i400s Calibration Information Introduction

Similar documents
Calibration Manual. Detachable Jaw True-rms Clamp Meter

323/324/325 Clamp Meter

373 Clamp Meter. Users Manual

365 Detachable Jaw True-rms Clamp Meter

Temperature Probe, Battery (installed), Instruction Sheet, Quick Reference Card and Warranty Card.

302+/303 Clamp Meter. Users Manual

Users Manual. Phase Rotation Indicator

Tachometer. English. Users Manual

80PK-2A Type K Immersion Probe

Infrared Thermocouple Probe

P5510. Users Manual. Pneumatic Comparison Test Pump

362 Clamp Meter. Users Manual

362 Clamp Meter. Users Manual

Temperature Calibrator

Users Manual. MegOhmMeter

The following table shows the symbols used on the product and/or in this manual. Hazardous Voltage. Risk of electric shock.

BT508/BT510/BT520/BT521 Battery Analyzer Safety Sheet

P5514, P5514-V & P5514-EP Hydraulic Comparison Test Pump

i3000s/i2000 Flex AC Current Probe

Calibrator Capabilities Function Range Resolution Measure V dc 28 V 1 mv Measure ma dc Source ma dc

233 True-rms Remote Display Digital Multimeter Safety Information

1587 FC/1587/1577. Insulation Multimeter. Safety Information. Warnings

Visit us at

1662/1663/1664 FC. Electrical Installation Tester. Safety Information. Warnings

712 RTD Calibrator. Instruction Sheet. Introduction

716 Pressure Calibrator

INSTRUCTION MANUAL CM95/ CM100 TRUE RMS FLEXIBLE CLAMP METER. 1. SAFETY INFORMATION: Always read before proceeding. REMEMBER: SAFETY IS NO ACCIDENT

MegOhmMeter. Users Manual

Go to to register your product and find more information.

Instruction Manual PC104/PC105 3-PHASE SOCKET TESTER. Check out what else you can get from Martindale:

80PK-8 Type K Thermocouple Pipe Clamp Temperature Probe


RS-3 PRO RS-1007 PRO. CAT IV Analog Clamp meter Series. Users Manual. For detailed specifications and ordering info go to

INSTRUCTION MANUAL CM69 TRMS AC EARTH LEAKAGE CLAMP METER. 1. SAFETY INFORMATION: Always read before proceeding. REMEMBER: SAFETY IS NO ACCIDENT

Instruction Manual VT4 VOLTAGE TESTER. Check out what else you can get from Martindale:

717 Series. Pressure Calibrators. Instruction Sheet. Introduction

LH41A. User Manual. Clamp On Ammeter. For detailed specifications and ordering info go to

DT33B/C/D OPERATOR S MANUAL LIMITED WARRANTY AND LIMITATION OF LIABILITY

27 II/28 II. Digital Multimeters. Getting Started Manual. Specifications

PY-11 Voltage & Continuity Tester

Users Manual. Airflow Meter

R&S RT-ZC02 R&S RT-ZC03 AC/DC Current Probe User Manual

1551A Ex/1552A Ex. Calibration Manual. Stik Thermometer

1000FLT. Users Manual. Fluorescent Light Tester

PRM-4. Phase Sequence and Motor Rotation Tester. Users Manual. For detailed specifications and ordering info go to

INSTRUCTION MANUAL DT173 DT175 DIGITAL THERMOMETERS ALWAYS READ THESE INSTRUCTIONS BEFORE PROCEEDING

51 & 52 Series II. Users Manual. Thermometer. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

707Ex. Users Manual. ma Calibrator

719 Series. Pressure Calibrator with Electric Pump. Product Overview (E)

MY /2-DIGIT DIGITAL MULTIMETER Users Manual

PK-8270-X/PK-8370-X. Lines and Fittings Kit. Instructions. Introduction. How to Contact Fluke Calibration

5618B Platinum Resistance Thermometer User s Guide

80 Series V Digital Multimeter Safety Information

2-Year Limited Warranty. See the Users Manual for the full warranty.

