Metrohm. 743 Rancimat. Determination of the oxidation stability of Oils and fats Biodiesel Food and cosmetics

Similar documents
measuring vessel air inlet conductivity cell reaction vessel sample absorption solution (water) heating block (e.g. 120 C)

RSSOT New Standards ASTM D 7525 ASTM D 7545 ASTM D 7525 ASTM D Manufacturer since

Instruments for the Analysis of Petroleum Products. Oxidation Stability

This presentation focuses on Biodiesel, scientifically called FAME (Fatty Acid Methyl Ester); a fuel different in either perspective.

Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers. Versa HT3. Application Note. Abstract.

V.Venkatakranthi Teja. N S Raju Institute of Technology (NSRIT), Sontyam, Visakhapatnam, Andhra Pradesh , India.

Stabinger Viscometer Series. SVM Series

Biodiesel Process Unit EBDB

The preparation of biodiesel from rape seed oil or other suitable vegetable oils

Biodiesel is NOT raw vegetable oil or SVO (Straight Vegetable Oil) or refined oil or filtered used cooking oil.

Callisto 100. Cold Filter Plugging Point Tester

ADU 5. Automatic Distillation Unit

Pursuing high precision, speed, and ease of use

New Energy Activity. Background:

Free and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN and ASTM D6584

Production of Biodiesel from Used Groundnut Oil from Bosso Market, Minna, Niger State, Nigeria

What s s in your Tank?

The storage stability issue of Biodiesel and the additives solution. Long - Term Storage of Fuels Seminar Israeli Institute of Energy and Environment

Alternative Fuel Vehicle Quiz Questions

Background on Biodiesel

POLLUTION CONTROL AND INCREASING EFFICIENCY OF DIESEL ENGINE USING BIODIESEL

Hydrocarbons 1 of 29 Boardworks Ltd 2016

The Discussion of this exercise covers the following points:

Liquid Handling Station

The Analysis of Biodiesel for Trace Metals and the Development of Certified Biodiesel Standards

ISO 660 INTERNATIONAL STANDARD. Animal and vegetable fats and oils Determination of acid value and acidity

Stabinger Viscometer SVM 3001

MB3600-CH30 Laboratory FT-NIR analyzer for biodiesel applications Suitable for production optimization and product quality assessment

Edexcel GCSE Chemistry. Topic 8: Fuels and Earth science. Fuels. Notes.

Where We Are. Today: Finish up Chapter 4, hopefully! Discussion: Alternative fuels, the benefits of conservation Where to go next?

PMA 5. Pensky-Martens Flash Point Tester. ::: Volatility... Flammability

Determination of Free and Total Glycerin in B100 Biodiesel

POTENTIAL OF BIOGAS GENERATION THROUGH CO-DIGESTION OF MSW, SEWAGE SLUDGE AND GLYCEROL

Biodiesel from Algae: Challanges, oppurtunuties and the way forward

Solution Sheet 1.

Q1. Useful fuels can be produced from crude oil. Crude oil is a mixture of hydrocarbons.

Canadian Canola Growers Association. Ernie Doerksen, General Manager phone: (204)

ASTM D Standard Specification for Biodiesel Fuel (B 100) Blend Stock for Distillate Fuels

Biodiesel. As fossil fuels become increasingly expensive to extract and produce, bio-diesel is

Vauhkonen Ville**, Hiltunen Erkki*, Niemi Seppo*, Pasila Antti**, Salminen Heikki*, Lehtonen Jari*, Ventin Mikael*** and Nummela Ilona****

Shut-off valve as a discharging valve when extracting glycerin from the waste products compound

Project Reference No.: 40S_B_MTECH_007

PROJECT REFERENCE NO.: 39S_R_MTECH_1508

Impact of Biodiesel Fuel on Engine Parts

Low cost EDXRF elemental analyzer

Bomb Calorimetry and Viscometry: What Properties Make a Good Fuel?

ecotrans Lf 01/02 Microprocessor Transmitter / Switching Device for Conductivity

812 Valve Unit Instructions for Use / dö. CH-9101 Herisau/Schweiz Internet

Alternative Automotive Fuels - Biodiesel

Determining the Ethanol Content of Denatured Fuel Ethanol Using Near Infrared. Gulf Coast Conference Patrick Ritz PAC LP

Q1. The table shows how much carbon dioxide is produced when you transfer the same amount of energy by burning coal, gas and oil.

