1 Product Data Sheet PS March 2002 Micro Motion ELITE Mass Flow and Density Meters Now available with MVD Technology!

2 Micro Motion ELITE Meters Experience the most accurate Coriolis meters available today. Micro Motion ELITE meters are the leading flowmeter for precision flow measurement. And with good reason. ELITE meters offer the most accurate measurement available for virtually any process fluid, while exhibiting exceptionally low pressure drop. Seven sizes of ELITE meters offer direct mass flow, volume flow, density, and temperature measurement of liquids, gases, and slurries without the need for additional equipment, manual calculations or estimations. Micro Motion ELITE meters are designed for unsurpassed performance in even the most harsh operating environments. They have no moving parts, and no special mounting or flow conditioning requirements. Every ELITE meter features standard secondary containment, and is available with stainless steel or nickel alloy wetted parts and a wide variety of process connections to meet your every need. And, they require no maintenance saving you money over the course of their lifetime by helping you make the most of your time, people, and materials. ELITE meters carry hazardous area approvals for the U.S.A. and Canada, Europe, Australia, Japan and other areas in the Asia-Pacific region. 2 Micro Motion ELITE Meters

3 Special Applications Several ELITE meters have been designed for special applications. The CMF010 is our smallest meter, designed for remarkably high performance in low-flow applications. It features a single, continuous flow tube, and is also available in a high-pressure model, for applications up to 413 bar. Much larger in size is our CMF400. This 100 mm offers the most accurate measurement available in a high-capacity meter. The 80 mm CMF300A is a high-temperature meter. It provides the same accuracy and measurement capabilities as our other ELITE meters, at temperatures up to 343 C. Our CMF025, CMF050, and CMF100 meters are available with optional, flangeless, waferstyle process connections. MVD Technology Micro Motion ELITE meters are now available with MVD Technology a new, innovative, multivariable, digital signal processing capability. A core processor, integrally mounted on the ELITE sensor, works with our new Series 1000 and 2000 transmitters to improve ease of use, reduce downtime, and lower your flow metering costs. Meters with MVD Technology provide cleaner, noise-free digital signals. Digital signal processing means faster response times, enhanced diagnostic capabilities, higher accuracy, and better repeatability. System Integration Micro Motion offers a variety of microprocessorbased transmitters, any of which can be connected to our ELITE sensors. Depending on the transmitter model, features include milliampere and frequency/pulse outputs, transmitter and process control functions in a single device, and API outputs for crude oil and other hydrocarbon fluids. Other options include transmitters that can be installed in instrument racks or panels, or housed in NEMA or explosion-proof enclosures. Sensors and explosion-proof transmitters can be installed in the same hazardous area. Micro Motion ELITE meters can be installed as part of a Bell 202 multidrop network, an RS-485 digital communications network, or a Profibus- PA or FOUNDATION fieldbus system. Transmitters feature HART, Modbus, FOUNDATION fieldbus, Profibus-PA, and other communication protocols. And all our transmitters support Fisher-Rosemount PlantWeb field-based architecture, which uses the power of intelligent, interoperable field devices to improve plant performance. Micro Motion is known worldwide for increasing plant efficiency, production, and profitability. More than Micro Motion meters are installed and working in processes just like yours. Contact us, and discover the best precision flowmeters available today Micro Motion ELITE meters. Micro Motion Elite flowmeters support PlantWeb field-based architecture, a scalable way to use open and interoperable devices and systems to build process solutions of the future. Micro Motion ELITE Meters 3

4 Performance specifications Liquid Flow Mass Volume Nominal flow range kg/h l/h CMF010 0 to 82 0 to 82 CMF025 0 to to 1090 CMF050 0 to to 3400 CMF100 0 to to CMF200 0 to to CMF300 0 to to CMF400 0 to to Maximum flow rate kg/h l/h CMF CMF CMF CMF CMF CMF CMF Mass Flow Accuracy (2) 1700/2700 transmitter or direct host ±0,10% of rate (3) All other transmitters ±0,10% ± [(zero stability/ flow rate) x 100]% of rate Mass Flow 1700/2700 transmitter or direct host ±0,05% of rate (3) Repeatability (2) All other transmitters ±0,05% ± [½(zero stability/ flow rate) x 100]% of rate Zero stability kg/h CMF010 0,002 High-pressure CMF010P 0,004 CMF025 0,027 CMF050 0,163 CMF100 0,680 CMF200 2,18 CMF300 6,80 CMF400 40,91 Micro Motion has adopted the terminology "nominal flow range." The upper limit of this range is the flow rate at which water at reference conditions causes approximately 1 bar of pressure drop for ELITE sensors. (2) Accuracy includes the combined effects of repeatability, linearity, and hysteresis. All specifications for liquids are based on reference conditions of water at 20 to 25 C and 1 to 2 bar, unless otherwise noted. (3) When flow rate is less than (zero stability/0,001), accuracy equals ±[(zero stability/flow rate) x 100]% of rate and repeatability equals ±[½(zero stability/flow rate) x 100]% of rate. 4 Micro Motion ELITE Meters

5 Performance specifications continued Gas Flow (Air at 15 o C and 70 bar) Mass Standard Volume Nominal flow range (2) kg/h Nm 3 /h CMF010 0 to 32 0 to 25 CMF025 0 to to 355 CMF050 0 to to 870 CMF100 0 to to 3865 CMF200 0 to to CMF300 0 to to CMF400 0 to to Maximum flow rate kg/h Nm 3 /h CMF CMF CMF CMF CMF CMF CMF Mass Flow Accuracy (3) 1700/2700 transmitter or direct host ±0,35% of rate (4) All other transmitters ±0,50% ± [(zero stability/ flow rate) x 100]% of rate Mass Flow 1700/2700 transmitter or direct host ±0,20% of rate (4) Repeatability (3) All other transmitters ±0,25% ± [(zero stability/ flow rate) x 100]% of rate Zero stability kg/h Nm 3 /h CMF010 0,002 0,0015 High-pressure CMF010P 0,004 0,003 CMF025 0,027 0,021 CMF050 0,163 0,126 CMF100 0,680 0,526 CMF200 2,18 1,69 CMF300 6,80 5,26 CMF ,64 Air flows in Nm 3 /h (converted to normal conditions: 0 o C and 1,013 bara) when the meter is operated at 15 o C and 70 bara. (2) Micro Motion has adopted the terminology "nominal flow range." The upper limit of this range is the flow rate at which air at 15 C and 70 bar causes approximately 1 bar of pressure drop for ELITE sensors. (3) Accuracy includes the combined effects of repeatability, linearity, and hysteresis. All specifications for gas are based on reference conditions of air at 15 C and 70 bar, unless otherwise noted. (4) When flow rate is less than (zero stability/0,0035), accuracy equals ±[(zero stability/flow rate) x 100]% of rate, and repeatability is equal to ±[½(zero stability/flow rate) x 100]% of rate. Micro Motion ELITE Meters 5

6 Performance specifications continued Density with Model 2700, 3500, 3700, RFT9739, 5300, or RFT9709 transmitter with IFT9701 transmitter Accuracy kg/m 3 kg/m 3 High-pressure CMF010P liquid ±2,0 ±8,0 All other models liquid ±0,5 ±2,0 Repeatability High-pressure CMF010P liquid ±1,0 ±4,0 All other models liquid ±0,2 ±1,0 Range All models 0 to to 5000 Temperature Accuracy All models ±1 C ± 0,5% of reading in C Repeatability All models ±0,2 C Range (2) CMF010, CMF025, CMF050, CMF100, CMF200, CMF300M, CMF300H (3)(4) with core processor C 50 to 125 process fluid with junction box 240 to 204 process fluid High-temperature CMF300A (3) sensor 0 to 343 process fluid junction box 40 to 120 ambient CMF400 (5) Remote booster amplifier for CMF400 with integral booster amplifier with remote booster amplifier 40 to 60 process fluid 240 to 204 process fluid 40 to 60 ambient Accuracy includes the combined effects of repeatability, linearity, and hysteresis. All specifications for liquids are based on reference conditions of water at 20 to 25 C and 1 to 2 bar, unless otherwise noted. (2) For CENELEC/ATEX-compliant sensors, the CENELEC "T" rating and hazardous area classification depend on the maximum process fluid and ambient temperature. See page 11. (3) For CENELEC/ATEX-compliant sensors, the maximum ambient temperature is 55 C. Use of sensors with a junction box above 55 C is acceptable, provided the ambient temperature does not exceed the CENELEC "T" rating listed on page 11 for temperature of the fluid. (4) For CENELEC/ATEX-compliant sensors, the minimum process fluid and ambient temperature is 20 C. If the process fluid remains at or above 0 C, ambient temperature below 20 C is acceptable for sensors with a junction box, and ambient temperature of -40 C is allowed for sensors with a core processor. (5) For CENELEC/ATEX-compliant sensors, the maximum ambient temperature is 60 C and the minimum ambient temperature is -40 C. 6 Micro Motion ELITE Meters

7 Performance specifications continued Pressure ratings Flow tube rating PED approved Housing rating Short-term housing rating (3) bar Stainless steel sensors 100 Nickel alloy sensors 148 High-pressure CMF010P 413 Sensors comply to council directive 97/23/EC of 29 May 1997 on Pressure Equipment. bar CMF010 (2) 29 CMF CMF CMF CMF CMF CMF bar CMF010 (2) 38 CMF CMF CMF CMF CMF CMF Pressure ratings at 25 C, according to ASME B31.3. For higher operating temperatures, pressure needs to be derated as follows: Flow tubes Flow tubes Housing 316L sensors nickel alloy sensors all sensors All sensors 94 to 148 C none 2% derating none 149 to 204 C 7,2% derating 9,2% derating 7,2% derating High-temperature CMF300A 205 to 260 C 13,8% derating not applicable 11,4% derating 261 to 316 C 19,2% derating not applicable 16,2% derating 317 to 343 C 20,1% derating not applicable 18,0% derating (2) Optional rupture disks for high-pressure CMF010 will burst if pressure inside sensor housing reaches 27 bar. (3) Housing pressure rating for 10 hours or less, according to ASME B31.3. Micro Motion ELITE Meters 7