KMP Insulation-Continuity Tester. Users Manual

PD-20 HIGH VOLTAGE PROBE INSTRUCTION MANUAL

TIC 300 PRO. Users Manual Mode d emploi Bedienungshandbuch Manual d uso Manual de uso. Non Contact AC Voltage Detector. High Energy Tic Tracer

T2 Tester. Calibration Information. Introduction. Safety Information

User's Manual. High Voltage Megohmmeter. Analog Insulation Tester plus AC Voltage and Continuity Tests. Model

1 INTRODUCTION SPECIFICATIONS Electrical Data General Data... 3

P5515, P5515-V, & P5515-EP Hydraulic Comparison Test Pump

Phase Sequence Tester

Instruction Manual ICM A1 AC/DC Clamp Meter ICM A1

5616 Platinum Resistance Thermometer User s Guide

Low Amp Probe EETA308A

SOIL RESISTIVITY TESTER

PK-VLV-ABORT-100M, PK-VLV-ISO-40M, PK-VLV-ISO-100M

Instruction Manual VT25 / VT28 VOLTAGE TESTER ALWAYS READ THESE INSTRUCTIONS BEFORE PROCEEDING

Users Manual. Accessories Motor and Phase Rotation Indicator ships with the following items: 3 test leads 3 test probes 3 alligator clips 9 V battery

3-Year Limited Warranty. See the Users Manual for the full warranty.

DIGITAL INSULATION RESISTANCE TESTER (5KV, 1T )

TIME-10 Process Control Stopwatch

User's Manual. Earth Ground Resistance Tester. Model Introduction

LM-100 LM-120. Light Meters. Users Manual. For detailed specifications and ordering info go to

USER MANUAL. 200A AC Mini Clamp Meter. Model MA130. Please visit for user manual translations

INSP-3 Wiring Inspection Tester

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com INSTRUCTION MANUAL PSMR1

OPERATION MANUAL MILLIAMP PROCESS CLAMPMETER MODEL - KM 071 LIST OF PRODUCTS

5.2 PRE-OPERATION PROCEDURE WARRANTY

718 30G/100G. Users Manual. Pressure Calibrator. 99 Washington Street Melrose, MA Phone Toll Free

For technical questions and replacement parts, please call

TX900. Microwave Leakage Detector. 99 Washington Street Melrose, MA Phone Toll Free

AC50A. Leakage Clamp Meter. User Manual. For detailed specifications and ordering info go to

THWD-3 and TH-3 Relative Humidity and Temperature Meters

INSPECTOR LINE LOAD SIMULATOR INSTRUCTION MANUAL TASCO, INC.

4-20mA DC PROCESS CLAMP METER

HIGH VOLTAGE PROVING UNIT

718 Series. Users Manual. Pressure Calibrator

TMULD-300 ULD-300. Ultrasonic Leak Detector. Users Manual. 99 Washington Street Melrose, MA Phone Toll Free

OPERATING INSTRUCTIONS

THERMOCOUPLE CALIBRATOR Mode KM-CAL-710

Automatic Pressure Calibrator. 3-Year Limited Warranty. See the Users Manual for the full warranty.

ASTROAI USER MANUAL DIGITAL MULTIMETER. NOTE: Fully read and understand this manual before using this instrument.