White Paper. Improving Accuracy and Precision in Crude Oil Boiling Point Distribution Analysis. Introduction. Background Information

Innovation in cold behavior

Stirring Motors from CAT. Ingenieurbüro M. Zipperer GmbH. Our development makes your success. High power output up to 600 Ncm. Fuzzylogic control

FEATURE ARTICLE. Advanced Function Analyzers: Real-time Measurement of Particulate Matter Using Flame Ionization Detectors. Hirokazu Fukushima

AVL Particle Measurement System Aviation

PERFORMANCE AND EMISSION TEST OF CANOLA AND NEEM BIO-OIL BLEND WITH DIESEL

The table below gives information about milk bottles. Raw materials Sand, limestone, salt Crude oil. Bottle material Soda-lime glass HD poly(ethene)

SKF Grease injection system

Biodiesel Production and Analysis

Heating chamber with mechanical convection

833 IC Liquid Handling Unit

Heating chamber with mechanical convection

Expert Group on Maritime Policy

Fuels are materials that are used to create energy. They may be

Model ED 720 Drying and heating chambers Classic.Line with natural convection

Mandate to CEN on the revision of EN 590 to increase the concentration of FAME and FAEE to 10% v/v

973 Dew Point Mirror. Portable Industrial Chilled Mirror Hygrometer. Typical applications: REFLECTING YOUR STANDARDS

8/3/2012 SIF: Energy School 2012,Varenna. Omar Said

7.09 ULTRASONIC ANALYZER

Combustion Calorimeter System

Penetration. Test Methods

Safety drying ovens with expanded temperature range

RALF At a glance. Bioengineering RALF essentials. General

Model ED 400 Drying and heating chambers Classic.Line with natural convection

1. Exclusion list: Banned products ( )

Safety drying oven with expended temperature range

Heating chamber with mechanical convection

Heating chamber with mechanical convection

Hydrogen Station Equipment Performance Device (HyStEP Device) Specification

Safety drying ovens. Advantages: Areas of application: FDL series 115

Nexa 1200 Fuel Cell System

Closed Cup Flash Point Tester

Infinite F50 family INNOVATIVE EIGHT CHANNEL ABSORBANCE READERS WITH ADVANCED LED TECHNOLOGY. 21 CFR Part 11* 98/79/EC IVD-D*

Hot-film Air-mass Meter HFM 6

Biodiesel Production and Analysis

Synthetic Gear Lubricants Go Green

The Use of Microalgae Biodiesel in Diesel Engine : Production, Extraction and Engine Performance Assoc. Professor Dr. T. F. Yusaf Saddam H Al-lwayzy

Biodiesel and Renewable Fuels

FLOTTWEG SEPARATION TECHNOLOGY FOR THE PRODUCTION OF BIODIESEL

Fuel Related Definitions

SYNTHESIS OF BIODIESEL

Emission Analysis Of The Biodiesel From Papaya And Chicken Blends

This document is a preview generated by EVS

PRODUCT INFORMATION SHEET

FORCE TENSIOMETER K20 THE STAND-ALONE TENSIOMETER FOR QUALITY CONTROL

BIOGAS PRODUCTION ENHANCEMENT BY USING GLYCERINE AS CO SUBSTRATE

Bioengineering RALF Autoclavable Laboratory Bioreactors / Fermentors with glass vessels. Bioengineering RALF Basic + Advanced

Instruments for Chemical and Petroleum Products. Flash Point

Metrohm Preventive Maintenance

Transcription:

Ion Metrohm analysis 743 Rancimat Determination of the oxidation stability of Oils and fats Biodiesel Food and cosmetics

The 743 Rancimat in brief The 743 Rancimat is a modern, PC-controlled analytical instrument for the comfortable determination of the oxidation stability of oils and fats, also referred to as Oil Stability Index (OSI). The software of the third Rancimat generation comprises automatic data acquisition and evaluation as well as a database and permits the convenient management of large amounts of data. The unique temperature extrapolation allows a rough estimation of the storage stability of a product. Disposable reaction vessels reduce the amount of cleaning of the accessories to an absolute minimum. This saves both time and money. The accuracy and reproducibility of the measurements are also significantly improved. The GLP Set with certified temperature sensor allows to set the temperature very accurately and reproducibly, which in turn improves the reproducibility of the results. The most important applications Oxidation stability of natural oils and fats Oxidation stability of biodiesel (fatty acid methyl esters, FAME) Oxidation stability of food and cosmetics 2

