KV4-300 mk2. KV5-200 mk2. KV5-100T mk2. KV3-250 mk2

Similar documents
OPERATING AND MAINTENANCE MANUAL. Primary Current Injection Test Set. 750ADM-H mk2

Metrohm E3640 FLASH TESTER INSTRUCTION MANUAL. Martindale Electric Co Ltd.

1 This instrument must only be used by a competent and trained person and operated in strict accordance with the instructions.

BM kv Digital Insulation Tester USER MANUAL

& HIGH CURRENT DC POWER SUPPLIES INSTRUCTION MANUAL

OPERATIONS AND APPLICATIONS MANUAL MODEL NUMBER PM6. 6 KV DC HIGH POTENTIAL TESTER WITH MEGOHMMETER Version 2.0

OPERATING INSTRUCTION

R&S RT-ZC02 R&S RT-ZC03 AC/DC Current Probe User Manual

MAX. AC 600V HOLD V M H.V. LIVE CIRCUIT DIGITAL INSULATION TESTER DO NOT OPERATE IF LIT POWER DATA HOLD TEST ON / OFF 250V 500V 1000V OFF CAT.

GUIDELINES ON ELECTRICAL SAFETY AT WORKPLACES

RE-PR3-E-86&105 3-Phase Panel Mount 86 and 105kW

Installation, Operation and Maintenance Manual. EVC Controller (from February 2007)

MODEL 520 REMOTE START ENGINE MANAGEMENT SYSTEM

SDS Enclosed Star-Delta Starter User Guide

HT Dielectric Withstand Tester Volts AC Output 25mA Leakage Option

Unrestricted. S.Mukhopadhyay

KEWTECH. KT56 digital multi function tester. Instruction manual

Digitrip Retrofit System for ITE K-3000, K-3000 S, K-4000 and K-4000 S Breakers

HIB Enclosed Inverter User Guide

12 amp RMS Battery Charger

GUARD-LITE SERIES EMERGENCY UNIT EQUIPMENT

1.Safe testing IMPORTANT:

INSTRUCTION MANUAL ANALOGUE INSULATION TESTER KEW 3131M KYORITSU ELECTRICAL INSTRUMENTS WORKS,LTD., TOKYO JAPAN

INSTRUCTION MANUAL FOR THE THERMOCOUPLE ATTACHMENT UNIT (TAU)

Dielectric Withstand Tester

IN1001/IN1003 ANALOGUE INSULATION & CONTINUITY TESTERS INSTRUCTION MANUAL IN1001/IN1003 ANALOGUE INSULATION & CONTINUITY TESTERS INSTRUCTION MANUAL

In general this document presents specific safety procedures for systems with potentially lethal voltages on bare or exposed conductors.

MULTIVOLTAGE PORTABLE BATTERY CHARGER MVM

Open Chassis Star-Delta Starter User Guide

HIGH VOLTAGE PROXIMITY DETECTOR

Rental Industry Safety Tester Safe Check 5s

Monicon Instruments Co., Ltd. CHR-1285/2485 CHR-1285/2485 BATTERY CHARGER

PORTABLE POWER SUPPLY MODEL NUMBER VMS-1 VERSION 1.0

DEEP SEA ELECTRONICS PLC

Electrical test equipment for use on low voltage electrical systems

Electrical Safety. Electrical Safety Webinar. Electrical. Printing Industries Alliance Printing Industries Alliance 1

FLÄKTGROUP PM-MOTOR WITH INTEGRATED FC 106 FREQUENCY CONVERTER

Unified requirements for systems with voltages above 1 kv up to 15 kv

Chapter 14:08 Factories and Works (Electrical) Regulations, Chapter 14:08 Factories and Works (Electrical) Regulations, 1976

Distributor. Kewtech Corporation Limited 76 St. Catherine s Grove Lincoln LN5 8NA KEWTECH

Switching DC Power Supply

DEEP SEA ELECTRONICS PLC

Car Battery Charger Instructions for Use

High Frequency SineWave Guardian TM

SDM72 Start on Demand Modules

MIT300 Series Insulation and Continuity Testers

HT Dielectric Withstand Tester Volts AC Output Volts DC Output COMPLIANCE WESTUSA. Instruction Manual

DUAL 60V 20A POWER FLEX POWER SUPPLY INSTRUCTION MANUAL

Model 8 Mk7 AVOMETER USER GUIDE GUIDE DE L UTILISATEUR BEDIENUNGSANLEITUNG GUÍA DEL USUARIO

3 Phase Smart Controller

MARTINDALE INSTRUCTIONS ELITE FUSE FINDER KIT ELECTRIC. Trusted by professionals. 4.4 Storage Conditions

