Proving Liquid Ultrasonic Flowmeters Summary of Testing Conducted January 2011

Similar documents
Calibration, Fluid Property Effects Reproducibility and Long Term Stability. All in 20 minutes! Dr Gregor Brown Cameron

DFX Series Liquid Ultrasonic Flow Meters

Flow Research and Test Center. Creating resource technology for today and tomorrow.

Sentinel LCT. Panametrics Ultrasonic Flowmeter for Liquid Custody Transfer Measurement. GE Measurement & Control. Benefits.

Flow Research and Test Center

Sentinel LCT. Panametrics Ultrasonic Flowmeter for Liquid Custody Transfer Measurement. GE Sensing & Inspection Technologies. Benefits.

Traceability overview

DFX series. High-end Ultrasonic Flow Meters for Liquid Hydrocarbons Custody Transfer. 32 beams for ultimate performance

BASICS OF ULTRASONIC FLOW METERS

FH8500. Metering Solution for Custody Transfer Applications. One Step Ahead in Ultrasonic Flowmetering. Main Applications

HERMAN HOFSTEDE. Sales Director, KROHNE Oil & Gas

LIQUID MEASUREMENT STATION DESIGN with NGL CONSIDERATIONS

FUNDAMENTALS OF INSERTION TURBINE METERS Les Bottoms Thermo Electron Corporation, Flow Systems

Product Loss During Retail Motor Fuel Dispenser Inspection

Constant Known Conductivity Fuel Systems. & how they assist fuel conductivity management

LIQUID MEASUREMENT STATION DESIGN Class No

STEALTH INTERNATIONAL INC. DESIGN REPORT #1001 IBC ENERGY DISSIPATING VALVE FLOW TESTING OF 12 VALVE

COMPUTATIONAL FLOW MODEL OF WESTFALL'S 2900 MIXER TO BE USED BY CNRL FOR BITUMEN VISCOSITY CONTROL Report R0. By Kimbal A.

Applied Fluid Mechanics

Intercomparison Between Trapil, France and NEL, UK Using Oil Flow Turbine Meters

The Discussion of this exercise covers the following points:

FLUID CONTROL RESEARCH INSTITUTE

Fluid Flow Conditioning

Fundamentals of Natural Gas Measurement

Experiment No.3: Flow through orifice meter. Background and Theory

Flow Measurement SITRANS F S Clamp-on

High Performance Turbine Technology

Calibration of Pipe Provers in the field

FUNDAMENTALS OF VOLUME MEASUREMENT (TURBINE METERS) Tony Petitto. FMC Technologies Measurement Solutions, Inc.

DFX-MM series. High-end Ultrasonic Flow Meter for Liquid Hydrocarbons Custody Transfer. 32 beams for ultimate performance. L002G DFX-MM leaflet

Seals Stretch Running Friction Friction Break-Out Friction. Build With The Best!

Testing Of Fluid Viscous Damper

Compressor Noise Control

PROVISIONAL PATENT APPLICATION INVENTORS: RICHARD E. AHO CAVITATION ENGINE

Pump Control Ball Valve for Energy Savings

CHAPTER 5 ANALYSIS OF COGGING TORQUE

A COMPARISON OF THE PERFORMANCE OF LINEAR ACTUATOR VERSUS WALKING BEAM PUMPING SYSTEMS Thomas Beck Ronald Peterson Unico, Inc.

Pneumatic & Hydraulic SYSTEMS

A Practical Guide to Free Energy Devices

Exercise 4-1. Flowmeters EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION. Rotameters. How do rotameter tubes work?

FLUXUS HPI Series. Non-Intrusive Ultrasonic Flow Measurement for the Hydrocarbon Processing Industries. External Measurement of Internal Flow FLEXIM

KRAL Volumeter - Turbine Fuel and Water Injection Measurement

Non-Invasive Hydrocarbon Products Mass Flow Measurement, Media Identification and Leak Detection

Fundamentals of the Ultrasonic Flowmeter

RUN ACCUM. TOTAL STOP BAT LOW HIGH

A Different Approach to Gas Turbine Exhaust Silencing

Comparison between Fluid Viscous Dampers and Friction Damper Devices. Fluid Viscous Dampers (FVD) Friction Damper Device (FDD) Working principle:

SWIRL MEASURING EQUIPMENT FOR DIRECT INJECTION DIESEL ENGINE

Module7:Advanced Combustion Systems and Alternative Powerplants Lecture 32:Stratified Charge Engines

FUNDAMENTALS OF GAS TURBINE METERS Paul Honchar Sr. Product Manager Sensus 805 Liberty Boulevard DuBois, PA, USA 15801

