OD DIGITAL DISSOLVED OXYGEN CONTROLLER POLAROGRAPHIC CELL

Similar documents
OD DISSOLVED OXYGEN METER

OD OD DIGITAL DISSOLVED OXYGEN CONTROLLERS

OD 8382 OPTICAL DISSOLVED OXYGEN AUTOCLEAN

Dissolved Oxygen Measurement System with Air Blast Cleaner

Operating manual. ph / mv controller type M8832N

5001TCP SPEED CONTROLLER

5001TCP SPEED CONTROLLER

R & D SPECIALTIES ROTROL I USER'S MANUAL

TIME OF FLIGHT Components from:

TIME OF FLIGHT Components from:

Combined Ventilation Controller RVWS-T-224HA

2.0 CONSTRUCTION 3.0 OPERATION. SA-1 Generator Differential Relay - Class 1E 2.5 TRIP CIRCUIT

ProMinent. ProMinent D4a Metering Pumps with built-in proportional controllers

RESISTIVITY MONITOR/CONTROLLERS

3. OPERATION 2.1. RESTRAINT CIRCUIT 2.6. INDICATING CIRCUIT 2.2. OPERATING CIRCUIT 2.7. SURGE PROTECTION CIRCUIT 2.3.

WalkLAB Digital Dissolved Oxygen meter HD9030

HI 8510-HI 8512 HI 8710-HI 8711 HI 8720-HI HI HI

DISSOLVED OXYGEN MEASUREMENT

TECHNICAL DESCRIPTION

Time Electronics DC Voltage Calibrator. Technical Manual

Installation and Maintenance Instructions. World Leader in Modular Torque Limiters. PTM-4 Load Monitor

Bante220 Portable ph Meter Instruction Manual

BATTERY OPERATED DIGITAL MILLI-OHMMETER

SB 2000 PUSH TO SEARCH NEXT STAG E. Aerotech, Inc. FORM: QM 1320

Model DS1000/DT20 System

ABB MEASUREMENT & ANALYTICS DATA SHEET Hydrazine monitor

Compact System NRGS 11-2 NRGS Original Installation Instructions English

Application Note. Monitoring Bearing Temperature with ProPAC

MU MMMREV.2 09/01 UNI EN ISO ELECTRONIC POWER FACTOR REGULATOR type QR INSTRUCTION MANUAL PAG. 7

Harris IRT Enterprises Digital Resistance Tester Model XP

METROLOGIC INSTRUMENTS, INC. MX001 Industrial Control Interface Installation and User s Guide

Dissolved Oxygen, ph, and Conductivity

THUNDERBIRD. Agricultural Weighing Systems SS1000 Instructions

User Manual. T6 Tachometer. Online: Telephone: P.O. Box St. Petersburg, Florida 33736

BAPI-Stat 4 Room Humidity Sensor (BA/ B4 -H200 Series)

ENGINE GOVERNING SYSTEMS LSM672 LOAD SHARING MODULE. GOVERNORS AMERICA CORP. 720 Silver Street Agawam, MA , USA MEMBER

Installation Instructions Spector compact Conductivity Transmitter System LRGT 16-1

RS-3 PRO RS-1007 PRO. CAT IV Analog Clamp meter Series. Users Manual. For detailed specifications and ordering info go to

FC/FCA 12, 24, 32 & 48 VOLT, 6 & 10 AMP BATTERY CHARGER OPERATION & MAINTENANCE GUIDE

ZMT & ZFG2 Zirconia Oxygen Analyzer Systems. Superior technology and quality from the world leader in oxygen measurement

INSTALLATION MANUAL & OPERATING INSTRUCTIONS

MODEL 422 Submersible Pump Controller

P200 P/I Transducer. Installation, Operation, and Maintenance Instructions INTRODUCTION

Series 20 Installation Instructions

Temperature Controller. TC5+2V4SA Plus USER'S MANUAL

Inlet Controller TC5-ITA USER'S MANUAL. M rev. 02 K rev. 00

INSTRUCTION MANUAL 272-5X5 ANALOG TRANSMITTER (210 SERIES FLOW METERS) 272-5X7 ANALOG TRANSMITTER (220/240 SERIES FLOW METERS)

Battery Power Inverters

INDEX Section Page Number Remarks

Measuring Cells for free chlorine, chlorine dioxide, ozone

Transmitter CSA-591 Instruction Manual

AS4V DC ALARM BEFORE USE... CAUTION WARNING INSTRUCTION MANUAL AS4V MODEL. (dual or quad alarm trip; field-configurable)

Eclipse Solar Suitcase

AUTOMATIC BATTERY CHARGER WITH CONSTANT VOLTAGE AND CURRENT

METROTEC A15-N. Oxygen Sensor Type A15-N. *** Version 1.5 ***

Model Q45H/ Residual Chlorine Monitor...