MARTINDALE MARTINDALE IN2001/IN2003 INSULATION & CONTINUITY TESTERS INSTRUCTION MANUAL IN2001/IN2003 INSULATION & CONTINUITY TESTERS

5 IN 1 JUMP START OPERATION & MAINTENANCE INSTRUCTIONS MODEL NO: JS5IN1 PART NO: LS0810

RE-PR1-F 1-Phase Din-Rail Mount 1.5, 3 & 6kW

Model 340i / 680i / 1220i DC drive product manual HG iss 8 1

Triple Output Power Supply

AA4006 Speed Sensitive Trip Unit. User Manual

Installation and Maintenance Instructions. World Leader in Modular Torque Limiters. PTM-4 Load Monitor

Declaration of Conformity

IB PowlVac ITE-HK Remote Racking Device

An average of one worker is electrocuted on the job every day There are four main types of electrical injuries:

This instrument is manufactured in the United Kingdom. The company reserves the right to change the specification or design without prior notice.

OPERATING INSTRUCTION

Data Bulletin. Ground-Censor Ground-Fault Protection System Type GC Class 931

PREMIER POWER PACK INSTRUCTION MANUAL EN54-4 POWER SUPPLY UNIT INSTRUCTION MANUAL. GLT.MAN-138 Issue: /05/2016 N.R.P.J.

REVISED ATM72 Auto Transfer Module

MP V 8A Electronic Smart Charger. Instruction and Information Manual

99 Washington Street Melrose, MA Fax TestEquipmentDepot.com. Instruction Manual. Model 1672 Triple Output Power Supply

Continuing Education Course #206 Introduction to Designing Machine Control Systems Part 2

Electrical Awareness for Construction Activities. Overhead Power Lines & Underground Cables

BATTERY CHARGER-STARTER

The Management of Electrical Safety in Quarries, Associated Plant and Equipment

Valcom Failsafe Unit. 1620ESv2 SERIES. Operation and Maintenance Manual

MOdel No: MM19.V3 WARNING CAUTION 1. SAFETY INSTRUCTIONS

dv Sentry TM 208V 600V INSTALLATION GUIDE Quick Reference ❶ How to Install Pages 6 14 ❷ Startup/Troubleshooting Pages WARNING

Model ER-340XRi / ER-680XRi / ER-1220XRi DC drive product manual HG iss 9 1

PURE SINE WAVE DC TO AC POWER INVERTER

1200W INVERTER GENERATOR

HT-25kVdc Dielectric Withstand Tester Volts DC Output

EC225 Power Control System

ANALOGUE INSULATION-CONTINUITY and VOLTAGE METER

Energized Electrical Work

RP-4000 Reserve Power Control

SM-1700 Secondary Shunt Trip Coil Assembly. Powered by Safety

DRL-1 Series Instruction Manual

World Leaders in Diesel Fuel Injection Test Equipment. IFT-c. Injector Function Tester Controller. Operating and Servicing Manual

The 4HR-UPS 220V ac output is provided via a Pure Sine Wave inverter, allowing it to be backed up by 2x 12V standby batteries.

DRB-1 Series Instruction Manual

Lastolite Lastolite Lumen8 F200 Lumen8 F400 Stored Energy Guide Number (m/100iso) Recycling Time (100%) Flash Duration Flash Power

GP/GPA GROUND FAULT PROTECTION

PowerLogic High Density Metering System 4-Meter Enclosure Installation Guide

Parker AC10 Frequency Inverter (to 22kW) Easy Start Guide

Reviewed: DD Month University Code of Practice for Electrical Safety. PART B - Design and Construction of Electrical Equipment within the University

Wilkins Safety Group

SDM320D / SDM320A. Single phase din rail meter SDM320D / SDM320A user manual

Instruction Manual. Q Series 2 Motor Control Modules - OEM

ES5 Security Theory/Regulations Answer Schedule

INDUSTRY WIDE LABOR-MANAGEMENT SAFETY COMMITTEE

PSU1 Power Supply Unit

331-SV. User Manual THREE PHASE DUPLEX LIFT STATION CONTROL PANEL WITH STATIONVIEW CONTROLLER. Ashland, OH

Transcription:

