User Manual Electronic Titrator

Similar documents
N E W! The next generation Digital Burette.

Titrette. class A precision. The next generation Digital Burette!

Titrette. class A precision. 10 ml. New. size. The next generation Digital Burette!

Titrette. class A precision. The next generation Digital Burette!

User Manual dpette+ Electronic Pipette 2016-ver.1

Liquid Handling Easy Handling! Titrette. Gebrauchsanleitung. Operating Manual. Mode d'emploi. Instrucciones de manejo

The Measure of all Things. Volumetric Solutions from Merck

Operation Manual BOTTLE TOP DISPENSER

TITREX. One instrument for all your titration and dosing applications

JADE Bottle Top Dispenser

Digital Burette. Assembly and operating instructions. For capacities of 30ml & 50ml

Prospenser. User Manual Mode d emploi Bedienungsanleitung Manual Usuario. For capacities of 2.5 ml 5 ml 10 ml 30 ml and 50 ml

Operating Manual for. Dispenser. Read this manual thoroughly before operating the Labmax Bottle Top Dispenser!

Liquid Handling DIN EN ISO DE-M

Perfect Liquid Handling PRECISE RESULTS WITH INNOVATIVE TECHNOLOGY

PROPOSED REVISION OF THE GENERAL CHAPTER 1.11 COLOUR OF LIQUIDS. for The International Pharmacopoeia

Liquid Handling. iclab.com.au / iclab.co.nz

ECscan10H Pocket Conductivity Tester. Instruction Manual BANTE INSTRUMENTS LIMITED

For any application the right Dispensette. Dispensette III Dispensette Organic Dispensette HF

SAMPLING CONTAINERS & HOLDING TIMES. Parameters Method Holding Time 8 Minimum Volume Water Sampling Water Soil Water Soil Container Preservative 9

DURAN GROUP DISPENSER & DURAN GROUP DISPENSER PRO. A safe and precise dispensing system

Tempo PIPETTE CONTROLLER USER GUIDE. Masters of precision

Single and 8-channel Electronic Pipettes. Operation Manual Version: 2.9

User Manual Solar Charge Controller 3KW

TDSscan10M Pocket TDS Tester Instruction Manual

Dilutor Module for TriPlus RSH

PC60 Premium Multi-Parameter Tester (ph/ec/tds/salinity/temp.) Instruction Manual. APERA INSTRUMENTS, LLC V6.4

MM PISTON PUMP

8,1,8,1,2,7,2,7,6,3,6,3,4,5,4,5

BOHLENDER GmbH Phone: +49(0)9346 / Fax: +49(0)9346 / Internet:

ECscan40 Pocket Conductivity/TDS/Salinity Tester Instruction Manual

ECscan20 Pocket Conductivity Tester Instruction Manual

806 Exchange Unit. Manual EN

Eppendorf Repeater Pro Instruction Manual Manual de Instrucciones

Never charge the batteries with a charger other than that supplied by the manufacturer.

Marin Control Digital

Operation Manual. Pipette Aid. Concorde Liquid Handling Equipment. *This instrument is intended for laboratory use only.

ceramus HiClass in dispensing.

TEST REPORT: CHM12-01-LSM

INSTRUCTION MANUAL HI 921 AUTOSAMPLER

Filtertechnik Filtration, Purification & Separation Solutions

MiniLab EL Series for High Performance Engine Development and Maintenance

BlueSwan Pipette Controller

ECscan30 Pocket Conductivity/TDS Tester Instruction Manual

IAQ-CALC TM Indoor Air Quality Meter

High Purity Chemical Valve Compact Type. Space saving, compact type is available. Piping from 4 directions are possible.

THREE RANGE DS METERS Operation Instructions

SofTouch Electronic Precision Pipette Instruction Manual

1 Operating instructions for rail-mounted stand holder, variable

Competence in Labware. Liquid-Handling range

CappController Pipette Controller. Operation Manual Version: 1.2.