Time Electronics. Calibration, Test and Measurement. Portable Process and Pressure Calibration Instruments

INDUCTIVE CONDUCTIVITY SENSOR. Instruction Manual. Bürkert 2001 Subject to technical change without notice

USING STANDARD x1 and x10 OSCILLOSCOPE PROBES WITH THE TEKTRONIX 222PS OSCILLOSCOPE:

Advantage-D. Operating Instructions and Maintenance Manual. Central Vacuum Systems (Expandable/Modular Models) (Ver.

Model 930 Power Control System Instruction Manual

The Temperature Control Company SERIES 2000 / 3020 ANALOG TEMPERATURE CONTROLLERS. Instruction Manual

INSTRUCTION MANUAL MODEL PS VDC POWER SUPPLY

BAPI-Stat 3 Room Sensor

OPERATING INSTRUCTIONS

Virtual Ground for HV Boosters Calibration

Temperature Controller OVATION 214 User's Guide

SCR Power Controllers

BOILER BLOWDOWN CONDUCTIVITY CONTROLLER (CONTINUOUS METHOD)

2002 F-150 Workshop Manual

Installation, Operation & Maintenance Manual. Bulletin: IOMS-PUL CHEMAlarm LEAK DETECTIONI

Atlas ESR. User Guide. Capacitance and Equivalent Series Resistance Meter. Model ESR60 (Enhanced)

PRODUCT INFORMATION BULLETIN

COLD CA THODE TUBE TEST SET J94731 A DESCRIPTION AND APPLICATION

Continental Hydraulics Installation Manual CEM-AA-A

Differential Pressure Gauges Cryo Gauge Model , Cu-alloy Model , Stainless Steel Series

SECTION DC POWER SUPPLY/BATTERY CHARGER

Assembly Instructions for the KA Electronics Flat MM Phono Preamplifier

Model 2053 Conductivity Meter Instruction Manual

CONTENTS TABLE OF CONTENTS... 1 INTRODUCTION... 2 SEC 1 - SPECIFICATIONS... 3 SEC 2 - DESCRIPTION... 5 SEC 3 - OPERATING INSTRUCTIONS...

MODEL 905V OPERATING INSTRUCTIONS

MODEL AMMONIA GAS SENSOR MODULE 4-20 ma

ecotrans Lf 01/02 Microprocessor Transmitter / Switching Device for Conductivity

DCT SERIES CASE SIZE 3A FILTERED BATTERY CHARGER 12V/200A 24V/ A 48V/100A 120V/50A 240V/25A OPERATION & MAINTENANCE GUIDE

Fincor Series 2230 MKII/2240

Using the LCT100 for Load Cell Troubleshooting

CAPACITIVE LEVEL CONTROLS SC - SCD SERIES

USER MANUAL FOR AUTOMATIC A/C CHARGING STATIONS

User s Manual Rev 1.3 GME

1333 (SERIES B & C) TROUBLESHOOTING GUIDE

Visit us at

amperometric (currentbased) cell coupled with a potentiometric

ARCHITECTURAL CONTROL SYSTEMS, INCORPORATED ST. LOUIS, MISSOURI

EDDY CURRENT VARIABLE SPEED DRIVE CONTROLLER - TYPE PM-07

TABLE OF CONTENTS. 8.3 Online Sensor Specifications. Series TS (Models TS1, TSP, TSH, TSL,TSF, TSB1, TSB2) According to Electromatic factory procedure

User Instructions for WRC1021D. Doc

FurnaceDoctor -DPT Dewpoint Analyzer. User Manual Version 008

Boston Gear Ratiotrol

Technical Description

(typ.) (Range) Input Specifications Parameter Model Min. Typ. Max. Unit 12V Input Models Input Surge Voltage (100ms.

