Instructions for use. Quick couplings RQ-03, RQ-04, RQ-14, RQ-24, RQ-34

Similar documents
Instructions for use AM-25 LT AM-25 A, AM-25 A BC/RM AM-25 E, AM-25 E BC/RM AM-25, AM-25 BC/RM

Instructions for use. Prosthodontic screwdriver IA-400

Instructions for use of S5 Endo Motor

Turbines with ut a dark side

Instructions for Use. A-dec EA-30. Electric Motor System

Operating Instructions. International English. CCS Dust Extractor

Simple, fast, efficient.

CO 3-WAY PNEUMATIC VALVE INSTRUCTION MANUAL 2080

HEINE NT4 MED V

Instructions for Use. Edition USA SI-1015

362 Clamp Meter. Users Manual

Operating instructions in the back. PD1420LP

Operating instructions in the back. PD1820LF

People have Priority. Restoration and Prosthetics

323/324/325 Clamp Meter

Instructions for Use. A-dec EA-53 Electric Motor

INSTRUCTIONS FOR USE. Rx Only

Instructions for use Spare cartridge for COMPACTtorque 636 P, 636 CP and POWERtorque 646 B, 646 C , Always be on the safe side.

PURE SINE WAVE DC TO AC POWER INVERTER

Assembly and Maintenance Manual Type RSBW

Rechargeable laryngoscope handles Directions for use

User Manual TABLE MAGNIFIER TL-70. User-friendly Manual ID: #05007

Assembly and Maintenance Manual Type ASNU

Installation manual wall-mounted distributor

HD18F2M1.

SPARK PLUG CLEANING KIT OPERATING MANUAL

Operating Instruction INTRAmatic Prophy Handpiece 19 ES

Assistant s Instrumentation T

SURGICAL POWER & ACCESSORIES. Rx Only PM-X00-710/PM-X IR-12.13

Assembly and Maintenance Manual Type AS

LH41A. User Manual. Clamp On Ammeter. For detailed specifications and ordering info go to

Isolation Transformer for Medical Applications

1587 FC/1587/1577. Insulation Multimeter. Safety Information. Warnings

Users Manual. Phase Rotation Indicator

60V RECHARGEABLE LITHIUM-ION BATTERY

Installation manual portable distributors

40 V LITHIUM-ION BATTERY ATTACH YOUR RECEIPT HERE AB13786C 1. kobalttools.com ITEM # /

365 Detachable Jaw True-rms Clamp Meter

BW - BIG VOLUME BOOSTER INSTRUCTION MANUAL 2072

ACROBAT SWING STAND MODEL

302+/303 Clamp Meter. Users Manual

362 Clamp Meter. Users Manual

Armon Edero. User manual

Instruction Manual PC104/PC105 3-PHASE SOCKET TESTER. Check out what else you can get from Martindale:

Instructions for use GENTLEmini LUX 4500 BR - REF

233 True-rms Remote Display Digital Multimeter Safety Information

Go to to register your product and find more information.

1662/1663/1664 FC. Electrical Installation Tester. Safety Information. Warnings

Swing Piston Compressors and Vacuum Pumps

STAND MODEL FOR MACH M2

OWNER'S MANUAL. Royal Sovereign International Inc. RET Desk Series

USE, CARE & STERILIZATION INSTRUCTIONS. Consumable Accessories Burs Saw Blades K-Wires Steinmann Pins Twist Drills Rasps Ortho Pins/Drills Guide Wires

SPARKSCAN1 HIGH VOLTAGE CLAMP OPERATING MANUAL

Before installing this unit, read the operating instructions carefully. Strictly adhere to these operating instructions.

Catalyser Introduction. 3. Hazard Information. 2. Area of Application. 3.1 Symbology

Installation and Operating Manual for Tank and Equipment Cleaning Nozzles Series 5TM

Mod: KLD6-12/35XLAS-N

VSM480. Instruction Manual VARIABLE SPEED MIXER. Read instructions before operating this tool.