The most important advantages at a glance All instrument functions controlled from PC Unparalleled reliability and simple operation due to unique accessories - Favorably-priced disposable glass parts - Robust conductivity cell with integral electrical connections in cover Excellent data security by storing all measuring, method and instrument data Re-evaluation and recalculation of all measuring data Database functions such as filtering and sorting User administration with freely configurable access rights Convenient user interface GLP and validation functions Estimation of the storage stability of products Compact dimensions and low weight 2 heating blocks with 8 measuring positions per instrument. Up to 4 instruments can be connected to one PC. Independent start of each channel 3

Oxidation stability of natural oils and fats The determination of the oxidation stability of oils and fats is the classical application for the 743 Rancimat. The resulting induction time characterizes the resistance of oils and fats to oxidation. In addition to the term «induction time», the expression Oil Stability Index (OSI) is in common use. The Rancimat method is also called the automated Swift test or accelerated oxidation test and is in general use today as an automated version of the previously used and extremely complicated Active Oxygen Method (AOM). During the measurement a stream of air is passed through the oil or fat sample contained in a sealed and heated reaction vessel. This treatment results in oxidation of the oil or fat molecules in the sample, with peroxides initially being formed as the primary oxidation products. After some time the fatty acids are completely destroyed; the secondary oxidation products formed include low-molecular organic acids in addition to other volatile organic compounds. These are transported in the stream of air to a second vessel containing distilled water. The conductivity in this vessel is recorded continuously. The organic acids can be detected by the increase in conductivity. The time that elapses until these secondary reaction products appear is known as the induction time, induction period or Oil Stability Index (OSI). The Rancimat method yields standard quality control parameters for the production of oils and fats in the food industry or for incoming goods checks in further processing. In addition to oils and fats from vegetable sources, the oxidation stability of animal fats such as lard, tallow and fish oil can also be checked with the 743 Rancimat as a matter of course. Foodstuffs may contain added antioxidants in order to delay the oxidative decomposition of their oils and fats. With the 743 Rancimat it is possible to characterize the effectiveness of antioxidants. This opens up a further field of application for the instrument. 4

Standards The Rancimat method is included in various national and internationals standards, such as: AOCS Cd 12b-92 (AOCS American Oil Chemists Society) Sampling and analysis of commercial fats and oils: Oil Stability Index ISO 6886 Animal and vegetable fats and oils Determination of oxidation stability (accelerated oxidation test) 2.4.28.2-93 Fat stability test on Autoxidation. CDM, Japan Swiss Food Manual (Schweizerisches Lebensmittelbuch), section 7.5.4 Estimating the storage stability An extension of the previous possibilities is provided by a new function the so-called temperature extrapolation. This is an aid for estimating the shelf life of oils and fats. The extrapolation makes use of the relationship between the measured induction time and the temperature given by van t Hoff s law. Several measurements are made at different temperatures and then extrapolated to the storage temperature. The value obtained allows a rough estimation of the storage stability of the oil or fat. 5

Oxidation stability of biodiesel (fatty acid methyl esters, FAME) In addition to other alternative fuels such as ethanol, methanol or biogas (methane), fatty acid methyl esters are increasingly being found on the market; these are also known as biodiesel or FAME (fatty acid methyl esters). Fatty acid methyl esters are usually obtained from oil seeds and are mainly used in their pure form or mixed with conventional diesel fuel in the transport sector. During manufacture the vegetable oil is transesterified with methanol. This produces the methyl esters of the fatty acids present in the oil together with glycerol as a byproduct. Fatty acid methyl esters are relatively unstable on storage, as like all natural oils and fats they are slowly oxidized by atmospheric oxygen. The substances produced in this way could cause motor damage. This is why the oxidation stability is an important quality criterion for biodiesel; it is regularly determined during the production process. With the 743 Rancimat this determination can be carried out quickly and simply. The oxidation process can be delayed by the addition of antioxidants. The 743 Rancimat can also be used to determine the effectiveness of the antioxidants. The use of alternative fuels from renewable vegetable sources has become increasingly widespread in recent years, as the life cycle of biogenic fuels is neutral with respect to CO 2 release. The carbon dioxide released has already been assimilated from the atmosphere during plant growth. In contrast to fossil fuels, biogenic fuels neither contribute to the accumulation of CO 2 in the atmosphere nor to global warming. In addition, alternative fuels are biodegradable and thus less environmentally harmful. The oxidation stability of fatty acid methyl esters has already been included in various test methods as a standard parameter used to define the minimum quality requirements of biodiesel: EN 14214 «Automotive fuels Fatty acid methyl esters (FAME) for diesel engines Requirements and test methods» EN 14112 «Fat and oil derivatives Fatty acid methyl esters (FAME) Determination of oxidation stability (accelerated oxidation test)» 6

