AXD 530 MicroManometer

Similar documents
AXD 510 MicroManometer

Thermo-Anenometer Model Owner s Manual

MICROMANOMETER ALNOR MODEL AXD610 AIRFLOW INSTRUMENTS MODEL PVM610

Ventilation Testing. Micromanometer. Alnor Model AXD610 AIRFLOW TM Model PVM610. Operation and Service Manual

IAQ-CALC INDOOR AIR QUALITY METER MODEL 7515

Indoor Air Quality Meter

OWNER S MANUAL. Model 9880 Intrinsically Safe Air Velocity Meter

AIR VELOCITY METER ALNOR MODEL AVM410 AIRFLOW INSTRUMENTS MODEL TA410

DP-CALC TM Micromanometer

IAQ-CALC TM Indoor Air Quality Meter

IAQ-CALC TM Indoor Air Quality Meter

Model 8340 Intrinsically Safe VELOCICHECK Air Velocity Meter

VELOCICALC Air Velocity Meter

Rotating Vane Anemometer

Rotating Vane Anemometer

User's Manual. Model Hot Wire Thermo-Anemometer. Introduction

User's Guide. Vane Anemometer. Model Measures Air Velocity & Temperature. Degrees C/F switchable

Important Operating Instructions and Warranty Information On Your New Electronic AMSEC Safe

Exhaust Fan Flow Meter. Operation Manual. The ENERGY CONSERVATORY DIAGNOSTIC TOOLS TO MEASURE BUILDING PERFORMANCE

PVI 1800/PVI Residential/Commercial Grid-Tied Photovoltaic Inverter WARRANTY MANUAL. Subject to Change REV , Solectria Renewables

HTR-202 INSTRUCTION MANUAL. Portable Digital Height Rod. Befour, Inc. 102 Progress Drive Saukville, WI 53080

CM100. Carbon Monoxide Meter. 99 Washington Street Melrose, MA Phone Toll Free

SPC Series. Digital Scale. Operation Manual

INTRODUCTION. Backlight for operating in dark area Low battery indicator Auto power off function. CO levels and its effect

DAFM4. Digital Air Flow Meter INSTRUCTION MANUAL Fax: (503) ENGLISH DAFM4 VOL CFM

Refinery Supply Company, Inc.

Effective June 1, 2013 This guide supersedes all previous versions

PVI 60KW, PVI 82KW, PVI 95KW

FOR YOUR PERMANENT FILE

Model 3400KL / 3401KL / 3400KG / 3400KG-C Digital Wet Diaper / Lap Sponge / Organ Scale

Power. On Your Terms.

BEAMER Tilt Beam Sensor

Model 8345/8346/8347/8347A VELOCICALC Air Velocity Meters

User's Guide. EzFlex Thermo-Anemometer. Model EZ30 UNITS MAX MIN AVG VEL

FOR YOUR PERMANENT FILE

CTFRP Series Power Supplies

DT304. Digital Temperature Logger INSTRUCTION MANUAL

DMG100 Digital Micron Gauge Limited Warranty DMG100 DMG100 Digital Micron Gauge USA CANADA COPYRIGHT 2011 UEi

Compact Scales. Software Revision V1.25 & above

Instruction Manual AVTM for. Strip Chart Recorder Catalog Nos and

Series 2100 Operating Instructions

WARRANTY POLICY. Grid-Tied Photovoltaic Inverters. Revision D. 2014, Solectria Renewables, LLC DOCIN

SKY-WEIGH HELICOPTER CARGO HOOK LOAD MONITORING SYSTEM. User Guide Quality Industrial Weighing and Force Measurement Equipment

EXHAUST FAN FLOW METER

Intelligent Electronic Safe

Reach ins, Freeezers & Refrigerators Installation & Operation Manual

10 Year Limited Warranty

Megohmmeter Test Probe

DUSTTRAK AEROSOL MONITOR SOLAR POWER KIT MODEL

Harvil 5 Foot Air Hockey Table

User's Manual. Earth Ground Resistance Tester. Model Introduction

Care and use guide. Contents. page

Manual No: Revision: F. Mag Probe Assembly Manual

Larson Systems Inc. V200, V750 and V1500 Digital Valve Spring Tester User Manual. Current for Software Version 6.9

Harvil 4 Foot Air Hockey Table

XPC-EBP64 External Battery Pack User & Installation Manual Xtreme Power Conversion Corporation. All rights reserved.