HR-20P Pneumatically Controlled Pressure Regulator

BK BK A BK MA BK MA BK J93 MN93-BK MN193-BK MR193-BK SL261-BK SR193-BK

CP GermanCp 41_CP C. Operating Instructions. CP 30 CP 330 CP 1100 AC/DC current probe /7.10

OPERATING INSTRUCTIONS

Dawson DDM190D. Digital Multimeter User s Manual

IR-500 Infrared Thermometer / Voltage Detector / Flashlight

Digital Multimeter User's Manual

Transcription:

i400s AC Current Clamp Introduction XWWarning To avoid electric shock or injury, do not perform the performance tests or calibration procedures unless you are qualified to do so. The information provided in this manual is for the use of qualified personnel only. The i400s provides the information necessary to verify the performance and adjust the calibration of the Fluke i400s AC Current Clamp, hereafter known as the Current Clamp. The following information is included in this document: Safety Information and Electrical Symbols Specifications Maintenance Performance Tests Calibration Adjustment User-Replaceable Parts Warranty Statement See the i400s Instruction Sheet for complete operating instructions. Contact Information To contact Fluke, call: USA: 1-888-44-FLUKE (1-888-443-5853) Canada: 1-800-36-FLUKE (1-800-363-5853) Europe: +31 402-675-200 Japan: +81-3-3434-0181 Singapore: +65-738-5655 Anywhere in the world: +1-425-446-5500 USA Service: 1-888-99-FLUKE (1-888-993-5853) For additional information about Fluke, its products, and services, visit Fluke s web site at: www.fluke.com To register this product, go to register.fluke.com PN 2414327 February 2005 2005 Fluke Corporation. All rights reserved. Printed in U.S.A. 1

i400s Safety Information XW Read First: Safety Information To ensure safe operation and service of the Current Clamp, follow these instructions: Read the operating instructions before use and follow all safety instructions. Use the Current Clamp only as specified in the operating instructions, otherwise the clamp s safety features may not protect you. Adhere to local and national safety codes. Individual protective equipment must be used to prevent shock and arc blast injury where hazardous live conductors are exposed. Do not hold the Current Clamp anywhere beyond the tactile barrier, see Figure 1. Before each use, inspect the Current Clamp. Look for cracks or missing portions of the clamp housing or output cable insulation. Also look for loose or weakened components. Pay particular attention to the insulation surrounding the jaws. Never use the Current Clamp on a circuit with voltages higher than 1000 V CAT III or 600 V CAT IV. CAT III equipment is designed to protect against transients in equipment in fixed-equipment installations, such as distribution panels, feeders and short branch circuits, and lighting systems in large buildings. CAT IV equipment is designed to protect against transients from the primary supply level, such as an electricity meter or an overhead or underground utility service. Use extreme caution when working around bare conductors or bus bars. Contact with the conductor could result in electric shock. Use caution when working with voltages above 60 V dc or 30 V ac. Such voltages pose a shock hazard. Electrical Symbols The symbols in Table 1 appear in this document and on the Current Clamp. Table 1. Electrical Symbols Symbol Explanation, Application around and removal from hazardous live conductors is permitted. T W X Product is protected by double insulation. Risk of Danger. Important information. See Instruction Sheet. Hazardous voltage. ) Conforms to relevant Canadian Standards Association directives. P Conforms to relevant European Union directives. 2

AC Current Clamp Safety Specifications Tactile Barrier bbc03f.eps Figure 1. Tactile Barrier Safety Specifications Category Rating: CAT III 1000 V and CAT IV 600 V per EN61010-1, Pollution Degree 2 EMC: EN 61326-1, FCC for emission and immunity ): Tested to US and Canadian standards for compliance to UL61010-1 and CAN/CSA C22.2 No. 101.1:2004 P: IEC 61010-1 2 nd Edition IEC 61010-02-032 Electrical Specifications Reference Conditions: 23 ± 5 C, 20 to 75 % RH; conductor centered in jaw opening; no DC component; no adjacent conductor. 40 A Range 400 A Range Measurement Range 0.5 A to 40 A 5 A to 400 A Output 10 mv/a 1 mv/a Accuracy 45 Hz to 400 Hz 2 % + 0.015 A 2 % + 0.04 A Phase Shift: (45 Hz to 400 Hz) 0.5 A to 1 A 1 A to 5 A 5 A to 10 A 10 A to 20 A 20 A to 40 A 40 A to 400 A Unspecified 4 3 3 2 NA NA NA Unspecified 2 2 1.5 Crest Factor 3 3 to 300 A 2.5 to 400 A Typical Bandwidth: 5 Hz to 10 khz Working Voltage: 1000 V ac rms, in compliance with EN61010 Common Mode Voltage: 1000 V ac rms from earth ground, in compliance with EN61010 3

