Type D0800. Bantam Cartridge Holder OPERATION MANUAL AND PARTS LIST SERIES 494. Mo de l # Vo ltag e D D /12/08

Similar documents
Type Maxi Mix III Rotary Shaker

Lab Rotators and Low Cost Orbital Shakers

S25500 MAXI STIRRER OPERATION MANUAL AND PARTS LIST SERIES 1352

Type Maxi Mix III Rotary Shaker OPERATION MANUAL AND PARTS LIST SERIES 1046 BARNSTEAD THERMOLYNE

S25500 MAXI STIRRER OPERATION MANUAL AND PARTS LIST SERIES 849

Operating Instructions

OPERATION MANUAL AND PARTS LIST

NANOpure Infinity UV Module Upgrade

BARNSTEAD THERMOLYNE ROTO MIX TYPE

Operating and Installation Instructions

Installation Manual DIAPHRAGM WELL TANK

Manufacturing Facility is ISO9001:2000. Welded Steel Pipe Conform to ASTM A53. Threads Conform to ASME B1.20.1

STAINLESS STEEL FORGED FITTINGS 3000lb 304/304L & 316/316L Forged Stainless Steel Fittings Product Specifications

ALITA LINEAR AIR PUMP OPERATION & MAINTENANCE MANUAL. AL- Model Number Date Code / Serial Number Date of Purchase

Operators Manual. Model 3370 Air Cooled Recirculator rev.8/98

II DISTRIBUTION & SUBSTATION TYPE C

PT-516 Pumping Trap Installation and Maintenance

PVI 1800/PVI Residential/Commercial Grid-Tied Photovoltaic Inverter WARRANTY MANUAL. Subject to Change REV , Solectria Renewables

Owner s Manual GLASSLINED PUMP TANK

BC Brake Caliper. (i) MEX (55) QRO (442) MTY (81) DIST. AUTORIZADO

MV Series Motors Operation & Parts Manual. Models M3V, M5V, M5V-US

Broadband PowerShield. External Battery Pack 20 Ah 12 V Battery. User Manual

Veterinary Anesthesia Gas Scavenging System (Active)

OWNER S MANUAL SUBMERSIBLE UTILITY PUMP

7.3L POWERSTROKE BANJO BOLT KIT Fits L Powerstroke Diesel. Installation Guide

DRAIN WATER TEMPERING KIT

HALLMARK INDUSTRIES INC

INSTALLATION INSTRUCTION & OWNER S MANUAL

PVI 60KW, PVI 82KW, PVI 95KW

Single Post Caliper Brake VC500

EF SERIES PUMPS OPERATION & PARTS MANUAL

TM SERIES MIXER TMS OPERATION & PARTS MANUAL

INSTRUCTIONS SMOKE BLOWER WITH HONDA ENGINE PART NUMBER

SUNC1200 / ITEM #40882 SUBMERSIBLE UTILITY PUMP OPERATIONS MANUAL

Watts Series CSM-61. Flow Measurement/Balancing Valves Sizes: 1 1 4", 1 1 2", 2", 2 1 2", and 3" (32, 40, 50, 65 and 80mm) Installation Instructions

XLR-IR850 Long-Range Infrared Illuminator

PUMP SPY INSTALLATION AND OPERATING INSTRUCTIONS

LESTRONIC II BATTERY CHARGER BUILT-IN OR PORTABLE CHARGERS

Model P-40 & Model P-25 POWER PUSHER

OWNERS MANUAL HF4263

WARRANTY POLICY. Grid-Tied Photovoltaic Inverters. Revision D. 2014, Solectria Renewables, LLC DOCIN

READ THIS MANUAL CAREFULLY BEFORE USING THE PUMP

Full Flow Drinking Water System Model: 3MFF100

Lubricator Gun: 10,000 psi (700 bar) Maximum Delivery Pressure when disconnected from Dispenser

LUBRICATOR GUN INSTRUCTIONS-PARTS LIST. 10,000 psi (700 bar) Maximum Delivery Pressure. Detachable-type

MOVE ON TO THE REAR BAR INSTALLATION

MaxLite StaxMax LED Flood Lights

DeZURIK AFR3 Filter Regulator Used on Pneumatic Cylinder Actuators

Adjustable Base. CONTENTS Advisory Setup and Installation Innova Wired Hand Control year Warranty 1-10

INSTALLATION NOTES: nbeckon-26

OWNER S MANUAL SELF-PRIMING PORTABLE UTILITY PUMP

MEX (55) QRO (442) Web Controls

Pressurized oil drain for collection of used lubricants and anti-freeze.