8 Performance specifications continued Typical accuracy, turndown, and pressure drop in mass with 1700/2700 transmitter or direct host CMF010N CMF010M CMF025 2,5 2,0 1,5 100:1 20:1 10:1 1:1 Turndown 500:1 100:1 20:1 10:1 1:1 Accuracy (±%) liquid 1,25 0,25 0,10 0,10 0,10 gas 3,35 0,65 0,35 0,35 0,35 Pressure drop liquid (bar) ~0 ~0 0,01 0,01 1,0 Accuracy, % 1,0 0,5 0 0,5 1,0 gas (bar) ,01 0,88 1,5 liquid 2,0 gas 2, Nominal flow rate, % CMF010P CMF050 CMF100 CMF200 CMF300 Turndown 500:1 100:1 20:1 10:1 1:1 Accuracy (±%) liquid 2,40 0,50 0,10 0,10 0,10 gas 7,35 1,50 0,35 0,35 0,35 Pressure drop liquid (bar) ~0 ~0 0,01 0,01 0,93 gas (bar) 0 0 0,01 0,02 0,97 Accuracy, % 2,5 2,0 1,5 1,0 0,5 0 0,5 1,0 1,5 2,0 2,5 100:1 20:1 10:1 liquid gas 1: Nominal flow rate, % To determine accuracy, turndown, and pressure drop using your process variables, contact your local Micro Motion representative. 8 Micro Motion ELITE Meters

9 Performance specifications continued Typical accuracy, turndown, and pressure drop in mass with 1700/2700 transmitter or direct host CMF400 Turndown 500:1 100:1 20:1 10:1 1:1 Accuracy (±%) liquid 5 1,0 0,20 0,10 0,10 gas 16,25 3,25 0,65 0,35 0,35 Pressure drop liquid (bar) ~0 ~0 <0,01 0,01 1,0 gas (bar) <0,01 1,0 Accuracy, % 2,5 2,0 1,5 1,0 0,5 0 0,5 100:1 20:1 10:1 1:1 1,0 1,5 liquid 2,0 gas 2, Nominal flow rate, % To determine accuracy, turndown, and pressure drop using your process variables, contact your local Micro Motion representative. Micro Motion ELITE Meters 9

10 Functional specifications Environmental effects Process temperature effect Process temperature effect is defined as the worst-case zero offset due to process fluid temperature change away from the zeroing temperature for flow rate and calibration temperature for density. Process temperature effect % of nominal flow rate per C density accuracy per C CMF010 ±0,00025 ±0,015 kg/m 3 CMF025 ±0,00025 ±0,015 kg/m 3 CMF050 ±0,00025 ±0,015 kg/m 3 CMF100 ±0,00025 ±0,015 kg/m 3 CMF200 ±0,001 ±0,015 kg/m 3 CMF300 ±0,001 ±0,015 kg/m 3 CMF400 ±0,001 ±0,015 kg/m 3 Pressure effect Pressure effect is defined as the change in sensor flow due to process pressure change away from the calibration pressure. Pressure effect can be corrected. Pressure effect on flow accuracy % of rate per bar CMF010 none CMF025 none CMF050 none CMF100 0,003 CMF200 0,012 CMF300 0,009 CMF400 0,15 Power supply for CMF400 booster amplifier Input voltage Maximum current 85 to 265 VAC, 50/60 Hz 0,5 A Maximum power 50 watts Nominal flow rate is the upper limit of the nominal flow range. 10 Micro Motion ELITE Meters

11 Functional specifications continued Hazardous area classifications UL is a U.S.A. approvals agency, CSA is a Canadian approvals agency that provides approvals accepted both in the U.S.A. and in Canada. CENELEC and ATEX are European standards organizations, and SAA is an Australian approvals agency. UL and CSA Class I, Div. 1, Groups C and D Class I, Div. 2, Groups A, B, C, and D Class II, Div.1, Groups E, F, and G SAA CMF025 Ex ib IIC T5 CMF050 Ex ib IIC T5 CMF100 Ex ib IIC T6 CMF200 Ex ib IIB T6 CMF300 Ex ib IIB T6 ATEX (2) sensors with core processor except CMF400 CMF010, CMF025, CMF050, CMF100 CE 0575 II 2G EEx ib IIC T1...T5 CMF200, CMF300 CE 0575 II 2G EEx ib IIB T1...T5 sensors with junction box, except CMF300A and CMF400 CMF010, CMF025, CMF050, CMF100 CE 0575 II 2G EEx ib IIC T1...T6 CMF200, CMF300 CE 0575 II 2G EEx ib IIB T1...T6 max. ambient temp. ( C) max. ambient temp. ( C) de-rate at slope = -,093 C ambient per C fluid T5 T4 T3 T1-T2 sensor fluid temp. ( C) T6 T5 T4 T3 T1-T sensor fluid temp. ( C) CMF300A CE 0575 II 2G EEx ib IIB T1...T6 max. ambient temp. ( C) 55 T6 T5 T4 T3 T2 T sensor fluid temp. ( C) Standard-temperature model only. CMF300A high-temperature model is not SAA approved. (2) ATEX "T" rating depends on the maximum temperature shown in the graphs above. Ambient temperature limits are 20 C and 55 C for all sensors except CMF400. CMF400 ambient temperature limits are -40 C to 60 C. Micro Motion ELITE Meters 11

12 Functional specifications continued Hazardous area classifications continued CENELEC CMF400 with integral booster amplifier and integral core processor EEx de[ib] ibiib T1...T5 max. ambient temp. ( C) T1-T5 sensor fluid temp. ( C) CMF400 with integral booster amplifier and junction box EEx de[ib] ibiib T1...T6 max. ambient temp. ( C) T1-T6 sensor fluid temp. ( C) CMF400 with remote booster amplifier EEx de[ib] ibiib T1...T6 max. ambient temp. ( C) T6 T5 T4 T3 T1- T sensor fluid temp. ( C) CENELEC "T" rating depends on the maximum temperature shown in the graphs above. Ambient temperature limits are 20 C and 55 C for all sensors except CMF400. CMF400 ambient temperature limits are -40 C to 60 C. 12 Micro Motion ELITE Meters

13 Physical specifications Materials of construction Wetted parts Housing Junction box Core processor not available with CMF300A Booster amplifier housing (for CMF400 only) Stainless steel Nickel alloy CMF L or 304L Inconel alloy 686 CMF L or 304L Hastelloy C-22 CMF L or 304L Hastelloy C-22 CMF L or 304L Hastelloy C-22 CMF L or 304L Hastelloy C-22 CMF L or 304L Hastelloy C-22 High-temperature CMF300A 316L not available CMF L not available 304L stainless steel Epoxy-coated aluminum 316L stainless steel Epoxy-coated aluminum Weight (2) With junction box With core processor kg kg CMF CMF CMF CMF CMF CMF CMF400 with integral booster amplifier CMF400 with remote booster amplifier (sensor only) Remote booster amplifier 6 7 General corrosion guides do not account for cyclical stress, and therefore should not be relied upon when choosing a wetted material for your Micro Motion sensor. Please refer to Micro Motion s corrosion guide for proper material compatibility information. (2) Weight of sensor with ANSI 150 lb weld neck raised face flanges. Micro Motion ELITE Meters 13

14 CMF010 dimensions Dimensions in mm CMF010 with core processor FLOW 84 2x 40 temperature extender option 205 2x 1/2" 14 NPT female purge fitting (optional) Ø /2" 14 NPT female or M20 x 1,5 female CMF010 with junction box FLOW 2x 40 temperature extender option x 1/2" 14 NPT female purge fitting (optional) Ø /4" 14 NPT female CMF010 with core processor or junction box Back view with union fittings Side view with rupture disk Flange dimensions 2x 35 Dim. A face to face ± Dim. A face to face ± x 8 thru 180 Dim. Ø B 2x rupture disks (optional) Micro Motion ELITE Meters

15 CMF010 process fittings Fittings for 316L stainless steel sensors Fitting code Dim. A (mm) Face-to-face Dim. B (mm) Outside diam. 1/2-inch 150 lb ANSI weld neck raised face flange /2-inch 300 lb ANSI weld neck raised face flange /2-inch 600 lb ANSI weld neck raised face flange /2-inch sanitary fitting mm DIN PN40 weld neck, DIN 2635, type C face mm DIN PN100 weld neck, DIN 2637, type E face /4-inch NPT female union fitting /4-inch tube compression fitting mm tube compression fitting Fittings for 304L stainless steel sensors 1/2-inch 150 lb ANSI weld neck raised face flange /2-inch 300 lb ANSI weld neck raised face flange mm DIN PN40 weld neck, DIN 2635 type C face Fittings for nickel alloy sensors 1/2-inch 150 lb ANSI lap joint flange /2-inch 300 lb ANSI lap joint flange mm DIN PN40 lap joint, DIN /4-inch NPT female union fitting Fittings for high-pressure sensors 1/4-inch NPT female union fitting /4-inch tube compression fitting mm tube compression fitting Fittings listed here are standard options. Other types of fittings are available. Contact your local Micro Motion representative. Micro Motion ELITE Meters 15

