Productivity Improvements for Dual Column Applications: USP <467> and Blood Alcohol Analysis

Similar documents
Forensic Blood Alcohol Determination with the Intuvo 9000 GC

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application

Using the PSD for Backflushing on the Agilent 8890 GC System

Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593

Same Column and Gas Type Try Different/Faster Velocities. Same Column, Switch He to H2 Carrier Then Try Faster Velocities

Application. Gas Chromatography June 1995

Think GC/MS Couldn t Get Any Better? Think Again!

Detection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector

Application Note. Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D using AC OXYTRACER

High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph

Agilent and ASTM. Update on Recent Activities. Page 1

Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel

GC Inlets. An Overview. Simon Jones GC Applications Engineer

Technical Procedure for Gas Chromatography (GC-FID)

Oxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels

Fast GC. Dial for e-seminar Audio. Slide 4 SPEEDY GC

High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System

Troubleshooting Tips & Tricks for your GC Analyzer & CFT Application

Optimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System. Jim McCurry Roger Firor Agilent Technologies Wilmington, DE

Application Note. Abstract. Authors. Environmental Analysis

Refinery Gas. Analysis by Gas Chromatography WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL

Dual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector

Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)

ASTM D for Denatured Fuel Ethanol Automating Calculations and Reports with Empower 2 Software

Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011

GC Analysis of Total Fatty Acid Methyl Esters (FAME) and Methyl Linolenate in Biodiesel Using the Revised EN14103:2011 Method

Agilent Solutions for the Analysis of Ethanol- Based Fuels Derived From Biomass. James D. McCurry, Ph.D. Senior Scientist

For more information about Agilent Technologies services please visit our web site using the following URL

C2, C3, C4 Monomer Analysis

Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS)

Agilent 7693A Automated Liquid Sampler

Agilent 6850 Series II Network GC System G2630A Performance Specifications

Beverage Grade Carbon Dioxide

GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585

ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column

Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers. Versa HT3. Application Note. Abstract.

Application Note. Authors. Abstract. Energy & Chemicals

Application Note. Author. Abstract. Energy & Chemicals - Petrochemicals. Edgar Naegele, Agilent Technologies, Inc. Waldbronn, Germany

Agilent 7693A Automated Liquid Sampler

Agilent G2855A Deans Switching System

Setting up SilFlow for BackFlush in your GC

Rapid Qualitative GC-TOFMS Analysis of a Petroleum Refinery Reformate Standard

ASTM D 6730 Detailed Hydrocarbon Analysis

Page 1. Alternate Carrier Gas Considerations and Faster GC Analysis

Agilent Multimode Inlet for Gas Chromatography

Application Note. Author. Introduction. Energy and Fuels

The Stability of Sulfur Compounds, Low Molecular Weight Gases, and VOCs in Four Air Sample Bag Materials

Method Development for Capillary GC Systems. Slide 1

Simulated Distillation Analyzers, Software, Standards, Consumables, Training

[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS

Fausto Munari e Andrea Cadoppi ThermoFisher - Italy

Free and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN and ASTM D6584

Analysis of Petroleum Fractions by ASTM D2887

GC Method Compliance and Large Valve Oven Application

CP-3800 GAS CHROMATOGRAPH

DANI Transformer Oil Gas Analyzer

GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS

Detection of Volatile Organic Compounds in Gasoline and Diesel Using the znose Edward J. Staples, Electronic Sensor Technology

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

Agilent 9000 Intuvo GC Applications. Matthew Giardina, Ph.D. James McCurry, Ph.D. Rebecca Veeneman, Ph. D.

