Brandon Coolview XM Models Installation & Instructions for Use Brandon Medical Company Ltd Holme Well Road Leeds LS10 4TQ England Tel. +44 (0)113 277 7393 Fax. +44 (0)113 272 8844 E-mail enquires@brandon-medical.com Web WWW.brandon-medical.com Issue: A1 Print Date: 24-Sep-04
CVXM-C Ceiling Mounted Coolview XM. Caution: Only to be installed by qualified personnel. Check that the supply voltage matches the voltage shown on the rating plate. Ensure that the building mounting is secure and rigid. 1. Drill 4 x 10mm holes to match the attached template. Where fixing to the concrete sofit, drill the holes to a depth of 75mm and blow out the dust. Attach the ceiling base plate assembly to the building using the four anchor bolts or suitable alternative. 2. On a flat and level surface prepare to fit the horizontal arm to the end of the stem column as shown in Fig 2. Pull the cable through the stem column. Insert the sliding key into the groove (see detailed view Fig 2.1). Slide the horizontal arm into the end of the stem column and lock the sliding key in place with the grub screw provided. Fully tighten the grub screw and then slacken it by quarter of a turn. 3. Place the O ring and the ceiling cover over the other end of the ceiling stem. Insert the stem tube into the ceiling base plate assembly and fix
it temporarily by tightening the four bolts. Align the stem column to ensure that the stop points away from the main direction of usage. Tighten the four bolts to secure the stem column into the ceiling base plate assembly. Connect the mains supply in accordance with the wiring diagram. Push the cover and O ring up to cover the ceiling base plate assembly. 4. Now complete the head assembly section. CVXM-W Wall Mounted Coolview XM. Same procedure as above apart from section 3:
3. Place the O ring and cover over the wall mounting tube. Insert the tube into the wall base plate assembly and fit the locking screw (Figure 3 - Part 5) from below. Tighten the four bolts to secure the wall mounting tube into the wall base plate assembly. Connect the mains supply in accordance with the wiring diagram. Push the cover and O ring up to cover the ceiling base plate assembly. 4. Now complete the head assembly section. CVXM-M Mobile Coolview XM.
1. Insert the column 1 st section into the castor base and align it so that the hole for the cable holder faces backwards (See Figure 4). Tighten the four bolts to secure the column 1 st section into the castor base assembly. 2. Fit the spring mounted arm and column 2 nd section onto the column 1 st section. Fit the cable holder and tighten the fixing screw to secure the cable holder and column. 3. Now complete the head assembly section. Coolview XM Head Assembly.
1. Connect the spade terminals, as shown in Figure 5. Align the lamp head interface with the end of the spring arm as shown in Figure 6. Tighten the srews to secure the lamp head to the end of the spring arm.
Coolview XM Care and Maintenance. Changing the Bulb. The Coolview XM uses a specialist bulb only replace with original parts to prevent damage to the luminaire. Let the luminaire cool before attempting to change the bulb. Lift the protective cover and remove the allen key which is stored under the cover near the hinge. Remove the 3 screws that secure the cover ring and reflector. Remove the reflector to allow access to the bulb. Pull the old bulb out and discard. Fit a replacement bulb, using a soft cloth to ensure that you do not touch the glass envelope. Ensure that the bulb is pushed securely home into the bulb holder and re-assemble.
Adjustment of spring arm balance. The weight compensation of the spring arm will have been factory set. However transportation, use and the passage of time can allow the spring to settle in which case it will require the spring balance to be adjusted. Should this become necessary, fist remove the access cover on the underside of the spring arm. See Figure 6. Move the arm until the capstan wheel (round nut with holes in its periphery) appears in the window. Use a 4mm bar, screw or hex driver to turn the capstan wheel to increase or decrease the spring tension. Cleaning and Disinfection. The Coolview XM should be cleaned with a moistened cloth using one of the following cleansers: Dismozon, Lysoformin, Hexaquart plus or Sagrotan rapid disinfectant cleanser.