DPM1B. Operators Manual. Pneumatic Transducer Tester

9' & 12' CHALLENGER SHUFFLEBOARD TABLE ASSEMBLY INSTRUCTIONS

AC/DC Powered Peak Detection Fast Charger for 4- to 8-Cell Ni-Cd & Ni-MH Batteries

Smart-UPS RC External Battery Pack Tower/Rack-Mount 4U

Smart-UPS RT External Battery Pack Stack/Rack-Mount 6U

INSTRUCTION MANUAL DT173 DT175 DIGITAL THERMOMETERS ALWAYS READ THESE INSTRUCTIONS BEFORE PROCEEDING

Phase Sequence Tester

OPERATIONS MANUAL LEVER CHAIN HOIST

Barnant Company (Type J) (Type K) (Type T)

MODEL MVX-2011 TANK MOUNT SPEEDOMETER/TACHOMETER

HDI inch Choke Position Indicator User s Manual Rev Indicator Gauge Linear Potentiometer

Operating Instructions - Revision 4.1 January 2016 RJ11 PORT

User's Manual. High Voltage Megohmmeter. Analog Insulation Tester plus AC Voltage and Continuity Tests. Model

User's Guide. Model RPM10 Laser Photo / Contact Tachometer with IR Thermometer. Introduction

DMG200. Digital Micron Gauge. 99 Washington Street Melrose, MA Phone Toll Free

Users Manual. Airflow Meter

Tachometer. English. Users Manual

Model DS7060 Handrail Scale. User Manual

Product Parts Information

Model P772 Conductivity Meter. User Manual

Harvil 4 Foot Air Hockey Table

User's Guide. CO 2 Meter. Model CO250

INSTALLATION & OPERATING INSTRUCTIONS: REVOLUTION SPINEBOARD ATTACHMENT WARNING

BroadBand PowerShield. 20 AHr Battery. User Manual

Model Battery Powered Milliohm Meter

Right Weigh. Load Scales. Right Weigh Digital Load Scale. Model 201-EDG-01(B) Installation and Operation Manual

AD-7 THERMAL INDICATING AMMETER

Cable Length Meter INSTRUCTION MANUAL

Thermistor Thermometer MODEL NO

Digital/Analog Megohmmeter Model 1026 USER MANUAL

3000 SERIES DRY-BLOCK HEAT SOURCE

Thermocouple Thermometer Models

Blue Air. Commercial Refrigeration Inc. Installation & Operation Manual Chef Bases

Owner s Manual & Safety Instructions

Dimensions 12/800N 12/1200N D. DC to AC Power Inverters. OWNERS MANUAL for Models: OWNERS MANUAL April ISO 9001:2000 Certified Company

Mechanical Spiral Sensor

6.0 SPECIFICATIONS CONTENTS. Calibration. According to factory procedureeeeeeeeeeeeeee Accuracy*

Exterior Digital Load Scale 201-EDG-01(B) Installation and Operation Manual Please read carefully before installation

Where to Find Us. Pressure Washer. Model Number Revision Serial Number. Date Purchased

AD-7. Instruction Manual. Thermal Indicating Ammeter HD ELECTRIC COMPANY 1475 LAKESIDE DRIVE WAUKEGAN, ILLINOIS U.S.A.