16 CMF025 dimensions Dimensions in mm CMF025 with core processor FLOW 84 2X 1/2" 14 NPT purge fitting (optional) 2X 58 temperature extender option 246 Ø /2" 14 NPT female or M20 X 1,5 female CMF025 with junction box FLOW 2x X 1/2" 14 NPT purge fitting (optional) temperature extender option 246 Ø /4" 14 NPT female CMF025 with core processor or junction box Back view with wafer style fittings Back view with union fittings Flange dimensions Dim A ±3 Dim. Ø B 24 2X 46 Dim. A face to face ±3 Dim. A face to face ±3 1/2" NPT female Dim. B Micro Motion ELITE Meters

17 CMF025 process fittings Fittings for 316L stainless steel sensors Fitting code Dim. A (mm) Face-to-face Dim. B (mm) Outside diam. Wafer style, 1/2" ANSI (150 lb, 300 lb, 600 lb bolt kit) Wafer style, 15 mm DIN 2635, type C facing (PN40 bolt kit) Wafer style, 15 mm DIN 2635, type N grooved facing (PN40 bolt kit) Wafer style, 15 mm DIN 2637, type E facing (PN100 bolt kit) Wafer style, 15 mm DIN 2637, type N grooved facing (PN100 bolt kit) /2" ANSI 150 lb weld neck raised face flange /2" ANSI 300 lb weld neck raised face flange /2" ANSI 600 lb weld neck raised face flange /2" NPT female union fitting /2" sanitary fitting mm DIN PN40 weld neck, DIN 2635, type C facing mm DIN PN40 weld neck, DIN 2635, type N grooved facing mm DIN PN100 weld neck, DIN 2637, type E facing mm DIN PN100 weld neck, DIN 2637, type N grooved facing Fittings for 304L stainless steel sensors 1/2" ANSI 150 lb weld neck raised face flange /2" ANSI 300 lb weld neck raised face flange mm DIN PN40 weld neck, DIN 2635 type C face Fittings for nickel alloy sensors 1/2" ANSI 150 lb lap joint flange /2" ANSI 300 lb lap joint flange mm DIN PN40 lap joint flange, DIN Fittings listed here are standard options. Other types of fittings are available. Contact your local Micro Motion representative. Micro Motion ELITE Meters 17

18 CMF050 dimensions Dimensions in mm CMF050 with core processor FLOW 111 2X 63 temperature extender option 313 2X 1/2" 14 NPT purge fitting (optional) Ø /2" 14 NPT female or M20 X 1,5 female CMF050 with junction box FLOW 111 2X 63 temperature extender option 305 2X 1/2" 14 NPT purge fitting (optional) Ø /4" 14 NPT female Back view with wafer style fittings CMF050 with core processor or junction box 55 Back view with union fittings 225 Flange dimensions Dim. A ±3 24 2X 49 Dim. A face to face ±3 Dim. A face to face ±3 Dim. Ø B Dim. B 3/4" NPT female Micro Motion ELITE Meters

19 CMF050 process fittings Fittings for 316L stainless steel sensors Fitting code Dim. A (mm) Face-to-face Dim. B (mm) Outside diam. Wafer style, 1/2" ANSI (150 lb, 300 lb, 600 lb bolt kit) Wafer style, 15 mm DIN 2635, type C facing (PN40 bolt kit) Wafer style, 15 mm DIN 2635, type N grooved facing (PN40 bolt kit) Wafer style, 15 mm DIN 2637, type E facing (PN100 bolt kit) Wafer style, 15 mm DIN 2637, type N grooved facing (PN100 bolt kit) /2" ANSI 150 lb weld neck raised face flange /2" ANSI 300 lb weld neck raised face flange /2" ANSI 600 lb weld neck raised face flange /4" NPT female union fitting /4" sanitary fitting mm DIN PN40 weld neck, DIN 2635, type C facing mm DIN PN40 weld neck, DIN 2635, type N grooved facing mm DIN PN100 weld neck, DIN 2637, type E facing mm DIN PN100 weld neck, DIN 2637, type N grooved facing Fittings for 304L stainless steel sensors 1/2" ANSI 150 lb weld neck raised face flange /2" ANSI 300 lb weld neck raised face flange mm DIN PN40 weld neck, DIN 2635 type C face Fittings for nickel alloy sensors 1/2" ANSI 150 lb lap joint flange /2" ANSI 300 lb lap joint flange mm DIN PN40 lap joint flange, DIN Fittings listed here are standard options. Other types of fittings are available. Contact your local Micro Motion representative. Micro Motion ELITE Meters 19

20 CMF100 dimensions Dimensions in mm CMF100 with core processor FLOW 2X X 1/2" 14 NPT purge fitting (optional) temperature extender option Ø /2" 14 NPT female or M20 X 1,5 female CMF100 with junction box FLOW 2X X 1/2" 14 NPT female purge fitting (optional) temperature extender option 409 Ø 32 3/4" 14 NPT female Back view with wafer style fittings Dim. A ±3 Dim. Ø B 33 Flange dimensions Dim. A face to face ±3 Dim. B Micro Motion ELITE Meters

21 CMF100 process fittings Fittings for 316L stainless steel sensors Fitting code Dim. A (mm) Face-to-face Dim. B (mm) Outside diam. Wafer style, 1" ANSI (150 lb bolt kit) Wafer style, 1" ANSI (300 lb, 600 lb bolt kit) Wafer style, 25 mm DIN 2635, type C facing (PN40 bolt kit) Wafer style, 25 mm DIN 2635, type N grooved facing (PN40 bolt kit) Wafer style, 25 mm DIN 2637, type E facing (PN100 bolt kit) Wafer style, 25 mm DIN 2637, type N grooved facing (PN100 bolt kit) " ANSI 150 lb weld neck raised face flange " ANSI 300 lb weld neck raised face flange " ANSI 600 lb weld neck raised face flange /2" ANSI 600 lb weld neck raised face flange " sanitary fitting mm DIN PN40 weld neck, DIN 2635, type C facing mm DIN PN40 weld neck, DIN 2635, type N grooved facing mm DIN PN100 weld neck, DIN 2637, type E facing mm DIN PN100 weld neck, DIN 2637, type N grooved facing Fittings for 304L stainless steel sensors 1" ANSI 150 lb weld neck raised face flange " ANSI 300 lb weld neck raised face flange mm DIN PN40 weld neck, DIN 2635 type C face Fittings for nickel alloy sensors 1" ANSI 150 lb lap joint flange " ANSI 300 lb lap joint flange mm DIN PN40 lap joint flange, DIN mm DIN PN40 lap joint flange, DIN Fittings listed here are standard options. Other types of fittings are available. Contact your local Micro Motion representative. Micro Motion ELITE Meters 21

22 CMF200 dimensions Dimensions in mm CMF200 with core processor FLOW Ø 32 1/2" 14 NPT female or M20 X 1,5 female 2X 1/2" 14 NPT purge fitting (optional) temperature extender option 178 2X CMF200 with junction box FLOW Ø 32 3/4" 14 NPT female 2X 1/2" 14 NPT female purge fitting (optional) temperature extender option 178 2X Back view and flange dimensions Dim. A ±3 38 Dim. Ø B Micro Motion ELITE Meters

23 CMF200 process fittings Fittings for 316L stainless steel sensors Fitting code Dim. A (mm) Face-to-face Dim. B (mm) Outside diam. 1 1/2" ANSI 150 lb weld neck raised face flange /2" ANSI 300 lb weld neck raised face flange /2" ANSI 600 lb weld neck raised face flange " ANSI 150 lb weld neck raised face flange " ANSI 300 lb weld neck raised face flange " ANSI 600 lb weld neck raised face flange /2" sanitary fitting " sanitary fitting mm DIN PN40 weld neck, DIN 2635, type C facing mm DIN PN40 weld neck, DIN 2635, type N grooved facing mm DIN PN100 weld neck, DIN 2637, type E facing mm DIN PN100 weld neck, DIN 2637, type N grooved facing mm DIN PN40 weld neck, DIN 2635, type C facing mm DIN PN40 weld neck, DIN 2635, type N grooved facing mm DIN PN100 weld neck, DIN 2637, type E facing mm DIN PN100 weld neck, DIN 2637, type N grooved facing Fittings for 304L stainless steel sensors 1 1/2" ANSI 150 lb weld neck raised face flange /2" ANSI 300 lb weld neck raised face flange " ANSI 150 lb weld neck raised face flange " ANSI 300 lb weld neck raised face flange mm DIN PN40 weld neck, DIN 2635, type C face mm DIN PN40 weld neck, DIN 2635, type C face Fittings for nickel alloy sensors 1 1/2" ANSI 150 lb lap joint flange /2" ANSI 300 lb lap joint flange " ANSI 150 lb lap joint flange " ANSI 300 lb lap joint flange mm DIN PN40 lap joint flange, DIN mm DIN PN40 lap joint flange, DIN Fittings listed here are standard options. Other types of fittings are available. Contact your local Micro Motion representative. Micro Motion ELITE Meters 23

24 CMF300 dimensions Dimensions in mm CMF300 with core processor FLOW Ø 32 1/2" 14 NPT female or M20 X 1.5 female 2X 1/2" 14 NPT purge fitting (optional) temperature extender option X FLOW CMF300 with junction box Ø 32 3/4" 14 NPT female 2X 1/2" 14 NPT female purge fitting (optional) temperature extender option X Back view and flange dimensions Dim. A ±3 84 Dim. Ø B Micro Motion ELITE Meters

