Nickel Catalyst Tube on a 6820 GC, Accessory G4318A

Similar documents
Valve Box and Valve Accessories G4315A, G4316A, G4317A

Agilent G1554A On-Column Capillary Inlet with EPC

Nickel Catalyst Tube Accessory 19205A

Agilent G2738A Upstream Capillary Interface

Installing a Purged Packed (PP) Inlet

Installing a PTV Inlet

Installing a Purged Packed Inlet

Installing a Cool On-Column Inlet

Installing the Flame Ionization Detector EPC Flow Control Manifold

Installation Guide Thermal Conductivity Detector Accessory 19232E

Installing a PTV Inlet

Installation Guide. Flame Ionization Detector on a 6890 GC. Accessories G1561A, G1562A, G1591A, G1598A

Installation Guide. Flame Ionization Detector on a 6850 GC Accessory G2621

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

Installation Guide. Nitrogen-Phosphorus Detector on 6890 GC Accessories G1575A, G1576A, G1594A, G1599A

Installation Guide. Purged Packed Inlet, Electronic Pneumatics Control on 6890 GC Accessory G1551A

Installing the 6850 Direct Valve Column Connector Kit

Installation Guide. Thermal Conductivity Detector on a 6850 GC Accessory G2623A

Installing an Auxiliary Pressure Control Module

Agilent N2763A Rack Mount Kit for 4000 X-Series Oscilloscopes

Installing the CO 2 Cryogenic Oven Cooling Kit

260 Solvent Vapor Exit Accessory

Installation Guide. Auxiliary Pressure Control Manifold on 6890 GC Accessory G1570A

Installation Guide. Heater Interface Cable for Mass Sensitive Detector Accessory G1590A

Installing an Auxiliary Pressure Control Module

Agilent G3510A/G3511A Multimode Inlet

Agilent N5470A Series Oscilloscope Rack Mount Kit

Installation Guide. Volatiles Interface for 6890 Gas Chromatograph Accessory G2319A

29 Installation Step 1. Unpacking the GC Cable diagrams Step 2. Placing the GC system on the benchtop Step 3. Turning the power on

Agilent 6890 Gas Chromatograph

Installation Guide NPD/ECD Chemical Filters Accessory 19199N

Hot Injection Adapter Kit - Installation Guide

Alcatel-Lucent 7310 DC Metro Cell AWS (B4) Kit MC-DBR1, MC-DBR2, MC-DBR3, MC-DBR4, MC-DBR5. Installation Manual

Agilent 490 Micro GC Internal and External Mount Genie Filter Brackets

Deans Switch Microfluidics

Installing the G2743A/G2745A Valve Actuator Kits

AssayMAP 96AM Wash Station. Maintenance Guide

6890 Installing ¼ Glass Columns

Agilent G2647A (Single) and G2648A (Dual) Flame Photometric Detector Update Kits

Agilent 1200 Series Manual Injector

Agilent 1260 Infinity Manual Injector

Agilent G3348B Flame Photometric Detector

In addition to this instruction sheet, the following are included in this kit (some additional parts are included for future maintenance):

RIGID ENTRY BOOT INSTALLATION GUIDE

1350 Valves. 1 of 14. Jun Illustrated Parts Breakdown Agilent 6890 Gas Chromatograph Service Manual

FLEX-ING TM DISPENSER HOSES INSTALLATION GUIDE

Maintenance of the Electronic ( EPC ) Pressure Controllers on SRI GCs

Agilent G2333B (Single) and G2334B (Dual) Flame Photometric Detectors

NPD to FID conversion on SRI 8610C or 310C GC July2012

FLEX CONNECTOR HEAT SHRINK SLEEVE INSTALLATION GUIDE

Installation instructions

Agilent 210/218 - Isocratic Pump

EVO TM 200 AND EVO TM 400 LCD DISPLAY INSTALLATION GUIDE

INSTALLATION INSTRUCTIONS FOR PCBM-36C CONVERSION KIT From Natural Gas to Propane/LP Gas