Lumina 28 Traction Drive Model: Battery Burnisher M28036TDQP OPERATION SERVICE PARTS CARE

OPERATORS MANUAL FOR USE WITH MAGNETIC/HEATED STIRRER MG200

IAQ-CALC INDOOR AIR QUALITY METER MODEL 7515

How to Use the Chemical Compatibility Chart. Vials and Closures >> Chemical Compatibility Chart

884 Professional VA. Manual Short Instructions EN /

PHOTOSWITCH Liquid Level and Resistance Sensing Control Bulletin 13DJ3-3000

Arch HTH Water Chemicals Commercial Equipment Trouble Shooting Guide

ASTM Methods that are feasible with Metrohm instruments

Autoplex Specifications

WalkLAB Digital Dissolved Oxygen meter HD9030

Competence in Labware. Liquid-Handling range

Dispensette. Testing Instructions (SOP) 1. Introduction. May 2009

4&5 RANGE CONDUCTIVITY METERS Operation Instructions

AUTOMATIC BURNISHER MODEL FURY 21 HSB

Instruction Manual. Vortex Mixers

Valor 1000W Series Instruction Manual. Serie Valor 1000W Manual de Instrucciones. Série Valor 1000W Guide de I utilisateur

Dynaform Fiberglass Structural Shapes. Design Guide. High Performance Composite Solutions. Corrosion Resistant. Fire Retardant.

DUAL RANGE DIALYSATE METER Operation Instructions User Manual for Models D-2

150 Series Brass Horizontal Float Valve. Technical Information

IAQ-CALC TM Indoor Air Quality Meter

MICROMANOMETER ALNOR MODEL AXD610 AIRFLOW INSTRUMENTS MODEL PVM610

Finnpipette Novus. Single Channel & Multichannel. User Manual. Instructions for Use

Foam Output 50 gal/min (190 ltr/min) 65 gal/min (246 ltr/min) Foaming Distance 30 feet (9 meters) 35 feet (10 meters) alcohols, ketones.

N) manual. Register your instrument! Eppendorf Reference 2. Operating manual

Register your instrument! Eppendorf Reference 2. Operating manual

Safety Products brochure 11 Int INDEX. Exhaust Filters 2. Air Inlet Valve 5. GL45 Caps 6. DIN50 Caps 8. GL38 Caps 10. DIN60 Caps 11.

D Dräger Alcotest 3000 Breath Alcohol Measuring Device

ATTENTION! READ BEFORE ATTACHING THE AIR HOSE

9. WARRANTY INFORMATION

FULL AUTOMATIC ASPHALTENE ANALYZER APD-500A USER S GUIDE

Air Operated Double Diaphragm Pump. M-Pump ½ Metallic Non Metallic Pump INSTALLATION, OPERATION & MAINTENANCE MANUAL

ISO-9001:2008 DISPENSERS CATALOG. Phone: Fax: Web:

Techcon Systems TS500R-PC Controller For Positive Displacement PC Pump

Indoor Air Quality Meter

Sediment strainer (Type Y)

Model 2053 Conductivity Meter Instruction Manual

DOSING UNIT WITH TIME CONTROL INSTRUCTIONS MANUAL WARNING!

CRD610 Automatic Fitting Inserter

Small on price, big on boot

Fibreglass Reinforced Plastic

Contents. Zippette Pro Bottle-top Dispensers Safetypette Pipet Filler...4. Zippette Pro Bottle-top Dispenser Accessories...

Rosemount 402 and 402VP

CRD600 Automatic Fitting Inserter

Operating Instructions. Magnetic Drive For EasyMax

3/E3x ual Repeater E3/E3x

SIP Direct Drive Oil-Lube Air Compressors - Operating & Maintenance Instructions

AIR VELOCITY METER ALNOR MODEL AVM410 AIRFLOW INSTRUMENTS MODEL TA410

Techcon Systems TS6500 CIM Automatic Techkit Mixer

AMB100L-im-issue INSTRUCTION MANUAL 240V DIESEL MINI BOWSER AMB100L AMB100L

Transcription:

User Manual Electronic Titrator BDI-TIT PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com MRC.3.17

Content 1. Unpacking... 1 2. Overview... 2 3. Parts Description... 3 4. Assembly Instruction... 10 5. Operation... 16 6. Function Setting... 21 7. Accessories... 23 8. Calibration... 29 9. Cleaning and Maintenance... 30 10. Trouble Shooting... 35 11. Storage... 37 12. Warranty... 38 13. Limitations and Compatibility... 38

Safety Reminder CAUTION:Possible damage to instrument. Caution notes indicate any condition or practice, which if not strictly observed or remedied, could result in damage or destruction of the instrument.

