product data sheet Usage: Duty cycle: Max 10% or 2 minutes continuous use followed by 18 minutes not in use Ambient temperature: +5 to + 40 C

Actuator LA44. Data sheet LINAK.COM/MEDLINE-CARELINE

ACTUATOR LA31 PRODUCT DATA SHEET

Actuator LA43. Data sheet LINAK.COM/MEDLINE-CARELINE

Actuator LA23. Data sheet LINAK.COM/MEDLINE-CARELINE

Actuator LA34. Data sheet LINAK.COM/MEDLINE-CARELINE

Actuator LA27. Data sheet LINAK.COM/MEDLINE-CARELINE

Actuator LA23 DESKLINE. Data sheet LINAK.COM/DESKLINE

Actuator LA32. Data sheet LINAK.COM/TECHLINE LINAK.COM/MEDLINE-CARELINE

ACTUATOR LA31 PRODUCT DATA SHEET

Actuator LA27 HOMELINE Analogue. Data sheet LINAK.COM/HOMELINE

Actuator LA30. Data sheet LINAK.COM/TECHLINE LINAK.COM/MEDLINE-CARELINE

Actuator LA40. Data sheet LINAK.COM/MEDLINE-CARELINE

ACTUATOR LA28 PRODUCT DATA SHEET

Actuator LA29. Data sheet LINAK.COM/HOMELINE

Actuator LA40. Data sheet LINAK.COM/MEDLINE-CARELINE

ACTUATOR LA44 PRODUCT DATA SHEET

Control Box CB12. Data sheet LINAK.COM/MEDLINE-CARELINE

Control Box CBJ1 / CBJ2. Data sheet LINAK.COM/MEDLINE-CARELINE

LIFTING COLUMN LC2 PRODUCT DATA SHEET

Battery Li-Ion BA21. Data sheet LINAK.COM/MEDLINE-CARELINE

ACTUATOR LA30 PRODUCT DATA SHEET

DESKLIFT TM DL6. Data sheet LINAK.COM/DESKLINE

ACTUATOR LA30 PRODUCT DATA SHEET

ACTUATOR LA30 WE IMPROVE YOUR LIFE. LA30 : 12/24V DC. Max. thrust 3500 N LA30L : 24V DC. Max. thrust 6000 N (LA30LK)

Actuator LA35. Data sheet LINAK.COM/TECHLINE

Kick & Click TM. Data sheet LINAK.COM/DESKLINE

Kick & Click TM. Data sheet LINAK.COM/DESKLINE

LIFTING COLUMN BL3 PRODUCT DATA SHEET

Actuator systems for hospital beds Perfection in movement LINAK.COM/MEDLINE-CARELINE

TA1. series. Comfort Motion Care Motion Auto Motion Ergo Motion. Product Segments

Actuator systems for patient lifts

Focus on marine LINAK.COM/TECHLINE

LIFTING COLUMN LC2 PRODUCT DATA SHEET

MOVE ON GO ELECTRIC! A Comparison of motion control systems: - Hydraulic - Electric

Actuator LD3. Product Data Sheet Feature

Better Solutions - Better Care. Product Overview LINAK.COM/MEDLINE-CARELINE

Focus on Ventilation and Farming Solutions LINAK.COM/TECHLINE

Actuator LA14. Data sheet LINAK.COM/TECHLINE

Components of an Electric Linear Actuator

Actuator ID10. Product Data Sheet Feature

* depending on drive configuration at constant 24/29 V DC without load

Actuator LA36. Data sheet LINAK.COM/TECHLINE

SERIES PA SWING ARM CLAMPS for the Automotive Industry

Preface... 4 Valid for... 5 Important information... 6 Safety instructions... 6 Manufacturer s declaration... 8

ACME SCREW LINEAR ACTUATOR OVERALL DIMENSIONS. Motor cable length 0.3 m. T [mm]

ACTUATORS POSITION SENSOR

BALL VALVE - 340BA SERIES Pneumatic actuator - El-O-Matic

guide unit, double acting Technical details

Kuhnke Technical Data. Contact Details

XG*-***-****-***-****

Linear actuator. Product features. Overview. Wide range of force outputs. Wide selection of column adapters. Wide selection of flanges

Series ZX. Cylinders > rodless cylinders > Technical details. Versions. Order code ZX- ** - ** - **** - **

ACME SCREW LINEAR ACTUATOR. OVERALL DIMENSIONS La =Lc + Stroke Lc T Capacitor

PNEUMATIC ACTUATOR FIG. PD - PE

PNEUMATIC ACTUATOR FIG. PD - PE

Compact cylinders series NYD Double acting with magnetic piston, ISO M5 to G1/8 Piston Ø 20 to 100 mm