i400s Input Load Impedance (of host instrument): > 1 M Ω in parallel with up to 47 pf Maximum Non-destructive Current: 1000 A Duty Cycle: 0.5 A to 400 A continuous Influence of Adjacent Conductor: < 9.0 ma/a Influence of Conductor Position in Jaw Opening: ± 1.0 % of reading + 0.05 A General Specifications Output Cable Length: 2.5 m (8.2 ft) Maximum Conductor Size: 32 mm (1.25 in) Storage Temperature: -20 C to 60 C (-4 F to 140 F) Operating Temperature: 0 C to 50 C (32 F to 122 F) Relative Humidity: 10 C to 30 C: 95 % (50 F to 86 F) 30 C to 40 C: 75 % (86 F to 104 F) 40 C to 50 C: 45 % (104 F to 122 F) Temperature Coefficient: 0.01 % X (specified accuracy)/ C (< 18 C or > 28 C) 0.01 % X (specified accuracy)/ F (< 64.4 F or > 82.4 F ) Altitude: Operating: 2000 m; 2000 m to 4000 m, derate category rating to 1000 V CAT II/600 V CAT III Non-operating: 12000 m Dimensions: (LxWxH) 150 x 70 x 30 mm (5.09 x 2.75 x 1.18 in) Weight: 114 g (4 oz) Maintenance XW Warning To avoid possible electric shock or personal injury: Before each use, inspect the Current Clamp. Look for cracks or missing portions of the clamp housing and output cable insulating cover and for loose or weakened components. Pay particular attention to the insulation surrounding the clamp jaws. Do not use a damaged Current Clamp. If a clamp is damaged, tape it shut to prevent unintended operation. A damaged clamp under warranty will be promptly repaired or replaced (at Fluke s discretion) and returned at no extra charge. If the Current Clamp does not work or perform properly, use the following steps to help isolate the problem: 1. Inspect the jaw-mating surface for cleanliness. If any foreign material is present, the jaw will not close properly and measurement errors will result. 2. Verify that the function selection and range on the Multimeter are correct and adjusted to the sensitivity of the Current Clamp. 4

AC Current Clamp Performance Tests Cleaning Periodically wipe the case with a damp cloth and mild detergent. W Caution To avoid damaging the Current Clamp, do not use abrasives or solvents to clean the clamp. Open the jaws and wipe the magnetic pole pieces with a lightly oiled cloth. Do not allow rust or corrosion to form on the magnetic core ends. Performance Tests XWWarning To avoid electric shock or injury, do not perform the performance tests or calibration procedures unless you are qualified to do so. The information provided in this manual is for the use of qualified personnel only. The following performance tests verify the complete operation of the Current Clamp and checks the accuracy of each function against the Clamp s specifications. If the Current Clamp fails any part of the test, calibration adjustment and/or repair is indicated. In the performance tests, the Current Clamp is referred to as the unit under test (UUT). Required test equipment is defined in Table 2. Table 2. Required Test Equipment Equipment Required Characteristics Recommended Model Calibrator Multimeter AC Current: 0-8 A Max. Accuracy: 0.15 % Frequency: 45 Hz - 400 Hz AC Voltage: 20 mv - 400 mv Accuracy: 0.45 % Frequency: 45 Hz - 400 Hz Fluke 5520A Multi-Product Calibrator Fluke-45 Digital Multimeter (DMM) Toroid Coil Turns: 50 Fluke 5500A/Coil Current Accuracy Tests 1. Connect the equipment as shown in Figure 2. 2. Center the UUT jaws around the 5500A/Coil windings. 3. Set the UUT range and 5520A output as called out in Table 3. Check that the Multimeter readings are within the specified limits in Table 3. 5