OWNERS MANUAL HF

CRD610 Automatic Fitting Inserter

INSTALLATION AND MAINTENANCE MANUAL FORM #PM-122 REV A 12/09

"WHERE TUBING AND FITTINGS COME TOGETHER"

APCO CRF-100A RUBBER FLAPPER SWING CHECK VALVES

BroadBand PowerShield. 20 AHr Battery. User Manual

BARNSTEAD THERMOLYNE CORPORATION MEL-TEMP 3.0 OPERATING INSTRUCTIONS. Model No LT1186X4 5/1/00

Elgin Hydraulic Clutch-Brake ECB-240, Product Number FORM NO. L F FORM NO. L F-0704

Operating and Installation Instructions This Product is Protected by U.S. Patent

Wall mount PMV pump with suction kits and hoses

60 PSI Boost Gauge. For Product Numbers: MT-DV01_60, MT-WDV01_60

DUAL WIDEBAND AIR/FUEL RATIO GAUGE Product Numbers: GS-W702W_Dual, GS-C702W_Dual, GS-T702W_Dual

AMERICAN AVK COMPANY TABLE OF CONTENTS

Nor East. Instructions. Safety Messages. Inspection. Parts Nor East Controls. Service. Type 05 Air-O-Motor Actuator Used With Globe Valves

AQUA2000 ADVANCED WATER TREATMENT SYSTEMS OWNER S MANUAL INCLUDES: Restrictions. Installation. Operation. Maintenance. Changing the Filter

Grinnell Grooved Fire Protection Products Grooved Fittings

COMPRESSED AIR LOADER MODEL: AL-1

FCB-450, LCB-600, MCB-800

V200D VERIS Verabar (Double Rod) Installation and Maintenance Manual

HP30 Slab Saw Owner s Manual and Operating Instructions

AWE AirGate Carbon Intake VW / Audi MQB 1.8T 2.0T. AWE website here

SOLUTION DISTRIBUTORS BOOK NO. IM AA UA

IMPORTANT OPERATING CONDITIONS. Failure to comply with any of these conditions invalidates the warranty. STANDARD CONFIGURATIONS

Installation Power Management Unit Battery Cables and Battery Harness

SUPER PUMP OUT SYSTEM

CRD400 Fitting Inserter OPERATIONS MANUAL

Smart Battery Charger GPC-35-MAX GPC-45-MAX GPC-55-MAX GPC-75-MAX GPC-100-MAX. Owner s Manual

SUNC3000 / Item #40885

MV Series Motors Operation & Parts Manual

INSTRUCTION MANUAL. Rotary Unidirectional Piezoelectric Motor Evaluation Kit. (Model: UPM-28) Made in USA

OWNERS MANUAL HF4550 PREACHER CURL. Customer Service (800) (858) Fax (858) RECORD SERIAL NUMBER HERE

8" - 12" Hydraulic Steel Squeeze Off Tool

EF SERIES MOTORS OPERATION & PARTS MANUAL

Tri-Clover Manual and Air Actuated Fractional Valves

Operating and Installation Instructions

TBP SERIES PUMPS OPERATION & PARTS MANUAL

User s Manual. Item no. 2561

B-pure Pressure Cartridge System

OWNERS MANUAL HF PAIR VERTICAL DUMBBELL RACK. Customer Service (800) (858) Fax (858) RECORD SERIAL NUMBER HERE

GENERAL WARRANTY TERMS & CONDITIONS FOR EICS PERMANENTLY SEALED CATALYSTS HAVING DIAMETERS OF 9.5 in, 11 in, & 13 in.

DAP-625S and DAP-875S

PEDESTAL SUMP PUMP. MODEL # Español p. 11. Zoeller is a registered trademark of Zoeller Co. All Rights Reserved.

AQUA2000 From O3PURE

TB SERIES PUMPS TBS OPERATION & PARTS MANUAL

CLEAN ROOM DEVICES, LLC "WHERE TUBING AND FITTINGS COME TOGETHER"

Installation and Maintenance Instructions

LESTRONIC II BATTERY CHARGER MODEL 19740

Transcription:

Type D0800 Bantam Cartridge Holder OPERATION MANUAL AND PARTS LIST SERIES 494 Mo de l # Vo ltag e D0800 120 D0805 240 60119 3/12/08

Table of Contents Safety Information...3 Alert Signals...3 Warnings...3 General Specifications...4 Cartridge Specifications...5 High Capacity...5 Ultrapure (Mixed Resin)...5 Installation...6 Unpacking...6 Choosing a Site...6 Mounting...6 Plumbing Service Connections...6 Elbow-Tubing Assembly...7 Cartridge Installation...8 Operation...9 Purity Meter Operation...9 Maintenance and Servicing...10 Calibration...10 Exploded View...11 Replacement Parts List...12 Ordering Procedures...13 Warranty...16 2