PRODUCT SPECIFICATION. Product Specification Crystal 9000 Gas Chromatograph. Chromatec Crystal Laboratory Gas Chromatography System

6890 CHECKOUT PROCEDURE TCD (THERMAL CONDUCTIVITY) DETECTOR

Detailed Hydrocarbon Analysis Featuring Rtx -1 PONA Columns

High-Temperature Simulated Distillation System Based on the 6890N GC Application

LIABLE RESULTS AND RELIABLE RESULTS AND EXCEPTIONAL REPRODUCIBILITY

Performing ASTM 6584 free and total glycerin in BioDiesel using an SRI Gas Chromatograph and PeakSimple software

Alliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE

Safety Products brochure 11 Int INDEX. Exhaust Filters 2. Air Inlet Valve 5. GL45 Caps 6. DIN50 Caps 8. GL38 Caps 10. DIN60 Caps 11.

Safety Products brochure 11 Int INDEX. Exhaust Filters 2. Air Inlet Valve 5. GL45 Caps 6. DIN50 Caps 8. GL38 Caps 10. DIN60 Caps 11.

gc applications Hydrocarbons C1-C5 Paraffins Hydrocarbons, Sulfur Gases C1-C2 Hydrocarbons Gases

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard

Practical Steps in GC Troubleshooting

Impurity Testing of Fixed-Dose Combination Drugs Using the Agilent 1290 Infinity II HDR-DAD Impurity Analyzer Solution

The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector

TECHNICAL REPORT. Introduction. Agilent 1100 HPLC system. Figure 1: HALO columns exhibit UHPLC-like performance at conventional HPLC pressure

Restek Corporation 110 Benner Circle, Bellefonte, Pa Barry L. Burger, Neil Johansen, Valerie Gamble, Donald Rhoades

ACQUITY UPLC I-Class System with 2D Technology

Determination of Sudan Dyes I IV in Curry Paste

Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis

AppNote 1/2010 KEYWORDS ABSTRACT. Biodiesel, Automation, ASTM D

Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC

Model 8610C Gas Chromatograph. GC Chassis Types ECD PID FID / DELCD FPD. Mounts up to Six Detectors and Five Injectors

SELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS

Operation and Applications of Differential Flow Modulation

Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI)

DETECTORS Photo Ionization Detector - PID. Overview

Agilent Multimode Inlet

Determination of fuel system icing inhibitor content of aviation turbine kerosine by HPLC

Analysis and. Separation of Oxygenates in Hydrocarbon Matrices. Simon Jones Application Engineer Folsom, CA

Liquefied Gas Injector. Solution for the Sampling and Analysis of Liquefied Gases

REFINED RESTEK. Detailed Hydrocarbon Analysis. innovative petrochemical solutions. Featuring Rtx -DHA Columns

The next level of True Blue Performance

Refinery Support from the R&D Laboratory Perspective using Fast & Micro Gas Chromatography

Analysis of Glycerin and Glycerides in Biodiesel (B100) Using ASTM D6584 and EN Application. Author. Abstract. Introduction

High-Performing Restek PAL SPME Fibers

AppNote 6/2006. Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) KEYWORDS

Agilent 6890 Gas Chromatograph

Proof of Long-Term, Leak-Free Performance for a Novel Self-tightening GC Column Nut

Transcription:

Productivity Improvements for Dual Column Applications: USP <467> and Blood Alcohol Analysis Rebecca Veeneman, PhD Applications Chemist Agilent Technologies

The GC Usability Gap Usability lags features and performance Over the years, GC features and performance have matured and largely serve today s needs Usability has lagged and does not meet today s expectations, especially as expertise more scarce Time Features and performance Usability

A Time for Change Shifting Demographics Seasoned operators retiring; highly skilled replacements harder to find GC troubleshooting skills not necessarily located at point of use Operators assuming multiple responsibilities Challenging Economics Budgets squeezed, expectations increased Capital purchase decisions made on business NOT technical basis

The GC Community Voice We listened and responded Improve the User Experience Install and setup Use and maintain Make GC more practical for today s busy lab enterprise Prepare for Next Generation of Users Its not only about better analytical performance its about better lab outcomes

Make GC Easier

Innovating a New Path to GC Productivity A whole new way to GC Easier Faster Smaller Smarter Greener