INSTALL GUIDE Dodge/RAM 5.7L HEMI

DC to AC Power Inverters

THWD-5. Relative Humidity Temperature Meter. Users Manual. For detailed specifications and ordering info go to

Transcription:

OWNER S MANUAL AXD 530 MicroManometer A TSI Company

Table of Contents Features 2 Using the AXD530 Safely 3 Getting Started 4 Installing Batteries 4 Preparing the Instrument 4 Attaching the Optional Pitot-Static Probe 4 Start-Up Sequence 5 Zeroing the Instrument 5 The Display 7 Modes 7 AXD530 Keypad 8 The AXD530 Keypad 8 Key Functions 8 Using the AXD530 9 Basic Functions 9 To Change Units 9 Measurement Modes 9 Pressure Selection 9 Velocity Selection 10 Maintenance & Troubleshooting 11 Repair Information 13 Instructions for Return 14 Ownership/Calibration Log 16 1

Features The AXD530 MicroManometer is a meter that measures differential pressure. It is also capable of measuring and displaying air velocity. Functions are selected through a 3-key pad. The AXD530 displays measurement results on a liquid-crystal display (LCD), with additional display segments for units of measure and low battery indication. It features two built-in pressure ports for attachment to the pressure source you will measure. Optional hoses and Pitot-static probes are available. The AXD530 uses three (3) AA size alkaline batteries for low replacement cost. 2

! WARNING NOTICE Using the AXD530 Safely When using the instrument to check air flow in an elevated workplace, make certain that you can safely raise and hold the instrument while making measurements. This is especially important when you are working on a ladder. Avoid catching hoses or attachments in moving machinery. Use the instrument only for measurements using air. Avoid any corrosive or other dangerous or explosive gas mixtures. Please dispose of used batteries in a responsible manner. 3

Getting Started Installing Batteries The AXD530 uses three (3) AA size non-rechargeable batteries. The unit was shipped with batteries not installed. You will find the batteries in the instrument package. To install the batteries: Back view of instrument LO BATT Low battery displays Loosen the battery cover screw on the rear of the instrument. Slide the cover up and lift it out. Install the batteries over the pull strap, taking care to observe each battery s polarity. Replace the cover and tighten the screws. Note: When the batteries become low, the LCD will show [LO BATT] above the reading. From that point, you will have about one hour of normal use left. If [ ] comes on continuously and replaces the reading, the battery voltage is too low for accurate measurements and the meter will turn itself off. Always keep a replacement set of new batteries available. Preparing the Instrument Attach hoses to the pressure ports at the top of the instrument. Do not exceed maximum pressure. Attaching the Optional Pitot-Static Probe To attach the probe for measuring air velocity: Remove the Pitot-static probe and rubber tubing from their carrying case. Remove the protective caps from the probe and save them for re-storing the probe later. 4

Pitot-static probe TOTAL PRESSURE HOLE STATIC PRESSURE HOLES STATIC PRESSURE PORT (TO - INPUT ON MANOMETER) TOTAL PRESSURE PORT (TO + INPUT ON MANOMETER) Attach one section of flexible tubing to the probe s static pressure port, and another to the probe s total pressure port. Identify the (+) and (-) pressure ports at the top of the AXD530. Attach the tubing from the total pressure port of the Pitot-static probe to the (+) port of the MicroManometer. Attach the tubing from the static pressure port of the Pitot-static probe to the (-) port of the MicroManometer. Check to ensure that all tubing connections are tight. Start-Up Sequence Each time you turn the instrument on, some basic functions are tested. The following will occur: All the LCD segments will turn on. The current software revision level will be displayed. The message [ ] will appear. The last unit of measurement will be displayed along with a measurement value. [ ] followed by [ ] will appear. The meter will begin to measure. Zeroing the Instrument Before you begin taking measurements, the AXD530 must be zeroed by following these steps: Most hand-held meters using differential pressure sensors require a warm-up time. Turn on your Alnor AXD530 a minimum of 5 minutes before zeroing and taking your first measurements. 5

Make sure there is no air flow past a Pitot-static probe (also verify that none of the tubing is pressurized by accidental squeezing). Press the ZERO key until [ ] appears on the display. The instrument will automatically zero its readings. Repeat this procedure if necessary, because at high resolutions the AXD530 will read small pressure changes. 6