Safety Information Alert Signals Warning Warnings alert you to a possibility of personal injury. Caution Cautions alert you to a possibility of damage to the equipment. Note Notes alert you to pertinent facts and conditions. Hot Surface Hot surfaces alert you to a possibility of personal injury if you come in contact with a surface during use or for a period of time after use. Important Information This manual contains important operating and safety information. The user must carefully read and understand the contents of this manual prior to the use of this equipment. Safety Information Your Thermo Scientific Barnstead Bantam Cartridge Holder has been designed with function, reliability, and safety in mind. It is the userʼs responsibility to install it in conformance with local electrical codes. For safe operation, please pay attention to the alert signals throughout the manual. Warnings To avoid electrical shock, always: 1. Use a properly grounded electrical outlet of correct voltage and current handling capacity. 2. Disconnect from the power supply prior to maintenance and servicing. To avoid personal injury: 1. Do not use in the presence of flammable or combustible materials; fire or explosion may result. This device contains components which may ignite such materials. 2. Refer servicing to qualified personnel. 3

General Specifications Note Pure water exposed to the atmosphere will absorb CO 2. This will lower resistivity and ph. Max imum flow rate Temperature range E x change capacity SI U Plumbing service connections nits E nglish 1 38 LPH 0 GPH 1 1 4-49 C 40-120 F see below see below I nlet ~ 1/4" OD tubing O utlet ~ 1/4" OD tubing D rain ~ 1/4" OD tubing Overall dimensions Width Depth Height 152 mm 6" 213 mm 8-3/8" 724 mm 28-1/2" E lectrical req uirements Model D0800 D0805 120V, 50/60 Hz, 2 1 phase 240V, 50/60 Hz, 2 1 phase 120V, 50/60 Hz, 2 1 phase 240V, 50/60 Hz, 2 1 phase 1 Using average tap water. Delivers deionized water in continuous flow at atmospheric pressure. Although the cartridge holder can be operated at inlet pressures up to 700 kpa (100 psig), the cartridge holder cannot without any back pressure. Ensure that maximum flow rate is not exceeded, the outlet is open to atmosphere, and there are no tubing restrictions. 2 Provided with 1.7 m (5-1/2 ) electrical cord and plug. 4

Cartridge Specifications High Capacity This cartridge (catalog no. D0803) has the capacity to remove 113 grams (1740 grains) of dissolved ionized impurities (as NaCl) to an endpoint of 50,000 ohm-cm. This high capacity configuration produces water with an electrical resistivity of about 175,000 ohm-cm. Each cartridge will purify about 660 liters (175 gallons) of water containing 170 mg/liter (10 grains/gallon, as NaCl). The pressure drop across the cartridge is less than 7 kpa (about 1 psig) at 38 L/hr (10 GPH) when the cartridge is new. Ultrapure (Mixed Resin) This cartridge (catalog no. D0809) has the capacity to remove 65 grams (750 grains) of dissolved ionized impurities (as NaCl). About 3/4 (75%) of the capacity is capable of producing water with an electrical resistivity greater than 1 megohm/cm. Each cartridge will purify about 380 liters (about 100 gallons) of water containing 170 mg/liter (10 grains/gallon, as NaCl) to 20,000 ohms. The ph is 6.8 to 7.2, and the water produced is largely free of silica. The water is also free of CO2 until exposed to atmosphere. The pressure drop across the cartridge is less than 7 kpa (about 1 psig) at 38 L/hr. (10 GPH) when the cartridge is new. 5

Installation Unpacking Unpack the Bantam Holder carefully so that none of the parts will be marred or damaged. Ensure that any loose parts such as tubing and fittings for the plumbing service connections are not discarded with the packing materials. Choosing a Site Move the Bantam Holder to the desired operating location. The holder should be installed in a location accessible to a water supply (raw water faucet). Also, the unit should be accessible to an electrical outlet. The holder is equipped with a 1.7m (5-1/2') electrical cord. Mounting The Bantam Holder is shipped assembled for bench mounting. Four screws are provided for wall mounting if desired. Plumbing Service Connections The Bantam Holders are supplied with polypropylene elbows, inlet and drain tubing, and straight connectors. The plumbing service connections are accomplished as follows: 1. Remove the fitting and screen from the raw water faucet opening. Note It is suggested that the Bantam Holder inlet be connected to a raw water faucet. Other means of connecting the holder inlet to a raw water supply can be used, if desired. 2. Install a customer-supplied 3/8" NPT x 1/8" NPT brass bushing in the faucet opening. The 1/8" NPT portion of the brass bushing is used to connect the 1/8" NPT x 1/4" OD straight connector provided to connect the Bantam Holder inlet to the raw water supply. If the raw water supply service connection is not 3/8" NPT, use another brass fitting as required. 6