BT508/BT510/BT520/BT521 Battery Analyzer Safety Sheet

Assembly and maintenance manual Type FSO, FSO-GR, FS, HPI

Instructions for use DURAtec Prophy REF

Safety information on VIKING type AAI lithium-ion batteries for use in imow robotic mowers

Calibrator Capabilities Function Range Resolution Measure V dc 28 V 1 mv Measure ma dc Source ma dc

Register your instrument! HeatSealer S100. Operating manual

373 Clamp Meter. Users Manual

Charder. Medical scale. User instruction. Important information. This user instruction must be read before the medical scale is used.

KD1E Range Potential Indicators

MANN+HUMMEL ENTARON XD. Installation and Maintenance Manual. Version 0709

02OWNER S MANUAL 06 TROUBLESHOOTING 08LIMITED WARRANTY

AXIAL FANS Axis / Tubo OPERATION MANUAL

PRM-4. Phase Sequence and Motor Rotation Tester. Users Manual. For detailed specifications and ordering info go to

Battery Charger JCB- SCH20LI.2

Installation and Operation Manual

Rev ENGLISH..2

Instructions for: FOLDING ENGINE CRANE PREMIER 2 TONNE. MODEL No: SPC2000

Operating Instructions K Always on the safe side.

LUBRICATOR GUN INSTRUCTIONS-PARTS LIST. 10,000 psi (700 bar) Maximum Delivery Pressure. Detachable-type

Model 2310 Series. Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority


28W WORK LAMP Model: CTL28 Part No:

Installation and Operational Instructions for ROBATIC -clutch Type and Type Sizes 3 9

EN Operating manual. Motorised zone valve. Three-way, 22 mm & 28 mm 3PV2, 3PV8 & VRMH3

MV Series Motors Operation & Parts Manual

User s Guide DSC. electronic. DENTALE MEDIZINISCHE PRODUKTE Hans Karl Matysiak HKM


PIX. Operation - Replacement of Fuses Replacement of Voltage Transformers Technical manual. Medium Voltage Distribution

Rev Miniature / Miniature K. Silent Power Evo 2 / 2L / 4L / 4LK / ENGLISH...2

FL 10 DIAPHRAGM PUMP INSTALLATION INSTRUCTIONS. Before operating the pump, please read the Installation Instructions and safety precautions.

Digital Apex Locator. ipex OPERATION MANUAL OM-E0285E

Excel Wall and Deck Mounted Bath Shower Mixers Installation and User Guide

OWNER S MANUAL TOLL-FREE HELPLINE: (Monday - Friday / 8am - 4:30pm CST)

2.0Ah BATTERY INSTRUCTION MANUAL

Gavita HortiStar 1000 DE EU/US

Instructions for use GENTLEsilence 6500B - REF

User manual Pipe notcher Type: AL 1-2E

51 & 52 Series II. Users Manual. Thermometer. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

i400s Calibration Information Introduction

Operating Instructions

SCOPELITE TIMING LIGHT OPERATING MANUAL

Universal Bevel Drives Service Manual

Transcription:

Instructions for use Quick couplings RQ-03, RQ-04, RQ-14, RQ-24, RQ-34

Contents W&H Symbols...4 1. Introduction...5 7 2. Before use Safety notes...8 9 3. Product description...10 11 Roto Quick coupling (10), Cooling water lines (11) 4. Operation...12 19 Connecting the Roto Quick coupling (12), Disconnecting the Roto Quick coupling (13), Checking for leaks (14), Changing air-driven instruments (15 16), Regulating the spray water (17), Test run (18), Changing O-rings (19) 2

5. Hygienic maintenance...20 25 6. W&H Accessories...26 7. Technical data...27 8. Recycling and disposal...28 29 Letter of indemnity...30 Authorized W&H service partner...31 3

W&H Symbols WARNING! (Risk of injury) ATTENTION! (to prevent damage occurring) General explanations, without risk to persons or objects Sterilizable up to the stated temperature Data Matrix Code for product identification, e.g. in hygienic maintenance process 4