Innovating a New Path to GC Productivity A whole new way to GC Direct heating

Innovating a New Path to GC Productivity A whole new way to GC No-trim column Direct heating

Innovating a New Path to GC Productivity A whole new way to GC Ferrule-free click-andrun connections No-trim column Direct heating

Innovating a New Path to GC Productivity A whole new way to GC Disposable Guard chip Ferrule-free click-andrun connections No-trim column Direct heating

Innovating a New Path to GC Productivity A whole new way to GC Modular Intuvo flow chips Disposable Guard chip Ferrule-free click-andrun connections No-trim column Direct heating

Flexible Compatible Design Configurable to any application SSL, MMI, GSV, LSV inlets FID, TCD, ECD, NPD, FPD, NCD, SCD detectors SQ and TQ mass spectrometers Headspace sampler, CTC PAL 16-, 50-, 150-position auto-injectors and trays Software: OpenLAB and MassHunter

Innovating the GC Flow Path Conventional flow path Inlet Gold seal Nuts and ferrules Classic capillary column To detector

Innovating the GC Flow Path Conventional flow path Intuvo flow path Inlet Inlet Guard chip Gold seal Nuts and ferrules Classic capillary column To detector

Innovating the GC Flow Path Conventional flow path Intuvo flow path Inlet Inlet Click-andrun direct connections Guard chip Gold seal Nuts and ferrules To detector Flow chip Classic capillary column Intuvo planar column To detector

Click-and-Run Direct Connections Eliminating connection uncertainty No more ferrules Direct face seal connections Audible and tactile click lets you know connection is made Easier to train Less unplanned downtime Fewer batch reruns and precious samples lost

Autonomous Leak Checking Avoiding unplanned downtime Microfluidic-enabled 6 th generation EPC modules allow hands-free leak checking to confirm and document leak-free operation autonomously.

Intuvo Graphical User Interface System status

Intuvo Graphical User Interface System status Real-time chromatograms pa 3000 2500 2000 1500 1000 500 0 0.6 0.8 1.0 1.2 1.4 1.6 1.8 min.

Intuvo Graphical User Interface System status Real-time chromatograms Step-by-step user maintenance and troubleshooting

Intuvo Graphical User Interface System status Real-time chromatograms Find Inlet Parts Replace Inlet Liner Step-by-step user maintenance and troubleshooting Inlet Liner Part Number 8001-0101 Finding parts fast

Internet of Things A full suite of customer support Intuvo serves up a web homepage User help for Maintenance, troubleshooting and diagnostics Accessible through any approved PC or mobile device securely on internal network

Agilent CrossLab Services for Intuvo A streamlined service experience Intuvo 9000 GC Health Report Installation and familiarization Compliance services Education

A simplified approach to dual column analyses USP <467> and Blood Alcohol Analysis

Dual Column Configuration Headspace to SSL Inlet splitter flow chip equally splits to two columns Two columns to two FID detectors Inlet FID 1 FID 2 Guard Chip Inlet splitter flow chip Column 2 Column 1

USP <467> Residual Solvent Requirements Analysis of residual solvent is a critical application in the pharmaceutical industry. The choice of solvent during manufacturing can improve yield or affect chemical properties of the product synthesized. Solvents do not enhance the product s efficacy and must be removed as completely as possible to meet product specification and good manufacturing practices. USP467 specifies a single column analysis A secondary analysis is performed if the solvent is found above limit detection. With an Intuvo 9000 GC configured with an inlet splitter both analyses can be performed in one analytical run. USP 40, general chapter USP <467> Residual Solvents https://hmc.usp.org/sites/default/files/documents/hmc/gcs-pdfs/c467.pdf