25 CMF300 process fittings Fittings for 316L stainless steel sensors Fitting code Dim. A (mm) Face-to-face Dim. B (mm) Outside diam. 3" ANSI 150 lb weld neck raised face flange " ANSI 300 lb weld neck raised face flange " ANSI 600 lb weld neck raised face flange " ANSI 150 lb weld neck raised face flange " ANSI 300 lb weld neck raised face flange " ANSI 600 lb weld neck raised face flange " sanitary fitting mm DIN PN40 weld neck, DIN 2635, type C facing mm DIN PN40 weld neck, DIN 2635, type N grooved facing mm DIN PN100 weld neck, DIN 2637, type E facing mm DIN PN100 weld neck, DIN 2637, type N grooved facing mm DIN PN40 weld neck, DIN 2635, type C facing mm DIN PN40 weld neck, DIN 2635, type N grooved facing mm DIN PN100 weld neck, DIN 2637, type E facing mm DIN PN100 weld neck, DIN 2637, type N grooved facing Fittings for 304L stainless steel sensors 3" ANSI 150 lb weld neck raised face flange " ANSI 300 lb weld neck raised face flange mm DIN PN40 weld neck, DIN 2635, type C face Fittings for nickel alloy sensors 3" ANSI 150 lb lap joint flange " ANSI 300 lb lap joint flange mm DIN PN40 lap joint flange, DIN Fittings listed here are standard options. Other types of fittings are available. Contact your local Micro Motion representative. Micro Motion ELITE Meters 25

26 High-temperature CMF300A dimensions and process fittings Dimensions in mm FLOW 1163 Dim. A face to face ±3 84 Junction box Dim. B 976 2X 1/2" NPT female purge fitting (optional) 2X 279 Remote-mount junction box 2X flexible conduit Minimum bend radius: X Ø 6 thru for mounting 1/2" NPT female Fittings Fitting code Dim. A (mm) Face-to-face Dim. B (mm) Outside diam. 3" ANSI 150 lb weld neck raised face flange " ANSI 300 lb weld neck raised face flange " ANSI 600 lb weld neck raised face flange " ANSI 900 lb weld neck raised face flange " ANSI 150 lb weld neck raised face flange " ANSI 300 lb weld neck raised face flange " ANSI 600 lb weld neck raised face flange " ANSI 900 lb weld neck raised face flange mm DIN PN40 weld neck, DIN 2635, type C facing mm DIN PN40 weld neck, DIN 2635, type N grooved facing mm DIN PN100 weld neck, DIN 2637, type E facing mm DIN PN100 weld neck, DIN 2637, type N grooved facing mm DIN PN40 weld neck, DIN 2635, type C facing mm DIN PN40 weld neck, DIN 2635, type N grooved facing mm DIN PN100 weld neck, DIN 2637, type E facing mm DIN PN100 weld neck, DIN 2637, type N grooved facing Fittings listed here are standard options. Other types of fittings are available. Contact your local Micro Motion representative. 26 Micro Motion ELITE Meters

27 CMF400 dimensions Dimensions in mm CMF400 with integrally mounted booster amplifier and core processor 183 normal flow direction Dim. A ±5 137 Dim. øb 250 includes 33 mm clearance for cover removal 165 3/4"-14 NPT female* 971 1/2"-14 NPT female or M20 x 1, *Note: Adapters available in 1/2"-14 NPT female or M20 x 1,5 for CENELEC-approved models CMF400 with integrally mounted booster amplifier and junction box normal flow direction Dim. A ± Dim. øb 250 includes 33 mm clearance for cover removal /4"-14 NPT female* 3/4"-14 NPT female *Note: Adapters available in 1/2"-14 NPT female or M20 x 1,5 for CENELEC-approved models Micro Motion ELITE Meters 27

28 CMF400 dimensions and process fittings Dimensions in mm CMF400 for remotely mounted booster amplifier normal flow direction 2x 350 Dim. A ± includes 33 mm clearance for cover removal Dim. øb 971 3/4"-14 NPT female* *Note: Adapters available in 1/2"-14 NPT female or M20 x 1,5 for CENELEC-approved models Fittings Fitting code Dim. A (mm) Face-to-face Dim. B (mm) Outside diam. 4" ANSI 150 lb weld neck raised face flange " ANSI 300 lb weld neck raised face flange " ANSI 600 lb weld neck raised face flange " ANSI 150 lb weld neck raised face flange " ANSI 300 lb weld neck raised face flange " ANSI 600 lb weld neck raised face flange mm DIN PN40 weld neck, DIN 2635, type C facing mm DIN PN40 weld neck, DIN 2635, type N facing mm DIN PN100 weld neck, DIN 2637, type E facing mm DIN PN100 weld neck, DIN 2637, type N facing mm DIN PN40 weld neck, DIN 2635, type C facing mm DIN PN40 weld neck, DIN 2635, type N facing mm DIN PN100 weld neck, DIN 2637, type E facing mm DIN PN100 weld neck, DIN 2637, type N facing Fittings listed here are standard options. Other types of fittings are available. Contact your local Micro Motion representative. 28 Micro Motion ELITE Meters

29 CMF400 remote booster amplifier Dimensions in mm Remote mount booster amplifier with core processor 3 meter of cable supplied by Micro Motion 2 1/2"-14 NPT female or M20 x 1, /4"-14 NPT female* 210 2x x 5 80 includes 33 mm clearance for cover removal 2x wall mount 2x 36 2x x ø9 thru ø50 pole mount *Note: Adapters available in 1/2"-14 NPT female or M20 x 1,5 for CENELEC-approved models Remote mount booster amplifier with junction box meter of cable supplied by Micro Motion 3/4"-14 NPT female or M20 x 1,5 76 3/4"-14 NPT female* x 124 2x 5 80 includes 33 mm clearance for cover removal 2x wall mount 38 ø50 pole mount 2x 36 2x x ø9 thru *Note: Adapters available in 1/2"-14 NPT female or M20 x 1,5 for CENELEC-approved models Micro Motion ELITE Meters 29

30 Ordering information ELITE sensor model number matrix Model CMF010 CMF025 CMF050 CMF100 CMF200 CMF300 CMF400 Code M L H N A P Sensor Model ELITE 3 mm sensor ELITE 6 mm sensor ELITE 15 mm sensor ELITE 25 mm sensor ELITE 50 mm sensor ELITE 80 mm sensor ELITE 100 mm sensor Pressure, temperature and wetted material Standard pressure, standard temperature, 316L stainless steel Standard pressure, standard temperature, 304L stainless steel not available with CMF400 Standard pressure, standard temperature, Nickel Alloy Not available with CMF010 or CMF400 Standard pressure, standard temperature, Inconel 686 CMF010 only Standard pressure, high temperature, 316L stainless steel CMF300A High pressure, standard temperature, Inconel 686 nickel alloy and 316L stainless steel CMF010 only Code Process Connections ### See fittings tables on pages Code Case Options N Standard pressure containment P Purge fittings (two 1/2-inch NPT female, not available with CMF400) D Rupture disks (two 28 bar disks) CMF010 with pressure/material code P only Code Electronics Interface A Local core processor for remotely mounted Series 1000/2000 not available with CMF300A or CMF400 B Local core processor with extended mount for remotely mounted Series 1000/2000 transmitter not available with CMF300A or CMF400 D MVDSolo; local core processor for direct host connections (for OEM s) E MVDSolo; local core processor with extended mount for direct host connection (for OEM s) not available with CMF300A or CMF400 R 9 wire junction box not available with CMF400 H 9-wire junction box with extended mount not available with CMF300A or CMF400 For CMF400 only K Integral Booster Amp with local core processor to remote mount Series 1000/2000 transmitters L Integral Booster Amp with MVDSolo, local core processor for direct host connection (for OEM's) M Integral Booster Amp with 9 wire junction box N Remote Booster Amp with local core processor to remote mount Series 1000/2000 transmitters O Remote Booster Amp with MVDSolo, local core processor for direct host connection (for OEM's) P Remote Booster Amp with 9 wire junction box Code Conduit Connections Electronics Interface Codes 'A', B, D, E, 'K', 'L', 'N' and 'O' (local core processor) B 1/2 inch NPT no gland C 1/2 inch NPT with brass nickel cable gland (cable dia. 8,5 mm to 10 mm) D 1/2 inch NPT with stainless steel cable gland (cable dia. 8,5 mm to 10 mm) E M20-no gland F M20 brass nickel cable gland (cable dia. 8,5 mm to 10,0 mm) G M20 stainless steel cable gland (cable dia. 8,5 mm to 10,0 mm) continued on next page 30 Micro Motion ELITE Meters

31 Ordering information (continued) ELITE sensor model number matrix (continued) Code Conduit connections (continued) Electronics Interface Code 'R', H, 'M' and 'P' (9-wire junction box) A 3/4" NPT no gland H 3/4" NPT with brass nickel gland J 3/4" NPT with stainless steel gland Code Approvals (for all sensor types) M Micro Motion Standard (no approval) N Micro Motion standard/ped compliant U UL Z ATEX Equipment Category 2 (Zone 1)/PED compliant (CENELEC Increased Safety/PED compliant CMF400 only) F CENELEC Flameproof/PED compliant (CMF400 only) Code Language A Danish Quick Reference and English Manual D Dutch Quick Reference and English Manual E English Quick Reference and English Manual F French Quick Reference and French Manual G German Quick Reference and German Manual H Finnish Quick Reference and English Manual I Italian Quick Reference and English Manual N Norwegian Quick Reference and English Manual P Portugese Quick Reference and English Manual S Spanish Quick Reference and Spanish Manual W Swedish Quick Reference and English Manual Code Connection to Booster Amp - CMF400 only A 3/4-inch NPT conduit opening B M20 Brass Nickel Adapter (2) C M20 Stainless Steel Adapter (2) D 1/2-inch NPT Brass Nickel Adapter (2) E 1/2-inch NPT Stainless Steel Adapter (2) Code Measurement Application Software Z None A API (3) Code Future Options Z Reserved for future use - not available on CMF400 Typical Model Number: CMF010M 313 N A B U E Z Z Safe area, no approvals (2) CENELEC approval only (3) Available with electronics interfaces D, E, L, and O. For all others, order with Model 2700 transmitter. Micro Motion ELITE Meters 31