OPERATING AND MAINTENANCE MANUAL Product: Type: High Voltage AC Test Set KV3-250 mk2 KV4-300 mk2 KV5-100 mk2 KV5-100T mk2 KV5-200 mk2 DESIGNED AND MANUFACTURED BY: T & R Test Equipment Limited 15-16 Woodbridge Meadows, Guildford, Surrey, GU1 1BJ, United Kingdom Telephone: 01483 207428 e-mail: sales@trtest.com Fax.: 01483 511229 Web: www.trtest.com 0

1

GENERAL SAFETY STATEMENT The following safety precautions should be reviewed to avoid injury to the user and damage to the product (and other products connected to it). To avoid potential hazards only use this product as specified. Only suitably qualified personnel should use this equipment. Servicing of this product should only be carried out by suitably qualified service personnel. The high voltage generated by this unit is extremely dangerous and may be fatal. This unit is designed for operation in a designated high voltage test area. To Avoid Fire Hazards and Personal Injury Use the correct power supply lead. Only use a suitably rated and approved power supply lead for the country of use. Ensure that systems that the unit is to be connected to are dead. Do not connect and disconnect leads whilst outputs are switched on. Ensure that the product is grounded. To avoid electric shock it is essential that the grounding conductor is connected to the earth ground. An additional earth terminal is provided on the unit for connection to a local earth if necessary. Ensure that the unit is properly grounded before making any connections to inputs or outputs. Output ratings must be observed to prevent fire hazards and risk of injury to the operator. Consult the product manual for ratings information before making any connections to the unit. It is ESSENTIAL to consult the product manual for rating information before making any connection to a terminal or terminal group marked with a warning triangle. Only use fuses of a type and rating specified for this product. Do not operate the unit out of its case or with any covers or panels removed. Do not touch exposed connections and components when power is present. Do not operate the product if any damage is suspected. Refer the unit to qualified service personnel to be checked. Do not operate the unit in wet or damp conditions Do not operate the unit in an explosive atmosphere This unit is not designed for unattended operation If any further queries occur regarding the usage and maintenance of the equipment detailed in this manual, please refer these to the supplier of the equipment in the first case or to: T & R Test Equipment Limited 2

HIGH VOLTAGE SAFETY It is essential to follow safe working procedures when working with high voltage. Information on accepted codes of practice should be obtained from your local heath and safety regulatory body. IEEE standard 510-1983 (IEEE Recommended Practices for Safety in High-Voltage and High- Power Testing) provides a working framework for establishing safe procedures, but must be read in conjunction with local regulations and accepted codes of practice. The following excerpts are taken from IEEE 510 All ungrounded terminals of the test equipment or apparatus under test should be considered as energised. Common ground connections should be solidly connected to both the test set and the test specimen. As a minimum, the current capacity of the ground leads should exceed that necessary to carry the maximum possible ground current. The effect of ground potential rise due to the resistance and reactance of the earth connection should be considered. Precautions should be taken to prevent accidental contact of live terminals by personnel, either by shielding the live terminals or by providing barriers around the area. The circuit should include instrumentation for indicating the test voltages. Appropriate switching and, where appropriate, an observer should be provided for the immediate de-energisation of test circuits for safety purposes. In the case of dc tests, provisions for discharging and grounding charged terminals and supporting insulation should also be included. High-voltage and high-power tests should be performed and supervised by qualified personnel. Consideration should be given to safety regulations which may apply to specific circumstances; for example, HSE, company, or government regulations. In the use of signal-gathering equipment, each device should be used in such a manner that it will not present a personnel hazard should it inadvertently become a part of the high-voltage circuit, or fail to function properly. 3

SAFETY TERMS AND SYMBOLS The following safety symbols appear on the equipment: CAUTION/WARNING Refer to manual DANGER High voltage Mains off Mains on The following safety symbols appear in this manual: CAUTION This action or procedure may be dangerous if not carried out correctly, and may cause damage to the equipment or connected equipment. WARNING This action or procedure may be cause injury or death to the operator or other personnel if not carried out correctly using applicable safety procedures. 4

CONTENTS 1 DESCRIPTION OF EQUIPMENT 7 1.1 Electrical Specification 7 1.1.1 Supply Requirements 7 1.1.2 Output Specifications 7 1.1.3 Zero Voltage Interlock 7 1.1.4 External Interlock 7 1.2 Output Voltage Control 8 1.3 Overload Protection 8 1.3.1 Variable overload 8 1.4 Metering 9 1.5 Fault Burn Facility (except KV5-200 mk2) 10 1.5.1 Fault Burning 10 1.5.2 Burn Switch 10 1.6 High Voltage Connection 10 1.7 Construction 10 1.8 Timer Operation (KV5-100T mk2 only) 11 2. OPERATION 12 2.1 Safety 12 2.1.1 Installation 12 2.1.2 Operation 12 2.2 Front Panel Layout 13 2.2.1 Front Panel Control Functions 15 2.2.2 Rear Panel Control Functions 15 2.3 Method of Operation 16 2.3.1 Connections 16 2.3.2 Operation 16 2.3.3 Timer Operation (KV5-100T mk2 only) 16 5

