Customer Information. Service Engineer s Responsibilities. Agilent Cary 7000 UMS/UMA/Solids Autosampler
|
|
- Harvey Morgan
- 5 years ago
- Views:
Transcription
1 Agilent Preventive Maintenance provides factory recommended service for your analytical systems to assure reliable operation and the accuracy of your results. Delivered by highly-trained and certified service engineers using genuine Agilent parts and supplies, Agilent Preventive Maintenance provides everything you need to reduce unplanned downtime and keep your systems operating at their peak. For more information about services please visit our web site using the following URL Customer Information Customers should provide all necessary operating supplies upon request of the engineer. A customer representative should be available to the engineer while performing the preventive maintenance procedures. Any parts, not included in the Parts Lists section of this document, are not part of the recommended Preventive Maintenance service, nor are they included in the price of this service. If a system requires the use of additional or special procedures and/or parts for the instrument service, then these must be ordered separately and charged as a repair, which may incur additional costs. Service Engineer s Responsibilities Only complete/printout pages that relate to the system or module being serviced. Complete empty fields with the relevant information. Complete the relevant checkboxes in the checklist using a X or tick mark in the checkbox. Complete Not Applicable check boxes to indicate services not delivered, as needed. Complete the PM service in the order of the tasks listed. Complete the Service Review section together with the customer. If the PM service is performed prior to a qualification service, then ignore this checklist and use the qualification procedure detailed in ACE. Agilent Cary 7000 UMS/UMA/Solids Autosampler This documentation covers Agilent part numbers; o G6873AA Cary 7000 UMS o G6875AA Cary Universal Measurement Accessory (UMA) o G6876A Solids Autosampler Accessory Page 1 of 7
2 System Information Guidance Check this box if an instrument configuration report is attached instead of completing the table. Instrument system name and ID Instrument system site and location List system component product numbers List the serial numbers of each component Cary 7000 UMS: Parts List Table Part Description Part Number Product or Model# where used Quantity Deuterium lamp Cary Visible lamp Cary Preparation Discuss any specific issues with the customer prior to starting. Review the instrument logbook. Perform general inspection of system for cleanliness. Check for proper installation of safety-related parts, assemblies, sensors etc. Check for required firmware updates and verify with customers if they would like it installed. Run the relevant performance tests prior to starting any maintenance procedure Page 2 of 7
3 Cary 7000 instrument Perform Validate instrument performance tests. Inspections With all spectrophotometer covers removed; Remove dust from optical casting. Remove dust from internal components, electronic printed circuit boards and surrounding areas. Clean all monochromator related opto-couplers. Clean sample compartment windows. Clean sample compartment area. Inspect the following items. Make a note in the comments section as a possible follow up action if any part requires replacement. Source mirror optics inspection. Wavelength micrometer inspection. Micrometer pulley inspection. Wavelength drive belt inspection. Wavelength motor coupling inspection. Instrument cooling fan inspection. Restore Instrument Replace and clean spectrophotometer external cover. Return instrument to initial operating condition. Replacements and Adjustments Replace and optimize UV source lamp. Replace and optimize Visible source lamp. Adjust wavelength drive belt tension. Operational Checks Upgrade firmware if applicable. Use Validate to run all applicable calibrations. Instrument Performance Tests Page 3 of 7
4 Universal Measurement Accessory (UMA) Perform Validate UMA performance tests. Clean and Inspect With the UMA cover removed; Wipe dust from optics enclosure. Wipe dust from UMA sample stage and surrounding areas. Examine all I/O cables and ensure connections are secure. Restore Instrument Wipe external cover. Return accessory to initial operating condition. UMA Optical Checks and Adjustments Optical axis of rotation (concentricity). Beam height. Beam tilt. Operational Checks Upgrade UMA firmware if applicable. UMA detector calibration. UMA sample stage calibration. Automated Polarizer calibration. Sample holder tilt. UMA Performance Tests. Fused silica test Page 4 of 7
5 Solids Autosampler Accessory Inspections Cable inspection. Control Box fuse. Restore Instrument Return accessory to initial operating condition. Operational Checks Control Box indicators. Autosampler positional accuracy verification. Autosampler stage calibration. Autosampler tilt. Fused silica test Page 5 of 7
6 Service Review Attach available reports/printouts of all tests to this documentation. Record the PM service activity in the customer's instrument records/logbook. Affix the PM sticker to the system or instrument logbook based on the customer's request. Complete the Service Review Comments section below if there are additional comments. Review the service and any test results with the customer. If the Instrument firmware was updated, record the details of the change in the Service Engineer's Comments box below. Cary 7000 instrument: Test Results Table UV source lamp conditions Result Signal (%T) Energy Beam Mode Slit Height Ave time (sec) Wavelength (nm) SBW (nm) Single Front Full 0.1sec 235nm 1.0nm Vis source lamp conditions Result Signal (%T) Energy Beam Mode Slit Height Ave time (sec) Wavelength (nm) SBW (nm) Single Front Full 0.1sec 500nm 1.0nm Page 6 of 7
7 Service engineer comments (optional) If there are any specific points you wish to note as part of performing the preventive maintenance or other items of interest for the customer, please write in this box. Other Important Customer Web Links How to get information on your product: Literature Library - Need to know more? - Need technical support, FAQs? - Need supplies? - Service completion Service request number Date service completed Agilent signature Customer signature Document part number: G Page 7 of 7
For more information about Agilent Technologies services please visit our web site using the following URL
Agilent Preventive Maintenance provides factory recommended service for your analytical systems to assure reliable operation and the accuracy of your results. Delivered by highly-trained and certified
More informationPlease exercise care when using solvents close to the instrument. It could damage the vapor sensor present in the ELSD.