BALL SCREW LINEAR ACTUATOR OVERALL DIMENSIONS. MASS [Kg] STROKE T STROKE LENGTH STROKE LENGTH. standard brake CODE

ACME SCREW LINEAR ACTUATOR OVERALL DIMENSIONS. Actuator with. Attachment A1. Lc [mm] Stroke Stroke T [mm] Stroke Stroke

INTEGRATED HYDRAULIC BRAKE ACTUATORS

Focus on. Mining Equipment.

bar -20 C C ( C for high temperature version) 1000 mm

BALL VALVE HIT SERIES Pneumatic actuator - El-O-Matic

Quarter turn actuators Index

CPLab / CP Factory. Application module magazine. Data sheets

0 = 0 mm rear cover panel edging (standard) Number = rear cover panel edging according to table

HEAVY LIFTERS FOR MATERIAL HANDLING

ISO Technical details

BUTTERFLY VALVE SERIES Pneumatic actuator - El-O-Matic

Data Sheet. AmLok.

C-SERIES S-SERIES. Metric Machine Screw Jacks EMA LINEAR ACTUATORS

AIR CYLINDER Series A12, A13

Product features & Overview. KA Linear actuators

Kuhnke Technical Data. Contact Details

BUTTERFLY VALVE 27 SERIES Pneumatic actuator - El-O-Matic


BALL VALVE SERIES Pneumatic actuator - El-O-Matic

BUTTERFLY VALVE - N1234 SERIES Pneumatic actuator - El-O-Matic

Pneumatic cylinders, piston-ø mm Double acting with magnetic piston DIN ISO 15552

P1E Series ISO Cylinders

END FITTINGS 126. Clevis M6 LOCKABLE GAS Cat.no: S1130 SPRING 130. Clevis M8 Cat.no: S Ballcup M6, Ø 10 Cat.no:

Mounting and operating instructions EB 5801 EN. Electric Actuators Type 5801 (Rotary Actuator) Type 5802 (Linear Actuator)

INTEGRATED HYDRAULIC BRAKE

Product features & Overview. N1 - N4 A Rotary and part-turn actuators

BALL SCREW LINEAR ACTUATOR OVERALL DIMENSIONS. Length [mm] Stroke Stroke LENGTH

FUSERBLOC Fuse combination switches with rear connection for industrial fuses up to 400 A

Pressure switches hex 24

Ergonomic Height Adjustable Systems

MINI-SOLENOID VALVES ISO (CNOMO, size 15) interface direct operated, pad mounting body connector size 15

bar -30 C C ( C for high temperature version)

LINEAR ACTUATORS INDEX

BALL VALVE SERIES Pneumatic actuator - El-O-Matic

Keystone F89C Pneumatic Quarter-Turn Actuator Chemical Industry / Hazardous area

PNEUMATIC ACTUATOR FIG. PD - PE

TECHLINE Solar Catalogue

Linear Actuator with Ball Screw Series OSP-E..S. Contents Description Overview Technical Data Dimensions 79

BUTTERFLY VALVE - 18/28 SERIES Pneumatic actuator - El-O-Matic

SERVOMECH linear actuators are electromechanical cylinders able to transform a rotary movement into a linear motion.

Die changing technique

Datasheet. Electrical Switching / Locking

Transcription:

Actuator LA31 Data sheet LINAK.COM/MEDLINE-CARELINE

LA31 The LA31 is a compact, quiet and powerful actuator designed for a variety of applications in the MEDLINE & CARELINE segment, such as hospital beds, couches and nursing home beds. The standard LA31 actuator features known parts such as piston rod eye with slot, and comes in different variations with e.g. fast motor, hall positioning and manual lowering. The LA31 actuator has exchangeable cables and is ideal in combination with OpenBus control boxes. The LA31 actuator has an ingress protection of IPX6 and is available in version with up to 6000N in push and 4000N in pull. Features and options: 24 V DC permanent magnet motor Thrust up to 6000 N in push and 4000 N in pull Electric chromated steel piston rod eye with slot High-strength plastic housing protects motor and gears Elegant and compact design with small installation dimensions Standard protection class: IPX1, IPX4, IPX6 Colour: light grey (RAL 7035), dark grey (RAL 7016), black (RAL 9005) 2.25 m straight cable Built-in limit switches (not adjustable) Scratch and wear-resistant powder painting on outer tube Ø30 mm Zinc alloy back fixture Strong wear and corrosion resistant stainless steel inner tube Noise level 48dB(A); measuring method DS/EN ISO 3743-1, actuator not loaded. Winding isolation class F (155 degrees) Flexible back fixture Reed switch (8 pulses per spindle revolution) for exact positioning of memory control and compatibility with CB9P) Hall positioning (for use with e.g.cb6 OBF, CB16 OBF, CB20) Fast motor Mechanical splines (the actuator can only push) Jack plug for use with other control boxes than CB9 Mini-fit cable for use with OpenBus IPX6: only with cable placement standard and forward (only when actuator fully retracted) 0.2/0.4 m coiled cable with DIN or Jack plug Safety nut in push Mechanical end stop in both directions is possible as an option Exchangeable cables (also for DIN Reed) Manual lowering Hall Potentiometer (not with EL or mechanical spline). Cannot be used with LINAK control boxes Can be mounted together with CB7 or CB9 (without cable) Brake in either push or pull is possible as an option Usage: Duty cycle: Max 10% or 2 minutes continuous use followed by 18 minutes not in use Cycles: The LA31 life cycle test has been performed with a stabilised power supply (10% duty cycle) on a 200 mm stroke actuator at max. load for 10,000 cycles (at ambient temperature) Ambient temperature +5 C to +40 C Approvals: IEC 60601-1, ANSI/AAMI ES60601-1 and CAN/CSA-22.2 No 60601-1 LINAK.COM/MEDLINE-CARELINE

