel and are shown in Table 1. Although methodology for sulfur does not currently include inductively coupled plasma optical emission spectroscopy (ICP-OES), it is a modern technique analytically capable of this determination. There are several ASTM methods specifying ICP-OES for a variety of elements in similar matrices, such as D7151-05, D7111-05, D5185-05, and D4951-06. (3-6) This paper will demonstrate the measurement of sulfur using ICP-OES for biodiesel. Experimental The Optima 7000 (PerkinElmer, Inc., Shelton, CT, USA) was used for all measurements. Samples were obtained from a biodiesel producer and run with the addition of yttrium (Conostan, Ponca City, OK). Sulfur in Diesel Fuel series, blank, 5, 10, and 15 mg/kg calibration standards and multielement standards (Conostan, Ponca City, OK) were combined to provide calibration standards. Instrument conditions are shown in Table 2. Table 2. Instrumental ICP-OES Conditions for Biodiesel Analysis. Instrument RF Power Nebulizer Flow Auxiliary Flow Plasma Flow Sample Pump Flow Plasma Viewing Processing Mode Auto Integration Read Delay Rinse Replicates 3 Background Correction Spray Chamber Nebulizer Injector Total Analysis Time per Sample optima 7000 ICP-OES 1500 W 0.50 L/min 0.8 L/min 16.0 L/min 0.5 ml/min Radial Area 2 sec min-5 sec max 30 sec 30 sec one or two points Cyclonic MicroMist Ceramic 1.2 mm id ~2 minutes 2

The radial mode of viewing the plasma was used for all elements because of sufficient sensitivity. Because the sample is an organic liquid, little improvement in detection limit is seen using axial viewing because of the enhancement of the organic structure. Yttrium was chosen as an internal standard because it is not usually observed in biodiesel samples, has a clean spectrum with no interferences and does not interfere with the analytes of interest. Other elements, such as cobalt, may also be a suitable internal standard element. The MicroMist nebulizer (Glass Expansion, USA) was chosen because of the low uptake rate and excellent precision. Two wavelengths were evaluated for determination of sulfur in biodiesel. The wavelengths selected and the resulting profiles are shown in Figure 1 for the suite of blank and calibration standards in the radial mode. Although both wavelengths demonstrate a baseline free from structure, the S 180.669 nm wavelength is more sensitive. This wavelength was further evaluated for issues arising from typical samples of diesel and biodiesel, shown in Figure 2. No overlaps are observed on the wavelength of interest and a rugged location was identified for a background correction point as shown by the small cross on the baseline at a slightly higher wavelength. Results and discussion To ensure the accuracy of the method developed, NIST 2723 Sulfur in Diesel Oil, was prepared in duplicate. Table 3 shows the results for the analysis. The standard deviation and relative standard deviation of the three replicate measurements was very good. The relative percent difference between the two duplicate preparations and measurements was 3.5% which is well below the 20% usually deemed acceptable for environmental programs. The reference value of the standard was recovered within the tolerance of the certified value. Figure 1. Evaluation of two wavelengths for sulfur determination, 180.669 nm and 181.975 nm. Table 3. Results for Sulfur (180.669 nm) from Measurement of a NIST Standard Reference Material. Sample ID Mean (mg/kg) SD (mg/kg) %RSD RPD NIST 2723 I 11.3 0.10 0.89 NIST 2723 II 10.9 0.07 0.61 Mean of Duplicates 11.1 3.5 NIST Certified Value 11.0 ±1.1 Figure 2. Calibration standards, NIST material and samples at 180.669 nm. 3