USP <467> Residual Solvent Methods Class 1, Class 2A, and Class 2B residual solvents were prepared at their limit concentrations in water. Class 1 (5190-0490) was prepared in the following manner: 1. 1.0 ml stock and 9 ml of DMSO was diluted to 100 ml with water 2. 1.0 ml from step 1 was diluted to 100 ml with water 3. 10 ml from step 2 was diluted to 100 ml with water Class 2A (5190-0492) was prepared by diluting 1 ml of stock to 100 ml with water Class 2B (5190-0513) was prepared by diluting 1 ml of stock to 100 ml with water Headspace vials were prepared with 1 ml of each solvent and 5 ml of water. Samples were prepared in 10 ml HS vials

7697 Headspace parameters 7697 Headspace Sampler Setpoint Oven 85 C Loop 85 C Transfer line 100 C Vial equilibration time Injection duration Vial size Vial shaking Vial fill mode Vial fill pressure Loop ramp rate Loop final pressure Loop equilibration time 40 min 0.5 min 10 ml Level 2, 25 shakes/min Default (50 ml/min to 15 psi (0.1 min)) 15 psi 20 psi/min 0 psi 0.05 min

Intuvo Parameters Intuvo 9000 GC Oven Setpoint 40 C (5 min) 15 C/min to 180 C (3 min) Split/Splitless Inlet Split 5:1, 140 C DB-624 Select UI (123-0334UI-INT) 30m x 320µm x 1.8µm DB-WAX UI (123-7032UI-INT) 30m x 320µm x 0.25µm Constant flow 2 ml/min Controlled by column 1 FID (Front and Back) 250 C H2 Air 30 ml/min 400 ml/min N2 (makeup) 25 ml/min Jumper chip 140 C Bus 200 C Front/Back Signal 20 Hz Jumper chip set to the inlet temperature Bus temperature set to default

1.795 5.380 2.879 7.246 8.873 1.912 8.208 peak@8.208min 10.280 12.350 12.310 13.702 Response [pa] 4.327 peak@4.327min 8.535 peak@8.535min 17.061 8.014 peak@8.014min 8.463 peak@8.463min Class 1, DB-624 Select UI x10 1 DFID_Mix1_01 FID3A 1. 1,1-dichloroethane 2. 1,1,1-trichloroethane 3. carbon tetrachloride 4. benzene 5. 1,2-dichloroethane 1.025 1.02 1.015 1.01 1.005 1 0.995 0.99 0.985 0.98 0.975 0.97 0.965 0.96 0.955 0.95 0.945 0.94 0.935 0.93 0.925 0.92 0.915 0.91 0.905 0.9 0.895 0.89 0.885 0.88 0.875 0.87 0.865 0.86 0.855 0.85 0.845 0.84 0.835 0.83 0.825 0.82 0.815 0.81 1 3 2 4 5 0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5 9 9.5 10 10.5 11 11.5 12 12.5 13 13.5 14 14.5 15 15.5 16 16.5 17 Retention time [min]

1.888 15.136 Response [pa] 2.264 peak@2.264min 3.135 peak@3.135min 6.736 peak@6.736min 3.892 peak@3.892min Class 1, DB-WaxUI 1. 1,1-dichloroethane 2. 1,1,1-trichloroethane 3. carbon tetrachloride 4. benzene 5. 1,2-dichloroethane x10 1 1.0175 1.015 1.0125 1.01 1.0075 1.005 1.0025 1 0.9975 0.995 0.9925 0.99 0.9875 0.985 0.9825 0.98 0.9775 0.975 0.9725 0.97 0.9675 0.965 0.9625 0.96 0.9575 0.955 0.9525 0.95 0.9475 0.945 0.9425 0.94 0.9375 0.935 0.9325 0.93 0.9275 0.925 0.9225 0.92 0.9175 0.915 0.9125 0.91 0.9075 0.905 0.9025 0.9 0.8975 0.895 0.8925 0.89 0.8875 0.885 0.8825 0.88 0.8775 0.875 DFID_Mix1_01 FID4B 1 2,3 4 5 0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5 9 9.5 10 10.5 11 11.5 12 12.5 13 13.5 14 14.5 15 15.5 16 16.5 17 Retention time [min]