1. Unpacking Apart from the user manual, the BDI-TIT package should contain the following items. Electronic Titrator Include: Titrating pipe X1 Titrating pipe cover X1 BDI-TIT X1 AC Adapter X 1 Controller X 1 Controller cable USB X 2 Bottle Adapter X 5(GL32; GL38; GL28; GL25; S40) Magnetic Stirrer X 1 Remote Titrating pipe X 1 Remote Control Handle X 1 Stirrer Bar(20mm)X 1 Filling valve X 1 Dispensing valve X 1 Filling pipe X 2 Installation tools X 1 Stander Please check that all the items are present and inform your supplier immediately if any of the above is missing. 1

2. Overview A digital titrating instrument, BDI-TIT delivers accurate and precise bottle-top titration. Please refer to "Limitations and Compatibility" for liquid compatibility prior to operation. 2.1 Specification Volume Range Precision Velocity Battery 0.01mL-99.99mL Increment 10μL R= 0.2% CV=0.07% 16 Stages Capacity:4000mA/h Fully charged in 4 hours (please use standard charger) working time:about 5 hours 2.2 Limitations of Use Temperature:15 ~ 40 C Vapor pressure:<500mbar Viscosity:<500mm²/s Humidity:20~90% 2

3. Parts Description 3.1 BDI-TIT 1 2 3 4 5 6 7 No. 1 2 3 4 5 6 7 Description Liquid Level Observation (in piston running state) Return Valve (to adjust the liquid direction of dispensing ) Titrating pipe Titrating pipe cover Titrating pipe Tip Bottle Adapter Filling pipe 3

1 2 3 No. 1 2 3 Description Controller Port (Micro USB) Main Body lock Air Admission Cap (pressure balance) 3.2 Remote Control Handle Control Knob Data Connector 4

3.3 Controller Allows for BDI-TIT control and function setting. 1 2 3 6 5 4 No. 1 2 3 4 5 6 Description LCD Display(show BDI-TIT running state) Filling(press and hold for filling,release it to stop) Dispensing(press and hold for fast titration,release it to stop) Setting/magnetic stirrer(short press switching magnetic stirrer on/off; press and hold 2s into setting interface) Pre-Filling(press and hold 2s for piston to complete a aspirating and dispensing process) Knob(Turn Knob for dropwise titrating, short press Knob for reset the liquid volume record to zero) 5

Control Panel can be fully adjustable up to an angle of 45 3 2 1 4 No. Description 1 Magnetic Stirrer Data Port(USB) 2 Locating Slot(to ensure magnetic stirrer assembly in place) 3 Sensor Holder Assembly Slot 4 Magnetic Stirrer Fastening Slot(to fixation with controller) 6

1 5 2 3 4 No. Description 1 Power Switch(symbol O indicates Off, - indicates On) 2 Remote control handle Port 3 Communication port(non-function) 4 Charging/Communication port 5 Main Instrument Port 7

3.4 Magnetic Stirrer 4 1 2 3 No. Description 1 Stirring Plate(Max. Bar 20mm) 2 Communication port 3 Fastening key 4 Location key(to ensure magnetic stirrer assembly in place) 8

3.5 Display Magnetic Stirring Indication Battery Indication Indication Arrow (piston movement) Liquid Volume Unit 9

4. Assembly Instruction Step - 1: Turn the Main Instrument Lock clockwise to connect titrating tip with Main Body. Step - 2: Press the titrating pipe tail end into the location slot. Location slot 10

Step - 3: Push the titration pipe tail end with a certain length. Step - 4: Connect the titration pipe tail end with the Dispensing valve. 11

Step - 5: locking the titration pipe. CAUTION:Titrating pipe are made of FEP. Please confirm compatibility prior to use(refer to chapter12 Limitations and Compatibility ). 12

Step - 6: Connect filling pipe with filling valve. Step - 7: Choose a suitable bottle adapter, then connect it with BDI-TIT main body. 13

Step - 8: Turn bottle adapter to fasten main body and bottle. CAUTION: 1 Please enable that the adapter is fastened prior to each use. 2 For perfect titration, please do not move or touch Main Instrument during operation to avoid physical damage to your BDI-TIT. 14

Step - 9: Use USB cable to connect Main Instrument with Controller. Step - 10: Turn Return Valve to direction 1 If liquid is needed to be emptied from the barrel, turn Return Valve to direction 2. BDI-TIT basic system was assembled. 1 2 15

5. Operation CAUTION: Do a complete process of aspirating and dispensing before the first time work. 5.1 Titration Step - 1: Power on. Step - 2: Waiting self-check complete. 16

Step - 3: Long press Pre-filling button( ) 2 seconds to let air out, leaving the piston at the bottom of the barrel finally. Step - 4: Press and hold Filling button ( )to fill targeted liquid until Indication Arrow showing that the piston is moving downward. Long press knob 2 seconds, do the same operation also. 17