LA31 MEDLINE & CARELINE Ordering example XX 001 000 A P 00 0 0-1 0 1 0 0 XXXX 2 1 S 0 0 0 Not used Not used Safety factor: 0 = 2.0 1 = 1.5 2 = 2.5 Cable: C = Coiled E = Exchangable (female) S = Straight X = Special Plug Type: A = JACK E = (Molex) B = DIN F = DL plug C = Flying leads V = Angled Minifit D = Mini-fit X = Special Fire Category: Install. Dim.: Brake: 1 = V0 3 = HB XXXX = mm (see later pages) 1 = Brake PUSH 2 = Brake PULL Option Position: C = Manual lowering A = Internal QR *) D = Free Wheel B = QUICK-RELEASE *) X = Other Piston Rod Eye: Back fixture: 1 = With slot 3 = LP 2 = Solid X = Special 1 = 0 degrees 3 = LP 2 = 90 degrees X = Special Colour: 1 = Black RAL 9005 3 = Dark grey RAL7016 2 = Grey RAL7035 X = Special IP: 1 = IPx1 6 = IPx6 4 = IPx4 Motor Type: A = 12 V DC Normal D = 12 V DC fast B = 24 V DC Normal E = 24 V DC fast X = Other F = 24 V DC Double worm PCB Options: Positioning: 0 H = Hall R = Reed Z = Hall potentiometer X = Special Safety: E = Mechanical end stop A = Safety nut F = Mechanical end stop + Safety nut G = Mechanical end stop + Spline C = Spline H = Mechanical end stop + Spline + Safety nut D = Spline + Safety nut. X = Special Stroke Length: XXX = mm 0-250 mm max. 6000N (see later pages) 250-350 mm max. 4000N (see later pages) Spindle Pitch: Actuator Type: 030 = 3 mm 040 = 4 mm 050 = 5 mm 060 = 6 mm 090 = 9 mm 120 = 12 mm XX = LAxx *) QR versions are not available for sale to new projects. We recommend to use LA40 QR. N.B. Always configure the LA31 in the configurator Econ Smart Client before ordering.

Installation dimensions with different combinations of features/back fixtures LA31 standard LA31 Standard Length 115 115 < Length 250 250 < Length 300 300 < Length 350 Max. load = 6000 N 6000 N 4000 N 4000 N Back fixture BID (minimum) BID BID BID 1/2 and 7/8 288 mm S + 173 mm S + 192 mm S + 212 mm 5/6 288 mm S + 173 mm S + 192 mm S + 212 mm A/B 291 mm S + 176 mm S + 195 mm S + 215 mm Explanation of table: Min. installation dimension = 288 mm when the stroke is 115 mm. Min. installation dimension is S + 173 when the stroke length is 115<Length 250 and S + 176 with A/B type back fixture. Min. installation dimension is S + 192 when the stroke >250 mm and S + 195 with A/B type back fixture. Min. installation dimension is S + 212 when the stroke > 300 mm and S + 215 with A/B type back fixture. BID = Built-in dimension LA31 with mechanical splines LA31 mechanical splines Length 100 100 < Length 250 250 < Length 300 300< Length 350 Max. load = 6000 N 6000 N 4000 N 4000 N Back fixture BID (minimum) BID BID BID 1/2 and 7/8 289 mm S + 189 mm S + 189 mm S + 211 mm 5/6 289 mm S + 189 mm S + 189 mm S + 211 mm A/B 292 mm S + 192 mm S + 192 mm S + 214 mm LA31 with mechanical end stop LA31 mechanical end stop Length 95 95 < Length 250 250 < Length 300 300< Length 350 Max. load = 6000 N 6000 N 4000 N 4000 N Back fixture BID (minimum) BID BID BID 1/2 and 7/8 288 mm S + 193 mm S + 193 mm S + 212 mm 5/6 288 mm S + 193 mm S + 193 mm S + 212 mm A/B 291 mm S + 196 mm S + 196 mm S + 215 mm