INSTALLATION 3. Cut a piece of tubing from the 2.1m (7') of 1/4" OD tubing (part no. TU1494X1) provided to make the plumbing service connection between the raw water faucet and the plumbing service connection marked inlet on the Bantam Holder. Ensure that sufficient tubing is left to drain the holder through the plumbing service connection marked drain on the Bantam Holder to the sink and to make the plumbing service connection between the holder effluent connection at the top of the unit and the point of distribution. 4. Thread the 1/8" NPT x 1/4" OD straight connector provided (part no. 15451) into the brass bushing at the faucet opening and connect the tubing between this connection and the plumbing service connection marked inlet on the holder. Caution Do not restrict the tubing and run the tubing in an upward direction from the top of the unit. Any restriction on the outlet can result in a cartridge bursting and flooding. 5. Ensure that the 1/8" NPT x 1/4" OD polypropylene elbow (part no. 15847) is threaded into the effluent connection at the top of the holder and connect tubing between this connection and the point of distribution. See Elbow-Tubing Assembly in this manual to ensure that the connection is made properly to prevent leakage. 6. Connect the remaining tubing to the plumbing service connection marked drain and extend the tubing into the sink to allow the holder to be drained in an unrestricted manner. Elbow-Tubing Assembly Assembling the tubing and polypropylene elbow is accomplished as follows to ensure against leakage. 1. Remove the adapter nut, being careful not to lose the o-ring, backup ring, and stainless steel grab ring. Elbow Tubing Installation 2. Remove the o-ring from the elbow. 3. Install the metal grab ring and backup ring in the adapter nut, and install the nut on the elbow. Do not install the o-ring at this time. 7

INSTALLATION 4. Push the tubing through the nut until it bottoms out in the elbow. Caution Do not apply a wrench to the nut the nut should be hand-tightened only to prevent leakage. 5. Remove the adapter nut, then install the o-ring on the tubing. Do not push the backup or metal grab rings back on the tubing when installing the o-ring on the tubing. 6. Install the adapter nut on the elbow and handtighten the nut. Caution Do not over-tighten the tightening wheel. Note Water purified by the ion-exchange method is not necessarily free from odor and color. In some instances, color or odor could be introduced by trace quantities of free amines or low molecular weight polymers which are present in fresh untreated resin. Generally, the amine odor lessens or disappears after passing a few gallons of water through the holder. Cartridge Installation Install a cartridge in the holder as follows: 1. Install the large o-ring (part no. 06419) on the cartridge holder. A small o-ring should be on the bottom of the cell. The large o-ring is packaged separately during shipment. 2. Install the fresh cartridge in an upright position, the large opening toward the bottom. The lower tightening wheel should be turned down to allow the cartridge to be centered between the top and bottom o-ring. 3. Turn the lower tightening wheel until the cartridge seats against the top o-ring. 8

Operation Caution Do not exceed a flow rate of 38 LPH (10 GPH). Leakage will occur if this flow is exceeded. Caution Do not over-tighten the tightening wheel. Operate the Bantam Holder as follows: 1. Close the drain valve. 2. Open the inlet valve one to two turns and allow a couple of minutes for the cartridge to fill. If leakage occurs at an o-ring, close the inlet valve, loosen the lower tightening wheel enough to rotate the cartridge 90, and retighten. Open the inlet valve and recheck for leakage. 3. When water issues from the tubing, adjust the inlet valve to the desired flow rate (see Cartridge Technical Characteristics). Purity Meter Operation Operate the Direct Reading Meter by setting the on/off switch to the on position. When the purity meter drops below the acceptable minimum purity, or when the purity meter needle registers in the 0.05 to 0.025 megohm red range, the cartridge should be replaced. The meter provides a constant indication of the water purity. The meter scale is calibrated 0 to 19 PPM (ionizable impurities) with equivalent indications in megohms resistance from 0.025 to 18, making possible constant monitoring of the water purity. To stop the meter from operating, set the on/off switch to the off position. 9