Transcription:

UNI EN ISO 9001:2000 CERT. N. 9115 BCEL INSTRUCTION MANUAL OD 265.2 DIGITAL DISSOLVED OXYGEN CONTROLLER POLAROGRAPHIC CELL Rev. B - 03/07 B&C Electronics Srl - Via per Villanova 3-20040 Carnate (Mi) - Italy - P.IVA 00729030965 Tel +39 039 63 1721 - Fax +39 039 607 6099 bc@bc-electronics.it - www.bc-electronics.it

Index 1 GENERAL INFORMATION... 3 2 PRINCIPLES... 3 3 FUNCTIONAL DESCRIPTION... 4 4 PHYSICAL DESCRIPTION... 5 5 SPECIFICATIONS... 6 6 ELECTRICAL INSTALLATION... 7 6.1 Connecting the power... 7 6.2 Warnings... 7 6.3 Connecting the cell... 7 6.4 Connecting alarms, pumps, valves... 7 6.5 Connecting a recorder... 8 6.6 Connecting the RTD... 8 6.7 Checking... 8 7 OPERATING THE SYSTEM... 9 7.1 Pre-operation check... 9 8 NORMAL OPERATION... 10 9 CALIBRATING THE CONTROLLER... 11 9.1 Electric zero... 11 9.2 Zero cell calibration... 11 9.3 Slope calibration... 11 9.4 Polarization voltage of sensor... 11 10 TEMPERATURE COMPENSATION... 12 10.1 Manual... 12 10.2 Automatic... 12 11 PREVENTIVE MAINTENANCE... 13 11.1 Controller... 13 11.2 Sensor... 13 12 TROUBLESHOOTING GUIDE... 14 Instruction Manual - Rev. B - 03/07-2 -

1 GENERAL INFORMATION This manual applies to the OD 265.2 digital bench controller. It explains the purpose of the equipment, describes the components of the series and the procedures for installing the equipment, operating it and calibrating it. Some suggestions are also given for its maintenance. 2 PRINCIPLES In a polarographic cell, if a constant Voltage is applied to the electrodes, the reduction of the Oxygen ions on the cathode generates a current proportional to the oxygen's partial Pressure. The intensity of the current produced at a constant Temperature and at a given Oxygen Pressure depends on the quantity and composition of the electrolyte, the cathode's area and the sample's state. Response time depends on the electrolyte's and diaphragm s characteristics. The value of the applied Voltage to the anode is chosen in a way that only Oxygen can react. In absence of Oxygen, the current produced is not exactly zero. In fact, impurities and other factors generate a "dark current" which can be offset by the amplifier. The electrode consumes the Oxygen diffused through the permeable membrane and the electrolyte's film, therefore a concentration gradient is established between cathode and sample; the electrodes with a big cathode and with a high output current, consume a lot of Oxygen and are more sensitive to the sample's flow and therefore request a large flow and an adequate stirring of the sample. The Temperature dependence of the output current is compensated by the automatic or manual Temperature compensation of the meter. Instruction Manual - Rev. B - 03/07-3 -

3 FUNCTIONAL DESCRIPTION This series of bench controllers is specifically designed for automatic activation of metering pumps and valves in fermentation processes. The controller provides a digital readout of the dissolved Oxygen in aqueous solutions. In order to satisfy the many requests of such measurements, four different ranges have been designed and may be selected on the meter: - % Oxygen saturation - % air saturation - mm Hg - mg/l (ppm) RTD automatic Temperature compensation and manual compensation are employed to provide highly accurate measurements over the entire ranges of dissolved Oxygen and Temperature values. The Temperature value and the Temperature coefficient value are both visualized on the display by pushing the corresponding lever on the front panel. A basic O2 monitoring system consists of two elements: a monitoring unit, and a probe or O2 cell. The system can be expanded by adding accessories for field applications; recorders, secondary regulators, proportional regulators, special probes and Temperature probes to suit particular applications. The controller contains electronic circuits to control the operations of the entire system. It provides a digital readout of Oxygen on a 3 1/2 digit display. There is included an internal circuitry for controlling two external alarms, valves, pumps etc. Set-points are independent and programmed by a front-panel control to trigger at any level within the range of the meter. Red Leds on the panel indicate if the measured value is below the set point A level, or if it is above the set point B level. Red Leds also indicate the switching of corresponding relays. The set points have the feature of being inverted and the two regulators have a delay action within 5 seconds by means of a rear trimmer The set point value is visualized by pushing the button 6 on the front panel and the switching function of the relay is inverted by switches on the rear side of the meter. When the relay is switched on, an AC Voltage is available on the rear outlet for driving pumps and valves. The controller provides an output of 0/20 ma or 4/20 ma (selected by a switch on the rear panel) proportional to the meter reading, for driving a recorder, remote readout or regulators having a not grounded input. Zero and slope are calibrated by trimmers on the front panel. Automatic Temperature compensation is carried out by means of a RTD Pt 100. The meter's input is by a BNC connector for the electrodes and by a phone Jack for the RTD. The digital Temperature readout, both in the manual and the automatic Temperature operation, is obtained by pushing a lever 10 on front panel, as measuring and controlling functions of the instrument are not altered. The manual Temperature compensation is adjusted by the trimmer 11. The digital readout of Temperature coefficient ( % / C ), both in the manual and the automatic operation is obtained by pushing a lever 12, its value being adjusted by the trimmer 13. The scale is chosen by the selector 14 and the corresponding Led 15 shows which scale is operating. Instruction Manual - Rev. B - 03/07-4 -