3. MAINTENANCE 17 4. STANDARD ACCESSORIES 18 5. OVERALL PERFORMANCE SPECIFICATION 19 6. REVISION 20 6

1 DESCRIPTION OF EQUIPMENT 1.1 Electrical Specification 1.1.1 Supply Requirements KV5-100 mk2 115/230V 10% 600VA max KV3-250 mk2 115/230V 10% 900VA max KV4-300 mk2 230V 10% 1400VA max KV5-200 mk2 230V 10% 1100VA max 1.1.2 Output Specifications The output voltage, current and duty cycle are as follows: Output current Unit Output voltage 5 min on/ 15 min off 1 min on/ 15 min off 40 seconds on/ 15 min off KV5-100 mk2 0-5kV 100mA - - KV3-250 mk2 0-3kV 125mA 250mA 300mA KV4-300 mk2 0-4kV 200mA 300mA - KV5-200 mk2 0-5kV 100mA 200mA - One side of the high voltage transformer output is connected to the earth system of the equipment. 1.1.3 Zero Voltage Interlock The equipment is fitted with a zero volt interlock system on the output control. This interlock prevents the output being energised unless the output control is in the minimum output position. 1.1.4 External Interlock The equipment is fitted with a external interlock system. The interlock socket on the back panel must have a short circuit across it terminals to enable the output to be energised. Note: the open circuit voltage across the interlock terminals is 240Vac and is connected directly to the mains supply. The unit is supplied with a plug fitted to the external interlock socket on the rear of the unit. The plug contains a shorting link to enable the unit. If external interlocks are used, the interlocks should replace the shorting link. Please note that the output will not switch on without either the link fitted in the plug or a closed external interlock circuit. 7

1.2 Output Voltage Control The output from the equipment is controlled from zero by means of the output control. To increase the output voltage turn the control knob in a clockwise direction. Note: The output cannot be energised unless the regulator knob is set to zero, thus operating the zero voltage interlock switch. 1.3 Overload Protection The equipment is fitted with a fuse and variable overload protection circuits as standard. 1.3.1 Variable overload The variable overload protection system senses current changes in the high voltage circuit. The trip levels are adjustable in two ranges each of 6 steps giving 2-12% and 20-120% rated current. Unit X1 Trip current ranges X10 KV5-100 mk2 2, 4, 6, 8, 10, 12mA 20, 40, 60, 80, 100, 120mA KV3-250 mk2 5, 10, 15, 20, 25, 30mA 50, 100, 150, 200, 250, 300mA KV4-300 mk2 6, 12, 18, 24, 30, 36mA 60, 120, 180, 240, 300, 360mA KV5-200 mk2 4, 8, 12, 16, 20, 24mA 40, 80, 120, 160, 200, 240mA The range change and selector switches are located on the front panel of the instrument. The circuit will activate when the load current exceeds that set by the trip level selector switch. WARNING The variable overload trip circuit does not limit the output current. 8

1.4 Metering The equipment is provided with the following metering ranges using analogue moving coil meters: Metering Ranges Unit Voltage Current Low High X1 X10 KV5-100 mk2 3kV 6kV 10mA 100mA KV3-250 mk2 1.6kV 3.2kV 30mA 300mA KV5-200 mk2 3kV 5kV 20mA 200mA The ranges on both instruments are selected by means of switches located on the front panel. 9