Infinity II - Standard Preventive Maintenance Service Agilent Preventive Maintenance (PM) provides factory recommended service for your systems to assure reliable operation and the accuracy of your results.
More informationCustomer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard
Agilent Preventive Maintenance provides factory recommended service for your analytical systems to assure reliable operation and the accuracy of your results. Delivered by highly-trained and certified
More informationCustomer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard
7890 GC Standard Agilent Preventive Maintenance provides factory recommended service for your analytical systems to assure reliable operation and the accuracy of your results. Delivered by highly-trained
More informationGenesys 20 Calibration and Maintenance
NOTICE: This document is version controlled and was produced as a part of the GEX Information Program which requires that all Series 100 documents be reviewed periodically to maintain currency and continuity
More informationAgilent InfinityLab Quick Change Valve G4237A
Agilent InfinityLab Quick Change Valve G4237A Instructions Agilent G4237A Valve Kit Instructions This technical note describes the installation and application of the Agilent InfinityLab Quick Change 4
More informationS 1125 HPLC Pump System
Routine S 170 EXPERT HPLC HPLC System SYSTEMS 500 The Sykam ROUTINE HPLC System S 500 is intended for all routine analysis. The system configuration is highly variable and several upgrade options make
More informationAtomic Absorption Spectroscopy
Atomic Absorption Spectroscopy Atomic absorption spectroscopy (AAS) is a widely used technique for determining a large number of metals. In the most common implementation of AAS, a liquid sample containing
More informationAgilent G4236A 2D-LC Valve Installation Guide
Agilent G4236A 2D-LC Valve Installation Guide Technical Note Contents Delivery Checklist 2 Installing Valve Heads 3 Removing the Transportation Lock and the Valve Dummy 3 Installing the Valve Head and
More informationAgilent G5632 Valve Kit Instructions
Agilent G5632 Valve Kit Instructions Agilent G5632 Valve Kit Instructions Contents Typical Applications 2 Alternating Column Regeneration (2 position/10 port valves only) 2 Dual column selection 3 Sample
More informationC196-E081C. Ultra High Performance Liquid Chromatograph Nexera X2. Specifications
C196-E081C Ultra High Performance Liquid Chromatograph Nexera X2 Specifications Specifications System Controllers CBM-20A / 20Alite CBM-20A (228-45012-XX) CBM-20Alite (228-45011-XX) Connectable units Solvent
More informationAgilent G4239C Valve Kit - Instructions
Agilent G4239C Valve Kit - Instructions Agilent G4239C Valve Kit - Instructions Technical Information about the Agilent Valve Kit G4239C. Contents Typical Applications 2 Multicolumn Selection 2 Method
More informationInstallation of the 2 position/10 port valve into the 1100 series thermostatted column compartment
Agilent G1316A Thermostatted Column Compartment Technical Note Installation of the 2 position/10 port valve into the 1100 series thermostatted column compartment System Requirements 2 Performance Specifications
More informationDeuterium and Tungsten-Halogen Hybrid Light Source. Operation Manual