32 Visit us on the Internet at Toll-free numbers (UK only) Tel / Fax Due to Micro Motion's commitment to continuous improvement of our products, all specifications are subject to change without notice. Micro Motion, ELITE, and MVD are registered trademarks of Micro Motion, Inc., Boulder, Colorado. Plantweb is a registered trademark, and Fisher-Rosemount is a trademark of Fisher-Rosemount, Clayton, Missouri. HART is a registered trademark of the HART Communication Foundation, Austin, Texas. Modbus is a registered trademark of Schneider Automation, Inc., North Andover, Massachusetts. Foundation is a trademark of the Fieldbus Foundation, Austin, Texas. Hastelloy is a registered trademark of Haynes International, Inc., Kokomo, Indiana. Inconel is a registered trademark of Inco Alloys International, Inc., Huntington, West Virginia. Emerson Process Management Fisher-Rosemount Flow Groeneveldselaan AZ Veenendaal The Netherlands Tel +31 (0) Fax +31 (0) Toll-free numbers (UK only) Tel Fax Customer Service Tel +31 (0) Fax +31 (0) , Micro Motion, Inc. All rights reserved PS (03/02)

Now available with MVD Technology! ELITE Mass Flow and Density Meters

Now available with MVD Technology! ELITE Mass Flow and Density Meters Product Data Sheet PS-00374 July 2001 Now available with MVD Technology! ELITE Mass Flow and Density Meters Micro Motion ELITE Meters Experience the most accurate Coriolis meters available today. Micro

More information

Micro Motion ELITE Mass Flow and Density Meters

Micro Motion ELITE Mass Flow and Density Meters Product Data Sheet PS 00374, Rev. A July 2003 Micro Motion ELITE Mass Flow and Density Meters Available with MVD Technology! TM Micro Motion Micro Motion ELITE mass flow and density meters Experience the

More information

Micro Motion T-Series Straight-Tube Mass Flow and Density Meters. Product Data Sheet PS July 2001

Micro Motion T-Series Straight-Tube Mass Flow and Density Meters. Product Data Sheet PS July 2001 Micro Motion T-Series Straight-Tube Mass Flow and Density Meters Product Data Sheet PS-00371 July 2001 Micro Motion T-Series Straight-Tube Meters Micro Motion is proud to offer you the first single-straight-tube

More information

Product Data Sheet PS February Model D and DT. Mass Flow and Density Sensors.

Product Data Sheet PS February Model D and DT. Mass Flow and Density Sensors. Product Data Sheet PS 00388 February 2002 Model D and DT Mass Flow and Density Sensors www.micromotion.com Model D and DT sensors Micro Motion Model D sensors set the standard for Coriolis flowmeters nearly

More information

Micro Motion F-Series Coriolis Flow and Density Meters

Micro Motion F-Series Coriolis Flow and Density Meters Product Data Sheet PS-00603, Rev. F December 2007 Micro Motion F-Series Coriolis Flow and Density Meters Micro Motion F-Series Coriolis meters offer highly accurate mass flow, volume flow, and density

More information

Micro Motion Model D and DL Coriolis Flow and Density Meters

Micro Motion Model D and DL Coriolis Flow and Density Meters Product Data Sheet PS-00388, Rev. F April 2013 Micro Motion Model D and DL Coriolis Flow and Density Meters Micro Motion Model D and DL sensors provide flow and density measurement for liquids, gases,

More information

Micro Motion F-Series Coriolis Flow and Density Meters

Micro Motion F-Series Coriolis Flow and Density Meters Product Data Sheet PS-00629, Rev. E May 2007 Micro Motion F-Series Coriolis Flow and Density Meters Micro Motion F-Series Coriolis meters offer highly accurate mass flow, volume flow, and density measurement

More information

Micro Motion ELITE High Capacity Coriolis Flow and Density Meters

Micro Motion ELITE High Capacity Coriolis Flow and Density Meters Product Data Sheet PS-001041, Rev. F February 2010 Micro Motion ELITE High Capacity Coriolis Flow and Density Meters Micro Motion ELITE High Capacity Coriolis meters offer unparalleled flow and density

More information

Micro Motion ELITE High Capacity Coriolis Flow and Density Meters

Micro Motion ELITE High Capacity Coriolis Flow and Density Meters Product Data Sheet PS-001041, Rev. L April 2013 Micro Motion ELITE High Capacity Coriolis Flow and Density Meters Micro Motion ELITE High Capacity Coriolis meters offer unparalleled flow and density measurement

More information

Micro Motion Model CNG050 Compressed Natural Gas Flowmeters

Micro Motion Model CNG050 Compressed Natural Gas Flowmeters Product Data Sheet PS-00408, Rev. D April 2011 Micro Motion Model CNG050 Compressed Natural Gas Flowmeters Micro Motion Model CNG050 meters are specifically designed for the CNG industry to meet the challenges

More information

Micro Motion Model CNG050 Compressed Natural Gas Flowmeters

Micro Motion Model CNG050 Compressed Natural Gas Flowmeters Product Data Sheet PS-00408, Rev. C February 2008 Micro Motion Model CNG050 Compressed Natural Gas Flowmeters Micro Motion Model CNG050 meters are specifically designed for the CNG industry to meet the

More information

Micro Motion ELITE Coriolis Flow and Density Meters

Micro Motion ELITE Coriolis Flow and Density Meters Micro Motion ELITE Coriolis Flow and Density Meters Product Data Sheet PS-00374, Rev V April 2013 Ultimate real world performance Unchallengeable ELITE performance on liquid mass flow, volume flow, and

More information

Micro Motion T-Series Coriolis Flow and Density Meters

Micro Motion T-Series Coriolis Flow and Density Meters Micro Motion T-Series Coriolis Flow and Density Meters Product Data Sheet PS-00371, Rev K April 2018 Superior flow measurement in a single straight tube flow meter Built-in balance bar provides the best

More information

Micro Motion R-Series Coriolis Flow and Density Meters

Micro Motion R-Series Coriolis Flow and Density Meters Micro Motion R-Series Coriolis Flow and Density Meters Product Data Sheet PS-00363, Rev N June 2018 Multi-variable flow and density measurement Basic liquid mass flow, volume flow, and density measurement

More information

Micro Motion Technical Overview and Specification Summary

Micro Motion Technical Overview and Specification Summary Product Data Sheet PS-00232, Rev. J June 2009 Micro Motion Technical Overview and Specification Summary Emerson s world-leading Micro Motion Coriolis flow and density measurement devices have set the standard

More information

Micro Motion. CMF400 Sensor. Installation Instructions. Quick Reference Guide P/N , Rev. D May 2003

Micro Motion. CMF400 Sensor. Installation Instructions. Quick Reference Guide P/N , Rev. D May 2003 Quick Reference Guide P/N 3006075, Rev. D May 2003 Micro Motion CMF400 Sensor Installation Instructions For online technical support, use the EXPERT 2 system at www.expert2.com. To speak to a customer

More information

Micro Motion F-Series Flow and Density Meters

Micro Motion F-Series Flow and Density Meters Product Data Sheet PS-00603, Rev AB January 2019 Micro Motion F-Series Flow and Density Meters High accuracy real world performance Best-in-class performance on liquid mass flow, volume flow, and density

More information

Micro Motion F-Series Flow and Density Meters

Micro Motion F-Series Flow and Density Meters Micro Motion F-Series Flow and Density Meters Product Data Sheet PS-00603, Rev Z December 2017 High accuracy real world performance Best-in-class performance on liquid mass flow, volume flow, and density

More information

Micro Motion ELITE Coriolis Flow and Density Meters

Micro Motion ELITE Coriolis Flow and Density Meters Micro Motion ELITE Coriolis Flow and Density Meters Product Data Sheet PS-00374, Rev AE July 2017 Ultimate real world performance Unchallengeable ELITE performance on liquid mass flow, volume flow, and

More information

Micro Motion ELITE Coriolis Flow and Density Meters

Micro Motion ELITE Coriolis Flow and Density Meters Micro Motion ELITE Coriolis Flow and Density Meters Product Data Sheet PS-00374, Rev AG June 2018 Ultimate real world performance Unchallengeable ELITE performance on liquid mass flow, volume flow, and

More information

Technical Data Sheet PS , Rev C May Micro Motion R-Series Coriolis Flow and Density Meters

Technical Data Sheet PS , Rev C May Micro Motion R-Series Coriolis Flow and Density Meters Technical Data Sheet PS-001945, Rev C May 2017 Micro Motion R-Series Coriolis Flow and Density Meters R-Series Coriolis Flow Meters May 2017 Detailed Specifications R-Series accuracy Model Liquid mass

More information

OPTIMASS 3000 Technical Datasheet

OPTIMASS 3000 Technical Datasheet OPTIMASS 3000 Technical Datasheet Sensor for mass flow First choice for low flow applications Certified secondary pressure containment A common footprint for all three sizes, in Hastelloy or Stainless

More information

Micro Motion ELITE Coriolis Flow and Density Meters

Micro Motion ELITE Coriolis Flow and Density Meters Micro Motion ELITE Coriolis Flow and Density Meters Product Data Sheet PS-00374, Rev AH December 2018 Ultimate real world performance Unchallengeable ELITE performance on liquid mass flow, volume flow,

More information

Micro Motion F-Series Coriolis Flow and Density Sensors. Installation Manual , Rev CC December 2010