For patient lifts the following correlation between stroke length and built-in dimension is needed in order to live up to the required safety factor of 1.5 according to EN10535. The complete application must be tested in order to comply with Safety Factor 1.5. LA31 with mechanical splines+ml+mechanical end stop to patient lifts LA31 mechanical splines+ml+mechanical end stop to patient lifts Length 100 100 < Length 250 250 < Length 300 Max. load = 6000 N 6000 N 6000 N Back fixture BID (minimum) BID BID 1/2 and 7/8 374 mm S + 274 mm S + 274 mm 5/6 374 mm S + 274 mm S + 274 mm A/B 377 mm S + 277 mm S + 277 mm With manual lowering the installation dimension is increased as shown above, include SPLINE, MECHANICAL END STOP + 50 mm OVERLAP (20 mm from mechanical endstop is included in the 50 mm). LA31 mechanical splines + mechanical end stop - to patient lifts LA31 mechanical splines + mechanical end stop to patient lifts Length 100 100 < Length 250 250 < Length 300 Max. load = 6000 N 6000 N 6000 N Back fixture BID (minimum) BID BID 1/2 and 7/8 309 mm S + 209 mm S + 219 mm 5/6 309 mm S + 209 mm S + 219 mm A/B 312 mm S + 212 mm S + 222 mm Side load is max.1000n with stroke lengths up to 300 mm.

Back fixtures: 1 2 5 6 7 8 A B

Piston rod eyes: Standard solid: Dimensions LA31 Installation dimension (± 2 mm) = Stroke + addition N.B Tolerance information ± 2 mm - Installation dimension ± 2 mm - Stroke length

Motor current versus load Ampere LA31 24V Std. motor current v s load 4,5 4 3mm 3,5 5mm brake 5mm 3 9mm brake 6mm brake 12mm brake 2,5 2 1,5 1 0,5 0 0 1000 2000 3000 4000 5000 6000 7000 Load (N) 4mm brake 30 LA31 24V Std. motor speed v s load 25 12mm brake Speed (mm/s) 20 15 6mm brake 9mm brake 10 5mm brake 5mm 4mm brake 5 3mm 0 0 1000 2000 3000 4000 5000 6000 7000 Load (N) LA31 24V Fast motor current v s load 6 5 4mm brake Ampere 4 3 2 12mm brake 9mm brake 6mm brake 5mm brake 5mm 3mm 1 0 0 1000 2000 3000 4000 5000 6000 7000 Load (N) 35 LA31 24V Fast motor speed v s load 30 12mm brake Speed (mm/s) 25 20 15 9mm brake 10 5mm brake 6mm brake 5mm 4mm brake 5 3mm 0 0 1000 2000 3000 4000 5000 6000 7000 Load (N)

LA31 with manual lowering: The purpose of the ML (Manual Lowering) is in case of a power failure, to be able to mechanically lower a patient by turning the ML part in the clockwise direction until the actuator is fully lowered. The ML part substitutes the std. piston rod eye, and its function is intended for patient hoists. In normal operation the ML part will work as a piston rod eye, but in a failure condition the actuator can be lowered by turning the outer ring manually. Dimensions: Features: Mechanical lowering of actuator Zinc piston rod eye - slotted With Ø10.1 POM bushings Max. 6000 N in push Only 3 mm and 4 mm pitch actuators IPx4 Grey - RAL 7035 Drawing no. 03111202 Copyright LINAK 2018.12 MA M9-02-058-K. Chapter 5.8.2 Terms of use The user is responsible for determining the suitability of LINAK products for specific application. LINAK takes great care in providing accurate and up-to-date information on its products. However, due to continuous development in order to improve its products, LINAK products are subject to frequent modifications and changes without prior notice. Therefore, LINAK cannot guarantee the correct and actual status of said information on its products. While LINAK uses its best efforts to fulfil orders, LINAK cannot, for the same reasons as mentioned above, guarantee the availability of any particular product. Therefore, LINAK reserves the right to discontinue the sale of any product displayed on its website or listed in its catalogues or other written material drawn up by LINAK. All sales are subject to the Standard Terms of Sale and Delivery for LINAK. For a copy hereof, please contact LINAK. FOR MOUNTING INSTRUCTIONS AND GUIDANCE IN USAGE, PLEASE SEE THE RELEVANT USER MANUAL