Dispenser Pan Sensors & Containment Sump Sensors

Installation Note for Solvent Selection Valve Upgrade Kit (G1360A)

Agilent 7820A Gas Chromatograph

5000TOC Sensor Service Manual

A g i l e n t A G a s C h r o m a t o g r a p h Service Manual

Installation instructions

Installation instructions

Prodigy HDLV Generation II Pump Panel

AssayMAP 96AM Wash Station. Maintenance Guide

ANTI-SYPHON VALVE INSTALLATION GUIDE

Installation instructions

SEALING DIP CAP INSTALLATION GUIDE

Agilent 210/218 - Binary Pump

Installation instructions

Installation instructions

Installation Instructions

Agilent 5975T. Mobile Lab Hardware Installation. Agilent Technologies

BEFORE USING THE PRODUCT

Lamp House with 2.5 kw lamp

DIESEL EXHAUST FLUID PUMPING ACCESSORIES INSTALLATION GUIDE

Installation instructions

Prodigy HDLV Pump. Customer Product Manual Part A02

ASSEMBLY and OPERATING INSTRUCTIONS Mission Oaks Blvd. / Camarillo, CA 93011

Installation instructions

Installation instructions

Agilent Valve Kit G4231A/B and G4232A/B/C

DISPLACEMENT PUMP INSTRUCTIONS-PARTS LIST Rev. K. Model , Series A Model , Series B Model , Series A

Passive Check Valve Installation Instructions for H-Class Systems / Revision A Copyright Waters Corporation 2014 All rights reserved

Installation Operation Maintenance ADDENDUM

Installation instructions

Dispenser Pan Sensors & Containment Sump Sensors

Installation instructions

Installation instructions

Installation instructions

Installation instructions

Model 4430 Photoionization Detector (PID) Operator s Manual

3-PHASE SMART CONTROLLER STP-SCIIIC INSTALLATION GUIDE

KIT CONTENTS. Multimeter or test light Wire crimpers or soldering equipment. Wire terminals & connectors 10A inline fuse

Agilent SD-1 Purification System ml/min

PMA INTAKE FILTER SCREEN INSTALLATION GUIDE

Agilent 1100 Series Thermostatted Column Compartment. Reference Manual

STL Mounting Brackets

RENA AF371Feeder Operating Manual. Feeder. Operating Manual. Manual Part #: M AF371 Operations Rev

Disassembly and Assembly

Installation instructions

Model Combustible Gas Sample Draw Detector Head Operator s Manual

Transcription:

Nickel Catalyst Tube on a 6820 GC, Accessory G4318A Installation Guide The Nickel Catalyst tube (NCT) is used with the 6820 Gas Chromatograph (GC) for trace analysis of CO and CO 2 using a Flame Ionization Detector (FID). The gas sample is separated on a column and passed over a hot catalyst which, in the presence of hydrogen, converts any CO and/or CO 2 present to methane. Methane (CH 4 ) is readily detected by the FID. Before following this procedure, refer to the safety information at the end of this document. Sample Hydrogen Air Carrier gas Gas sample valve Column Nickel catalyst FID Requirements The accessory is intended to be installed in the back inlet position. If you remove an existing back inlet system to make space, be sure to cap its carrier gas supply. Two nut plates, shown installed inside the oven in this procedure, are provided separately with Heated Valve Box kits. Agilent Technologies

Parts list Nickel catalyst tube (NCT) assembly Porapak Q column Inlet tube with bulkhead fitting Torx screw for attaching flow manifold to the pneumatics carrier Insulation cup and insulation Top cover and insulation Torx screws, swaging nuts and ferrules, and other hardware Tools Configuration considerations Steps Electrostatic protection such as a grounded wrist strap 7-mm nut driver T-20 Torx screwdriver Suitable tool to properly cut 1/16-inch stainless steel tubing In installing the NCT, to achieve a standard factory configuration, it may be necessary to move or remove an already existing inlet. For a two-inlet configuration, the factory placement rule is as follows: A Purged Packed inlet is always placed in the front location. 1 Preparing the GC 2 Installing the NCT 3 Connecting the NCT hydrogen supply tube 4 Restoring the GC to operating condition 2 Installation Guide: Nickel Catalyst Tube