Micro Motion F-Series Coriolis Flow and Density Sensors. Installation Manual , Rev CC December 2010 Micro Motion F-Series Coriolis Flow and Density Sensors Installation Manual 20002298, Rev CC December 2010 Safety and approval information This Micro Motion product complies with all applicable European

More information

VersaFlow Coriolis 100 Mass Flow Sensor Specifications

VersaFlow Coriolis 100 Mass Flow Sensor Specifications VersaFlow Coriolis 100 Mass Flow Sensor Specifications 34-VF-03-09 May 2011 The Universal Solution for the Process Industry VERSAFLOW is the only sensor for mass flow in its class with secondary pressure

More information

7000 Series OPTIMASS Mass Flowmeters. Titanium Hastelloy Stainless Steel

7000 Series OPTIMASS Mass Flowmeters. Titanium Hastelloy Stainless Steel KROHNE 04/2004 7.02445.24.00 GR 7000 Series OPTIMASS Mass Flowmeters with single straight measuring tube Titanium Hastelloy Stainless Steel Status: 10/99 Status: 10/99 One tube no limits No limits with

More information

Micro Motion 7827 Digital Viscosity Meter

Micro Motion 7827 Digital Viscosity Meter Product Data Sheet PS-001217, Rev. C April 2013 Micro Motion 7827 Digital Viscosity Meter Micro Motion viscosity meters are built to tackle the most demanding process and control applications. Rugged and

More information

OPTIMASS 2000 Technical Datasheet

OPTIMASS 2000 Technical Datasheet OPTIMASS 2000 Technical Datasheet Sensor for bulk mass flow Large diameter for bulk measurement and custody transfer of liquids and gases Stainless Steel measuring tubes (NACE Compliant) Modular electronics

More information

Armstrong In-Line and Insertion Vortex Flow Meters

Armstrong In-Line and Insertion Vortex Flow Meters Armstrong In-Line and Insertion Vortex Flow Meters Designs, materials, weights and performance ratings are approximate and subject to change without notice. Visit for up-to-date information. 1 In-Line

More information

RHM40. Coriolis Mass Flow Meter for Demanding Applications. Features. Applications. Benefits

RHM40. Coriolis Mass Flow Meter for Demanding Applications. Features. Applications. Benefits RHM40 Coriolis Mass Flow Meter for Demanding Applications Features Standard pressure ratings up to 433 bar (6280 psi) Temperature ratings from -196 to 350 C (-320 to 662 F) Mass flow uncertainty down to

More information

Coriolis Flow Solutions for a Global Marketplace

Coriolis Flow Solutions for a Global Marketplace Coriolis Flow Solutions for a Global Marketplace www.tricorflow.com DECADES of EXPERIENCE = EXCEPTIONAL RESULTS When you combine decades of experience with truly unique talents, great things can happen.

More information

RHM30. Coriolis Mass Flow Meter for Plant and Loading Flows. Features. Applications. Benefits

RHM30. Coriolis Mass Flow Meter for Plant and Loading Flows. Features. Applications. Benefits RHM30 Coriolis Mass Flow Meter for Plant and Loading Flows Features Standard pressure ratings up to 798 bar (11574 psi) Temperature ratings from -196 to 350 C (-320 to 662 F) Mass flow uncertainty down

More information

Transmitting Transducer. Gas/Air Velocity Range:

Transmitting Transducer. Gas/Air Velocity Range: RNG Series Insertion-Style Gas Flow Meter The RNG Series meter is an insertion-style vortex shedding flow meter designed to offer high accuracy measurements over an extremely large flow range. The meters

More information

Installation Manual , Rev DD April Micro Motion ELITE Coriolis Flow and Density Sensors

Installation Manual , Rev DD April Micro Motion ELITE Coriolis Flow and Density Sensors Installation Manual 20002158, Rev DD April 2012 Micro Motion ELITE Coriolis Flow and Density Sensors Safety and approval information This Micro Motion product complies with all applicable European directives

More information

Coriolis Flow Solutions for a Global Marketplace

Coriolis Flow Solutions for a Global Marketplace Coriolis Flow Solutions for a Global Marketplace www.tricorflow.com DECADES of EXPERIENCE = EXCEPTIONAL RESULTS When you combine decades of experience with truly unique talents, great things can happen.

More information

Flow Measurement SITRANS F C

Flow Measurement SITRANS F C Overview Benefits It is compact and light, fitting neatly into dense piping arrangements Effective separation of measurement from plant vibration Reliable measurements due to high signal to noise ratio

More information

SITRANS F flowmeters SITRANS F US. SITRANS FUT1010 Ultrasonic flowmeter. Overview. Application

SITRANS F flowmeters SITRANS F US. SITRANS FUT1010 Ultrasonic flowmeter. Overview. Application Overview SITRANS F flowmeters Application Liquid applications Gas applications Pipelines Custody transfer, allocation, Upstream Production wells, line balance, gathering, separation interface/densitometer

More information

Installation Manual P/N , Rev. DA June Micro Motion ELITE Coriolis Flow and Density Sensors

Installation Manual P/N , Rev. DA June Micro Motion ELITE Coriolis Flow and Density Sensors Installation Manual P/N 20002158, Rev. D June 2009 Micro Motion ELITE oriolis Flow and Density Sensors Planning Mounting Wiring Grounding Safety and pproval Information This Micro Motion product complies

More information

Installation Manual , Rev CG January Micro Motion F-Series Coriolis Flow and Density Sensors

Installation Manual , Rev CG January Micro Motion F-Series Coriolis Flow and Density Sensors 20002298, Rev CG January 2019 Micro Motion F-Series Coriolis Flow and Density Sensors Safety and approval information This Micro Motion product complies with all applicable European directives when properly

More information

Micro Motion F-Series Coriolis Flow and Density Sensors. Installation Manual , Rev CF January 2018

Micro Motion F-Series Coriolis Flow and Density Sensors. Installation Manual , Rev CF January 2018 Micro Motion F-Series Coriolis Flow and Density Sensors Installation Manual 20002298, Rev CF January 2018 Safety and approval information This Micro Motion product complies with all applicable European

More information

Level instruments. Continuous level measurement - Capacitance transmitters SITRANS LC300. 5/262 Siemens FI Overview.

Level instruments. Continuous level measurement - Capacitance transmitters SITRANS LC300. 5/262 Siemens FI Overview. Overview Configuration is an inverse frequency shift capacitance continuous It is ideal for standard industrial applications in chemical, hydrocarbon processing, food and beverage and mining, aggregate

More information

Glass tube flowmeters Series 6000 Variable area flowmeter for liquids and gases

Glass tube flowmeters Series 6000 Variable area flowmeter for liquids and gases Glass tube flowmeters Series 6000 Variable area flowmeter for liquids and gases Easy installation Local indication by means of direct reading Low pressure drop Scales directly in l/h, m 3 /h, kg/h, l/min,

More information

VLM20 In-line Vortex Flowmeter

VLM20 In-line Vortex Flowmeter ocal regulations may restrict the use of this product to below the conditions quoted. In the interests of development and improvement of the product, we reserve the right to change the specification without

More information

LCMass Series Mass Flow Sensor

LCMass Series Mass Flow Sensor LCMass 100-140 Series Mass Flow Sensor Technical Data Sheet HIGH ACCURACY The LCMass 100 achieves a ±0.15% accuracy on liquids. COMPACT ECONOMIC DESIGN Straight twin-tube design allows for an overall smaller

More information

Vortex Flowmeters VFM 3100 F-T VFM 3100 W-T

Vortex Flowmeters VFM 3100 F-T VFM 3100 W-T KROHNE 10/2001 7.02313.23.00 GR/OP Vortex Flowmeters F-T W-T Variable area flowmeters Vortex flowmeters Flow controllers Electromagnetic flowmeters Ultrasonic flowmeters Mass flowmeters Level measuring

More information

Coriolis Mass Flowmeter for low flow measurements

Coriolis Mass Flowmeter for low flow measurements RHM 03 - Flexible Coriolis Mass Flowmeter for low flow measurements The RHM 03 can measure flow rates up to 5 kg/min (11 lb/min) with temperatures in excess of 400 C and pressures to 868 bar. This model

More information

VersaFlow Vortex 100 Vortex Flow Meter Specifications

VersaFlow Vortex 100 Vortex Flow Meter Specifications VersaFlow Vortex 100 Vortex Flow Meter Specifications 34-VF-03-05 May 2011 The All-In-One Solution The VERSAFLOW is the only vortex flowmeter with integrated and temperature compensation in 2- wire technology.