1. Introduction Customer satisfaction is the main priority under the W&H quality policy. This W&H product has been developed, manufactured and subjected to final inspection according to legal regulations, quality norms and industry standards. For your safety and for the safety of your patients Prior to initial use please read the Instructions for use. These explain how to use your W&H product and guarantee a smooth and efficient operation. 5

Intended use Connection piece for the transmission of media (air, water, electricity and light) between the supply hose of the dental unit and the air drives. Misuse can damage the Roto Quick coupling and involves risk and danger to user and third parties. Production according to EU directive The Roto Quick coupling is a medical product according to the EU directive 93/42/EEC. 6

Skilled application The Roto Quick coupling is intended only for skilled application in dentistry according to its purpose of use in compliance with the valid health and safety at work regulations and the valid accident prevention regulations as well as in compliance with these Instructions for use. The Roto Quick coupling should be prepared for use and maintained by staff who have been trained in procedures for infection control, personal safety and patient safety. Improper use, (e.g. through poor hygienic maintenance), noncompliance with our instructions or the use of accessories and spare parts which are not approved by W&H, invalidates all claims under warranty and any other claims. Service In case of service or repair contact an authorized W&H service partner (see Page 31). Service and maintenance work must only be carried out by an authorized W&H service partner. 7

2. Before use Safety notes > The operation of the Roto Quick coupling with light is permitted only on dental units which correspond to the standards IEC 60601-1 (EN 60601-1) and IEC 60601-1-2 (EN 60601-1-2) and satisfy the following requirements to be guaranteed by the system assembler: > Double insulation for the highest expected supply voltage must be provided between the primary and secondary power circuits. > Double insulation for the highest expected secondary voltage must be provided between the secondary voltage and earth (PE). > The secondary voltage must be limited to a maximum of 3.3 V AC/DC. > The leakage currents of the applied part must be kept. > The secondary circuits must be protected against short circuits and overloads. 8

> The user is not allowed to touch the ring contacts and the patient at the same time. > Only use supply hoses which comply with EN ISO 9168. > Always ensure that the correct operating conditions exist and that the coolant is working properly. > As cooling air, only use air which has been processed by dental compressors and which is filtered, free of oil and cooled. > Before using the coupling, always check it for damage and loose parts. Hygienic maintenance prior to initial use The Roto Quick coupling is in a clean condition and sealed in PE-film when delivered. Sterilize the Roto Quick coupling. > Sterilization Page 24 9

3. Product description ➊ O-rings ➋ Retention sleeve ➌ Spray regulation ring (for RQ-14, RQ-34) 10

Cooling water lines All Roto Quick couplings are equipped with a non-retraction valve which prevents contaminated cooling water from being sucked back into the turbine and the turbine hose. The non-retraction valve is integrated in the cooling water supply system. In the event of blocked or incorrectly routed cooling water lines, please contact an authorized W&H service partner (see page 31). The cooling water lines must not be cleaned with sharp objects! (This could damage the sealing element and prevent the non-retraction valve from working.) 11

4. Operation Connecting the Roto Quick coupling ➊ Connect the Roto Quick coupling to the turbine hose. ➋ Firmly tighten the union nut of the turbine hose by hand in a clockwise direction to ensure there are no leaks. This method of assembly provides a connection for drive air, chip air, water and return air (only for RM). 12

Operation Disconnecting the Roto Quick coupling ➊ Unscrew the union nut of the turbine hose by hand in an anticlockwise direction. ➋ Carefully remove the Roto Quick coupling from the turbine hose. 13

Checking the Roto Quick coupling for leaks > Check for leaks between the Roto Quick coupling and turbine hose. > Block the water outlet on the tip of the Roto Quick coupling using a piece of rubber or leather. > Turn on the spray water. After being blocked for 10 seconds, no water must escape between the coupling and turbine hose or at the spray regulator. > If you observe problems (e. g. vibrations, unusual noise and/or leaks), stop the Roto Quick coupling immediately and contact an authorized W&H service partner (see Page 31). 14