7

Oxidation stability of foods and cosmetics Just like the pure substances, the oils and fats contained in foods are also subject to oxidation, which contributes to their spoilage. In such cases the 743 Rancimat can be used to determine the oxidation stability of foods containing oils and fats. Meltable foods with a high fat content, such as butter, margarine, lard or tallow, can be analyzed directly without any further sample preparation. For liquid or semi-liquid foods, such as salad dressings or mayonnaise, it is better to split the emulsion and analyze the separated fat phase. For solid, non-meltable foods it is also necessary to separate off the fat phase. In this case the fat is normally cold-extracted with petroleum ether and the isolated fat is then analyzed. Examples of typical applications are: Mayonnaise Salad dressings Biscuits Waffles Grain flakes Nuts Chocolate Bacon Sausages As well as for determining the oxidation stability of foods, the 743 Rancimat is also used for testing cosmetics or cosmetic additives. With meltable products such as lipsticks this can be done without any further sample preparation. With creams or salves the fat phase must normally be separated off before the analysis. 8

9

The 743 Rancimat in detail Two heating blocks with two temperatures The 743 Rancimat has two independent heating blocks that allow up to eight samples to be analyzed at one or two temperatures. Up to four Rancimats can be connected to one PC, so that the maximum number of samples that can be analyzed in parallel can be increased to 32. Each measuring position can be started individually. As soon as the measurement has been completed the measuring position is immediately ready for a new sample, which means that the instrument can be used to its full capacity. 10

Simple reaction vial handling Favorably-priced disposable reaction vessels are used with the 743 Rancimat. Weighing out the sample and assembling the reaction vessel are extremely simple and safe. As the vessel does not need to be expensively cleaned at the end of the measurement this means that the personnel are available for other tasks. This reduces the analysis costs. Measuring vessels Easy-to-clean polycarbonate beakers are used for the automatic conductivity measurement. Glass beakers are available as an alternative. Cover with built-in conductivity cell The conductivity cell is incorporated in the measuring vessel cover. When the cover is placed in position the cell is immersed in the water. At the same time electrical contact is made to the electronics in the instrument. The use of fragile glass conductivity electrodes with lengthy connecting cables went out of fashion a long time ago. The new conductivity cell is also very easy to clean. Connections In order to make operation as simple as possible there are no controls at all on the instrument. All its functions are controlled from the PC. Apart from the power switch, the only features you will find on the instrument are the RS-232 socket for connection to a PC and a socket for connecting the Pt-100 temperature sensor. Validation with the GLP Set The optionally available GLP Set facilitates the validation of your 743 Rancimat. It contains a certified Pt-100 temperature sensor with accessories that can be used for testing the temperature regulation of the heating block. A test plug for checking the conductivity measurement inputs is also supplied. Air inlet filter and molecular sieve The air used for the measurement is aspirated through a filter that prevents particles from entering the instrument. The molecular sieve removes water vapor from the aspirated air; as water contributes to the hydrolytic decomposition of the fat molecules, it could interfere with the measurement. Air supply line The amount of air that passes through the sample is automatically controlled via the rotation rate of the built-in pump according to the method settings. A separate supply of compressed air is not necessary. 11

Software functions All functions of the 743 Rancimat are controlled by the Rancimat software, which excels by its user-friendliness. All the functions are clearly arranged in just a few windows, the operation is intuitive. Rancimat control This is where the measuring parameters can be called up and edited. The instrument functions are controlled directly from here, the measurements are also started and shown in the live display field. The arrangement of this window corresponds to a view of the instrument from above. This means that the assignment of sample information and measuring position is perfectly clear. The timer function can be used to automatically switch on the heating blocks before the start of work, so that it is no longer necessary to wait while they warm up. The functions in a nutshell: Individual start/stop for each position Live display Temperature display Method definition Instrument controls Calculation formulas, automatic result transformation to other temperatures Timer function Rancimat results At the end of each determination the measuring data is stored in a database and can be viewed by the user in the results window. Sample information and results are shown in tabular form and can be exported in various formats. The measured curves can be shown individually or in groups. It is also possible to edit the automatic evaluation and recalculate the results. The temperature extrapolation function for estimating the storage stability is available in this section of the software. You can, of course, sort or filter all the displayed data and adapt the display to meet your own requirements. Overview table Curve display: individual or multiple plots Re-evaluation: induction time, stability time and manual tangent method Report printout Temperature extrapolation (estimation of storage stability) Database functions: filtering, sorting Data export 12