Class 1 Repeatability Compound DB-624 RT RSD DB-624 Area RSD DB-Wax RT RSD DB-Wax Area RSD 1,1-dichloroethane 0.031% 3.3% 0.074% 2.1% 1,1,1- trichloroethane carbon tetrachloride 0.026% 2.6% 0.037% 1.8% 0.11% 4.8% Coelutes with 1,1,1- trichloroethane Coelutes with 1,1,1- trichloroethane benzene 0.017% 2.2% 0.055% 1.5% 1,2-dichloroethane 0.016% 3.9% 0.067% 1.8% Excellent RT repeatability for both columns Area RSD is 5% or better

2.902 peak@2.902min 8.459 5.238 peak@5.238min 5.747 peak@5.747min 7.746 peak@7.746min 9.662 peak@9.662min 10.801 7.344 peak@7.344min 9.499 peak@9.499min 12.716 peak@12.716min 8.093 peak@8.093min Response [pa] 10.653 peak@10.653min 12.199 peak@12.199min 12.326 peak@12.326min Class 2A, DB-624 Select UI 1. methanol 2. acetonitrile 3. methylene chloride 4. trans-1,2-dichloroethene 5. cis-1,2-dichloroethene 6. tetrahydrofuran 7. cyclohexane 8. methylcyclohexane 9. 1,4-dioxane 10.toluene 11.chlorobenzene 12.ethylbenzene 13.m-xylene 14.p-xylene 15.o-xylene x10 3 1.7 1.65 1.6 1.55 1.5 1.45 1.4 1.35 1.3 1.25 1.2 1.15 1.1 1.05 1 0.95 0.9 0.85 0.8 0.75 0.7 0.65 0.6 0.55 0.5 0.45 0.4 0.35 0.3 0.25 0.2 0.15 0.1 0.05 0 DFID_Mix2a_01 FID3A 1 2 3 4 5 6 7 8 9 13,14 10 12 15 11-0.05 0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5 9 9.5 10 10.5 11 11.5 12 12.5 13 13.5 14 14.5 15 15.5 16 16.5 17 Retention time [min]

1.895 3.747 peak@3.747min 5.156 peak@5.156min 6.644 peak@6.644min 17.220 2.413 peak@2.413min 2.949 peak@2.949min 4.888 peak@4.888min 8.717 peak@8.717min 7.784 peak@7.784min 9.254 peak@9.254min 2.251 peak@2.251min Response [pa] 6.054 peak@6.054min 8.028 peak@8.028min Class 2A, DB-Wax UI 1. methanol 2. acetonitrile 3. methylene chloride 4. trans-1,2-dichloroethene 5. cis-1,2-dichloroethene 6. tetrahydrofuran 7. cyclohexane 8. methylcyclohexane 9. 1,4-dioxane 10.toluene 11.chlorobenzene 12.ethylbenzene 13.m-xylene 14.p-xylene 15.o-xylene x10 3 1.275 1.25 1.225 1.2 1.175 1.15 1.125 1.1 1.075 1.05 1.025 1 0.975 0.95 0.925 0.9 0.875 0.85 0.825 0.8 0.775 0.75 0.725 0.7 0.675 0.65 0.625 0.6 0.575 0.55 0.525 0.5 0.475 0.45 0.425 0.4 0.375 0.35 0.325 0.3 0.275 0.25 0.225 0.2 0.175 0.15 0.125 0.1 0.075 0.05 0.025 0-0.025-0.05 DFID_Mix2a_01 FID4B 7 8 4 5 1 3,6 2 10 13 12 9 14 15 0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5 9 9.5 10 10.5 11 11.5 12 12.5 13 13.5 14 14.5 15 15.5 16 16.5 17 Retention time [min] 11