The Knowledge Bank at The Ohio State University. Ohio State Engineer. Electrolysis in Underground Structures

Best Practice Variable Speed Pump Systems

three different ways, so it is important to be aware of how flow is to be specified

PVP Field Calibration and Accuracy of Torque Wrenches. Proceedings of ASME PVP ASME Pressure Vessel and Piping Conference PVP2011-

VORTAB FLOW CONDITIONERS. For Flow Meters, Pumps and Other Flow Profile Critical Equipment

Development of Noise-reducing Wheel

Lecture 3.3. Velocity, motion, force and pressure sensors

Condensed Version of Operator s Manual For The VI/VP700 Series Flowmeter and VI800 Series Flowmeter

FUNDAMENTALS OF ORIFICE METERING Ken Embry FMC Measurement Solutions

GE Panametrics Ultrasonic Flow Meter

Legal Metrology Control of a Liquefied Petroleum Gas Transfer Standard and Field Practice

Jet Dispensing Underfills for Stacked Die Applications

Advancements in Compressor Anti-surge Control Valve Solutions

Consideration of Anti-Vibration Performance Improvement of a Servo Motor

4. Manual Sampling of Petroleum and Petroleum Products Introduction Recommended Procedures for Manual Sampling of Tanks...

United States Patent (19) 11 Patent Number: 5,780,736 Russell 45) Date of Patent: Jul. 14, 1998

Permanent Multipath Clamp-On Transit Time Flow Meter

Troubleshooting, Service Tips, And Major Improvements For Hydrostatic Transmissions (Special Edition){3200}

Fundamentals of Turbine Meters. Paul Honchar Sensus Metering Systems

Differential Expansion Measurements on Large Steam Turbines

Multistage gearbox failure

...particularly for OEM projects

37 th Gas-Lift Workshop Houston, Texas, USA February 3 7, 2014

EXPERIMENTAL INVESTIGATION OF THE FLOWFIELD OF DUCT FLOW WITH AN INCLINED JET INJECTION DIFFERENCE BETWEEN FLOWFIELDS WITH AND WITHOUT A GUIDE VANE

Minimizing Noise. Marybeth G. Nored Dr. Klaus Brun Eugene Buddy Broerman Augusto Garcia Hernandez Dennis Tweten. Southwest Research Institute

hydraulic flow meter Author:

Cane Creek Double Barrel Instructions

High Speed Reciprocating Compressors The Importance of Interactive Modeling

FEASIBILITY STYDY OF CHAIN DRIVE IN WATER HYDRAULIC ROTARY JOINT

HTM Series Helical Turbine Meters

MISHAAL ALANZI Gas Engineer, KOC

Effect of Compressor Inlet Temperature on Cycle Performance for a Supercritical Carbon Dioxide Brayton Cycle

SILENCER SELECTION INSTRUCTIONS

MECHATRONICS LAB MANUAL

IMPACT REGISTER, INC. PRECISION BUILT RECORDERS SINCE 1914

EZ-IN Series Turbine Flowmeter

Preparing the New EuroLoop Liquid Flow Facility for Accreditation. Jos van der Grinten - NMi EuroLoop

Discussed in this chapter are

Fundamental Training. Flow Con t

Claim Cost Reduction. The planned down-time of a turbine is typically. If dual fuel turbines gauges or billing from the. seeing more penalty clau-

Effect of Sample Size and Method of Sampling Pig Weights on the Accuracy of Estimating the Mean Weight of the Population 1

F-4600 INLINE ULTRASONIC FLOW METER Installation and Operation Guide

High Capacity Tapered Roller Bearings

Vibration product training Gearbox tutorial

FINAL REPORT MARCH 2008

Chapter 11. Control System Instrumentation

OBSERVATIONS ABOUT ROTATING AND RECIPROCATING EQUIPMENT

Orifice Fittings and Meter Tubes

Influence of Micro-Bubbles within Ejected Liquid on Behavior of Cavitating Flow inside Nozzle Hole and Liquid Jet Atomization

Commercial/Industrial Ultrasonic Gas Meter Operation & Diagnostics Western Regional Gas Conference Bayview Ballroom 2 8/29/2017 9:00 9:45

Transcription:

M E A S U R E M E N T S Y S T E M S Technical Paper No. 162 Rev. 0 Proving Liquid Ultrasonic Flowmeters Summary of Testing Conducted January 2011 LEFM 280CiRN