1.5 Fault Burn Facility (except KV5-200 mk2) The KV5-100 mk2, KV3-250 mk2 and KV4-300 mk2 are provided with an internal current limit circuit which limits the output current from the unit under short circuit conditions. This can be used to burn faults. 1.5.1 Fault Burning Fault burn is where the arc is maintained through the failed test object. The product does this by limiting the current and preventing the unit from tripping. Fault burning should be approached very cautiously as it can cause serious damage to the test object, and is totally unsuitable for many test objects. 1.5.2 Burn Switch When the burn switch is in the out position, the unit behaves normally and the output current is not limited. Switching the burn switch to in enables the current limiting The current is limited to the following values at maximum output voltage when the burn is enabled (the burn switch is in the in position): Output Voltage Maximum current KV5-100 mk2 5kV 90-100mA KV4-300 mk2 4kV 280 300mA KV3-250 mk2 3kV 160-180mA At lower output voltages, the burn current limit reduces in proportion to the output voltage. 1.6 High Voltage Connection The equipment is supplied with crocodile clips on the output. These clips are not insulated for high voltage use, and under no conditions must the high voltage connection be approached or touched whilst the unit is switched on. The test object must be earthed before making or breaking connections after testing. A prohibition zone of at least 30mm must be maintained around the high voltage connection (from EN50191:2000) WARNING This equipment must be used either with an interlocked enclosure for the object under test or in a suitable high voltage test area. Adequate safety clearances must be maintained between the operator and all high voltage parts at all times. 1.7 Construction The unit is housed in a robust case with a carrying handle. The handle will also act as a tilt support stand which is locked in various positions by pressing the release buttons at each end of the handle. 10

1.8 Timer Operation (KV5-100T mk2 only) The KV5-100T mk2 is fitted with a 5 second-5 min test duration timer. The test timer has two controls; Select Test Time and Alarm Cancel. The timer has nine switch-selectable ranges of 5, 10, 15, 20, and 30 seconds and 1, 2, 3 and 5 minutes. There is an OFF position for when the timer operation is not required. The timer is started when the output HV ON push button is pressed. The timer indicator then flashes per second for the duration of the time selected. When the timer has reached the selected time an intermittent alarm sounds, this is cancelled by pressing the Timer cancel push button. The audible alarm is cancelled using the Timer cancel push button. WARNING The timer does not switch the high voltage output off it only warns the operator when the required time has elapsed. At the end of the test time, the output control should be run down to zero before switching the output off. The high voltage output remains until HV OFF button is pressed, TIMER CANCEL does not turn the output off. The timer is cancelled if there is an over current trip condition, should this occur the RESET button must be pressed to clear the trip and silence the audible alarm. Timer operation may be aborted by pressing the TIMER CANCEL push button. Adjusting the TIMER select switch during a timed operation will have one of the following effects: Change to a longer setting will extend the period to the new value. Change to a shorter setting will limit the period to the new value. Change to a setting shorter than has already elapsed terminates the timing. Change to the OFF setting triggers the timer alarm 11

2. OPERATION 2.1 Safety The outputs from all units in the range are very dangerous, and if used incorrectly could be fatal. The unit must only be installed, operated, and maintained by suitably qualified and trained personnel. It is essential to follow accepted safety procedures and local health and safety regulations and guidelines when installing and operating high voltage equipment. A risk assessment should be undertaken on both the installation and the working procedures to ensure the safety of test personnel and all other personnel. 2.1.1 Installation The unit is designed to be installed and used in an indoor environment. 2.1.1.2 Test Area The unit must be installed in a suitable high voltage test area completely enclosed by walls or some type of physical barrier. Appropriate controls and safety measures must be applied to this area including interlocks connected to the supply or HV unit interlock to ensure that the unit cannot be switched on unless the area is secure. Refer to BS EN50191:2001 and IEEE Std 510-1983 for further details of suitable test enclosures. The test area must also be identified with suitable signs. 2.1.1.3 Mains supply The unit must be connected to a suitable supply via an approved and suitably rated mains connector with earth connection. 2.1.1.4 Earthing The unit must be securely earthed. The earth terminal on the rear of the unit may be connected to a low impedance local earth if required. 2.1.2 Operation It is essential that safe working practices are maintained when conducting high voltage testing. Safe working procedures must be implemented to accepted standards. 2.1.2.1 Interlocks The KV5-100 mk2 series are provided with an external interlock circuit that may be used to link to interlock switches on the high voltage test area. An interlock should be considered to be a safety back-up feature. An interlock should not be regarded as a substitute for adequate safety rules and proper operator vigilance. 12

2.2 Front Panel Layout Figure 2.1 KV5-100 mk2 front panel A B C D E 1 2 ma 0.5 1 3 kv.ac 1.5 F G H J K L M N P Figure 2.2 KV3-250 front panel 13