Deuterium and Tungsten-Halogen Hybrid Light Source - ASBN-D1-WXXX / ASBN-D2-WXXX- Operation Manual Important Safety Notices 1. Never look directly into the light beam, including through the cooling fan
More informationAgilent InfinityLab Quick Change Valves G4231A/C and G4232C/D
Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D Instructions Agilent Valve Kit G4231A/C and G4232C/D - Instructions This technical note describes the installation and application of the Agilent
More informationAvaLight-DHS Deuterium Halogen Light Source Operating manual
AvaLight-DHS Deuterium Halogen Light Source Operating manual Version 1.2 February 2006 14-02-06 AvaLight-DHS manual.doc 1 AvaLight-DHS Deuterium Halogen Light Source Safety Instructions Warning: All warnings
More informationAgilent Online-SPE Direct Inject Kit G4744A
Agilent Online-SPE Direct Inject Kit G4744A Installation Guide Agilent Online-SPE Direct Inject Kit G4744A - Installation Guide Contents Application 2 Delivery Checklist 4 Online SPE Direct Inject Kit
More informationNEW HPLC - SYSTEM LINE
NEW HPLC - SYSTEM LINE S c h a m b e c k S F D G m b H Drieschweg 13A D-53604 Bad Honnef/Germany Phone: +49 2224 9239-0 E-Mail: info@schambeck-sfd.com Internet: www.schambeck-sfd.com S 9425 HPLC solvent
More informationDH-2000 Deuterium-Halogen Light Source Installation and Operation Manual
DH-2000 Deuterium-Halogen Light Source Installation and Operation Manual For Products: DH-2000 / DH-2000-BAL / DH-2000-DUV / DH-2000-S / DH-2000-S-DUV / DH-2000-FHS, DH-2000-FHS-DUV /DH-2000-S-DUV-TTL
More informationAgilent InfinityLab Quick Change Valve G4235A Instructions
Agilent InfinityLab Quick Change Valve G4235A Instructions Contents Typical Application 2 Delivery Checklist 3 Specifications 4 Installing the Valve Heads 5 Agilent Technologies Typical Application This
More informationMB3000. Versatile FT-IR analyzer designed for your industry The most reliable FT-IR, the lowest cost of ownership.
MB3000 Versatile FT-IR analyzer designed for your industry The most reliable FT-IR, the lowest cost of ownership. MB3000 FT-IR Analyzer designed for research and development Our company needs versatility
More informationone filter having K between 1.5 and 2.5 m -1
CHAPTER IV : CONFORMTY OF PRODUCTION PROCEDURE FOR TESTING OF SMOKE METER. 1 Physical Check It shall consist of checking - 1) Suitability and label on the instrument for the intended use. 2) Identification
More informationElectro Optical Components, Inc Skylane Boulevard, Suite D, Santa Rosa, CA Toll Free: 855-EOC
Electro Optical Components, Inc. 5464 Skylane Boulevard, Suite D, Santa Rosa, CA 95403 Toll Free: 855-EOC-6300 www.eoc-inc.com info@eoc-inc.com [ user manual ] Fiber Coupler id1-fc Dual-mirror Fiber Coupler
More informationSpectrAA Report. 2:04 AM 11/25/2014 Page 1 of 22
SpectrAA Report. 2:04 AM 11/25/2014 Page 1 of 22 Analyst Date Started Worksheet Comment Methods Computer name Serial Number: 9:21 PM 11/24/2014 CH461 - Pb 11_21_14 cp 02 Pb,Pb E3ED16988D8C4E9 AA1110M047
More informationTHERMAL SHOCK TEST REPORT
Report No. :DH1509300024A Version:A THERMAL SHOCK TEST REPORT Company :FLIP CHIP OPTO. INC. Address :No.185, Kewang Rd., Longtan Dist., Taoyuan City 325, Taiwan (R.O.C.) Model Name :Apollo 600-B Lot Number
More informationMotion System Components Diagram. Note: #2 Mirror Cover and X-Axis Motor Cover have been removed for visibility. Maintenance.