4 PHYSICAL DESCRIPTION The controller's case consists of an anodized aluminium front panel and an epoxidic-coated metallic case, to ensure the maximum anticorrosion characteristics. Connections to the power supply, loads, recorders, RTD's, electrodes or probes are carried out by means of sockets or connectors on the rear side. Instruction Manual - Rev. B - 03/07-5 -

5 SPECIFICATIONS Voltage: Input: Output: Scales: 110 or 220 Volts (as requested) +/-10 % 50/60 Hz from polarographic O2 cell 0/20 ma dc, non isolated, 300 ohm max. 4/20 ma or others on request 0.0/199.9 % O2 0.0/199.9 % air saturation 0.0/199.9 mm Hg 0.0/19.99 ppm Zero: adjustment +/- 15% Slope: fine adjustment +/- 20% coarse adjust. 5 to 50 namp air saturation at 20 C Cell current in air: Cell Voltage Cell connector: Factory calibration: approx. 0.030 microamp. 675 mv BNC half scale with 0.030 microamp input. Temperature: 0/50 C Temperature compensation: manual and automatic 0/50 Celsius Temperature coefficient: 0 to 6 %/ C Temperature sensor: RTD Pt100 Temperature readout: -20/+120 C Regulators: Switching time: Regulator delay: Regulator output: Outlet fuse: Circuit fuse: +/- 0.25 % hysteresis (others as requested) < 0.5 Seconds 0 to 5 Seconds 220 V or 110 V (same as the power supply) 5 Amp. resistive load 2 Amp. T 80 ma T (110 V) internal 32 ma T (220 V) internal Power: Power fuse: Weight: Dimensions: 3 VA + power of the loads 5 A 2,140 g 215 x 95 x 200 mm Instruction Manual - Rev. B - 03/07-6 -

6 ELECTRICAL INSTALLATION The electrical installation consists of: - connecting the power supply to the meter; - connecting the cell or the probe to the meter; - connecting alarms, pumps, valves if necessary; - connecting the monitor output to the recorder or similar devices if required; - connecting other optional accessories (RTD P.I.D. regulators). 6.1 CONNECTING THE POWER - Connect the power cable to the connector on the rear panel; - the unit is factory-equipped with a 32 ma or 80 ma T internal fuse depending on the power Voltage required (80 ma T fuse if power Voltage is 110 V or 32 ma T fuse if power Voltage is 220 V); - in the power cable socket is installed a 5 A fuse; - before connecting the controller to primary power ensure that the line Voltage matches the Voltage requirement and the fuse value of the unit. 6.2 WARNINGS - Power the device by means of grounded wall outlet; - avoid mains Voltage from a branch point with heavy inductive loads; - control the mains Voltage value. 6.3 CONNECTING THE CELL - Cell cabling is a critical part of the whole system; - use a low noise coaxial cable on the overall length between the sensor and the input terminals of the meter; - avoid interruption of the cable if a coaxial connector and a high insulation terminal strip are not available; - keep the cable away from power wires on the overall length; - ask to your supplier for a cell with a BNC connector and follow the electrode manufacturer's instructions. 6.4 CONNECTING ALARMS, PUMPS, VALVES The output connections are made at sockets on the rear panel that are powered by line Voltage through the two independent relay contacts corresponding to Regulator "A" and Regulator "B". The output is protected by a fuse installed in each channel. Remember that relay is activated or deactivated according to the position of the switching lever marked min/max on the rear panel. The action may also be delayed from 0 to 5 Seconds by the trimmer marked DEL. (rear side) Instruction Manual - Rev. B - 03/07-7 -