Operation Changing air driven instruments > Observe the instructions in the relevant instructions for use. > Only connect air driven instruments with Roto Quick connections to the Roto Quick coupling. The user accepts sole responsibility if other air driven instruments are used. We accept no liability in such cases. 15

Operation Changing air driven instruments Never connect or disconnect the air driven instrument while the system is running! > Attach the air driven instrument onto the Roto Quick coupling until it audibly engages. > Pull axially along the Roto Quick coupling to check that the air driven instrument is securely attached. > Pull the retention sleeve of the Roto Quick coupling back. > To disconnect the air driven instrument, pull it off in axial direction. 16

Regulating the spray water For Roto Quick couplings RQ-14 and RQ-34: Adjustment directly at the spray adjuster ring or at the adjustable valves of the unit. Roto Quick couplings RQ-03, RQ-04 and RQ-24 can only be adjusted at the unit. Minimum spray flow: Maximum spray flow: Line marking on the spray adjuster ring and point on the retention sleeve correspond with each other. Turn the spray adjuster ring to the right or left. The water flow can only be varied using this spray adjuster ring. It cannot be stopped completely. 17

Test run > Connect the Roto Quick coupling to the turbine hose. > Attach the air driven instrument onto the Roto Quick coupling until it audibly engages. > Start the air driven instrument. > If you observe problems (e. g. vibrations, unusual noise and/or leaks), stop the Roto Quick coupling immediately and contact an authorized W&H service partner (see Page 31). 18

Operation Changing O-rings Immediately replace any damaged or leaking O-rings. Never use sharp tools! ➊ Press the O-ring firmly together with your thumb and index finger until it forms a loop. ➋ Pull off the O-ring. ➌ Slide the new O-ring back on. 19

5. Hygienic maintenance > Follow your country-specific directives, standards and guidelines for cleaning, disinfection and sterilization. > Wear protective clothing. > Remove the Roto Quick coupling from the turbine hose. > Clean and disinfect the Roto Quick coupling immediately after every treatment. > The Roto Quick coupling is sterilizable > Manual internal and external cleaning is to be carried out before each sterilization. 20

Pre-disinfection > If heavily soiled: Clean first with disinfectant cloths. Only use disinfectants that have no protein-fixing effects. 21

Manual cleaning internal and external ➊ Rinse under demineralized water (< 38 C / < 100 F) with the aid of a brush. ➋ Remove any liquid residues (absorbent cloth, blow dry with compressed air). Do not place the Roto Quick coupling in liquid disinfectant or in an ultrasonic bath! 22

Manual disinfection > Disinfection with disinfectants, ➊ wiping disinfection is recommended. > Use only disinfectants which do not contain chlorine and which are certified by officially recognized institutes. > Note the manufacturer s specifications for the use of the disinfectants. Storage in the surgery For storage, the Roto Quick coupling is normally firmly screwed between the air driven instrument (turbine, air motor, air scaler) and the turbine hose and stored on the turbine hose. 23

Sterilization and storage W&H recommends sterilization according to EN 13060, class B > Note the instructions of the unit manufacturer. > Clean and disinfect before sterilizing. > Wrap the Roto Quick coupling in sterile goods packing according to EN 868-5. > Make sure, that you only remove dry sterile goods. > Store sterile goods dust-free and dry. 24

Approved sterilization procedures Follow your country-specific directives, standards and guidelines. > Steam sterilization class B (pre and post vacuum) with sterilizers in accordance with EN 13060. Sterilization holding time a minimum of 3 minutes at 134 C (273.2 F). or > Steam sterilization class S with sterilizers in accordance with EN 13060. The sterilizer manufacturer must give their express approval for the sterilization Roto Quick couplings. Sterilization holding time a minimum of 3 minutes at 134 C (273.2 F). 25