Preparing the GC WARNING Hydrogen (H 2 ) is both flammable and an explosion hazard when mixed with air in an enclosed space (for example, the oven). In any application using H 2, switch off the supply at its source before working on the instrument. WARNING Hazardous voltages are present in the mainframe when the GC is connected to electrical power. Avoid potentially dangerous shock hazards by disconnecting the power cord before removing side panels. 1 Switch off electrical power to the GC and disconnect the power cord. Allow time for the oven and heated zones to cool. Then switch off supply gases at their sources. 2 Remove column(s) and any associated hardware from inside the column oven. 3 Lift the hinged GC top cover at its front edge to expose the detector area. Remove the cover by raising it to vertical, lifting its left hinge pin from its bracket, and then sliding the cover to the left to free its right hinge pin. 4 Remove the inlet cover by removing its three screws. 5 If (and only if) an inlet must be removed from the back location, remove two screws along the lower edge of the left side (flow control) panel. Slide the panel slightly to the rear and lift it off. 6 In the same manner, remove two screws along the lower edge of the right side (electronics) panel. Slide the panel slightly to the rear and lift it off. CAUTION Electronic components can be damaged by static electricity; use a properly grounded static control wrist strap when removing the electronics panel. Installation Guide: Nickel Catalyst Tube 3

7 Remove the back panel by removing two screws at its bottom edge, loosening two screws at its top edge, and the lifting off the two screws. 8 Finally, remove the rear top cover by removing four screws, two at each end. 9 At this time, if necessary to achieve a standard factory configuration, completely remove the back inlet. Carefully store the removed inlet assembly and all associated parts in a safe place for possible future use. Installing the NCT These instructions assume you are installing the NCT in the back inlet location: CAUTION GC insulation is made of refractory ceramic fibers. Ventilate your work area. Wear long sleeves, gloves, safety glasses, and a disposable dust/mist respirator. Dispose of unneeded insulation in a sealed plastic bag. 1 If present, remove the cover plate from the back inlet mounting location and carefully remove pre-cut insulation below it. 4 Installation Guide: Nickel Catalyst Tube

2 Prepare the NCT assembly for installation by bending both its hydrogen supply tube and its exit tube along the side of the catalyst chamber. Hydrogen supply tube Heater/sensor cable Catalyst chamber Exit tube (to FID) Column fitting 3 While positioning the NCT into the cavity, use both supplied insulation and removed oven insulation to pack around the bottom and sides of the NCT. The goal is to fill all voids around the NCT body as it is placed into the cavity. 4 Verify the NCT assembly is installed with its column fitting and exit tube inside the oven. Avoid getting insulation inside the exit tube. Feed the heater/sensor cable out the top. 5 Remove the center plug from one of the supplied insulation rings. Place the insulation ring on top of the NCT. Install the cover plate on top of the ring and secure it with three screws. 6 Remove center plugs from the remaining supplied insulation rings. Place the rings inside the insulation cup. Inside the oven, install the cup over both the NCT exit tube and bottom fitting. Again, avoid getting insulation inside the exit tube. Secure the cup to the oven with two screws. 7 From the left side of the GC, connect the NCT heater/sensor plug to the square receptacle closest to the NCT ("BI" for back inlet location). Installation Guide: Nickel Catalyst Tube 5