When the lever is in the "min" position, the output is activated until the Oxygen values are below the set-point. When the lever is in position "max" the output is activated until the Oxygen values exceed the setpoint. 6.5 CONNECTING A RECORDER The recorder output is factory-set for the 0/20 ma mode. If the 4/20 mode is desired, select it using the lever marked 0-20 / 4-20 on the rear panel. Series connection is required for driving more loads having a total input resistance lower than 300 ohm. Output drives ground-isolated loads only. Connect the recorder's input to the controller's recorder output terminals marked " + " (Current out) and " - " (Current return) on the rear panel. 6.6 CONNECTING THE RTD The model OD 265.2 has the manual and the automatic Temperature compensation carried out by means of a RTD Pt100 device. The Temperature sensor has to be placed in the same solution being measured, close to the electrode in the pipe-line or in the tank. When the RTD is not connected, the jack socket on the rear side marked "ATC" Keeps in action the internal Temperature compensation value adjusted at 20 C (degree Celsius). This Temperature value may be visualized using the lever 10 and manually adjusted by trimmer 11. To operate the automatic Temperature compensation, plug the jack connector of the RTD in to the socket marked "ATC" A three wire connection is suggested to achieve an accurate compensation over long distances between the RTD and the controller. The RTD connection as described above allows the OD 265.2 controller to provide a digital readout of Temperature values. The Temperature value of a solution is shown by pushing the lever 10 under the display on the front panel. The readout of Temperature does not interrupt the measuring and regulating function of the controller. 6.7 CHECKING Before connecting the system to the power supply: - check that all cables are properly fastened to prevent strain on the connections - check that the fuse value is correct. Instruction Manual - Rev. B - 03/07-8 -

7 OPERATING THE SYSTEM 7.1 PRE-OPERATION CHECK The system's controls and indicators are all located on the front panel (see Fig.1). The meter has a digital display 1 that, together with the green Led 4, indicate that unit is on. Select the scale of the meter using the selector 14. The red Leds 15 will indicate the scale selected. Push the levers 10 and 12 in order to check the Temperature and Temperature coefficient values. Push the button 6 and rotate the set-point control 8. The display will show the set point values. The regulators have a red Leds 7 set-point check wich are part of the circuitry that powers the relay. When the monitored value is below the set-point value, the Reg A Led is alight and the corresponding relay is turned on. When the monitored value is above the set-point value, the Reg B Led is alight and the corresponding relay is turned on. The set-points have the feature of being inverted using the levers marked min/max on the rear panel. The controllers' cards are adjusted at the factory. If sensors and probes have been connected correctly, as described in the above sections, the system should function correctly needing only the zero calibration to compensate the "dark" current of the cell and the slope calibration. Instruction Manual - Rev. B - 03/07-9 -

8 NORMAL OPERATION To operate the system, simply switch on the controller by means of the lever 5 (power switch), and observe the measured dissolved Oxygen of the solution on the meter. Adjust the control A set-point and the control B set-point to the setting required for each particular application. Instruction Manual - Rev. B - 03/07-10 -

9 CALIBRATING THE CONTROLLER 9.1 ELECTRIC ZERO This operation may be done only when the cell is not connected to the amplifier. - remove the cell connections from the amplifier; - rotate the trimmer 2 marked "zero" in order to read the value 0.00 on the display; - repeat the calibration when the scale is changed. 9.2 ZERO CELL CALIBRATION This operation may be effected after the electric zero calibration as described above and when the cell is connected to the amplifier, in order to read zero on the indicator. This instrumental zero cancels out the residual current of the cell zero point when the cell works in absence of Oxygen. Immerse the cell in a freshly prepared 2 % Sodium Metabisulphyte solution or in an equivalent solution with a total absence of Oxygen and operate as follows: - after a balance is reached, adjust the trimmer 2 marked "zero" in order to read 000.0 on the display - repeat the calibration when the scale is changed. 9.3 SLOPE CALIBRATION Zero point adjustment must precede slope calibration. Electrode manufacturers supply, with the sensor, the operating instructions for each application. In particular, the slope calibration procedures under partial Pressure, in % of atmospheric Oxygen, in % of Oxygen and in ppm are described. Calibration is usually effected at an Oxygen-saturation level under operating Pressure since it is the simplest. To calibrate, the reading is adjusted to 100 % saturation. In determining Oxygen concentration in ppm, the O2 content of the calibrating solution must be accurately known. The values applying to pure water, adjusted to the prevailing Pressure, are described on Oxygen saturation tables. The controller is supplied with "coarse" and "fine" adjustment of the sensitivity. The coarse adjustment must be effected when installing the cell if necessary, while the fine adjustment is effected during the normal operation. The coarse adjustment is carried out by the rear panel trimmer marked "12", which compensates for the different sensitivity of the cells. When adjusting the coarse sensitivity, check that the "fine" trimmer 3 marked "sens" on the front panel is set in the middle position. 9.4 POLARIZATION VOLTAGE OF SENSOR The amplifier is designed to maintain a constant polarization Voltage to the electrode for its correct operating. This value is calibrated at 675 mvolts and can be adjusted, if necessary, from 0 to 1.200 mvolts by an internal trimmer marked "R 87". Use a mvoltmeter with high Resistance input to check the polarization Voltage between "( - )recorder output" and "BNC shield" Instruction Manual - Rev. B - 03/07-11 -