Professional Laser System PLS3.75, PLS4.75, PLS6.75 and PLS6.150D Keeping the laser system clean will ensure the highest quality engraving. A clean laser system is the best performing laser system. The
More informationAnalysis of biodiesel oil (as per ASTM D6751 & EN 14214) using the Agilent 5100 SVDV ICP-OES
Analysis of biodiesel oil (as per ASTM D6751 & EN 14214) using the Agilent 5100 SVDV ICP-OES Application note Petrochemical Author Neli Drvodelic Agilent Technologies Melbourne, Australia Introduction
More informationAgilent Valve Kit G4231A/B and G4232A/B/C
Agilent Valve Kit G4231A/B and G4232A/B/C Instructions Contents Typical Applications 2 Dual column selection (G4231A/B or G4232A/B/C) 2 Sample enrichment and sample cleanup (G4231A/B or G4232A/B/C) 2 Alternating
More informationPhysiological Calibration Standard II Calibration standard physiological fluid CALIBRATION STANDARD PROTEIN
80-6000-32 Physiological Calibration Standard II 80-2037-62 Calibration standard physiological fluid 80-2037-61 CALIBRATION STANDARD PROTEIN HYDROLYSATE 80-2104-53 1/8" FLANGELESS FITTING (PK 5) 80-2108-41
More informationHI Ethylene Glycol Refractometer INSTRUCTION MANUAL
HI96831 Ethylene Glycol Refractometer INSTRUCTION MANUAL Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using this instrument. This
More informationHPLC System S 600 Series
HPLC System S 600 Series The Sykam HPLC System S 600 Series is intended for the ambitious analyst. The system configuration is highly variable and several upgrade options make this system suitable for
More informationEND OF YEAR WAREHOUSE CLEARANCE SALE! Ex Demo & Superseded Instrumentation
END OF YEAR WAREHOUSE CLEARANCE SALE! Ex Demo & Superseded Instrumentation Stock. Ends 31st Dec Dynamica Halo VIS-10 Scanning Spectrophotometer Wavelength range: 330-999nm Single Beam Concave diffraction
More informationAgilent 1290 Infinity Autosampler
Agilent 1290 Infinity Autosampler Installation of the Large Volume Injection Kit General Information This note describes how to install the large volume injection kit into an Agilent 1290 Infinity Autosampler.
More informationLuzchem Solar Simulator SolSim2 Updated September 2018
Luzchem Solar Simulator SolSim2 Updated September 2018 Luzchem s SolSim2 provides solar simulating radiation based on a powerful 300W ceramic xenon lamp with built-in mirrors for exceptional luminous output.
More information020: 2013 CEB SPECIFICATION MINIATURE CIRCUIT BREAKER (MCB)
020: 2013 CEB SPECIFICATION MINIATURE CIRCUIT BREAKER (MCB) CEYLON ELECTRICITY BOARD SRI LANKA Telephone: +94 11 232 0953 Fax: +94 11 232 3935 CONTENTS Page 1.0 Scope 3 2.0 System Parameters 3 3.0 Service
More informationUser manual. Fiber Probe Coupler FPC-6M. Senary-mirror Fiber Probe Coupler suitable for Nicolet is10(50), Nicolet 6700 & Avatar 360 FTIR spectrometers
User manual Fiber Probe Coupler FPC-6M Senary-mirror Fiber Probe Coupler suitable for Nicolet is10(50), Nicolet 6700 & Avatar 360 FTIR spectrometers 1 art photonics GmbH Rudower Chaussee 46 12489 Berlin,
More informationDeep UV Deuterium Light Source. Operation Manual
Deep UV Deuterium Light Source - ASBN-D130 / ASBN-D230 - Operation Manual Important Safety Notices 1. Never look directly into the UV light, as this can cause eye damage. 2. Do not remove or modify any
More informationHPLC SYSTEM S 500 SERIES
HPLC SYSTEM S 500 SERIES The Sykam HPLC System S 500 Series is intended for all routine analysis as well as for the ambitious analyst. The system configuration is highly variable and several upgrade options
More information30W High Stability Tungsten- Halogen Light Source - ASB-W Operation Manual
30W High Stability Tungsten- Halogen Light Source - ASB-W-030 - Operation Manual Important Safety Notices 1. Never look directly into the light beam, including through the cooling fan while light is on,
More informationHigh Power, RF Excited EUV Flow Light Source EUV X LHP500FL Operating Manual
Document RD 15 11 No: High Power, RF Excited EUV Flow Light Source EUV X LHP500FL Operating Manual Manual for EUV X LHP500FL 1 Table of Contents Table of Contents Overview Calibration System Diagram Flow
More informationAgilent Valve Kit G4231A/B/C and G4232B/C/D
Agilent Valve Kit G4231A/B/C and G4232B/C/D Instructions Agilent Valve Kit G4231A/B/C and G4232B/C/D - Instructions This Technical Note describes the installation of the Agilent G4231A/B/C and G4232B/C/D
More informationLAR Training. Series 8000 Pneumatic Safety Shut Off Valves Mark Lampe May,
LAR Training Series 8000 Pneumatic Safety Shut Off Valves Mark Lampe May, 18 2015 Series 8000 Pneumatic Valve Unique vertical space saving design Maintenance free gas and fluid seal that eliminates packing