The Display The AXD530 uses a 4-digit, 7-segment LCD. As a result, measurements exceeding 9,999 of a given unit of measure will be displayed in a scientific notation format of XXEY. The actual reading will be rounded to the nearest thousand, XX. The Y indicates the number of zeros after XX. For example, 12,664 is displayed as [ ]. Modes The AXD530 operates and takes measurements in the RUN mode. You may be in the RUN, ZERO, or UNIT select modes. 7

AXD530 Keypad The AXD530 Keypad The internal clock rate of the AXD530 software may cause the meter to react slowly to a key push. The reaction time of a key push will vary with different functions (1 3 seconds). Maintain pressure on the desired key until you see the required response on the display. Key Functions: ON OFF ZERO UNIT ON/OFF This key is used to turn the instrument on or off. ON OFF ZERO UNIT ZERO The AXD530 will zero the instrument any time this key is pushed. To get a proper zero, make sure no pressure difference is applied across the ports and no kinks are in the hoses. (See also Zeroing the Instrument) ON OFF ZERO UNIT UNIT This key is used to select the desired unit of measurement. 8

Pa inh 2 0 m/s fpm Display with annunciator and units of measure. Using the AXD530 Basic Functions To Change Units: Push and hold the UNIT key. The display annunciator will move to the left of the LCD beginning from the right. Release the UNIT key at the desired location. For example: To change the units of measurement from feet per minute (fpm) to meters per second (m/s): Push the UNIT key. The display annunciator moves to m/s, selecting meters per seconds. Release the key. Measurement Modes You can set the AXD530 for pressure or velocity measurement by pressing the UNIT key. Pressure Selection: The AXD530 is a pressure measuring instrument. Its factory default setting is to measure pressure in inches of water (in.h 2 O) or pascals (Pa). 9

Velocity Selection: Air velocity measurement can be accomplished using a Pitot-static probe. The following equations are used for pressure to velocity conversions: Velocity (fpm) = 4005 Pressure where Pressure is in inches H 2 0 or Velocity (m/s) = 1.29 Pressure where Pressure is in Pa The velocity calculations assume standard conditions. Use Appendix D to determine the correct multiplier if conditions other than standard exist. Refer to Appendix B for pressure to velocity conversion. 10

Maintenance & Troubleshooting Maintenance Periodic maintenance should be performed on the meter. Batteries should be replaced when the low battery indication is on or when the unit turns itself off after power-up. Batteries should be removed from the meter if it will not be used for two (2) months or longer. Calibration checks are recommended every 12 months. Keep the meter clean by using a mild detergent on its case. DO NOT use abrasives or solvents. Also, DO NOT allow liquids to enter the meter s case. Dry thoroughly after cleaning. Use an eyeglass cleaner and lens paper to clean the LCD window. 11

Troubleshooting Symptom Check DISPLAY SHOWS [ ] The measured value is beyond the instrument s range. Be sure there is no applied pressure, then press ZERO. DISPLAY SHOWS DISPLAY DOES NOT CHANGE METER TURNS OFF MEASUREMENT IS ERRATIC ERRONEOUS READINGS The internal connection between the meter and [ ] the sensor is faulty. Turn the meter off, then turn it back on again. Press the ON/OFF key until the unit turns off. Then turn it on and try again. Your input might simply be very stable. Test your meter by pinching the hoses. The batteries are too low for your completing start-up meter start. Watch for the [ ] cycle indication just before the meter turns off, then replace the batteries. The input is not steady. Check to see that your Pitot-static probe is held steady. Make sure that your Pitot-static probe is perpendicular to the air flow you are measuring. Clean your Pitot-static probe it may be dirty. Take measurements at a different location. If any of these problems persist, call the factory for assistance. 12

Repair Information Contact the distributor in your country, or Alnor Instrument Company directly, before returning your instrument. See INSTRUCTIONS FOR RETURN. Follow the procedure carefully as it will expedite processing. Failure to follow the procedure may cause return of the unit unrepaired. Send your instrument to the factory transportation prepaid. To assure fast turn-around time, photocopy and fill out this form with as much detail as possible and attach it to the instrument. Instrument Model Serial Number Date of Purchase Where Purchased Describe Malfunction Describe Environment Return Instrument to: Name (Your name or company) Address Telephone Address Correspondence to: Name Address Telephone 13