Several biodiesel samples from different batches were measured in the same fashion and the results are shown in Table 4. The starting material for the samples is most likely soybeans, but the sampling point in the manufacturing process (whether completely cleaned) is unknown. Table 4. Biodiesel Sulfur Results. Sample ID Mean (mg/kg) SD (mg/kg) %RSD B8-10-1 17.5 0.059 0.33 B8-10-2 1.56 0.11 7.13 Lot 95081 4.9 0.096 1.95 The results show measurements for a variety of concentrations, with good precision, even at the 1.5-mg/kg concentration level. The detection limit for sulfur was measured to be 0.01 mg/kg in the original fuel. The most efficient analysis would be to measure all necessary elements in biodiesel samples at the same time. ICP-OES is a multielement technique, suited for this purpose and elements were added to the existing sulfur method. After optimization the detection limits were measured using blank oil and calculated for the original starting material (method detection limits). Detection limits are shown in Table 5 and show that we can detect far below the compliance limit, ensuring confidence at higher concentrations. Table 5. Method Detection Limits for Biodiesel Analysis. Element S 180.669 0.01 P 178.221 0.09 Na 589.592 0.1 K 766.490 0.1 Ca 317.029 0.05 Mg 285.213 0.01 Biodiesel Detection Limit (mg/kg) Table 6 shows the results for the entire suite of elements necessary to meet the U.S. and European requirements, measured in a single run for duplicate samples. Table 6. Results for Duplicate ICP-OES Measurement of the Full Suite of Elements in Biodiesel. Sample 1 Sample 1 Dup Relative Percent Element (mg/kg) (mg/kg) Difference S 180.669 0.64 0.60 9 P 178.221 6.9 6.8 1 Na 589.592 3.0 2.7 10 K 766.490 < DL < DL Ca 317.029 7.7 7.7 0.6 Mg 285.213 3.4 3.6 5 The agreement between two duplicate sample preparations is good, shown in the relative percent difference calculations in the last column. Values that differ by less than 20% are usually acceptable. A variety of samples are shown in Table 7, showing biodiesel from two types of starting material, soybeans for the commercial biodiesel and used frying oil for the batch made at a nearby university. The blend combines commercial biodiesel with petroleum-based diesel fuel in a blend of 20% biodiesel. Table 7. Analisis Results for Several Biodiesel Samples and a Blend. commercial Academic Academic Element Biodiesel Batch Batch-washed B20 Blend S 180.669 0.8 0.9 0.9 4.44 P 178.221 7.6 1.5 1.2 1.6 Na 589.592 1.9 23.0 18.1 0.58 K 766.490 < DL < DL < DL < DL Ca 317.933 7.4 0.7 < DL 1.54 Mg 285.213 2.8 < DL < DL 0.63 The values measured for Na and K would add up to 1.9 mg/kg for the commercial biodiesel, within the acceptable limit for the ASTM and EU standards. Sodium in the academic biodiesel exceeds the limit specified of a total of 5 mg/kg for Na + K. The washed version has lower sodium, but may require additional washing to reduce the sodium content below the value specified in the standard. The measurement of Ca + Mg in the commercial sample adds up to 10.2 mg/kg, exceeding the allowable limit of 5 mg/kg. This biodiesel may be a sample taken prior to washing the final product. Table 8 shows the results for sample spikes to further ensure the method is operating properly. The samples were spiked at two levels, 5 mg/kg and 20 mg/kg. Recoveries are generally acceptable if they fall within 80-120% recovery. The results are excellent for both starting materials and the different spiking concentrations. Table 8. Spike Recoveries at Two Concentrations. commercial Commercial Academic Biodiesel Biodiesel Batch (5 ppm) (20 ppm) (20 ppm) S 180.669 98 90.6 88.6 P 178.221 94.5 108 102 Na 589.592 111 107 104 K 766.490 * 113 109 Ca 317.933 91.7 101 95.4 Mg 285.213 99.4 104 101 * K inadvertently omitted from the spiking standard. 4

Conclusion ICP-OES is shown to measure sulfur in biodiesel fuel with good accuracy and precision in standard reference materials and in samples. ICP-OES is a multielement technique that can measure sulfur and other elements together in one run for an economical analysis, about 2 minutes per sample. The instrument detection capability is sufficient for the regulatory limits currently in place and lower limits that might be required in the future. The long linear range of the instrument allows for a wide range of concentrations to be measured, which may differ through out the process from raw material to finished product. References 1. EN14214 Automotive Fuels Fatty Acid Methyl Esters (FAME) for Diesel Engines Requirements and Test Methods, available from www.bsi-global.com 2. ASTM 6751-07b Standard Specification for Biodiesel Fuel Blend Stock (B100 for Middle Distillate Fuels, available from www.astm.org 3. D7151-05 Standard Test Method for Determination of Elements in Insulating Oils by Inductively Coupled Plasma Atomic Emission Spectrometry (ICP-AES), available from www.astm.org 4. D7111-05 Standard Test Method for Determination of Trace Elements in Middle Distillate Fuels by Inductively Coupled Plasma Atomic Emission Spectrometry (ICP-AES), available from www.astm.org 5. D5185-05 Standard Test Method for Determination of Additive Elements, Wear Metals, and Contaminants in Used Lubricating Oils and Determination of Selected Elements in Base Oils by Inductively Coupled Plasma Atomic Emission Spectrometry (ICP-AES), available from www.astm.org 6. D4951-06 Standard Test Method for Determination of Additive Elements in Lubricating Oils by Inductively Coupled Plasma Atomic Emission Spectrometry, available from www.astm.org Acknowledgement Thank you to the student affiliates of the ACS at Roger Williams University, Rhode Island, for the sample of biodiesel. PerkinElmer, Inc. 940 Winter Street Waltham, MA 02451 USA P: (800) 762-4000 or (+1) 203-925-4602 www.perkinelmer.com For a complete listing of our global offices, visit www.perkinelmer.com/contactus Copyright 2009, PerkinElmer, Inc. All rights reserved. PerkinElmer is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners. 008591_01