More informationInstructions for Mounting and Operation of ARC Model TS Watt Tungsten-Halogen Light Source
15 Discovery Way, Acton, MA 01720 Phone: (978)263-3584, Fax: (978)263-5086 Web Site: www.acton-research.com www.piacton.com Instructions for Mounting and Operation of ARC Model TS-428 250 Watt Tungsten-Halogen
More informationEnergy efficient low voltage dichroic mirror halogen reflector lamps
GE Lighting ConstantColor Precise IR Energy efficient low voltage dichroic mirror halogen reflector lamps DATA SHEET Description Precise MR16 lamps are low voltage tungsten halogen reflector-mounted lamps
More informationTABLE OF CONTENTS NOTES. 1.0 Operating principle... 02
OI-001 NOTES TABLE OF CONTENTS 1.0 Operating principle.............................................. 02 2.0 Overview....................................................... 03 2.1 Operating elements 2.2
More informationInstructions for Mounting and Operation of the ARC Model XS Wall Xenon Light Source System
530 Main Street, Acton, MA 01720 Phone: (978)263-3584, Fax: (978)263-5086 Web Site: www.acton-research.com Instructions for Mounting and Operation of the ARC Model XS-432 75 Wall Xenon Light Source System
More informationCNC Tool Room Milling Center MODEL: VM-760
CNC Tool Room Milling Center MODEL: VM-760 MACHINE FEATURES 7.5HP high torque spindle motor, low noise and vibration BT40/CAT40/DIN40 6000 rpm spindle speed, ( Optional 8,000, 10,00, 12,000 or 15,000rpm
More informationBallast Water Management Surveyor checklist
Ballast Water Management Surveyor checklist Last updated: July 2017 This document is uncontrolled if printed, please refer to the Maritime New Zealand website for the latest version. Use this checklist
More informationTHE ISAAC NEWTON GROUP OF TELESCOPES INGRID AT CASS
THE ISAAC NEWTON GROUP OF TELESCOPES INSTRUMENT CHANGE CHECKLIST INGRID AT CASS Revision: 1.2 Date: 24/09/2002 Authors: P Jolley, M Blanken, S Rees Description: Location: Storage: ING Infra Red Detector
More informationUV Transilluminator Dual Wavelength (Large)
BT505 A Geno Technology, Inc. (USA) brand name UV Transilluminator Dual Wavelength (Large) Cat. No. BT505 1-800-628-7730 1-314-991-6034 info@btlabsystems.com INTRODUCTION BT Lab Systems UV Transilluminators
More informationInstallation Note. Agilent Technologies E4440A Spectrum Analyzer Option BAB, APC 3.5 Input Connector Retrofit Kit
Installation Note Agilent Technologies E4440A Spectrum Analyzer Option BAB, APC 3.5 Input Connector Retrofit Kit Part Number E4440-90569 Supersedes: E4440-90092 Printed in USA August 2003 Notice. The information
More informationTECHNICAL REPORT. Introduction. Agilent 1100 HPLC system. Figure 1: HALO columns exhibit UHPLC-like performance at conventional HPLC pressure
TECHNICAL REPORT Modifying Agilent 1100 HPLC Systems to Achieve UHPLC-like performance with HALO Fused-Core Columns Agilent 1100 HPLC system With a few modifications, an Agilent 1100 HPLC can produce UHPLC-like
More informationFWT-200 OPTI-CHROMIC READER SYSTEM
FWT-200 OPTI-CHROMIC READER SYSTEM User s Manual Covers Installation and Windows software for ISA and PCI cards September, 2002 Far West Technology, Inc. 330 South Kellogg Ave. Suite D Goleta, CA 93117
More informationOperation Manual LAM 10M IR 3A LASER AIMING MODULE / TARGET POINTER. 105 Sparks Ave., Toronto, ON, M2H 2S5, Canada
Operation Manual LAM 10M IR 3A LASER AIMING MODULE / TARGET POINTER 105 Sparks Ave., Toronto, ON, M2H 2S5, Canada LASER RADIATION AVOID EXPOSURE TO THE BEAM INFRARED LASER OUTPUT POWER
More informationThis is a new permit condition titled, "2D.1111 Subpart ZZZZ, Part 63 (Existing Non-Emergency nonblack start CI > 500 brake HP)"
This is a new permit condition titled, "2D.1111 Subpart ZZZZ, Part 63 (Existing Non-Emergency nonblack start CI > 500 brake HP)" Note to Permit Writer: This condition is for existing engines (commenced
More informationConsumables and Supplies Required for Repair
Consumables and Supplies Required for Repair Specification When your system needs repair, you want complete coverage, without having to generate a separate purchase order for the consumables or supplies
More informationPULP & PAPER INDUSTRY
Alignment & Positioning in the PULP & PAPER INDUSTRY Fast, easy and accurate alignment of paper machines, pumps, drives, foundations, etc. To compete in today s marketplace, you have to outperform your
More informationSnapshot LX10 USER MANUAL. OK on Dimmer Outdoor OK Sound Activated DMX512 Master/Slave 115V/230V Switch Replaceable Fuse User Serviceable Duty Cycle
LX10 Snapshot OK on Dimmer Outdoor OK Sound Activated DMX512 Master/Slave 115V/230V Switch Replaceable Fuse User Serviceable Duty Cycle USER MANUAL Chauvet, 3000 N 29 th Ct, Hollywood, FL 33020 U.S.A.