Instructions for Return Damaged in Transit All shipments are carefully examined by Alnor Quality Assurance and carefully packed for shipment. On receipt, if the shipping container appears to have been damaged during shipment, the instrument should be thoroughly inspected. The delivering carrier s papers should be signed noting the apparent damage. If the instrument itself has been damaged, a claim should be promptly filed against the carrier by the customer. The selling agent will assist the customer by supplying all pertinent shipping information; however, the claim must be filed by the insured. If the instrument is damaged beyond use, a new order should be placed with Alnor while awaiting reimbursement from the carrier for the damaged instrument. Call Alnor directly for assistance if necessary. Repair/Calibration Please follow these steps should you require factory service or repair of your Alnor Instrument: 1. Contact the Alnor distributor in your country for cost of repair or recalibration and shipping instructions. Obtain a Purchase Order or purchase order number from your Purchasing Department showing instrument model number and cost of repair and/or recalibration. Securely package your instrument in a strong container surrounded by at least two inches (5 cm) of suitable shock-absorbing material. Include the Purchase Order, or Reference the Purchase Order on your packing slip. 2. Contact the Customer Service at Alnor for cost of repair or recalibration, Return Material Authorization (RMA) number, and shipping instructions. On instruments being returned to the factory for recalibration, determine whether you prefer: 14

Certificate of Calibration, no data (states instrument checked and found to be within accuracy claimed and specifications listed in current literature). Certificate of Traceability to National Institute of Standards and Technology (NIST, formerly NBS) with data (states instrument compared with factory standard traceable to NIST, lists of calibration data, i.e., Standards reads... and This instrument reads... ) Obtain a Purchase Order or purchase order number from your Purchasing Department showing instrument model number and cost of repair and/or recalibration. Securely package your instrument in a strong container surrounded by at least two inches (5 cm) of suitable shock-absorbing material. Include the Purchase Order, or Reference the Purchase Order on your packing slip. Mark the outside of the shipping container with the RMA Control number. 3. Forward the instrument prepaid. See back cover for factory addresses. Please note that instruments received improperly marked or without an accompanying Purchase Order may be returned at your expense. Please also note that if an RMA number has been assigned, it will be canceled if the instrument is not received by ALNOR within 60 days. 15

Ownership/Calibration Log Equipment Log Date of Purchase Calibration Record 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. Notes: 16

Warranty Alnor Instrument Company (Alnor) warrants this product to be free of defects in material and workmanship for a period of two years from the date of original purchase. If the product should become defective during the warranty period, Alnor will repair it or elect to replace it free of charge under the following conditions: 1. Product is returned postpaid per the instructions for return, located in the owner s manual. 2. Owner submits proof of original date of purchase. 3. Alnor will inspect product for defects in material and workmanship. Alnor s decision as to existence of defect, and in the case of defect, to repair or replace will be final. THIS WARRANTY IS EXCLUSIVE. THERE ARE NO OTHER WARRANTIES EITHER EXPRESSED OR IM- PLIED. SPECIFICALLY AND WITHOUT LIMITATION, THERE IS NO WARRANTY, EXPRESSED OR IMPLIED, OF MERCHANTABILITY OR FITNESS FOR PARTICU- LAR PURPOSE. This warranty is void if product is misused, used contrary to procedures set forth in the owner s manual, or if product is serviced by anyone other than Alnor s authorized service. This warranty does not cover consumables such as light bulbs, paper, batteries, etc. Alnor s liability for this product is limited to the above stated warranty and shall not in any event exceed the cost of the product. In no event will Alnor be liable for any direct or consequential damages, including but not limited to lost profits, loss of use, inaccuracies, loss of data, dismantling or reinstallation. ALNOR INSTRUMENT COMPANY 7555 North Linder Avenue September 1996 Skokie, Illinois, 60077 USA Printed in USA Toll-Free (800) 424-7427 Part No. 116-159-151 Rev. 4 Telephone (847) 677-3500 Copyright 1996 Alnor Instrument Company Fax (847) 677-3539 17