Measurement Systems Introduction Proving Liquid Ultrasonic Flowmeters Summary of Testing Conducted January 2011 Ultrasonic flowmeters have been used for precise liquid petroleum measurement for nearly two decades. They have several advantages over conventional meters that have been instrumental in securing their place in the industry. Some of these advantages include their ability to be made in large sizes to measure high volume flowstreams. Because ultrasonic flowmeters are all electronic, there are no moving parts to wear out. The meter has no intrusions into the flowstream to catch debris and therefore, ultrasonic flowmeters have a very low pressure drop. Some ultrasonic flowmeters can be characterized over a range of Reynolds Numbers allowing then to provide accurate measurements regardless of flowrate or liquid viscosities that fall within the characterized range. However, one of the shortcomings of the ultrasonic flowmeter has been a relatively poor repeatability when being proved. Ultrasonic flowmeters measure flow by sending pulses of ultra-high frequency sound back and forth across the flowstream at an angle so that in one direction the pulses must travel upstream, while in the other direction they must travel downstream. The difference is these transit times allows a measure of the average flowstream velocity along the path. Figure 1 is a clear view of an ultrasonic flowmeter showing an array of acoustic transducers forming four paths across the flowstream. Figure 1 - Caldon LEFM 240Ci Ultrasonic Flowmeter (Standard) Unless the liquid has a very high viscosity and/or a very low velocity, where the flowstream can be in the laminar region (Reynolds Number under around 10,000), there will be turbulence. This turbulence is made up of eddy currents with a wide spectrum of frequencies and amplitudes. The highest intensity and lowest amplitude eddies are near the inside surface of the pipe wall while the lowest intensity and highest amplitude eddies are in the central portion of the flowstream - See Figure 2 TP162 Rev. 0 1

Measurement Systems Figure 2 - Turbulence within the Flowstream for Re > 10,000 The transit times of acoustic pulses passing through this turbulence, especially the eddies that are larger than the diameter of the acoustic beam, are caused to vary, depending upon the particular alignment of the eddy. This transit time variation is the root cause of poor proving repeatability. Testing Figure 3 shows the performance of a 6 inch four-path ultrasonic flowmeter as tested on 2.0 cst oil at the Caldon Ultrasonic Technology Center in Coraopolis, Pennsylvania. The prover was a Brooks 18 Compact prover with a base volume of 0.72 BBL. The proving was conducted with 5 passes per run and 5 runs per proving. Consider the prover volume in this case to be 3.6 BBL (5 x 0.72 BBL). During the linearity testing there were 29 provings at 5 different flowrates of which only 1 met the repeatability criteria of 0.05%. The average repeatability of the 29 provings was 0.149%. Most of the provings (26 of 29) had repeatability under 0.20%. It has been established that proving repeatability is a function of the reciprocal of the square root of prover volume 1, i.e.: ͳ ݕݐ ݐ Where V is the prover volume Using this relationship to estimate the prover volume required to meet the criteria of 5 runs agreeing within 0.05% repeatability, the repeatability will need to improve by a factor of 4 (from 0.20% to 0.05%), the prover volume would need to increase by a factor of 4 2 = 16. In order to increase the apparent prover volume by 16 times, it will be necessary to have about 70 passes per run ((16 x 3.2)/0.72). Figure 4 shows a diagram of an 8-path ultrasonic flowmeter where the meter bore is located downstream of a compound reducing nozzle. After the acoustic transit times are measured, the bore is gradually increased back to the original size which recovers most of the pressure drop caused by the restriction. Figure 5 shows a 6 LEFM 280CiRN ultrasonic flowmeter that employs the features described above. This meter differs from the one above in two fundamental ways; it has 8 acoustic paths (the LEFM 240Ci Standard meter has 4 acoustic paths) and within the throat of the meter there is a reducing nozzle that constricts the flowstream (the Standard meter has a full diameter bore). 1 Don Augenstein, Class 2430 Proving Liquid Ultrasonic Flow Meters, 2007 ISHM, Oklahoma City, OK TP162 Rev. 0 2

Measurement Systems 1.0060 1.0040 6" LEFM 240Ci (Standard) 0.72 BBL Small Volume Prover Repeatability of 5 Runs (5 Passes per Run) 0.50% 0.40% MeterFactor 1.0020 1.0000 0.30% 0.20% Repeatability 0.9980 0.10% 0.9960 0 500 1000 1500 2000 2500 3000 Flowrate (BPH) 0.00% Figure 3-6" Caldon LEFM 240Ci Linearity and Repeatability with an 0.72 BBL SVP Figure 4-8-Path Reducing Nozzle and Recovery Cone TP162 Rev. 0 3