Figure 2.3 KV5-100T front panel Figure 2.4 KV5-200 front panel 14

2.2.1 Front Panel Control Functions Figure 2.5 KV4-300 mk2 front panel A kv Meter Displays output voltage B ma Meter Displays load current C ma x1/x10 selector switch Selects ma meter range D Burn switch Selects burn facility E Trip reset switch Resets the output current trip F HV Output lead HV output to object under test G kv meter range switch Selects kv meter range H Earth output lead Earth output to object under test J HV Output ON pushbutton Switches HV output on. Illuminated red when output is on. K Variable trip level select Selects output current trip level L HV Output OFF pushbutton Switches HV output off. Illuminated green when output is off. M Mains supply switch (key operated) Key operated mains on/off switch O Emergency stop button Switches HV output off. N Supply on indicator Supply on indicator P Output level control Set output voltage Q Timer reset pushbutton (T only) Resets timer and cancels alarm Timer running indicator T Test time select (T only) Selects test time U Output trip Flashover trip KV5-200 mk2 only 2.2.2 Rear Panel Control Functions Mains voltage selector Interlock socket Earth terminal Selects mains input voltage For connection to an external interlock system For connection to a low impedance local earth 15

2.3 Method of Operation 2.3.1 Connections WARNING Before making any connections please ensure that you are aware of all hazards relating to the system and environment in which it is operating. Ensure that the key switch is in the off position and the key is removed before making any connections. The test object must be earthed before making or breaking any connections after testing. Ensure that the key switch is in the off position and the key is removed before making connections to the object under test or connecting the unit to the mains supply. Connect the mains input lead to a suitable three pin supply outlet. 2.3.2 Operation Switch on the key-operated mains switch. The following will now occur: a. Power ON lamp will be lit b. HV OFF lamp will be lit c. The variable overload reset lamp will be lit Press the variable overload reset push button. This will arm the trip circuit. Select the desired voltage range on the kv meter and set the desired overload trip level. Before commencing testing ensure that all external interlocks are closed, the emergency stop button is released and that the regulator is in the fully anti-clockwise position. Press the HV ON pushbutton. The HV output will now energise, and the red HV on indicator in the pushbutton will light up. The output voltage can now be increased to the desired level as indicated on the kv meter. In the event of a test object failure the HV output will be automatically switched off by the protection system. When the test is completed, turn the regulator control knob fully anti- clockwise and switch off the HV output and then the mains supply. In case of emergency, the emergency stop push button is provided, switching off the HV output when engaged. 2.3.3 Timer Operation (KV5-100T mk2 only) The timer is started when the output is switched on. The indicator in the timer cancel button flashes for the duration of the time selected indicating that the timer is in operation. An alarm sounds to indicate the timer reaches the selected time duration. The audible alarm is cancelled using the Timer cancel push button. WARNING The timer does not switch the high voltage output off it only warns the operator when the required time has elapsed. At the end of the test time, the output control should be run down to zero before switching the output off. 16

3. MAINTENANCE Maintenance and repair of the KV5-100 mk2 series must only be carried out by suitably qualified and trained personnel. Potentially lethal voltages are present inside the unit and on the output leads. WARNING Ensure that the unit is disconnected from the mains before removing any covers. It is advisable to check the carbon brush on the variable transformer for signs of wear on a regular basis. To gain access to the variable transformer remove the top cover of the unit (secured by four screws - two in each side of the case). Replacement brushes are available from T&R Test Equipment. 17

4. STANDARD ACCESSORIES The following items are provided with the equipment: a. Mains input lead. b. 2 keys (for mains ON/OFF switch). c. Operating & Maintenance Manual. d. 1x F5A 1¼ inch fuse and 1x T4A 20mm fuse: KV5-100 mk2, KV3-250 mk2. 1xT5A 1¼ inch fuse and 1xT8A 20mm fuse: KV4-300 mk2. 1 off T6.3A: KV5-200 mk2. 18

5. OVERALL PERFORMANCE SPECIFICATION Insulation resistance at 500V DC: >10 megohms between mains input and frame. Applied voltage test: 2.0kV RMS for 1 minute between mains input and frame. Accuracy of instruments kv meter: ±1.5% of FSD ma meter: ±2.5% of FSD Instrument calibration is traceable to national standards. 19

6. REVISION Product / Type: File: Author: KV5-100 mk2 series High Voltage AC Test Systems KV5-100mk2 series manual v5.doc P Cole Issue / Date: 5 / 24.06.2013 Modified By: TC Checked By: PRC Date: 24.06.13 20

Circuit diagram A2/001910 KV5-100 mk2 Other unit circuit diagrams available on request A2/000925 KV3-250 A2/001924 KV4-300 mk2 A2/001948 KV5-100T A2/001708 KV5-200 21