Class 2A Repeatability Compound DB-624 RT RSD DB-624 Area RSD DB-Wax RT RSD DB-Wax Area RSD methanol acetonitrile methylene chloride 0.23% 1.3% 0.22% 1.3% 0.023% 2.1% 0.039% 2.0% 0.018% 0.98% 0.033% 0.78% trans-1,2- dichloroethene 0.016% 0.85% 0.023% 0.72% cis-1,2- dichloroethene 0.012% 0.76% 0.039% 0.78% tetrahydrofuran cyclohexane methylcyclohexane 1,4-dioxane toluene chlorobenzene ethylbenzene m-xylene p-xylene o-xylene coleutes with 0.018% 0.95% methylene chloride coelutes with methylene chloride 0.011% 0.96% 0.013% 0.96% 0.0087% 0.98% 0.018% 0.96% 0.012% 1.9% 0.025% 1.7% 0.0073% 0.81% 0.029% 0.80% 0.0061% 0.69% 0.024% 0.69% 0.0060% 0.84% 0.016% 0.86% 0.0061% 0.82% 0.017% 0.82% coleutes with m-xylene coelutes with m-xylene 0.015% 0.83% 0.0059% 0.81% 0.026% 0.81% Excellent RT repeatability for both columns Area RSD is 2% or better

1.922 7.815 * peak@7.815min 8.473 * peak@8.476min 12.324 14.871 10.645 peak@10.642min 6.229 peak@6.230min 9.245 peak@9.247min 11.335 peak@11.334min Response [pa] 16.464 peak@16.463min Class 2B, DB-624 Select UI 1. n-hexane 2. nitromethane 3. chloroform 4. 1,2-dimethoxyethane 5. trichloroethane 6. pyridine 7. 2-hexanone 8. tetralin x10 1 3.3 3.25 3.2 3.15 3.1 3.05 3 2.95 2.9 2.85 2.8 2.75 2.7 2.65 2.6 2.55 2.5 2.45 2.4 2.35 2.3 2.25 2.2 2.15 2.1 2.05 2 1.95 1.9 1.85 1.8 1.75 1.7 1.65 1.6 1.55 1.5 1.45 1.4 1.35 1.3 1.25 1.2 1.15 1.1 1.05 1 0.95 0.9 0.85 0.8 0.75 DFID_Mix2b_01 FID3A 1 2 3 4 0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5 9 9.5 10 10.5 11 11.5 12 12.5 13 13.5 14 14.5 15 15.5 16 16.5 17 Retention time [min] 5 6 7 8

2.257 3.715 peak@3.705min 4.930 peak@4.930min 5.677 peak@5.676min 6.046 7.065 peak@7.065min 8.011 peak@8.012min 9.196 peak@9.194min 17.216 peak@2.028min Response [pa] 12.934 peak@12.936min Class 2B, DB-Wax UI 1. n-hexane 2. nitromethane 3. chloroform 4. 1,2-dimethoxyethane 5. trichloroethane 6. pyridine 7. 2-hexanone 8. tetralin x10 1 5.7 5.6 5.5 5.4 5.3 5.2 5.1 5 4.9 4.8 4.7 4.6 4.5 4.4 4.3 4.2 4.1 4 3.9 3.8 3.7 3.6 3.5 3.4 3.3 3.2 3.1 3 2.9 2.8 2.7 2.6 2.5 2.4 2.3 2.2 2.1 2 1.9 1.8 1.7 1.6 1.5 1.4 1.3 1.2 1.1 1 0.9 0.8 0.7 DFID_Mix2b_01 FID4B 1 1.895 2.030 4 5 3 7 2 0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5 9 9.5 10 10.5 11 11.5 12 12.5 13 13.5 14 14.5 15 15.5 16 16.5 17 Retention time [min] 6 8