More informationCAUTION CONTENTS 5.0 WARRANTY
5.0 WARRANTY ELECTROMATIC Equipment Co., Inc. (ELECTROMATIC) warrants to the original purchaser that this product is of merchantable quality and confirms in kind and quality with the descriptions and specifications
More information! Warning, refer to accompanying documents.
About this Manual To the best of our knowledge and at the time written, the information contained in this document is technically correct and the procedures accurate and adequate to operate this instrument
More informationMorgan Schaffer. Calisto Monitors. Maintenance. Presented by: Eric Thibodeau. Technical Support Specialist
Morgan Schaffer Calisto Monitors Maintenance Presented by: Eric Thibodeau Technical Support Specialist Flushing of the oil lines Purging the oil lines Step 1: All valves are closed Calisto Pump OFF OUTLET
More informationAnalysis of Petroleum Samples Using the Teledyne Leeman Labs
APPLICATION NOTE: TECHNOLOGY: INDUSTRY: AN1802 ICP Petroleum Analysis of Petroleum Samples Using the Teledyne Leeman Labs Prodigy Plus ICP-OES John Condon, Applications Chemist and Bruce MacAllister, Applications
More informationCONTINUOUS IMMERSION AND DUST TEST REPORT
Report No. :HS1404020081A Version:A CONTINUOUS IMMERSION AND DUST TEST REPORT Company Address :Taiway Electronics Components Co. :4F., No. 4, 6, 8, Lane 359, Sec. 2, Zhongshan Rd., Zhonghe Dist., New Taipei
More informationPreventive Maintenance Checklist of Industrial Control and Drive System Equipment
Service Bulletin Preventive Maintenance Checklist of Industrial Control and Drive System Equipment Use the following checklist as a guide in performing preventive maintenance. ATTENTION: Servicing energized
More informationRe: Online Reporting of Backflow Assembly Tests and Updates to Backflow Tester Records
Re: Online Reporting of Backflow Assembly Tests and Updates to Backflow Tester Records Dear Backflow Assembly Tester: Please note that there are significant changes to the City of Mercer Island's Testing
More information1.0 Introduction Mounting Procedure Verifying Calibration Accuracy Specifications Tension Ranges...
5.0 WARRANTY ELECTROMATIC Equipment Co., Inc. (ELECTROMATIC) warrants to the original purchaser that this product is of merchantable quality and confirms in kind and quality with the descriptions and specifications
More informationElectrophoresis File Run Summary
2100 expert DE04103803_2015-04-14_12-07-50.xad Page 1 of 19 C:\..._DE04103803_2015-04-14_12-07-50.xad Electrophoresis File Run Summary Instrument Information: Instrument Name: DE04103803 Firmware: C.01.069
More informationDSL-240 MkIII Single Pass Dust Monitor Measures mg/m 3 using DDP
Ideal for monitoring dust levels in the exhaust gas of industrial combustion or air filtration processes. Innovative Dynamic Detection Principle (DDP) measurement technique Immune to gradual reductions
More informationWhite Paper: Ensure Accuracy and Maximize the Value of Your Cable Testing Equipment with Calibration from Fluke Networks
White Paper: Ensure Accuracy and Maximize the Value of Your Cable Testing Equipment with Calibration from Fluke Networks As a cable installation contractor, you understand the importance of warranties
More informationIMEX 66R ROTATING LASERS
IMEX 66R ROTATING LASERS OPERATION MANUAL OPERATING MANUAL ROTATING LASER IMEX 66R Introduction Congratulations on your purchase of a quality Imex rotating laser. These lasers have been designed for the
More informationUser Manual. Snapshot. Use on Dimmer. Outdoor Use. Sound Activated DMX 512. Master/Slave. 115/230V Power Switch. Replaceable Fuse.