More information

Rosemount 8750WA Magnetic Flowmeter System For Water and Wastewater Industries

Rosemount 8750WA Magnetic Flowmeter System For Water and Wastewater Industries Product Data Sheet January 214 813-1-475, Rev FA Rosemount 875WA Magnetic Flowmeter System For Water and Wastewater Industries THE 875WA MAGNETIC FLOWMETER Rosemount reliability in a customized offering

More information

Pro-V Multivariable Flowmeter Model M24 In-line Vortex

Pro-V Multivariable Flowmeter Model M24 In-line Vortex Pro-V Multivariable Flowmeter Model M24 In-line Vortex Pro-V TM Advantage: VorTek Instruments Pro-V multivariable flowmeters utilize three primary sensing elements a vortex shedding velocity sensor, an

More information

4.30 [109] 5.43 [138] FCI FlexMASSter ST98 Series: The Best Value in a 8.00 [203] High Performance. Flange. Mounting Hole. Positioning to Split Flow

4.30 [109] 5.43 [138] FCI FlexMASSter ST98 Series: The Best Value in a 8.00 [203] High Performance. Flange. Mounting Hole. Positioning to Split Flow ST 98 METER.31 [236] 9.31 [236] /D [2.54] / Compression Fitti Teflon or Metal Ferrul Teflon or Metal Ferrule [2.54] 7.50 [191] [171] Ø.80 Enclosure meets explosion proof, water & dust tight approvals (See

More information

POSITIONERS and Controls for Complete Valve Automation

POSITIONERS and Controls for Complete Valve Automation CONTROLS A Division of AT Controls Pneumatic & Electro-Pneumatic POSITIONERS and Controls for Complete Valve Automation 1 SS SERIES SS Features Easy and quick auto-calibration Detecting RA (reverse acting)

More information

RHM04. Compact Coriolis Flow Sensor With Enhanced Performance. Features. Applications. Benefits

RHM04. Compact Coriolis Flow Sensor With Enhanced Performance. Features. Applications. Benefits RHM04 Compact Coriolis Flow Sensor With Enhanced Performance Features Standard high pressure versions up to 1067 bar (15475 psi) Temperature ratings from -196 to 350 C (-320 to 662 F) Mass flow uncertainty

More information

RHM08. Versatile Coriolis Mass Flow Meter. Features. Applications. Benefits

RHM08. Versatile Coriolis Mass Flow Meter. Features. Applications. Benefits RHM08 Versatile Coriolis Mass Flow Meter Features Standard pressure ratings up to 1254 bar (18188 psi) Temperature ratings from -196 to 350 C (-320 to 662 F) Mass flow uncertainty down to 0.10% Repeatability

More information

Series Level-Trol Electronic Liquid Level Transmitters

Series Level-Trol Electronic Liquid Level Transmitters 0- Level-Trol Series 0- Series Level-Trol Electronic Liquid Level Transmitters W0-/IL Non-Interactive Zero and Span Adjustments; Dry Span Calibration Allows Field Recalibration without Removing Wiring

More information

SATRON VDt Differential Pressure Transmitter

SATRON VDt Differential Pressure Transmitter SATRON VDt differential pressure transmitter belongs to V-transmitter family. The series V transmitters have both analog and smart properties. SATRON VDt is used for -,1kPa...-15 MPa ranges. It is a 2-wire

More information

VORTEX IN-LINE FLOW METERS

VORTEX IN-LINE FLOW METERS VORTEX IN-LINE FLOW METERS Principles of Operation Vortices are created when a fluid passes around a bluff body as shown below. Vortices are alternately shed on each side of the body, degrees out of phase

More information

Mobrey MLT100. Displacer Level Transmitter. Product Data Sheet February 2015 IP119, Rev CA. Level, contents or interface measurement transmitter

Mobrey MLT100. Displacer Level Transmitter. Product Data Sheet February 2015 IP119, Rev CA. Level, contents or interface measurement transmitter Product Data Sheet February 2015 IP119, Rev CA Displacer Level Transmitter Level, contents or interface measurement transmitter Direct or external chamber mounting 4 20 ma HART output ATEX Intrinsically

More information

Vortex In Line Flow Meter

Vortex In Line Flow Meter Principles of Operation Vortices are created when a fluid passes around a bluff body as shown below. Vortices are alternately shed on each side of the body, 0 degrees out of phase to each other, resulting

More information

JUMO dtrans p20. Process pressure transmitter. Type Brief description. Key features. Block diagram

JUMO dtrans p20. Process pressure transmitter. Type Brief description. Key features. Block diagram Data Sheet 403025 Page 1/10 JUMO dtrans p20 Process pressure transmitter Brief description The JUMO dtrans p20 pressure transmitter with HART interface combines maximum precision with easy operation. It

More information

FasTrak High Capacity Smart Positioner

FasTrak High Capacity Smart Positioner FasTrak High Capacity Smart Positioner 2 CCI provides the next level of control valve performance CCI provides the next level of control valve performance with FasTrak, a high capacity digital positioner

More information

RHM015. Ultimate Performance Coriolis Mass Flow Meter for Critical Ultra Low Flow Applications. Features. Applications. Benefits

RHM015. Ultimate Performance Coriolis Mass Flow Meter for Critical Ultra Low Flow Applications. Features. Applications. Benefits RHM015 Ultimate Performance Coriolis Mass Flow Meter for Critical Ultra Low Flow Applications Features Standard pressure ratings up to 752 bar (10907 psi) Temperature ratings from -196 to 350 C (-320 to

More information

ORIMAS Mass Flowmeter

ORIMAS Mass Flowmeter Typical Applications ORIMAS Mass Flowmeter The ORIMAS mass flow meters measure the mass flow of most liquids and gases within the process industries, including chemical, petro-chemical, pharmaceutical

More information

JUMO dtrans p20 DELTA

JUMO dtrans p20 DELTA Data Sheet 403022 Page 1/9 JUMO dtrans p20 DELTA Differential pressure transmitter Brief description The JUMO dtrans p20 DELTA differential pressure transmitter with HART interface combines maximum precision

More information

KROHNE 10/ Optimass 70 & 71 Single Tube Coriolis Mass Flowmeters for the Chemical Industry

KROHNE 10/ Optimass 70 & 71 Single Tube Coriolis Mass Flowmeters for the Chemical Industry KROHNE 10/02 7.02443.21.00 GR Optimass 70 & 71 Single Tube Coriolis Mass Flowmeters for the Chemical Industry Optimass One Tube... Optimass is a brand new family of Coriolis mass flowmeters from KROHNE.

More information

Installation Manual , Rev DI March Micro Motion ELITE Coriolis Flow and Density Sensors

Installation Manual , Rev DI March Micro Motion ELITE Coriolis Flow and Density Sensors Installation Manual 20002158, Rev DI March 2014 Micro Motion ELITE Coriolis Flow and Density Sensors Safety and approval information This Micro Motion product complies with all applicable European directives

More information

Rosemount 285 Annubar Primary Element Series

Rosemount 285 Annubar Primary Element Series 0013-0100-02, Rev CA Rosemount 25 Annubar Primary Element Series Designed for general purpose applications Increased plant uptime with the maintenance-free design Energy savings gained through minimal

More information

Flow Measurement SITRANS F C. Flow sensor MASS 2100 DI 3 to DI 40. 4/180 Siemens FI

Flow Measurement SITRANS F C. Flow sensor MASS 2100 DI 3 to DI 40. 4/180 Siemens FI Siemens AG Flow sensor MASS DI 3 to DI 0 Overview MASS DI 3 to DI 0 is suitable for accurate mass flow measurement of a variety of liquids and gases. The sensor offers superior performance in terms of

More information

SATRON VT Pressure Transmitter

SATRON VT Pressure Transmitter SATRON VT pressure transmitter belongs to the series V-transmitters. SATRON VT is used for 0-1.4 kpa...0-100 MPa ranges. It is a 2-wire transmitter with HART standard communication. In pressure measuring

More information

VersaFlow Vortex 100 Vortex Flow Meter Technical Datasheet. Specification. The all-in-one solution. Highlights. Industries.

VersaFlow Vortex 100 Vortex Flow Meter Technical Datasheet. Specification. The all-in-one solution. Highlights. Industries. VersaFlow Vortex 100 Vortex Flow Meter Technical Datasheet 34-VF-03-05 September 3rd, 2008 Specification The all-in-one solution The VERSAFLOW is the only vortex flowmeter with integrated pressure and

More information

Vortex In Line Flow Meter

Vortex In Line Flow Meter Vortex In Line Flow Meter Principles of Operation Vortices are created when a fluid passes around a bluff body as shown below. Vortices are alternately shed on each side of the body, 0 degrees out of phase

More information

SATRON VVFe Pressure Transmitter

SATRON VVFe Pressure Transmitter SATRON VVFe pressure transmitter belongs to Vtransmitter family. SATRON VVFe is used for 0 4 kpa...0500 kpa ranges. It is a 2wire transmitter with HART standard communication. SATRON VVFe pressure transmitter

More information

JUMO dtrans p20. Process pressure transmitter. Brief description. Special features. Block diagram. Approvals/approval marks (see "Technical data")

JUMO dtrans p20. Process pressure transmitter. Brief description. Special features. Block diagram. Approvals/approval marks (see Technical data) Page 1/15 JUMO dtrans p20 Process pressure transmitter Brief description The device with the HART interface combines maximum precision with simple operation. It is used to measure the system pressure of

More information

In Situ Oxygen Transmitter

In Situ Oxygen Transmitter Product Data Sheet PDS 106-340.A01 January, 2005 Oxymitter Now Offered With Remote Electronics! In Situ Oxygen Transmitter Outstanding accuracy Electronics mounted to probe or separate Adaptable to any

More information

BGN. All Metal Variable Area Flow Meter. Technical Data Sheet

BGN. All Metal Variable Area Flow Meter. Technical Data Sheet All Metal Variable Area Flow Meter BGN Technical Data Sheet Extremely robust through star guiding system instead of guiding rod Flow metering of liquids, gases and steam Available up to nominal size of

More information

Rosemount 1810 Flush Mount Pressure Transmitter ROSEMOUNT 1810 FEATURES:

Rosemount 1810 Flush Mount Pressure Transmitter ROSEMOUNT 1810 FEATURES: Rosemount 1810 Rosemount 1810 Flush Mount Pressure Transmitter ROSEMOUNT 1810 FEATURES: A cleanable and compact stainless steel design An innovative patented venting configuration that prevents gauge reference

More information

PanaFlow MV80. In-line Multivariable Flowmeter for Mass, Temperature, and Pressure. GE Sensing. Applications. Key Benefits

PanaFlow MV80. In-line Multivariable Flowmeter for Mass, Temperature, and Pressure. GE Sensing. Applications. Key Benefits Key Benefits Multivariable vortex flowmeter for measuring volumetric flow, temperature, pressure, density, and mass flow using a single meter Measures most gases, liquids and steam without the need to

More information

XMP ci. Process Pressure Transmitter with HART -communication. Ceramic Sensor. accuracy according to IEC 60770: 0.1 % FSO.