Class 2B Repeatability Compound DB-624 RT RSD DB-624 Area RSD DB-Wax RT RSD DB-Wax Area RSD hexane 0.021% 4.4% 0.039% 4.5% nitromethane 0.42% 4.9% 0.018% 2.5% chloroform 0.0099% 4.8% 0.026% 3.2% 1,2- dimethoxyethane 0.013% 3.6% 0.089% 3.6% trichloroethane 0.010% 3.4% 0.028% 3.1% pyridine 0.020% 1.7% 0.034% 1.7% 2-hexanone 0.0081% 2.7% 0.020% 0.74% tetralin 0.0078% 2.1% 0.016% 2.1% Excellent RT repeatability for both columns Area RSD is 5% or better

Blood Alcohol Concentration Application Requirements Determination of blood alcohol concentration requires rigorous control. Many labs use flame ionization detection (FID) which lacks identification capabilities Often a second system with a column having different retentive properties is used to confirm analyte identification Blood alcohol concentration determination and confirmation can be achieved simultaneously with the Intuvo 9000 GC. An Intuvo 9000 GC system was equipped with a 7697A Headspace Sampler and configured with an inlet splitter allowing dual column, dual FID analysis. NEW BAC UI columns (123-9334UI-INT and 123-9434UI- INT) were used which improve resolution of critical analytes

7697 Headspace parameters 7697 Headspace Sampler Set point Oven 70 C Loop 70 C Transfer line 90 C Vial equilibration time Injection duration Vial size Vial shaking Vial fill mode Vial fill pressure Loop ramp rate Loop final pressure Loop equilibration time 7 min 0.5 min 20 ml Off Default (50 ml/min to 15 psi (0.1 min)) 15 psi 30 psi/min 1.5 psi 0.05 min

Intuvo Parameters Intuvo 9000 GC Oven Set point 40 C (6.5 min) Split/Splitless Inlet Split 10:1, 110 C DB-BAC1 UI (123-9334UI-INT) 30m x 320µm x 1.8µm DB-BAC2 UI (1230-9434UI-INT) 30m x 320µm x 1.2µm Constant pressure 21 psi Controlled by column 1 FID (Front and Back) 250 C H2 Air 30 ml/min 400 ml/min N2 (makeup) 25 ml/min Jumper chip 110 C Bus Default (On 200 C) Front/Back Signal 20 Hz Jumper chip set to the inlet temperature Bus temperature set to default

Response Response Calibration Results Calibration standards were made in house 0.8% to 0.01% methanol, acetaldehyde (not quantified), acetone, ethanol, and isopropanol. Headspace vials were made in triplicate at each level consisting of 450 µl of internal standard (0.3% n- propanol) and 50µL of standard. Calibration curves for methanol, ethanol, acetone, and isopropanol are shown right. Ethanol calibration curves yield 0.9996 or better for the dual column ensemble. Slope difference between the two column/detector pairs is 6.3% 1000 900 800 700 600 500 400 300 200 100 0 1000 900 800 700 600 500 400 300 200 100 0 BAC UI 1 y = 2.0265x + 4.254 R² = 0.999 y = 1.1191x + 2.5958 R² = 0.9887 y = 0.3281x + 0.9655 R² = 0.9973 0 100 200 300 400 500 600 700 800 900 Concentraion (mg/dl) BAC UI 2 y = 2.1435x + 1.2863 R² = 0.9991 y = 1.2208x + 2.8953 R² = 0.9868 y = 0.3549x + 0.5807 R² = 0.9983 0 100 200 300 400 500 600 700 800 900 Concentraion (mg/dl) y = 0.616x - 1.3505 R² = 0.9996 Isopropanol Acetone Ethanol Methanol y = 0.6554x - 1.1943 R² = 0.9997 Isopropanol Acetone Ethanol Methanol