GLP functions By using the built-in monitoring of the validation intervals, nothing can go wrong with your quality management procedure. The software requests you to carry out the validation in good time. The wizards for temperature and conductivity validation help you to do this. Monitoring the validation intervals Wizards for temperature and conductivity Electronic validation logbook User administration The software includes a flexible and comfortable user administration with login function; this can be used to define access rights for groups of users and individuals in great detail. Definition of user access rights Login function 13

Technical specifications Heating blocks 2 aluminum heating blocks; electrically heated; can be set to different temperatures Number of samples 8 sample (4 measuring positions per heating block) Temperature control and measurement Temperature range 50...220 C, adjustable in 1 C steps Temperature correction 9.9...+9.9 C, adjustable in 0.1 C steps Reproducibility of set temperature Typically better than ±0.2 C* Temperature variation Typically <0.1 C* Temperature difference between different measuring positions Typically <0.3 C* Instrument heating-up time from 20 C to 120 C Approx. 45 min (to ±0.1 C temperature stability) to 220 C Approx. 60 min (to ±0.1 C temperature stability) Outer temperature of instrument <50 C (at an operating temperature of 220 C) Response temperature of thermal protection device 260 C * When operating temperature has been reached, with inserted reaction vessels with an identical filling and 20 L/h air throughput. Air throughput Pump Output range Conductivity measurement Measuring range Electrodes Diaphragm pump 7...25 L/h 0...400 μs/cm 6.0913.130 conductivity cell with double steel-pin electrode built into vessel cover Ambient temperature Nominal working range +5...+40 C (at 20...80% relative humidity) Storage 20...+70 C Transport -40...+70 C Line power Voltage 2.743.0014 / 2.743.0114: 230 V (220...240 V ±10%) 2.743.0015 / 2.743.0115: 115 V (100...120 V ±10%) Frequency 50...60 Hz Power consumption <450 VA (depending on heating power) Dimensions Width Height Depth 405 mm 268 mm (without accessories) 353 mm (with accessories) 466 mm Weight 27.6 kg (with accessories) PC requirements Processor Operating system Memory RAM working memory Graphics resolution Interface Printer Pentium III with 700 MHz or higher Windows TM NT, Windows TM 2000 or Windows TM XP 20 MB for program files, 200 MB recommended for measuring data storage 128 MB, recommended 256 MB or higher (particularly for Windows TM XP) min. 800 x 600, recommended 1024 x 768 or higher 1 free RS-232C interface (COM) All printers supported by Windows TM 14

Ordering information 743 Rancimat Instrument for the automatic determination of the oxidation stability of natural oils and fats with conductometric indication. Including Rancimat software. 2.743.0014 743 Rancimat (230 V, 50...60 Hz) for oils and fats 2.743.0015 743 Rancimat (115 V, 50...60 Hz) for oils and fats Instrument for the automatic determination of the oxidation stability of biodiesel with conductometric indication. Including Rancimat software. 2.743.0114 743 Rancimat (230 V, 50...60 Hz) for biodiesel 2.743.0115 743 Rancimat (115 V, 50...60 Hz) for biodiesel Options 6.2059.000 Turntable for 743 Rancimat 6.1428.020 Glass conductivity measuring vessel for 743 Rancimat (8 pcs. are required) 6.2757.000 Exhaust collection tube for 743 Rancimat (with 8 stoppers) 6.1816.010 Silicone tubing, i.d. 6 mm, o.d. 9 mm; length 22 cm (8 pcs. are required) 6.5616.000 GLP Test Set for 743 Rancimat; consisting of 6.1111.010 Pt-100 temperature sensor 16 cm 6.1253.000 Measuring insert for heating block 6.2109.030 Test plug 10 kohm 6.2301.060 KCl conductivity standard, c(kcl) = 0.1 mol/l, 250 ml Consumables 6.1429.040 Reaction vessels for 743 Rancimat (117 pcs.) 6.2418.100 Air inlet tube for oil measurements (117 pcs.) 6.1428.100 Polycarbonate conductivity measuring vessel for 743 Rancimat 6.1816.010 Silicone tubing for fat/oil measurements, i.d. 6 mm, o.d. 9 mm; length 22 cm 6.1839.000 Iso-Versinic tubing for biodiesel measurements, i.d. 6 mm, o.d. 9 mm; length 22 cm 6.2724.010 Dust filter for 743 Rancimat 6.2811.000 Molecular sieve 250 g 15

Ion analysis Metrohm Ltd. CH-9101 Herisau/Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 www.metrohm.com info@metrohm.com Subject to modifications Layout by Ecknauer+Schoch ASW, printed in Switzerland by Metrohm Ltd., Herisau 8.743.6013 2006-03