Snapshot Use on Dimmer Outdoor Use Sound Activated DMX 512 Master/Slave 115/230V Power Switch Replaceable Fuse User Serviceable Duty Cycle User Manual 3000 N 29 th Ct, Hollywood, FL 33020 U.S.A. (800)
More informationSnapshot LX5 USER MANUAL. OK on Dimmer Outdoor OK Sound Activated DMX512 Master/Slave 115V/230V Switch Replaceable Fuse User Serviceable Duty Cycle
LX5 Snapshot OK on Dimmer Outdoor OK Sound Activated DMX512 Master/Slave 115V/230V Switch Replaceable Fuse User Serviceable Duty Cycle USER MANUAL Chauvet, 5200 NW 108th Avenue, Sunrise, FL 33351 U.S.A.
More information3 Maintenance. 3 Maintenance. 3.1 Introduction. 3.1 Introduction
3.1 Introduction 3 Maintenance 3.1 Introduction Structure of this chapter This chapter describes all the maintenance activities recommended for the IR 1200. It is based on the maintenance schedule found
More informationHow to fill in a TIR Carnet
How to fill in a TIR Carnet This brochure is made available by the IRU in order to provide guidelines to transport operators, TIR Associations and Customs officers 2 How to fill in a TIR Carnet How to use
More informationSnapshot Ok on Dimmer Outdoor OK Sound Activated DMX512 Master/Slave 115V/230V Switch Replaceable Fuse User Serviceable Duty Cycle
SX Scope Snapshot Ok on Dimmer Outdoor OK Sound Activated DMX512 Master/Slave 115V/230V Switch Replaceable Fuse User Serviceable Duty Cycle USER MANUAL Chauvet, 3000 N 29 th Ct, Hollywood, FL 33020 U.S.A.
More informationSX-16 Nightsun Searchlight Safety and Service Bulletin # SL
SX-16 Nightsun Searchlight Safety and Service Bulletin # SL 0898-02 Issue Date: 09/25/98 Amended Date: 02/09/06 Subject: SX-16 Nightsun Junction Box Wiring Modification Affected Products: All SX-16 Nightsun
More informationAE-770 Electrostatic Field Meter
AE-770 Electrostatic Field Meter Palm-sized electrostatic measuring device ideal for checking ion balance and Sensor for polarity of static charge. Static Super Scale Features: Measuring polarity and intensity
More informationTable of Contents. Timer Identification Timer ID BLU-U Features: 1K 6K BLU-U Features 1K 6K
DUSA Pharmaceuticals, Inc. Table of Contents Go to Chart # Timer Identification Timer ID BLU-U Features: 1K 6K BLU-U Features 1K 6K BLU-U Features: 10K BLU-U Features 10K BLU-U Symptom Fans Running, Timer
More informationArtisan Technology Group is your source for quality new and certified-used/pre-owned equipment
Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment FAST SHIPPING AND DELIVERY TENS OF THOUSANDS OF IN-STOCK ITEMS EQUIPMENT DEMOS HUNDREDS OF MANUFACTURERS SUPPORTED
More informationSTANDARD OPERATING PROCEDURE (SOP)
Page 1 of 23 FORSYTH COUNTY OFFICE OF ENVIRONMENTAL ASSISTANCE AND PROTECTION STANDARD OPERATING PROCEDURE (SOP) Calibrators Signature Page Section 12 Page 2 of 23 By the signatures below, the Forsyth
More informationHelena Snyman - Technical Manager and Sydney Bhengu - Customer Support Technologist Elster Kent Metering (Pty) Ltd. South Africa
Helena Snyman - Technical Manager and Sydney Bhengu - Customer Support Technologist Elster Kent Metering (Pty) Ltd. South Africa Electricity meter installation and installation verification Safety NRS
More informationAn LED collimator consists of a collimating lens and an LED
HIGH POWER LED COLLIMATOR SOURCES An LED collimator consists of a collimating lens and an LED emitter. The LED emitter is placed at the focal plane of the collimating lens. The collimating lens thus images
More informationMagnetic Torpedo Laser Level with Base Model No Instruction Manual
40-6174 4/8/05 5:26 PM Page 1 Magnetic Torpedo Laser Level with Base Model No. 40-6174 Instruction Manual Congratulations on your choice of this Magnetic Torpedo Laser Level. We suggest you read this instruction
More informationSeries S847 Changeover switches featuring wiping, galvanically isolated, double-break contacts and positive opening operation Catalogue D47.