Step - 5: Long press Dispensing button ( )for fast titration. Turn Knob for dropwise titrating. The dispensing liquid volume will been continue recording and the volume will be show on the screen. NOTE: When recorded volume reaches the maximum value of 99.99ml, the instrument will stop.please press the knob to reset Liquid Volume to zero. CAUTION: There will may be some air bubble in the barrel during the operation. These bubble dose not effect the actual use. If the bubble is bigger to effect the actual use, please 18

running several times aspirating and dispensing. If this solution not works, please contact with the dealer or manufacturer. 5.2 Liquid Emptying If liquid is needed to be emptied from the barrel. Step - 1: Turn Return Valve to direction 2. 1 2 19

Step - 2: Long press Dispensing button ( ),until the piston run to the bottom of the barrel, make the liquid had been emptying. Step - 3: Turn Return Valve to direction 1. Emptying operation was completed. 1 2 20

6. Function Setting Press Setting button( )for about 2 seconds to enter Function Setting interface. Press Setting button( )to flick through pages. Long press Setting button( )for about 2 seconds on any page to exit Function Setting interface. Page I: Aspiration speed within a range of 1-16. Turn Knob to adjust and press for confirmation. 21

Page II: Dispensing speed within a range of 1-16. Turn Knob to adjust and press for confirmation. Page III: Magnetic stirrer speed within a range of off-5. Turn Knob to adjust and press for confirmation. NOTE: This function is effective after connecting the magnetic stirrer only. 22

7. Accessories 7.1 Remote Control Handle The Control Handle is fully map the operation of Control Panel, easy to operate over a long distance. Turn Knob for dropwise titrating, short press for reset liquid volume record to zero, long press 2 second for filling liquid. MAX. Length:90 cm 7.2 Magnetic Stirrer 23

Magnetic stirrer convenient user for mixing during the titration experiments. Step - 1: Connect Control Panel Fastening Slot with Magnetic Stirrer Fastening Slot. Step - 2: Turn the Magnetic Stirrer into horizontal position until the Location Hole into the Location Slot. Step - 3: Connect the USB Port under the Magnetic Stirrer with the Control Panel. 24

7.3 Assemble Sensor Holder. Step - 1: Fasten the Holder into place. Step - 2: Press black button of the clamp and release until reaching the appropriate altitude. 25

Assembly diagram 7.4 Remote Titrating pipe 26

Remote titrating pipe can effectively extend the titration distance. MAX. Length 1.5m CAUTION: Titrating pipe are made of FEP. Please confirm compatibility prior to use(refer to chapter12 Limitations andcompatibility ). Step - 1: Connect the titration pipe tail end with the Dispensing valve. 27

Step - 2: locking the titration pipe. Step - 3: Follow the figure to assemble the adapter and remote titrating pipe. 28

1 2 8. Calibration Calibration should take place at 20-25, kept constant within 29

±0.5. A dedicated calibration software will write calibration values in your BDI-TIT, after the distilled water has been repeatedly weighed up at least five times. Hardware needed:: Electronic balance with readability of 0.01 mg Distilled water X86-or X64-architectured PC with pre-loaded Windows (XP/Vista / 7/8/10 )operating system Software needed: Dedicated calibration software of BDI-TIT (For more information, please contact with your nearest distributor.) CAUTION: If your BDI-TIT can not work properly after calibration, please contact your nearest distributor for assistance. 9. Cleaning and Maintenance CAUTION: BDI-TIT cannot be autoclaved. 30

9.1 Cleaning the Outer Surface The outer surface of your BDI-TIT is made of ABS, ideal for easy cleaning with simply clean water. 9.2 Cleaning the Barrel BDI-TIT barrel cleaning is repeatedly inhale row clear water for cleaning. Aspiration and dispensing at least 5 times, according to user's actual situation to increase or decrease. To ensure emptying remained in the barrel, the operation reference liquid emptying Step - 1: Turn Return Valve to direction1, long press Pre-filling button make the piston stop at the barrel bottom 31