8 Connect the NCT exit tube to the FID column inlet. Following proper swaging procedures described in the 6820 Getting Started manual, assemble a 1/8-inch stainless steel nut with back and front ferrules onto the tube. Insert the tube, ferrules, and nut into the FID base. 9 Tighten the nut finger-tight, then, using two wrenches against each other to avoid twisting the exit tube, tighten the nut an additional 1/2- to 3/4-turn. NCT exit tube Gas sample valve fitting Nut plate NCT inlet tube Nut plate Retaining nut Porapak Q column 10 Install the bulkhead fitting end of the NCT inlet tube to the nut plate on the right side of the oven. Secure with the retaining nut. 11 Connect the NCT inlet tube to the fitting on the bottom of the NCT. Again, following proper swaging procedures described in the 6820 Getting Started manual, assemble a 1/8-inch stainless steel nut with back and front ferrules onto the tube. Insert the tube, ferrules, and nut onto the NCT fitting. 12 Tighten the nut finger-tight, then, using two wrenches against each other to avoid twisting the inlet tube, tighten the nut an additional 1/2- to 3/4-turn. 6 Installation Guide: Nickel Catalyst Tube

Connecting the NCT Hydrogen Supply Tube 13 Install the Porapak Q column between the gas sample valve fitting on the inlet side of the oven and the NCT inlet tube on the detector side of the oven. WARNING Hydrogen (H 2 ) is both flammable and an explosion hazard when mixed with air in an enclosed space (for example, the oven). In any application using H 2, turn off the supply at its source before working on the instrument. 1 At the top of the oven, route the NCT hydrogen supply tube to the area behind the oven top. 2 Verify hydrogen is switched off at its source. Precisely identify the tube used for hydrogen. Starting from the selected FID flow control manifold block, look for a red tag on one of the 3 tubes connected to the block. The tag identifies the hydrogen tube for the given FID. Red tag and FID H 2 supply tube 3 Carefully follow the identified tube into the area behind the oven top and choose a location to cut the tube about midway along its length behind the oven top. Make sure there will be enough free tubing from the FID flow manifold block to easily join to the NCT hydrogen supply tube (step 1). Route the Installation Guide: Nickel Catalyst Tube 7

tubes as needed to be sure they can be joined after the cut is made. 1/16-inch blind union (hidden behind oven top) 1/16-inch union NCT H 2 supply tube FID H 2 supply tube NCT installed Red tag 4 Carefully cut the tube such that the end of each resulting portion is open, round, and free of burrs and surface scratches which might prevent proper flow and/or might create leak paths. 5 For the next three steps, if necessary, refer to your Getting Started manual for proper swage techniques. Also, in making the described connections, use two wrenches against each other to avoid twisting the tubing: a b For the tube portion leading to the FID, connect it with a blind" tube (plugged, part number 07680-81150) by a 1/16-inch union to completely seal the tube. For the tube portion leading from the FID flow manifold block, install a 1/16-inch swage nut, back ferrule, front ferrule, and normal union. c Finally, connect the NCT hydrogen supply tube to the union: install a 1/16-inch swage nut, back ferrule, and front ferrule onto the tube, then tighten the assembly into the union. 6 As needed, use cable ties to secure tubing to existing wiring and tubing running behind the oven top. 8 Installation Guide: Nickel Catalyst Tube

Restoring the GC to Operating Condition Repacking the Catalyst 1 Restore gas supplies and, following leak test procedures described in your Maintenance and Troubleshooting manual, check for leaks, particularly at all new connections. 2 Reinstall GC covers and panels. 3 Restore electrical power. 4 Switch off electrical power, then switch it on again to ensure the new configuration is properly retained in GC memory. 5 See your Operation manual with respect to properly reconfiguring the GC for the installed NCT. 6 Verify performance by following procedures described in your Maintenance and Service manual. The nickel catalyst material can be damaged by exposure to air. If conversion performance becomes significantly degraded, repack the catalyst tube following directions found in your Maintenance and Service manual. Installation Guide: Nickel Catalyst Tube 9

Agilent Technologies Notices Agilent Technologies, Inc. 2003 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws. Manual Part Number G4318-90000 Edition First edition, July 2003 Printed in China Agilent Technologies, Inc. 412 Ying Lun Road Waigaoqiao Free Trade Zone Shanghai, 200131 P. R. China Safety Notices CAUTION A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met. WARNING A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met. G4318-90000