Measurement Systems Figure 5-6" Caldon LEFM 280CiRN In essence, it is like there are two 4-path meters measuring the same flowstream, so it is reasonable to take the root-mean-square of the two readings which improves the repeatability by the factor of the square root of 2 (1.414). It is also believed that the reducing nozzle tends to stretch the turbulent features in the axial direction, reducing their amplitude. This reduction tends to reduce the variations in the transit times and results in an improvement in the repeatability as well. The test results (Figure 6) show this meter has better repeatability than the standard meter. The average repeatability improved to 0.092% and nine (9) of the thirty (30) proving had repeatabilities under 0.10%. Twenty-nine (29) of thirty (30) provings had repeatabilities under 0.15%. The explanation for this improvement is because the 8-path meter is making more measurements of the flowstream. In order to reduce the variation in transit times due to turbulent features within the flowstream, research led to the development of a turbulence conditioner 2. This device, also called an RIE (Repeatability Improving Element) consists of a matrix of small apertures, like a honey comb, that break up large turbulent feature, see Figure 7. The RIE is placed immediately upstream of the meter with the idea that turbulent eddies will be reduced in size momentarily and if they are smaller in size than the acoustic beam they will be dissolved and cease to cause, or certainly reduce the variations in the transit times, thereby improving the repeatability of proving runs. 2 Gregor Brown, et al, Turbulence Conditioner for Ultrasonic Flow Meter and Method U.S. Patent 7,823,462, issued Nov. 2, 2010 TP162 Rev. 0 4

Measurement Systems 6" LEFM 280CiRN 0.72 BBL Small Volume Prover 1.0060 0.50% 1.0040 0.40% 1.0020 0.30% 1.0000 0.20% 0.9980 0.10% 0.9960 0.00% 0 500 1000 1500 2000 2500 Flowrate (BPH) Figure 6 - LEFM 280CiRN Test Results Show Improved Repeatability Figure 7-6" Repeatability Improving Element (RIE) TP162 Rev. 0 5 3000 Repeatability Repeatability of 5 Runs (5 Passes per Run)

Measurement Systems Figure 8 shows the test result for the 6 LEFM 280CiRN with the RIE installed immediately upstream of the meter. As can be seen, the repeatability has substantially improved with the average being 0.058% with twenty-six (26) of forty-two (42) proving having repeatability under 0.05%. Thirty-six (36) of fortytwo (42) proving had repeatabilities under 0.10%. These proving results demonstrated that the Caldon LEFM 280CiRN fitted with the RIE can be proved with an ordinary sized small volume prover with only 25 strokes of the displacer, i.e., 5 runs with 5 passes per run. When compared to the Standard meter that was calculated to require 350 strokes of the displacer (5 runs with 70 passes per run) in order to meet the 0.05% repeatability requirement, this is a truly significant improvement. Many in the industry would agree that test results show that this meter is nearly as easy to prove and a turbine meter. 6" LEFM 280CiRN w/rie 0.72 BBL Small Volume Prover 1.0020 0.50% by Master Meter w/~5% Flowrate Variation Repeatability of 5 Runs (5 Passes per Run) 0.40% 1.0000 0.30% 0.9980 0.20% 0.9960 0.10% 0.9940 0.00% 0 500 1000 1500 2000 2500 Repeatability 1.0040 3000 Flowrate (BPH) Figure 8-6" LEFM 280CiRN with RIE Test Results In order to establish that the linearity results were not being biased because of the SVP, it was decided to prove the ultrasonic meter using a turbine meter that had been proved with the same SVP. Figure 8 shows that the meter factors at three rates fall in line with the meter factors taken directly using the SVP. It has been established within the industry that flowrate changes during a proving run can result in bias errors in meter factor, particularly when using a SVP, since the prove times are often quite small. This error can be summarized as follows3. ݎ ݎݎܧ Where: Meter s Time Constant ( ʹ ͳ) ݒ ݎ ݐ Q2 = Flowrate as measured by the meter at the end of the proving pass Q1 = Flowrate as measured by the meter at the beginning of the proving pass What is interesting about the bias error phenomenon is that all of the variables needed to determine the maginitude of the error are known by the meter, i.e., the meter s time constant, the flow rate at the end of the proving pass and the flow rate at the beginning of the proving pass. Therefore, it is possible to correct 3 Don Augenstein, Class 2430 Proving Liquid Ultrasonic Flow Meters, 2007 ISHM, Oklahoma City, OK TP162 Rev. 0 6