Snap-action switches Changeover switches featuring wiping, galvanically isolated, double-break contacts and positive opening operation Catalogue D47.en Snap-action switches, S847 Series Dual changeover
More informationMonitor Chlorine in Crude at Sub-ppm Levels
better analysis counts Monitor Chlorine in Crude at Sub-ppm Levels Benchtop and Online Analysis Solutions Total Chlorine Analysis in Liquid Hydrocarbons Clora is a compact analyzer to measure total chlorine
More informationFULL AUTOMATIC ASPHALTENE ANALYZER APD-500A USER S GUIDE
FULL AUTOMATIC ASPHALTENE ANALYZER APD-500A USER S GUIDE Please read this manual carefully to ensure correct use of this analyzer before you start to use it. COSMO TRADE & SERVICE CO., LTD. TABLE OF CONTENTS
More informationHYDRAULIC HOSES AND HOSE KITS
HYDRAULIC HOSES AND HOSE KITS Liebherr Hydraulic Hoses and Hose Kits Liebherr-Australia Hose Production has been manufacturing high pressure hose assemblies in the Adelaide Head Office complex since 1999,
More informationWarrant of Fitness Inspection Checksheet
Warrant of Fitness Inspection Checksheet Printed by the Motor Trade Association for members involved in the inspection of motor vehicles for a Warrant of Fitness Suitable for motorcycles and trailers DATE
More informationS826, S926 Series Dual changeover switches with positive opening operation and wiping contacts Catalogue D26.en
2 Snap-action switches S826, S26 Series Dual changeover switches with positive opening operation and wiping contacts Catalogue D26.en 2 Snap-action switches S826 Series Dual changeover switches with positive
More informationInfraPRO Illuminator. 850nm PoE++ High Power LED model. 940nm PoE++ High Power LED model. FEATURES SPECIFICATIONS ACCESSORIES
InfraPRO Illuminator 850nm PoE++ High Power LED model 15-IL38P10 / 15-IL38P30 / 15-IL38P60 / 15-IL38P12 940nm PoE++ High Power LED model 15-IL38P910 / 15-IL38P930 / 15-IL38P960 / 15-IL38P912 FEATURES Autodetect
More informationVIBRATION AND MECHANICAL SHOCK TEST REPORT
Report No. :DH1701190024A-1 Version:A VIBRATION AND MECHANICAL SHOCK TEST REPORT Company :OTO Photonics Inc. Address :9F-5,No. 27, Guanxin Rd., East Dist., Hsinchu City 30072, Taiwan(R.O.C.) Sample Name
More informationMaintenance Adjustments
4 Maintenance and Adjustments Chapter Contents Cleaning the Printer and Paper Handling Accessories..... 158 Cleaning the HP Digital Copier....................... 161 Cleaning ADF and Glass............................
More informationFilterMax F3 and FilterMax F5 Multi-Mode Microplate Reader
FilterMax F3 and FilterMax F5 Multi-Mode Microplate Reader User Guide 5008528 A September 2010 This document is provided to customers who have purchased Molecular Devices, Inc. ( Molecular Devices ) equipment,
More informationEU RO Mutual Recognition Technical Requirements
1. PRODUCT DESCRIPTION... 1 2. DESIGN EVALUATION... 2 3. PRODUCTION REQUIREMENTS... 6 4. MARKING REQUIREMENTS... 6 5. TYPE APPROVAL CERTIFICATE CONTENT... 7 6. APPROVAL DATE AND REVISION NUMBER... 7 7.
More informationLS-1 Series Tungsten Halogen Light Sources
LS-1 Series Tungsten Halogen Light Sources The LS-1 Series of tungsten halogen light sources are versatile, white-light lamps optimized for use in the VIS- Shortwave NIR range (360-2000 nm). These light
More informationReliability & Qualification Report for MEMS Based Variable Optical Attenuators
Reliability & Qualification Report for MEMS Based Variable Optical Attenuators P/N: MMVOA-1-1550-S-9/125-XX-0.25-1 Page 1 of 21 Revision History Rev Approval Date ECM/ECR Description of Change 00 10/18/2008
More informationHolographic plane gratings
Scientific diffraction gratings / Custom gratings Holographic plane gratings Commitment to quality HORIBA Jobin Yvon offers complete customer service, including expert technical advice for optimizing system
More informationOutside of the Walls. Protocol & SOP Changes & Audit Checklists. Outside The Walls - Section 5 1
Outside of the Walls Protocol & SOP Changes & Audit Checklists Outside The Walls - Section 5 1 The Protocol It tells what is going to be done it DRIVES the study It should be signed & dated by the Study
More informationSelf-Leveling Cross Line Laser
Self-Leveling Cross Line Laser Self-Levelling Thank you for purchase of our laser tool. Please read these instructions thoroughly before operations. Contents * Safety * Applications * Product Overview
More informationUltrasonic heat meter
Ultrasonic heat meter Installation and user guide Flow sensor-vmc-p Energy integrator Compact heat meter Temperature senor CONTENTS Kapitel Please read this manual prior to installation Page 3 2! T! Important
More information