i400s Table 3. Accuracy Tests i400s Range Nominal Value 5520A Output LCOMP = ON Lower Multimeter Reading Limits Upper 1 mv/a 20 A, 50 Hz 0.4 A, 50 Hz 19.56 20.44 1 mv/a 20 A, 400 Hz 0.4 A, 400 Hz 19.56 20.44 1 mv/a 400 A, 50 Hz 8.0 A, 50 Hz 391.96 408.04 1 mv/a 400 A, 100 Hz 8.0 A, 100 Hz 391.96 408.04 10 mv/a 2 A, 50 Hz 40 ma, 50 Hz 19.45 20.55 10 mv/a 2 A, 400 Hz 40 ma, 400 Hz 19.45 20.55 10 mv/a 40 A, 50 Hz 0.8 A, 50 Hz 391.85 408.15 10 mv/a 40 A, 100 Hz 0.8 A, 100 Hz 391.85 408.15 Figure 2. Current Accuracy Test Connections bbc01f.eps 6

Calibration Adjustment 1. Using the equipment connections shown in Figure 2, set the 5520A output for 2 A, 45 Hz. AC Current Clamp Calibration Adjustment 2. Referring to Figure 3, slide the rubber boot on the UUT handle back to access the adjustment point, VR1. 3. Using a small flat-head screwdriver, adjust VR1 for a meter reading of 99.9 mv to 100.1 mv. VR1 Figure 3. i400s Calibration Point bbc02f.eps User-Replaceable Parts The Current Clamp contains no user-replaceable parts. The following user documentation is available for this product. Description Part No Qty i400s Instruction Sheet 2282671 1 i400s (this document) 2414327 1 7

i400s LIMITED WARRANTY AND LIMITATION OF LIABILITY Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is one year and begins on the date of shipment. Parts, product repairs, and services are warranted for 90 days. This warranty extends only to the original buyer or end-user customer of a Fluke authorized reseller, and does not apply to fuses, disposable batteries, or to any product which, in Fluke's opinion, has been misused, altered, neglected, contaminated, or damaged by accident or abnormal conditions of operation or handling. Fluke warrants that software will operate substantially in accordance with its functional specifications for 90 days and that it has been properly recorded on non-defective media. Fluke does not warrant that software will be error free or operate without interruption. Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price. Fluke reserves the right to invoice Buyer for importation costs of repair/replacement parts when product purchased in one country is submitted for repair in another country. Fluke's warranty obligation is limited, at Fluke's option, to refund of the purchase price, free of charge repair, or replacement of a defective product which is returned to a Fluke authorized service center within the warranty period. To obtain warranty service, contact your nearest Fluke authorized service center to obtain return authorization information, then send the product to that service center, with a description of the difficulty, postage and insurance prepaid (FOB Destination). Fluke assumes no risk for damage in transit. Following warranty repair, the product will be returned to Buyer, transportation prepaid (FOB Destination). If Fluke determines that failure was caused by neglect, misuse, contamination, alteration, accident, or abnormal condition of operation or handling, including overvoltage failures caused by use outside the product s specified rating, or normal wear and tear of mechanical components, Fluke will provide an estimate of repair costs and obtain authorization before commencing the work. Following repair, the product will be returned to the Buyer transportation prepaid and the Buyer will be billed for the repair and return transportation charges (FOB Shipping Point). THIS WARRANTY IS BUYER'S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL, OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THEORY. Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this Warranty is held invalid or unenforceable by a court or other decision-maker of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision. Fluke Corporation P.O. Box 9090 Everett, WA 98206-9090 U.S.A. Fluke Europe B.V. P.O. Box 1186 5602 BD Eindhoven The Netherlands 11/99 To register your product online, visit register.fluke.com 8