Calibration Accuracy Verification Ethanol Standard Calculated Concentration DB-BAC1 UI Pass/Fail Calculated Concentration DB-BAC2 UI Pass/Fail Agilent Ethanol Standards were evaluated on the Intuvo 9000 GC. The concentration of the ethanol standards was calculated based on the calibration curves and compared to the expected concentration. Pass or fail was determined with ± 6% error tolerance. 20 mg/dl (5190-9756) 50 mg/dl (5190-9757) 80 mg/dl (5190-9758) 100 mg/dl (5190-9759) 150 mg/dl (5190-9760) 200 mg/dl (5190-9761) 300 mg/dl (5190-9762) 400 mg/dl (5190-9763) 19.8 mg/dl Pass 19.3 mg/dl Pass 50.0 mg/dl Pass 47.1 mg/dl Pass 79.3 mg/dl Pass 76.8 mg/dl Pass 96.7 mg/dl Pass 94.4 mg/dl Pass 152 mg/dl Pass 149 mg/dl Pass 197 mg/dl Pass 193 mg/dl Pass 302 mg/dl Pass 302 mg/dl Pass 384 mg/dl Pass 386 mg/dl Pass

Area Repeatability Area repeatability for the 80mg/dL standard as well as the Agilent Blood Alcohol Checkout mix (5190-9765) was determined for five replicate headspace vials. Analyte DB-BAC1 UI DB-BAC2 UI Ethanol 80 mg/dl standard 3.70% 2.80% Methanol 4.10% 1.40% Acetaldehyde 2.80% 3.00% Ethanol 2.30% 1.10% Isopropanol 3.30% 1.90% t-butanol 2.80% 2.70% Propanal 3.40% 3.00% 4.1% or better! n-propanol 3.10% 2.10% Acetone 3.40% 2.90% Acetonitrile 2.30% 2.80% 2-Butanol 2.00% 3.00% Ethyl acetate 3.20% 3.10% 2-Butanone 3.10% 3.00%

Retention Time Repeatability Retention time repeatability for the 80mg/dL standard and the Agilent Blood Alcohol Checkout mix (5190-9765) was determined for five replicate headspace vials. Analyte DB-BAC1 UI DB-BAC2 UI Ethanol 80 mg/dl standard 0.04% 0.10% Methanol 0.01% 0.02% Acetaldehyde 0.01% 0.02% Ethanol 0.02% 0.05% Isopropanol 0.02% 0.04% t-butanol 0.03% 0.04% Propanal 0.01% 0.02% 0.1% or better! n-propanol 0.03% 0.04% Acetone 0.02% 0.03% Acetonitrile 0.02% 0.03% 2-Butanol 0.04% 0.04% Ethyl acetate 0.02% 0.03% 2-Butanone 0.02% 0.03%

DB-BAC1 UI Blood Alcohol Checkout Mix New columns resolve t-butanol and n-propanol from analytes of interest 2 Average resolution = 2.6 6 7 1. Methanol 2. Acetaldehyde 3. Ethanol 4. Isopropanol 5. t-butanol 6. Propanal 7. n-propanol 8. Acetone 9. Acetonitrile 10.2-Butanol 11.Ethyl Acetate 12.2-Butanone 5 8 11 12 1 3 4 9 10

DB-BAC2 UI Blood Alcohol Checkout Mix Elution order changes allow confirmation of analyte identification 2 6 8 5 7 12 11 1. Methanol 2. Acetaldehyde 3. Ethanol 4. Isopropanol 5. t-butanol 6. Propanal 7. n-propanol 8. Acetone 9. Acetonitrile 10.2-Butanol 11.Ethyl Acetate 12.2-Butanone 1 3 9 4 10

Conclusions Dual column analyses can be easily accomplished with Intuvo Analysis of USP <467> and blood alcohol can be easily accomplished with Intuvo Inlet splitting capability enables analysis and confirmation in a single run on two columns of different phases Can be configured as an Agilent Analyzer as well Includes a factory method and report template New columns (DB-624 Select UI, DB-Wax UI, DB-BAC 1 UI and DB-BAC 2 UI) yield excellent resolution of the analytes of interest in their application space. Excellent linearity, retention time repeatability, and area repeatability were achieved with both applications configured with the inlet splitter.

Thank you for your attention Any questions?