Maintenance and Servicing Warning To avoid electrical shock, always disconnect the power supply before maintenance and servicing. Refer servicing to qualified personnel. Maintenance of the Bantam Holder should be minimal, involving cartridge replacement and o-ring seal replacement. Also, an occasional calibration check of the purity meter might be necessary. Cartridges should be stored away from heat. Calibration Calibrate the Direct Reading Meter as follows: 1. Set the on/off switch to the off position. 2. Check the zero alignment of the meter needle to the meter scale. If required, rotate the zero adjustment screw until the meter needle is aligned with the line indicating the beginning of the meter scale. The zero adjustment screw is located on the face of the meter below the scales. 3. To calibrate the meter further, connect a known resistance across the cell terminals. A 50,000 ohm resistance should correspond to 0.05 on the meter scale. 4. With the on/off switch set to on, adjust the potentiometer (screwdriver adjustment) at the bottom of the meter case to make the scale reading correspond to the simulated resistance. 10

Exploded View 11

Replacement Parts Model Type: D0800 Product Name: Bantam Cartridge Holder Series Number: 494 K ey Part # (Q ty. ) D escription 1 15847 Elbow, 1/8" NPT x 1/4" tube 2 15570 Cell, BD-1 3 06411 O-ring 4 15482 Ring, retaining 5 06419 O-ring 6 03570 Valve, brass, 1/4" OD x 1/8" NPT 7 21538 (115V) 33019 (230V) Meter 8 15597 Cartridge holder 9 15896 Nipple 10 15502 Lower Tightening Wheel 11 05302 Tee D0803 or D0809 Cartridge, high capacity or Ultrapure, respectively (Purchased separately) 12

Ordering Procedures Please refer to the Specification Plate for the complete model number, serial number, and series number when requesting service, replacement parts or in any correspondence concerning this unit. All parts listed herein may be ordered from the Thermo Scientific dealer from whom you purchased this unit or can be obtained promptly from the factory. When service or replacement parts are needed we ask that you check first with your dealer. If the dealer cannot handle your request, then contact our Customer Service Department at 563-556-2241 or 800-553-0039. Prior to returning any materials, please contact our Customer Service Department for a Return Materials Authorization number (RMA). Material returned without an RMA number will be refused. 13

14

15

One Year Limited Warranty This Thermo Scientific product is warranted to be free of defects in materials and workmanship for one (1) year from the first to occur of (i) the date the product is sold by the manufacturer or (ii) the date the product is purchased by the original retail customer (the Commencement Date ). Except as expressly stated above, the MANUFACTURER MAKES NO OTHER WARRANTY, EXPRESSED OR IMPLIED, WITH RESPECT TO THE PRODUCTS AND EXPRESSLY DISCLAIMS ANY AND ALL WARRANTIES, INCLUDING BUT NOT LIMITED TO, WARRANTIES OF DESIGN, MERCHANT ABILITY AND FITNESS FOR A PARTICULAR PURPOSE. An authorized representative of the manufacturer must perform all warranty inspections. In the event of a defect covered by the warranty, we shall, as our sole obligation and exclusive remedy, provide free replacement parts to remedy the defective product. In addition, for products sold within the continental United States or Canada, the manufacturer shall provide free labor to repair the products with the replacement parts, but only for a period of ninety (90) days from the Commencement Date. The warranty provided hereunder shall be null and void and without further force or effect if there is any (i) repair made to the product by a party other than the manufacturer or its duly authorized service representative, (ii) misuse (including use inconsistent with written operating instructions for the product), mishandling, contamination, overheating, modification or alteration of the product by any customer or third party or (iii) use of replacement parts that are obtained from a party who is not an authorized dealer of Thermo Scientific products. Heating elements, because of their susceptibility to overheating and contamination, must be returned to the factory and if, upon inspection, it is concluded that failure is due to factors other than excessive high temperature or contamination, the manufacturer will provide warranty replacement. As a condition to the return of any product, or any constituent part thereof, to the factory, it shall be sent prepaid and a prior written authorization from the manufacturer assigning a Return Materials Number to the product or part shall be obtained. IN NO EVENT SHALL THE MANUFACTURER BE LIABLE TO ANY PARTY FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, OR FOR ANY DAMAGES RESULTING FROM LOSS OF USE OR PROFITS, ANTICIPATED OR OTHERWISE, ARISING OUT OF OR IN CONNECTION WITH THE SALE, USE OR PERFORMANCE OF ANY PRODUCTS, WHETHER SUCH CLAIM IS BASED ON CONTRACT, TORT (INCLUDING NEGLIGENCE), ANY THEORY OF STRICT LIABILITY OR REGULATORY ACTION. For the name of the authorized Thermo Scientific product dealer nearest you or any additional information, contact us: 2555 Kerper Blvd., Dubuque, Iowa, 52004-0797 Phone: 563-556-2241 or 1-800-553-0039 Fax: 563-589-0516 E-mail: mkt@thermofisher.com Web: www.thermo.com