Measurement Systems for this error when proving the meter4. Also shown in Figure 8 is a test point at high flow rate where the proving run time is short and the potential for bias error is greatest, the flowrate was deliberately varied rapidly by about 5%. As can be seen, since the bias correction was employed, the meter factor determined was in line with those previously determined. Figure 9 shows the 16 Caldon LEFM 280CiRN installed in the test loop. Unfortunately, a 16 in. standard meter was not available for this testing; however, the current version of API MPMS Chapter 5.8 shows that in order to prove a 16 ultrasonic meter in 5 proving runs, a 522 BBL prover would be required. Figure 9-16" Caldon LEFM 280CiRN Ultrasonic Flowmeter Figure 10 shows the test results of a 16 Caldon LEFM 280CiRN meter proved using 2 cst oil with the 63 BBL uni-directional pipe prover at the Caldon Ultrasonic Technology Center. A linearity curve was generated at five flowrates involving 62 proving (5 runs per proving). In twenty-nine (29) of thirty-six (36) provings the repeatability criteria of 0.05% was met. The average repeatability was 0.040%. In other words, it is possible to prove the 16 Caldon LEFM 280CiRN using the criteria of 5 proving agreeing within a repeatability of 0.05% over 80% of the time. This is about what would be expected with a 16 turbine meter on this size of prover. 4 Donald R. Augenstein, et al, Method, Apparatus and Computer Medium for Correcting Transient Flow Errors in Flowmeter Proving Data, U.S. Patent 7,366,625, issued Apr. 29, 2008 TP162 Rev. 0 7

Measurement Systems 16" LEFM 280CiRN 63 BBL Prover 1.0060 0.25% 1.0040 0.20% 1.0020 0.15% 1.0000 0.10% 0.9980 0.05% 0.9960 Repeatability Repeatability of 5 Runs 0.00% 0 2000 4000 6000 8000 10000 12000 14000 Flowrate (BPH) Figure 10-16" Caldon LEFM 280CiRN Test Results Figure 11 shows a 16 RIE. Notice that the aperture size is still about the same size as that used for the 6 RIE. This is because the acoustic transducers used on the 16 meter are the same diameter as those of the 6 meter and therefore, the acoustic beam is the same size. If larger diameter acoustic transducers were installed in the meter, the apertures of the RIE could also be larger. Figure 10-16" RIE Figure 12 shows the test results of the 16 LEFM 280CiRN with the RIE installed immediately upstream. Notice that the prover size is now only 42 BBL. This is because the previous test had already established TP162 Rev. 0 8

Measurement Systems that this meter could meet the 5 run, 0.05% criteria. Therefore, in this test with the RIE, the prover volume was reduced. The prover volume reduction is possible without actually changing the prover because the 63 BBL uni-directional prover is fitted with intermediate detector switches at 21 BBL and 42 BBL and these volumes each carry their independent calibrations. A linearity curve was determined at 5 flow rates over forty-eight (48) proving. The average repeatability was 0.045% with thirty-four (34) of forty-eight (48) (over 70%) meeting the 0.05% criteria. Most in the industry would consider this success ratio acceptable. The ability of the 16 LEFM 280CiRN fitted with the RIE to be proved meeting the 5 run, 0.05% criteria would certainly match, and possibly exceed that expected from a 16 turbine meter being proved on a 42 BBL prover. 16" LEFM 280CiRN w/rie 42 BBL Prover 1.0040 0.25% 0.20% 1.0000 0.15% 0.9980 0.10% 0.9960 0.05% 0.9940 0.00% 0 2000 4000 6000 8000 10000 12000 Repeatability Repeatability of 5 Runs 1.0020 14000 Flowrate (BPH) Figure 11-16" Caldon LEFM 280CiRN with RIE Test Results Conclusions Many in the industry have expressed reluctance to use ultrasonic flowmeters in critical applications because of the difficulty in proving brought about by poor proving repeatability. The solution to this proving difficulty offered in API MPMS Chapter 4.8 of opening the repeatability tolerance and taking more proving runs has not be satisfactory. First of all, it takes more time and causes more wear on the prover. Also, it is well established that there are poor repeatability causes unrelated to the meter, e.g., electronic noise, chipped prover ball, leaking four-way valve, product flashing, etc. These alternate causes of poor repeatability can be masked when the repeatability limit is increase. What is really needed is an ultrasonic flowmeter that can be easily proved using the traditional technique of 5 runs with a maximum repeatability of 0.05%. While the testing described above was conducted by Cameron at the Caldon Ultrasonic Technology Center near Pittsburgh, Pennsylvania, it was planned and witnessed by a joint industry group (JIG) representing four major oil and pipeline companies and one service Company. As can be seen from the data presented above, the results conclusively establish that the Caldon LEFM 280CiRN meter is not only significantly easier to prove that the standard ultrasonic meter but can be proved in 5 runs with a prover no larger in volume that that used by the conventional PD meter or turbine meter. TP162 Rev. 0 9