6. W&H Accessories Use only original W&H accessories and spare parts! Suppliers: W&H partners 10400300 Roto Quick coupling RQ-03 with resuction stop, with Borden 2/3-hole connection 10400400 Roto Quick coupling RQ-04 with resuction stop, for Ritter-Midwest 4-hole connection 10401400 Roto Quick coupling RQ-14 with resuction stop, spray regulation, for Ritter-Midwest 4-hole connection 10402400 Roto Quick coupling RQ-24 with resuction stop, standard 4-hole connection and 2 electrical contacts 10403400 Roto Quick coupling RQ-34 with resuction stop, spray regulation, for standard 4-hole connection and 2 electrical contacts 02060100 O-ring, large 02060200 O-ring, small 26

7. Technical data Hose type: ISO 9168 type 1 ISO 9168 type 3 Recommended voltage range: 3.2 ± 0.1 V AC/DC Physical characteristics Temperature for storage: Air humidity for storage: Ambient temperature during operation: Air humidity during operation: -40 C to +70 C (-40 F to +158 F) 8 % to 80 % (relative), non-condensing at +40 C (+104 F) +10 C to +40 C (+50 F to +86 F) 15 % to 80 % (relative), non-condensing at +30 C (+86 F) 27

8. Recycling and disposal Recycling W&H considers that it has a special duty towards the environment. The Roto Quick coupling equipment along with its packaging has been designed to be as environmentally friendly as possible. Disposal of the Roto Quick coupling Follow your country-specific laws, directives, standards and guidelines for the disposal of used electrical devices. Ensure that the parts are not contaminated on disposal. 28

Disposal of the packaging material All packaging materials have been selected according to environmentally compatible and disposal aspects and can be recycled. Please send old packaging materials to the relevant collection and reprocessing system. In this way, you will contribute to the recycling of raw materials and the avoidance of waste. 29

L e t te r o f i n d e m n i t y This W&H product has been manufactured with great care by highly qualified specialists. A wide variety of tests and controls guarantee faultless operation. Please note that claims under warranty can only be validated when all the directions in the instructions for use have been followed. As manufacturer, W&H is liable for material or manufacturing defects within a warranty period of 24 months from the date of purchase. We accept no responsibility for damage caused by incorrect handling or by repairs carried out by third parties not authorized to do so by W&H! Claims under warranty accompanied by proof of purchase must be sent to the vendor or to an authorized W&H service partner. The provision of service under warranty extends neither the warranty period nor any other guarantee period. 24 m o n th s wa r ra n t y

Authorized W&H service partner Find you nearest W&H service partner at http://wh.com. Simply go to the menu option»service«for full details. Alternatively please contact: W&H (UK) LIMITED, 6 Stroud Wood Business Centre, Park Street, St Albans, Herts AL2 2NJ, England t +44 1727 874990, f +44 1727 874628, E-Mail: technical.uk@wh.com W&H Impex Inc., 6490 Hawthorne Drive, Windsor, Ontario, N8T 1J9, Canada t +1 800 2656277, f +1 519 9746121, E-Mail: service.ca@wh.com W&H Impex Inc., 14300 Henn Rd., Dearborn, MI 48126, USA t +1 800 2656277, f +1 519 9746121, E-Mail: service.us@wh.com W&H DENTAL-MECHANIK (ASIA) PTE-LTD., 18 Boon Lay Way, #10-94 Tradehub 21, 609966 Singapore t + 65 63168370, f + 65 63168371, E-Mail: service.sg@wh.com A-DEC AUSTRALIA CO. INC., Unit 8, 5-9 Ricketty Street, Mascot NWS 2020, Australia t +61 2 83324000, f +61 2 83324099, E-Mail: craig.young@a-dec.com.au 31

Manufacturer W&H Dentalwerk Bürmoos GmbH Ignaz-Glaser-Straße 53, 5111 Bürmoos, Austria t + 43 6274 6236-0, f + 43 6274 6236-55 office@wh.com wh.com Form-Nr. 50610 AEN Rev. 006 / 28.03.2011 Subject to alterations