1 2 Step - 2: Press Filling and Dispensing button to aspiration and dispensing simply clean water at least 5 times. Step - 3: long press Pre-filling button make the piston stop at the barrel bottom. Step - 4: Ensure that into the tube is not submerged in a liquid, Press filling button make piston run to the top of the barrel. Step - 5: Turn Return Valve to direction2, press Dispensing button make piston run to the bottom of the barrel. Step - 6: Cleaning work is finish,turn Return Valve to direction1 CAUTION: User is not recommended to remove and 32

cleaning of BDI-TIT barrel, if the barrel cleaning operation fail to meet the cleaning requirements of users. please contact the dealer or manufacturer professional services personnel for cleaning. Ensure BDI-TIT empty without liquid residue before delivery to service personnel and inform details of last liquid handling. 9.3 Filling and Dispensing valve Replacement Use the Installation tools to disassemble the old valve, replace the 33

new valve to the same position. Valve has no fixed replacement cycle, problems after the replacement. The issue that could be has involvement with valve, please checking the Trouble Shooting CAUTION: The following operation must to use installing tools to do. Before disassemble, ensure to remove the dispensing and filling pipe. Disassemble Dispensing valve 34

Disassemble Filling valve 10. Trouble Shooting Issue Possible Cause Solution 35

Piston overflows with liquid Piston moves with difficulty Failure to filling Failure to refill; refilling sucks back into the titration tip. Bubbles in the instrument/ Dispensing volume is less than the volume displayed. Piston wears out. Piston or its parts are contaminated or damaged due to crystallization and sedimentation. Filling valve is clogged. Dispensing valve is contaminated or titration tip damaged. Filling pipe is loose or damaged. Filling pipe is away from the liquid. Return pipe is not installed or wrongly 36 Contact with manufacturer 1 Do Cleaning the Barrel 2 Contact with manufacturer 1 Replace filling valve 2 Contact with manufacturer Replace filling pipe Chicking filling pipe Contact with manufacturer

installed. The instrument is not fully refilled. Checking Operation Checking filling No display Filling valve is clogged or damaged. Battery dead Connection fail valve Replace filling valve charging USB cable connection checking 11. Storage During storage periods at constant temperature and humidity, 37

the recommended temperature range is from 0-40 and humidity no more than 80%. Please every month to charging for BDI-TIT if being unused in long time, make sure there are 50% power in battery at least. 12. Warranty BDI-TIT electronic titrators are covered by one-year warranty against defects in workmanship and materials. Please contact us or your nearest distributor. ANY WARRANTY WILL, HOWEVER, BE DEEMED AS VOID WITH NORMAL WEAR AND TEAR OR FOR OPERATIONS CONTRARY TO THE INSTRUCTIONS GIVEN IN THIS MANUAL. Each and every BDI-TIT has been calibrated and tested in compliance with ISO8655-6 / DIN12650 when manufactured, ensuring safe and comfortable operation. 13. Limitations and Compatibility It is recommended to confirm reagent s compatibility with this 38

instrument when applied for special purposes, trace analysis for example. The liquid-path construction of your BDI-TIT is made of borosilicate glass, FEP and PTFE. Do not apply it in handling liquids like hydrofluoric acid. The instrument would be clogged or damaged by solid particles in turbid liquid like activated carbon. The plastic parts of your BDI-TIT would be in swelling condition if concentrated acid and alkaline, and methylbenzene, benzene and other nonpolar organic solvents are put into use. Keep your BDI away from the highly combustible carbon disulfide. BDI-TIT cannot be autoclaved. Do not put your BDI-TIT in contact with corrosive gas like HCL smog. Compatibility(Max. Conc. 1 mol/l) Acetic acid 39

Alcoholic potassium hydroxide solution Ammonium iron (II) sulfate solution Ammonium thiocyanate solution Barium chloride solution Bromide bromate solution Cerium (IV) sulfate solution EDTA solution Hydrochloric acid Hydrochloric acid in Acetone Iodine solution* Iodide Iodate solution* Iron (II) sulfate solution Nitric acid Oxalic acid solution Perchloric acid Perchloric acid in glacial acetic acid 40

Potassium bromate solution Potassium bromate bromide solution Potassium dichromate solution Potassium hydroxide solution Potassium iodate solution Potassium permanganate solution* Potassium thiocyanate solution Silver nitrate solution* Sodium arsenite solution Sodium carbonate solution Sodium chloride solution Sodium hydroxide solution Sodium nitrite solution Sodium thiosulfate solution Sulfuric acid Tetra-n-butylammonium hydroxide sol. 41

Triethanolamine in Acetone* Zinc sulfate solution CAUTION: This compatibility is against parts which are directly in contact with liquid, if any of above solution needs to be applied, contact with manufacturer for consultation. 42