XMP ci. Process Pressure Transmitter with HART -communication. Ceramic Sensor. accuracy according to IEC 60770: 0.1 % FSO. Process Pressure Transmitter with HART -communication Ceramic Sensor accuracy according to IEC 60770: 0.1 % FSO Nominal pressure from 0... 160 mbar up to 0... 20 bar Output signals 2-wire:... 20 ma others

More information

COUNTER CLOCKWISE FOR CLARITY (SEE MANUAL FOR SPECIFIC WATER & DUST TIGHT AP LOCAL ELEMENT ENCLOSURE ROTA FLAT PARALLEL TO FLOW.

COUNTER CLOCKWISE FOR CLARITY (SEE MANUAL FOR SPECIFIC WATER & DUST TIGHT AP LOCAL ELEMENT ENCLOSURE ROTA FLAT PARALLEL TO FLOW. U-LENGTH HORIZ ORIENTA SHOWN ENCLOSURE MEETS WATER & DUST TIGHT AP (SEE MANUAL FOR SPECIFIC A LOCAL ELEMENT ENCLOSURE ROTA COUNTER CLOCKWISE FOR CLARITY 9.85 (250.2) 1" NPT X" 3.00 (76.2) LC OPTIONAL DISPLAY

More information

RHM 03 a1 - (low stress in welds and increased wall thickness against abrasion) RHM bar APPLICATIONS. Batching Dosing. Injections Filling

RHM 03 a1 - (low stress in welds and increased wall thickness against abrasion) RHM bar APPLICATIONS. Batching Dosing. Injections Filling RHM 03 a1 - RHM 03 0.05 ~ 5 kg/min (11 lb/min), -200 ~ 400 C, APPLICATIONS Batching Dosing Injections Filling FEATURES 868 bar 0.005 kg/min ~ 5 kg/min (11 lb/min) 0.1% 0.05% process connection. ADVANTAGES

More information

SITRANS P measuring instruments for pressure

SITRANS P measuring instruments for pressure Overview Application The pressure transmitter is designed for the special requirements of the food, pharmaceutical and biotechnology industries. The use of high-grade materials guarantees compliance with

More information

Installation Manual , Rev DH October Micro Motion ELITE Coriolis Flow and Density Sensors

Installation Manual , Rev DH October Micro Motion ELITE Coriolis Flow and Density Sensors Installation Manual 20002158, Rev DH October 2013 Micro Motion ELITE Coriolis Flow and Density Sensors Safety and approval information This Micro Motion product complies with all applicable European directives

More information

STEEMCO-MAS Flowmeter for Steam

STEEMCO-MAS Flowmeter for Steam STEEMCO-MAS Flowmeter for Steam Typical Applications The STEEMCO-MAS flow meters measure the mass flow of superheated steam within the process industries, including chemical, petro-chemical, pharmaceutical

More information

In Situ Oxygen Transmitter

In Situ Oxygen Transmitter Product Data Sheet PDS 06-340.A0 June, 2004 Oxymitter 4000 In Situ Oxygen Transmitter Now Offered With Remote Electronics! Outstanding accuracy Electronics mounted to probe or separate Adaptable to any

More information

Ar-Mite TM Low Flow Armored Flowmeter

Ar-Mite TM Low Flow Armored Flowmeter Data Sheet Brooks Ar-Mite TM MT 3750C Ar-Mite TM Low Flow Armored Flowmeter Ar-Mite TM Model MT 3750 Metal Tube Flowmeter Ar-Mite TM Model MT 3750 Metal Tube Flowmeter with Transmitter or Inductive Alarm

More information

Mid-West Instrument. Diaphragm Type Model 240. Hazardous Locations. Indicating / Non-Indicating Differential Pressure Switch or Transmitter

Mid-West Instrument. Diaphragm Type Model 240. Hazardous Locations. Indicating / Non-Indicating Differential Pressure Switch or Transmitter BULLETIN NO. 240/11 (SUPERSEDES BULLETIN NO. 240/06) Mid-West Instrument Diaphragm Type Model 240 Hazardous Locations Indicating / Non-Indicating Differential Pressure Switch or Transmitter Field wireable

More information

REVOLUTIONIZE YOUR DENSITY AND CONCENTRATION MEASUREMENT TECHNOLOGY

REVOLUTIONIZE YOUR DENSITY AND CONCENTRATION MEASUREMENT TECHNOLOGY REVOLUTIONIZE YOUR DENSITY AND CONCENTRATION MEASUREMENT TECHNOLOGY Micro Motion Fork Density Meter INTRODUCING EMERSON S MICRO MOTION FORK DENSITY METER Our Fork Density Meter (FDM) delivers accurate

More information

Mobrey MSM400 Ultrasonic sludge density measurement system

Mobrey MSM400 Ultrasonic sludge density measurement system Technical specification sheet IP257 October 2008 Mobrey MSM400 Ultrasonic sludge density measurement system Level Features Measuring range up to 50% solids Simple keypad operation HMI menu based programming

More information

Rosemount Orifice Series Differential Pressure Flow

Rosemount Orifice Series Differential Pressure Flow Rosemount Orifice Series Differential Pressure Flow Innovative Offering for Better Measurement Revolutionizing the Industry Workhorse Orifice Plate Technology It s no coincidence that orifice plates are

More information

Resistance thermometer Model TR10-L, Ex d, for additional thermowell

Resistance thermometer Model TR10-L, Ex d, for additional thermowell Electrical temperature measurement Resistance thermometer Model TR10-L, Ex d, for additional thermowell WIKA data sheet TE 60.12 for further approvals see page 6 Applications Chemical industry Petrochemical

More information

The EPSILON is based on the principle of measuring velocity in the pipe line, therefore the flow measurement is volumetric.

The EPSILON is based on the principle of measuring velocity in the pipe line, therefore the flow measurement is volumetric. Typical Applications EPSILON Flowmeter The EPSILON flow meters measure the flow of most liquids and gases within the process industries, including chemical, petro-chemical, pharmaceutical and the power

More information

Mid-West Instrument. Piston Type Model 220. Hazardous Locations. Indicating / Non-Indicating Differential Pressure Switch or Transmitter

Mid-West Instrument. Piston Type Model 220. Hazardous Locations. Indicating / Non-Indicating Differential Pressure Switch or Transmitter BULLETIN NO. 220/11 (SUPERSEDES BULLETIN NO. 220/06) Mid-West Instrument Piston Type Model 220 Hazardous Locations Indicating / Non-Indicating Differential Pressure Switch or Transmitter Low cost piston

More information

Level Measurement. Point level measurement Capacitance switches. Pointek CLS200 Standard. icenta Controls Ltd

Level Measurement. Point level measurement Capacitance switches. Pointek CLS200 Standard. icenta Controls Ltd Pointek CLS200 Standard Overview Configuration Installation Pointek CLS200 (standard version) is a versatile inverse frequency shift capacitance level switch with optional rod/cable choices liquids, solids,

More information

REVOLUTIONIZE YOUR GAS DENSITY AND PURITY MEASUREMENT TECHNOLOGY

REVOLUTIONIZE YOUR GAS DENSITY AND PURITY MEASUREMENT TECHNOLOGY REVOLUTIONIZE YOUR GAS DENSITY AND PURITY MEASUREMENT TECHNOLOGY Micro Motion Gas Density Meter INTRODUCING EMERSON S MICRO MOTION GAS DENSITY METER Our Gas Density Meter (GDM) delivers fast-response online

More information

Hot Tap Insertion Vortex

Hot Tap Insertion Vortex Hot Tap Insertion Vortex Product Features The same insertion meter can measure Steam, Gases or Liquids. Standardize on an insertion bar that can measure line sizes 2 48 Heavy Duty & Maintenance Free Design

More information

MASSMAX 1000 Series Straight twin tube Coriolis Mass Flowmeter

MASSMAX 1000 Series Straight twin tube Coriolis Mass Flowmeter MASSMAX Series Straight twin tube Coriolis Mass Flowmeter OUTLINE MASSMAX series is the coriolis mass flowmeters developed on the theme of the straight tube model which is excellent in cost performance.

More information

SKINNER Intrinsically Safe Series Four-Way

SKINNER Intrinsically Safe Series Four-Way SKINNER Four-Way Two-Position Valves SPECIFICATIONS Mechanical Characteristics Standard Materials of Body Aluminum Seals FKM, NBR. Other diaphragm materials available upon request. Compatible Fluids Air

More information

Electromagnetic Flowmeters. Product Overview.

Electromagnetic Flowmeters. Product Overview. Electromagnetic Flowmeters Product Overview www.lcmeter.com Premier products, premier performance Liquid Controls offers a broad selection of electromagnetic (Mag) flow meters for industrial and water

More information

SATRON VTe pressure transmitter

SATRON VTe pressure transmitter SATRON VTe pressure transmitter belongs to the series V-transmitters which will have both analog and smart properties. SATRON VTe is used for 0-4 kpa...0-00 MPa ranges. The transmitter communicates in

More information

VIM20 Vortex Insertion Flowmeter

VIM20 Vortex Insertion Flowmeter VIM20 Principle of Operation Vortex flowmeters measure flows of liquid, gas and steam by detecting the frequency at which vortices are alternately shed from a bluff body. According to proven laws of physics,

More information

Rosemount 3051CFA Annubar Flowmeter

Rosemount 3051CFA Annubar Flowmeter December 2013 Rosemount 3051 Rosemount 3051CFA Annubar Flowmeter The Rosemount 3051CFA Annubar Flowmeter utilizes the patented T-shaped sensor design that delivers best in class accuracy and performance.

More information