10 TEMPERATURE COMPENSATION 10.1 MANUAL The RTD is not connected. - Push the lever 10 in order to visualize the Temperature on the display; - adjust the Temperature value using the trimmer 11; - push the lever 12 in order to visualize the Temperature coefficient on the display; - adjust the coefficient value using the trimmer 13; - the value has to be 0.00 for non-compensated operation; - the Temperature coefficient is approximately 5.40 %/ C for measuring in ppm, approximately 2.30 %/ C for measuring in % O2, % air, mm Hg. More information is given by the cell manufacturer. 10.2 AUTOMATIC Connect the RTD to the phone jack connector marked "ATC" on the rear panel. - Push the lever 12 in order to visualize the Temperature coefficient on the display; - adjust the coefficient value using the trimmer 13; - the value has to be 0.00 for non-compensated operation; - the Temperature coefficient is approximately 5.40 %/ C for measuring in ppm, approximately 2.30 %/ C for measuring in % O2, % air, mm Hg. More information is given by the cell manufacturer; - push the lever 10 in order to visualize the Temperature value of the solution. Instruction Manual - Rev. B - 03/07-12 -

11 PREVENTIVE MAINTENANCE 11.1 CONTROLLER Quality components are used to give the controller a greater reliability. The Maintenance frequency depends on the nature of each particular application. As for any electronic equipment, the mechanical components, such as switches, relays, potentiometers and connectors, are the most subject to damage. - check for damage of the electrolytic capacitors if the meter is exposed to temperatures above 80 C;. - check for damage in all the electronic components if the meter is subjected to excessive Voltage; - check for damage of the electronic and mechanical components if the meter is dropped; - periodically repeat the pre-operation checklist; - check that all the connections are free from moisture and dirt. Disconnect the power supply to the monitor before performing the following procedures: - use moisture free air and blow out the interior of the case and terminal board connections as necessary; - inspect the printed circuit boards for dirt and corrosion; clean as necessary and blow dry - tighten all the terminal-board connections and mounting screws - replace the front panel circuit board or the base circuit board 11.2 SENSOR The condition of the cell's surface is critical for the normal operation of the system and should be inspected more frequently when using water containing oil and grease, and in bio-applications. See general instructions given by the cell manufacturer. Instruction Manual - Rev. B - 03/07-13 -

12 TROUBLESHOOTING GUIDE Symptoms Probable cause Remedy Neither Led lit nor meter reading No meter reading, but Leds light Meter reading too high/low Meter reading does not change Alarm circuit does not operate, meter reading OK Slope not sufficient Recorder does not operate -line not connected -incorrect power wiring -fuse interrupted -I.C. failure -inside connector -cell dirty -meter not calibrated -electrode damage -short circuit -outlet fuse -circuit failure -cell damage -temperature compensation -recorder not connected -output circuits damage -check power -check wiring -replace fuse -replace I.C. -replace circuit board -clean cell -adjust "sens" -electrode replacement -replace fuse -replace base -return to factory -cell refilling or replacement -check the T.coeff. or RTD -check wiring -replace base circuit board See the peculiar troubleshooting guide of each cell employed. Instruction Manual - Rev. B - 03/07-14 -

FRONT PANEL Fig.1 Instruction Manual - Rev. B - 03/07-15 -

REAR PANEL Fig. 2 Instruction Manual - Rev. B - 03/07-16 -

N O T E S Instruction Manual - Rev. B - 03/07-17 -